thermox wdg-v, wdg-vc, wdg-vcm combustion analyzer · thermox® wdg-v, wdg-vc, wdg-vcm combustion...

130
Thermox ® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road Pittsburgh, PA 15238 PN 9000-133-VE Rev. K

Upload: others

Post on 29-Jul-2020

41 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Thermox®

WDG-V, WDG-VC, WDG-VCMCombustion Analyzer

User Manual

Thermox 150 Freeport Road

Pittsburgh, PA 15238PN 9000-133-VE Rev. K

Page 2: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

© 2014 AMETEK

This manual is a guide for the use of the Thermox WDG-V (VC / VCM). Data herein has been verified and validated and is believed adequate for the intended use of this instrument. If the instrument or procedures are used for purposes over and above the capabilities specified herein, con-firmation of their validity and suitability should be obtained; otherwise, AMETEK does not guarantee results and assumes no obligation or liability. This publication is not a license to operate under, or a recommendation to infringe upon, any process patents.

OfficesUSA - Pittsburgh 150 Freeport Road Pittsburgh, PA 15238 Ph. 412-828-9040Fax 412-826-0399

FRANCEAMETEK Precision Instruments France Rond Point de l’épine des champsBuroplus Bat D78990 Elanbourt Ph. 33 1 30 68 89 20Fax 33 1 30 68 89 29

USA - Delaware455 Corporate Blvd.Newark, DE 19702Ph. 302-456-4400Fax 302-456-4444

GERMANYAMETEK GmbH Rudolf-Diesel-Strasse 16 D-40670 Meerbusch Ph. 49 21 59 91 36 0 Fax 49 21 59 91 3680

USA - Texas4903 West Sam Houston Parkway NorthSuite A-400 Houston, TX 77041Ph. 713-466-4900Fax 713-849-1924

MIDDLE EAST – DubaiAMETEKPO Box 17067Jebel Ali Free ZoneDubai, UAEPh. 971-4-881 2052Fax 971-4-881 2053

CANADA2876 Sunridge Way N.E.Calgary, AB T1Y 7H9Ph. 403-235-8400Fax 403-248-3550

SINGAPOREAMETEK Singapore Pte. Ltd.No 43, Changi South Avenue 2, #04-01Singapore 486164Ph. 65-64842388Fax 65-64816588

CHINAAMETEK Shanghai (SPL)Room 408, Metro Tower#30 Tian Yao Qiao RoadShanghai 200030Ph. 86 21 6426 8111Fax 86 21 6426 7054

INDIAIndia Private Ltd.Plot 148, EPIP Zone Phase 2Featherlite Tech Park, 1st FloorWhitefield, Bangalore 560066Tel: 91-80-6782-3200Fax: 91-80-6780-3232

AMETEK Beijing CITIC Building, Room 2305 19, Jianguomenwai Dajie Beijing 100004 Ph. 86 10 8526 2111 Fax 86 10 8526 2141

BRAZILRod. Eng. Ermênio de Oliveira Penteado, Km 57Tombadouro, IndaiatubaSao Paulo, Brazil 13337-300Tel: 55 19 3825-8775Fax: 55 19 3935-8773

Page 3: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Contents | iii

ContentsOffices ..................................................................................................................... iiSafety Notes .......................................................................................................... viiImportant Notice To Users ...................................................................................viiiEnvironmental Information (WEEE) ..................................................................... ixGeneral Safety Summary ........................................................................................ xProduct Damage Precautions ................................................................................. xiDeclaration Of Conformity ................................................................................... xiiWarranty And Claims ..........................................................................................xiii

CHAPTER 1 Specifications Sensor: WDG-V ..................................................................................................1-1Sensor: WDG-VC ...............................................................................................1-2Sensor: WDG-VCM ............................................................................................1-3Remote Calibration Unit (RCU) ..........................................................................1-5

CHAPTER 2 System OverviewSensor Operations ................................................................................................2-1Sensor .................................................................................................................2-2

Basic Elements of the Sensor ........................................................................2-2The Oxygen Measuring Cell .........................................................................2-3Hydrocarbons ................................................................................................2-4

Sensor Operations - WDG-VCM .........................................................................2-5Close-Coupled Extrative Flue Gas Analyzer .......................................................2-5Detectors ..............................................................................................................2-6

Combustibles .................................................................................................2-6Methane .........................................................................................................2-7

Flow Sensor .........................................................................................................2-7Overview .......................................................................................................2-7Alarm Limits .................................................................................................2-8Troubleshooting ............................................................................................2-8

Temperature Sensors ............................................................................................2-9Cell Temperature Thermocouple ...................................................................2-9Combustible/Box Temperature RTD .............................................................2-9

Heaters .................................................................................................................2-9Cell Furnace Heater ......................................................................................2-9 Box Heater ....................................................................................................2-9

Electronics .........................................................................................................2-10I/O .............................................................................................................2-10 Analog Outputs ..................................................................................... 2-11 Alarm Relay Contacts ........................................................................... 2-11 Contact Input ........................................................................................2-12 RCU Interface .......................................................................................2-12 Communications Interface ....................................................................2-12

Page 4: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

iv | Thermox WDG-V / VC / VCM

Flame Arrestors ..................................................................................................2-13 Start-Up Checklist .............................................................................................2-14Common Operator Errors ..................................................................................2-15Technical Support ..............................................................................................2-16

CHAPTER 3 Installation GuideAnalyzer Location ...............................................................................................3-2Inspect Shipping Contents ...................................................................................3-2Mechanical Installation ........................................................................................3-3

Sample Inlet Probe Installation .....................................................................3-3Probe Heater Installation ...............................................................................3-6Exhaust Tube Installation ..............................................................................3-7Sensor Mounting ...........................................................................................3-7Calibration/Aspirator Air ............................................................................3-13Z-Purge Connections ...................................................................................3-14Remote Calibration Unit Mechanical Installation ......................................3-18

Wiring ................................................................................................................3-24General Wiring and Conduit Requirements ................................................3-25Mains Supply Connections .........................................................................3-26Current Output Connections .......................................................................3-29

RCU Connections ..............................................................................................3-31

CHAPTER 4 AMEVision Display User Interface or CommunicationsDefinition of the AMEVision Display User Interface & Purpose........................4-1AMEVision Display User Interface & Keypad ...................................................4-2Setting the Analyzer Address ...............................................................................4-3Connecting to the WDG-V ..................................................................................4-4

Connecting a Single Sensor to the AMEVision ............................................4-4Connecting Multiple Sensors to the AMEVision ..........................................4-4Customer I/O Interconnecting .......................................................................4-5Maximum Distance Between Sensor & AMEVision .............................. 4-6

Troubleshooting ...................................................................................................4-6 No Communications ...............................................................................4-6 Intermittent Communications .................................................................4-6

WDG-V PC Configurator Software .....................................................................4-7

CHAPTER 5 Flow Setting

CHAPTER 6 LED Status Indicator

Page 5: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Contents | v

CHAPTER 7 Troubleshooting General Troubleshooting .....................................................................................7-2

Leak Check ...................................................................................................7-2Plugged Plumbing Check ..............................................................................7-3Aspirator Air Not Pulling Sample from Process ...........................................7-3

Diagnostics Checks ..............................................................................................7-4Wiring Checks ...............................................................................................7-5Interconnecting Wiring Problem ...................................................................7-5Thermocouple Checks ...................................................................................7-5Shorted/Failed Thermocouple .......................................................................7-5Reversed Thermocouple Wires .....................................................................7-5Thermocouple Compensation ......................................................................7-5Circuit Check ................................................................................................7-5Calibration/Aspirator Setup Checks ..............................................................7-7AC Power Checks .........................................................................................7-8Furnace Checks .............................................................................................7-8Process Pressure Checks ...............................................................................7-8Cell Checks ...................................................................................................7-8

Alarm Messages (Red Bar on AMEVision Display User Interface) ...................7-9Cell T/C Failure .............................................................................................7-9Cold Junction Compensator Failure ..............................................................7-9Cell Temperature Control ..............................................................................7-9Cell Over Tem ...............................................................................................7-9Cell Under Temp ...........................................................................................7-9Cell Temp Rise Failure ..................................................................................7-9Over Temp Relay Tripped .............................................................................7-9RTD Failure ...................................................................................................7-9Box Temp High .............................................................................................7-9Box Temp Low ..............................................................................................7-9Box Temp Rise Failure ..................................................................................7-9Low Sample Flow .......................................................................................7-10Cell Failure (Open) .....................................................................................7-10Combustible Detector Open ........................................................................7-10Methane Detector Open ..............................................................................7-10Cell Mv Mismatch ......................................................................................7-10Analog Output 1 Failure .............................................................................7-10Flow Sensor Failure ....................................................................................7-10Cell T/C Measurement Mismatch ...............................................................7-10

Page 6: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

vi | Thermox WDG-V / VC / VCM

Warning Messages (Yellow Bar on AMEVision Display User Interface) ......... 7-11Cell Life Nearing its End ............................................................................ 7-11Comb Detector Life Nearing its End .......................................................... 7-11Methane Detector Life Nearing its End ...................................................... 7-11High Cell Mv .............................................................................................. 7-11Analog Output 1 Out of Range ................................................................... 7-11Analog Output 2 Out of Range ................................................................... 7-11Analog Output 3 Out of Range ................................................................... 7-11Last O2 Span Calibration Failed .................................................................. 7-11Last O2 Zero Calibration Failed .................................................................. 7-11Last Comb Calibration Failed ..................................................................... 7-11Last CH4 Calibration Failed ....................................................................... 7-11Last Flow Calibration Failed ....................................................................... 7-11Oxygen Calibration Required ..................................................................... 7-11Combustible Calibration Required .............................................................. 7-11CH4 Calibration Required ........................................................................... 7-11

Chapter 8 Service and Parts Warnings & Cautions ...........................................................................................8-2Wiring Diagram ...................................................................................................8-3Cell Replacement/Cleaning .................................................................................8-5Thermocouple Replacement ................................................................................8-7Detector Replacement ..........................................................................................8-9

Combustibles Detector Replacement ............................................................8-9Methane Detector Replacement ....................................................................8-9

Flow Sensor Replacement ................................................................................. 8-11Furnace Replacement ........................................................................................8-13Heater Replacement ...........................................................................................8-14

Box Heater Replacement ............................................................................8-14Combustible Heater Replacement ...............................................................8-16

Electronics Replacement ...................................................................................8-18Recommended Maintenance Schedule ..............................................................8-20

Calibration ...................................................................................................8-20Replacement Parts List ......................................................................................8-21

Chapter 9 Drawings and Custom Instructions

Chapter 10 Customer Modbus Map

Appendix A French Warnings / Français Avertissements

Page 7: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

| vii

Safety Notes

WARNINGS, CAUTIONS, and NOTES contained in this manual emphasize critical instructions as follows:

An operating procedure which, if not strictly observed, may result in personal injury or environmental contamination.

An operating procedure which, if not strictly observed, may result in damage to the equipment.

Important information that should not be overlooked.

Burn hazard. Hot surface. Do not touch, allow to cool before servicing.

Electrical Safety

Up to 5 kV may be present in the analyzer housings. Always shut down power source(s) before performing maintenance or troubleshooting. Only a qualified electrician should make electrical connections and ground checks.

Any use of the equipment in a manner not specified by the manufacturer may impair the safety protection originally provided by the equipment.

Grounding

Instrument grounding is mandatory. Performance specifications and safety protection are void if instrument is operated from an improperly grounded power source.

Verify ground continuity of all equipment before applying power.

NOTE

Page 8: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

viii | Thermox WDG-V / VC / VCM

Important Notice to Users

The following applies to the WDG-V Division 2 Models:

Power, input, and output (I/O) wiring must be in accordance with Class I, Division 2 wiring methods [Article 501-4(b) of the National Electric Code, NFPA 70] and in ac-cordance with the authority having jurisdiction.

WARNING - EXPLOSION HAZARD - SUBSTITUTION OF COMPO-NENTS MAY IMPAIR SUITABILITY FOR CLASS I, DIVISION 2.

WARNING - EXPLOSION HAZARD - WHEN IN HAZARDOUS LO-CATIONS, TURN OFF POWER BEFORE REPLACING OR WIRING MODULES.

WARNING - EXPLOSION HAZARD - DO NOT DISCONNECT EQUIP-MENT UNLESS POWER HAS BEEN SWITCHED OFF OR THE AREA IS KNOWN TO BE NONHAZARDOUS.

THIS EQUIPMENT IS SUITABLE FOR USE IN CLASS I, DIVISION 2, GROUPS A, B, C, AND D OR NONHAZARDOUS LOCATIONS ONLY.

The maximum ambient temperature for the analyzer is 60°C (140°F).

The WDG-V sensor and AMEVision Display User Interface is a complex piece of equipment that should only be serviced by a qualified service technician with expertise in instrument technology and electrical systems. AMETEK recommends that all equip-ment requiring service be sent back to the factory. You should only attempt to repair or service this equipment after receiving training from an AMETEK/P&AI Division train-ing representative. If you decide to service this equipment be aware that high voltages, high temperatures, and other potentially hazardous conditions may arise.

Page 9: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

| ix

Environmental Information (WEEE)

This AMETEK product contains materials that can be reclaimed and recycled. In some cases the product may contain materials known to be hazardous to the environ-ment or human health. In order to prevent the release of harmful substances into the environment and to conserve our natural resources, AMETEK recommends that you arrange to recycle this product when it reached its “end of life”.

Waste Electrical and Electronic Equipment (WEEE) should never be disposed of in a municipal waste system (residential trash). The Wheelie Bin marking on this prod-uct is a reminder to dispose of the product properly after it has completed its useful life and been removed from service. Metals, plastics, and other components are recyclable and you can do your part by doing one of the following steps:

• When the equipment is ready to be disposed of, take it to your local or regional waste collection administration for recycling.

• In some cases, your “end of life” product may be traded in for credit towards the pur-chase of new AMETEK instru-ments. Contact your dealer to see if this program is avail-able in your area.

• If you need further assistance in recycling your AMETEK product, contact our office listed in the front of the instruc-tion manual.

Page 10: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

x | Thermox WDG-V / VC / VCM

General Safety Summary

Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it.

Use Proper Wiring To avoid fire hazards, use only the wiring specified in the Installation Chapter of this user’s manual.

Avoid Electrical OverloadTo avoid electrical shock or fire hazard, do not apply a voltage to a terminal that is outside the range specified for that terminal.

Ground the ProductFollow the grounding instructions provided in the Installation Chapter of this user’s manual. Before making connections to the input or output terminals of this product, ensure that the product is properly grounded.

Do Not Operate without CoversTo avoid electric shock or fire hazard, do not operate this product with covers or panels removed.

Use Proper FuseTo avoid fire hazard, use only the fuse type and rating specified for this product.

Do Not Operate in Explosive AtmosphereTo avoid injury or fire hazard, do not operate this product in an explosive atmosphere unless you have purchased options that are specifically designed for these environments.

Page 11: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Contents | xi

Product Damage Precautions

Use Proper Power SourceDo not operate this product from a power source that applies more than the voltage specified.

Do Not Operate with Suspected FailuresIf you suspect there is damage to this product, have it inspected by qualified service personnel.

Use Caution When LiftingThe analyzer weighs approximately 20 kg (44 lb). Use caution when lifting it from its crate.

Use Proper AttireEquippement is hot, user should wear protective groves while handling the equipment.

