Thermo Scientific Orion 2230 Silica Analyzer User Guide
Ross and the COIL trade dress are trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries.
AQUAfast, AQUASensors, BOD AutoEZ, ionplus, KNIpHE, LogR, No Cal, ORION, perpHect, PerpHecT, pHISA, pHuture, Pure Water, Sage, ROSS, ROSS Ultra, Sure-Flow, Titrator PLUS, and TURBO2 are registered trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries.
A+, All in One, Aplus, AUTO-BAR, AUTO-CAL, Auto-ID, AUTO-READ, AUTO-STIR, Auto-Test, AutoTration, CISA, digital LogR, DuraProbe, EZ Startup, ISEasy, Low Maintenance Triode, Minimum Stir Requirement, MSR, NISS, Optimum Results, Orion Dual Star, Orion Star, SAOB, SMART AVERAGING, SMART STABILITY, Star LogR, Star Navigator 21, Stat Face, Triode are trademarks of Thermo Fisher Scientific, Inc. and its subsidiaries.
Guaranteed Success and The Technical Edge are service marks of Thermo Fisher Scientific, Inc. and its subsidiaries.
© 2010 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific, Inc. and its subsidiaries.
The specifications, descriptions, drawings, ordering information and part numbers within this document are subject to change without notice.
This publication supersedes all previous publications on this subject.
Thermo Scientific Orion 2230 Silica Analyzer User Guide �
Table of Contents
General �nformation: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11.1SafetyPrecautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11.2SymbolsUsedinthisManual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11.3Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
�ntroduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32.1UnpackingtheInstrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 2.1.1 StartupProcedure. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42.2GeneralProductDescription. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 2.2.1 UserInterfaceGeneralDescription......................6 2.2.2 MenusDisplay . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7 2.2.3 MeasurementModeDisplay. . . . . . . . . . . . . . . . . . . . . . . . . . .8 2.2.4 AnalysisReactionPrinciple. . . . . . . . . . . . . . . . . . . . . . . . . . . .9 2.2.5FluidsStorage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 2.2.6 CalibrationStandard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 2.2.7 ReagentsRequired. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 2.2.8 ReagentConsumption . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
�nstallation Environmental Considerations . . . . . . . . . . . . . . . . . . . . . . . . . 113.1MountingtheInstrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .113.2PlumbingConnections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14 3.2.1 InstallingPinchValveTubing . . . . . . . . . . . . . . . . . . . . . . . . .15 3.2.2 InstallingPeristalticPumpHeadTubing. . . . . . . . . . . . . . . . .163.3InstallingtheReagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .173.4ConnectingPowertotheAnalyzer . . . . . . . . . . . . . . . . . . . . . . . . . .17 3.4.1 AnalogOutputsConnections . . . . . . . . . . . . . . . . . . . . . . . . .19 3.4.2 AlarmConnections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20 3.4.3 InputsSignalConnections . . . . . . . . . . . . . . . . . . . . . . . . . . .23 3.4.4 CommunicationSignalConnections. . . . . . . . . . . . . . . . . . . .23
Startup and Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 254.1PlacingtheAnalyzerintoService. . . . . . . . . . . . . . . . . . . . . . . . . . . .264.2ValidatingaCalibrationStandard . . . . . . . . . . . . . . . . . . . . . . . . . . .264.3RunninganAnalysis(AnalyzeOnce) . . . . . . . . . . . . . . . . . . . . . . . .274.4InstrumentOperationModes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .274.5SampleAnalysisandValidation. . . . . . . . . . . . . . . . . . . . . . . . . . . . .274.6Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .284.7Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .284.8PostWash . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .284.9Priming. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
Chapter 3
Chapter 4
Chapter 2
Chapter 1
�� Thermo Scientific Orion 2230 Silica Analyzer User Guide
Analyzer Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 295.1OpticsCalibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .295.2OpticsCalibrationSteps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .295.3HeaterTemperatureCalibration. . . . . . . . . . . . . . . . . . . . . . . . . . . .305.4PumpsCalibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .315.5PumpsCalibrationMethod. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .315.6AnalogOutputsCalibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
Calibration of the �nstrument for Measurement Analysis . . . . . . . . . . . . 336.1 TheoryofCalibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .336.2 MethodsofCalibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .346.3 SinglePointCalibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .346.4 BaseLineCalibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .346.5 HiRangeDilution-FactorCalibration . . . . . . . . . . . . . . . . . . . . . .356.6 TheoryofMeasurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .366.7 ModeofMeasurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .366.8 AutoRestartMode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .376.9 ManualRestartMode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .376.10 MeasurementCycle. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38 6.10.1CellWash . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38 6.10.2SampleCapture . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38 6.10.3OpticsTesting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39 6.10.4CheckingReactionTemperature . . . . . . . . . . . . . . . . . . . . .39 6.10.5BlankMeasurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40 6.10.6AddReagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40 6.10.7CollectData. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41 6.10.8PostWash . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41 6.10.9MeasurementData. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42 6.10.10PrintingResults. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .42
Menu Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 437.1 SetupMenu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .437.2 MeasurementMenu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .457.3 ViewDataandAlarmsMenu. . . . . . . . . . . . . . . . . . . . . . . . . . . . .457.4 ServiceMenu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .467.5 SystemAdministrationMenu . . . . . . . . . . . . . . . . . . . . . . . . . . . . .477.6 HardwareSetupMenu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .477.7 PrimeFluidicsMenuandNewReagentsBlank. . . . . . . . . . . . . . . .487.8 DiagnosticsMenu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .487.9 DefaultandLostorForgottenPasswords . . . . . . . . . . . . . . . . . . . .497.10UserInterfaceTree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .51
Terms Definition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
Chapter 5
Chapter 6
Chapter 7
Chapter 8
���Thermo Scientific Orion 2230 Silica Analyzer User Guide
Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 619.1ChangingReagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .61 9.1.1 InstallingTubingMaintenanceKit . . . . . . . . . . . . . . . . . . . . .629.2CleaningtheSampleFlowChamber . . . . . . . . . . . . . . . . . . . . . . . . .649.3Stand-ByModeandExtendedShutdown. . . . . . . . . . . . . . . . . . . . .64ErrorMessages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65Measurement/Calibration/ValidationMessages . . . . . . . . . . . . . . . . . . . .66CalibratingOpticsMessages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67CalibratingThermistorMessages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67PartsandAccessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .67NoticeofCompliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .68DeclarationofConformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .69WarrantyInformation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .70WarrantyShipments/Returns/Adjustments. . . . . . . . . . . . . . . . . . . . . . .71Consumables. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .71Assistance................................................71
Chapter 9
Thermo Scientific Orion 2230 Silica Analyzer User Guide 1
Chapter 1 General �nformation
Thisproducthasbeendesignedwithcarefulattentiontosafety.Itcomplieswithformalsafetydesignrulesestablishedbyinternationalregulatoryagencies.Observeallsafetywarningsmarkedontheinstrument.Thesewarningsidentifyareasofpotentialhazard,whichcouldresultinpersonalinjury,orlossoflife.
Toensurethattheprotectionprovidedbythisequipmentisnotimpaired,donotuseorinstallthisequipmentinanymannerotherthanthatwhichisspecifiedinthismanual.
•Donotusethisproductforanypurposeotherthandescribedinthismanual.
•DisconnectpowerbeforeservicingandpriortomakinganyterminalconnectionswithintheAnalyzerenclosure.
