MEGOHMMETER 6505

User Manual

Page 2: User Manual - Test Equipment Depot: Test & Measurement

Statement of Compliance

Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments certifies that this instrument has been calibrated using standards and instruments traceable to international standards.

We guarantee that at the time of shipping your instrument has met its published specifications.

An NIST traceable certificate may be requested at the time of purchase, or obtained by returning the instrument to our repair and calibration facility, for a nominal charge.

The recommended calibration interval for this instrument is 12 months and begins on the date of receipt by the customer.

Serial #: ________________________________

Catalog #: 2130.18

Model #: 6505

Please fill in the appropriate date as indicated:

Date Received: _________________________________

Date Calibration Due: _______________________

Chauvin Arnoux®, Inc. d.b.a AEMC® Instruments

Page 3: User Manual - Test Equipment Depot: Test & Measurement

Table of Contents

INTRODUCTION ..................................................................................... 41.1 Symbols ....................................................................................41.2 DefinitionofMeasurementCategories .....................................51.3 ReceivingYourShipment ..........................................................51.4 OrderingInformation .................................................................6

1.4.1 Accessories ..................................................................61.4.2 ReplacementParts .......................................................6

PRODUCT FEATURES ............................................................................. 72.1 Description ................................................................................72.2 ControlFeatures .......................................................................82.3 SwitchFunctions .......................................................................92.4 ButtonFunctions .......................................................................92.5 Display ....................................................................................10

2.5.1 DigitalDisplay .............................................................102.5.2 BargraphDisplay ........................................................102.5.3 DisplaySymbols .........................................................10

SPECIFICATIONS ................................................................................. 113.1 ReferenceConditions .............................................................113.2 Voltage ....................................................................................113.3 InsulationResistance ..............................................................113.4 Capacitance ............................................................................143.5 PowerSupply ..........................................................................153.6 EnvironmentalSpecifications ..................................................153.7 MechanicalSpecifications ......................................................163.8 SafetySpecifications ..............................................................163.9 VariationsinOperatingRange ................................................16

OPERATION ......................................................................................... 174.1 SerialNumber .........................................................................174.2 SoftwareVersion .....................................................................174.3 VoltageMeasurement .............................................................17

Page 4: User Manual - Test Equipment Depot: Test & Measurement

4.4 InsulationMeasurement ........................................................184.5 PIMeasurement .....................................................................20

4.5.1 AdjustmentsofthePI ..................................................214.6 AdjustmentoftheVariableTestVoltage .................................224.7 AdjustmentoftheMaximumTestVoltage ...............................224.8 ErrorMessages ......................................................................23

MAINTENANCE ................................................................................... 245.1 RechargingtheBattery ...........................................................245.2 FuseReplacement ..................................................................255.3 Cleaning ..................................................................................255.4 Storage ...................................................................................25

Page 5: User Manual - Test Equipment Depot: Test & Measurement

4 Megohmmeter Model 6505

CHAPTER 1

INTRODUCTION

WARNINGThisinstrumentisprotectedagainstaccidentalvoltagesofnotmorethan1000VwithrespecttoearthinmeasurementCategoryIII.Theprotectionprovidedby the instrumentmaybe compromised if it isusedinawaynotspecifiedbythemanufacturer.• Makenomeasurementsonconductorslikelytobeconnectedtoalivesource.

• Complywiththeratedvoltageandmaximumcurrentandthemea-surementcategory.

• Neverexceedtheprotectionlimitsindicatedinthespecifications.• Complywiththeconditionsforuse:temperature,humidity,altitude,degreeofpollutionandplaceofuse.

• Donotuse the instrumentor itsaccessories if theyseemdam-aged.

• Useonly theaccessoriesdeliveredwith theunit,compliantwithsafetystandards(IEC61010-2-031).

• Respectthevalueandtypeofthefuse(see§5.2)toavoiddamag-ingtheinstrumentandvoidingthewarranty.

• SettheswitchtoOFFwhentheinstrumentisnotinuse.• Repairs and metrological verifications must be carried out byapproved,qualifiedpersonnel.

