title: investigations concerning chromium plating from ... · pdf filenodules (figure 3-4),...
TRANSCRIPT
1
Authors and contributors: Juergen Fischer, Eric Nagel, Michael Mann
Date (written): 12-20-2012
Report number: 120808 JKF
Approved:
Title: Investigations concerning chromium plating from electrolytes containing chromium-III
chloride and ionic liquid.
1 Introduction 2 Literature review 3 Chromium plated sample from the University of Leicester, UK 4 Instrumentation and electrochemical cell setup 5 Investigating electrolytes made from choline chloride and chromium-III
chloride hexahydrate 6 Investigating electrolytes made from choline chloride, chromium-III
chloride hexahydrate and additions 6.1 Adding water 6.2 Changing the chromium to ionic liquid ratio in the electrolyte 6.3 Adding lithium chloride 6.4 Adding potassium chloride 7 Investigating electrolytes made with 1-ethyl-3-methylimidazolium chloride
([EMIm]Cl), anhydrous chromium chloride 7.1 Varying the current density 7.2 Adding lithium chloride and changing the current density 8 Summary, conclusions and recommendations for future research Attachments
A References
B Cr plating tests
AEM Center
2
Investigations concerning chromium plating from electrolytes
containing chromium-III chloride and ionic liquid.
1 Introduction The investigations were done by the Advanced Engineered Materials Center (AEM Center) of the University of North Dakota (UND), Grand Forks, ND. The purpose of these experiments was to rapidly assess the feasibility of using ionic liquids to electroplate steel with chromium metal from a non-aqueous chromium-III plating bath. The chromium metal coating will function as a corrosion barrier, so the coating must be contiguous with no cracks; high thickness uniformity and minimal inclusion of impurities is also desired. Published literature [1] indicates crack-free coatings of chromium have been deposited using choline chloride and chromium (III) chloride hexahydrate eutectic mixtures. The experimental tasks are meant to use the published methodologies for electrodeposition of chromium with ionic liquids [1-4] as a base-line to develop a commercially useful process for plating chromium onto steel with an electrolyte containing chromium-III salt and ionic liquid. Later we decided to do some testing in 1-ethyl-3-methylimidazolium chloride ([EMIm]Cl) with anhydrous chromium chloride to avoid the possibility of introducing hydrogen embrittlement into the substrate.
3
2 Literature review The references mentioned are listed in attachment A. A good overview of the aqueous chromium plating is given by M. Schlesinger and M. Paunovic [5]. The basic information about electrodeposition from ionic liquids can be found in a book from F. Endres [6]. Newer information about chromium plating with ionic liquids is provided in [1, 2, 4, 7 to 10]. The most promising approach seems to use a mixture of choline chloride, chromium-III chloride and water as published in 2010 by E. Smith et. al. [4]: The exact electrolyte composition is proprietary, but chromium(III) chloride hexahydrate, choline chloride and water are the only components. The bath water content is monitored by electrolyte conductivity. The plating temperature is kept at 40 oC using a cold water cooling coil and the current density is set at 600 mA/cm2. The plating time is generally 45 to 60 minutes. The current efficiency for black chromium can be up to 90 %, where for hard chromium it is in the region of 40 %. The overall current efficiency depends strongly on the composition of the electrolyte and the anode material. The pathways of lost current density are mostly known and well defined, but are related to the proprietary electrolyte composition and cannot be detailed. The amount of micro cracking found in the chromium layer varies with current density and plating time. The adherence of the chromium layer needs improvement. The absolute current density of 600 mA/cm2 and/or the general plating time seems very high for a current efficiency of 40 or 90 %. For chromium A chromium plated steel tube sample from the IONMET consortium was requested, received and analyzed (see Chapter 3). In 2007, P. Benaben [7] reported about a similar electrolyte: At about 40 oC the current density was 150 mA/cm2 and the deposition rate around 50 µm/hour. This suggests a current efficiency of 10 % (see Chapter 6.1 and 6.2). A. Abbott et. al. published 2004 the electrodeposition of chromium black from an electrolyte consisting of choline chloride, chromium chloride hexahydrate and lithium chloride (molar ratio = 1:2:2.8) at 60 oC and a very low current density of -0.345 mA/cm2 (see Chapter 6.3). S. Eugénio et. al. [8, 9, 10] published findings of precipitation of black chromium which consists of chromium oxide, chromium hydroxide and metallic chromium. The electrolyte consisted of 1-butyl-3-methylimidazolium tetrafluoroborate and chromium(III) chloride hexahydrate. The electrodeposition occurred in two steps involving Cr(III) and Cr(II) electroactive species. In 1997, M. Ali et. al. [11] investigated the electrodeposition of aluminum–chromium alloys from electrolytes made from chromium(II) chloride, aluminum chloride and N-(n-butylpyridinium chloride.
4
3 Chromium plated sample from the University of Leicester, UK The AEM Center requested from the University of Leicester a steel sample plated with a crack free, 10 micrometer chromium layer, using their proprietary electrolyte consisting of choline chloride, chromium-III chloride and water for the plating. The sample piece from Leicester is shown in Figure 3-1 was evaluated with the scanning electron microscope (SEM) and energy-dispersive X-ray spectroscopy (EDS). It is a copper tube with 1” outer diameter. The plated chromium layer is not well adhered to the surface and was partially flaked off (Figure 3-1). The chromium layer has a grey metallic shine, is 20 to 25 µm thick (Figure 3-2), shows cracks (Figure 3-3 and 3-4) and nodules (Figure 3-4), and the typical EDS surface analysis gives 42 at.% chromium with 35 at.% oxygen and carbon (Figure 3-5). In some of our experiments we came close to these values but we only produced “black chromium” powder, not metallic chromium.
Figure 3-1: 1” outer diameter copper tube plated with chromium by the University of Leicester, UK using an electrolyte probably containing chromium-III chloride, water and choline chloride.
Chromium coating flaked off and the oxidized copper beneath was revealed. Chromium plating with cracks (see next three SEM pictures and the following EDS analysis)
5
Figure 3-2: SEM picture of chromium layer flakes with thickness measurements.
Figure 3-3 and 3-4: SEM pictures (100X and 1000X) of the middle region of the chromium plated sample shown in Figure 3-1.
Figure 3-5: Typical EDS spectrum with analysis of the chromium layer shown in Figure 3-3.
6
4 Instrumentation and electrochemical cell setup All electrodeposition experiments were conducted in a nitrogen-filled glove box seen in Figure 4-1. The atmosphere in the glove box was maintained at less than 0.5 ppm H2O and O2.
Figure 4-1: Glove box containing an inert gas atmosphere; atmosphere maintained at <0.5 ppm H2O and O2. The electrochemical cell for the deposition experiments consisted of a 25 mL round-bottomed, three-necked flask, into which two chromium anodes and a cylindrical cathode were inserted. The cathode was made of copper, 420 steel, or nickel, depending on the experiment. A picture of the electrochemical cell setup is shown in Figures 4-2 and 4-3, while a close-up view of the setup used to ensure electrical contact with the rotating cathode is shown in Figure 4-4. The rotating cathode was used to agitate the electrolyte and ensure a constant availability of Cr-ions to the surface of the cathode.
7
Figure 4-2: 50 mL electrochemical cell with temperature probes and electrical contacts.
Figure 4-3: Electrochemical cell setup
1. Stirrer motor (app. 600 rpm for all experiments) 2. Flexible coupling 3. Contact for the cathodic stirrer (see Figure 5.3 for details) 4. PTFE covered stirring rod in glass bearing 5. Cylindrical copper, steel or nickel cathode in 50 mL three neck round flask with PTFE stoppers, thermoelement and two chromium anodes with electrical platinum contact to rectifier 6. Aluminum bowl with copper shot on aluminum plates 7. Thermoelement controlled hotplate
1. Adapter with the PTFE covered stirrer. 2. Thermoelement in the electrolyte 3. Thermoelement in the copper shot to control the temperature 4. Electrical contacts for both anodes 5. Cathode screwed to the stirrer in the electrolyte
8
Figure 4-4: Rotating cathode electrical connection setup The electrolyte used in the experiments consisted of approximately 25 mL of a choline chloride/chromium-III chloride hexahydrate mixture. All electrochemical cell measurements were carried out with a Solartron 1287 potentiostat/galvanostat controlled by CorrWare software. A pass-through fitting was employed to connect the instrument from outside the glove box to the electrochemical cell inside the glove box. A picture of the Solartron instrument is shown in Figure 4-5.
Figure 4-5: The Solartron 1287 potentiostat/galvanostat used for all electrodeposition experiments with pass-through fitting for the glove box. The electrodeposition experiments began by attempting to first deposit chromium onto a nickel cathode. Cathode material was later changed to 420 steel. Because 420 steel contains 13 w.% chromium we encountered complications during EDS analysis of the precipitated layer and switched to copper cathodes.
All experiments were conducted galvanostatically, with current densities varied between -0.345 mA cm-2 to -154 mA/cm2. Two chromium anodes and the copper, steel, or nickel cathode was weighed prior to the start of the experiment. The two 1.0 cm by 1.5 cm pure chromium anodes were positioned on opposite sides of the electrochemical cell,
1. End of flexible coupling 2. Copper leads for spring loaded carbon contact brushes 3. Contact brushes for the cathodic stirrer 4. Electrical contacts for brushes 5. Aluminum housing with holder
9
with the 3 cm2 rotating cylindrical cathode in the center of the cell. The rate of rotation of the cylindrical cathode was 600 rpm. This equates to a cathode surface speed relative to the electrolyte of about 30 cm/s. The chromium anodes were connected to the working electrode of the Solartron, while the cathode was connected to the counter electrode. Reference electrodes were connected directly to the working and counter electrodes. In each experiment, the cell potential versus time was recorded.
After the various deposition time, the anodes and cathode were removed from the plating cell, removed from the glove box, and rinsed in ethanol and acetone. The mass of each anode and cathode was recorded, and the mass difference from before the experiment was noted. The surface of each cathode and anode were then analyzed in the scanning electron microscope (SEM).
