tissue equivalent survey meters - jrt associatesjrtassociates.com/pdfs/microrem.pdf · the...

3
The tissue-equivalent scintillator used in the Micro Rem and Micro Sievert gives them a nearly flat, rem energy response. This rem response is based on the deep dose equivalent index for 1 cm (0.39”) depth, uniparallel directional beam as calculated on the ICRU standard sphere. These instruments give tissue- equivalent photon response for x-ray and gamma radiation from environmental levels of 0-20 μrem/h (o-0.2 μ Sv/h) full scale up to normal survey levels of 200 mrem/h (2 mSv/h) full scale. Rugged construction and quality components make the Micro Rem and Micro Sievert durable, and they are easy to service. Internal components are laid out on modular circuit boards. Span, HV and calibration pots (one for each range) are clearly marked. The HV check assures you that the high voltage required for proper operation of the detector is supplied. This is important for accurate readings. Two 9-volt batteries power the instrument. You can check their condition with a control switch setting. Applications - Routine low/medium level gamma dose rate surveys - Confirming radiation boundaries - Monitoring items for unrestricted release - X-ray surveys, using the expanded low energy range option Options You can “customize” your instrument with practical options to match the requirements of your survey situation. Options can be combined for maximum versatility. The expanded low energy response option extends the instruments low energy cutoff to 17 keV (vs 40 keV for the standard instruments). The extended detector option mounts the internal detector so that its sensitive area extends out from the front of instrument case bottom. This makes it easier for you to survey certain hard-to-reach locations. Product Specifications Micro Rem / Micro Sievert Tissue equivalent survey meters The Micro Rem and Micro Sievert models are lightweight, portable survey meters for applications where accurate dose rate measurements of low radiation levels are required. Accurate down to background levels, the Micro Rem and Micro Sievert read absorbed dose rate directly so no conversion from mR/h is required. - Flat energy response - Tissue equivalent scintillator - Gamma and x-ray detection - Optional audio counts and alarms - Extended detector option - Expanded low-energy response option - HV/ battery check

Upload: tranhuong

Post on 10-Feb-2018

221 views

Category:

Documents


3 download

TRANSCRIPT

Page 1: Tissue equivalent survey meters - JRT Associatesjrtassociates.com/pdfs/microrem.pdf · The tissue-equivalent scintillator used in the Micro Rem and Micro Sievert gives them a nearly

The tissue-equivalent scintillator

used in the Micro Rem and Micro

Sievert gives them a nearly flat, rem

energy response. This rem response

is based on the deep dose

equivalent index for 1 cm (0.39”)

depth, uniparallel directional beam

as calculated on the ICRU standard

sphere.

These instruments give tissue-

equivalent photon response for x-ray

and gamma radiation from

environmental levels of 0-20 μrem/h

(o-0.2 μ Sv/h) full scale up to normal

survey levels of 200 mrem/h

(2 mSv/h) full scale.

Rugged construction and quality

components make the Micro Rem

and Micro Sievert durable, and they

are easy to service. Internal

components are laid out on modular

circuit boards. Span, HV and calibration

pots (one for each range) are clearly

marked.

The HV check assures you that the high

voltage required for proper operation of the

detector is supplied. This is important for

accurate readings.

Two 9-volt batteries power the

instrument. You can check their condition

with a control switch setting.

Applications- Routine low/medium level gamma

dose rate surveys

- Confirming radiation boundaries

- Monitoring items for unrestricted release

- X-ray surveys, using the expanded low

energy range option

OptionsYou can “customize” your instrument with

practical options to match the requirements

of your survey situation. Options can be

combined for maximum versatility.

The expanded low energy response

option extends the instruments low energy

cutoff to 17 keV (vs 40 keV for the standard

instruments).

The extended detector option mounts the

internal detector so that its sensitive area

extends out from the front of instrument

case bottom. This makes it easier for you to

survey certain hard-to-reach locations.

Product Specifications

Micro Rem / Micro SievertTissue equivalent survey meters

The Micro Rem and Micro Sievertmodels are lightweight, portable survey meters for applications whereaccurate dose rate measurements oflow radiation levels are required.Accurate down to background levels,the Micro Rem and Micro Sievertread absorbed dose rate directly sono conversion from mR/h is required.