• Weartheappropriateprotectivegear(insulatedboots&gloves).

1.1 Symbols

Signifies that the instrument is protected by double or reinforced insulation.

CAUTION - Risk of Danger! Indicates a WARNING and that the operator must refer to the user manual for instructions before operating the instrument in all cases where this symbol is marked.

Risk of electric shock. The voltage at the parts marked with this symbol may be dangerous.

Page 6: User Manual - Test Equipment Depot: Test & Measurement

Megohmmeter Model 6505 5

Ground/Earth

Important instructions to read and understand completely

Important information to acknowledge

Battery

Fuse

USB socket

The CE marking guarantees conformity with European directives and with regulations covering EMC.

The trashcan with a line through it means that in the European Union, the product must undergo selective disposal for the recycling of electric and electronic material, in compliance with Directive WEEE 2002/96/EC.

1.2 Definition of Measurement Categories

CAT II: Formeasurements performed on circuits directly connected totheelectricaldistributionsystem.Examplesaremeasurementsonhouseholdappliancesorportabletools.

CAT III: For measurements performed in the building installation atthe distribution level such as on hardwired equipment in fixedinstallationandcircuitbreakers.

CAT IV: For measurements performed at the primary electrical supply(<1000V) such as on primary overcurrent protection devices,ripplecontrolunits,ormeters.

1.3 Receiving Your ShipmentUponreceivingyourshipment,makesurethatthecontentsareconsistentwiththeorderinginformation.Notifyyourdistributorofanymissingitems.Iftheequipmentappearstobedamaged,fileaclaimimmediatelywiththecarrierandnotifyyourdistributoratonce,givingadetaileddescriptionofanydamage.Save thedamagedpackingcontainer tosubstantiateyourclaim.

Page 7: User Manual - Test Equipment Depot: Test & Measurement

6 Megohmmeter Model 6505

1.4 Ordering InformationMegohmmeter Model 6505................................................Cat. #2130.18Includes extra large tool bag, set of 3 color-coded 6 ft leads [5000V] (red, blue, black), set of three color-coded alligator clips 600V CAT IV (red, blue, black), one blue guard terminal jumper lead, fuse 0.1A 380V, rechargeable battery pack (installed), US 115V power cord and user manual.

1.4.1 Accessories

Lead-Setof3,color-coded10ft(5kV)Safetywith 3color-codedalligatorclips(600VCATIV) ......................... Cat. #2119.76

Lead-Setof3,Color-coded10ft(5kV)SafetywithIntegralHippoClips(JUMPER LEAD INCLUDED) ........................... Cat. #2119.85

Lead-Setof3,Color-coded25ft(5kV)SafetywithIntegralHippoClips(JUMPER LEAD NOT INCLUDED) ..................... Cat. #2119.86

Lead-Setof3,Color-coded45ft(5kV)SafetywithIntegralHippoClips(JUMPER LEAD NOT INCLUDED) ..................... Cat. #2119.87

1.4.2 Replacement Parts

Setof3,color-coded6ft(5kV)safetyleads .......................... Cat. #2119.77

1ftjumperlead .................................................................... Cat. #2119.78

Fuse–Setof3,0.1A,380V,5x20,.10kA .......................... Cat. #2119.84

ExtraLargeToolBag ...........................................................Cat. #2133.73

Safetyalligator(red)600VCATIV ......................................Cat. #2140.52

Safetyalligator(black)600VCATIV ...................................Cat. #2140.53

Safetyalligator(blue)600VCATIV .....................................Cat. #2140.54

9.6VrechargeableNiMHbatterypack.................................Cat. #2960.21

PowerCord115VUSPlug ..................................................Cat. #5000.14

Order Accessories and Replacement Parts Directly OnlineCheck our Storefront at www.aemc.com/store for availability