Page 12: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

xii | Thermox WDG-V / VC / VCM

Declaration of Conformity

Manufacturer’s Name: AMETEK/Thermox®

Manufacturer’s Address: Process & Analytical Instruments Division 150 Freeport Road Pittsburgh, PA 15238 declares that the products:

Product Names: Thermox WDG series of Flue Gas Analyzers using AMEVision Display User Interface: WDG-V WDG-VC WDG-VCM

Conform to the following standards:

EMC compliance: EMC Directive 2004/108/EC Immunity: EN 50082-2:1995 Generic - Heavy Industrial Equipment EN 50082-1:1997 Generic - Residential, Commercial, Light Industrial Equipment EN 61000-6-2:1999 Generic - Heavy Industrial Equipment EN 61326:1997 Measuring Equipment (Heavy Industrial) Emissions: EN 55011:1998 Class A, ISM Equipment Safety Compliance: UL 3101-1 UL Listed Inspection and Measuring Electrical Equipment CAN/CSA 22.2 No. 1010.1-92 cUL Listed Inspection and Measuring Electrical Equipment

Low Voltage Directive 73/23/EEC: EN61010-1:1993, Amendment A2: 1995 Electrical Equipment for Measurement, Controls, and Laboratory Use

Ingress Protection: IP65

Manufacturer’s Address in Europe: AMETEK Precision Instruments Europe GmbHRudolf-Diesel-Strasse 16D-40670 Meerbusch, Germany

June 2013

Page 13: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Contents | xiii

WARRANTY AND CLAIMS

We warrant that any equipment of our own manufacture or manufactured for us pursuant to our speci-fications which shall not be, at the time of shipment thereof by or for us, free from defects in material or workmanship under normal use and service will be repaired or replaced (at our option) by us free of charge, provided that written notice of such defect is received by us within twelve (12) months from date of shipment of portable analyzers or within eighteen (18) months from date of shipment or twelve (12) months from date of installation of permanent equipment, whichever period is shorter. All equipment requiring repair or replacement under the warranty shall be returned to us at our factory, or at such other location as we may designate, transportation prepaid. We shall examine such returned equipment, and if it is found to be defective as a result of defective materials or workmanship, it shall be repaired or replaced as aforesaid. Our obligation does not include the cost of furnishing any labor in connection with the installation of such repaired or replaced equipment or parts thereof, nor does it include the responsibility or cost of transportation. In addition, instead of repairing or replacing the equipment returned to us as aforesaid, we may, at our option, take back the defective equipment, and refund in full settlement the purchase price thereof paid by Buyer.

The warranty shall not apply to any equipment (or part thereof) which has been tampered with or altered after leaving our control or which has been replaced by anyone except us, or which has been subject to misuse, neglect, abuse or improper use. Misuse or abuse of the equipment, or any part thereof, shall be construed to include, but shall not be limited to, damage by negligence, accident, fire or force of the elements. Improper use or misapplications shall be construed to include improper or inadequate protection against shock, vibration, high or low temperature, overpressure, excess voltage and the like, or operating the equipment with or in a corrosive, explosive or combustible medium, unless the equipment is specifically designed for such service, or exposure to any other service or environment of greater severity than that for which the equipment was designed.

The warranty does not apply to used or secondhand equipment nor extend to anyone other than the original purchaser from us.

THIS WARRANTY IS GIVEN AND ACCEPTED IN LIEU OF ALL OTHER WARRANTIES, WHETHER EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION AND WARRAN-TIES OF FITNESS OR OF MERCHANTABILITY OTHER THAN AS EXPRESSLY SET FORTH HEREIN, AND OF ALL OTHER OBLIGATIONS OR LIABILITIES ON OUR PART. IN NO EVENT SHALL WE BE LIABLE UNDER THIS WARRANTY OR ANY OTHER PROVISION OF THIS AGREEMENT FOR ANY ANTICIPATED OR LOST PROFITS, INCIDENTAL DAMAGES, CONSEQUENTIAL DAMAGES, TIME CHANGES OR ANY OTHER LOSSES INCURRED BY THE ORIGINAL PURCHASER OR ANY THIRD PARTY IN CONNECTION WITH THE PURCHASE, INSTALLATION, REPAIR OR OPERATION OF EQUIPMENT, OR ANY PART THEREOF COVERED BY THIS WARRANTY OR OTHERWISE. WE MAKE NO WARRANTY, EXPRESS OR IMPLIED, INCLUDING WITHOUT LIMITATION ANY WARRANTIES OF FIT-NESS OR OF MERCHANTABILITY, AS TO ANY OTHER MANUFACTURER’S EQUIPMENT, WHETHER SOLD SEPARATELY OR IN CONJUNCTION WITH EQUIPMENT OF OUR MANU-FACTURE. WE DO NOT AUTHORIZE ANY REPRESENTATIVE OR OTHER PERSON TO ASSUME FOR US ANY LIABILITY IN CONNECTION WITH EQUIPMENT, OR ANY PART THEREOF, COVERED BY THIS WARRANTY.

Page 14: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

xiv | Thermox WDG-V / VC / VCM

This page intentionally left blank.

Page 15: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Specifications | 1-1

SPECIFICATIONS

Sensor : WDG-V

Output Range From 0-1% to 0-100%Accuracy ± 0.75% of measured value or ± 0.05% oxygen, whichever is greaterResponse 90% of a process step change < 6 secs.

Drift < 0.1% of cell output per month (< 0.005% O2 per month with 2% O2 applied)

Aspirator AirRequirements ≈ 3 scfh (1.4 L/min) at 15 to 100 psig (1.05 to 7.04 kg/cm2)

Max. Flue Gas Temp / Probe Type / Lengths1300°F (704°C)1875°F (1024°C)3000°F (1649°C)

316 SS310 SSHexoloy®

36” - 108” (91 cm - 271 cm)36” - 108” (91 cm - 271 cm)24” - 72” (60 cm-182 cm)

Max. SampleDewpoint

392°F (200°C) standard.

Sample Pressure ± 6 inch water gauge

Environment Ambient Temp.: -13°F to 149°F (-25°C to 65°C)

-4°F to 140°F (-20°C to 60°C) for Purged Units Relative Humidity: 10% to 90%, non-condensing Max. Altitude: 5000 metersEnclosure Hinged IP65, Optional Div II

Power Requirements 115 VAC, ± 10%, 47-63 Hz, 600VA max.

230 VAC, ± 10%, 47-63 Hz, 600 VA max.

Calibration Gas Requirements Use calibration gases @ 10 PSIG, 3.0 SCFH (1.4 L/min.)

O2 Span Gas: Air or from 1.0 to 100% O2, balance N2O2 Zero Gas: 2% or from 0.1 to 10% O2, balance N2

System Compliance

EMC Compliance: Safety Compliance:

2004/108/EC73/23/EEC

NOTES1. All static performance characteristics are with operating variables constant. 2. System accuracy referenced to 0.1 to 10% calibrated range. 3. Response is to calibration gas.

Page 16: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

1-2 | Thermox WDG-V / VC / VCM

Sensor : WDG-VC

Output Range From 0-1% to 0-100% Oxygen Combustibles

From 0-1% to 0-100%From 0-1,000 ppm to 0-10,000 ppm or from 0-1% to 0-5%

Accuracy Oxygen ± 0.75% of measured value or ± 0.05% oxygen, whichever is greater Combustibles ± 2% of full scale output rangeResponse O2 90% of a process step change < 6 secs.

Drift < 0.1% of cell output per month (< 0.005% O2 per month with 2% O2 applied)

Aspirator AirRequirements ≈ 3 scfh (1.4 L/min) at 15 to 100 psig (1.05 to 7.04 kg/cm2)

Max. Flue Gas Temp / Probe Type / Lengths1300°F (704°C)1875°F (1024°C)3000°F (1649°C)

316 SS310 SSHexoloy®

36” - 108” (91 cm - 271 cm)36” - 108” (91 cm - 271 cm)24” - 72” (60 cm-182 cm)

Max. SampleDewpoint 392°F (200°C)

Sample Pressure ± 6 inch water gauge

Environment Ambient Temp.:

-13°F to 149°F (-25°C to 65°C)-5°F to 140°F (-20°C to 60°C) with Div. 2 Option

Relative Humidity: 10% to 90%, non-condensing Max. Altitude: 5000 metersEnclosure

Power Requirements 115 VAC, ± 10%, 47-63 Hz, 600VA max. 230 VAC, ± 10%, 47-63 Hz, 600 VA max.

Calibration Gas Requirements Use calibration gases @ 10 PSIG, 3.0 SCFH (1.4 L/min.) O2 Span Gas: Air or from 1.0 to 100% O2, balance N2

O2 Zero Gas & Comb Zero Gas:

2% or from 0.1 to 10% O2, balance N2

Comb Span Gas: 60 to 80% (ppm ranges) of the selected comb. recorder output range in certified mixtures of CO + H2, 3-4% O2, balance N2.

System Compliance

EMC Compliance: Safety Compliance:

2004/108/EC73/23/EEC

NOTES: 1. All static performance characteristics are with operating variables constant.2. System accuracy referenced to 0.1 to 10% calibrated range. 3. Response is to calibration gas (without flame arrestors).

Page 17: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Specifications | 1-3

Sensor : WDG-VCM

Output Range From 0-1% to 0-100% Oxygen Combustibles Methane

From 0-1% to 0-100%From 0-2,000 ppm to 0-10,000 ppm0 to 5%

Accuracy Oxygen ± 0.75% of measured value or ± 0.05% oxygen, whichever is greater Combustibles ± 2% of full scale output range Methane ±5% of full scale output rangeResponse O2 90% of a process step change < 6 secs.

Drift < 0.1% of cell output per month (< 0.005% O2 per month with 2% O2 applied)

Aspirator AirRequirements ≈ 3 scfh (1.4 L/min) at 15 to 100 psig (1.05 to 7.04 kg/cm2)

Max. Flue Gas Temp / Probe Type / Lengths1300°F (704°C)1875°F (1024°C)3000°F (1649°C)

316 SS310 SSHexoloy®

36” - 108” (91 cm - 271 cm)36” - 108” (91 cm - 271 cm)24” - 72” (60 cm-182 cm)

Max. SampleDewpoint 392°F (200°C)

Sample Pressure ± 6 inch water gauge

Environment Ambient Temp.:

-13°F to 149°F (-25°C to 65°C);-5°F to 140°F (-20°C to 60°C) with Div. 2 Option

Relative Humidity: 10% to 90%, non-condensing Max. Altitude: 5000 metersEnclosureHinged NEMA 3R, weather resistant, stainless steel. Optional hinged NEMA 4X (IP56), explo-sion-proof, purged, Div II.

Power Requirements 115 VAC, ± 10%, 47-63 Hz, 700VA max. (750 VA max. floor mount option)

230 VAC, ± 10%, 47-63 Hz, 1950 VA max. (2000 VA max. floor mount option)

Page 18: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

1-4 | Thermox WDG-V / VC / VCM

Calibration Gas Requirements for Combustibles DetectorUse calibration gases @ 10 PSIG, 3.0 SCFH (0.70 kg/cm2, 0.7 L/min.) O2 Span Gas: Air or from 1.0 to 100% O2, balance N2

O2 Comb & CH4 Zero Gases: From 0.1 to 10% O2, balance N2

Comb Span Gas: 60 to 80% (ppm ranges) or 40 to 60% (% ranges) of the selected comb. recorder output range in certified mixtures of CO + H2, 3-4% O2, balance N2.

CH4 Span Gas 2% CH4, 8% to 10% O2, balance N2

System Compliance

EMC Compliance: Safety Compliance:

2004/108/EC73/23/EEC

NOTES:

1. All static performance characteristics are with operating variables constant.2. System accuracy referenced to 0.1 to 10% calibrated range.3. Response is to calibration gas (without flame arrestors).

Page 19: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Specifications | 1-5

Remote Calibration Unit (RCU) O2 Only RCU O2 & Comb. RCU O2, Combustibles & Methane RCU

Enclosure UL Type 4X (NEMA 4X (IP56), Div IIEnvironment Ambient Temperature: Humidity: Max Altitude: IEC Installation IEC Pollution Degree 2

-18°C to 60°C 0 to 90%, non-condensing2000 MetersCategory II

System ComplianceEMC Compliance: 2004/108/ECSafety Compliance: 73/23/EEC

Page 20: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

1-6 | Thermox WDG-V / VC / VCM

This page intentionally left blank.

Page 21: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

System Overview | 2-1

SYSTEM OVERVIEW

Sensor Operations

Designed for fast response in a wide range of flue gas applications, the sensor mounts directly on the combustion process to provide continuous measurement of oxygen, or oxygen and combustibles, or oxygen, combustibles and methane.

A sample is drawn from the process stream by means of an air-operated aspira-tor and is immediately returned to the process. A portion of this gas rises past the oxygen measuring cell and combustibles detector and returns to the primary loop. All sample wetted parts are heated to 200°C maintain the sample above the dewpoint of the gas.

Follow these cautions when working on the sensor:

Do not turn on the aspirator until the analyzer has been turned on and is hot - preferably 24 hours. Also, if you turn off the analyzer or the process is shut down, turn off the aspirator air to avoid plugging problems! Remove AC power from the sensor and allow the sensor to cool for at least one hour before performing any maintenance or trouble-shooting activities. The outside of the sensor cover and all sensor assembly components are hot during normal operation (up to 500°F, 260°C inside the cover). Allow sensor components to cool for at least an hour before working inside the sensor. Use caution and wear appropriate gloves when handling components or when touching the sensor cover!

Page 22: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

2-2 | Thermox WDG-V / VC / VCM

Sensor

Basic Elements of the Sensor

The WDG-V series analyzer consists of the following basic systems:

The Plumbing All inlet and outlet tubing, O2 cell housing, and the combustibles flow block.

Also includes the calibration gas inlet and aspirator assembly, flow sensor, and the aspirator used to pull the sample into the sensor.

The Measuring System The oxygen cell, combustibles detector, methane detector, and flow sensor.

The Temperature System The electrical cell heater (furnace), sensor case heaters, the type “K” thermo-

couple (maintains cell operating temperature). The sensing cell operates at a constant temperature. The circuit board in the sensor electronic box switches power to the furnace from the AC mains connected to the sensor. This board also provides cold junction compensation to the thermocouple circuit.

The WDG-V is a complete standalone unit. No separate controller is required. All required I/O is contained within the WDG-V including:

• Qty (3) configurable 4-20 mA outputs

• Qty (2) system alarm relays (Service and Data Valid)

• Qty (3) configurable process alarm relays

• Qty (1) RS485, 2 wire, MODBUS RTU interface (57.6 Kbaud)

• Qty (1) Contact input (for remote auto calibration trigger)

• Valve controls for Remote Calibration Unit (RCU)

The WDG-V is configured, calibrated and monitored via the MODBUS RTU interface. AMETEK provides two options for communicating with the WDG-V:

• AMEVision Display User Interface

• WDG-V Configurator Software via PC (Provided with each unit)

Since the interface is industry standard MODBUS, custom user software can also be written. For this purpose, consult the factory for the MODBUS register defini-tions.

Page 23: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

System Overview | 2-3

The Oxygen Measuring Cell

The sensing element itself is a closed-end tube made from ceramic zirconium oxide stabilized with an oxide of yttrium or calcium. Porous platinum coatings on the inside and outside serve as a catalyst and as electrodes. At high temperatures (generally above 1200°F/650°C), oxygen molecules coming in contact with the platinum electrodes near the sensor become ionic. As long as the oxygen partial pressures on either side of the cell are equal, the movement is random and no net flow of ions occurs. If, however, gases having different oxygen partial pressures are on either side of the cell, a potentiometric voltage is produced (See Figure 2-1). The magnitude of this voltage is a function of the ratio of the two oxygen partial pressures. If the oxygen partial pressure of one gas is known, the voltage produced by the cell indicates the oxygen content of the other gas. A reference gas, usually air (20.9% O2), is used for one of the gases.

Figure 2-1. Zirconium oxide cell principle of operation.

Since the voltage of the cell is temperature dependent, the cell is maintained at a constant temperature. The oxygen content is then determined from the Nernst equation:

E = InRT4F

O1O2

where R and F are constants, T is absolute temperature, and O1 and O2 are the oxygen partial pressures on either side of the cell.

