using anisotropic conductive adhesives in advanced medical … - ito corp.pdf · 2013-03-27 ·...
TRANSCRIPT
Using Anisotropic Conductive Adhesives in Advanced Medical Applications
Peter J. Opdahl, Ito Corporation
www.ito-group.com www.autoacf.com
www.bonding-equipment.com
Vocabulary
ACA: Anisotropic Conductive Adhesive ACF: Anisotropic Conductive Film ACP: Anisotropic Conductive Paste
Lamination: Placing the ACF on a substrate. Mounting: Alignment and placing the two sides
of the assembly together prior to Bonding. Bonding: The final application of heat and
pressure necessary to flow and cure the ACA.
What are Anisotropic Conductive Adhesives? A system used to make both the mechanical and electrical
connection between two devices. Mechanical connections at peel strengths up to 20N/cm. Electrical connections at pitches down to 0.035mm pitch
Minimum 0.010mm space between contacts. Minimum 1500µm2 contact patch (Typically 30µm x 50µm). Typical 5-7mΩ contact resistance for Au-Au bonds. 10-8Ω (10-7Ω after aging) electrical isolation. Good electrical characteristics at >20Ghz frequencies.
Adhesives are typically epoxy for medical applications, but are also available in acrylic/epoxy hybrids and as hotmelt adhesives.
Particles can either be solid metal particles or Ni-Au plated compressible plastic spheres.
Anisotropic Conductive Film (ACF)
Conductive particles (SEM)
ACF( Optical Microscope)
Binder (Thermoset Adhesive)
Conductive Particle
AuNi-Plated Polymer Core
Au-Plated Ni Particle
(3 -10um Diameter)
( 6um Diameter)
(10um Diameter) Pb Free Solder Particles
Low Temperature Low Pressure
Short Time
Substrate
Remove Release Liner
ACF Lamination
Device Alignment and Mounting
Completion Final bonding
Hotbar
Hotbar High Temperature
High Pressure Longer Time
Hotbar
Using ACF
Inside the Bond
Flex Copper Pad
" As the two sides move together, the particles are trapped and crushed between them. This creates the Z-axis connection while maintaining isolation in the X and Y axis.
Conductive Particles Substrate Pad
Substrate
Typical ACF Assemblies COG
(Chip On Glass) COB
(Chip On Board) COF
(Chip On Flex)
ACF
LCD
IC
LCD
ACF
FOB (Flex On Board)
FOG (Flex On Glass)
FOF (Flex on Flex)
Reliability Data
Current Carrying Capacity Gold-Gold bond 0.1mm2 contact area 1A current 85C/85% Rh Environment
Res
ista
nce
(Ω)
High Speed Signaling Extensive testing to greater than
30Ghz shows no breakdown. S21 Transmission Coefficient
S11 Reflection Coefficent
Eye Diagrams
Initial 250hr 500hr
85C/85Rh 100C/-40C
The Next Generation of Medical Detector
Principles of X-ray Detector:
Active Matrix Flat Panel Imagers (AMFPI) imaging system diagram: From Medical Imaging: Principles, Detectors, and Electronics, edited by Krzysztof Iniewski [3]
Next Generation X-Ray Detectors
COF Module: Chip/Die on Flex by ACF for X-Ray Detector Readout COF Module: ASIC/Die Size 12mmx5mm Bump Size: 80 µm x 80 µm Total > 220 I/Os(Analog Input 128 Channels, Digital Output 16 Channels)
Flex Size: 70x15 mm60 μm pitch
Gate Driver, COF Module:ASIC/Die Size 12mmx0.90mm Bump Size: 30 µm x 60 µm Total > 450 I/Os(Digital Output 263 Channels)
Flex Size: 20x20mm40 μm pitch
Scintillator"
Readout"ADC" Addressing"
Active Matrix Flat Panel Imagers (AMFPI)
ACF Bond to readout input pads and X-ray panel.
Output/Input Digital Signal: ZIF Connector or ACF direct bond to board
ACF-COF (Chip-on-Flex)
The Next Generation of Medical Detector
Next Generation PET Imaging
Principles of PET Imaging: Configurations
Next Generation PET Imaging Sensor
Future Generation of PET: Potential Module Configurations
Next Generation PET/SPECT Sensors
ACF-COF Module Chip-on-Flex for PET/SPECT Detector
ACF-COF: Chip/Die/ASICs on Flex
SMT Component on Flex
ZIF Connector: Digital Output Signal
-- or -- Direct Flex-on-Board
with ACF
Silicon pixel array on flex. Current parameters 135C/0.8Mpa/15sec
Future parameters 130C/0.35MPa/15sec (For CZT, others.)
Next Generation Ultrasound Transducers
Principles of Ultrasound System Diagram
Next Generation Ultrasound Transducers
Miniaturization: Ultrasound Probe/Catheter
Migration of functionality in the ultrasound machine: From Medical Imaging: Principles, Detectors, and Electronics, edited by Krzysztof Iniewski [208]
Next Generation Ultrasound Transducers
Miniaturization: Ultrasound Probe/Catheter
Catheter-based side-looking ultrasound probe: From Medical Imaging: Principles, Detectors, and Electronics, edited by Krzysztof Iniewski [208]
Next Generation Ultrasound Transducers
ACF-COF/ACF-COB Module for Ultrasound Probe/Catheter Proposed package has outer dimensions 1.0mm larger than cMUT or
PZT Array size.
Readout"Addressing"
cMUT or PZT array and COF: Chip on Flex-Rigid PCB for Ultrasound Probe/Catheter
cMUT or PZT Array
ACF-COF:Chip-on-Flex -- or -- ACF-COB: Chip on Board
Fold Flex
Final Product
Using Anisotropic Conductive Adhesives in Advanced Medical Applications
Thank you very much
By Peter J. Opdahl, Ito Corporation www.ito-group.com www.autoacf.com
www.bonding-equipment.com