•Donotoperatewiththeelectronicsenclosuredooropen.
wArninG
WARN�NG This symbol alerts the user about possible hazards that can cause serious damage to the instrument, and/or personal injury.
CAUTiOn
CAUT�ON This symbol alerts the user about possible faults from incorrect operation, which if ignored, could result in damage of the instrument.
HOT This symbol indicates that the marked area can be hot and should not be touched without care.
EARTH (Ground) This symbol, when noted on the product, identifies the location of the connection for Protective Earth (ground).
NOTE: indicates important information that requires special attention.
1 .2 Symbols Used in this Manual
1 .1 Safety Precautions
2 Thermo Scientific Orion 2230 Silica Analyzer User Guide
General
Display Graphic LCD 120 x 92 mm (4.7 x 3.6 inches)
Enclosure Fluidics iP65 (nEMA 4X); Electronics iP52 (nEMA 12) rated for indoor installation
Dimensions (H x w x D) (698 x 406 x 162 mm) 27 x 16 x 6 inches
weight Approximately 16 kg (35 lb) without reagents
Mounting wall or Panel mount
Measuring Method Heteropoly-molybdenum blue
Measuring interval 15, 20, or 30 minutes (selectable settings and user programmable)
Electrical
Power requirements 110 watts. Automatic voltage recognition from 100 – 240 VAC, 110 w (VA), 50 – 60 Hz, Single Phase
Connection Three wire terminal block (18 – 12 AwG) through a 19 mm (3/4-inch) bulkhead on side of the electronics enclosure
Safety Class i, protective grounding
Fuse rating 250 VAC, 4 A (FAST BLOw)
relay Contacts Four potential free relays with 250 VAC, 2A resistive contact ratings. Each relay may be configured separately to a list of programmable alarm functions
Outputs Two-isolated analog Output: 0/4-20mA, max. Load 900 ohms Four Programmable Alarm relays, rS-232
inputs 2 programmable functions digital control inputs
Environmental
Storage Temperature range 5 – 45 °C (41 – 113 °F)
Operating Temperature range 5 – 45 °C (41 – 113 °F)
relative Humidity 90% at 40 °C (104 °F), non-Condensing
Altitude Up to 2000 m
Sample Requirements
inlet Pressure to instrument 1 – 5 psig
Sample Flow rate 50 – 1000 ml/min with < 100 microns particles size (user provides filter)
Sample Temperature 5 – 40 °C (41 – 104 °F)
Sample Fitting 6mm (1/4-inch) OD tubing
Drain Fitting 9.6mm (3/8-inch) OD tubing
1 .3 Specifications
Thermo Scientific Orion 2230 Silica Analyzer User Guide 3
Chapter 2 �ntroduction
TheThermoScientificOrion2200seriesanalyzersarecompactwet-chemistryanalysissystems,designedforcontinuouson-lineanalyzeringofseveralapplicationsinthewaterindustries.
TheOrion2230SilicaAnalyzerusesthecolorimetricdetectionmethodthatreliesonspecificreagentsthatareaddedtothesampletoproduceareactedcolorcompound.
Theintensityofthecolordeterminestheconcentrationinreferencetoaknownreactedstandard.ThecoloredcompoundabsorbsenergyataspecificwavelengthinaccordancewiththeBeer-LambertLaw.
TheinstrumentusesLightEmittingDiodes(LEDs)thatemitlightatanarrowbandwidthselectedforthespecificapplication.
TheSilicaanalysismethodusesthenearinfraredLED(NIR)at810nm.
RemovetheAnalyzerfromitsshippingcartonandinspectitfordamage.Verifythatallthepartslistedbelowinthestartupkitdetailareincluded.
Ifdamageisevidentoriftheshipmentisincomplete,pleasecontacttheCustomerServicedepartment.Donotdestroytheshippingcontainerduringunpackingsothatyoucansaveitforpossiblefutureuse.
NOTE: The Analyzer may be shipped with a translucent protective cover on the LCD. Carefully peel off this translucent layer and discard it.
wArninG
DO NOT APPLY ELECTR�CAL POWER TO V�S�BLY DAMAGED COMPONENTS, AS �NJURY OR FURTHER EQU�PMENT DAMAGE MAY OCCUR .
2 .1 Unpacking
the �nstrument
4 Thermo Scientific Orion 2230 Silica Analyzer User Guide
Thissectiondescribesnecessarystepstobeginusingthe2230SilicaAnalyzer.RefertoChapter3fordescriptionsofmountingtheinstrument,makingtheplumbingconnections,andconnectingelectricalpower.Moredetailscanbefoundinsubsequentchapters.
OncetheAnalyzerissecurelymountedinafixedlocation,wiredtoelectricalpower,wiredtocommunications,(ifnecessary)andplumbedtoprocesswater,performthefollowinginstrumentStartupProcedure:
1. Poweroninstrument.2. Toallowlongerlifetimeofunit,thepinchvalveandpump-headtubing
arenotfullyengagedwhentheunitisshipped.RefertoSection3.2. a. ResettubesinPumpsheads.Note: thisstepmustberepeated
forallthreePumpheads. i. Removetape. ii. Slidetuberetainerintoplacealongkeyway. iii. Installtubingcoverbysqueezingbothsides,andsliding intogrooves. b. Restreartubinginpinchvalves.Note: Besuretubingisproperly
seatedanddoesnotpinch.3. InstallReagents.Verycarefullymatchreagentnumbersandcolors
intheirproperorder.Refertonextpage,andSection3.3.
4. Primethetubing.Pleaserepeatthefollowingsequence3times.(SeeSection3.3)
a. PressMenu b. SelectServices c. SelectPrimeFluidics d. SelectPrimeAll Note: Thisprocedurewillrunautomaticallyandtake
approximately10minutespercycle.5. RefertoChapter4formoredetailsonstartupandoperation.
Oncetheabove4stepshavebeencompleted,(4.a,b,c,d)pressStop,SelectRun.
Allowtheunittooperateovernightorforatleast12hours,tobecomestabilizedinthelocalenvironment.
6. Foradditionalinformationon2230OrionSilicaProcessAnalyzerOperation,refertoChapter4.
2 .1 .1 Startup Procedure
NOTES:
it is recommended to wear latex gloves whenever handling or changing reagents.
Turn the reagent cap clockwise while holding bottle vertical and level.
inspect the tubes to be sure none have gotten pinched.
5Thermo Scientific Orion 2230 Silica Analyzer User Guide
AdditionalDetailcanbefoundinthefollowingchapters:Chapter4-InstrumentOperationChapter5-AnalyzerHardwareConfigurationandCalibrationChapter6-CalibrationoftheInstrumentforMeasurementAnalysisChapter7-MenuOverviewChapter8-TermsDefinitionChapter9-MaintenanceSchedule
ThisReferenceManualprovidestheinformationnecessarytoobtainthemaximumbenefitsfromtheOrion2230SilicaAnalyzer.Alloperatinginstructions,functionaldescriptions,illustrations,andotherrelevantinformationarecontainedinthismanual.Descriptionsofthesystemhardwareanduser-interfacesoftware(thecontrolpanel)areincluded,asisanoverviewoftheanalyticaltechniquesemployedintheAnalyzer.Step-by-stepinstructionsguideyouthroughalloperatingproceduresandfieldprogramming.
TheAnalyzerisdesignedforeasyoperation.EachAnalyzerinthisfamilyrunsinafullyautomatedmodeforroutine,day-to-dayanalyzering.Theneedforoperatorinteractionisminimal.Accordingly,theinformationrequiredbytheroutineoperatorofthisAnalyzerismodest,andthatinformationisprovidedinChapter4:StartupandOperation,ofthismanual.