Page 24: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

2-4 | Thermox WDG-V / VC / VCM

For measuring oxygen in non-combustibles gases, the calibration of an analyzer is obtained from the formula:

AT = 48.0 at 680°CE = A*T*Log 20.9%O2 Unk%

Where A is a constant, T is the cell temperature on an absolute scale (°C + 273) and O2 Unk% is the unknown oxygen concentration of the gas to be analyzed, and which is calculated by the analyzer.

The cell produces zero voltage when the same amount of oxygen is on both sides, and the voltage increases as the oxygen concentration of the sample decreases. The voltage created by the difference in the sample gas and the reference air is carried by a cable to the microprocessor control unit, where it is linearized to an output signal.

Because of the high operating temperature of the cell, combustible gases that are present may burn. When this occurs, the cell will generate high millivolts and cause the display to indicate less oxygen than is actually in the gas (net oxygen content).

Hydrocarbons

When hydrocarbons are present in the gas sample, a combustion process occurs when this gas sample is exposed to the high temperature of the zirconia cell. An indication that hydrocarbons may be present in the gas sample is that the oxygen reading will be lower than expected. For example, if a calibration gas cylinder has an oxygen value of 20 PPM and 5 PPM of hydrogen (balance nitrogen), the oxygen analyzer will read 17.5 PPM oxygen. The reduction of oxygen is due to the combustion process where 5 PPM of hydrogen will combine with 2.5 PPM of oxygen to form water. Thus, the oxygen analyzer will read 17.5 PPM rather than the actual 20 PPM oxygen.

The amount of oxygen reduction is dependent on the type of hydro-carbons present in the sample gas.

NOTE

NOTE

Page 25: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

System Overview | 2-5

Sensor Operations - WDG-VCMClose-Coupled Extrative Flue Gas Analyzer

The WDG-VCM analyzer is ideal for natural gas-fired power boilers or for those using gas during start-up and shutdown. With this analyzer, you can monitor oxygen and combustibles for maximum fuel efficiency. In addition, the methane detector can be used for purge and lightoff cycles during start-up and shutdown.

WDG-VCM analyzers with a combustibles detector are used for processes that use natural gas for start-up and oil, fuel gas or coal for primary fuel. Process gas is drawn into the analyzer’s primary sample loop by an aspirator and returns to process. Gas enters the split flow block that is separted by a baffle and contacts the hot-wire catalytic combustibles and methane detector. It then passes the zirco-nium oxide cell, where the oxygen concentration is measured, before returning to the main sample loop and then back to the process.

Detectors are very delicate and should be handled with care.NOTE

Page 26: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

2-6 | Thermox WDG-V / VC / VCM

Detectors

Combustibles

The Combustibles Detector is a dual element device. The elements differ only in that one is coated with a catalyst. The catalyst causes oxidation to occur at a low-er-than-normal temperature. The temperature of the catalyzed element changes as the combustible mixture burns. The temperature change causes the resistance of the catalyzed element to change. The resistance change is interpreted by the microprocessor and the corresponding combustibles reading is displayed.

The Thermox catalytic combustibles detector will detect combustible gas present in a sample. The sample must, however, contain enough oxygen to fully burn the combustible present. The combustibles detector responds to all unburned com-bustibles gases. This includes gases such as CO and H2. You should attempt to match your calibration gas with a mixture of CO and H2 that most closely match-es the combustibles mix in your process. We recommend using an even mixture of CO and H2 as the combustibles component within your span gas.

Examples:

• Flue gas containing 1000 PPM CO, 1000 PPM H2 and 1% O2 with an H2 O, CO2 and N2 balance will be sensed to contain 2000 PPM combustibles by the catalytic detector (provided the combustibles calibration gas consists of equal parts CO and H2 ).

• Flue gas containing 1000 PPM CO, 1000 PPM H2 and 500 PPM O2 with a H2 O, CO2 and N2 balance will be sensed to contain only 1000 PPM Comb. by the catalytic detector, since there is less than stoichiometric oxygen present. In this case, the combustibles detector output will be set to full scale/

• If there is little or no other combustible component in the flue gas, the com-bustibles sensor can be calibrated on a mixture of CO and air to give a CO meaasurement.

Page 27: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

System Overview | 2-7

Methane

The Methane Detector is a dual element device. The elements differ only in that one is coated with a catalyst. The catalyst causes oxidation to occur at a lower-than-normal temperature. The temperature of the catalyzed element changes as the mixture burns. The temperature change causes the resistance of the catalyzed element to change. The resistance change is interpreted by the microprocessor and the corresponding reading is displayed.

The Thermox catalytic methane detector will detect methane gases present in a sample. The sample must, however, contain enough oxygen to fully burn the methane present.

The methane span gas should be 2% CH4 with 8 - 10% oxygen present in the mixture.

Flow Sensor

Overview

The aspirator creates a vacuum and pulls the process gas into the analyzer. There are two paths for the gas to take.

1. Immediately back to the process gas.

2. Through the sample loop, past the cell and/or combustible detector.

Calibrated at the factory, the flow sensor measures the sample flow through the sample loop. The flow measurement is updated every 90 seconds. This delay is used to allow the sensor to effectively auto-zero to prevent false alarms.

Page 28: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

2-8 | Thermox WDG-V / VC / VCM

Alarm Limits

The flow sensor reports flow as relative to ideal (ie. 100% is ideal flow, 50% is half of flow). An alarm is generated when the sample flow is less than 45% ideal.

The flow measurement is directional, with reverse flow generating an alarm.

Troubleshooting

The flow sensor contains two resistive elements. Each element should be ap-proximately 8-10 Ω.

A Flow Sensore failed alarm will be generated if either element is open.

To Measure:

- Turn off the instrument.

- Measure the resistance of each element. (Flow Heater, Flow Sense) at the terminals inside the electronics enclosure.

- If resistance is much lower or higher than 8-10 Ω, the flow sensor must be replaced.

NOTE

Flow Sensor Heater Element (Black Wire)Flow Sensor Sense Element (Brown Wire)

Page 29: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

System Overview | 2-9

Temperature Sensors

There are two tightly controlled temperature zones in the WDG-V. The Oxygen Cell Temperature and the combustible block temperature (CM units only). (On Oxygen only units the box temperature is controlled with an RTD mounted on the aspirator block.)

Cell Temperature

The Oxygen cell is controlled to 680°C. The cell temperature is measured by dual, Type K thermocouples. Dual thermocouples are used for self check of the cell temperature. The thermocouples are collocated in a single assembly. There is no distinction between the two thermocouples as they are effectively at the same location in the assembly.

Combustible/Box Temperature

The combustible block/ box temperature is controlled to 225°C. This temperature is measured by 100 ohm platinum RTD. The RTD is embedded in the combus-tible block. On oxygen only units, the RTD is mounted on the aspirator block. All internal plumbing components are maintained above 200°C.

Heaters

There are two main heaters in the WDG-V, the cell heater and box heater. In CM units, an additional heater is used for fine temperature control of the combustible block.

Cell Heater

The Oxygen Cell heater is a 300W coil heater encompassing the cell housing and providing precision heat to the oxygen cell.

Box Heater

The box heater is a 350W cartridge heater located in the aluminum heater block at the bottom of the aspirator. This heater provides the heat for the enclosure to maintain all plumbing components above 200°C. The combustible heater is a 50W heater used for precise temperature control of the combustible Block.

Page 30: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

2-10 | Thermox WDG-V / VC / VCM

Electronics

• One assembly - mounted to a plate

• The electronics/software completely control the sensor. - No external controllers are required.

• The configuration/monitoring interface to the electronics is RS485 MODBUS RTU.

• Sensor wiring and Customer wiring (i.e. power, I/O) are directly connected to the main electronics. - Connections are pluggable. - Both plugs and headers are laser printed for easy matching.

I/O

The WDG-V has the following I/O available:

• Qty (3) configurable 4-20 mA outputs

• Qty (2) system alarm relays (Service and Data Valid)

• Qty (3) configurable process alarm relays

• Qty (1) RS485, 2 wire, MODBUS RTU interface (57.6 Kbaud)

• Qty (1) Contact input (for remote auto calibration trigger)

• Valve controls for Remote Calibration Unit (RCU)

Page 31: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

System Overview | 2-11

Analog Outputs

There are 3 analog outputs in the WDG-V. Each analog output can be configured for Mode, Function, and Span, Zero.

Mode options include, 0-20ma, 4-20ma and NAMUR.

Functions include; Oxygen, cell millivolts, combustibles*, methane*

The analog outputs can be powered internally or externally. External power must be greater than 26V.

The NAMUR outputs are implemented as follows:

4-20 mA signal Analyzer Condition0 mA Analyzer unpowered, or completely failed3.5 mA Critical Alarm - analyzer reading unusable (factory default)3.8 mA Reading Under Range (Example: user sets range to 2-10%. Current reading is 1.9%4 to 20 mA Normal Operation20.5 mA Reading Over Range (Example: range is 0-10%. Current reading is 12%

* For combustible and/or methane units only.

Alarm Contacts

There are 5 alarm contacts in the WDG-V, 1 service alarm, 1 data valid alarm, and 3 configurable process alarms.

Service Alarm Relay

This relay is used to determine if there is a problem with the analyzer. This relay is normally open. Contacts are closed when in normal mode. The relay will be open when there is a problem with the analyzer or on loss of power.

Data Valid Relay

This relay is used to determine if the output is representative of the process. This relay is normally open. Contacts are closed when in normal mode. Contacts are open when there is a problem with the analyzer, when the analyzer is calibrating, warming up, when the analyzer is in diagnostic mode, or on loss of power.

Page 32: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

2-12 | Thermox WDG-V / VC / VCM

Contact Input

There is one Contact Input peripheral on the WDG-V.

An auto calibration can be initiated using the contact input peripheral. (Only with the RCU option).

Auto calibration must be selected in the sensor configuration menu.

Remote Calibration Unit Interface

The WDG-V has 5 solenoid valve drivers that are compatible with AMETEK’s Remote Calibration Unit (RCU).

See wiring section for interface details.

Communications Interface

There is one communications interface on the WDG-V.

• RS485 MODBUS RTU Configuration: - Two wire -57.6 Kbaud - No parity - 1 stop bit

The WDG-V is configured, calibrated and monitored via the MODBUS RTU interface. AMETEK provides two options for communicating with the WDG-V:

• AMEVision Display User Interface (Reference separate manual)

• WDG-V Configurator Software via PC (Provided with each unit)

Since the interface is industry standard MODBUS, custom user software can also be written. For this purpose, refer to Chapter 11 for the MODBUS MAP or consult the factory for the MODBUS register definitions.

Page 33: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

System Overview | 2-13

Flame Arrestors

The WDG-V sensor contains two flame arrestors to prevent it from being an igni-tion source to the process for short periods of high combustible levels (25% of LEL - Lower Explosive Limit) in the process. The flame arrestors are not, howev-er, intended to protect the process where the combustibles’ levels are constantly high.

Page 34: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

2-14 | Thermox WDG-V / VC / VCM

Start-Up Checklist

Review the installation chapter in the User’s Manual.

Install the sensor on the process.

- If the sensor is on the pipe nipple, insulate the pipe from the process wall to the mounting flange on the sensor.

- Install the probe.

- Install the probe heater, if required.

This type of sensor requires a clean, dry instrument air supply for the aspira-tor. Connect the air supply to the aspirator air supply inlet but DO NOT turn on the air until the sensor is at operating temperature. If the analyzer is turned off or the process is shut down, the aspirator air should be turned off to avoid plugging problems.

If used, interconnect the AMEVision Display User Interface and the sensor per the appropriate interconnecting drawing.

- The specified cable must be used in order to ensure proper operation of the systems.

Connect AC mains supply to the sensor. Make sure that the AC voltage is within specified limits.

Page 35: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

System Overview | 2-15

Common Operator Errors

If you follow the steps below, your analyzer will operate with a minimum of maintenance and troubleshooting.

Connect the air supply to the aspirator air supply inlet but DO NOT turn on the air until the sensor is at operating temperature. If the analyzer is turned off or the process is shut down, the aspirator air should be turned off to avoid plugging problems.

If your process is running and the analyzer is installed on the process, the analyzer must have power applied to it to prevent plugged plumbing and sen-sor component damage. The case heaters must be on if the process is running to maintain sensor plumbing above 200°C.

Do not use pipe dope or any other contaminant that gives off combustible vapor on any joints of the sample tubing.

For O2 calibration gases, do not use calibration gases to check Thermox ana-lyzers if they contain a mixture of oxygen and combustibles. Note, however, that this is acceptable for combustibles span gases (used only if you pur-chased the combustibles option).

Always introduce calibration gases at the recommended flow rate of 3.0 scfh.

When working on the plumbing inside the sensor cabinet, turn the power off.

Use caution if performing maintenance on the sensor while the process is running. Removing any part of the sensor can allow process gases and gases of high temperature to escape into the sensor.

Do not handle the cell excessively. Do not try to clean the cell except by rinsing. If you need to handle the cell, grasp by touching the seal fitting at the top; never touch the bare part of the cell.

Do not remove a cell or type “K” thermocouple that you may want to use again when the inside of the furnace is still hot - severe thermal shock can be destructive to either of them.

Always replace the metal cell O-ring when replacing the oxygen cell.

Page 36: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

2-16 | Thermox WDG-V / VC / VCM

Technical Support

AMETEK/Thermox is committed to providing the best technical support in the industry. If you need service or application assistance, please call AMETEK at (412) 828-9040, or your local AMETEK/Thermox representative.

Before you call the factory for technical support, run test gases and record the following values (you may be asked by the factory to provide this information when receiving service):

Cell millivolts

Thermocouple millivolts

Cell temperature

If you need to return equipment, you will be asked to provide the following infor-mation before obtaining a Return Material Authorization (RMA) number.

Billing and shipping address

Model number

Serial number

Purchase order number

Telephone number

Before returning material, you must get an RMA number from the factory.

NOTE

Page 37: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Installation Guide | 3-1

INSTALLATION GUIDE

The operations in this chapter should be performed only by qualified service personnel experienced in electrical safety techniques. Never service the sensor unless power has been removed from the ensor, and the sensor has been allowed to cool for at least one hour. Also, always use gloves when working on the sensor.

Do not turn on the aspirator until the sensor has been turned on and is hot - preferably 24 hours. If you turn off the analyzer or the process is shut down, turn off the aspirator air to avoid plugging problems!

NOTE

Page 38: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

3-2 | Thermox WDG-V / VC / VCM

Analyzer Location

Observe the following guidelines when selecting an analyzer installation loca-tion:

1. Select a readily accessible position for the analyzer to allow for routine main-tenance. Comfort levels for maintenance personnel should be considered in placement of the sensor and AMEVision Display User Interface.

2. The installation location should be free from excessive vibration and the ambient temperature is required to be within the limits listed in the specifica-tions. If the ambient temperature is outside the specified limits or the vibra-tion is excessive, please contact Thermox Sales or Service Department at (412) 828-9040. We have special options to address ambient temperatures outside the listed specifications.

This chapter shows you how to install your WDG-V sensor components and AMEVision Display User Interface, and includes the following sections:

• Inspection of Shipping Contents• Mechanical Installation• Wiring

If you have an oxygen-only analyzer without the combustibles option, the combustibles and methane options described in this chapter do not apply. These combustibles and methane sections are clearly identified. Likewise, if you have the combustibles option, but not the methane option, you can skip all sections pertaining to methane.

Inspect Shipping Contents

Remove any packing material from the WDG-V sensor. Check for damage. If any is found, notify the shipper.