10
5 Investigating electrolytes made from choline chloride and chromium-III chloride hexahydrate
This chapter describes the investigation of electrodeposition of chromium from electrolytes made from choline chloride and chromium-III chloride hexahydrate. All plating experiments in this report were conducted galvanostatically, so as to compare the performance of the electrolyte at different current densities and concentrations. For the tests 1Cr and 2Cr a molar ratio of 3.1:1 for chromium-III chloride hexahydrate to choline chloride was used. In 8Cr the composition was 2:1. All the chromium plating experiments are listed in Attachment B. Table 5-1 lists the tests relevant for this chapter. In the first test we plated onto a nickel cylinder cathode before we switched to steel 420 and later to copper cylindrical cathodes. Pure chromium plates were the anodes in all reported plating attempts. Figure 5-1 provides information about the cell voltage versus plating time for the three test. The maximum voltage of the galvanostat (15 to 16 V) was reached fast if the current density was high. In the 1Cr-test run, the desired current density of -150 mA/cm2 was never reached. An average current density of -140 mA/cm2 was established under these conditions. At lower absolute current densities (10.34 mA/cm2 in average) the desired current density (15.00 mA/cm2) could be maintained for only the first 3000 seconds. It seems as though an insulating layer builds up on one or both electrodes under these conditions. At higher absolute current densities (>9 mA/cm2) the cathodes turn black (Figures 5-2 and 5-9). After plating, the electrodes were always rinsed with ethanol and acetone and then dried in the air. The 1Cr-cathode, which was plated with -140mA/cm2, is covered with a crust of black powder (Figures 5-3, 5-5, 5-10, 5-12) containing 35 at.% chromium and 55 at.% oxygen (Figure 5-4). As reported in Chapter 3, the EDS analysis of the chromium plated sample from the University of Leicester gave 42 at.% chromium and 35 at.% oxygen. Plating with an average of 10.3 mA/cm2 resulted in a crust with 13 at.% chromium and 68 at.% oxygen (Figure 5-11). A current density of -0.34 mA/cm2 resulted in corrosion of both electrodes (Table 5-1) but no plating on the copper cathode (Figure 5-15 till 5-18). The anode lost weight in all cases and looked etched. (Figure 5-6, 5-7, 5-13, 5-18, 5-19, 5-21). The EDS of the “pure” chromium anode surface showed 67 +- 3 at.% chromium with 26 +-2 at.% oxygen (Figure 5-8, 5-14, 5-20).
11
Table 5-1: Chromium plating experiments 1Cr, 2Cr and 8Cr Chromium Plating Experiment
Number 1 Cr 2 Cr 8 Cr
Electrolyte compositionCrCl3*6H2O : Cholin
chloride = 3.10:1
CrCl3*6H2O : Cholin
chloride = 3.10:1
CrCl3*6H2O :
Cholin chloride =
2:1
Electrolyte history Newly prepared 1Cr New
Plating temperature oC 58-70oC (no cooling) 72-46 60
Stirrer setting 1-10 5 = 645-675 rpm 5 = 634-658 rpm 5
Start weight cathode g 2.8111 2.0039 2.2122
Start weight anode 1 g 2.1837 2.1450 2.2289
Start weight anode 2 g 2.6460 2.6134 1.8544
Plating time sec 1200 12000 7200
Area of cathode cm2 3.15568 2.67257 2.594484293
Current mA -441.66 -27.6343738 -0.8950978
Cathode material Nickel Type 420 steel (13%Cr) Copper
Current density mA/cm2 -139.96 -10.34 -0.35
Plating remark 1
At the beginning the
cell voltage was the
Solartron maximum
and the current too
low.
The cell voltage was
the Solartron
maximum and the
current too low.
positive voltage
was observed
Plating remark 2
The wanted current
was -473 mA. The
average current was
only -442 mA.
The wanted current
was-40.1 mA. The
average current was
only -27.6 mA.
End weight cathode g 2.8239 2.0270 2.212
End weight anode 1 g 2.1498 2.1255 2.2231
End weight anode 2 g 2.6174 2.5969 1.8499
Weight gain (cathode) g 0.0128 0.0231 -0.0002
Weight loss (anodes) g 0.0625 0.0360 0.0103
How did the electrolyte appearance change? no change no change no change
How did the cathode look like?Covered with black
powder
Unevenly covered
with black powderno plating
How did the anode look like? Partially etched Partially etched no change
Cathodic current efficiency % 13.44 38.78 -17.28
Anodic current efficiency % 65.65 60.43 889.71
Open circuit cell voltage V
Characteristic cell voltage level V -15.6 -16 ±0.03
Analysis of cathode surface in atomic %
% C 2.734 14.5 34.6
% N 2.687 3.2 0.0
% O 54.925 67.8 3.2
% Cl 4.504 1.0 0.1
% Fe 0.142 *Ni not Fe 0.2 0.0
% Cu 0 0.0 61.8
%Cr 35.0 13.4 0.2
Analysis of cathode surface 2 in atomic % n/a n/a n/a
% C n/a n/a n/a
% N n/a n/a n/a
% O n/a n/a n/a
% Cl n/a n/a n/a
% Fe n/a n/a n/a
% Cu 0 0.0 n/a
%Cr n/a n/a n/a
Analysis of cathode surface 3 in atomic % n/a n/a n/a
% C n/a n/a n/a
% N n/a n/a n/a
% O n/a n/a n/a
% Cl n/a n/a n/a
% Fe n/a n/a n/a
% Cu 0 0.0 n/a
%Cr n/a n/a n/a
Analysis of anode surface in atomic %
% C 5.7 2.3 2.851
% N 2.2 2.1 1.836
% O 27.0 26.8 24.885
% Cl 0.2 0.2 0.07
% Fe 0.243 *Ni not Fe 0.0 0
% Cu 0 0.0 0.233
%Cr 64.7 68.7 70.1
12
Figure 5-1: Cell voltage versus plating time for the 1st test series
Figure 5-2: Electrodes after test 1Cr. Figure 5-3: 1Cr-cathode (100X)
13
Figure 5-4: Spectrum with analysis of the 1Cr-cathode
Figure 5-5: 1000X SEM picture of the 1Cr-cathode
Figure 5-6: 1Cr-anode (100X) Figure 5-7: 1Cr-anode (1000X)
14
Figure 5-8: Typical spectrum with analysis of the 1Cr-anode
Figure 5-9: 2Cr-cathode Figure 5-10: 100X SEM picture of the 2Cr-cathode
Figure 5-11: Typical spectrum with analysis of the 2Cr-cathode
15
Figure 5-12: 2Cr-cathode crust (1000X) Figure 5-13: 2Cr-anode (100X)
Figure 5-14: Typical spectrum with analysis of the 2Cr-anode
Figure 5-15: 8Cr-cathode Figure 5-16: 100X SEM picture of the 8Cr-cathode
16
Figure 5-17: Typical spectrum with analysis of the 8Cr-cathode
Figure 5-18: 8Cr cathode (1000X) Figure 5-19: 8Cr anode (100X)
Figure 5-20: Typical spectrum with analysis of the 8 Cr-anode
17
Figure 5-21: 8 Cr anode (1000X) 6 Investigating electrolytes made from choline chloride, chromium-III
chloride hexahydrate and additions 6.1 Adding water This chapter describes the investigation of electrodeposition of chromium from electrolytes made from choline chloride, chromium-III chloride hexahydrate and additional water. All plating experiments in this chapter were conducted galvanostatically at -15 mA/cm2. The intent of this chapter is to compare the performance of the electrolyte at different water contents (2Cr till 4Cr) and different plating times (4Cr and 5Cr). A molar ratio of 3.1 to 1 for chromium-III chloride hexahydrate to choline chloride was used. For every mol chromium-III chloride hexahydrate 1.5 mol of water was added for 3Cr and 3 mol of water for 4Cr. 5Cr is a repeat of 4Cr with longer plating time. Unfortunately, the cathode for 4Cr, 5Cr and 6Cr was left for 65 hours in the electrolyte and lots of the plated substance fell off during that time. Only small amounts remained at the surface for EDS analysis. All the chromium plating experiments are listed in Attachment B. Table 6.1-1 lists the tests relevant for this chapter. We plated onto a steel 420 as cathodes and with pure chromium plates as anodes. Figure 6.1-1 informs about the cell voltage versus plating time for the four tests. As mentioned in Chapter 5, the desired absolute current density of 15 mA/cm2 was never reached in the 2Cr-test run. An average current density of -10.3 mA/cm2 was established under these conditions with a -16 V cell voltage limit. The higher the water content the lower the absolute cell voltage at 15 mA/cm2. The 5Cr-curve matches nicely with the 4Cr-curve like it should be because 5Cr was a repeat of 4Cr with fresh electrolyte and longer plating time. We decided to use the longer plating time because we saw the possibility that we create Cr(II) compounds at the electrodes. With longer plating times we tried to better the outcome of the plating.
18
Table 6.1-1: Chromium plating experiments 2Cr till 5Cr Chromium Plating Experiment
Number 2 Cr 3 Cr 4 Cr 5 Cr
Electrolyte compositionCrCl3*6H2O : Cholin
chloride = 3.10:1
CrCl3*7.5H2O : Cholin
chloride = 3.10:1
CrCl3*9H2O : Cholin
chloride = 3.10:1
CrCl3*9H2O : Cholin
chloride = 3.10:1
Electrolyte history 1Cr 1-2Cr and water 1-3Cr and water 1-4Cr
Plating temperature oC 72-46 48.0 50 50
Stirrer setting 1-10 5 = 634-658 rpm 5 5 5
Start weight cathode g 2.0039 1.6053 1.4488 1.567
Start weight anode 1 g 2.1450 1.9622 1.9291 1.8279
Start weight anode 2 g 2.6134 2.397 2.3754 2.2654
Plating time sec 12000 12000 12000 60300
Area of cathode cm2 2.67257 2.6246436 2.506991 2.5533043
Current mA -27.6343738 -39.37 -37.6048641 -38.29956445
Cathode material Type 420 steel (13%Cr) Type 420 steel (13%Cr) Type 420 steel (13%Cr) Type 420 steel (13%Cr)
Current density mA/cm2 -10.34 -15.00 -15.00 -15.00
Plating remark 1
The cell voltage was
the Solartron
maximum and the
current too low.
