Journal of Research of the National Bureau of Standar ds-A. Physics and Chemistry Vol. 63 A, No. 2, September-October 1959 Tritium-Labeled Compounds I. Radioassay of Tritium- Labeled Compounds in "Infinitely Thick" Films With"a Windowless, Gas-Flow, Proportional Counter I Horace S. Isbell , Harriet 1. Frush, and Ruth A. Peterson (May 5 , 1959) J:. simple, sensitive, reliable techniqu e has been devi se d for t he ra dioassay of non- vol at ile, water-soluble t ntmm compounds. The s ub stan ce t o be analyzed is dissolved in an aqueous solution of a t hi ckening agent, preferably s odium O-( ca rboxymethyl) ce llulose or so dium algin ate. The solution is pl aced in a shallow pl a n ch et, a nd after eva porat ion of the water, resulting film, whi ch is t hi ck " t o tri t ium beta particles, is counte d with a 271", wmdowless, gas-flow, proportlOnal counter. By means of an e mpirica l factor d eter- mi ned with a. s ubstance of known radioact ivity, the count s are conv erte d to In a film a t hi ckn ess of 0.7 mg/c:n 2 , co unt ing e ffi ciency is a bout 4 per cent; the s ta nda rd devl at lOn fr om the mean, ob tamed m a sen es of rou tine measurements, was less than 2. An say. can be made .triti uIl!--con ta ini ng m aterial having 0.01 m ICroc une of radlOact lv lty. 'Ihe m et hod, whICh IS applIca ble to n on vo lat ile wate r- so tuble solids, so lut ions, or liquids, is sui ta ble f or routine analyses. ' 1. Di scussion Tritium is on e of th e eheap e t and mo st v er sa tile radioi sotop es for use as a tracer in chemi cal reactions. Howev er , its wid es pread use has been hamp ered by a la ck of convenient m et hods of analysis. In con- nection with the deve lopment of m et hod for the pr eparation of position-labeled, radio ac tive carbo- hydr ates containing tritium, a simpl e, ra pid tech- mqu e w3: s needed for the assay of radioactivi ty of trit ium-labeled mat erial s. Liquid scin- eo. un ters [ 1], 2 althou gh very satisfact ory for dI ssolved sampl e, arc both expensive a nd unavailable to many workers. Th e conversion of mat erials to gases, for counti ng in ioni zaLion cham- bers [1] is laborious for rou tine an aly es . Several work ers have coun ted t ritium-l abeled ma teri als from "infinitely thick" films with propor tional coun ters [2 to. 7] . A tritium-containing solid, if available in relatIvely large quantity, may be packed in to a c!lPped planchet and count ed [2 , 3, 4] ; this low-effi- CIen cy method has chiefly been used for comparisons rath er than for the de termination of ab solute iti es. Altern ativ el y, an infinitely thick film of a nonvolatile liquid been employed [5], or a solid film has b ee n depOS Ited on a planchet by ev apor at ion of th e solvent from a solution [6 7] . The l ast- men. tioned met hod is pr obabl y th e' most ver satile, but It suffers from th e fa ct that crystallization on th e planch et can r en der the surface ar ea of the film un - and vari able [5, 7], a nd may even cause con- t amll1a t lOn of th e co un ter during flushing [6 ]. For routin e radioassay of nonvola tile wat e r- soluble,. t ri t ia ted m aterials in. a windowless, gas-fl ow, coun ter, a techmqu e ha s been developed m thIS lab oratory t ha t avoids difficulty cau se d by I 'part of a projcct on the development of methods for the product io n of radio· actIve carbohydrates, sponsored by tho Div ision of Research of the Atomic Energy Commission. 2 Figures in brackets iudicate the literature refe rences at the end of this paper. cr yst alliz ation and permi ts the coun ting, from uni- form, infinitely thick fi lms, of ma terial s availabl e in small quan tity. Th e radioa ctive sub stance is dis- solve d in an aq.ueous solution of a thi ckening agent., prefer abl y SOdIum O- (c arboxym et hyl)cellulose or odium algin ate. 3 Th e amount of sampl e and t hi cken- ing solution arc so a djusted t hat the t hi ckn ess of the film formed on the plan chet aft er ev apora tion of the water is more than th e m aximum ran ge of the tr itium beta particles, (inflllitely thick, a ppro ximately 0.7 mg/cm 2 fo r the m aterial used [9 ]). Th e films adhere well to. pl anchets, a nd the thi ck en er prevents crystallIzatlOn. The C OUlI ts arc converted to mi cro- curies (MC) by m eans of an e mpiri cal fact or determ in ed und er th e sam e conditions with a s ampl e of known radio act ivity.4 Alt hough the coun ting eff icien cy of a film having a thickn ess of 0.7 mg/cm 2 is only abou t 4 percen t, the repro du cibili ty of th e m ethod cor- res ponds to a s tand ard devia tion of less th an 2 per cent. A satisfa ctory determin ation can be mad e with as little as 0.01 MC of a t ritium-labeled m aterial on a 2-in. pl anchet. Because of i ts simplici ty the method is suitabl e for rou t ine analyses. Th e ac- cur acy of the m e thod depend s on th e absolute act iv ity of the ma te rial used as stand a rd . For the work repor te d h er e, D-glu cose -l -t was standard ized by co mp a ring it with the NBS tritiunl oxid e stand a rd . In the cour se of t he work , a st udy was made with films of less than infinite thi ckn ess. Although hi ah counting effi cien cies can be ob tain ed, un ce rtain ty ari in g from uneven distrib ut ion of the sample makes coun ting from thin films less satisfa ctory t han coun t- ing from infinitely t hi ck films by the me thod described. 3 A critical st u dy of the energy loss and penetrat ion range of beta particles is given in [81. In general, the range increases with tbe atomic n umher of the mate- rial penetrated. In th e film-co unting met bo d described bere, tbe fil ms are oC organic materials, and have approximately tbe same elemental composition. In the presence of elements of higber atomic num ber, the greater depth of pene- tration would require a tbicker film for infinite th ickness, and usc of a different factor fo r converting counts to microcur ies. • Tbe co unting of carbon·14 from thin fi lms of agar bas been report ed by Mc- Cready [10]. See also a modifi cation of the method [11]. 171