ImportantstandardAnalyzercapabilitiesinclude: •Samplingtheprocessstreamautomatically. •Calculatingconcentrationofyoursampleandreporting
itinunitsofconcentration. •Recalibratingautomaticallyatpreprogrammedintervalsto
compensateforanychangesincapturedsamplevolume,pumpflowratesandreagentcomposition.
•Cleaningautomatically(whenprogrammed)atuserspecifiedintervalstopreventprecipitationbuild-up,coatingoftheopticsorbacterial/fungalgrowth.
2 .2 General Product
Description
reagent Bottle numbers in Proper Order
6 Thermo Scientific Orion 2230 Silica Analyzer User Guide
TheAnalyzerbringscolorimetriclaboratoryprecisiontotherealworldofunattendedprocessanalyzering.Itisdesignedforlonglifeundercontinuous,rigoroususe.
The2200seriesaredesignedforaccommodatingseveralapplications.Eachapplicationrequiresaspecificconfigurationinhardwareandsoftwarethatissetatthefactory.Theproductassuppliedwillhaveaconfigurationsheetdetailingthesetupandthedetailedtestsandvalidationthatwereconductedatthefactory.ThisAnalyzermodelisintendedtouseforwater analysisonly.
Theuser-interfaceconsistsofakeypadandagraphicalLCDdisplay.TheKeypadislocatedtotherightofthedisplay.TheF1 F4functionkeys(softkeys)takeondifferentmeaningsdependingonthemenuselection.Thecolumnontheleftofthesoftkeyslabelsthekeyswiththeirappropriatefunctions.TheRUNkeystartstheanalysisandtheSTOPkeystopsthecurrentfunctioningoftheInstrument.The+and¯keyshelpinnavigationbetweenpages.TheENTERkeyconfirmsselectionsmadeinthemenus.Additionalnavigationisprovidedbythesoft-keys(inappropriatemenus),forleft/rightorup/downcursormovement.
ThegraphicalLCDdisplayshowsthedate,analysislabel,time,thenavigationalmenus,information&helptotheuser,thecurrentuser,instrument-workingmode,currentstatusoftheinstrumentandthesoft-keylabels.
TheMENUkeyisusedtoaccessthemenus.
2 .2 .1 User �nterface
General Description
7Thermo Scientific Orion 2230 Silica Analyzer User Guide
Display Fields
1. DATE FIELDshowsthecurrentdate. 2. CURSORindicatescurrentselection. 3. INSTRUMENT NAME FIELDshowsinstrumentnametag. 4. TIME FIELDshowspresenttime. 5. DISPLAY FIELDshowsmenus/measurements. 6. INFOfieldshowsinformation/instructionstonavigate. 7. USERfieldshowscurrentusername. 8. MODEfieldshowswhethertheinstrumentisworking
inAuto,ManualorRunmode. 9. STATUS fieldshowscurrentworkingstatusoftheinstrument. 10. FUNCTIONAL KEYS ASSIGNMENT FIELD
showsthefunctionsallottedtothefunctionalkeys. 11. FUNCTION KEYSselectfunctions. 12. OPERATIONAL KEYS
aremeantforoperationoftheInstrument.
Symbols used:
÷Escape (Cancel/Back) ß Cursor
Logged out (locked) Logged in
ÈScroll Up ÍScroll Down
Print Delete
ÁScroll Left ËScroll right
˛Accept ˝reject
A“+”beforeamenuitemdenotesthattheitemhasmultipleoptions.
TheUp/DownArrowkeysscrollupordownthroughoptionsandENTERisusedtoviewdetailsoftheoptions.
PressingtheMENUkeydisplaystheinstrument’sMAINMENU.SystemparameterscanbeconfiguredfromtheMainMenu.
NOTE: when the instrument is performing a measurement, the user is restricted from entering the SErViCES menu.
2 .2 .2 Menus Display
8 Thermo Scientific Orion 2230 Silica Analyzer User Guide
Inmeasurementmode,theinstrumentdisplayscurrentandpreviousreadings,units,dateandtime.ThemeasurementresultscanbedisplayedasLargesizefontsorNormalsizefonts.“F1”keywilltogglebetweenLargeandNormalfonts.
NOTE: in AUTO mode, the user can navigate between different menus even though the instrument is performing a measurement or a calibration.
2 .2 .3 Measurement Mode
Display
Figure 1 Orion 2230 Silica
Analyzer SiO2 layout
DAnGEr
Handling chemical samples, standards, and reagents can be dangerous .
Review the necessary Material Safety Data Sheets and become familiar with all safety procedures before handling any chemicals .
CAUTiOn
REAGENT BOTTLES MUST BE kEPT T�GHTLY CLOSED .
Observe all health and safety procedures for handling chemicals. wash any spillage with Distilled water. Allow enclosure to vent any trapped fumes before servicing. Air purge of enclosure is recommended .
EMPTY ALL REAGENTS BEFORE SHUTT�NG UN�T DOWN .
9Thermo Scientific Orion 2230 Silica Analyzer User Guide
Thecolorimetricmethodisbasedonthereactionofreagentswiththesampletoproduceacolor.Thiscolorismeasuredandcomparedtoastandardcolor,whichproportionallydefinesitsconcentration.
Thecolorcompound,whichisthereactedmixtureofthesamplewiththereagents,absorbsenergyataspecificwavelengthinaccordancetoBeer-LambertLaw.
Storeallreagentsandstandardsinadarkplaceatroomtemperature.
Standardpart#223020,thestandardrequiredforthisapplicationisa200ppb.Eachcalibrationcycleconsumesapproximately50ml.Eachbottlecontainsenoughstandardsolutionforapproximately10calibrations.
ReagentKit:2230REContainsall3reagents.
TheAnalyzerrequiresthreereagentsandonestandardsolution.Allmaybetypicallyhousedinthefluidicsenclosurecompartment.
For proper operation: EnsurethatthereagentsarecorrectlyinstalledandtheAnalyzertubingiscorrectlyroutedinthepumpsandpinchvalvesandthattheAnalyzerdraintubeisconnectedtoanatmosphericdrain.Allofthereagentsandstandardventtubesneedtoexhausttotheoutsideoftheenclosure,tominimizetrappinganyvaporsinsidetheenclosure.
1. ConnectsampletotheAnalyzerandensurethatitisflowing throughthesamplechamberbyobservingtheflowfromthe samplereturnline.
2. Turnpowerontotheinstrument.Afterthebootingsequence (thismaytakeseveralminutes)pressSTOPandthenMENUkey:
3. SelectSERVICES MENUoptionandthenPrime Fluidics ➔ AUTO PRIME ➔ ENTER
TheinstrumentwillgothroughaprimingsequenceofalltheavailablesolutiontotheAnalyzer.FollowthestartupproceduredescribedinChapter4.YoumayrepeattheAUTOPRIMEsequenceasnecessarytopurgeoutanytrappedbubblesinthelines.
TheSilicaanalysisrequiresthreereagents.Thereagentusagedependsonthefrequencyofanalysis.Typically0.20mlofeachreagentisusedperanalysiscycle.
Basedon4analysesperhour,the1Literreagentsshouldlastapproximately45days.