Page 39: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Installation Guide | 3-3

Mechanical Installation

This section describes how to mount and connect gases to your WDG-V sensor. This includes the following:

• Sample Inlet Probe Installation

• Probe Heater Installation (optional)

• Exhaust Tube Installation (optional)

• Sensor Mounting

• Remote Calibration Unit Mounting and Plumbing (optional)

Sample Inlet Probe Installation

For some applications, you must first connect the sample inlet probe to the sensor before mounting the sensor. The same applies if you are installing the probe heater or exhaust tube. Therefore, these subjects are discussed before the sections on how to mount the sensor to the process.

The sample inlet probe is how the sample gas gets to the sensor. Aspirator air within the sensor is used to pull process gas from the sample probe into the sen-sor. The sample inlet probe is identified by the threads on both ends.

• Figure 3-1 shows how to install the standard 1/8” NPT probe.

• Figure 3-2 shows how to install a ceramic or long probe.

• Figure 3-3 shows port locations on the rear of the sensor.

You can connect the probe to the sensor before mounting the sensor unless the probe is either very long or made of ceramic. In that case you should mount the sensor first and then install the probe.

NOTE

Page 40: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

3-4 | Thermox WDG-V / VC / VCM

Standard 1/8” NPT probe installation

Connect a standard 1/8” NPT sample inlet probe to the sensor as follows:

• If your probe is 4 feet or less, simply screw the probe into the back of the sample probe port on the rear of the sensor, then mount the sensor - see Fig-ure 3-1.

Exhaust extension3/8"T x 10-3/8"Lg(optional). Threadedextension screwsinto fitting.

Backplate

Insert PlateBushing3/4"NPT(F) x 3/8"NPT(F)(NA for long probes)

Fitting3/8"P x 3/8"Ttapped 1/8"NPT(NA for long probes)

Probe Inlet Hook

TOP VIEW

Probe Support54-1/8"longAssembly replacesbushing for probes5', 6 and 8'

Standard Probe lengths3', 4', 5', 6' and 8'Threaded probes screwinto fitting

Figure 3-1. WDG-V Probe installation.

Page 41: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Installation Guide | 3-5

Ceramic probe or long probe installation

If you ordered a long probe or a ceramic probe, connect the probe to the sensor as follows:

1. Mount the sensor. See the “Sensor Mounting” section later in this chapter for help. Note, however, that if you are installing the probe heater or exhaust tube, you must connect these items to the sensor before mounting the sensor.

2. Remove the probe inlet hook from the sensor as shown in Figure 3-2. To do this, loosen the fitting connecting the inlet hook to the aspirator block and pull the hook out of the sensor.

3. Unscrew the bushing that was plugged into the sensor inlet port and discard. The probe comes with its own bushing and fitting.

4. Thread the sensor probe into the sensor inlet port on the sensor backplate. Do not let the probe hit against the Thermox-provided pipe nipple as you insert the probe into the process. If inserting a ceramic probe, you must slowly insert it into a hot process to prevent the probe from cracking due to thermal shock.

5. Reconnect the probe inlet hook to both the sensor backplate and the aspirator block and tighten.

BackplateBushing

Probe Inlet Hook TOP VIEW

Aspirator Block

Figure 3-2. WDG-V ceramic probe installation - top view.

Page 42: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

3-6 | Thermox WDG-V / VC / VCM

Probe Heater Installation

Do not install the probe heater if your sample gas contains a poten-tially explosive mixture of combustibles; the probe heater can heat the flue gas to the point of ignition.

The probe heater must be connected to the sensor before mounting the sensor to the process.

For installations where a standard setup allows the sample gas to cool below its acid dewpoint before reaching the sensor, you can attach the optional probe heat-er to the probe heater port on the backplate of the sensor to heat the sample gas - see Figure 3-3. This will prevent plugged plumbing due to condensation of the sample gas. For example, when a fuel containing only a few thousand parts per million of sulfur is burned with a high level of excess oxygen (over 4 percent), the resulting sulfuric acid mist may condense at temperatures as high as 175°C.

The probe heater comes shipped in an envelope, is 8.5” long, has a 3/8 NPT bushing and wires leading out of it.

Sample ProbeInlet Port

Exhaust Port

Probe Heater Port(Remove Plug)

Terminal Box

Figure 3-3. Port locations on rear of sensor - front view.

NOTE

Page 43: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Installation Guide | 3-7

Follow the steps below to install the probe heater:

1. Remove and discard the plug from the probe heater port as shown in Figure 3-3. This port is plugged during shipment and should remain plugged if no probe heater is installed.

2. Screw the probe heater into the 3/8” NPT probe heater port.

3. Route the wires from the probe heater to the electronics enclosure.

4. Remove the electronics box cover and connect the wires to an AC power source.

Exhaust Tube Installation

If installing an exhaust tube, you must connect it to the sensor before mounting the sensor to the process.

An optional stainless steel exhaust tube is available (PN 70619KE), which allows you to extend the aspirator exhaust.

To install the exhaust tube, thread the exhaust tube into the exhaust port on the rear of the sensor as shown in Figure 3-3.

Sensor Mounting

• The sensor ambient temperature range is -25°C to 65°C.

• For purged units, the sensor ambient temperature range is -20°C to 60°C.

• When installing a probe heater or exhaust tube along with the sensor, attach it to the back of the sensor before mounting the sensor to the process.

Do not open a Z-purge, hinged-sensor enclosure door until you first verify that the area has been classified as nonhazardous.

NOTE

Page 44: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

3-8 | Thermox WDG-V / VC / VCM

Methods for mounting the sensor

There are three methods for mounting a WDG-V sensor to the process:

• Flush with the process,

• Using a customer-supplied pipe nipple,

• Using a customer-supplied flange.

Each of these methods is explained in the sections that follow. The flush mount method is preferred because it positions the sensor as close to the process as pos-sible and minimizes the chances of the sample gas cooling below its dewpoint.

Mounting dimensions for the hinged style enclosure, including the Z-purge op-tion, are provided in Figure 3-4.

If the protection nipple provided by AMETEK/Thermox extends past the inside of the refractory wall on a process that has a gas tem-perature over 1200°F (650°C), the nipple should be cut down before installation in order to remain flush with the inside wall.

NOTE

Page 45: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Installation Guide | 3-9

5.

2313

2.95

3.80

96.5

9

2.5163.86

18

.61

472.

59

3.6492.56

InstrumentAir Inlet

MeteringValve

CheckValve

ExhaustRestrictor

PressureGauge

Terminal Box

CL

`

HingeSide

4.39111.44

8.48

215.

50

Probe

Figure 3-4. Mounting dimensions - Hinged sensor.

A 7/16” wrench is required to open both doors to the sensor.

NOTE

Page 46: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

3-10 | Thermox WDG-V / VC / VCM

Sensor flush mount (Preferred Method)

Weld the 7” x 7 5/8” (17.78 cm x 19.37 cm) mounting plate (P/N 70626SE) sup-plied with the sensor to the process wall over a 3 3/4” (9.52 cm) diameter hole as shown in Figure 3-5. Then bolt the sensor to the mounting plate. This method positions the sensor as close to the process as possible and minimizes any chance that the sample gas will cool below its dewpoint.

Sensor bolted to mounting plate afterplate is welded to wall by customer.

Gasket (by Ametek)

Mounting plate supplied by Ametekwith 2 7/8"OD x 9"long nipple (7.30cmx 22.86cm) welded to wall over 3 3/4"od(9.52cm) hole by customer.

Process wall

Probe

SIDE VIEW

Single stud at12 o'clock position

Proper orientation formounting plate to flange

Figure 3-5. WDG-V sensor flush mount (preferred).

Customer-supplied pipe nipple sensor mounting

If you can’t mount the sensor flush with the process as described in the previous section (for example, the process wall is curved sharply, covered with insulation, or obstructed with reinforcing members) you can instead:

1. Weld a short 3 1/2” diameter pipe nipple to the process wall and make sure to extend the nipple through the skin of the process (see Figure 3-6).

Page 47: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Installation Guide | 3-11

Sensor bolted to mounting plate afterplate is welded to nipple by customer.

Gasket (By Ametek)

Process wall

Probe

Required 1" (25.4cm) minimuminsulation around exposed portionof nipple

Pipe nipple(supplied by customer)

SIDE VIEW

Mounting plate (supplied by Ametek)1 7/8"OD with 9"long nipple (7.30cmx 22.89cm) welded over 3 3/4"OD(9.52cm) hole by customer.

Single stud at12 o'clock position

Proper orientation formounting plate and flange

Figure 3-6. Customer-supplied pipe nipple sensor mounting.

2. Weld the 7” x 7 5/8” (17.78 cm x 19.37 cm) supplied plate to the end of the nipple, taking care to center it over the hole

3. Wrap the nipple with at least 1” (2.54 cm) weatherproof insulation and heat trace it.

4. Then bolt the sensor to the plate.

If mounting the sensor using this method, you can also install the probe heater to minimize the possibility of the sample gas cooling below its dewpoint.

Page 48: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

3-12 | Thermox WDG-V / VC / VCM

Flange sensor mounting method

You can also mount the sensor to a 3” x 300# customer-supplied flange. In this case, you will not need the AMETEK-supplied mounting plate with attached protective pipe nipple. If you have a different type of flange, AMETEK supplies a number of flange adapters. SB

MOUNTING & DIMENSIONSWDG-V PURGE SENSOR

ACK

BORIGINAL ISSUE

NTS 12

A

B B

A

ORIG. ISSUE

ENGR. APPR.

DOC. CONT.

REVDWG. NO.SIZE

DRAWN BY

MATERIAL

150 FREEPORT ROADPERMISSION BY AMETEK P & AI DIVISION

DATE

2

10-8-13

ADD MISSING PARTS

1

CAM

SB06-11-13

MG

THERMOX

B06-11-13 202-645-310-8-13

10-14-13 MG

REVISION DESCRIPTIONREV. DR APPR

PROCESS & ANALYTICAL INSTRUMENTS DIVISION

AMETEK

SURFACE ROUGHNESS 125

PRODUCT LINE

1

1

A

PROCESS & ANALYTICAL INSTRUMENTS DIV.

PITTSBURGH, PENNSYLVANIA 15235 USA

OFSHEETSCALE

CONFIDENTIALREPRODUCTION PROHIBITED WITHOUT

.005

MATERIALS

SB 06-11-13

DIMENSIONS ARE IN INCHESTOLERANCES:FRACTIONAL 1/32"ANGULAR: MACH 30' TWO PLACE DECIMAL .010THREE PLACE DECIMAL

SEE BILL OF

CAM

CUSTOMERS 3"-300# FLANGE (REDUCINGFLANGE IF REQUIRED) CONNECTED TO2 1/2" MINIMUM SCH 40 PIPE ATTACHED

TO WALL BY CUSTOMER.

GASKET BY THERMOX

3"-300# FLANGE MOUNTING

PROCESS WALL

1" MIN INSULATION AROUNDEXPOSED PORTION OF NIPPLE

Figure 3-7. WDG-V sensor flange mount.

Page 49: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Installation Guide | 3-13

Calibration/Aspirator Air

Required calibration gases and tubing

See Chapter 2 for information on calibration gases. The span gas is the high gas; the zero gas the low gas.

The span gas must be 10 times greater than the zero gas. For example, if the zero gas is 1% O2, the span gas must be 10% O2 or higher.

O2 Span Gas Instrument air (20.9%) or from 1.0% to 100% O2.

O2 Zero Gas From .1 to 10% O2, balance N2

Combustibles Span Gas If using a WDG-VC analyzer, you must also have a combustibles span gas as

follows:

• 60 to 80% of the full combustibles operating range in certified mixtures of CO, with an equal amount of O2, balance N2.

For example, if using an operating range of 0 to 2000 ppm combustibles, use 1200 ppm CO, 1 to 2% O2, balance N2.

Methane Span Gas

If using a WDG-VCM analyzer, you also must have a methane span gas as follows:

• 2% CH4, 8 to 10% O2, balance N2.

NOTE

Page 50: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

3-14 | Thermox WDG-V / VC / VCM

Z-Purge Connections (Optional)

Connect instrument air to the sensor Z-purge inlet. Maintain the pressure and flow as directed on the Z-purge warning label.

• You must provide a protective gas supply with an alarm for a loss of pressure indication in order to meet regulatory requirements.

• You must not exceed the maximum temperature rating listed in the specifica-tions.

• Process pressure is assumed to always be subatmospheric (vacuum) unless the High-pressure option is installed.

• The protective gas supply shall be equipped with an alarm to indicate failure of the protective gas supply to maintain the minimum pressurized enclosure pressure.

• The indicator shall be located for convenient viewing.

• The indicator shall indicate the enclosure pressure.

• The sensing point for the indicator shall be located to take into account the most onerous conditions of service.

• There shall be no devices between the pressurized enclosure and the protec-tive gas supply alarm other than an isolating valve and/or a pressure of flow controlling mechanism.

• Any isolating valve shall - Be marked WARNING - PROTECTIVE GAS SUPPLY VALVE - FOLLOW INSTRUCTIONS BEFORE CLOSING - Be capable of being sealed or secured in the open position. - Have an indication of whether it is open or closed. - Be located immediately adjacent to the pressurized enclosure. - Be used only during servicing of the pressurized enclosure.

• Any pressure or flow controlling mechanism, if adjustable, shall require a tool to operate it.

• No filters shall be fitted between the pressurized enclosure and the protective gas system alarm.

• No isolating valve shall be fitted between the indicator and the pressurized enclosure.

Page 51: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Installation Guide | 3-15

• The WDG-V sensor contains two flame arrestors to prevent it from being an ignition source to the process for short periods of high combustible levels (25% of LEL - lower explosive limit) in the process. The flame arrestors are not, however, intended to protect the process where the combustibles levels are always high.

Do not install the probe heater if your sample gas contains a poten-tially explosive mixture of combustibles; the probe heater can heat the flue gas to the point of ignition.

Z-Purge Shutdown Procedure

Hot internal parts are above the ignition temperature of combustible gases. Power must be disconnected from the sensor for 90 minutes while maintaining purge air flow before the door is opened, unless the area is demonstrated to be nonhazardous. The Z-purge apparatus provided by Thermox meets NFPA 496. Power must not be restored after the enclosure has been opened until the enclosure has been purged for 65 minutes at a pressure of 0.4” of water.

Always use tubing that is free of oil and dirt. Stainless steel tubing is recom-mended.

Figure 3-8. Z-purge setup.

Page 52: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

3-16 | Thermox WDG-V / VC / VCM

Purge Air Inlet1/4"T Compression

Fitting

InstrumentAir Inlet

1/4"NPT (F)

Calibration Gas Inlet Port1/4"T Compression Fitting

Figure 3-9. Sensor inlet connections.

Page 53: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Installation Guide | 3-17

Manual calibration and aspirator air connections

If you have a remote calibration unit (RCU), skip the remainder of this section and proceed to the “RCU Mechanical Installation” sec-tion.

Aspirator air connection

If you have an AMEVision Display User Interface, you can use the flow gauge to set the flow as described in Chapter 5.

• Connect aspirator air to the aspirator air inlet on the sensor (see Figure 3-9). The recommended initial aspirator pressure setting is 3 psi. This will vary depending on the pressure/vacuum of the process for which the WDG-V is installed. If you purchased an AMEVision Display User Interface, you can use the Flow Sensor output diagnostic screen to set the flow to the ideal rate. If you do not have an AMEVision Display User Interface, it is recommended to adjust the aspirator until you get a response from the analyzer, then in-crease the pressure by 0.25 to 0.5 psi. Do not turn on the aspirator until the sensor has been turned on and is hot (preferably 24 hours from a cold start; 1 hour after a restart). See page 5-1 for furter detail.

Calibration gas connection

• A calibration gas inlet port is provided to allow you to calibrate the system as shown in Figure 3-9. During calibration, you turn off the aspirator air sup-ply and inject your calibration gases into the calibration gas inlet port on the sensor. During normal operations, the calibration gas inlet port should be plugged.