Magnetic stirrer in the
electrolyte caused
scratches on cathode.
Plating remark 2
The wanted current
was-40.1 mA. The
average current was
only -27.6 mA.
Cathode was looking
metallic grey but was
left for 65 hours in the
electrolyte. See 6Cr
remark.
End weight cathode g 2.0270 1.6561 1.444 1.7256
End weight anode 1 g 2.1255 1.9298 1.8488 1.712
End weight anode 2 g 2.5969 2.3759 2.329 2.1008
Weight gain (cathode) g 0.0231 0.0508 -0.0048 0.1586
Weight loss (anodes) g 0.0360 0.0535 0.1267 0.2805
How did the electrolyte appearance change? no change no change no change no change
How did the cathode look like?Unevenly covered
with black powderDark. Black. Powdery
Looked metallic grey
after 65h in
electrolyte.
Unevenly covered
with black powder
How did the anode look like? Partially etched Etched. Etched. Etched.
Cathodic current efficiency % 38.78 59.86 -5.92 38.23
Anodic current efficiency % 60.43 63.04 156.30 67.61
Open circuit cell voltage V
Characteristic cell voltage level V -16 -5.3 -3.3 -3.5
Analysis of cathode surface in atomic % smooth, shiny outer crust
% C 14.5 12.8 15.1 19.7
% N 3.2 3.2 0.1 4.2
% O 67.8 64.8 5.7 50.8
% Cl 1.0 1.6 0.1 3.5
% Fe 0.2 0.2 62.9 -
% Cu 0.0 0.0 0.0 0.0
%Cr 13.4 17.4 16.1 21.8
Analysis of cathode surface 2 in atomic % n/a n/a black spot inner crust
% C n/a n/a 9.171 16.785
% N n/a n/a 2.227 3.556
% O n/a n/a 61.387 55.191
% Cl n/a n/a 3.291 2.07
% Fe n/a n/a 0.438 -
% Cu 0.0 0.0 0.0 0.0
%Cr n/a n/a 23.5 22.4
Analysis of cathode surface 3 in atomic % n/a n/a n/a n/a
% C n/a n/a n/a n/a
% N n/a n/a n/a n/a
% O n/a n/a n/a n/a
% Cl n/a n/a n/a n/a
% Fe n/a n/a n/a n/a
% Cu 0.0 0.0 0.0 0.0
%Cr n/a n/a n/a n/a
Analysis of anode surface in atomic %
% C 2.3 2.7 0.0 4.8
% N 2.1 3.9 1.8 1.9
% O 26.8 30.3 20.1 20.0
% Cl 0.2 0.1 0.2 0.1
% Fe 0.0 0.3 0.1 -
% Cu 0.0 0.0 0.0 0.0
%Cr 68.7 62.7 77.8 73.2
19
Figure 6.1-1: Cell voltage versus plating time for the 2nd test series
Figure 6.1-1: 3Cr-electrodes Figure 6.1-2: 3Cr-cathode crust (100X) The results regarding test 2Cr are reported in Chapter 5. Figure 6.1-1, 6.1-7, 6.1-16 shows the electrodes for 3Cr till 5Cr. All cathodes were covered with black powder after the plating attempts. The 3Cr cathode was partially touching a PTFE-covered stirring magnet which compressed the plating in this area. The 4Cr-cathode and the 5Cr-cathode lost most of the powder because they were allowed to sit for 65 hours (over the weekend) in the electrolyte after plating. During rinsing more of the powder came off. Nevertheless we found the remaining powder (Figure 6.2, -4, -5, -8,-10, -11) with the SEM and using EDS analysis we determined the chromium content (Figure 6.1-3, -12, -18). The chromium content in the precipitate increased with the water content in the electrolyte (Table 6.1-1) from 13 at.% to 28 at.% (Figure 6.1- 12). The anodes lost weight during plating. EDS analysis showed 63 till 78 at.% Cr on the anodes.
20
Figure 6.1-3: Typical spectrum with analysis of the 3Cr-cathode
Figure 6.1-4: 3Cr-cathode crust (1000X) Figure 6.1-5: 3Cr-anode (100X)
Figure 6.1-6: Typical spectrum with analysis of the 3Cr-anode
21
Figure 6.1-7: 4Cr-electrodes Figure 6.1-8: 4Cr-cathode (100X)
Figure 6.1-9: Typical spectrum with analysis of the 4Cr-cathode without crust
Figure 6.1-10: 4Cr-cathode with crust (100X) Figure 6.1-11: 4Cr-cathode crust
22
Figure 6.1-12: Typical spectrum with analysis of the 4Cr-cathode precipitate
Figure 6.1-13: 4Cr-anode (100X) Figure 6.1-14: 4Cr-anode (1000X)
Figure 6.1-15: Typical spectrum with analysis of the 4Cr-anode
23
Figure 6.1-16: 5Cr-cathode Figure 6.1-17: 5Cr-cathode crust (100X)
Figure 6.1-18: Typical spectrum with analysis of the 5Cr-cathode precipitate
Figure 6.1-19: 5Cr-cathode crust (1000X) Figure 6.1-20: 5Cr-anode (100X)
24
Figure 6.1-21: Typical spectrum with analysis of the 5Cr-anode in a heavily etched area with high current density (see Figure 6.1-20) 6.2 Changing the chromium to ionic liquid ratio in the electrolyte This chapter describes the investigation of electrodeposition of chromium from electrolytes made from choline chloride, chromium-III chloride, and water in a molar ratio of 1: 3.1: 27.9 and 1:2:18. All plating experiments in this chapter were conducted galvanostatically, at -15 mA/cm2, only for 7Cr -60 mA/cm2 was used. All the chromium plating experiments are listed in Attachment B. Table 6.2-1 lists the tests relevant for this chapter. Normally we plated onto steel 420 as cathodes, and with pure chromium plates as anodes. In test 7Cr a copper cathode was used. Figure 6.2-1 shows the cell voltage versus plating time for all four platings. In the previous chapter we discussed the results for 4Cr and 5Cr. The curve for 6Cr gives a higher absolute cell voltage suggesting that the absolute cell voltage increases with lower chromium chloride concentration. The curve for 7Cr in Figure 6.2-2 goes down to an absolute cell voltage twice that of 7Cr because of the four-fold increase in current density. The visual appearance of the 6Cr and 7Cr is documented in Figures 6.2-3 and -7. The crust is shown in Figures 6.2-4, -6, -11. The EDS analysis of the crust on the 6Cr 420 steel cathode, which itself contains around 14 at.% chromium at the surface, resulted in chromium contents from 25 to 40 at.% chromium. To avoid the uncertainty of the influence of the chromium content in the substrate, we decided to switch to copper cathodes. The 6Cr anode had 43 at.% chromium at the surface (Figure 6.2-9), which is a relatively low value. At a -60 mA/cm2 current density, we produced a crust with 34 at.% chromium on the 7Cr cathode. (Figure 6.2-12)
25
Table 6.2-1: Chromium plating experiments 4Cr till 7Cr Chromium Plating Experiment
Number 4 Cr 5 Cr 6 Cr 7 Cr
Electrolyte compositionCrCl3*9H2O : Cholin
chloride = 3.10:1
CrCl3*9H2O : Cholin
chloride = 3.10:1
CrCl3*9H2O : Cholin
chloride = 2:1
CrCl3*9H2O : Cholin
chloride = 2:1
Electrolyte history 1-3Cr and water 1-4Cr New 6 Cr
Plating temperature oC 50 50 50 40
Stirrer setting 1-10 5 5 5 5
Start weight cathode g 1.4488 1.567 1.458 2.4684
Start weight anode 1 g 1.9291 1.8279 2.2828 2.2513
Start weight anode 2 g 2.3754 2.2654 1.9011 1.8728
Plating time sec 12000 60300 9000 1800
Area of cathode cm2 2.506991 2.5533043 2.524112 2.8043427
Current mA -37.6048641 -38.29956445 -37.872168 -168.26056
Cathode material Type 420 steel (13%Cr) Type 420 steel (13%Cr) Type 420 steel (13%Cr) Copper
Current density mA/cm2 -15.00 -15.00 -15.00 -60.00
Plating remark 1
Problems with the
electrical connection.
Bad connection
between rod and
cathode?
lots of gas evolution
Plating remark 2
Cathode was looking
metallic grey but was
left for 65 hours in the
electrolyte. See 6Cr
remark.
Cathode was covered
with black powdery
crust which
disappeared after 65
hours in the
electrolyte.
End weight cathode g 1.444 1.7256 1.4524 2.4763
End weight anode 1 g 1.8488 1.712 2.2524 2.2299
End weight anode 2 g 2.329 2.1008 1.8743 1.8551
Weight gain (cathode) g -0.0048 0.1586 -0.0056 0.0079
Weight loss (anodes) g 0.1267 0.2805 0.0572 0.0391
How did the electrolyte appearance change? no change no change no change no change
How did the cathode look like?
Looked metallic grey
after 65h in
electrolyte.
Unevenly covered
with black powder
Covered with
powdery black crust.
Looked metallic grey
after 65h in
Dark. Black. Powdery
How did the anode look like? Etched. Etched. Etched. Etched.