2 .2 .4 Analysis Reaction
Principle
2 .2 .5 Fluids Storage
2 .2 .6 Calibration Standard
2 .2 .7 Reagents Required
2 .2 .8 Reagent Consumption
10 Thermo Scientific Orion 2230 Silica Analyzer User Guide
Thermo Scientific Orion 2230 Silica Analyzer User Guide 11
Chapter 3 �nstallation Environmental Considerations
Theinstrumentisdesignedforgeneral-duty,indoorinstallations.TheelectronicsenclosureenvironmentalratingisNEMA12andtheFluidicsenclosureisaNEMA4Xfiberglasswithacleardoorwindow.
TheOrion2230SilicaAnalyzershouldbeinstalledinawell-ventilatedlocation,withambienttemperaturebetween5°Cand45°C(41°to113°F);andrelativehumiditynottoexceed90%at40°C.Theareamustbefreefromdust,corrosivegases,vibrationandshocks;shelteredfromdirectsunlight,andshieldedfromdrippingwater.Powershouldbeinstrumentgradewithovervoltageprotection,andaprovisionfordisconnectingpower.
Theinstrumentcaseisdesignedforwallorpanelmounting. •Usefour#10screwsorequivalentforpropersupport. •RefertoFigure2andFigure3fordimensions
(inmmandinches). •Makesurethatthewallthatyouchoosecansupporttheweight
oftheinstrument(approximately20kgor44lbs.)SeeFigure2. •Locatetheinstrumentasclosetothesamplingpointaspractical
toensurecompletepurgingofthesamplelineduringeachcycle. •Allowapproximately50cm(20in.)ofclearanceattheleftside
fordoorswing.Alsoallowenoughclearanceontherightsideforwiringconduit.
•LocatetheAnalyzerheightsuchthatthedisplayisateyelevelheightforanoperator,allowingalsoforappropriatesampleanddrainconnections.
3 .1 Mounting the
�nstrument
12 Thermo Scientific Orion 2230 Silica Analyzer User Guide
Figure 2 Mounting Dimensions
SAMPLE CHAMBER
C AUT ION Air purge of enclosure is recommended.
SAMPLE IN 1/4 ''SAMPLE RETURN 1/4 ''
DRAIN 3/8 ''
VENT LINES
AIR PURGE
AIR PURGE AIR VENT
I/O SIGNALSPOWER IN
ALLOW 20 INCHES(510 mm) CLEARANCEFOR DOOR SWING
70mm(2.75 ")
304.8mm (12.0 ")
674.
4 m
m (2
6.5
")
132mm (5.19 ")
25mm (0.98 ")30mm (1.18 ")
330.
3mm
(13.
0 ")
41mm (1.62 ")
162mm (6.4")
698m
m (2
7.5"
)
152m
m (6
")
406mm (16")
13Thermo Scientific Orion 2230 Silica Analyzer User Guide
NOTES:
1. it is recommended that you install an external sample shut off valve to allow control while accessing or service to the sample chamber.
2. Air Purging of upper and lower enclosures is recommended.
3. Each Analyzer MUST have its own drain tube vented to atmosphere.
4. The installation of a power circuit breaker is highly recommended.
Figure 3 Typical External
Plumbing Connections
SAMPLE IN
SAMPLE RETURN
FLOW DIRECTION
FILTERED AND OIL FREE AIR PURGE SUPPLY LINE
CONNECT TO ATMOSPHERIC DRAIN
FLOW DIRECTION
SHUT-OFFVALVE
AC MAINS
1
0
POWER CIRCUIT BREAKER MUST BE CLEARLY IDENTIFIED
C AUTION Air purge of enclos ure is recommended.
INSTALL HEIGHT TO OPERATOR EYE LEVEL
PROCESS LINE USER MAY NEED TO INSTALL A FILTER BASED ON USER SAMPLE CONDITIONFLOW RATE : 50 TO 1000 ML/MIN
AC POWER
Flow rate Control not more than 5 psig
14 Thermo Scientific Orion 2230 Silica Analyzer User Guide
Sampleinlet,samplereturnanddrainconnectionsaremadeonthebottomsideoftheinstrument.ThesampleinletandsamplereturnfittingsonthebottomofAnalyzerrequire¼-inchOD(6mmOD)flexibletubing.ThedrainfittingatthebottomoftheAnalyzerrequires3/8-inchOD(9.6mmOD)flexibletubing.TheAirpurgeinletsrequire¼-inchODtubing.TheAirpurgesupplymustbefilteredexternallyandfreeofoilandanyotherparticulates.
DAnGEr
The 3/8-inch (9.6 mm) DRA�N MUST BE VENTED TO ATMOSPHERE. The sample drain contains analysis wastes, which include both sample and chemical reagents. Please check with local authorities concerning proper disposal of waste from this product.
NOTES:
The enclosure drain produces no waste under normal use. However, in the event of spills or leaks, some waste may be produced.
The Analyzer is shipped with all of the internal tubing attached. Before starting up the instrument, verify that the tubing lines are secured.
3 .2 Plumbing Connections
Figure 4internal Tubing and Valves
Tube routing interconnections
10
8
10
8
20
16
17
1511
7
12
125 6
1 2 3 4
9
21
22
1413
18
19
15Thermo Scientific Orion 2230 Silica Analyzer User Guide
Thevalvesoperatebyapplyingpressuretoclosethetubethatrunsthroughthevalveports:ThefrontportisNormallyOpen(NO)andthebackportisNormallyClosed(NC).ToinserttubinginthebacksideportsapplypressurewithfingerasshowninFigure5.Releasethefingerpressureandinsertthenexttubeinfrontport.
3 .2 .1 �nstalling Pinch Valve
Tubing
Figure 5installing the
Tubing in Pinch Valves
NCNO
16 Thermo Scientific Orion 2230 Silica Analyzer User Guide
PresstheSTOPkey.
DisengagetheUshapepressureplateofthepumpheadassemblybypressingwithtwofingersandgentlymoveupsideasshowninFigure6.SlidethetubefittingretainersateachsideasshowninFigure6.Replacetheoldtubebyslidingoffthebarbedfitting.Ifnecessary,cleanthepumprollers.Lubricatethetubingwithapea-sizedquantityofsiliconeoil.Fitthenewtubeandreplaceinthepressureplate.Notethatthepressureplatesidesmustclickproperlyinthepumpheadassemblyandalignonbothsides.
Repeatabovestepsforallpumps.YoucantestthepumpunderDiagnosticsmenuforproperoperation:
MENU ➔ SERVICES ➔ DIAGNOSTICS ➔ PUMPS, VALVES & STIRRER
3 .2 .2 �nstalling Peristaltic
Pump Head Tubing
Figure 6 installing Peristaltic
Pump Head Tubing - (Dual Channel)
17Thermo Scientific Orion 2230 Silica Analyzer User Guide
TheAnalyzerreagentsstartupkitcontainsallthenecessaryreagentsforthespecificapplication.Thereagentsandcalibrationstandardsaresuppliedinready-to-usebottlesthatcanbereplaceddirectlyontothecapassemblies.
FollowthemarkingstoinstalleachsolutionintotherespectivefeedlineintheAnalyzer.
CAUTiOn
Operational Safety and General Precaution
Onlytrainedtechnicalpersonnelshouldperforminstallation,electricalwiringandmaintenanceofthisproduct.ProperhandlingisrequiredtoavoidESD(ElectricalStaticDischarge),astheinternalelectroniccomponentsonthecircuitboardscanbedamagedbystaticelectricity.Itisrecommendedthatbeforetouchinganyelectroniccomponents,(suchasprintedcircuitcardsandthecomponentsonthem)youshoulddischargestaticelectricityfromyourbody.Thiscanbeaccomplished
Figure 7 installing Peristaltic
Pump Head Tubing - (Single Channel)
3 .3 �nstalling the
Reagents
3 .4 Connecting Power
to the Analyzer
18 Thermo Scientific Orion 2230 Silica Analyzer User Guide
bytouchinganearth-groundedmetalsurfacesuchasthechassisofaninstrument,metalconduit,orpipe.