NOTE

NOTE

Page 54: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

3-18 | Thermox WDG-V / VC / VCM

Remote Calibration Unit Mechanical Installation

Remote calibration unit (RCU) mechanical installation includes the following:

• Mounting the RCU.

• Plumbing calibration gases to the RCU.

• Plumbing the RCU to the sensor.

This section is divided into instructions for standard RCU and combustibles RCU mechanical installation. Go to the section that applies to the type of RCU you purchased.

Do not turn on the aspirator until the sensor has been turned on and is hot - preferably 24 hours. Also, if you turn off the analyzer or the process is shut down, turn off the aspirator air to avoid plugging problems!

Standard RCU mechanical installation

Figure 3-10 shows mounting dimensions for a standard RCU. Use two #10 screws to mount the RCU. Note from Figure 3-10 that only the top left and lower right holes are used; the other holes hold the RCU to its mounting plate. The ambient temperature range for the RCU is -20° to 70°C. Mount the RCU as close to the sensor as possible. Shorter calibration plumbing improves response times, reduces calibration gas expense, and reduces the chance of contaminants in the calibration gas plumbing.

Page 55: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Installation Guide | 3-19

Figure 3-10. RCU mounting dimensions - O2-only.

Page 56: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

3-20 | Thermox WDG-V / VC / VCM

Standard RCU

Make the following calibration gas connections to the standard RCU. Figure 3-10 shows standard RCU calibration gas connections:

1. Connect instrument grade air that is free of oil or dirt to the instrument air inlet connection on the left side of the RCU. If you are also using instrument air as your span gas, you can skip Step 2 below because instrument air will be used for both aspirator air and the span calibration gas.

2. If you wish to use a span gas other than instrument air, you must purchase the optional alternate span gas kit. Connect the span gas to the alternate span gas inlet on the RCU if your span gas is other than instrument air.

3. Connect the O2 zero calibration gas to the O2 zero gas inlet connection on the RCU.

4. Connect the calibration gas outlet on the right side of the RCU to the cali-bration gas inlet on the sensor -see Figure 3-9 for the calibration gas inlet connection on the WDG-V sensor. Also, be sure to install the supplied check valve as close to the sensor as possible - note the orientation of the valve.

5. Connect the aspirator air outlet on the RCU to the aspirator air inlet connec-tion on the sensor.

Combustibles RCU

Figure 3-11 shows mounting dimensions for the combustibles RCU. Mount the RCU as close to the sensor as possible. Shorter calibration plumbing improves response times, reduces calibration gas expense, and reduces the chance of con-taminants in the calibration gas plumbing. The ambient temperature range for the combustibles RCU is -20° to 70°C.

Make the following calibration gas connections to the combustibles RCU.

Page 57: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Installation Guide | 3-21

COMBUSTIBLES (SPAN 2)CALIBRATION GAS INLET

1/4” COMPRESSION FITTING

Figure 3-11. RCU mounting dimensions - Combustibles.1. Connect the O2 span gas, if other than instrument air, to the alternate O2 span

gas inlet.

2. Connect the O2 zero calibration gas to the O2 zero gas inlet connection on the RCU.

3. Connect the combustibles span calibration gas to the combustibles span gas inlet on the RCU. For a WDG-VCM analyzer with RTD-type combustibles detector, the combustibles span calibration gas would also include the meth-ane span calibration concentration.

4. Using appropriate tubing, connect the calibration gas outlet on the right side of the RCU to the calibration gas inlet on the sensor - see Figure 3-8 for sen-sor calibration gas inlet connection.

5. Using appropriate tubing, connect the aspirator air outlet on the right side of the RCU to the aspirator air inlet connection on the sensor.

Page 58: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

3-22 | Thermox WDG-V / VC / VCM

Figure 3-12. RCU flow diagram - Single Span.

Figure 3-13. RCU flow diagram - 2 Span - Combustibles.

Page 59: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Installation Guide | 3-23

Figure 3-14. RCU plumbing - 3 Span - Combustibles & Methane.

Page 60: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

3-24 | Thermox WDG-V / VC / VCM

Wiring

Remove AC mains power from the controller before performing wiring.

Connections to the sensor unit are made through the wiring card. The wiring card is located on the front-bottom of the sensor unit.

Any screw terminals on the wiring card not described in this section are reserved for future use.

This wiring section shows you how to make the following connections:

• AC mains supply wiring to WDG-V sensor

• WDG-V unit to remote calibration unit

• WDG-V unit to alarm devices

• WDG-V unit to current output devices

• WDG-V unit to AMEVision Display User Interface (RS-485 communica-tions)

• AC mains supply wiring to AMEVision Display User Interface

In addition, this wiring section provides mandatory EMC grounding, shielding, and noise protection requirements.

NOTE

Page 61: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Installation Guide | 3-25

(4) 3/4"NPTConduit EntriesSupplied with

Nema Conduit Plugs42454JE

(2) 1/2"NPTConduit EntriesSupplied with

Nema Conduit Plugs42464JE

SS ExternalGround StudM5 Ring Terminal12-10AWG

General Wiring and Conduit Requirements

This section describes general wiring and conduit requirements.

• Sensor wiring conductors must be rated at a minimum of 80°C. All other wiring conductor ratings should be for the minimum temperature required for the equipment being connected to the analyzer, but not less than 60°C.

• Use only the applicable NEMA-approved conduit fittings or cable fittings to maintain the NEMA rating for the sensor enclosure or WDG-V. If not using a conduit entry, leave the factory NEMA-approved plugs intact. Never leave any holes unplugged.

• Follow all applicable electrical codes for your location.

• Follow proper grounding, shielding and noise protection practices as de-scribed in this section.

• For all sensor and signal wiring use twisted-pair cable, 18 to 22 AWG (0.82 mm2 to 0.33 mm2 ), with an overall braided shield, or twisted-pair cable in rigid metal conduit.

• For AC mains supply wiring, use between 12 and 14 American Wire Gauge (AWG) or equivalent metric between 3.3 mm 2 and 2.1 mm2.

• Use the conduit entry point closest to the connections you are making. Do not add any additional conduit entry holes!

Page 62: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

3-26 | Thermox WDG-V / VC / VCM

WDG-V Mains Supply Connections

Do not run control unit AC mains supply wiring in the same conduit with other AC line power wires. By keeping this wiring separated, you prevent transient signals from reaching the control unit.

The WDG-V can operate using between 115 VAC, ± 10% or 230 VAC, ± 10%. There is no power switch or circuit breaker on the sensor, and it must be protected by installing it on a circuit-protected line, maximum 15 amperes, with a switch or circuit breaker in close proximity to the control unit and within easy reach of an operator. Mark the switch or circuit breaker as the control unit disconnecting device.

Mains supply connections to the control unit are as follows:

L - Line connection

N - Neutral connection (USA)

Chassis Stud - Equipment ground (protective conductor)

Use the 1/2” conduit entry hole in the WDG-V for AC mains supply wiring. Use the chassis stud next to the 1/2” conduit entry hole for equipment ground (protec-tive conductor).

AC (L) and (N) markings are provided by the terminal block for connection of AC power. These markings are for reference purposes only, such as for use on system wiring diagrams, etc. The system/product has or needs no specific LINE or NEUTRAL connection for any function, safety or otherwise. The (N) terminal is not internally grounded, nor needs to be. The system will operate normally re-gardless of what AC input terminal (L or N) the AC Line or Neutral is connected to, or, if there is a Neutral used at all (i.e., 208 VAC US power connection).

Page 63: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Installation Guide | 3-27

EMC grounding, shielding, and noise protection

For EMC purposes, under no circumstances should you leave cable shields disconnected at one end or both ends of the cable (sensor or control unit or other device).

You must use twisted-pair cable in rigid metal conduit or use twisted pair cable with an overall braided shield. All cable shields or conduits connecting to the sensor unit must be chassis grounded.

SHIELD RING METHOD

GROUND STUD METHOD

Shield Terminal Ring

Conduit Housing Cable Shield(s)Max. Length 1 inch

Conduit Nut

Quick Disconnect (1/4”)

Capacitor

Conduit Nut

Cable Shield(s)Max. Length 2 inches

Nut

Washer

Ring Terminal

Conduit Housing

Capacitor

Figure 3-25. Direct shield grounding methods

Page 64: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

3-28 | Thermox WDG-V / VC / VCM

Transient and RFI interference

This section describes transient and RFI interference precautions:

• Although there are transient and noise protectors on all sensor unit I/O con-nections (communications, current outputs, sensor, etc.), this protection is intended to act as a last line of defense against unwanted transient and RFI interference.

• Proper installation practices to prevent the introduction of transients and noise into the system must be followed. Inductive loads connected to the sensor unit must have transient suppressors installed at the inductive loads. Be sure to place the transient suppressor as close to the load as possible. Examples of transient suppressors include MOVs, TRANSORBs, and RC snubbers.

• AC mains supply wiring should not be run in the same conduit with mains supply wiring that feeds heavy inductive loads.

• Avoid running signal wiring in the same cable or conduit with wires that power inductive loads unless all the cables within the conduit are shielded, the inductive loads are small, and transient suppressors are used at the loads.

• Do not run signal lines in the same cable or conduit with high voltage lines.

• For optimum noise protection, AMEVision Display User Interface mains supply wiring should be connected to a circuit separate from any circuit that could introduce transients into the system. As an example, do not run motors, blowers, or air conditioners using the same mains supply circuit or conduit as the sensor unit’s mains supply circuit or conduit.

Page 65: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Installation Guide | 3-29

Current Output Connections

Standard current outputs

There are three current outputs on the WDG-V sensor. The current output con-nections are labeled as follows on the sensor board terminals:

Analog Output #1 => +I1- Analog Output #2 => +I2- Analog Output #3 => +I3-

The current outputs are referenced as Analog Outputs 1, 2 and 3 in the AMEVi-sion Display User Interface and PC software. Be sure to observe the polarity when connecting current output devices to these terminals.

Each of the current outputs are capable of driving up to 1000 ohm loads.

The current outputs can be selected for the following ranges:

0-20mA 4-20mA NAMUR

The NAMUR outputs are implemented as follows:

4-20 mA signal Analyzer Condition0 mA Analyzer unpowered, or completely failed3.5 mA Critical Alarm - analyzer reading unusable (factory default)3.8 mA Reading Under Range (Example: user sets range to 2-10%. Current reading is 1.9%4 to 20 mA Normal Operation20.5 mA Reading Over Range (Example: range is 0-10%. Current reading is 12%

Page 66: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

3-30 | Thermox WDG-V / VC / VCM

Figure 3-26. Wiring

Page 67: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Installation Guide | 3-31

Alarm Contact Connections

This section describes how to make wiring connections for any alarm devices you wish to connect to the control unit.

Standard alarm connections

The WDG-V sensor provides five sets of normally open alarm contacts as fol-lows:

Terminal Block ID Description SVC - Service Alarm DV - Data Valid Alarm ALM1 - Process Alarm 1 (Configurable) ALM2 - Process Alarm 2 (Configurable) ALM3 - Process Alarm 3 (Configurable)

The process alarms are referenced as Relay 3, Relay 4 and Relay 5 in the AMEVision.

Relay 3 - ALM 1 Relay 4 - ALM 2 Relay 5 - ALM 3

Service Alarm - The service alarm indicates that a critical alarm exists and the analyzer needs repair. Readings are not valid when this alarm is active. The relay contacts are normally-open and are OPEN (de-energized) when a critical alarm exists.

Data Valid - The data valid alarm indicates if the concentration readings are rep-resentative of the process. When this alarm is active, concentration readings may not be representative of the process and should not be used for control/safety. Examples of this alarm include: Warm-up mode, in calibration, and diagnostics mode. This alarm differs from the service alarm in that there is not necessarily a problem with the analyzer when this alarm is active. The relay contacts are normally-open and are OPEN (de-energized) when the concentration readings are not valid.

Page 68: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

3-32 | Thermox WDG-V / VC / VCM

Remote Calibration Unit Connections

If you don’t have the remote calibration unit (RCU) option, skip this section.

Oxygen-only remote calibration unit connections

Oxygen-Only Remote Calibration Unit (RCU) connections on the WDG-V wir-ing card, and their RCU connections, are as follows (see Figure 3-27):

ZERO GAS - WDG-V Terminal Z to Pin 13 on RCU

ASPIRATOR - WDG-V Terminal A to Pin 14 on RCU

O2 SPAN - WDG-V Terminal S1 to Pin 15 on RCU

VALVE COMMON - WDG-V Terminal ‘V’ to pin 16 on RCU

Combustibles remote calibration unit connections

If you have the combustibles option (and a combustibles RCU), you must also make the following connection (see Figure 3-27):

COMBUSTIBLE SPAN - WDG-V Terminal S2 to Pin 17 on RCU

For a WDG-VCM with hot-wire detector, Remote Calibration Unit connects:

METHANE SPAN - WDG-V Terminal S3 to Pin 18 on RCU

Page 69: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Installation Guide | 3-33

Remote calswitch normally

open (drycontact closure)

All common wiresto pin 16 on RCUand valve commonson control unit

Overall braided shieldor rigid metal conduit

CIN

VA

LVE

CO

M

A SPA

N 3

SPA

N 4

Z SPA

N 1

VA

LVE

CO

M

SPA

N 2

Figure 3-27. RCU connections.

Digital input to initiate remote calibration unit

The digital input connections on the wiring card allow you to initiate a remote calibration from a remote location. For this option to work, you must have a re-mote calibration unit. The system monitors the digital input, and when the switch closes, it sends the system into an automatic calibration. The switch you connect to this digital input must be a normally open switch.

Digital input connections are labeled as follows on the wiring card:

CIN

Page 70: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

3-34 | Thermox WDG-V / VC / VCM

Notes About Ducting

Ducting Between Pressurized Enclosure and Inlet The intake ducting to a compressor should not normally pass through a hazard-ous area. If the compressor intake line passes through a hazardous area, it should be constructed of non-combustible material and protected against mechanical damage and corrosion. Adequate precautions should be taken to ensure that the ducting is free from leaks in case the internal pressure is below that of the external atmosphere (see Annex C). Additional protective measures, for example, combustible gas detectors, should be considered to ensure that the ducting is free of flammable concentrations of gas vapor. Additional Purge Time to Account for Ducting The purge duration should be increased by the time necessary to purge the free volume of those associated ducts which are not part of the equipment by at least five times their volume at the minimum flow rate specified by the manufacturer. Power for Protective Gas Supply The electrical power for the protective gas supply (blower, compressor, etc.) should be either taken from a separate power source or taken from the supply side of the electrical isolator for the pressurized enclosure. Enclosure Maximum Overpressure The user should limit the pressure as specified by the manufacturer.

Page 71: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

AMEVision Display User Interface or Communications | 4-1

AMEVISION DISPLAY USER INTERFACE OR COMMUNICATIONS

Purpose

The AMEVision Display User Interface provides a remote, color, graphical user interface as well as a communications suite for the WDG-V. The AMEVision Display User Interface provides the ability to easily configure, calibrate and monitor up to four WDG-V sensors from a single unit. The interface between the AMEVision Display User Interface and the WDG-V sensor is two-wire MOD-BUS RTU.

The AMEVision Display User Interface provides the following communication interfaces:

• MODBUS RTU

• Web Interface (TCP/IP)

• USB (Flash Drive)

Additional features:

• 4.2” Color, 1/4 VGA Display

• 19 Key Keypad

• Intuitive Menus

• Alarm and Event logging

• Data logging and trending. Data can be graphed on the local display and/or uploaded via USB flash drive

• Upgradable firmware for attached sensors and/or AMEVision Display User Interface USB flash drive

• Calibration history display and plot

• Sensor Diagnotics

Page 72: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

4-2 | Thermox WDG-V / VC / VCM

Figure 4-1. AMEVision Display User Interface

Reference the AMEVision Display User Interface manual for more information.