Page 8: User Manual - Test Equipment Depot: Test & Measurement

Megohmmeter Model 6505 7

CHAPTER 2

PRODUCT FEATURES

2.1 DescriptionTheMegohmmeterModel6505isaportableinstrumenthousedinaruggedfieldcaseandoperatesoneitherbatteryorlinepower.Itperformsvoltage,insulation,andcapacitancemeasurements.Thisinstrumentcontributestothesafetyofelectricalinstallationsandequipment.FeaturesincludeautomaticcalculationandpresentationoftheDielectricAbsorption Ratio (DAR) and Polarization Index (PI). The Model 6505displays the test voltage, insulation resistance and the leakage currentduringthetest.Capacitanceofthesampleanddischargevoltagepresentatthetestleadsisdisplayedattheconclusionofthetest.TheModel6505isdesignedwiththehighestlevelofbuilt-insafetyfea-tures.Thismeterincorporatestestinhibitcapabilitieswhichwillnotallowtestvoltagestobegeneratedifalivesampleisdetected.Thetesttermi-nalsarerecessedtoensureoperatingsafety.

Features: • Testvoltagecombinationof500V,1000V,2500Vand5000V• Insulationmeasurementsfrom30kΩto10TΩ

• Adjustableandprogrammabletestvoltage(40to5100V)• AutomaticcalculationofDARandPIvalues• DirectmeasurementanddisplayofCapacitanceandLeakage

Current• Displayofresistance,testvoltageandruntime• ProgrammabletestruntimesandPIratiotimes• Automatictestinhibitioniflivesample(>25)• Automaticdischargeanddisplayofdischargevoltage• Largedual-displaywithtime,voltageandmeasurementsshown• Ruggeddualwallweatherprooffieldcase• DesignedandbuilttoIECsafetystandards• EN61010-1,1000VCATIII

Page 9: User Manual - Test Equipment Depot: Test & Measurement

8 Megohmmeter Model 6505

2.2 Control Features

( )

FF 0.1A380V - 10kA

5x20mm

110 - 230V50/60 Hz

20 VA max

CALIBRATION

MEGOHMMETERMODEL 6505

1000V CAT III2500V

GDISPLAY

START/STOP(> 2 s : D A R - P I )

1

6

7

5

32 4

OFF

MΩ -1000V

MΩ -2500V

MΩ -5000V

MΩ 50-5000V

SET V VAR

SET V LOCK

MΩ - 500V

Figure 2-1

Item Description

1 Connection terminals “+”, “G” and “-” and access to the protective fuse

2 AC power plug (direct operation on AC and battery recharge)

3 Back-lit liquid crystal display (see § 2.5)

4 Serial interface male plug (9-pin) for calibration only

5 Rotary selector switch with 8 positions (see § 2.3)

6 START/STOP button

7 Two function buttons (see § 2.4)

Page 10: User Manual - Test Equipment Depot: Test & Measurement

Megohmmeter Model 6505 9

2.3 Switch Functions

Range Function

OFF Instrument powered down

MΩ - 500V Insulation measurement at 500V, up to 2TΩMΩ - 1000V Insulation measurement at 1000V, up to 4TΩMΩ - 2500V Insulation measurement at 2500V, up to 10TΩMΩ - 5000V Insulation measurement at 5000V, up to 10TΩ

MΩ - 50-5000V Insulation measurement with variable test voltage

SET VVARSets the user definable test voltage for the variable 50-5000V position

SET VLOCKSets the user definable maximum test voltage output irrespective of the insulation measurement positions

2.4 Button Functions

Button Function

ON/OFFThis button is pressed to start then stop the measurement.A long press starts the measurement of the DAR and of the PI.

DISPLAYBefore, during or after the measurement, pressing this key displays the various measurement parameters.

This function is available only in the SET positions of the switch. Increases the flashing parameter being displayed. To move about the list of interval insulation measurements, in the R(t) function.

This function is available only in the SET positions of the switch. Decreases the flashing parameter being displayed. To move about the list of interval insulation measurements in the R(t) function.

NOTE: If theandbuttonsarehelddown, themovementbetweenparametersisincreasedtoafasterrate.