TheAnalyzerisequippedwithathreeterminalpowerconnectorstriplocatedintheelectronicsenclosureasshowninFigure10.
IT IS hIGhLY RECOMMENDED ThAT PROVISIONS FOR AN ExTERNAL CUSTOMER-SUPPLIED POwER SwITCh OR bREAKER bOx (wITh COMPLETE Cb SPECIFICATION), CLEARLY MARKED FOR ThE ANALYzER AND wIThIN REACh OF ThE USER, bE PROVIDED TO ALLOw REMOVING POwER DURING SERVICE.
TheAnalyzerhasbeendesignedtobehard-wiredwithprovisionsfortheACpowerlinestoberoutedinconduitpipesconnectingtothe¾"FNPThubslocatedontheleftsideoftheelectronicsenclosure.SeeFigure9.
Whenusinglinecords,useonlycertifiedpowercordsof3core,min0.75mm2/18AWG,min.temperatureof75°C,H05VV-F.
Thepowersafetygroundmustbeprovidedforproperoperation.
bEFORE MAKING ANY CONNECTIONS: •Makesurethatthewiresarenotpoweredbeforeconnecting
totheinstrument. •Disconnectanypowerpreviouslyconnectedtotherelays. •ConnectthePhaseHotorL(line),EarthandNeutral
totheterminalblock.Notethelocationofthefuse. •Insertthewireintotheterminalconnectorandmakesure
totightenthescrew. •Routesignalconnectionstothebackhub,andpowerconnections
separatelytothefronthub. •Allconnectionsmustconformtolocalsafetystandards.
Figure 9 Electrical wiring Ports
19Thermo Scientific Orion 2230 Silica Analyzer User Guide
CAUTiOn
The Analyzer is intended for use only with single-phase power .
The analyzer is not equipped with a power switch, therefore a circuit breaking device such as a circuit breaker should conform to local safety standards and must be fitted before the final installation. The circuit breaker must be in close proximity to the analyzer and within reach of the user, and must be marked clearly as the disconnection device for the analyzer.
ALL CONDUCTORS NEED TO HAVE M�N�MUM �NSULAT�ON RAT�NG OF 75 °C .
TheAnalyzerisequippedwithtwoisolatedanalogoutputs.Theanalogoutputscanbeprogrammedtobeanyoneofthefollowing:0-20mAor4-20mAorreverseactingi.e.20–0mAor20–4mA.Therangeisalsoprogrammableandcanbescaledaccordingtoapplicationneeds.Theoutputshaveacommonisolationfromthemainsystemandcandriveuptoa900Ohmsload.
PleaserefertoFigure11,nextpage,andTable2,page22,forsignalterminalconnectiondetail.
3 .4 .1 Analog Outputs
Connections
Analyzer Terminal Destination North America Wire Color European Wire Color
HOT or L ( Line) Black Brown
Protective Earth Ground (PG) Green Green with yellow tracer
neutral (n) white Blue
Table 1AC Power Terminal
wire Connections
Figure 10 AC Power Connections
AC POWERCONNECT�ONS
20 Thermo Scientific Orion 2230 Silica Analyzer User Guide
TheAnalyzerhasfivelevelsensorinputsthatacceptadrycontactclosuresignaltoindicatethestateofthelevelsensors.Input5istypicallyassignedforsample.TheAnalyzerisequippedwithfourrelaysthathavevoltagefreecontacts.Twocontacts:NormallyOpen(NO)andNormallyClosed(NC)withaCommon(C).Therelaycontactsareratedforuseatamaximumof240Vand2A.
Theterminalconnectorsacceptwiregaugeupto14AWG.
Theterminalstripislocatedatthebottomoftheelectronicscabinet(Figure11).Table2showstheconnectororderandterminaldesignation. •Makesurenopowerissuppliedtotheinstrument. •Striptheinsulationoneachwireback¼-inch. •Insertthewireendsthroughtheuppermostopeningbelow
theterminalscrewandmakesurethatthestrippedwireendsareintheterminal,tightentheconnectorscrews.
3 .4 .2 Alarm Connections
•RELAYcontactsareratedat:250V,2AAC500W(non-inductiveload).
•Allinputs(exceptIN6&7)mustbeconnectedtoPOTENTIALFREEcontacts.
Figure 11Terminal Strips for Analog
and relay Outputs
21Thermo Scientific Orion 2230 Silica Analyzer User Guide
Figure 12 Communication Ports
ETHERNET
RS-232
22 Thermo Scientific Orion 2230 Silica Analyzer User Guide
Table 2 Main Terminal Strip
Pin Assignments
Terminal Number Designation Function
1 in1 Digital input 1 Live (Level Sensor 1 Live)
2 GnD Digital input 1 Ground (Level Sensor 1 Ground)
3 in2 Digital input 2 Live (Level Sensor 2 Live)
4 GnD Digital input 2 Ground (Level Sensor 2 Ground)
5 in3 Digital input 3 Live (Level Sensor 3 Live)
6 GnD Digital input 3 Ground (Level Sensor 3 Ground)
7 in4 Digital input 4 Live (Level Sensor 4 Live)
8 GnD Digital input 4 Ground (Level Sensor 4 Ground)
9 in5 Digital input 5 Live (Level Sensor 5 Live)
10 GnD Digital input 5 Ground (Level Sensor 5 Ground)
11 in6 Digital input 6 Live (External 1 control)
12 rTn Digital input 6 return (External 1 control return)
13 in7 Digital input 7 Live (External 2 control)
14 rTn Digital input 7 return (External 2 control return)
15 +24V 24V DC Live (limited to 20 mA source)
16 GnD 24V DC Ground
17 Earth Ground System Earth Ground
18 GnD Analog Output 1 Ground
19 OUT1 Analog Output 1 (900 Ohms Max)
20 Earth Ground System Earth Ground
21 GnD Analog Output 2 Ground
22 OUT2 Analog Output 2 (900 Ohms Max)
23 nC1 normally Closed Contact relay 1
24 nO1 normally Open Contact relay 1
25 C1 Common Contact relay 1
26 nC2 normally Closed Contact relay 2
27 nO2 normally Open Contact relay 2
28 C2 Common Contact relay 2
29 nC3 normally Closed Contact relay 3
30 nO3 normally Open Contact relay 3
31 C3 Common Contact relay 3
32 nC4 normally Closed Contact relay 4
33 nO4 normally Open Contact relay 4
34 C4 Common Contact relay 4
23Thermo Scientific Orion 2230 Silica Analyzer User Guide
Thetwoinputsareassignedforexternalcontrolcommands.Inputs6and7requireaDCvoltagebetween5and24Voltstoactivate.A24Vsupplyisavailablefromthemainterminalstrippins15and16thatcanbeusedforactivatingtheexternalcontrolinputs.
DigitalCONTROLinputs:
Pin Signal11 DigitalInput6Live12 DigitalInput6Return13 DigitalInput7Live14 DigitalInput7Return15 +24VDCLive16 24VDCGround
Input6isassignedbydefaulttoRUN/STOPfunctionwithClose/Openofadrycontact.Input7isleftunassignedbydefault.
