transductor, bdot and magnetic field inside reference magnet box spectrums x. zhang, k. triplett
TRANSCRIPT
![Page 1: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/1.jpg)
Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums
X. Zhang, K. Triplett
![Page 2: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/2.jpg)
Summary
• Using Gaussmeter, the peak field measured near the two GPMS buses on the wall is about 25Gauss.
• The peak file inside the reference magnet box for the proposed new transductor location is about 7 Guass (combined, likely due to the bus), which is satisfying the recommend external field limit 50 Gauss for precision measurement.
• The 720Hz signal are near the noise level of about 80dB lower than 15Hz, except for the Bdot signal monitor.
![Page 3: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/3.jpg)
Hall Probe Inside Ref. Magnet Box
![Page 4: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/4.jpg)
Hall Probe Inside Ref. Magnet Box
![Page 5: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/5.jpg)
Hall Probe Inside Ref. Magnet Box
![Page 6: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/6.jpg)
Hall Probe Inside Ref. Magnet Box
![Page 7: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/7.jpg)
Hall Probe Inside Ref. Magnet Box
![Page 8: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/8.jpg)
TRANSDUCTOR/BDOT BUFFER MONITOR
![Page 9: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/9.jpg)
Transductor Buffer Monitor
![Page 10: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/10.jpg)
Transductor Buffer Monitor
![Page 11: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/11.jpg)
Bdot Buffer Monitor
![Page 12: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/12.jpg)
Bdot Buffer Monitor
![Page 13: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/13.jpg)
Reference Magnet Field Measurement
![Page 14: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/14.jpg)
![Page 15: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/15.jpg)
Gaussmeter By Analog Output
![Page 16: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/16.jpg)
Gaussmeter By Spectrum
Peak Field:2645 Gs
![Page 17: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/17.jpg)
Gaussmeter By Spectrum
Peak Field:2645 Gs
![Page 18: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/18.jpg)
Gaussmeter Bx Spectrum
Peak Field:45.53 Gs
![Page 19: Transductor, Bdot and Magnetic Field Inside Reference Magnet Box Spectrums X. Zhang, K. Triplett](https://reader033.vdocuments.site/reader033/viewer/2022052603/56649d965503460f94a7f2b6/html5/thumbnails/19.jpg)
Gaussmeter Bx Spectrum
Peak Field:45.53 Gs