Upload: others

Post on 24-Jun-2020

5 views

Category:

Documents


1 download

TRANSCRIPT

Page 1: Tritium-labeled compounds I. Radioassay of tritium-labeled ... · The activity of tritium in the sample is given by the relationship: dps jJ.c = 37000, = cpsXmXk (1) where cps is

Journal of Research of the National Bureau of Standards-A. Physics and Chemistry Vol. 63A, No. 2, September-October 1959

Tritium-Labeled Compounds I. Radioassay of Tritium­Labeled Compounds in "Infinitely Thick" Films With"a

Windowless, Gas-Flow, Proportional Counter I

Horace S. Isbell, Harriet 1. Frush, and Ruth A. Peterson

(May 5 , 1959)

J:. s imple, sensit ive, ~nd reliable technique has been devised for t he radioassay of non­volatile, water-soluble t ntmm compounds. The substan ce t o be analyzed is dissolved in a n aqueous solut ion of a t hickening agent , preferably sodium O-(carboxymethyl) cellulose or sodium alginate. The solution is placed in a shallow plan chet, a nd af ter evaporation of t he water , ~he result ing film, which is "~nfinitely t hick" t o trit ium beta par ticles, is counted with a 271", wmdowless, gas-flow, proportlOnal coun ter. By means of an empirical factor deter­mined with a. subst ance of known rad ioactivity, t he counts are converted to micr~curies . I n a film hav~n& a t hi ckness of 0.7 mg/c:n2, t~e co unting e ffi ciency is about 4 percen t ; t he standard devlatlOn from t he mean, ob tamed m a sen es of rout ine m eas urements, was less t h a n 2. peree~ t. An ~ say. can re~dj]y be made ~th .triti uIl!--containing material having 0.01 mICrocune of radlOactlvlty. 'Ihe method, whICh IS applIcable to non volatile water-sotuble solids, solutions, or liquids, is sui table for rout ine a nalyses . '