Cathodic current efficiency % -5.92 38.23 -9.15 14.52
Anodic current efficiency % 156.30 67.61 93.42 71.87
Open circuit cell voltage V
Characteristic cell voltage level V -3.3 -3.5 -3.6 -7.3
Analysis of cathode surface in atomic % smooth, shiny outer crust smooth
% C 15.1 19.7 20.6 8.8
% N 0.1 4.2 0.0 2.3
% O 5.7 50.8 23.6 49.7
% Cl 0.1 3.5 0.1 5.0
% Fe 62.9 - 38.5 0.0
% Cu 0.0 0.0 0.0 0.5
%Cr 16.1 21.8 17.2 33.7
Analysis of cathode surface 2 in atomic % black spot inner crust smooth n/a
% C 9.171 16.785 15.6 n/a
% N 2.227 3.556 0.0 n/a
% O 61.387 55.191 22.8 n/a
% Cl 3.291 2.07 0.1 n/a
% Fe 0.438 - 43.1 n/a
% Cu 0.0 0.0 0.0 n/a
%Cr 23.5 22.4 18.4 n/a
Analysis of cathode surface 3 in atomic % n/a n/a crust n/a
% C n/a n/a 4.6 n/a
% N n/a n/a 3.5 n/a
% O n/a n/a 47.7 n/a
% Cl n/a n/a 3.1 n/a
% Fe n/a n/a 0.5 n/a
% Cu 0.0 0.0 0.0 n/a
%Cr n/a n/a 40.5 n/a
Analysis of anode surface in atomic %
% C 0.0 4.8 11.9 n/a
% N 1.8 1.9 5.5 n/a
% O 20.1 20.0 39.5 n/a
% Cl 0.2 0.1 0.3 n/a
% Fe 0.1 - 0.2 n/a
% Cu 0.0 0.0 0.0 n/a
%Cr 77.8 73.2 42.6 n/a
Analysis of crystals in electrolyte
26
Figure 6.2-1: Cell voltage versus plating time for the 3rd test series
Figure 6.2-2: Cell voltage versus plating time for the 4th test series
27
Figure 6.2-3: 6Cr-cathode with some Figure 6.2-4: 6Cr-cathode (100X) with remaining plating leftover plated crust
Figure 6.2-5: Spectrum with analysis of the 6Cr-cathode remaining crust shown in
Figure 6.2-4
Figure 6.2-6: 6Cr-cathode crust (1000X) Figure 6.2-7: 6Cr-anode (100X)
28
Figure 6.2-8: Spectrum with analysis of the 6Cr-cathode remaining crust shown in
Figure 6.2-6
Figure 6.2-9: Spectrum with analysis of the 6Cr-anode
Figure 6.2-10: 7Cr-cathode Figure 6.2-11: 7Cr-cathode (100X)
29
Figure 6.2-12: Spectrum with analysis of the 7Cr-cathode precipitation 6.3 Adding lithium chloride After test 5Cr, it was decided to finalize the test series with additional water in the electrolyte and to investigate the effect of lithium chloride as an addition. The basis for this decision was the publication from A. Abott et. al. [2]. They reported “that thick adherent, crack free deposits of chromium black can be electrodeposited from an ionic liquid containing Cr (III)”. An electrolyte containing choline chloride and chromium(III) chloride hexahydrate (molar ratio = 1 : 2) with 15 wt.% lithium chloride was used with copper cathodes at 60 oC for 2 hours at a current density of -0.345 mA/cm2. We used a copper cathode for our tests and similar process parameters. This chapter describes the investigation of electrodeposition of chromium from electrolytes made from choline chloride, chromium-III chloride, water, and lithium chloride in a molar ratio of 1:2:12:0 and 1:2:12:2.8. All plating experiments in this chapter were conducted galvanostatically at varying current densities. The intent of this chapter is to compare the performance of the electrolyte with and without the addition of lithium chloride (8Cr and 9Cr) in the 5th test series, as well as the effect of current density on plating solutions containing LiCl in the 6th test series. In the 5th test series we decided to use the low current density of -0.345 mA/cm2 recommended in the literature [2] and got no plating. All the chromium plating experiments are listed in Attachment B. Table 6.3-1 lists the tests relevant for this chapter. Copper cathodes were used in all experiments presented in this chapter. Figure 6.3-1 shows the cell voltage versus plating time for 8Cr and 9Cr. Because of the obvious influence of corrosion (see current efficiency in Table 6.3-1) the cell voltage is close to 0. The 8Cr curve documents some contact problems. No precipitate was found on the cathode (Figure 6.3-3).
30
Table 6.3-1: Chromium plating experiments 8Cr till 11Cr Chromium Plating Experiment
Number 8 Cr 9 Cr 10 Cr 11 Cr
Electrolyte composition
CrCl3*6H2O :
Cholin chloride =
2:1
CrCl3*6H2O :
Cholin chloride =
2:1 and 15 w.% LiCl
CrCl3*6H2O : Cholin
chloride = 2:1 and
15 w.% LiCl
CrCl3*6H2O : Cholin
chloride = 2:1 and
15 w.% LiCl
Electrolyte history New 8 Cr 8-9Cr 8-10Cr
Plating temperature oC 60 60 60 60
Stirrer setting 1-10 5 5 5 5
Start weight cathode g 2.2122 2.0055 2.1978 2.5046
Start weight anode 1 g 2.2289 2.1918 2.3143 1.8095
Start weight anode 2 g 1.8544 1.8353 1.8306 2.2849
Plating time sec 7200 7200 7200 7200
Area of cathode cm2 2.594484293 2.726616538 2.549213943 2.858676
Current mA -0.8950978 -0.940682706 -8.794788103 -28.58676
Cathode material Copper Copper Copper Copper
Current density mA/cm2 -0.35 -0.35 -3.45 -10.00
Plating remark 1positive voltage
was observed
large amounts of
carbon from
brushes present
Plating remark 2
End weight cathode g 2.212 2.0054 2.1988 2.5066
End weight anode 1 g 2.2231 2.1901 2.3097 1.7938
End weight anode 2 g 1.8499 1.8342 1.827 2.275
Weight gain (cathode) g -0.0002 -0.0001 0.0010 0.0020
Weight loss (anodes) g 0.0103 0.0028 0.0082 0.0256
How did the electrolyte appearance change? no change no change no change no change
How did the cathode look like? no plating no plating Black. Powdery
black w/ copper
showing through
How did the anode look like? no change no change Partially etchedno change / stained
Cathodic current efficiency % -17.28 -8.22 8.79 5.41
Anodic current efficiency % 889.71 230.14 72.09 69.24
Open circuit cell voltage V
Characteristic cell voltage level V ±0.03 -0.08 -4.1 range: -5 -> -9
Analysis of cathode surface in atomic % crust
% C 34.6 27.7 10.2 7.9
% N 0.0 0.0 3.1 3.3
% O 3.2 2.9 56.8 61.4
% Cl 0.1 0.2 1.1 2.6
% Fe 0.0 0.0 0.0 0.0
% Cu 61.8 68.8 3.3 0.1
%Cr 0.2 0.4 25.4 20.2
Analysis of cathode surface 2 in atomic % n/a n/a n/a smooth
% C n/a n/a n/a 11.569
% N n/a n/a n/a 3.828
% O n/a n/a n/a 57.59
% Cl n/a n/a n/a 1.522
% Fe n/a n/a n/a 0
% Cu n/a n/a n/a 19.004
%Cr n/a n/a n/a 6.5
Analysis of cathode surface 3 in atomic % n/a n/a n/a n/a
% C n/a n/a n/a n/a
% N n/a n/a n/a n/a
% O n/a n/a n/a n/a
% Cl n/a n/a n/a n/a
% Fe n/a n/a n/a n/a
% Cu n/a n/a n/a n/a
%Cr n/a n/a n/a n/a
Analysis of anode surface in atomic %
% C 2.851 n/a 3.2 1.7
% N 1.836 n/a 3.3 4.4
% O 24.885 n/a 33.0 30.0
% Cl 0.07 n/a 0.4 0.1
% Fe 0 n/a 0.0 0.0
% Cu 0.233 n/a 0.2 0.3
%Cr 70.1 n/a 59.9 63.6
31
Figure 6.3-1: Cell voltage versus plating time for the 5th test series
Figure 6.3-2: Cell voltage versus plating time for the 6th test series
32
Figure 6.3-3: 8Cr-cathode Figure 6.3-4: 9Cr-cathode
Figure 6.3-5: 10Cr-cathode Figure 6.3-6: 10Cr-cathode (100X)
Figure 6.3-7: Spectrum with analysis of the 10Cr-cathode surface in Figure 6.3-6
33
Figure 6.3-8: 10Cr-cathode (1000X) Figure 6.3-9: 10Cr-anode (100X)
Figure 6.3-10: Spectrum with analysis of the marked 10Cr-cathode precipitation in Figure 6.3-8
Figure 6.3-11: Spectrum with analysis of the 10Cr-anode in Figure 6.3-9
34
Figure 6.3-12: 11Cr-cathode Figure 6.3-13: 11Cr-cathode (100X)
Figure 6.3-14: 11Cr-cathode (1000X) Figure 6.3-15: 11Cr-anode (100X)
Figure 6.3-16: Spectrum with analysis of the marked 11Cr-cathode precipitation in Figure 6.3-13
35
Figure 6.3-17: Spectrum with analysis of the 11Cr-anode in Figure 6.3-14 In Figure 6.3-2 we see how the curve for the cell voltage versus plating time changes with the current density. Figures 6.3-4, -5, -12 record how the amount of precipitate on the copper cathode increases. The chromium content of the black powder (Figure 6.3-6, -8, -13, -14) is 25 at.% for 10Cr (Figure 6.3-10) and 20 at.% for 11Cr (Figure 6.3-16) combined with huge amounts of oxygen. The high oxygen content could be a result of the moisture in the rinsing liquids, the handling in air and/or the huge surface area of the powdery precipitate. The anodes (Figures 6.3-9, -15) show 60 to 64 at.% chromium at their surface with some oxygen (Figures 6.3-10, -14). 6.4 Adding potassium chloride This chapter describes the investigation of electrodeposition of chromium from electrolytes made from choline chloride, chromium-III chloride hexahydrate, and lithium chloride or potassium chloride. The intention was to create electrolytes in a molar ratio of 1:2:12:2.8 which results in an addition of 15 wt.% of lithium chloride and 26.38 wt.% of potassium chloride. The potassium chloride solution became saturated approximately at a concentration of 20 to 25 wt.%. All plating experiments in this chapter were conducted galvanostatically at -10 mA/cm2. The intent of this chapter is to compare the performance of the electrolyte with the addition of lithium chloride to that of the addition of potassium chloride. For the tests a molar ratio of 2 to 1 for chromium-III chloride hexahydrate to choline chloride was used. For every mol of chromium-III chloride hexahydrate, 2.8 mol of lithium chloride or 2.4 mol of potassium chloride was added for 11Cr and 12Cr, respectively. All the chromium plating experiments are listed in Attachment B. Table 6.4-1 lists the tests relevant for this chapter. Copper cathodes were used in all experiments presented
36
in this chapter. Figure 6.4-1 documents the cell voltage versus plating curves for the 7th test series and shows problems with the contact in test 12Cr. It is evident that the absolute cell voltage is much smaller with potassium chloride, the precipitate is increased (Figures 6.4-2, -3, -4), it contains more chromium (31 at.% see Figure 6.4-6) and the anodes (Figure 6.4-5) are cleaner with 76 at.% chromium (Figure 6.4-7). The results for 11Cr were shown in chapter 6.3. After these tests we decided to continue the experiments with entirely water free chromium chloride electrolytes. The solubility of chromium(III) chloride was tested in several ionic liquids, but only EMIm Cl showed promising results. The tests are described in chapter 7.
