Bruker Daltonics Abstract An ultrafast gradient LC separation method was developed to separate a 5-component drug mixture in 30 seconds, with peak widths of 1 second. maXis mass accuracy at sub-ppm levels and true isotopic pattern of the spectra from the peaks leads to a confident elemental formula assignment for each drug compound with the SmartFormula™ algorithm. Introduction Modern discovery applications demand definitive MS results on ever more complex samples. maXis is the only mass spectrometer able to deliver the maximum MS performance specification at the very highest speeds made available by modern ultrafast liquid chromatography. Here, a complete gradient LC separation of typical drug molecules is achieved in 30 seconds, with each component detected with sub-ppm mass accuracy. Technical Note # TN-28 maXis High Resolution LC/MS Makes the Most of Ultrafast LC Separations Experimental A drug mixture (norharmane, sulfamethazine, sulfamethoxazole, oxybutin and terfenadine) was separated with a Waters Acquity UPLC TM system with a 1x50mm Acuity BEH C18 column at using a flow rate of 0.5 mL/min and a gradient from 4% acetonitrile to 95% acetonitrile in 0.5mins. The maXis TM electrospray LC/MS dataset was acquired with an acquisition rate of 10Hz. The dataset was automatically recalibrated in the Compass Data Analysis TM 4.0 processing software with several masses of salt clusters resulting from a loop injection of lithium formate solution (10mM in water) at the end of the separation. OH O O N C 0.3 0.4 0.5 Time [m 12 3 45 0.9 sec 1.0 sec 1.0 sec 0.9 sec 1.0 sec N CH 3 CH 3 High resolution mass spectrometry Unequivocal formula determination fast LC separation

Upload: others

Post on 08-May-2020

7 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Unequivocal formula maXis High Resolution LC/MS Makes the ... · Conclusion The maxis is an ideal instrument for maximum MS performance at the very highest speeds delivered by modern

Bruker Daltonics

Abstract

An ultrafast gradient LC separation method was developed to separate a 5-component drug mixture in 30 seconds, with peak widths of 1 second. maXis mass accuracy at sub-ppm levels and true isotopic pattern of the spectra from the peaks leads to a confident elemental formula assignment for each drug compound with the SmartFormula™ algorithm.

Introduction

Modern discovery applications demand definitive MS results on ever more complex samples. maXis is the only mass spectrometer able to deliver the maximum MS performance specification at the very highest speeds made available by modern ultrafast liquid chromatography. Here, a complete gradient LC separation of typical drug molecules is achieved in 30 seconds, with each component detected with sub-ppm mass accuracy.

Technical Note # TN-28

maXis High Resolution LC/MS Makes the Most of Ultrafast LC Separations

Experimental

A drug mixture (norharmane, sulfamethazine, sulfamethoxazole, oxybutin and terfenadine) was separated with a Waters Acquity UPLCTM system with a 1x50mm Acuity BEH C18 column at using a flow rate of 0.5 mL/min and a gradient from 4% acetonitrile to 95% acetonitrile in 0.5mins. The maXisTM electrospray LC/MS dataset was acquired with an acquisition rate of 10Hz. The dataset was automatically recalibrated in the Compass Data AnalysisTM 4.0 processing software with several masses of salt clusters resulting from a loop injection of lithium formate solution (10mM in water) at the end of the separation.

OH

O O

N CH3

CH3

0.0 0.1 0.2 0.3 0.4 0.5 Time [min]0

1

2

3

4x10Intens.

1 2 3 4 5

0.9 sec

1.0 sec

1.0 sec

0.9 sec

1.0 sec

1 Norharmane 2 Sulfamethazine 3 Sulfamethoxazole 4 Oxybutynin5 Terfenadine

OH

O O

N CH3

CH3

High resolution mass spectrometry

Unequivocal formula determination

fast LC separation

Page 2: Unequivocal formula maXis High Resolution LC/MS Makes the ... · Conclusion The maxis is an ideal instrument for maximum MS performance at the very highest speeds delivered by modern

Results

The separation of the 5 drug mix gave baseline resolution of each compound within 30 seconds (Figure 1). For each peak more than 15 spectral measurements were made across each peak, to give reproducible peak areas suitable for quantitative analysis (Figure 2).