1. Discussion

Tritium is one of the eheape t and most versa tile radioisotopes for use as a tracer in chemical reactions. However , its widespread use has been hampered by a lack of convenien t methods of analysis. In con­nection with th e development of method for the preparation of position-labeled, radioactive carbo­hydrates containing tritium, a simple, rapid tech­mque w3:s needed for the assay of radioactivity of l~onv.olatile, tri tium-labeled materials. Liquid scin­tillatl~)l1 eo.un ters [1],2 although very sa tisfactory for assayl~g dIssolved sample, arc both expensive and unavailable to many workers. The conversion of materials to gases, for counting in ionizaLion cham­bers [1] is laborious for rou tine analy es. Several workers have counted tritium-labeled materials from " infinitely thick " films wi th propor tional counters [2 to. 7] . A tritium-containing solid, if available in relatIvely large quantity, may be packed in to a c!lPped planchet and counted [2 , 3, 4] ; this low-effi­CIency method h as chiefly been used for comparisons rather than for the determination of absolute activ~ ities. Al terna tively, an infinitely thi ck film of a nonvolatile liquid ~as been employed [5], or a solid film has been depOSIted on a planchet by evaporation of the solven t from a solution [6 7] . The last­m en.tioned m ethod is probably th e' m ost versatile, but It suffers from the fact that crys tallization on the planchet can render the surface area of the film un­cer t~in and variable [5, 7], and may even cause con­t amll1atlOn of th e coun ter during flushing [6].

For routine radio assay of nonvolat ile water­soluble,. tri tia ted materials in. a windowless, gas-flow, prop~rtlOnal counter , a t echmque has been developed m thIS laboratory that avoids difficulty caused by

I 'part of a projcct on the development of methods for the production of rad io· actIve carbohydrates, sponsored by tho Division of Research of the Atomic Energy Commission.

2 Figures in brackets iudicate the literatu re references at the end of this paper.

crystallization and permits the counting, from uni­form, infini tely thick films, of materials available in small quan ti ty. The radioactive substance is dis­solved in an aq.ueous solution of a thickenin g agent., prefer ably SOdIum O- (carboxymethyl)cellulose or odium alginate.3 The amount of sample and thicken­

ing solution a rc so adjust ed that the thickness of the film formed on the planchet after evaporation of the water is more than the m aximum range of the tritium beta particles, (inflllitely thick, approximately 0.7 mg/cm2 for the m aterial used [9]). The films adhere well to. th~ planch ets, and the thickener prevents crystallIzatlOn. The COUlI ts arc conver ted to micro­curies (MC) by m eans of an empirical factor de termined under the sam e conditions with a sample of known radioactivity.4 Although the coun ting efficien cy of a film having a thickness of 0.7 mg/cm2 is only abou t 4 percen t, the reproducibili ty of the method cor­responds to a s tandard devia tion of less than 2 percen t . A sa tisfactory determina tion can be made with as li ttle as 0.01 MC of a t ritium-labeled material on a 2-in. plan chet. Because of i ts simplici ty the m ethod is suitable for routine analyses. The ac­curacy of the m ethod depends on the absolute activity of the material used as standard. For t he work reported here, D-glucose-l -t was standardized by comparing i t with the N BS tri tiunl oxide standard.

In the course of t he work, a study was made with films of less t han infinite thickness. Although hiah counting efficien cies can be ob tained, uncer tainty ari ing from uneven distribution of the sample m akes counting from thin films less satisfactory than count­ing from infini tely thick films by the method described .