Page 11: User Manual - Test Equipment Depot: Test & Measurement

10 Megohmmeter Model 6505

2.5 Display

2.5.1 Digital Display

Main DisplayIndicatesthevalueof:• Insulationmeasurement(resistance,DARandPIorcapacitance).

Small DisplayIndicates:• Voltagemeasuredorappliedbytheinstrument.• Elapsedtimeortheoutputvoltage,duringinsulationmeasurement.

2.5.2 Bargraph Display

Indicates:• Activeduringinsulationmeasurement(0.1MΩto1TΩ).• Indicatesthebatterychargeatstart-up.

2.5.3 Display Symbols

Dangerous voltage generated; V >120V.

External voltage present, symbol is activated after pressing START, if V >25VAC ± 3V or > 35VDC.Indicates the duration of the measurement, or the time remaining in the case of a PI measurement.Indicates the battery is low and must be recharged (see § 5.1).

Page 12: User Manual - Test Equipment Depot: Test & Measurement

Megohmmeter Model 6505 11

CHAPTER 3

SPECIFICATIONS

3.1 Reference ConditionsInfluence Quantity Reference Values

Temperature 23 ± 3°CRelative humidity 45 to 55% RHSupply voltage 9 to 12VFrequency range DC and 15.3 to 65HzCapacity in parallel on resistor 0µFElectrical field nilMagnetic field < 40 A/m

3.2 Voltage

Measurement Range 1.0 to 99.9V 100 to 999V 1000 to 2500V 2501 to 5100V

Frequency Range* DC and 15 to 500Hz DC

Resolution 0.1V 1V 2V 2V

Accuracy1% of Reading

± 5cts1% of Reading ± 3cts

Input Impedance 750kΩ at 3MΩ depending on measure voltage

*Over 500Hz, the small display indicates “- - - -” and the main display gives only an assessmentof the peak value of the measured voltage.

3.3 Insulation ResistanceMethod: Voltage-currentmethodaccordingtoEN61557-2(ed.02/97)Nominal Output Voltage:500,1000,2500,5000VDC(oradjustablefrom40to5100V)

No Adjustment in Variable Mode: 10Vfrom40to1000V100Vfrom1000to5100V

Nominal Current:>1mADC

Page 13: User Manual - Test Equipment Depot: Test & Measurement

12 Megohmmeter Model 6505

Short-circuit Current: <1.6mA±5%

Load Current: 3mADCapproxwhenstartingmeasurement

Max. Acceptable Voltage: Upeak=0.4Un

TestVoltage

500V - 1000V - 2500V - 5000V

Range10 to 999kΩ

1.000 to 3.999MΩ 4.00 to 39.99MΩ 40.0 to 399.9MΩ 0.400 to 3.999GΩ

Resolution 1kΩ 10kΩ 100kΩ 1MΩ

Accuracy ±5% of Reading + 3cts

TestVoltage

500V - 1000V - 2500V - 5000V1000V - 2500V

5000V2500V5000V

Range4.00 to

39.99GΩ40.0 to

399.9GΩ0.400 to 1.999TΩ

2.000 to 3.999TΩ

4.00 to 9.99TΩ

Resolution 10MΩ 100MΩ 1GΩ 10GΩ

Accuracy ±5% of Reading + 3cts ±15% of Reading + 10cts

Accuracy in variable modeRmeasured=Un/250pA

Test voltage 40 to 160V 170 to 510V 520 to 1500V 1600 to 5100V

R measured min 10kΩ 10kΩ 10kΩ 10kΩ

R measured max160.0GΩ to 640.0GΩ

640.0GΩ to 2.040TΩ

2.080TΩ to 6.000TΩ

6.400TΩ to 10.00TΩ

Toobtaintheaccuracyinvariablevoltagemode,calculatefromtheaccuraciesofthefixedvoltagesabove.