Figure 6.4-1: Cell voltage versus plating time for the 7th test series
Figure 6.4-2: 12Cr-cathode Figure 6.4-3: 12Cr-cathode (100X)
37
Table 6.4-1: Chromium plating experiments 11Cr and 12Cr Chromium Plating Experiment
Number 11 Cr 12 Cr
Electrolyte composition
CrCl3*6H2O : Cholin
chloride : LiCl = 2:1:
2.8
CrCl3*6H2O : Cholin
chloride : KCl = 2:1:2.5
Electrolyte history 8-10Cr Fresh
Plating temperature oC 60 60
Stirrer setting 1-10 5 5
Start weight cathode g 2.5046 1.9826
Start weight anode 1 g 1.8095 1.7892
Start weight anode 2 g 2.2849 2.2701
Plating time sec 7200 7200
Area of cathode cm2 2.858676 2.725959
Current mA -28.58676 -27.25959
Cathode material Copper Copper
Current density mA/cm2 -10.00 -10.00
Plating remark 1
Plating remark 2
End weight cathode g 2.5066 1.9894
End weight anode 1 g 1.7938 1.7576
End weight anode 2 g 2.275 2.235
Weight gain (cathode) g 0.0020 0.0068
Weight loss (anodes) g 0.0256 0.0667
How did the electrolyte appearance change? no change no change
How did the cathode look like?
black w/ copper
showing through
Green&black
How did the anode look like?no change / stained no change / stained
Cathodic current efficiency % 5.41 19.29
Anodic current efficiency % 69.24 189.18
Open circuit cell voltage V
Characteristic cell voltage level V range: -5 -> -9 -1.75
Analysis of cathode surface in atomic % crust nodules
% C 7.9 7.9
% N 3.3 2.5
% O 61.4 49.9
% Cl 2.6 7.7
% Fe 0.0 0.0
% Cu 0.1 1.2
%Cr 20.2 30.9
Analysis of cathode surface 2 in atomic % smooth smooth-some plating
% C 11.569 6.6
% N 3.828 1.9
% O 57.59 54.0
% Cl 1.522 10.7
% Fe 0 0.0
% Cu 19.004 6.6
%Cr 6.5 20.3
Analysis of cathode surface 3 in atomic % n/a smooth-little/no plating
% C n/a 9.9
% N n/a 1.4
% O n/a 41.3
% Cl n/a 7.0
% Fe n/a 0.0
% Cu n/a 31.7
%Cr n/a 8.7
Analysis of anode surface in atomic %
% C 1.7 0.5
% N 4.4 2.7
% O 30.0 20.4
% Cl 0.1 0.1
% Fe 0.0 0.0
% Cu 0.3 0.2
%Cr 63.6 76.1
38
Figure 6.4-4: 12Cr-cathode (1000X) Figure 6.4-5: 12Cr-anode (100X)
Figure 6.4-6: Spectrum with analysis of the 11Cr-cathode precipitation in Figure 6.4-4
Figure 6.4-7: Spectrum with analysis of the 11Cr-anode in Figure 6.4-5
39
7 Investigating electrolytes made with 1-ethyl-3-methylimidazolium chloride ([EMIm]Cl), anhydrous chromium chloride
Because it was decided to investigate electrolytes formulated without any water, we tried unsuccessfully to dissolve anhydrous chromium chloride in [MMPIm][TFSI] and choline chloride. [MMPIm][TFSI] showed virtually no solubility for chromium chloride. At 320 oC the choline chloride decomposed without dissolving chromium chloride, but we were able to create electrolytes with 10 wt.% chromium chloride in [EMIm]Cl. 7.1 Varying the current density This chapter describes the investigation of electrodeposition of chromium from electrolytes made from anhydrous chromium(III) chloride and 1-ethyl-3-methylimidazolium chloride ([EMIm]Cl) in a molar ratio of 1:9.7. Both plating experiments in this chapter were conducted galvanostatically at -10 mA/cm2 and -25 mA/cm2 respectively. We used copper cathodes rotated at 660 rpm and a plating temperature of 120 oC. The intent of this chapter is to compare the plating performance of this electrolyte at different current densities. All the chromium plating experiments are listed in Attachment B. Table 7.1-1 lists the two tests relevant for this chapter. Figure 7.1-1 shows the cell voltage versus plating time for the 8th test series. The curve for 13Cr reveals some contact problems. Curve 14Cr seems ideal. As expected, there are increases in the voltage with increased current density. At 25 mA/cm2 and 120 oC, the absolute cell voltage decreases from 1.8 V and levels out at around 1.45 V. The visual appearance of the 13Cr- and the 14Cr-cathode is given in Figures 7.1-2 and -8. The 13Cr-cathode is covered with a thin grey layer (Figures 7.1-3 and -4). The EDS analysis (Figure 7.1-4) gives only 10 at.% chromium and 42 at.% copper besides oxygen and carbon. The black powdery layer on the 14Cr-cathode (Figures 7.1-9 and -11) shows an even lower chromium content in the EDS analysis (Figure 7.1-10): 3 at.% chromium and 46 at.% copper besides oxygen and carbon. Because the cathodic current efficiency was so low (Table 7.1-1), it is assumed that corrosion is occurring besides the plating and/or that instead of plating occurring, soluble Cr(II)-salts were produced. To eliminate the effect of the substrate on the analysis we scrapped the black precipitate off the 14Cr-cathode (Figure 7.1-14) and analyzed the black powder without the copper substrate. The EDS-analysis (Figure 7.1-15) gave us 64 at.% copper and 3 at.% chromium besides oxygen and carbon. Both pairs of anodes (Figures 7.1-6, -7 and -12, -13) are nicely etched. Because the anodic current efficiency was too high (Table 7.1-1), it is assumed that corrosion is occurring along with the electrochemical oxidation and/or that soluble Cr(II)-salts were produced. In the 9th test series we investigated the effects of a lithium chloride addition and in the 10th test series a longer plating time.
40
Table 7.1-1: Chromium plating experiments 13Cr and 14Cr Number 13 Cr 14 Cr
Electrolyte compositionCrCl3 : EMImCl =
1:9.7 = molar ratio
CrCl3 : EMImCl =
1:9.7 = molar
ratio
Electrolyte history Fresh 13Cr
Plating temperature oC 120 120
Stirrer setting 1-10 5 5
Start weight cathode g 3.3775 2.0708
Start weight anode 1 g 2.2325 1.6937
Start weight anode 2 g 1.7545 2.1667
Plating time sec 7200 7200
Area of cathode cm2 3.3115 2.61534468
Current mA -33.115 -65.3836117
Cathode material Copper Copper
Current density mA/cm2 -10.00 -25.00
Plating remark 1Good looking
graph
Plating remark 2
End weight cathode g 3.3607 2.0718
End weight anode 1 g 2.1717 1.6219
End weight anode 2 g 1.6988 2.088
Weight gain (cathode) g -0.0168 0.0010
Weight loss (anodes) g 0.1165 0.1505
How did the electrolyte appearance change?no change purple to
brownish-black
How did the cathode look like?