- Flat energy response

- Tissue equivalent scintillator

- Gamma and x-ray detection

- Optional audio counts and alarms

- Extended detector option

- Expanded low-energy response option

- HV/ battery check

Janet
Text Box
This product is available through: JRT Associates 5 Nepperhan Avenue, Suite 2B Elmsford, NY 10523 800-221-0111
Page 2: Tissue equivalent survey meters - JRT Associatesjrtassociates.com/pdfs/microrem.pdf · The tissue-equivalent scintillator used in the Micro Rem and Micro Sievert gives them a nearly

Complete Flat Energy Response

The rem response and accuracy set the Micro Rem and Micro Sievert apart from conventional "micro R" meters which use Nal(TI)

detectors. Nal(TI) detectors over-respond to low energies and produce erroneously high readings. Also, with with conventional “micro R”

meters the user must to convert μR/h readings to μrem/h to get absorbed dose.

Micro Rem / Micro Sievert Specifications

Power Reactors: Because of its sensitivity and accurate readings, you can use the instrument for setting the boundaries of specific dose rate areas. The over-response of micro R meters often leads to unrealistic boundaries (through buildings, past a fenceline, etc.).

X ray Device Monitoring: A MICRO REM or MICRO SIEVERT, with its expanded low energy response option, meets or exceeds the instrumentation requirements of 21 CFR1 020.10 and 21 CFR1 020.40. You can use oneto survey color TV monitors and all cabinet X-ray systems.

Uranium Processing/Mining Cleanup Operations: The tissue equivalency (rem response) of a MICRO REM or MICRO SIEVERT produces much

more accurate environmental measurements than do micro R meters.Additional Applications: The instrument has many uses for environmental measurement companies, consultants,

geophysical companies, laboratories and accelerators.Radiation Detected: Gamma and x-ray, 40 keV to 1.3 MeVDetector: Internal, tissue equivalent, organic scintillator.Range: Five linear ranges:

MICRO REM: MICRO SIEVERT:Range μrem/h μSv/hX0.1 0 to 20 0 to 0.2X1 0 to 200 0 to 2X10 0 to 2000 0 to 20X100 0 to 20,000 0 to 200x1000 0 to 200,000 0 to 2000

Accuracy: Within 10% of reading for 137Cs between 20% and 100% of full scale on any range.Energy Response: See energy response curve.High Voltage: Electronically stabilized, factory set during calibration, with check band on the meter.Warmup Time: None.Response Time: Optimized for each range, 0 to 90% of final reading as follows:

Range: Time:X0.1 <15 secondsX1 <15 secondsX10 <5 secondsX100 <2 secondsX1000 <2 seconds

Temperature: Operational from -20 °C to 50 °C (-4 °F to 122 °F).Humidity: <5% change in reading from 10% to 95% RH.Battery Complement: Two 9-volt, MN1604 or equal.Battery Life: >100 Hours.Geotropism: Within + 2% of full scale.Shock: 100 g per lightweight machine of MIL-STD 202 C, method 202 B.Vibration: 5 g in each of three mutually orthogonal axes at one or more frequencies from 10 to 33 Hz.Meter Reset Button: Allows meter to reset to “0” for repeat measurement.Display: Ruggedized, recessed, high-torque 1 mA meter protected by impact resistant Lexan

polycarbonate window with 85.1 mm (3.35”) scale marked:MICRO REM: 0-200 μSv/h, with ‘Bat. ok’, ‘HV ok’ check bands.MICRO SIEVERT: 0-2.0 μSv/h, with ‘Bat. ok’, ‘HV ok’ check bands.

Control: Eight position rotary switch: Off; Bat. (checks the condition of the batteries); HV (Audits the detector high voltage); X1000, X100, X10, X1, Xo.1 (Counting ranges = multiply the meter reading by these figures).

www.thermo.com/rmp

©2007 Thermo Fisher Scientific Inc. All rights reserved. Kapton is a registered trademark of of E.I. du Pont de Nemours and Company. All other trademarks are the property of Thermo Fisher

Scientific Inc. and its subsidiaries. Results may vary under different operating conditions. Specifications, terms and pricing are subject to change. Not all products are available in all countries.