Measurement of the test voltage after a capacitive insulation measurement

Measurement Range 25 to 5000V

Resolution 0.2% Un or 1ct

Accuracy 5% of Reading ± 3cts

Page 14: User Manual - Test Equipment Depot: Test & Measurement

Megohmmeter Model 6505 13

Calculation of terms DAR and PI

Specified Range 0.02 - 50.00

Resolution 0.01

Accuracy 5% of Reading ± 1ct

Typical change curve for test voltages according to load:

Range 500V

600

V

500

400

300

200

100

00.01 0.1 1 MΩ

Range 1000V

1200

V

1000

800

600

400

200

00.1 1 MΩ

Page 15: User Manual - Test Equipment Depot: Test & Measurement

14 Megohmmeter Model 6505

Range 2500V

3000

V

2500

2000

1500

1000

500

00.1 1 10 MΩ

Range 5000V

6000

V

5000

4000

3000

2000

1000

00.1 1 10 MΩ

3.4 CapacitanceThismeasurement ismadeattheendofeachinsulationmeasurement,whilethecircuitisbeingdischarged.

Measurement Range 0.001 to 9.999 µF 10.00 to 49.99 µF

Resolution 1 nF 10 nF

Accuracy 10% of Reading ± 1ct 10%

Page 16: User Manual - Test Equipment Depot: Test & Measurement

Megohmmeter Model 6505 15

3.5 Power Supply• 9.6VrechargeableNiMHbatterypack• LineVoltage:85to256V/50-60Hz

Minimum Battery Charge Life(perNFEN61557-2)

Test Voltage Nominal ChargeNumber of Measurements

5s on nominal charge (with 25s pause between each measurement)

500V 500kΩ 6500

1000V 1MΩ 5500

2500V 2.5MΩ 4000

5000V 5MΩ 1500

Average Battery Life:Theoperatingtimewillbe15daysor3weeks,basedupona10minutelongPImeasurement.Recharge Time:Chargingmustbedonebetween20and30°C6hoursfor100%capacity(10hoursifthebatteryiscompletelydrained)0.5hoursfor10%capacity(chargelife:2daysapproximately)

NOTE: It is possible to recharge the batteries while performing insulation measurements provided that the values measured are higher than 20MΩ. In this case, the recharging time is higher than 6 hours and depends on the frequency of the measurements.

3.6 Environmental SpecificationsOperating Range: 14°to104°F(-10°to40°C)duringrechargingofbatteries 14°to131°F(-10°to55°C)duringmeasurement10to80%RHStorage:-40to158°F(-40to70°C);10to90%RHAltitude:<2000mUseindoorsoroutdoors

Page 17: User Manual - Test Equipment Depot: Test & Measurement

16 Megohmmeter Model 6505

3.7 Mechanical SpecificationsCase Dimensions:10.63x9.84x7.09"(270x250x180mm)

Weight:9.5lbs(4.3kg)approx.Mechanical ProtectionIP53perNFEN60529(Ed.92)IK04perNFEN50102(Ed.95)

3.8 Safety SpecificationsElectricalsafetyaccordingtoEN61010-1(Ed.2for2001),EN61557(Ed.2005)

1000V CATIII PollutionDegree2

Electromagnetic Compatibility:NFEN61326-1(Ed.97)+A1,industrialenvironmentcategory

3.9 Variations in Operating Range

InfluentialQuantity

Range ofInfluence

QuantityInfluenced*

Influence

Typical Max.

BatteryVoltage

9V to 12VV

MΩ<1ct<1ct

2cts3cts

Temperature -10° to +55°CV

MΩ0.15% R/10°C0.20% R/10°C

0.3% R/10°C ± 1ct1% R/10°C ± 1ct

Humidity 20 to 80% RHV

MΩ (10kΩ to 40GΩ)MΩ (40GΩ to 10TΩ)

0.2% R0.2% R3% R

1% R ± 2cts1% R ± 5cts

15% R ± 5cts

Frequency 15 to 500Hz V 0.3% R 0.5% R ± 1ct

AC voltagesuperimposedon test voltage

0% to 20% Vn MΩ 0.1% R / % Vn0.5% R / % Vn

± 5cts

*The terms DAR, PI and the capacity and current leak measurements are included in the quantity “MΩ”.