Enter/ Menu Key

Navigation Buttons

Color 1/4 VGA

Help

Cancel/Escape

Number Pad

Page 73: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

AMEVision Display User Interface or Communications | 4-3

Setting the Analyzer Address

The interface bewteen the AMEVision Display User Interface and the WDG-V sensor is a two-wire MODBUS RTU. Up to four WDG-V sensors can be con-nected to a single AMEVision Display User Interface unit. Each attached sensor must have a unique address. The address on the WDG-V sensor is set via the dip switch on the front of the electronics PCB.

Figure 4-2. Dip Switch Locations

321 Switch Address001 1010 2011 3100 4

1 = ON (UP)0 = OFF (DOWN)

Page 74: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

4-4 | Thermox WDG-V / VC / VCM

Connecting a Single WDG-V Sensor to the AMEVision

When only one sensor is connected, the WDG-V address should be set to “1”.

Connecting Multiple WDG-V Sensors to the AMEVision

Up to four sensors can be connected to a single AMEVision Display User Inter-face unit. The WDG-V sensor provides extra terminal connections to facilitate easy daisy chain connections.

Each analyzer must have a unique address (reference address setting section).

NOTE

NOTE

Figure 4-3. Single WDG-V Sensor to the AMEVision Display User Interface

Figure 4-4. Multiple WDG-V Sensors to the AMEVision Display User Interface

Page 75: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

AMEVision Display User Interface or Communications | 4-5

Page 76: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

4-6 | Thermox WDG-V / VC / VCM

When connecting multiple sensors in a daisy chain configuration, the com-munication termination resistor should be enable on the last sensor in the chain (farthest from the AMEVision Display User Interface). The termination enable is controlled by switch SW-5 of the dip switch on the front of the electronics PCB.

SW-5 on: Termination Enabled

SW-5 off: Termination Disabled

The maximum distance between the AMEVision Display User Inter-face and the WDG-V analyzer is 4000 feet (1219.2 meters).

Troubleshooting

The interface between the WDG-V and the AMEVision Display User Interface is simple and robust. However, if you experience connectivity issues, here are some of the possible causes:

No Communications:

• Polarity reversed

• Address not set correctly

• Maximum distance exceeded

• Broken connection

Intermittent Communications:

• Poor wiring (i.e. lack of shielded twisted pair)

• Termination enabled on more than one unit (on multi-unit systems)

• Termination not enabled (on multi-unit systems)

• Maximum distance exceeded

NOTE

Page 77: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

AMEVision Display User Interface or Communications | 4-7

WDG-V PC Configurator Software

Overview

The WDG-V PC Configurator Software provides a convient way to configure, calibrate and monitor a single WDG-V analyzer. The interface between the con-figurator software and the WDG-V sensor is two-wire MODBUS RTU.

Connecting to the WDG-V

Using the optional USB to RS485 converter (P/N 1000-724-VE), attach as shown.

The address must be set per the Settings >> Address section.

An AMEVision Display User Interface and PC software cannot be simultaneously connected to the analyzer.

NOTE

1000-724-JE

WDG-VTERMINAL

MB+MB-

1000-724-JECONNECTOR

T/R+T/R-

USB TO COMPUTER

Note: Only one connection can be made to the sensor. The AmeVision and Configurator cannot be connected simultaneouslt.

Page 78: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

4-8 | Thermox WDG-V / VC / VCM

This page intentionally left blank.

Page 79: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Flow Setting | 5-1

FLOW SETTING

Setting the Flow

The recommended initial aspirator pressure setting is 3 psi. This will vary de-pending on the pressure/vacuum of the process for which the WDG-V is in-stalled. If you purchased an AMEVision Display User Interface, you can use the Flow Sensor output diagnostic screen to set the flow to the ideal rate. If you do not have an AMEVision Display User Interface, it is recommended to adjust the aspirator until you get a response from the analyzer, then increase the pressure by 0.25 to 0.5 psi.

Setting the flow using the AMEVision Display User Interface Flow Sensor Diagnostic

The internal flow sensor output is displayed is percent to ideal sample loop flow. For example, a reading of 100% indicates ideal flow whereas a reading of 50% indicates half the ideal flow rate. The update interval for the flow sensor is 90 seconds, so you must wait at least 90 seconds for the flow output value to change. It is recommended to operate the sensor between 80% and 120%. A low flow alarm is generated when the flow is 45% or lower.

To set the flow, first adjust the aspirator pressure regulator until you get a re-sponse from the analyzer. Wait 90 seconds and check the flow sensor output. Adjust the aspirator as necessary until the flow sensor reads approximately 100%. Small increments of 0.25 psi are recommended.

Page 80: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

5-2 | Thermox WDG-V / VC / VCM

This page intentionally left blank.

Page 81: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

LED Status Indicator | 6-1

LED STATUS INDICATOR

Main Status LED (Light Pipe Visible From the Front)

OFF – Power is off.

The status LED is NOT directly connected to the WDG-V power supply. In the unlikely event of an error condition, it is possible for the analyzer to be energized with dangerous voltages even though the status LED is off. Always verify power has been removed before servicing the instrument.

GREEN – Analyzer is in normal operating mode.

YELLOW – Analyzer is in warm-up state or diagnostic state.

RED – Analyzer has a service alarm and is in need of attention.

Figure 6-1. Analyzer is in normal operating mode.

NOTE

Page 82: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

6-2 | Thermox WDG-V / VC / VCM

Figure 6-2. Analyzer is in warm-up state or diagnostic state.

Figure 6-3. Analyzer has a service alarm and is in need of attention.

Page 83: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

LED Status Indicator | 6-3

Printed Circuit Board (PCB) Status LED’s

There are two small, surface-mount green LED’s on the PCB located on the back left corner of the PCB (when looking at the analyzer). You must look up into the electronics enclosure to view these LED’s, they are not visible when looking directly at the analyzer.

REAR LED – Heart Beat LED. Blinks once a second indicating the analyzer software is operating. If off or not blinking this indicates the power is off or the electronics are not functional.

FORWARD LED – Modbus Communications LED. Toggles upon receipt of a valid Modbus message. This LED will blink rapidly in short bursts when commu-nicating properly to the AMEVision Display User Interface.

Page 84: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

6-4 | Thermox WDG-V / VC / VCM

This page intentionally left blank.

Page 85: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Troubleshooting | 7-1

TROUBLESHOOTING

The operations in this chapter should be performed only by qualified service personnel experienced in electrical safety techniques.

Never service the sensor unless power has been removed from the sensor, and the sensor has been allowed to cool for at least one hour. Also, always use gloves when working on the sensor.

This chapter describes AMEVision Display User Interface and error messages. It also provides troubleshooting assistance. System and error messages scroll on the bottom line of the display at three second intervals until the condition is corrected or has ended. These system and error messages are listed in this chapter alpha-betically to make them easier to locate.

System or error messages will not appear on the display when you are navigating the menus. However a system alarm ICON (SHOW ICON) will appear in the top left hand corner of the screen (show display and icon).

NOTE

Page 86: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

7-2 | Thermox WDG-V / VC / VCM

General Troubleshooting

Your system may pass calibrations, yet still seem to be reading incorrect oxygen levels. If this is the case, you may want to check the following:

Leak Check

Leaks can lead to inaccurate readings, especially if operating under a significant pressure or vacuum.

• Check that all compression fitting and pipe thread connections are leak tight.

• Be sure that the mounting plate or mounting flange gasket on the rear of the sensor is in place (see the “Sensor Mounting” section in Chapter 2 for details on the placement of the gasket).

Sniffing for Leaks

For processes under vacuum you can check for leaks by sniffing the fittings with another gas (for example, nitrogen or pure oxygen), being sure to avoid the area over the top of the cell.

1. Use a piece of tygon or plastic tubing with a 1/8” stainless steel nozzle to ap-ply the gas from a cylinder (using stainless steel prevents any problems that might occur with plastic melting the nozzle on hot sensor components).

2. Monitor the response from the cell. When the cell millivolt reading changes, it indicates a leak in that area of the plumbing (the vacuum of the process pulls in the gas).

3. If not convenient to view the AMEVision display, you can also apply a volt-meter to the Cell terminals on the sensor board labeled (+O2-) to see if the cell millivolts change, indicating a leak.

Pressurizing for Leaks

1. Remove the sensor from the process (after allowing sensor to cool) and pres-surize it with 5 lbs. of air, plugging any exiting ports, the inlet probe, and exhaust tube holes. Then go over the sensor fittings with a leak detector fluid.

2. If you see bubbles, it indicates a leak. If using this method be sure to prevent the liquid from reaching the furnace. If the furnace does get wet, allow it suf-ficient time to thoroughly dry.

Page 87: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Troubleshooting | 7-3

Plugged Plumbing Check

Exercise care when working on the sensor. Allow the unit to cool and wear gloves.

1. Examine the inlet and exhaust for plugging problems. When possible, rod out the probe and exhaust.

2. If this doesn’t solve the problem, disassemble the analyzer to locate the plug. Clean the plumbing using hot water and a bottle brush.

3. When assembly is complete, check for leaks using a leak detecting liquid.

Aspirator Air Not Pulling Sample from Process

The aspirator will need cleaned with the Aspirator Clean Out Tool (P/N 7000-404-KE) per instructions.

Page 88: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

7-4 | Thermox WDG-V / VC / VCM

Diagnostics Checks

This section shows you how to check different sensor areas for possible prob-lems. See the “Error Messages” section for help on what checks you should perform based on the error message displayed. If you don’t see an error message, yet feel your readings are inaccurate, you may also want to check the “General Troubleshooting” section that follows. We recommend monitoring the thermo-couple millivolts and cell millivolts be displayed during troubleshooting as a troubleshooting aid. This information will be helpful should you need to contact the factory for assistance. Be sure to always include your analyzer model and se-rial number when calling the factory for technical support.

Diagnostic checks are broken down as follows:

Wiring Checks

Thermocouple Checks

Calibration/Aspirator Setup Checks

AC Power Checks

Furnace Checks

Process Pressure Checks

Cell Checks

Exercise care when working on the sensor. Turn off power, allow the unit to cool, and wear gloves.

Page 89: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Troubleshooting | 7-5

Thermocouple Checks

Open ThermocoupleRemove power to the sensor. Measure across terminals “+TC-” on the sensor board with an ohm meter. If an open is measured, replace the thermocouple.

Shorted/Failed Thermocouple

Check that the thermocouple leads are not shorted to chassis ground by using an Ohm meter to measure between terminal TC+ on the sensor board and chassis ground, and between terminal TC- on the sensor board and chassis ground. If shorted, replace the thermocouple.To verify the operation of the thermocouple itself, remove the thermocouple from the sensor and heat its ceramic tip to a known temperature. Measure the millivolt output with a proper temperature indicating meter (type K thermocou-ple). If the thermocouple reads inaccurately, replace it.

Reversed Thermocouple Wires

First, view the cell temperature via the Home screen or Diagnostic screen. If the thermocouple leads are reversed, the displayed temperature will be decreasing as the sensor warms up (this will usually happen at start-up or after you replace a thermocouple). This indicates that the thermo-couple wiring is reversed. If you just replaced a thermo-couple, check the leads from the thermocouple to the sensor board: yellow wire connects to terminal TC (+), red wire connects to terminal TC (-).

Page 90: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

7-6 | Thermox WDG-V / VC / VCM

Calibration/Aspirator Setup ChecksCalibration Gas Check

To check calibration gas values:Select the Cal Gas Value menu option from the Cali-brate key on the control unit and check that the calibra-tion gas values entered match the analyzed concen-tration of the cylinders.

Check that the correct calibration gas values have been en-tered into the analyzer. Check that the calibration gas cylinders are turned on and are not empty.Check for the proper flow rate and proper delivery pressure of calibration gases when the remote calibration unit has been activated (see the Inject Cal Gas Diagnostic option in the AMEVision Display User Interface manual for help).

Calibration Line Check

The best calibration gas to use for this check is an O2 zero calibration gas.

Ensure that your calibration line is not contaminated with such things as pipe dope, cutting fluid, oil, or solvents. All these contaminants produce hydrocarbon vapors that interfere with the proper calibration of your analyzer, resulting in lower than expected oxygen readings. To test for contaminated lines, you must temporarily bypass your current calibration line with a clean calibration line (directly from cylinder to the sensor calibration inlet port, using a flow meter to set the proper flow) and compare the response with that from the possibly contaminated line.

Remote Calibration Unit (RCU) Problem

The RCU contains one normally open solenoid-the aspirator air solenoid. All other solenoids are normally closed. Problems with the RCU are usually as follows:

Plumbing Leak To check for plumbing leaks, disconnect power from the RCU and pressurize the inlets. Apply a leak detecting liquid along the base of the solenoids and any plumbing fittings. Repair any leaks found.

Solenoid NotEnergizing

The solenoid drive signal is a 12 VDC signal. This is used to close the aspirator solenoid and open the appropriate calibra-tion gas solenoid. To test a solenoid valve, select Inject Cal Gas from the Calibrate key menu to energize that solenoid. Verify that the solenoid drive signal is present. The aspirator air sole-noid is closed when the drive signal is applied. Other solenoids are open when the drive signal is applied (cal gas flowing). With the drive signal present, verify the proper flow. If no signal is present, check the interconnecting wiring between the control unit and the RCU. If wiring is correct, replace the electronics. If the solenoid drive signal is present, it indicates a problem with one of the solenoid valves. Replace the solenoid valve, being sure that the solenoid O-ring seals are properly positioned.

Page 91: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Troubleshooting | 7-7

To check if a calibration gas solenoid is stuck open:

Verify that no drive signal is present and check for flow on the RCU flow meter. If flow is indicated when no solenoids are ener-gized, a solenoid is stuck open. Shut off your calibration gases (one at a time) until the flow drops to zero. This identifies the defective solenoid valve (replace the solenoid valve, being sure that the solenoid O-ring seals are properly positioned).

Calibration Gas Time Inadequate

If you are having problems running an auto calibration, you may not have allowed the calibration gases enough time to flow through the sensor and stabilize. To correct this problem, select Inject Cal Gas from the Diagnostic menu. Turn on each calibration gas and determine how long it takes for each gas to stabilize on the control unit display. Then add a one minute buffer to each of these times. Also make sure the cylinder regulators are set to the correct pressure. See the Flow Section of the System Interconnect Drawing for the system flow and pressure requirements. This drawing is included with your Installation package.Select Cal Gas Duration from the Calibration menu to set cali-bration gas times.

If performing a manual calibration, ensure that you are waiting for the reading to stabilize on the display before switching to the next calibra-tion gas.

AC Power ChecksLoss/Inadequate AC Voltage to the Sensor

Measure the AC voltage to the sensor board at terminals L and N. Ensure that this voltage is sufficient. Check the measure-ment technique used by the Volt meter (for example, RMS, average, peak, etc.). Specifications are based on RMS mea-surements.

Furnace ChecksOpen Furnace Disconnect power to the control unit and the sensor. With Ohm

meter, measure across Terminals “FURN” on the sensor board. The resistance of the furnace should be:

32 ohms (±10%)If the furnace resistance is not within allowable tolerances, replace the furnace.

Loss of AC Power to the Furnace

Verify the correct line voltage at L and N of the sensor board.With line voltage present, check the voltage at terminals FURN on the sensor board. If voltage is present and the system is not heating, remove power and check the furnace resistance (see “Furnace Checks” section). If no voltage is present at terminals FURN on the sensor board, replace the electronics.

NOTE

Page 92: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

7-8 | Thermox WDG-V / VC / VCM

Process Pressure Checks (Only if above 2 psig)To check that you entered the process pressure correctly, select Process Pressure from the configuration menu. Also ensure that the calibration process pressure equals normal operat-ing process pressure. Calibration should only be performed under these conditions for highest accuracy.