copper with some
grey
powdery black
How did the anode look like?no change / stained very shiny, some
dark spots
Cathodic current efficiency % -39.23 1.18
Anodic current efficiency % 272.01 177.97
Open circuit cell voltage V
Characteristic cell voltage level V -0.8 range: -1.8 -> -1.47
Analysis of cathode surface in atomic %
% C 12.7 17.0
% N 0.9 0.0
% O 34.3 32.8
% Cl 0.2 1.0
% Fe 0.0 0.0
% Cu 42.5 46.2
%Cr 9.6 3.0
Analysis of cathode surface 2 in atomic % n/a n/a
% C n/a n/a
% N n/a n/a
% O n/a n/a
% Cl n/a n/a
% Fe n/a n/a
% Cu n/a n/a
%Cr n/a n/a
Analysis of cathode surface 3 in atomic % n/a n/a
% C n/a n/a
% N n/a n/a
% O n/a n/a
% Cl n/a n/a
% Fe n/a n/a
% Cu n/a n/a
%Cr n/a n/a
Analysis of anode surface in atomic %
% C 1.5 0.0
% N 2.3 4.4
% O 20.1 31.9
% Cl 0.2 0.2
% Fe 0.0 0.0
% Cu 0.7 0.4
%Cr 75.2 63.2
41
Figure 7.1-1: Cell voltage versus plating time for the 8th test series
Figure 7.1-2: Picture of electrodes after 13Cr-test Figure 7.1-3: SEM picture of the 13Cr-cathode (100X)
42
Figure 7.1-4: Spectrum with analysis of the 13Cr-cathode
Figure 7.1-5: 1000X SEM picture of the 13Cr cathode
Figure 7.1-6: 100X SEM picture of the 13Cr Figure 7.1-7: 1000X SEM picture of the 13Cr anode anode
43
Figure 7.1-8: Electrodes after the 14Cr-test Figure 7.1-9: 14Cr-cathode (100X)
Figure 7.1-10: Spectrum with analysis of the 14Cr-cathode
Figure 7.1-11: 14Cr-cathode (1000X) Figure 7.1-12: 14Cr-anode (100X)
44
Figure 7.1-13: 14Cr anode (1000X) Figure 7.1-14: Powder from 14Cr-cathode
Figure 7.1-15: Spectrum with analysis of the 14Cr-powder from the 14Cr-cathode (Figure 7.1-14)
45
7.2 Adding Lithium Chloride and Changing the Current Density This chapter describes the investigation of electrodeposition of chromium from electrolytes made from anhydrous chromium-III chloride, lithium chloride, and 1-ethyl-3-methylimidazolium chloride ([EMIm]Cl) in a molar ratio of 1:1:9.7. All plating experiments in this chapter were conducted galvanostatically at 120 oC with varying current densities and plating times. The intent of this chapter is to compare the plating performance of this electrolyte at different current densities and plating times. A molar ratio of 1 to 9.7 for anhydrous chromium-III chloride to [EMIm]Cl was used for the tests. For every mol of anhydrous chromium-III chloride, 1 mol of lithium chloride was added. All the chromium plating experiments are listed in Attachment B. Table 7.2-1 lists the tests relevant for this chapter. A copper cathode was used in 15Cr while nickel cathodes were used in 16-20Cr. The addition of lithium chloride to the 13-14Cr-electrolyte resulted in a higher absolute cell voltage (Figure 7.2-1) and no plating (Figures 7.2-5 and -6), meaning no chromium (Figure 7.2-7) at the cathode. The anodes were nicely etched (Figures 7.2-5, -8, -9, -10). Because of the possibility that we created soluble Cr(II)-compounds, in test 16/17Cr we plated for 17.6 hours at 8.53 mA/cm2 onto a nickel rod hanging on a platinum wire and moved in a circle at 300 rpm by a magnetic field. The plating documented in Figure 7.2-2 was interrupted after 800 seconds for a few minutes and had contact trouble after 42,000 seconds. The electrolyte turned from violet to dark brown during the plating. The electrodes are shown in Figure 7.2-11, the resulting black plating at the cathode in Figures 7.2-12, -14, -15 and the analysis with 23 and 32 at.% chromium in Figure 7.2-13 and -16 . The appearance of the 16/17Cr-anodes is shown in Figures 7.2-11, -17, -18, and the analysis in Figure 7.2-19. We ran test 18/19Cr like 16/17Cr, but with lower absolute current density (5 mA/cm2). As recorded in Figure 7.2-3, we had similar contact problems as with 16/17Cr. The cell absolute voltage was higher than in the previous test with a higher current density. Most of the black precipitate fell off the cathode (Figures 7.2-20, -21, -23, -24) The analysis of the surface showed up to 15 at.% chromium (Figures 7.2-22, -25). In test 20Cr, we ran the highest absolute current density of this test series, 13.15 mA/cm2. To avoid problems with electrical resistance we attached the platinum wires of anodes and the copper wire of the cathode with silver epoxy to the electrodes and protected this area with PTFE foil to avoid contact with the electrolyte. Figure 7.2-4 shows again an unstable cell voltage. The plating at the cathode was a thin, black, non-adhering foil shown in Figures 7.2-26 and -27, consisting of mostly carbon which points to electrolyte destruction. The anodes turned dark during plating. We found 63 at.% chromium at their surface (Figure 7.2-31). More research is required to make more conclusive statements. One test is not sufficient to draw decent conclusions.
46
Table 7.2-1: Chromium plating experiments 15Cr till 20Cr Number 15 Cr 16+17 Cr 18+19 Cr 20 Cr
Electrolyte compositionCrCl3 : LiCl : EMImCl =
1:1:9.7 = molar ratio
CrCl3 : LiCl : EMImCl =
1:1:9.7 = molar ratio
CrCl3 : LiCl : EMImCl =
1:1:9.7 = molar ratio
CrCl3 : LiCl : EMImCl =
1:1:9.7 = molar ratio
Electrolyte history Fresh New New New
Plating temperature oC 120 120 100-117 125-137
Stirrer setting 1-10 5 magn. stir. 300 rpm magn. stir. 300 rpm magn. stir. 300 rpm
Start weight cathode g 2.3734 17.566 17.5555 17.8508
Start weight anode 1 g 1.6187 1.5408 1.221 1.1745
Start weight anode 2 g 2.0855 2.0351 1.5708 1.1887
Plating time sec 7200 63403 121000 57900
Area of cathode cm2 2.57459801 6.7273 6.7273 6.14
Current mA -25.7459801 -57.41 -34 -80.72
Cathode material Copper Nickel rod as cathode Nickel rod as cathode
Current density mA/cm2 -10.00 -8.53 -5.05 -13.15
Plating remark 1
Contact problems
during the first
minutes and the last
hours. Electrolyte
turned dark brown.
Run with lots of
problems. Anode fell
off the wire. Lsot
contacts. Electrolyte
turned brown.
Contacts with silver
epoxy secured and
area covered with
PTFE foil.
Plating remark 2
Potentiostat shut
down before the end
of the test.
Only app. One third
of the time the run
was normal. Plating
was mostly rinsed off
the cathode.
Cell voltage jumped
from 5 to 16 V - back
and forth - after 6400
seconds.
End weight cathode g 2.3699 17.5824 17.5567 17.7976
End weight anode 1 g 1.5746 1.2214 1.0551 1.1695
End weight anode 2 g 2.0438 1.5784 1.1737 1.1835
Weight gain (cathode) g -0.0035 0.0164 0.0012 -0.0532
Weight loss (anodes) g 0.0858 0.7761 0.5630 0.0102
How did the electrolyte appearance change?slightly darker Turned from pink to
dark brown.
Turned from pink to
dark brown.
Turned from pink to
dark brown.
How did the cathode look like?
copper w/
discoloration
Ni-rod turned black Ni-rod turned black
but lost crust during
rinsing.
Ni-rod turned black
but lost some foil
during rinsing.
How did the anode look like?darker, etched etched darker, etched dark
Cathodic current efficiency % -10.51 2.51 0.16 -6.34
Anodic current efficiency % 257.67 118.69 76.18 1.21
Open circuit cell voltage V -0.225 -0.235 0.3416
Characteristic cell voltage level V -1.6 -1.50 -1.5 till 2.5 -5
Analysis of cathode surface in atomic %
% C 12.8 32.5 43.3 94.2
% N 0.0 1.7 0.0 0.0
% O 6.4 41.4 25.9 0.2
% Cl 0.0 0.7 1.2 5.2
% Fe 0.0 - 20.8 at.% Ni 0.2 at.% Ni
% Cu 80.3 0.6 at.% Ni 1.3 0.0
%Cr 0.4 23.0 7.5 0.1
Analysis of cathode surface 2 in atomic % n/a n/a n/a n/a
% C n/a n/a n/a n/a
% N n/a n/a n/a n/a
% O n/a n/a n/a n/a
% Cl n/a n/a n/a n/a
% Fe n/a n/a n/a n/a
% Cu n/a n/a n/a n/a
%Cr n/a n/a n/a n/a
Analysis of cathode surface 3 in atomic % n/a n/a n/a n/a
% C n/a n/a n/a n/a
% N n/a n/a n/a n/a
% O n/a n/a n/a n/a
% Cl n/a n/a n/a n/a
% Fe n/a n/a n/a n/a
% Cu n/a n/a n/a n/a
%Cr n/a n/a n/a n/a
Analysis of anode surface in atomic %
% C 0.2 0.2 4.8
% N 0.1 3.5 2.5
% O 18.9 31.9 27.0
% Cl 0.2 0.2 0.2
% Fe 0.0 0.2 at.% Ni 0.3 at.% Ni
% Cu 0.2 0.3 0.6
%Cr 75.8 63.6 64.6
47
Figure 7.2-1: Cell voltage versus plating time for the 9th test series
Figure 7.2-2: Cell voltage versus plating time for test 16/17Cr
48
Figure 7.2-3: Cell voltage versus plating time for test 18/19Cr
Figure 7.2-4: Cell voltage versus plating time for test 20Cr
49
Figure 7.2-5: Electrodes after test 15Cr Figure 7.2-6: 15Cr-cathode (100X)
Figure 7.2-7: Spectrum with analysis of the 15Cr-cathode
Figure 7.2-8: 15 Cr-anode (100X) Figure 7.2-9: 15Cr-anode (1000X)
50
Figure 7.2-10: Spectrum with analysis of the 15Cr-anode
Figure 7.2-11: 16/17Cr-electrodes Figure 7.2-12: 16/17 cathode (100X SEM picture)
Figure 7.2-13: Spectrum with analysis of the 16/17Cr-cathode
51
Figure 7.2-14: 16-17Cr cathode (1000X) Figure 7.2-15: 16-17Cr cathode (1000X)
Figure 7.2-16: Spectrum with analysis of the cathode area shown in Figure 7.2-15
Figure 7.2-17: 16/17Cr-anode (100X) Figure 7.2-18: 16/17Cr-anode (1000X)
52
Figure 7.2-19: Spectrum with analysis of the cathode area shown in Figure 7.2-17
Figure 7.2-20: Cathode after 18/19Cr-test Figure 7.2-21: 18/19Cr-cathode (100X)
Figure 7.2-22: Spectrum with analysis of the 18/19Cr-cathode
53
Figure 7.2-2-23: 18/19Cr-cathode (1000X) Figure 7.2-2-24: 18/19Cr-cathode (1000X)
Figure 7.2-25: Spectrum with analysis of black crust in the marked area in Figure 7.2-24
Figure 7.2-26: 20Cr-electrodes Figure 7.2-27: 20Cr-cathode (35X)
54
Figure 7.2-28: Spectrum with analysis of the 20Cr-cathode
Figure 7.2-29: 20Cr-anode (100X) Figure 7.2-30: 20Cr-anode (1000X)
Figure 7.2-31: Spectrum with analysis of the 20Cr-anode (Figure 7.2-29)
55
8 Summary, conclusions and recommendations for future research The investigations regarding chromium plating from electrolytes containing chromium(III) chloride in ionic liquid were done in a technical setup with chromium anodes, a good bath movement and a variety of materials for cathodes. In the literature we found that the most promising electrolyte for hard chromium is a proprietary mixture of chromium(III) chloride, choline chloride and water. We started our investigations with chromium(III) chloride hexahydrate and choline chloride mixed in a molar ration of 3.1 to 1. The black precipitate contained more chromium (35 at.%) at higher absolute current densities (140 mA/cm2). In the layer was a high amount of oxygen (around 58 at.%). The absolute cell voltage was extremely high (16 V), even if the absolute current density was only 10 mA/cm2. At very low absolute current densities (0.35 mA/cm2) the plating was overpowered by the corrosion occurring at the electrodes. The anodes dissolved during the plating process. Adding more water to this kind of electrolyte lowered the cell voltage. Also here we saw that a higher current density gives us a higher chromium content in the black precipitate, but also leads to more gas evolution during plating. Future research could investigate the effect of the addition of more water and/or higher current densities with these electrolytes. It was decided to explore the effect of the addition of lithium and potassium chloride on the process. We found that both additions improved the process but potassium chloride additions are more promising than lithium chloride additions due to the superior absolute cell voltage drops and chromium content of the plated layer . Future research should investigate how far the chromium content can be increased in the plated layer with electrolytes saturated with potassium chloride and possibly with less than six water molecules per chromium ion. It was observed that water condensed in the cold parts of the plating cell when mixtures of choline chloride and chromium(III) hexahydrate were heated up. To create an electrolyte with less water we can pull a vacuum on the electrolyte and remove some of the water or we try to add anhydrous chromium(III) chloride. Water in the electrolyte increases the possibility of introducing hydrogen embrittlement into the substrate during plating. We decided to investigate water-free electrolytes from [EMIm] Cl and 10 wt.% anhydrous chromium(III) chloride. We saw very little or no plating during our first tests with a high copper content in the layer produced at the copper cathode. The chromium concentration was below 10 at.% on the cathode. Long term plating experiments with nickel cathodes resulted in high absolute cell voltages at some point of time during plating, which was first attributed to bad contacts, but the last test showed a thin carbon rich film on the cathode pointing to electrolyte destruction. More research in this area would focus on electrolytes with other ionic liquids and additions like potassium chloride.