Page 18: User Manual - Test Equipment Depot: Test & Measurement

Megohmmeter Model 6505 17

CHAPTER 4

OPERATION

Charge the instrument fully before use (see § 5.1)

4.1 Serial NumberToview theserialnumberof the instrument, keep theDISPLAYbuttonpressedandturntheswitchtotheMΩ-500Vposition.

4.2 Software VersionToviewtheinternalsoftwareversionoftheinstrument,keeptheDISPLAY buttonpressedandturntheswitchtotheMΩ-1000Vposition.

Technical documentation on Understanding Insulation Resistance Testing is available on our website at www.aemc.com/resources.

4.3 Voltage MeasurementAssoonastheswitchissettoaninsulationmeasurementposition,theinstrumentautomaticallymeasuresthepresenceofanyAC/DCvoltage.Thisvoltageismeasuredatalltimesandindicatedonthesmalldisplayunit.

TheinstrumentautomaticallydeterminesACorDC.TheACmeasurementisanRMSvalue.

Ifanexcessivelyhighexternalvoltageispresentontheterminals(>0.4Un), pressing theSTART/STOP buttonhasnoeffect andnomeasure-mentsaremade.Similarly,ifanexcessivelyhigherroneousvoltage(>0.4Un)isdetectedduringthemeasurement,themeasurement isautomati-callystopped.

Page 19: User Manual - Test Equipment Depot: Test & Measurement

18 Megohmmeter Model 6505

4.4 Insulation Measurement Dependingon themeasurements to bemade, there are threeways toconnecttheinstrument.

Inallcases,disconnectthedevicetobetestedfromthesource.Weak InsulationConnecttheredhigh-voltageleadbetweenearthandthe+terminaloftheinstrument.Connecttheblackhigh-voltageleadbetweenonephaseofthemotorandthe -terminaloftheinstrument.

M+

-

Strong InsulationForaveryhighinsulationvalue,connectthesmallbluehigh-voltageleadbetweentherearearthpick-upjackoftheblackleadandtheG terminaloftheinstrument.Thisservestoreduceanyexternalinfluenceandobtainamorestablemeasurement.

M+

-G

CableConnecttheredhigh-voltageleadbetweenthebraidandthe+terminaloftheinstrument.Connecttheblackhigh-voltageleadbetweenthecoreandthe-terminaloftheinstrument.

G

-

+

ExteriorInsulation

Braid

Insulation Core

Connectthebluehigh-voltageleadbetweentheinsulationandtheG terminaloftheinstrument.Theguardservestoeliminatetheeffectofsurfaceleakagecurrents.

Page 20: User Manual - Test Equipment Depot: Test & Measurement

Megohmmeter Model 6505 19

Oncetheconnectionshavebeenmade,choosethedesiredtestvoltageontherotaryswitch.

Whenpoweredup,theinstrumentdisplaysthefollowing:• Conditionofthebattery• Testvoltage• Voltagepresentontheobjecttobetested

• PresstheSTART/STOPbuttontostartthemeasurement.

• Duringmeasurements, the instrumentwillbeepevery10seconds toalerttheuserthatahighvoltageispresent.

• Press theSTART/STOPbuttonagain tostop themeasurement.Theinstrumentcontinuestomeasureexternalvoltagesbut the test resultremainsdisplayedonthemaindisplayunit.

To ensure your safety, the instrument will automatically discharge the circuit under test, allow for the voltage displayed to fall back below 25V before disconnecting the leads.

Press the DISPLAY button to display:

Before the measurement(2 presses)

• voltage present on the device to be tested• test voltage• surface leakage current

During the measurement(2 presses)

• test voltage• instantaneous insulation resistance value• duration of the measurement• current flowing in the resistance being measured

After the measurement(5 presses)

• voltage present on the device tested• insulation resistance value just before the

measurement was stopped• duration of the measurement• test voltage generated during the measurement• current that flowed in the resistance measured• surface leakage current• capacitance

Page 21: User Manual - Test Equipment Depot: Test & Measurement

20 Megohmmeter Model 6505

4.5 PI Measurement

MΩ -1000V MΩ -1000V

• Settheswitchtooneoftheinsulationmeasurementpositions.