Drycontact=voltagefreecontact.Thefunctionalityalongwiththestateoftheenablingsignalisprogrammablebytheuser.
TheAnalyzerhastwocommunicationconnectorslocatedattherightsideintheuppercabinet.SeeFigure12.Forconnectingtothem,usethecommunicationconduithubportontherightsideoftheinstrumentforwireentry.
3 .4 .3 �nputs Signal Connections
3 .4 .4 Communication
Signal Connections
24 Thermo Scientific Orion 2230 Silica Analyzer User Guide
Notes:
Thermo Scientific Orion 2230 Silica Analyzer User Guide 25
Chapter 4 Startup and OperationEachOrion2230SilicaAnalyzergetsthoroughlywet-testedandcalibratedatthefactory.Thesystemisreadytobeginanalyzingsamplestreams.TheresultsofthisactivityarerecordedinaQualityControlReportshippedtoyouwiththeAnalyzer.
Sincetheelapsedtimebetweenfactorycalibrationandyourstartupprocessisunpredictable,itisimportantthatyoucalibratetheAnalyzeragainbeforeyouplaceitinserviceforthefirsttime.
PleaserefertoSection4.6foradditionalinstructionsoncalibratingtheAnalyzer.ThereportedresultofacalibrationisanumberthattheAnalyzerusestodeterminethesampleconcentration.Theparticularvalueofthisnumberisnotimportant,butitshouldbewithin±5%ofthecorrespondingvalueintheQualityControlReport.Ifso,youcanbeconfidentthatallsubsystemsofyourAnalyzerareworkingproperly.
Beforeproceedingpleasemakesureofthefollowing: •Thereagentbottles1,2,and3arefullandproperlyconnected. •The200ppbSicalibrationstandard(1/2Lbottle)isfullwith
thestandardandproperlyconnected. •Thedrainandsamplelinesareconnected.Verifythesample
isflowingintothesamplechamberlocatedatthebottomrightsideofthefluidicsenclosure.
•Thelinesareprimedandtherearenotrappedairbubblesinanyofthelines.
Toprimethelines:
MENU ➔ SERVICES ➔ PRIME FLUIDICS ➔ AUTO PRIME ALL
Repeatautoprimeasnecessarytopurgealltheairfromthelines.Whendoneproceedtocalibrate:
MENU ➔ MEASUREMENT ➔ CALIbRATE ➔ CALIbRATE STANDARD [ENTER]
NOTE:
it is highly recommended to allow the 2230 Silica Analyzer to complete its measurement cycle rather than using the STOP key. if the STOP key is pressed before a measurement cycle is complete, it is highly recommended to perform both the AUTO PriME ALL and CLEAn sequences described in Chapter 4.
26 Thermo Scientific Orion 2230 Silica Analyzer User Guide
Theresultofthecalibrationwillbedisplayedasfollows:
STANDARD : 200ppbABSORBANCE : 0.200SLOPE : +1000.00INTERCEPT : -2.0
RefertoSection6.6foradetailedexplanationabouttheslopeandintercept.Theabsorbancevaluefoundcorrespondstothestandardconcentrationused.Ifthecalibrationabsorbancevalueiswithin+/-5%ofyoursystemQCreportthenthesystemisoperatingproperly,andreadyforrunningsampleanalysis.Ifthedataisdifferent,repeatthecalibrationcyclebeforeproceedingfurther.
YourAnalyzerisnowreadyforroutineoperation.ToplacetheAnalyzerinroutineservice,pressthecontrolpanelRUNkey.
TheAnalyzertypicallyrunsunattended,analyzingsamplestreamsandcalibratingitselfatregularintervals.Allrelevantparameterswereprogrammedatthefactorybaseduponyourapplication.Consequently,theAnalyzerisfullypreparedforroutineoperationonceyoucompletethesestartupprocedures.
Aqualifiedandauthorizedusercanchangeagreatmanyoftheoperatingparametersthroughthepassword-protected“TechnicianLevel”menuasdetailedintheUSERINTERFACEmenu.
YoucanchecktheAnalyzer’scalibrationaccuracyatanytimeusingthevalidationfunction.Validationconsistsofrunningasinglemeasurementcycleandreportingtheresultinconcentration.Thisreportedresultcanthenbecomparedtotheexpectedvalue,whichistheknownconcentrationofthevalidationsolution,todeterminetheaccuracyofsystemcalibration.
TheadvantageofusingthisfunctionovertheRunAnalysisfunctionisthatpriorknowledgeoftheconcentrationallowsvalidatingtheaccuracyoftheAnalyzer,withoutaffectingdatareportingtoyourcontrolcenterordevice.
NOTES:
Running a validation is identical to running an analysis: although the results are printed they are not stored, alarm conditions are not tested, and the analog outputs are not updated .
it is highly recommended to allow the 2230 Silica Analyzer to complete its measurement cycle rather than using the STOP key. if the STOP key is pressed before a measurement cycle is complete, it is highly recommended to perform both the AUTO PriME ALL and CLEAn sequences described in Chapter 4.
4 .1 Placing the
Analyzer into Service
4 .2 Validating a
Calibration Standard
27Thermo Scientific Orion 2230 Silica Analyzer User Guide
AlthoughtheAnalyzerisusuallysetuptoanalyzesamplestreamsonaregularschedule,youmaywishtoexecuteananalysisoncommandtoobtainanimmediateresult.Thisfunctionallowsyoutorunsingleormultipleanalysesonasamplewithoutrequiringthattheautomaticmodebeinitiated.TheAnalyzerthenrunsthespecifiednumberofanalyses,testsforalarmconditions,exercisestheanalogoutputs,andprintstheresultsexactlyasifitwereintheautomaticmode.
Torunmultipleanalyses(morethanone)setthenumberofrepeatsinANALYSISmenu:
PressMENU ➔ SETUP ➔ (+) togotopage2 ➔ ANALYSIS PARAMETERS ➔ ENTER ➔ selectNO. OF REPEATS
Therearetwoprimaryoperatingmodes:
1SAMPLEANALYSIS2CALIBRATION
Theinstrumentalsooperatescleaningandpostwashcycles,alongwithaprimingfunctionusefulforstartupandprimingfeedtubinglinestothereaction-measuringcell.
TheSAMPLE ANALYSISmodeissimilartothecalibrationmode,withtheexceptionthatduringcalibrationtheinstrumentcalculatestherelevantparameterscorrelatingtothestandardconcentrationsusedforcalibration.
DuringsampleanalysisthesystemreliesontheCURRENTcalibrationparameterstocalculatethesampleconcentration.Eachmeasurementcyclestartswithflushingthereactioncellwithfreshsample.Anautoblankstepmakessurethatthebackgroundcolorofthesampleismeasuredasabaseline.Typically,reagentsareaddedtodevelopthecolorofthefinalcompound,andtheintensityofthecolordevelopedismeasuredandcorrelatedtothesampleconcentration.
TheValidationcycleissimilartothemeasurementcycleexceptthattheresultisnotreportedontheanalogloopsandnotstoredinthesampleanalysisresultsdatastorage.TheresultofthevalidationanalysisisdisplayedandprintedontheRS-232port.ThevalidationcyclecanbeexecutedonanyavailablesolutionstandardorsampleselectedbytheuserintheValidationsetupparametersmenu.