56
Attachment A: References [1] A. Abbott, G. Capper, D. Davies, R. Rasheed (2004) Chem. Eur. J. 10, 3769. [2] A. Abbott, G. Capper, D. Davies, R. Rasheed, J. Archer, and C. John, (2004)
Trans. Inst. Met. Finish. 82, 14. [3] Tabereaux, A. T. (1994) JOM, 42, 30-34. [4] E. L. Smith, C. Fullarton, R. Harris, S. Saleem, A. P. Abbott, K. S. Ryder: Metal
finishing with ionic liquids: scale-up and pilot plants from IONMET consortium (2010) Institute of Metal Finishing.
[5] M. Schlesinger, M. Paunovic: Modern electroplating, 5th edition (Chapter 7: Electrodeposition of chromium) (2010) John Wiley & Sons, Inc.
[6] F. Endres, D. MacFarlane, A. Abbott: Electrodeposition from ionic liquids. (2008) Wiley-VCH Verlag GmbH & Co. KGaA, ISBN: 978-3-527-31565-9.
[7] IONMET – New ionic liquid solvent technology to transform metal finishing – products and processes Chapter 3: P. Benaben : Hard and decorative coating http://www.ionmet.eu/fileadmin/ionmet/Dissemination/IONMET_brochure_200711.pdf
[8] S. Eugénio, C. Rangel, R. Vilar: Departamento de Engenharia de Materiais, Instituto Superio Técnico, Universidade Técnica de Lisboa, Portugal
http://repositorio.lneg.pt/bitstream/10400.9/529/1/ELECTRODEPOSITSONIA.pdf [9] S. Eugénio, C. Rangel, R. Vilar, S. Quaresma: Electrochemical aspects of black
chromium electrodeposition from 1-butyl-3-methylimidazolium tetrafluoroborate ionic liquid (2011) Electrochimica Acta 56 10347-10352
[10] S. Eugénio, C. Rangel, R. Vilar, A. M. Botelho do Rego: Electrodeposition of black chromium spectrally selective coatings from a Cr(III)-ionic liquid solution (2011) Thin solid Films 519 1845-1850
[11] M. Ali, A. Nishikata, T. Tsuru: Electrochimica Acta 42 (1997) 2347
57
Attachment B: Cr-plating tests Chromium Plating Experiment
Number 1 Cr 2 Cr 3 Cr 4 Cr 5 Cr 6 Cr
Electrolyte compositionCrCl3*6H2O : Cholin
chloride = 3.10:1
CrCl3*6H2O : Cholin
chloride = 3.10:1
CrCl3*7.5H2O : Cholin
chloride = 3.10:1
CrCl3*9H2O : Cholin
chloride = 3.10:1
CrCl3*9H2O : Cholin
chloride = 3.10:1
CrCl3*9H2O : Cholin
chloride = 2:1
Electrolyte history Newly prepared 1Cr 1-2Cr and water 1-3Cr and water 1-4Cr New
Plating temperature oC 58-70oC (no cooling) 72-46 48.0 50 50 50
Stirrer setting 1-10 5 = 645-675 rpm 5 = 634-658 rpm 5 5 5 5
Start weight cathode g 2.8111 2.0039 1.6053 1.4488 1.567 1.458
Start weight anode 1 g 2.1837 2.1450 1.9622 1.9291 1.8279 2.2828
Start weight anode 2 g 2.6460 2.6134 2.397 2.3754 2.2654 1.9011
Plating time sec 1200 12000 12000 12000 60300 9000
Area of cathode cm2 3.15568 2.67257 2.6246436 2.506991 2.5533043 2.524112
Current mA -441.66 -27.6343738 -39.37 -37.6048641 -38.29956445 -37.872168
Cathode material Nickel Type 420 steel (13%Cr) Type 420 steel (13%Cr) Type 420 steel (13%Cr) Type 420 steel (13%Cr) Type 420 steel (13%Cr)
Current density mA/cm2 -139.96 -10.34 -15.00 -15.00 -15.00 -15.00
Plating remark 1
At the beginning the
cell voltage was the
Solartron maximum
and the current too
low.
The cell voltage was
the Solartron
maximum and the
current too low.
Magnetic stirrer in the
electrolyte caused
scratches on cathode.
Problems with the
electrical connection.
Bad connection
between rod and
cathode?
Plating remark 2
The wanted current
was -473 mA. The
average current was
only -442 mA.
The wanted current
was-40.1 mA. The
average current was
only -27.6 mA.
Cathode was looking
metallic grey but was
left for 65 hours in the
electrolyte. See 6Cr
remark.
Cathode was covered
with black powdery
crust which
disappeared after 65
hours in the
electrolyte.
End weight cathode g 2.8239 2.0270 1.6561 1.444 1.7256 1.4524
End weight anode 1 g 2.1498 2.1255 1.9298 1.8488 1.712 2.2524
End weight anode 2 g 2.6174 2.5969 2.3759 2.329 2.1008 1.8743
Weight gain (cathode) g 0.0128 0.0231 0.0508 -0.0048 0.1586 -0.0056
Weight loss (anodes) g 0.0625 0.0360 0.0535 0.1267 0.2805 0.0572
How did the electrolyte appearance change? no change no change no change no change no change no change
How did the cathode look like?Covered with black
powder
Unevenly covered
with black powderDark. Black. Powdery
Looked metallic grey
after 65h in
electrolyte.
Unevenly covered
with black powder
Covered with
powdery black crust.
Looked metallic grey
after 65h in
How did the anode look like? Partially etched Partially etched Etched. Etched. Etched. Etched.