In the mass spectrum, the mass accuracy for each peak of the isotopes is below 1 ppm, with a resolution of 43,700 - acquired at an acquisition rate of 10 Hz (Figure 3). SmartFormula analysis is automated in the CompassTM 1.3 data processing software. SmartFormulaTM automatically determines the possible elemental compositions of each compound LC peak, using a mass accuracy of < 2 ppm error with allowed chemical rules and elements (CnHnNnOnSn). Formulae are ranked according to the closest match factor between the measured isotopic pattern and the theoretical pattern for a given formula – the Sigma value, which should be close to zero. For these drug compounds, a unique SmartFormula result was produced within the allowed mass accuracy and Sigma values (Figure 4). Peak 4, Oxybutynin (Figure 5) was unequivocally confirmed in this experiment.

Chromatographic peak widths (FWHM) of one second or below

0.0 0.1 0.2 0.3 0.4 0.5 Time [min]0

1

2

3

4x10Intens.

1 2 3 4 5

0.9 sec

1.0 sec

1.0 sec

0.9 sec

1.0 sec

1 Norharmane 2 Sulfamethazine 3 Sulfamethoxazole 4 Oxybutynin5 Terfenadine

Fig. 1: Base peak chromatogram of the separation showing resolution of the 5 components in 30 seconds.

Fig. 2: More than 15 spectral datapoints were obtained for each LC peak, this example is peak #4 from figure 1.

Multiple mass spectra from each peak

26.5 27.0 27.5 28.0 28.5 29.0 29.5 Time [s]0.0

0.5

1.0

1.5

2.0

4x10Intens.

SEG71_1-A,4_01_93.d: EIC 358.237670±0.001500 +All MS FullScan, Smoothed (0.22,1,SG)

Page 3: Unequivocal formula maXis High Resolution LC/MS Makes the ... · Conclusion The maxis is an ideal instrument for maximum MS performance at the very highest speeds delivered by modern

Conclusion

The maxis is an ideal instrument for maximum MS performance at the very highest speeds delivered by modern ultrafast chromatography. Sub-ppm mass accuracy is shown for small molecule drugs, together with a resolution in excess of 40,000. Thus the instrument allows extraction of full quantitative information from high speed chromatography. Supported by SmartFormula, the maXis provides unequivocal molecular formula determination capabilities. For compounds with higher molecular weight (up to 1000 Da) SmartFormula 3D™ is using MS/MS fragment data for certainty in small molecule identifications [1].

Fig. 3: Sub-ppm mass accuracy was obtained for each isotope of the m/z peak for compound #4 (oxybutynin) together with >40K resolution.

Mass accuracy and resolution at speed

358.237959

359.241306

360.243507

4. +MS, 0.5-0.5min #(247-261), -Peak Bkgrnd

0.00

0.25

0.50

0.75

1.00

1.25

4x10Intens.

356 357 358 359 360 361 362 m/z

1.0 ppm

-0.9 ppm

-0.8 ppm

SmartFormula provides one single formula candidate

Fig. 4: SmartFormula result for peak #4 with a 2 ppm mass window, only one candidate for-mula meets the mass accuracy and isotopic pattern tolerance.

Fig. 5: Structure formula of Oxybutynin (C22H31NO3) with a M[M+H]=358.23767

OH

O O

N CH3

CH3

Acknowledgement

SmartFormula 3D was developed from an original concept by Don Richards, Pfizer.

References

[1] Bruker Daltonics Technical Note-23. Certainty in Small Molecule Identification by Applying SmartFormula 3D on a UHR-TOF Mass Spectrometer.

Authors

Catherine Stacey, Sebastian Götz, Carsten Bäßmann, Bruker Daltonics.

Keywords

ultra performance liquid chromatography

UHR-TOF

ab initio formula determination

Instrumentation & Software

maXis

SmartFormula

Compass

Waters, ACQUITY UPLC, and UPLC are trademarks of Waters Corporation.

maXis, Compass Data Analysis, Compass, SmartFormula and SmartFormula 3D are trademarks of Bruker Daltonics.

For research use only. Not for use in diagnostic procedures.

Page 4: Unequivocal formula maXis High Resolution LC/MS Makes the ... · Conclusion The maxis is an ideal instrument for maximum MS performance at the very highest speeds delivered by modern

Bruker Daltonik GmbH

Bremen · GermanyPhone +49 (421) 2205-0 Fax +49 (421) 2205-103 [email protected]

Bruker Daltonics Inc.

Billerica, MA · USAPhone +1 (978) 663-3660 Fax +1 (978) 667-5993 [email protected]

www.bdal.com

Bru

ker

Dal

toni

cs is

con

tinua

lly im

prov

ing

its p

rodu

cts

and

rese

rves

the

rig

ht

to c

hang

e sp

ecifi

catio

ns w

ithou

t no

tice.

© B

ruke

r D

alto

nics

05

-20

08,

TN

-28