4 .3 Running an Analysis
(Analyze Once)
4 .4 �nstrument
Operation Modes
4 .5 Sample Analysis
and Validation
28 Thermo Scientific Orion 2230 Silica Analyzer User Guide
Theusermayexecutemanualorsetautomaticcalibrationcyclesatuserdefinedintervals.IntheautomodetheAnalyzercalibratesperiodicallyandchecksthevalidityofthecalibrationperthespecifiedtolerancefortheREFERENCEslopeandoffsetparameters.NewcalibrationdataisonlyacceptedifitiswithinthelimitsofthespecifiedtolerancefortheREFERENCEcalibrationparameters.Thecurrentcolumnshowsthecurrentoperatingparametersthatareusedtocalculatethesampleconcentration.
Automaticflushingofthereactioncellmaybeprogrammed.
TheDefaultsetupfortheanalyzerusesthesampleasthecleaningsolution.Acleaningcyclecanbeperformedbypressing:
STOP ➔ MENU ➔ MEASUREMENT ➔ CLEAN
Thiscyclehelpsmaintaincleanreactioncell.
NOTE: To avoid damage and contamination to the reaction cell, use only cleaning solutions recommended by the factory.
Automaticpostwashcyclesaretypicallypartofeverymeasurementcycle.Attheendoftheanalysis,thepostwashcycleevacuatesthereactedsamplefromthecellandcleansittominimizecoatingandresiduedepositsinthecell.ThepostwashcyclesettingscanbeaccessedfromtheANALYSIS PARAMETERS menu:
MENU ➔ SETUP ➔ (+) ANALYSIS PARAMETERS ➔ (+)
Pleaseconsultfactorybeforemodifyinganyoftheparameters.
Primingoftheinstrumentlinesmaybedoneasauto(allthesolutions)ormanuallybyselectingthespecificreagentorstandard.Primingfillsthetubing,purgesthelinesofair,andpreparestheAnalyzerformeasurementmode.
ThePRIMEVOLUMEinmlappliestoallthelistedsolutionsintheprimingmenu.
The Prime Volume may be specified as a (-) number; this allows the user to pump back the reagents inside the tubing back into the reagents containers allowing for safer service when replacement of the tubing is needed. User may select to pull back all or any one of the reagents or standard.
4 .6 Calibration
4 .7 Cleaning
4 .8 Post Wash
4 .9 Priming
Thermo Scientific Orion 2230 Silica Analyzer User Guide 29
Chapter 5 Analyzer Hardware Configuration and Calibration
EachAnalyzerissetupatthefactory,andshippedreadytobecommissionedforon-lineanalysis.Thissectiondescribesthehardwaresetupincasemaintenanceorotherpostcalibrationadjustmentafterserviceisrequired.
TheOrion2230SilicaAnalyzerisbasedoncolorimetricdetectionprinciples.TheAnalyzerusesopticalcomponentsthatrequireinitialsetupatthefactoryandcanbefurthertunedorre-calibratedbyatrainedtechnician.
ThisChapterdescribestheoperationandcalibrationofthefollowingassemblies:Optics,Heater,PumpsandAnalogOutputs.
Orion2230SilicaAnalyzeropticsconsistsofaNIRLEDlightpathandaphoto-detector.TheanalyzeriscapableofdrivingthreeLEDandtheintensityofeachoftheLEDcanbeindividuallyvaried.Thelightpathispresetinthereactioncellsamplechamber.Thephotodetectorreceivingthelightenergyisconnectedtoanamplifierwithprogrammablegain.TheoutputofthephotodetectoramplifierisdigitizedandmeasuredinmV.OpticscalibrationintheAnalyzerisdoneintwostages.Inthefirststage,thelevelofstraylightisrecorded.TheLEDlightsourceisswitchedoffandtheoutputofthephotodetectorisrecorded.ThesecondstageofcalibrationconsistsofadjustingtheintensityoftheLEDandthegainofthephotodetectoramplifiertogetanoptimumlevelofoutputclosetoafixedreferencelevel.ThisisoptimizedforthespecificLEDandtheresultsofthecalibrationarestored.
Forproperopticscalibrationthecellmustbecleanandfilledwithde-ionizedwater.PresstheMENUkeyandnavigatetoCALIBRATEOPTICS:
MAIN MENU ➔ SERVICES ➔ hARDwARE SETUP ➔ CALIbRATE OPTICS
Thepromptsremindtheusertoaddde-ionizedwaterinthereactioncell,andthenallowsthemtoproceedforopticscalibrationuponconfirmation.
Theopticscalibrationisanautomatedsequencethatalsodetectsfaultconditions.Onceafaultisdetectedthesequenceisaborted.Forexample,theopticscalibrationofstraylightisabortedifthestraylightmeasured
5 .1 Optics Calibration
5 .2 Optics Calibration
Steps
30 Thermo Scientific Orion 2230 Silica Analyzer User Guide
isgreaterthanthefactoryallowedsetreferencevalue.TheopticscalibrationoptimizesthelightfromtheLEDpresentinthesystemalongwithtuningthefront-enddetectiontomaximizethesignallevelswithde-ionizedwaterinthecell.Thesystemstoresalltheparametersfortheapplication.
Thecalibrationresultsareavailableforediting.FromtheMAIN MENU,followthesequence:
MENU ➔ SERVICES ➔ hARDwARE SETUP ➔ EDIT CALIbRATED VALUES
ThestraylightmeasurementislistedasOPTICS DARK;theoptimumlightlevelislistedasOPTICS xREFfortheLEDspecifiedintheMEASUREMENTMETHOD.
InordertogetconsistentandrepeatablemeasurementsthemethodissetupsuchthatthereactioncellintheOrion2230SilicaAnalyzerisheatedtoaspecificsetpoint(25°C).Aheaterelementlocatedonthebackplateofthereactioncellcontrolsthetemperature.Thetemperatureofthereactioncellismeasuredbyathermistorembeddedinthereactioncellbody.Thetemperatureofthecellisdisplayedontheresultsscreen.
NOTE: Calibration is typically NOT required except in rare cases where the thermistor has been replaced and the default parameters are not functional .
TheOrion2230SilicaAnalyzerreliesonaPID(Proportional,Integral,Derivative)algorithmsettingtocontrolandmaintainthecellatthesetpoint.Thefollowingprocedureallowstuningandcalibrationofthethermistorincaseitwasreplacedorrequiredadjustment.
HeaterTemperatureCalibration:
MENU ➔ SERVICES ➔ hARDwARE SETUP ➔ CALIbRATE ThERMISTOR
PressF3keytoRESETthermistorcalibrationvaluetoFACTORYDEFAULTS.
Theoperatorisaskedtoreadthetemperatureofthereactionusinganexternaldigitalthermometerandentersthevalue.Theunitwillmakeappropriatecorrectionifneeded.
Editingthecalibrationresults:
Uponsuccessfulcompletionoftemperaturecalibration,thenewcalibratedtemperatureisfurtheravailableforediting.FromtheMAIN MENU,followthesequence:
5 .3 Reaction Cell Temperature
Calibration
31Thermo Scientific Orion 2230 Silica Analyzer User Guide
5 .4 Pumps Calibration
MENU ➔ SERVICE ➔ hARDwARE SETUP ➔ EDIT CALIbRATED VALUES
TheOrion2230SilicaAnalyzerusesperistalticpumpsformovingfluidsthroughthereactioncell.Thepumptubingwillrequirereplacementperiodically.PleaserefertoChapter9fordetails.
Themaximumrateofflowofthefluidisdependentonboththeinnerdiameterofthetubingconveyingthefluidandtherpmofthepump.Thepumpsareusedtomovethefollowingfluids:sample,standard,andreagents.