3 A critical study of the energy loss and penetration range of beta particles is given in [81. In general, the range increases with tbe atomic numher of the mate­rial penetrated. In the film-counting metbod described bere, tbe fil ms are oC organic materials, and have approximately t be same elemental composition. In the presence of elements of higber atomic n um ber, the greater dept h of pene­t ration would require a tbicker film for infinite th ickness, and usc of a different factor for converting counts to microcuries.

• Tbe count ing of carbon·14 from thin fi lms of agar bas been reported by Mc­Cready [10]. See also a modification of the method [11].

171

Page 2: Tritium-labeled compounds I. Radioassay of tritium-labeled ... · The activity of tritium in the sample is given by the relationship: dps jJ.c = 37000, = cpsXmXk (1) where cps is

2. Materials and Apparatus

2.1. Tritium-Labeled Materials

D-Glucose-l-t was prepared by sodium amalgam reduction of D-glucono-o-Iactone in tritiated water [12] in the presence of sodium acid oxalate [13]. The sample used as a tritium standard was assayed by comparing its activity in a phosphoric acid­phosphoric anhydride solution with that of the NBS standard tritium oxide sample No. 4926; the measure­ments were made with a proportional counter [14]. D-Mannitol-l-t was prepared by reducing D-mannose with tritiated lithium borohydride [14].

2.2. Solution of Sodium O-(Carboxymethyl)cellulose

The thickening agent was commercial , medium­viscosity sodium O-(carboxymethyl) cellulose, OMC-12MP, a product of Hercules Powder Company, Wilmington, Del. The CMO stock solution ordi­narily used for the radio assay of solid samples was prepared by dissolving 1 g of CMO, 0.5 g of anhy­drous D-glucose, and 10 mg of eosin 5 in sufficient water to give 100 ml of solu tion. The D-glucose serves as an organic "ballast" material and plasti­cizer. It was omitted entirely, or in part, in thicken­ing solu tions for use with large samples of weakly radioactive material. D-Glucitol has been used with equal success as ballast in place of D-glucose.

2.3. Solution of Sodium Alginate

The sodium alginate used was high-viscosity Algin, obtained from Kelco Company, San Diego, Calif. '1'he stock solu tion was prepared from 0.5 g of Algin, 1 g of anhydrous D-glucose, 10 mg of eosin, and sufficient water to give 100 ml of solution.

2.4. Counting Equipment

Radioactivity measurements were made with a 271' , windowless, gas-flow, proportional counter, Model PO- 3, obtained from Nuclear Measurements Corp., Indianapolis, Ind.; the slide holder of the instrwnent had been modified by inserting tubes for water circulation [15] . Commercial gas, consisting of 90 percent of argon and 10 percent of methane, was used in the coun ter. The gas was dried by suc­cessive passage through soda-lime and anhydrous calcimn sulfate. The 2-in. , stainless-steel, cupped, flat-bottomed, counting planchets used in this work were also obtained from Nuclear Measurements Corp.; they have an effective area of 21.4 cm2. Some of these planchets were modified by enclosing a 10-cm2 area with a circular groove beyond which the solution did not spread.

, The dye serves to show the uniform ity of th e film .