• Startthemeasurementbyalongpress(>2s)ontheSTART/STOPbutton.Thelongpressisacknowledgedbyanaudiblebeep.

• Themeasurementstartsthedefaultdurationof10min.Acountdowndisplaysthetimeremaining.Themeasurementstopsautomatically.

Press the DISPLAY button to display:

Before the measurement(2 presses)

• voltage present on the device to be tested• test voltage• leakage current present

During the measurement(4 presses)

• measurement time remaining• instantaneous insulation resistance value• test voltage• current flowing in the resistance being measured• value of PI (available at the end of 10 minutes)• value of DAR (available at the end of one minute)

After the measurement(6 presses)

• test voltage generated during the measurement• PI and DAR• duration of the measurement• insulation resistance value just before the

measurement was stopped• current that flowed in the resistance measured• voltage present on the device being tested• capacitance• surface leakage current

The values of PI and DAR are calculated as follows:PI=R10min /R 1min(2values toberecordedduringameasurementlasting10min)*DAR=R1min/R 30 sec (2valuestoberecordedduringameasurementlasting1min)

*For the calculation of the PI, the times of 1 and 10 minutes can be modified by the user, ifrequired, for a particular application. See § 4.5.1.

Page 22: User Manual - Test Equipment Depot: Test & Measurement

Megohmmeter Model 6505 21

Theyareespeciallyuseful formonitoring theageingof the insulationofrevolvingmachinesorofverylongcables.

Onitemsofthistype,themeasurementisinitiallyperturbedbyspuriouscurrents (capacitive charging current, dielectric absorption current) thatgradually cancelout.Tomeasure the leakagecurrent representativeoftheinsulationaccurately,itisthereforenecessarytomakemeasurementsoflongduration.

Thequalityoftheinsulationisafunctionoftheresultsfound.

DAR PI Condition of the Insulation

< 1.25< 1

Inadequate or even dangerous< 2

< 1.6 < 4 Good> 1.6 > 4 Excellent

4.5.1 Adjustments of the PI

ItispossibletomodifythePItimestomeetspecificneeds.

Reminder:PI=R10min/R 1min

ThefirstPItimeis1min.Itcanbesettovaluesfrom30secto30minin30secsteps.

• PressandholdtheDISPLAYbuttonandturntherotaryswitchtotheSET VVAR position.HoldtheDISPLAYbuttonuntilPI_1appearsinthedisplay.

• YoucanchangethefirstPItime(PI_1)usingtheandbuttons.

• Tosavechangessimply,turntheswitch.

• ThesecondPItime(PI_2)is10min.ItcanbesettovaluesfromPI_1upto59minin1minsteps.

• PressandholdtheDISPLAYbuttonandturntherotaryswitchtotheSET V LOCK position.

• YoucanmodifythesecondPItimeusingtheandbuttons.

• Tosavechangessimply,turntheswitch.

Page 23: User Manual - Test Equipment Depot: Test & Measurement

22 Megohmmeter Model 6505

4.6 Adjustment of the Variable Test Voltage

SET V VAR SET V VAR

MΩ 50-5000V MΩ 50-5000V

Thisfunctionmakesitpossibletousetestvoltagesotherthanthe4presetvaluesof500,1000,2500and5000.

• SettheswitchtoSET VVAR.

• Thetestvoltageflashes.

• Changethetestvoltageusingtheandbuttons.

• SettheswitchtoMΩ 50-5000Vtomakethemeasurement.

• Thisvalueisretainedinanon-volatilememory.

4.7 Adjustment of the Maximum Test VoltageTheusercansetamaximumgeneratedvoltagetopreventanyaccidentalover-voltagetestsbeingconductedinerror.