Please consult your local sales representatives for details. Literature Code LITMICROREM MICROSIEVERT 0407

Worldwide

Frauenauracher Strasse 96 +49 (0) 9131 909-0

D 91056 Erlangen, Germany +49 (0) 9131 909-205 fax

United Kingdom

Bath Road, Beenham, +44 (0) 118 971 2121

Reading RG7 5PR United Kingdom +44 (0) 118 971 2835 fax

United States +1 (508) 520-2815

27 Forge Parkway +1 (800) 274-4212 toll-free

Franklin, MA 02038 USA +1 (508) 428-3535 fax

Janet
Text Box
This product is available through: JRT Associates 5 Nepperhan Avenue, Suite 2B Elmsford, NY 10523 800-221-0111
Page 3: Tissue equivalent survey meters - JRT Associatesjrtassociates.com/pdfs/microrem.pdf · The tissue-equivalent scintillator used in the Micro Rem and Micro Sievert gives them a nearly

2011 Price List for the USA

All prices quoted herein are subject to change without prior notice.MicroREM

All prices are FOB: Origin. An additional charge will be added for FOB: Destination

Line Part Number Description Quantity Price

Micro Rem Tissue-Equivalent Survey Meters1.

MICROREMAO Micro Rem Survey meter for high sensitivity doserate measurements of gamma and X-radiation. Detector, internal tissue equivalent organic scintillator, 5 linear ranges from 0 - 20 µRem/hr up to 0 - 200 mRem/hr. with audio capability added

2. $4,420.00

MICROREMAOED Micro Rem Survey meter for high sensitivity doserate measurements of gamma and X-radiation. Detector, internal tissue equivalent organic scintillator, 5 linear ranges from 0 - 20 µRem/hr up to 0 - 200 mRem/hr. with audio capability added and extended dete

3. $4,650.00

MICROREMAOLE Low Energy version of the Micro Rem survey meter for high sensitivity doserate measurements of gamma and X-radiation. Detector, internal tissue equivalent organic scintillator, 5 linear ranges from 0 - 20 µRem/hr up to 0 - 200 mRem/hr. with audio

4. $4,630.00

MICROREMAOLEED Low Energy version of the Micro Rem survey meter for high sensitivity doserate measurements of gamma and X-radiation. Detector, internal tissue equivalent organic scintillator, 5 linear ranges from 0 - 20 µRem/hr up to 0 - 200 mRem/hr. with audio and ext

5. $4,900.00

Micro Sievert Tissue-Equivalent Survey Meters6.

MICROSIEVERTAO Micro Sievert survey meter for high sensitivity doserate measurements of gamma and X-radiation. Detector, internal tissue equivalent organic scintillator, 5 linear ranges from 0 - 0.2 µSv/hr up to 0 - 2 mSv/hr. with audio

7. $4,420.00

MICROSVAOED Micro Sievert survey meter for high sensitivity doserate measurements of gamma and X-radiation. Detector, internal tissue equivalent organic scintillator, 5 linear ranges from 0 - 0.2 µSv/hr up to 0 - 2 mSv/hr. with audio and extended detector

8. $4,630.00

MICROSVAOLE Low Energy version of the Micro Sievert Survey meter for high sensitivity doserate measurements of gamma and X-radiation. Detector, internal tissue equivalent organic scintillator, 5 linear ranges from 0 - 0.2 µSv/hr up to 0 - 2 mSv/hr. with audio

9. $4,630.00

MICROSV/AOLEED Low Energy version of the Micro Sievert Survey meter for high sensitivity doserate measurements of gamma and X-radiation. Detector, internal tissue equivalent organic scintillator, 5 linear ranges from 0 - 0.2 µSv/hr up to 0 - 2 mSv/hr. with audio and ex

10. $4,900.00

Extended Service Plans11.

MICROREM-DS3 3 Year Extended Depot Service for Micro Rem12. $590.00

Wednesday, December 08, 2010 Portables