Cell Checks

If the cell fails when you first begin to use the analyzer:

It is likely that there is a leak in the sensor plumbing or an improper calibration gas setup, and there is not a problem with the cell itself (see the “Leak Check” section later in this chapter for help on how to check for leaks; see the “Calibration/Aspirator Air Checks” section earlier in this chapter for help on checking your calibration gas setup).

If the analyzer has been operating for some time and you feel the oxy-gen reading is inaccurate:

First check by running a known calibration gas to veri-fy the analyzer’s response. If the analyzer responds to the calibration gas correctly, it indicates either leaking or plugged plumbing.

If your analyzer doesn’t respond properly to the calibration gas:

This may indicate a problem with the cell. Before re-placing the cell, check for leaks or plugged plumbing.

Page 93: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Troubleshooting | 7-9

Alarm and Warning Messages

ALARMSALARM DESCRIPTION Possible Causes

Cell T/C Failure Thermocouple is not connected or is open Bad T/C, Bad connection

Cold Junction Compensator Failure

The PCB temperature sensor has failed Bad Electronics

Cell Temperature Control A critical overtemperature condition has occurred

Shorted, open, or Bad T/C or Failed Temperature Control Circuit.Unit must be reset to clear this alarm

Cell Over Temp Cell temperature is 4 degrees or higher above the setpoint

Bad Thermocouple, Failed temperature control circuit

Cell Under Temp Cell temperature is 4 degrees or lower below the setpoint

Faulty heater, faulty heater connection, Faulty T/C, Faulty T/C connection, Low Power.Only after warm-up is complete

Cell Temp Rise Failure Cell heater failed to heat the cell Bad Cell furnace heater, Bad heater connection, Shorted T/C, Bad T/C con-nection.Unit must be reset to clear this alarm

Over Temp Relay Tripped The cell over-temperature relay has tripped Open Thermocouple, Over Temperature condition, Failed Temperature control circuit.Must cycle power to reset the relay

RTD Failure The Box Temp RTD Failed Open or Shorted RTD

Box Temp High Box temperature is 4 degrees or higher above the setpoint

Failed temperature control circuit

Box Temp Low Box temperature is 4 degrees or lower below the setpoint

Faulty heater, faulty heater connection, Faulty RTD, Faulty RTD connection, Low Power.Only after warm-up time expires

Box Temp Rise Failure Box temperature failed to heat Bad box heater(s), Bad heater connection, Bad RTD, Bad RTD connection

Page 94: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

7-10 | Thermox WDG-V / VC / VCM

Low Sample Flow Low sample flow detected Loss of aspirator air plugged inlet probe, plugged sample system.Active only after warm-up time expires

Cell Failure (Open) Cell resistance exceeds the normal limit Bad Cell, Open Cell Con-nection

Combustible Detector Open Comb detector is open or not connected Bad Detector, Bad elec-tronics.Only when configured as WDG-VC

Methane Detector Open CH4 detector is open or not connected Bad Detector, Bad elec-tronics.Only when configured as WDG-VCM

Cell Mv Mismatch The cell mv measurement does not match the real time cell meas

Bad A/D circuit (Bad elec-tronics)

Analog Output 1 Error The analog output measurement does not match the setting

Open 4-20mA loop or failed output or bad read-back circuit

Flow Sensor Failure The flow sensor failed (open) Bad flow sensor or circuit

Cell T/C Measurement Mis-match

The Redundant temperature measurement does not match the real time meas

Bad T/C or circuit

Page 95: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Troubleshooting | 7-11

WARNINGSCell Life Nearing Its End The ZrO2 cell is nearing its end of life. This

is based on calibration history.Low cell output

Comb Detector Life Nearing Its End

The combustible detector sensitivity is near its end of life. This is based on calibration history.

Low detector output

Methane Detector Life Nearing Its End

The methane detector sensitivity is near its end of life. This is based on calibration history.

Low Detector output

High Cell Mv The combustible measurement is no longer valid (set to full scale) due to lack of O2

Hydrocarbon upset

Analog Output 1 Out Of Range The analog output value is saturated (high or low)

Improperly set range

Analog Ouptut 2 Out Of Range The analog output value is saturated (high or low)

Improperly set range

Analog Ouptut 3 Out Of Range The analog output value is saturated (high or low)

Improperly set range

Last O2 Span Calibration Failed

The last oxygen span calibration failed Tank Empty, bad solenoid, didn’t wait long enough, Bad Cell

Last O2 Zero Calibration Failed The last oxygen zero calibration failed Tank Empty, bad solenoid, didn’t wait long enough, Bad Cell

Last Comb Calibration Failed The last combustible calibration failed Tank Empty, bad solenoid, didn’t wait long enough, bad detector

Last CH4 Calibration Failed The last methane calibration failed Tank Empty, bad solenoid, didn’t wait long enough, bad detector

Last Flow Calibration Failed The last flow sensor calibration failed Plugged sample system or bad sensor

Oxygen Calibration Required O2 Calibration is required Last Calibration failed or unit has never been cali-brated

Comb Calibration Required Combustible Calibration is required Last Calibration failed or unit has never been cali-brated

Methane Calibration Required CH4 Calibration is required Last Calibration failed or unit has never been cali-brated

Page 96: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

7-12 | Thermox WDG-V / VC / VCM

This page intentionally left blank.

Page 97: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Service and Parts | 8-1

SERVICE AND PARTS

The operations in this appendix should be performed only by quali-fied service personnel with a knowledge of electrical safety tech-niques. Never service the sensor unless power has been removed from the sensor, and the sensor has been allowed to cool for one hour.

This chapter is divided into the following sections:

• Cell Replacement

• Thermocouple Replacement

• Furnace Replacement

• Combustibles/Methane Detector Replacement (optional)

• Flow Sensor Replacement

• Box Heater Wiring

• RCU (Remote Calibration Unit) Solenoid Valve Replacement

• Electronics Board

• Recommended Maintenance Schedule

• Replacement Parts List

You can clean the outside of the sensor using normal household or commercial general purpose cleaners, cloths, or sponges. You can also use water. Always turn off power before attempting to clean the enclosure.

A wiring diagram for a standard sensor is provided in Figure 8-1. This figure shows the locations of all wiring connections on the sensor. If your sensor re-quires different wiring, this is provided in an appendix.

Page 98: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

8-2 | Thermox WDG-V / VC / VCM

Warnings

• If you turn off the analyzer or the process is shut down, turn off the aspirator air to avoid plugging problems.

• The outside of the sensor cover and all sensor assembly components are extremely hot (up to 500°F (260°C) inside the cover), even after a consider-able period from shutdown. Turn off the power to the sensor and control unit when working inside the sensor. Use caution and wear appropriate gloves when handling components.

• Be extremely careful when performing maintenance on the sensor while the process is running, especially if the process is under significant positive pres-sure. Removing any part of the sensor plumbing can allow process gases and gases of high temperature to escape into the sensor.

Cautions

• Always use a backup wrench when working on sensor plumbing. This helps to prevent damaging welds and distorting sensor plumbing.

• Never use pipe dope or any other contaminant that gives off combustible vapor on any joints of the sample tubing. Combustible vapor in the sample tubing can lead to erroneous readings.

Page 99: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Service and Parts | 8-3

CE

LL +C

ELL -

RTD

THE

RM

OC

OU

PLE

+TH

ER

MO

CO

UP

LE -

SIL TH

ER

MO

CO

UP

LE +

SIL TH

ER

MO

CO

UP

LE -

CO

MB

US

TIBLE

S D

ETE

CTO

RA

CTIV

E E

LEM

EN

TC

OM

BU

STIB

LES

DE

TEC

TOR

RE

FER

EN

CE

ELE

ME

NT

ME

THA

NE

DE

TEC

TOR

AC

TIVE

ELE

ME

NT

ME

THA

NE

DE

TEC

TOR

RE

FER

EN

CE

ELE

ME

NT

FLOW

SE

NS

OR

HE

ATE

R E

LEM

EN

TFLO

W S

EN

SO

R S

EN

SE

ELE

ME

NT

CE

LL FUR

NA

CE

AC

50W H

EA

TER

AC

(H

ALF W

AV

E R

EC

TIFIED

)

350W H

EA

TER

AC

CU

STO

ME

R C

ON

NE

CTIO

NS

LINE

NE

UTR

AL

Figure 8-1a. Wiring diagram for standard sensor.

Page 100: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

8-4 | Thermox WDG-V / VC / VCM

PR

OC

ES

S A

LAR

M(A

LAR

M 4)

AN

ALO

G O

UTP

UT

AN

ALO

G O

UTP

UT

CO

NTA

CT IN

PU

TFU

TUR

E

LINE

NE

UTR

AL

PR

OC

ES

S A

LAR

M(A

LAR

M 5)

SP

AR

E

AS

PIR

ATO

RZE

RO

GA

SS

PA

N 1

SP

AN

2

SO

LEN

OID

VA

LVE

CO

NTR

OL

AN

ALO

G O

UTP

UT

AN

ALO

G IN

PU

T

MO

DB

US

(DA

ISY

CH

AIN

)S

EN

SO

R M

OD

BU

S

GN

D S

CR

EW

AC

PW

RS

UP

PLY

PR

OC

ES

S A

LAR

M(A

LAR

M 3)

DA

TA V

ALID

ALA

RM

(ALA

RM

2)

SE

RV

ICE

ALA

RM

(ALA

RM

1)

SP

AR

E

12VD

C FO

R R

CU

SO

LEN

OID

DR

IVE

24 TO 36V

DC

AN

ALO

G O

UTP

UT

EX

TER

NA

L VO

LTAG

E S

UP

PLY

Figure 8-1b. Customer wiring diagram for standard sensor.

Page 101: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Service and Parts | 8-5

Cell Replacement/Cleaning

Disconnect power from the sensor and control unit and allow the sensor to cool before replacing parts inside the sensor.

Use Figure 8-2:

1. Open the sensor cover to expose sensor components.

2. Remove the cell clips.

3. While holding the top of the cell housing with a backup wrench, loosen the lower hex nut (not the top hex nut), then remove the entire cell assembly. (See the * in Figure 8-2).

Loosening the cell may require a good deal of torque.

4. If cleaning the cell, wash with water or alcohol. Dry the cell thoroughly be-fore reinstalling. Always use a new cell O-ring.

If replacing the cell, discard the old cell and cell O-ring and retrieve the new cell with supplied cell O-ring. Avoid touching the bare cell. Instead, hold the cell by one of its hex nuts.

5. Place the cell O-ring on the base of the cell as shown in Figure 8-2.

6. Insert the new cell into the cell housing. Do not touch the bare cell when reinserting. Tighten the cell into the cell housing using the lower hex nut (the upper hex nut is pre-tightened at the factory). The cell O-ring provides a seal for the system. Make sure it is aligned into its proper recess and evenly crushed when tightening the cell.

NOTE

Page 102: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

8-6 | Thermox WDG-V / VC / VCM

FURNACE

O-RING

CELL

HEX NUT

CELL HOUSING

WASHER

HOLD CELL HOUSINGUSING 7/8" OR 1" WRENCH(23mm or 26mm)

TURN HEX NUT COUNTERCLOCKWISE USING A7/8" WRENCH (23mm)

CELL CLIPS

CELL REPLACEMENTFigure 8-2. Cell replacement.

*

Page 103: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Service and Parts | 8-7

Thermocouple Replacement

The tip of the thermocouple is positioned in the furnace so that it is near, but not touching, the cell housing or the furnace heater coil.

Use Figure 8-3:

1. Open the sensor door to expose sensor components.

2. Disconnect thermocouple wires from terminals +TC- on the sensor board.

3. Remove the 2 speed clips from the thermocouple mounting tabs. Use needle-nose pliers to remove the clips.

4. Pull the thermocouple straight down and remove it from the furnace.

5. Insert the new thermocouple aligning the tab with the two posts on the bot-tom plate. Be careful not to damage the ceramic tip while inserting the as-sembly.

6. Attach one new speed nut to each post.

7. Attach the thermocouple wires to the +TC- terminals on the sensor card (yel-low wire is positive (+)).

8. Perform an oxygen calibration.

NOTE

Page 104: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

8-8 | Thermox WDG-V / VC / VCM

Figure 8-3. Thermocouple replacement

Page 105: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Service and Parts | 8-9

Combustibles/Methane Detector Replacement

The Combustible detector will always face outward, and the Methane detector will always face inward from the factory.

Combustibles Detector Replacement

Use Figure 8-4:

1. Disconnect the combustible detector (Black and Brown) wires on the sensor board.

2. Loosen the two 4.0 mm hex screws that holds the combustible detector in place and remove the detector from its housing. Save the screws from the detector to use when installing the new combustibles detector.

3. Always use a new O-ring when detector is removed.

4. Install the new combustible detector in place, aligning the two screws on the block. Be careful not to damage the detector element ends while inserting the assembly. Tighten screws until face is flush with the block and gasket is fully compressed.

5. Reconnect the detector leads to the sensor board.

Methane Detector Replacement

1. You must remove the block in order to access and replace the Methane detec-tor.

2. Disconnect the methane detector (Blue and White) wires on the sensor board.

3. Loosen the two 4.0 mm hex screws that holds the methane detector in place and remove the detector from its housing. Save the screws from the detector to use when installing the new combustibles detector.

4. Always use a new O-ring when detector is removed.

5. Install the new methane detector in place, aligning the two screws on the block. Be careful not to damage the detector element ends while inserting the assembly. Tighten screws until face is flush with the block and gasket is fully compressed.

6. Reinstall the block.

7. Reconnect the detector leads to the sensor board.

NOTE

Page 106: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

8-10 | Thermox WDG-V / VC / VCM

Figure 8-4. Combustibles detector replacement

Page 107: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Service and Parts | 8-11

Flow Sensor Replacement

Flow Sensor Replacement

Use Figure 8-5:

1. Disconnect the flow sensor Red wires (Heater) and the Yellow wires (Sense) on the sensor board.

2. Loosen the nut that connects the heater block and the flow sensor.

3. Loosen the nut that connects the furnace and the flow sensor. Brace the fur-nace assembly by using a second wrench attached to the sensor nut under the furnace. See Figure 8-5 for reference.

4. Remove the flow sensor.

Handle the new flow sensor with care. It is very delicate when not installed in the system.

5. Install the new flow sensor in place. Tighten the two connecting nuts.

6. Reconnect the Red wires (Heater) and the Yellow wires (Sense) on the sensor board.

NOTE

Page 108: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

8-12 | Thermox WDG-V / VC / VCM

Figure 8-5. Flow sensor replacement

Page 109: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Service and Parts | 8-13

Furnace Replacement

Furnace Replacement

Use Figure 8-6:

1. Disconnect cell furnace wires.

2. Remove the Thermocouple assembly from the furnace.

3. Remove the Cell.

4. Loosen the swage nut at the top and bottom of the furnace. You may need to remove the flow sensor to get the furnace assembly out.

5. Remove the Furnace assembly.

6. Install the new Furnace assembly in place.

7. Reconnect the Thermocouple, Flow Sensor and Cell.

8. Reconnect the wires on the sensor board.

Figure 8-6. Furnace Assembly

Page 110: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

8-14 | Thermox WDG-V / VC / VCM

Heater Replacement

There is only one heater in the O2 version. There are two heaters in the Combustible/Methane version of the WDG-V.