SET V LOCK SET V LOCK • SettheswitchtoSET VLOCK.

• Themaximumtestvoltageflashesandcanbeadjustedusingtheandbuttons.

• Turntheswitchtoaninsulationmeasurementsettingtomakemeasurements.

• Themaximumtestvoltagevalueisretainedin anon-volatilememory.Itwillbedisplayedfora fewsecondsonselectionofanaffectedrange (e.g.ifthemaximumvoltageis750V,itwillbe appliedanddisplayedonallswitchsettings from1000Vup).

Page 24: User Manual - Test Equipment Depot: Test & Measurement

Megohmmeter Model 6505 23

4.8 Error Messages

0.1M 1M 10M 100M 1G 10G 100G 1T

V

0.1M 1M 10M 100M 1G 10G 100G 1T

V

VAC

VAC

• Theinsulationresistanceistoolow.• Checkyourconnections.The+and

-terminaloftheinstrumentmaybeshort-circuited.

• Theinsulationresistanceisoutsidethemeasurementrange.

• Checkyourconnections.Oneoftheterminalsoftheinstrumentmaybedisconnectedorthevaluemeasuredis>4TΩ.

• Thevoltagepresentontheterminalsisgreaterthan25VACor35Vpeak.

• Theinstrumentalertsyoubutdoesnotpreventmakingthemeasure-ment.

• Thevoltagepresentontheterminalsistoohighforameasurementtobemade:peakV>0.4UnThetestvoltage,Un,isindicatedbythesettingoftheswitch.

• Eliminatethevoltageandrestartthemeasurement.

• IndicatesthattheprotectivefuseoftheGterminalisdefective.

• Replacethefuseasindicatedin§ 5.2.

Page 25: User Manual - Test Equipment Depot: Test & Measurement

24 Megohmmeter Model 6505

CHAPTER 5

MAINTENANCE

Use only factory specified replacement parts.AEMC® will not be heldresponsible for anyaccident, incident, ormalfunction followinga repairdoneotherthanbyitsservicecenterorbyanapprovedrepaircenter.

5.1 Recharging the BatteryIfthe symboldisplays,thebatteryneedstoberecharged.

Connect the instrument to the 115VAC power cord via the connector(chargingstartsautomatically).

• bAtonthesmalldisplayandchrGonthemaindisplaysignifiesfastcharginginprogress.

• bAtonthesmalldisplayandchrGflashingonthemaindisplaysignifiesslowcharging.

• bAtonthesmalldisplayandFULLonthemaindisplaysignifiesthatchargingiscomplete.

If the instrument is startedupand thebattery voltage is>8V, then thenormaluseofthedeviceispermitted.

NOTE: The battery should only be changed by an authorized repair facility recognized by AEMC® Instruments.

Page 26: User Manual - Test Equipment Depot: Test & Measurement

Megohmmeter Model 6505 25

5.2 Fuse ReplacementIfFUSE -G-flashesonthedisplay,thefusemustbereplaced,takingallthenecessaryprecautionswhenopeninguptheinstrument.

Make sure that none of the terminals are connected and that the selector switch is set to OFF.

• Thefuseislocatedontheleftsideofthefaceplate,indicatedbythe symbol.

• Usingacoinoraflatheadscrewdriver,unscrewthefuseholderandremovethefuse.

• Onlyreplacewiththetypeoffusespecifiedonthelabelinsidetheunit’scover:0.1A-FastActing-380V,5x20mm,10kA

NOTE: IfafterchangingthefusethedisplaystillindicatesFUSE -G-,theinstrumentmustbereturnedtothefactoryforservicing.

5.3 Cleaning

Disconnect the instrument from any source of electricity.

Useasoftcloth lightlydampenedwithsoapywater.Rinsewithadampclothandthendrywithadrycloth.Donotusealcohol,solventsorhydrocarbons.

5.4 StorageIftheinstrumentisnotusedforanextendedperiodoftime,it isrecom-mendedtochargetheinstrumenteverytwoorthreemonths.