3. Recommended Procedure for Radioassay of Tritium-Labeled Compounds in Infinitely Thick Films

The solution of the radioactive sample in the aqueous thickening agent can be prepared by any convenient method. Ordinarily, l'adioacti ve solids are weighed, and then dissolved in a lmown volume of the stock thickening solution. A radioactive liquid, usually less than 0.5 ml, is mixed with a measured amount of the thickening solution. How­ever, if the liquid is sufficiently radioactive, it is measured in a dilution pipet and diluted to a definite volume with the thickening solu tion . Approxi­mately 1 ml of the solution to be assayed is trans­ferred to a clean,6 2-in. cupped planchet (or 500 microliters is transferred to the 10-cm2 area of the planchet) , which is then placed under an infrared lamp. After the solvent has evaporated, the planchet is kept for about 1 hI' in a desiccator over a saturated solu tion of potassium acetate in contact with the solid phase. This rather unusual procedure gives films of reproducible moisture content, and does not lower the count substantially. The sample is then counted in the proportional counter for 3 periods, with a total of at least 10,000 co unts. Before each period the counter is flushed with the counting gas for 30 sec. It occasionally happens that successive counts show a marked trend, presumably caused by the accumulation of a static charge on the film. If this occurs, the sample is removed to the desiccator for a short time, and then recounted. The back­ground count is determined with a film-covered planchet placed in the counting chamber.

The voltage for optimum precision must be deter­mined under the conditions used. The optimum voltage for assay of solid samples with the equip­ment used in the present study was approximately 2,050 volts .

4. Discussion of the Procedure and Results

The activity of tritium in the sample is given by the relationship:

dps jJ.c = 37000 = cps X m X k , (1)

where cps is the net count, m is the total solids in a volume of solution containing the radioactive sample, and k is a constant determined by the same method for a compound of known tritium content.

The value of m must be determined accurately, although the exact amount of solution placed on the planchet is not critical if the film formed from it is infinitely thicle Because of hydration , the value of m calculated from the ingredients in the formula­tion may differ slightly from that determined by weighing the film produced, under the conditions of

6 Care must he used to obtain a grease-free surface on tbe planchet . In tbis laboratory the planchets are cleaned in a solution of sodium hydroxide or "Versene" immediately before use.

172

Page 3: Tritium-labeled compounds I. Radioassay of tritium-labeled ... · The activity of tritium in the sample is given by the relationship: dps jJ.c = 37000, = cpsXmXk (1) where cps is

~

I

I

Lhe method, from a definite volume of the solution. Determination of the weight of the film in each analysis is laborious. This step can be avo ided by determining the weight of th e film formed on a planchet from exactly 1 ml of the stocle solution of CMC or Algin under the conditions of the analysi . Because of the viscosity of the solut ion , a pipet, calibrated to contain 1 ml, is employed , and the solution adhering to the pipet is carefully rinsed ou t and added to the planchet. The film, obtained after evaporation of the water, is conditioned in the usual manner and weighed. Several determinations arc made, and the average value is used for calculating m, the weight of the film containing the radioactive su bstance, on the assumption that m equals the weight of the film from the amount of stock solution used plus the weight of the radioactive sample. All nonvolatile matter in the formulation must be included in the calculation of m.

The value of Ie depends on the surface area, back­scattering, efficiency of the countor , geometry, and other factors; it has the dimension of weight-I. To obtain accuraLe results in the analy is of a seri es of compounds, lc mus t be detennined by assa.v of a s tandard tritium compound under the conditions used for the unknowns. ,\Vith ma terials of similar character, the amount of radiation scattering is cons tant. Because sca ttering varies with the 01 -

frctive atomic number of the reflectin g matrrial, heavy clements such as barium increasr in trmal scat tering and may lrad to small errors. These can be avoided by adding, to th e sLandard , com­pounds of any h eavy clements that may be present in the maLerial to be assayed. Ions of certain metals, as, for example, lead , form in soluble com­pounds with both CMC and Algin . This undesirable effect can be overcome by addin g Versene to Lhe film.-forming mixture.

T able 1 presents r esults obtained with two samples each of D-glucosr-l-t and D-mannitol-l-t. Th e stand­ard deviations a re bBsed on the deviations in countin g

rates for the separate planchets fronl. the mean counting rate for the group of planchets. Sample G I was used as a standard, and its specific activity (0.0865 J,Lc/mg) was carefully determined by the comparative method described under materials and apparatus. Usc of this material as a control gave values o[ 7c ullder various conditions, i.e., for CMC and Algin films and for the 10-cm2 and 21.4-cm2 (2-i11. ) pl anche ls. By use oC these constants, the specifi c act ivities of G2, NIl, and }'12 were determined. The counting rates for film havin g thicknesses between 0.69 and 1.48 mg/cm2 showed no significant differences.