Box Heater Replacement

Use Figure 8-7:

1. Disconnect heater wires.

2. Remove the speed clip from the mounting tab. Use needle-nose pliers to remove the clip.

3. Remove the box heater by sliding it out.

4. Replace the box heater

5. Attach a new speed nut to the post.

6. Reconnect the wires on the sensor board.

NOTE

Page 111: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Service and Parts | 8-15

Figure 8-7. Box Heater Replacement

Page 112: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

8-16 | Thermox WDG-V / VC / VCM

Combustible Block Heater Replacement (Only on Combustible/Methane Units))

Use Figure 8-8:

1. Disconnect heater wires.

2. Loosen the 4.0 mm hex screw.

3. Remove the combustible block heater by sliding it out.

4. Replace the box heater

5. Tighten the hex screw into place.

6. Reconnect the wires on the sensor board.

Page 113: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Service and Parts | 8-17

Figure 8-8. Combustible Block Heater Replacement

Page 114: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

8-18 | Thermox WDG-V / VC / VCM

Electronics Replacement

Electronics Replacement

Use Figure 8-9:

1. Disconnect all wires.

2. Loosen the two Phillips screw located on the back of the electronics plate.

3. Slide the board out until the pegs at the top line up with the holes in the elec-tronics board plate and drop it down.

2. Replace the electronics board.

2. Properly tighten the screws.

Page 115: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Service and Parts | 8-19

Figure 8-9. Electronics Board Removal

Page 116: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

8-20 | Thermox WDG-V / VC / VCM

Recommended Maintenance Schedule

Calibration

Check calibration and/or recalibrate the analyzer every 90 days.

Page 117: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Service and Parts | 8-21

Replacement Parts List

Sensor

Assy, Furnace, WDG-V, 120 VAC 7000-719-SEAssy, Furnace, WDG-V, 240 VAC 7000-720-SEKit, WDG-V Replacement Thermocouple For O2 Cell Furnace 7000-839-TECombustibles Detector Kit - RTD Detector 7000-815-TE *Combustibles Detector Kit - Hot Wire Detector 7000-816-TE **Methane Detector Kit 7000-814-TEWDG-V Sensor Main Processor Board For RTD Comb. Detector 8000-119-SEWDG-V Sensor Main Processor Board For Hot Wire Comb. Detector 8000-148-SEAssy, WDG-V Boards and Plate For RTD Comb. Detector 7000-845-SE ***Assy, WDG-V Boards and Plate For Hot Wire Comb. Detector 7001-214-SE ***Kit, Combustibles Block w/Nut and Ferrules 7001-226-TE****Kit, Combustibles/Methane Block w/Nut and Ferrules 7001-227-TE****Kit, Flow Sensor, WDG-V 7000-817-TECell O-Ring Metal 42005JEKalrez O-Ring, Combustibles/Methane Detector Housing 3000-337-JEStandard Zirconia Cell 7000-568-SESevere Service Zirconia Cell - Consult Factory For Application 7000-733-SEKit, WDG-V Aspirator w/ Nut And Ferrules 7001-216-TEFlashback Assy, 1” Element 71212SEKit, Convection Loop w/ Nut and Ferrules, O2 Only (long tube for use without flame arresters) 7001-224-TEKit, Flame Arrester Convection Loop w/ Nut and Ferrules (short tube for use with flame arresters) 7001-225-TECell Clip Assembly - Inside 72329SECell Clip Assembly - Outside 72328SEAssy, WDG-V, Box Heater, 120VAC 7000-843-SE *****Assy, WDG-V, Box Heater, 240VAC 7000-844-SE *****Heater Plate WDG-V 7000-405-KEKit, Cartridge Heater 350W, 120V, Includes Heater and Push Nut 7000-846-TEKit, Cartridge Heater 50W, 120V, Includes Heater and Screw 7000-848-TEKit, Cartridge Heater 350W, 240V, Includes Heater and Push Nut 7000-847-TEKit, Cartridge Heater 50W, 240V, Includes Heater and Screw 7000-849-TEKit, WDG-V Replacement 115 VAC Heater Coil and Push Nut 7000-840-TEKit, WDG-V Replacement 240 VAC Heater Coil and Push Nut 7000-841-TERTD - Box Temp Control - 100 Ohm (O2 Only) 25384JERTD - Box Temp Control - 100 Ohm (Comb/Methane) 1000-630-JEThreaded Exhaust Tube 70619KESample Probe by serial number

Page 118: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

8-22 | Thermox WDG-V / VC / VCM

Remote Calibration Unit (RCU)

Air Regulator 39004JERegulator Gauge 37070JEAspirator Air Solenoid Valve (normally open) 36090JECalibration Gas Solenoid Valve (normally closed) 36088JEFlow Meter 37020JE

* The RTD detector is the standard detector for most appications. Requires sensor main processor board 8000-119-SE ** The Hotwire detector is the severe service detector. Consult factory for applications. Requires sensor main processor board 8000-148-SE *** Sensor assembly includes I/O connection board, main processor board, and mounting plate **** The block heater, RTD, and detector are not included ***** Includes heat plate, heater, push nut, and dowel pin

When ordering, provide the serial number of your analyzer to ensure proper parts are ordered:

AMETEKProcess & Analytical Instruments Division150 Freeport RoadPittsburgh, PA, USA 15238Phone: (412) 828-9040 Fax: (412) 826-0399

Page 119: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Drawings and Custom Instructions | 9-1

DRAWINGS AND CUSTOM INSTRUCTIONS

This appendix provides any custom drawings or instructions you may have ordered in addition to the standard WDG-V analyzer. If you didn’t order any custom options, the standard Interconnect Drawing is provided.

If you did order any special options, the drawings or special instructions provided here supercede any drawing or options provided elsewhere in this manual.

Page 120: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

9-2 | Thermox WDG-V / VC / VCM

This page intentionally left blank.

Page 121: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

Customer Modbus Map | 10-1

CUSTOMER MODBUS MAP

Sensor Host

Register Name Data Type Description

1 0 System Alarm Int 32 System alarms reported by the AMEVision Display User Interface Software

Read Only

1 2 Oxygen Float Oxygen concentration in percentage Writable

1 4 Combustibles Float Combusitble concentration in ppm Writable

1 6 Methane Float Methane concetration in percentage Read Only

1 8 Cell Temperature Float Cell temperature in degree C Read Only

1 10 Box Temperature Float Box temperature in degree C Read Only

1 12 Cell mV Float Oxygen cell in millivolts Writable

1 14 CJ Temperature Float Electronics Temperature (cold junction com-pensator temp) in degree C

Read Only

1 16 Process Alarm Mask Unsigned Int 32 Process alarm bit mask Writable

1 18 Command Unsigned Int 16 Command to the analyzer Writable

1 19 Command Response Unsigned Int 16 Response to a command

1 20 Oxygen Span Gas Float Oxygen span gas value Writable

1 22 Zero Gas Float Zero gas value Read Only

1 24 Combustible Span Gas

Float Combustible span gas value Read Only

1 26 Methane Span Gas Float Methane span gas value Read Only

1 28 Data Valid Unsigned Int 16 Datavalid (1 if valid, 0 if not) Read Only

1 29 State Unsigned Int 16 Current state of the analyzer Read Only

1 30 Event Mask 1 Unsigned Int 32 Event Mask 1 Writable

1 32 Event Mask 2 Unsigned Int 32 Event Mask 2 Read Only

1 34 Pressure Float Process Pressure Writable

Page 122: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

10-2 | Thermox WDG-V / VC / VCM

This page intentionally left blank.

Page 123: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

French Warnings / Français Avertissements | A-1

FRENCH WARNINGS / FRANÇAIS AVERTISSEMENTS

NOTES SUR LA SECURITE

LES AVERTISSEMENTS, LES PRÉCAUTIONS ET LES NOTES CONTE-NUS DANS CE MANUEL ATTIRENT L’ATTENTION SUR DES INSTRUC-TIONS CRITIQUES COMME SUIT :

Une procédure de fonctionnement qui n’est pas strictement suivie peut engendrer une blessure de personne ou une contamination de l’environnement.

Une procédure de fonctionnement qui n’est pas strictement suivie peut endommager l’équipement.

Informations importantes à ne pas laisser passer :

Risque de brûlure. Surface brûlante. Ne pas toucher, laisser refroidir avant tout travail.

NOTE

Page 124: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

A-2 | Thermox WDG-V / VC / VCM

Sécurité électrique

Une tension jusqu’à 5 kV peut être présente dans le boîtier. Coupez toujours les alimentations électriques avant d’effectuer une maintenance ou un dépannage. Seuls des techniciens qualifiés électriquement doivent effectuer les connexions électriques et les vérifications de terres.

Une utilisation quelconque de l’équipement de manière non spécifiée par le con-structeur peut compromettre la protection de la sécurité d’origine fournie par l’équipement.

Mise à la terre

La mise à la terre de l’instrument est obligatoire. Les spécifications de performance et la protection de sécurité sont invalides si le fonctionnement de l’instrument est effectué depuis une alimentation électrique à la terre impropre.

Vérifiez la continuité de la terre de tout équipement avant d’alimenter électriquement.

Page 125: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

French Warnings / Français Avertissements | A-3

Note importante aux utilisateurs

Il n’y a pas de composant à maintenir à l’intérieur même des composants de l’analyseur WDG-V. Ne pas retirer le capot d’un analyseur WDG-V. Se référer à du personnel qualifié pour la maintenance.

Ce qui suit s’applique aux modèles WDG-V Division 2 :

L’alimentation électrique, les entrées et sorties (E/S) doivent être en accord avec les méthodes de câblage en zone Class I Division 2 (se référer au code électrique national) et en accord avec l’autorité locale ayant juridiction.

AVERTISSEMENT – RISQUE D’EXPLOSION – LA SUBSTITU-TUON DE COMPOSANTS PEUT COMPROMETTRE LA CONFOR-MITE DE ZONE CLASS I DIVISION 2.

AVERTISSEMENT – RISQUE D’EXPLOSION – EN ZONE EXPLO-SIBLE, COUPEZ L’ALIMENTATION ELECTRIQUE AVANT DE REMPLACER OU CABLER UN MODULE.

AVERTISSEMENT – RISQUE D’EXPLOSION – NE DECON-NECTEZ PAS L’EQUIPEMENT SAUF SI L’ALIMENTATION ELEC-TRIQUE A ETE COUPEE OU SI LA ZONE EST SANS RISQUE.

CET EQUIPEMENT EST CONFORME A UNE UTILISATION EN ZONE CLASS I, DIVISION 2, GROUPES A, B, C ET D OU HORS ZONE EXPLOSIBLE SEULEMENT.

La température ambiante maximum ce cet analyseur est de 60°C (140°F).

L’unité de terrain WDG-V et son afficheur sont des pieces complexes qui doivent être maintenues uniquement par des techniciens qualifies ayant une expertise en technologie de l’instrument et en systèmes électriques. AMETEK recommande que tout analyseur nécessitant un dépannage soit renvoyé à l’usine. Vous devriez seulement essayer de travailler sur cet équipement après avoir reçu une formation d’un représentant AMETEK / Division P&AI. Si vous décidez de travailler sur cet analyseur, ayez conscience que des hautes tensions, des hautes températures et d’autres conditions potentiellement à risque peuvent surgir.

Page 126: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

A-4 | Thermox WDG-V / VC / VCM

Sommaire de Sécurité Générale

Lisez attentivement les précautions de sécurité suivantes pour éviter de se blesser et d’endommager l’équipement ou tout produit connecté à ce dernier.

Utilisez un câblage adéquatPour éviter les risques de feu, utilisez seulement le câblage spécifié dans le chapitre Installation de ce manuel de l’utilisateur.

Evitez les surcharges électriquesPour éviter une électrocution ou un risque de feu, n’alimentez pas électriquement des bornes qui sont hors de la gamme spécifiée pour ces bornes.

Mettez à la terre l’analyseurSuivez les instructions de mise à la terre fournies dans le chapitre Installation de ce manuel de l’utilisateur. Avant de connecter aux bornes d’entrée ou de sortie de l’analyseur, assurez-vous que l’analyseur est correctement mis à la terre.

Ne pas faire fonctionner sans capotPour éviter une électrocution ou un risque de feu, ne faites pas fonctionner l’analyseur sans capot ou avec un panneau retiré.

Utilisez des fusibles adéquatsPour éviter tout risque de feu, utilisez des fusibles de types et classes spécifiés pour cet analyseur.

Ne faites pas fonctionner en atmosphère explosiblePour éviter toute blessure ou risque de feu, ne faites pas fonctionner l’analyseur dans une atmosphère explosible, sauf si vous avez acheté les options spécifiquement conçues pour ces environnements.

Page 127: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

French Warnings / Français Avertissements | A-5

Precautions aux Dommages de L’Analyseur

Utilisez une alimentation électrique adéquateNe faites pas fonctionner l’analyseur avec une alimentation électrique qui fournit plus que la tension électrique spécifiée.

Ne faites pas fonctionner avec les défaillances suspectesSi vous suspectez que l’analyseur est endommagé, demandez à un technicien qualifié de l’inspecter.

Prenez des précautions en levant l’instrumentL’analyseur pèse environ 25 kg. Prenez des précautions lorsque vous le levez de sa caisse.

Utilisez des vêtements adaptésL’analyseur est brûlant, l’utilisateur doit porter des gants de protection lorsqu’il manipule l’analyseur.

Page 128: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

A-6 | Thermox WDG-V / VC / VCM

Les opérations dans ce chapitre doivent être effectuées seulement par un technicien qualifié et expérimenté en techniques de sécurité. Il n’y a pas de composant à maintenir à l’intérieur même des composants de l’analyseur WDG-V. Ne jamais travailler sur l’analyseur sans que l’alimentation électrique ait été coupée et que l’analyseur ait pu refroidir pendant au moins une heure. Utilisez aussi des gants si vous travaillez sur l’analyseur.

Ne pas installer le chauffage de la sonde si le gaz échantillon contient potentiellement un mélange explosive de combustibles ; le chauffage de la sonde peut chauffer le gaz jusqu’à son point d’inflammation.

Ne pas ouvrir une porte à charnière de l’unité de terrain purgée Z avant d’avoir vérifié que la zone n’est pas classée à risque.

Des pièces internes chaudes sont au-dessus de la température à laquelle s’enflamment les gaz combustibles. L’alimentation élec-trique doit être coupée sur les unités de terrain et de contrôle pen-dant 90 minutes en maintenant le debit d’air de purge avant d’ouvrir la porte, sauf s’il est prouvé que la zone est sure. L’alimentation électrique ne doit pas être remise après ouverture de l’enceinte jusqu’à ce qu’elle ait été purgée pendant 65 minutes à une pression de 10 mm H2O.

N’ouvrez pas l’unité de calibration déportée purgée Z jusqu’à ce que vous verifiez que la zone soit classée sans risque. Maintenez aussi la purge Z à la pression et au debit indiqués sur la plaque d’avertissement de l’unité.

NOTE

Page 129: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

French Warnings / Français Avertissements | A-7

Coupez l’alimentation électrique du secteur avant d’effectuer un câblage.

Ne placez pas les fils d’alimentation électrique de l’unité de contrôle dans les mêmes câbles que d’autres fils d’alimentation électrique. En séparant ces fils, vous empéchez des signaux transitoires d’atteindre l’unité de contrôle.

Dans un but de compatibilité électromagnétique, en aucune circon-stance vous ne devez laisser l’écran d’un câble déconnecté en une ou deux extrémités (unité de contrôle ou de terrain ou autre appareil).

Les operations dans ce chapitre doivent être effectuées seulement par un technicien qualifié et expérimenté en techniques de sécurité électrique.

Soyez prudent lorsque vous travaillez sur l’analyseur. Laissez l’analyseur refroidir et portez des gants.

Coupez l’alimentation électrique des unités de terrain et de contrôle et laissez l’analyseur refroidir avant de remplacer des pieces dans l’analyseur.

Page 130: Thermox WDG-V, WDG-VC, WDG-VCM Combustion Analyzer · Thermox® WDG-V, WDG-VC, WDG-VCM Combustion Analyzer User Manual Thermox 150 Freeport Road PN 9000-133-VE Rev. K Pittsburgh,

A-8 | Thermox WDG-V / VC / VCM

This page intentionally left blank.