Cathodic current efficiency % 13.44 38.78 59.86 -5.92 38.23 -9.15
Anodic current efficiency % 65.65 60.43 63.04 156.30 67.61 93.42
Open circuit cell voltage V
Characteristic cell voltage level V -15.6 -16 -5.3 -3.3 -3.5 -3.6
Analysis of cathode surface in atomic % smooth, shiny outer crust smooth
% C 2.734 14.5 12.8 15.1 19.7 20.6
% N 2.687 3.2 3.2 0.1 4.2 0.0
% O 54.925 67.8 64.8 5.7 50.8 23.6
% Cl 4.504 1.0 1.6 0.1 3.5 0.1
% Fe 0.142 *Ni not Fe 0.2 0.2 62.9 - 38.5
% Cu 0 0.0 0.0 0.0 0.0 0.0
%Cr 35.0 13.4 17.4 16.1 21.8 17.2
Analysis of cathode surface 2 in atomic % n/a n/a n/a black spot inner crust smooth
% C n/a n/a n/a 9.171 16.785 15.6
% N n/a n/a n/a 2.227 3.556 0.0
% O n/a n/a n/a 61.387 55.191 22.8
% Cl n/a n/a n/a 3.291 2.07 0.1
% Fe n/a n/a n/a 0.438 - 43.1
% Cu 0 0.0 0.0 0.0 0.0 0.0
%Cr n/a n/a n/a 23.5 22.4 18.4
Analysis of cathode surface 3 in atomic % n/a n/a n/a n/a n/a crust
% C n/a n/a n/a n/a n/a 4.6
% N n/a n/a n/a n/a n/a 3.5
% O n/a n/a n/a n/a n/a 47.7
% Cl n/a n/a n/a n/a n/a 3.1
% Fe n/a n/a n/a n/a n/a 0.5
% Cu 0 0.0 0.0 0.0 0.0 0.0
%Cr n/a n/a n/a n/a n/a 40.5
Analysis of anode surface in atomic %
% C 5.7 2.3 2.7 0.0 4.8 11.9
% N 2.2 2.1 3.9 1.8 1.9 5.5
% O 27.0 26.8 30.3 20.1 20.0 39.5
% Cl 0.2 0.2 0.1 0.2 0.1 0.3
% Fe 0.243 *Ni not Fe 0.0 0.3 0.1 - 0.2
% Cu 0 0.0 0.0 0.0 0.0 0.0
%Cr 64.7 68.7 62.7 77.8 73.2 42.6
58
Chromium Plating Experiment
Number 7 Cr 8 Cr 9 Cr 10 Cr 11 Cr 12 Cr
Electrolyte compositionCrCl3*9H2O : Cholin
chloride = 2:1
CrCl3*6H2O :
Cholin chloride =
2:1
CrCl3*6H2O :
Cholin chloride =
2:1 and 15 w.% LiCl
CrCl3*6H2O : Cholin
chloride = 2:1 and
15 w.% LiCl
CrCl3*6H2O : Cholin
chloride : LiCl = 2:1:
2.8
CrCl3*6H2O : Cholin
chloride : KCl = 2:1:2.5
Electrolyte history 6 Cr New 8 Cr 8-9Cr 8-10Cr Fresh
Plating temperature oC 40 60 60 60 60 60
Stirrer setting 1-10 5 5 5 5 5 5
Start weight cathode g 2.4684 2.2122 2.0055 2.1978 2.5046 1.9826
Start weight anode 1 g 2.2513 2.2289 2.1918 2.3143 1.8095 1.7892
Start weight anode 2 g 1.8728 1.8544 1.8353 1.8306 2.2849 2.2701
Plating time sec 1800 7200 7200 7200 7200 7200
Area of cathode cm2 2.8043427 2.594484293 2.726616538 2.549213943 2.858676 2.725959
Current mA -168.26056 -0.8950978 -0.940682706 -8.794788103 -28.58676 -27.25959
Cathode material Copper Copper Copper Copper Copper Copper
Current density mA/cm2 -60.00 -0.35 -0.35 -3.45 -10.00 -10.00
Plating remark 1 lots of gas evolutionpositive voltage
was observed
large amounts of
carbon from
brushes present
Plating remark 2
End weight cathode g 2.4763 2.212 2.0054 2.1988 2.5066 1.9894
End weight anode 1 g 2.2299 2.2231 2.1901 2.3097 1.7938 1.7576
End weight anode 2 g 1.8551 1.8499 1.8342 1.827 2.275 2.235
Weight gain (cathode) g 0.0079 -0.0002 -0.0001 0.0010 0.0020 0.0068
Weight loss (anodes) g 0.0391 0.0103 0.0028 0.0082 0.0256 0.0667
How did the electrolyte appearance change? no change no change no change no change no change no change
How did the cathode look like? Dark. Black. Powdery no plating no plating Black. Powdery
black w/ copper
showing through
Green&black
How did the anode look like? Etched. no change no change Partially etchedno change / stained no change / stained
Cathodic current efficiency % 14.52 -17.28 -8.22 8.79 5.41 19.29
Anodic current efficiency % 71.87 889.71 230.14 72.09 69.24 189.18
Open circuit cell voltage V
Characteristic cell voltage level V -7.3 ±0.03 -0.08 -4.1 range: -5 -> -9 -1.75
Analysis of cathode surface in atomic % crust nodules
% C 8.8 34.6 27.7 10.2 7.9 7.9
% N 2.3 0.0 0.0 3.1 3.3 2.5
% O 49.7 3.2 2.9 56.8 61.4 49.9
% Cl 5.0 0.1 0.2 1.1 2.6 7.7
% Fe 0.0 0.0 0.0 0.0 0.0 0.0
% Cu 0.5 61.8 68.8 3.3 0.1 1.2
%Cr 33.7 0.2 0.4 25.4 20.2 30.9
Analysis of cathode surface 2 in atomic % n/a n/a n/a n/a smooth smooth-some plating
% C n/a n/a n/a n/a 11.569 6.6
% N n/a n/a n/a n/a 3.828 1.9
% O n/a n/a n/a n/a 57.59 54.0
% Cl n/a n/a n/a n/a 1.522 10.7
% Fe n/a n/a n/a n/a 0 0.0
% Cu n/a n/a n/a n/a 19.004 6.6
%Cr n/a n/a n/a n/a 6.5 20.3
Analysis of cathode surface 3 in atomic % n/a n/a n/a n/a n/a smooth-little/no plating
% C n/a n/a n/a n/a n/a 9.9
% N n/a n/a n/a n/a n/a 1.4
% O n/a n/a n/a n/a n/a 41.3
% Cl n/a n/a n/a n/a n/a 7.0
% Fe n/a n/a n/a n/a n/a 0.0
% Cu n/a n/a n/a n/a n/a 31.7
%Cr n/a n/a n/a n/a n/a 8.7
Analysis of anode surface in atomic %
% C n/a 2.851 n/a 3.2 1.7 0.5
% N n/a 1.836 n/a 3.3 4.4 2.7
% O n/a 24.885 n/a 33.0 30.0 20.4
% Cl n/a 0.07 n/a 0.4 0.1 0.1
% Fe n/a 0 n/a 0.0 0.0 0.0
% Cu n/a 0.233 n/a 0.2 0.3 0.2
%Cr n/a 70.1 n/a 59.9 63.6 76.1
59
Chromium Plating Experiment
Number 13 Cr 14 Cr 15 Cr 16+17 Cr 18+19 Cr 20 Cr
Electrolyte compositionCrCl3 : EMImCl =
1:9.7 = molar ratio
CrCl3 : EMImCl =
1:9.7 = molar
ratio
CrCl3 : LiCl : EMImCl =
1:1:9.7 = molar ratio
CrCl3 : LiCl : EMImCl =
1:1:9.7 = molar ratio
CrCl3 : LiCl : EMImCl =
1:1:9.7 = molar ratio
CrCl3 : LiCl : EMImCl =
1:1:9.7 = molar ratio
Electrolyte history Fresh 13Cr Fresh New New New
Plating temperature oC 120 120 120 120 100-117 125-137
Stirrer setting 1-10 5 5 5 magn. stir. 300 rpm magn. stir. 300 rpm magn. stir. 300 rpm
Start weight cathode g 3.3775 2.0708 2.3734 17.566 17.5555 17.8508
Start weight anode 1 g 2.2325 1.6937 1.6187 1.5408 1.221 1.1745
Start weight anode 2 g 1.7545 2.1667 2.0855 2.0351 1.5708 1.1887
Plating time sec 7200 7200 7200 63403 121000 57900
Area of cathode cm2 3.3115 2.61534468 2.57459801 6.7273 6.7273 6.14
Current mA -33.115 -65.3836117 -25.7459801 -57.41 -34 -80.72
Cathode material Copper Copper Copper Nickel rod as cathode Nickel rod as cathode Nickel rod as cathode
Current density mA/cm2 -10.00 -25.00 -10.00 -8.53 -5.05 -13.15
Plating remark 1Good looking
graph
Contact problems
during the first
minutes and the last
hours. Electrolyte
turned dark brown.
Run with lots of
problems. Anode fell
off the wire. Lsot
contacts. Electrolyte
turned brown.
Contacts with silver
epoxy secured and
area covered with
PTFE foil.
Plating remark 2
Potentiostat shut
down before the end
of the test.
Only app. One third
of the time the run
was normal. Plating
was mostly rinsed off
the cathode.
Cell voltage jumped
from 5 to 16 V - back
and forth - after 6400
seconds.
End weight cathode g 3.3607 2.0718 2.3699 17.5824 17.5567 17.7976
End weight anode 1 g 2.1717 1.6219 1.5746 1.2214 1.0551 1.1695
End weight anode 2 g 1.6988 2.088 2.0438 1.5784 1.1737 1.1835
Weight gain (cathode) g -0.0168 0.0010 -0.0035 0.0164 0.0012 -0.0532
Weight loss (anodes) g 0.1165 0.1505 0.0858 0.7761 0.5630 0.0102
How did the electrolyte appearance change?no change purple to
brownish-black
slightly darker Turned from pink to
dark brown.
Turned from pink to
dark brown.
Turned from pink to
dark brown.
How did the cathode look like?
copper with some
grey
powdery black copper w/
discoloration
Ni-rod turned black Ni-rod turned black
but lost crust during
rinsing.
Ni-rod turned black
but lost some foil
during rinsing.
How did the anode look like?no change / stained very shiny, some
dark spots
darker, etched etched darker, etched dark
Cathodic current efficiency % -39.23 1.18 -10.51 2.51 0.16 -6.34
Anodic current efficiency % 272.01 177.97 257.67 118.69 76.18 1.21
Open circuit cell voltage V -0.225 -0.235 0.3416
Characteristic cell voltage level V -0.8 range: -1.8 -> -1.47 -1.6 -1.50 -1.5 till 2.5 -5
Analysis of cathode surface in atomic %
% C 12.7 17.0 12.8 32.5 43.3 94.2
% N 0.9 0.0 0.0 1.7 0.0 0.0
% O 34.3 32.8 6.4 41.4 25.9 0.2
% Cl 0.2 1.0 0.0 0.7 1.2 5.2
% Fe 0.0 0.0 0.0 - 20.8 at.% Ni 0.2 at.% Ni
% Cu 42.5 46.2 80.3 0.6 at.% Ni 1.3 0.0
%Cr 9.6 3.0 0.4 23.0 7.5 0.1
Analysis of cathode surface 2 in atomic % n/a n/a n/a n/a n/a n/a
% C n/a n/a n/a n/a n/a n/a
% N n/a n/a n/a n/a n/a n/a
% O n/a n/a n/a n/a n/a n/a
% Cl n/a n/a n/a n/a n/a n/a
% Fe n/a n/a n/a n/a n/a n/a
% Cu n/a n/a n/a n/a n/a n/a
%Cr n/a n/a n/a n/a n/a n/a
Analysis of cathode surface 3 in atomic % n/a n/a n/a n/a n/a n/a
% C n/a n/a n/a n/a n/a n/a
% N n/a n/a n/a n/a n/a n/a
% O n/a n/a n/a n/a n/a n/a
% Cl n/a n/a n/a n/a n/a n/a
% Fe n/a n/a n/a n/a n/a n/a
% Cu n/a n/a n/a n/a n/a n/a
%Cr n/a n/a n/a n/a n/a n/a
Analysis of anode surface in atomic %
% C 1.5 0.0 0.2 0.2 4.8
% N 2.3 4.4 0.1 3.5 2.5
% O 20.1 31.9 18.9 31.9 27.0
% Cl 0.2 0.2 0.2 0.2 0.2
% Fe 0.0 0.0 0.0 0.2 at.% Ni 0.3 at.% Ni
% Cu 0.7 0.4 0.2 0.3 0.6
%Cr 75.2 63.2 75.8 63.6 64.6