5. Assay of Tritium Compounds in Films of Less Than Infinite Thickness

Table 2 and figUl' e 1 show the relative counting rate of a preparation of D-glucose-t as a function of source thickness. Each sample was spread on a fre hly cleaned, 2-i11., stainless-steel planchet, and the solvent was evaporated under an infrared lamp. The planchet was cooled in a desi.ceator over anhy­drous calcium chloride, and then counted with the equipment deseri.bed before. The first seven deter­minations were made on films from solutions of D-glucose-t i.n water, and the last three on films from solutions of D-glucose-t conta ining sufficient Algin to give the weight recorded. The radioactivity in thin fi lms can be es timat ed by use of the data of table 2 and figure 1.

6 . Discussion of Absolute Radioactivity

Th e counting rate of a tritium source in an effi cient, windowless, gas-flow, proportional counter can be represented by an equat ion of the type [6]:

TABLE 1. Counting "ales of tritium cornp01mds in "infinitely thick" film s

Sample Film Counting D ata

Substance ' Wei~ht b 1'hicken- Area T otal solids 'l ' hickncss Ko. of Average Standard (m') ing agent (m) planchets count devi ation 0

111.{/ cm2 my 7na/cm2 G 1 _____ _ _ ._. ______ 0.435 CMC 10.0 7.7 0.77 G 1 ________________ .87 Cl\lC 21. 4 15.4 .n G 1 ______________ _ _ .415 Algi'" 10. 0 8. I .81 G 1 ________________ .83 Algin 21. 4 16.2 .76

G, _____________ ___ .82 C l\f C 21. 4 14.8 .69 G2 __________ ______ 1. 23 CMC 21. 4 22.1 l.03 G, ________________ 0. 82 CMC 10.0 14.8 1. 48

G, ________________ .82 Algin 21. 4 15.7 0. 73 G, _______________ _ 1. 23 Algin 21. 4 23.5 1. JO M I __ _____________ 0.036 CiVrC 21. 4 14.0 0.65 M , __ _____________ 2.50 Ci\JC 2 1. 4 16.7 .78

a 0 1 an d 02 are different preparat ions of D-glucosc-l-t ; lVI l and nI 2 are D-mannitol-l-t. b m' is the weight of the radioactive sample contained in the film of weight m. , I '£d2 X 100

'V n-l d Jlc=cp!:-iXmXk. Values in p arentheses were used as standards.

173

cps % 5 50. I 1. 04 5 J08.7 0.86 4 45. 1 .58 5 95.3 .87

5 101.3 .44 5 JOO.5 .63

II 47. 6 1.15

5 95.0 1. 57 3 94.5 1.71

4 289.4 0.14 5 34.1 1. 40

Spec ific activity d k X lO' of sample

p.c/rn' (0. 0856) 9.65 (.0856) 4. 45 (. 0856) 9. n (.0856) 4.60

.08 14 (4. 45)

.080-1 (4. 45)

. 0829 (9. 65)

.0837 (4.60)

.0831 (4.60)

5. 01 (4. 45) 0.010 (4.45)

Page 4: Tritium-labeled compounds I. Radioassay of tritium-labeled ... · The activity of tritium in the sample is given by the relationship: dps jJ.c = 37000, = cpsXmXk (1) where cps is

(2)

where dps is the activity of a source, and ig, fb, and i. are correction factors for the geometry of the detector, for backscattering, and for self-absorption, respectively. The geometry factor , ig, with a flat planchet is approximately 0.5. The backscattering factor, ib, is a function of the atomic weight of the backing material and depends somewhat on the ch aracter of the radiation and on the geometry of the counter r16]. But with an infinitely thick source, scattering from the backing material may

