transmitter calibration
DESCRIPTION
CalibrationTRANSCRIPT
-
Isolator Calibration Example
Input is 4-20mA and output is 4-20mA
1. Input 4mA into the unit and
adjust the output with the
OFFSET (zero) control until the
output is also 4mA.
2. Input 20mA into the unit and
adjust the output with the SPAN
(gain) control until the output is
also 20mA.
3. Repeat step 1 and 2 if
necessary.
4. Check the output at half scale.
Input 12mA and check that the
output is 12mA.
T r a n s mi t t e r C a l ib r a t io n , O f fs e t ( Ze r o ) , S p a n
Most transmitters have two adjustments in common which are termed OFFSET (zero) and SPAN(gain). These adjustments allow the output s ignal to be varied considerably, generally 20% of
the transmitter range.
Offset
This adjustment sets what is in effect the value of output
when the input is zero. Figure 1 shows the effect of
altering the offset on a transmitter.
Span
As can be seen from
Figure 2, altering the
SPAN changes the
slope of therelationship between
input and output s ignals.
Calibration Principles
The principle used in the calibration of transmitters is that
if two points on a straight line are established then the
line itself is established. By examining Figure 2 it can be seen that at low magnitudes of input s ignal
the effect of span adjustment is very small. Consequently the firs t calibration point should be selected close to the bottom of the input
range. The offset adjustment is then used to give the required output s ignal for the input s ignal being fed to the transmitter.
Because alteration of the span results in a higher output deviation at the top end of the input range it follows that the second calibration
point should be at this top end, say in the top 25%. Thus the calibration at this point is achieved by imposing an input s ignal of suitable
magnitude and adjusting the span potentiometer to give the correct output.
In practice it is sometimes difficult to eliminate the interaction effects between offset and span so it is recommended that the procedure for
setting these two points be repeated until both points are obtained without the need for further adjustments.
Analogue Process Control Services is a division of NESS Corporation