tof analysis of bay water surrounding the battleship texas
DESCRIPTION
TOF Analysis of Bay Water surrounding the Battleship Texas. Presented at: SETAC 2013 South Regional Conference Work By: Sue D’Antonio, Joan Stevens, Andre Szczesniewski , Lynne Marshall, John Kimmel and Ken Helmreich . Agilent 6230 TOF LC/MS. - PowerPoint PPT PresentationTRANSCRIPT
TOF Analysis of Bay Water surrounding the
Battleship Texas
Presented at: SETAC 2013 South Regional Conference
Work By: Sue D’Antonio, Joan Stevens, Andre Szczesniewski , Lynne Marshall, John Kimmel and Ken Helmreich
Agilent 6230 TOF LC/MS
TOF and QTOF techniques (non-target compounds)• Identifies and differentiates compounds that were
previously “hidden” under background• Accurate mass TOF instruments:
– Can determine empirical formula based on the mass of the compound in question (accurate mass)
Novel data mining techniquesAccurate Mass Databases, Mass Profile Processional Software • Discover more compounds/contaminants, faster
and with greater precision
6/27/2013
API-TOF Ion Optical Design
6/27/2013Page 3
API Ion Source
Beam Shaper
TOF Mass Analyzer
Two StageIon Mirror
Optically CoupledIon Detector
Effective Flight PathLength of 2.0m
Formulae for TOF Performance
6/27/2013
Resolution = (m/z)/w1/2
1000/0.1 = 10,000
922/0.087 = 10,597
Mass accuracy (mass error)
(1000.001-1000.000)/1000)x106=1 ppm
Therefore: 1 ppm @ 1000 m/z is 0.001Da (1 mDa)
921.5 922.0 922.5 923.0 923.5 924.0 924.5m/z, amu
923.0121
W1/2 = 0.087
+TOF MS: 5.474 to 5.547 min from steroid2TOFap42.wiff Agilent, subtracted (4.696 to ... Max. 1.2e5 counts.
120 140 160 180 200 220 240 260 280 300 320 340 360 380 400m/z, amu
0.00
1.00e4
2.00e4
3.00e4
4.00e4
5.00e4
6.00e4
7.00e4
8.00e4
9.00e4
1.00e5
1.10e5
1.16e5
Inte
ns
ity, c
ou
nts
289.2159
271.2051
269.1893
287.1997
Epitestosterone
Resolution 6890
meas = 289.2159
actual = 289.2162
-1.06 ppm
922.0092
610actual
actual - measured accuracy Mass ´÷øö
çèæ=
How Much Accuracy is Needed?(Newer concept for LC/MS: ppm accuracy)
Reserpine (C33H40N2O9) has a protonated ion at 609.28066
Single quad reports mass to +/- 0.1 = 165 ppm
Number of possible formulas using only C, H, O & N:• 165 ppm209• 10 ppm 13• 5 ppm 7• 3 ppm 4• 2 ppm 2
Accurate mass reduces number of candidates and risk of investing effort with the wrong molecule
6/27/2013Page 5
6
SPE Sample Prep
6/27/2013
7 6/27/2013
8 6/27/2013
9
Analysis of Water Samples
6/27/2013
10
Choose Your Elements: Find Compounds by Molecular Feature and Generate Formulae
6/27/2013
11
Molecular Feature Generator
6/27/2013
12
Compound Chromatogram
6/27/2013
13
Deductive Compound Choices
6/27/2013
14 6/27/2013
15 6/27/2013
16
Conclusions
Utilizing Agilent’s Time of Flight accurate mass, high resolution, superior isotopic fidelity and the unique features of Mass Hunter software, such as Molecular Feature Extractor and Molecular Formula Generator, we were able to identify over 100 compounds found in the samples.
6/27/2013