TABLE 2. Counting rate of D-glucose-t as a function of film thickness

Volume Counting n-Glu· of film- Total Film Average Standard efficiency cose-t a forming solids thickness count b deviation cpsjdps

solu tion

mg ml mg mg/em' eps % % 0.02 0.02 0. 02 0. 000934 30.9 2.3 48. 8 .05 .05 . 05 .00234 74.3 2.9 46.9 .10 . 10 .10 .00467 142.2 3.4 44.9 .20 .20 .20 .00034 232.7 5.7 36.7 . 50 .50 .50 .0234 499 4.4 31. 5

1. 00 1.00 1.00 .0467 750 4.0 23.7 2.00 2.00 2.00 _ 0934 1103 3.4 17. 4

1. 00 1. 00 • 9.00 0. 420 d 1971 2.1 6.9 1.00 1. 00 17.00 .794 2108 2.0 3.9 1. 00 1.00 25.00 1.170 2108 0.6 2.7

• The activity of the n-glucose-t wa~ 0.0856 !,cjm g_ The" effective area" of the planchet was 2l.4 cm'.

b Average of three planchets, each coun ted to 10,000 counts. • In tbe last t hree experiments, the amount of n-glucose-t wa~ constan t, and

the weight of film was increased hy use of an Algin solution_ d The observed cps and the total dps In the I >st th ree experiments were mnlti­

plied by the dilu tion ratiO, i.e., by 9, 17, and 25, respectively, to reduce all results to a common specific activity.

55.---------,----------.---------,----------,

50 •

~5 - •

~ 0

• 35 ..

>-~ 30 -w Q "-~ 25 -

'" ;;: ... z 20 :::> 0 u

15

10

0.0001 0.001 0.0 1 0.1 1.0 FIL M T HICK NESS , mg/cm-2

FIGURE L Counting efficiency as a function of film thickness.

be neglected because the only reflected radiation t.hat r eaches the counter originates in the source itself. With sources composed of carbon, hydrogen , and oxygen , the scattering factor is of the order of 5 percent, and can be neglected in obtaining an approximate estimate of the true activity. Prior workers have shown that the self-absorption factor, i s, for the beta radiation is given by the relationship :

is (3)

where p, is the self-absorption coeffici ent and AT is the radiation from a source containing a unit weight of the radioactive material with a thickness T; Ao is the limiting value for AT as T approaches zero. If the weight of the source is m mg and the area is

a cm2, T= m and eq (3) can be rewritten: a

(4)

When T is infinite with respect to beta particles, as in the analytical method, 1-e-I'T=1, and eq (4) becomes:

mAT a T=~· (5)

\iVhen T aproaches zero, i . approaches 1, and from eq (2) and the definition of Ao:

A =(CPS) = dps X ! X i. o m m g b

f,-'>1

(6)

Since mAT is the cps of a source weighing m milli­grams, eq (5), by substitution for mAT and Ao becomes:

a p,

(7)

Assuming 211" geometry (fu = 0.5) and no back­scattering (fb= l ), values of 1/1'- can be calculated from eq (7) for all of the measurements recorded in table 1. However, there is considerable uncertainty in the value of a for determinations made with the 2-in., commercial, flat-bottomed planchets, because of the curved edge. Hence, the best value of 1/ J.L is that obtained from the measurements for the stand­ardized sample G1 in the 10-cm2 areas. In the first experiment of table 1, a = 10.0, cps = 50_1 , and (dps /m ) Xig= (0.435 X 0_0856 X 37,000/7.7) X 0.5. From these values there is obtained:

1 / 2 -=0.0560 mg em . I'-

The value of l i p, can be used to obtain T 1I2 (the half-value thickness of the radiation) and R (the

174

Page 5: Tritium-labeled compounds I. Radioassay of tritium-labeled ... · The activity of tritium in the sample is given by the relationship: dps jJ.c = 37000, = cpsXmXk (1) where cps is

range) from the relationships [9, 17] T I/2= 0.693 /fJ. and R = 5/J.l- . With the above value of 1/J.l-, T1/2= 0.039 mg/cm2 and R = O.28 mg/cm2• This value of R can be compared with experimental values, reported in the literature, of 0.23 mg/cm2 [1 8], 0.5 mg/cm2 [19],1 0.6 mg/cm 2 [6], and with the value of 0.7 mg/cm2 based on an empirical energy-range relationship [9]. The variation in the reslllts arises from experimental difficulties and from uncertainty in the absolute disintegration rate of the source.

It is of interest to note that extrapolation of the observed counting rates of figure 1 to zero thickness gave a value of approximately 50 percent of the abso­lute disintegration rate, based on the NBS tritium standard. With a stainless-steel planchet, under conditions some'what comparable to those used here, Rydberg's data [6] for thin films indicate a back­scattering factor of about l.25. Thus the count would be expected to be approximately 62 percent of the absolute rate, with a stainless-steel planchet, 271" geometry, and no self-absorption. A low value might arise from unequal distribution of the material on the planchet. This factor is of no importance with thick samples, but could give rise to large errors with thin samples. Such errors make accurate assay of tritium materials in thin samples impracticable.

Assay of infinitely thick films is, how ever, highly satisfactory, and is recommended for routine, labora­tory analysis. In eq (1), the empirical constant k includes corrections for the area of the source, geometry of the counter , self-absorption, backscat­tering, and conversion of total dps to micro curies. By use of eq (1) and (7), le may be relaLed to a/ fJ. Lhus:

However, because of uncertainties in the values for a, f g, and f b, the constant le , for the calculation of ab olute radioactivity from observed counts, should be experimentally determined under the conditions of use.

7 . References [1] Nucleonics 16, No.3, 62 (195 ). [2] W. A. Jenkins, Anal. Chern. 25, 1477 (1953) . [3] T. E. Banks, J . C. Crawhall , and D . G. myth, Biochem.

J . 64, 4ll (1956) . [41 W. A. Jenkins and D . M. Yost, J. Chem. Phys. 20, 538

(1952) . [5] F. L . Jackson and H. W. Lampe, Anal. Chem. 28, 1735

(1956) . [6] J . Rydberg, Acta Chem. Scand. 12, 399 (1958). [7] M. L. Eidenoff and J . E. Knoll, Science 112, 250 (1950). [8] A. T . Nelms, Energy loss and range of electrons and

protons, NBS Circ. 577 (1956); Supplement (1958). [9] L . E . Glendenin, Nucleonics 2, No.1, 12 (1948).

[10] C. C. McCready, Nature 181, 1406 (1958). [ll] H . Campbell , H. A. Glastonbury, and M. D. Stevenson,

Nature 182, llOO (1958). [12] H . S. I sbell an d coworkers (unpublished work). [13] H. . Isbell , J . V. Karabinos, H . L . Frush , N . B. Holt,

A. Schwebel, and T . T . Galkowski, J. R esearch NBS 48, 163 (1952) RP2301.

[14] H . S. I sbell an d J . D . Moyer (unpublished work) . [15] A. Schwebel, H. S. Isbell , and J. D. Moyer, J . R esearch

NBS 53, 221 (1954) RP2537. [16] W. B. Mann and H . H. Seliger, Prepa ra tion, main te­

nan ce, and applicat ion of s tand a rds of radioactivi ty, NBS Circ. 594 (1958).

[171 W. F. Libby, Anal. Chern. 19, 2 (1947) . [18] S. C. Brown, Phys. R ev. 59, 954 (1941) . [19] L . M. Dorfman, Phys. Rev. 95, 393 (1954).

, 7 Value derived from the average of two half-thickness values experimentally ; WA SHING 'I'ON, D.C. determined by Dorfman. (Paper 63A2- 11) .