WHENREPLACINGWITHSAMETYPEFACTORYTUBING,USERMAYBYPASSANYCALIBRATIONANDUSEFACTORYDEFAULTVALUES.IFSUBSTITUTETUBINGISUSED,PUMPCALIBRATIONISREQUIREDTOINSUREACCURATEDOSING.
FromtheMAIN MENU,followthesequence:
MENU ➔ SERVICES ➔ hARDwARE SETUP ➔ CALIbRATE PUMPS
Selectthepumptobecalibrated.Theuserisnowaskedtoprimethepumpwithde-ionizedwaterandtorespondwhenready.Thepumpcalibrationprocedureisabortediftheuserfailstoconfirmreadinesstopumpde-ionizedwaterwithin90seconds.
Thesequencestartswithaprimingsteptoallowfluidtopurgetheairfromthetubing.Theusermayrespondassoonasready.Theprimestepwillstopafter60seconds.Thenextsteppromptaskstheusertobepreparedtocollectthewaterinameasuringcylinder.Thisstepmustbeallowedtocompleteandistimedat60secondstoallowtheusertoenterthevolumecollectedin60seconds(or1minute).Ifthemeasuredvolumeamountisnotenteredwithin90secondsthepumpcalibrationprocedureisaborted.
Uponsuccessfulcompletionofthepumpcalibrationprocedure,theuserhasthechoicetocalibratetheremainingpumpsortoeditthecalibratedvalues.
5 .5 Pumps Calibration
Method
32 Thermo Scientific Orion 2230 Silica Analyzer User Guide
TheAnalyzerhastwoisolatedAnalogoutputs.Theoutputsarerated0/4-20mAwithamaximumloadof900ohms.Theycanbeindependentlycalibrated.
FromtheMAIN MENU,followthesequence:
MENU ➔ SERVICES ➔ hARDwARE SETUP ➔ CALIbRATE ANALOG OUTPUTS
Theuserisaskedtoconnecta900ohmsresistortoeachtwoanalogoutputs.Thecalibrationrequirestheuseofacurrentmeter(0–100mArange)placedin serieswiththeoutputundertest.
Theusermayselectoneoutputatatimeforcalibration.Thecalibrationstartsbyadjustingthefullscaleto20mA.The-/+buttonsareusedtoadjusttheoutputto20mA.TheinstrumentwaitsfortheusertoconfirmthecalibrationbypressingENTER.Ifthereisnouserresponsewithin90seconds,theanalogoutputprocedurecalibrationisaborted.
Uponsuccessfulcalibrationofoneoftheanalogoutputs,theuserhasthechoiceofcalibratingthesecondanalogoutput.
Thecalibratedvaluesfortheanalogoutputsarenotavailableforediting.
Fortestingthecalibratedoutputsuse:
MENU ➔ SERVICES ➔ DIAGNOSTICS ➔ ANALOG OUTPUT ➔ ChANNEL –1:
BypressingENTERtheusermayselectthesignallevelfromthefollowingsettings:
0,4,10,12,and20mA
5 .6 Analog Outputs
Calibration
Thermo Scientific Orion 2230 Silica Analyzer User Guide 33
Chapter 6 Calibration of the �nstrument for Measurement Analysis
IntheAnalysismode,theOrion2230SilicaAnalyzermeasurestheabsorbanceoflightthroughthesample,andcomparestheresultwiththeabsorbanceoflightthroughastandardsolutionofknownconcentration.
TheMethodofAnalysisdefinesthedetectionwavelengthforSilica.Itissetat810nm.
Usingtwoknownstandards,oneoflowconcentrationandtheotherofhighconcentration,theabsorbanceforeachconcentrationismeasuredandtheslopeandinterceptarecalculatedfromthecalibrationcurveasshownbelow:
TheMeasuredAbsorbanceistheresultabsorbanceaftersubtractingtheBlankAbsorbance.
6 .1 Theory of Calibration
Concentration of High Standard
Concentration of Low Standard
Absorbance of Low StandardAbsorbance of High Standard
Intercept
Absorbance
Slope =
AbsorbanceAbsorbance
Concentration
Concentration
Conc
entra
tion
∆
∆
∆∆
À
À
34 Thermo Scientific Orion 2230 Silica Analyzer User Guide
TwomethodsofcalibrationareprovidedintheOrion2230SilicaAnalyzer.
SINGLEPOINTcalibration,usinghighconcentrationstandardalone,andBASELINEcalibrationusingeitheraveryloworzerostandards.
Thebaselinecalibrationisprovidedtoimproveaccuracyforlowlevelmeasurementsandtraceanalysis.
Forsettingupthecalibrationmethod,fromtheMAIN MENU,followthesequence:
MAIN MENU ➔ SETUP ➔ CALIbRATION PARAMETERS
SettheMODEforSINGLEPOINTorBASELINE.
ForCalibration,fromtheMAIN MENU,followthesequence:
MAIN MENU ➔ MEASUREMENT ➔ CALIbRATE ➔ CALIbRATE STANDARD
Forsinglepointcalibration,typicallytheinterceptvalueiszero.Therefore,Concentration=SlopexAbsorbance.
Thetechnicianmaymodifytheinterceptvalue.
Theanalyzerusesa200ppbconcentrationstandardsolutionforcalibration,theabsorbanceofthestandardsolutionismeasuredandtheslopeiscalculatedandstoreduponuser’sconfirmation.
Aftercalibratingusingthesinglepoint,improvedaccuracyaroundthebaseorzeropointmaybefurthertunedbyexecutingabaselinecalibration.Alowconcentrationsolutionisusedanditsabsorbanceismeasuredintheinstrument.Onlytheinterceptisrecalculatedas:
Intercept=Concentration(Low)–[SlopeXAbsorbance(Low)]Thisisusedalongwiththeslopeforsinglepointcalibration.
Thecalibrationdataisavailableforediting.FromtheMAINMENU,followthesequence:
MENU ➔ SETUP ➔ CALIbRATION PARAMETERS
ThemenuscreenshowstwocolumnslabeledREFERENCEandCURRENT.Thedatainthecurrentcolumnistheactivedatabeingusedforslopeandintercept,andthedatainthereferencecolumnisthedatathatthesystemreliesonwhencomparingthenewcalibrationresulttocheckitsalloweddeviationperthespecifiedlimitspecifiedforthetolerancevalue.
6 .2 Methods of Calibration
6 .3 Single Point
Calibration
6 .4 Base Line Calibration
35Thermo Scientific Orion 2230 Silica Analyzer User Guide
TheAnalyzerusesadilutionmethodformeasurementofhighconcentrations.Thecycleisautomaticandhasbeenprogrammedatthefactory.Inmostcasesthefactorydefaultvaluesaresufficientforthespecifiedrangeoftheproduct.Forimprovedaccuracy,theusermaycalibratethedilutionfactorusingahighconcentrationstandardlikea3000ppb.Thiscalibrationisonlyallowedinmanualmode,withuserpromptsandconfirmationrequiredattheendofthecycle.
Extendedrangecalibration:MENU ➔ SETUP ➔ (+) CALIbRATION PARAMETERS
SettheSTANDARDCONC=3000ppb(fora0to5000ppbrange).Andexecuteamanualcalibration:
MAIN MENU ➔ MEASUREMENT ➔ CALIbRATE ➔ CALIbRATE STANDARD
Whenthesystemdetectshighabsorbancelevelofthereactedsolution,abovethesetlimit,itinitiatesthedilutioncycleusingthestandard.Attheendofthecyclethedilutionfactoriscalculatedbasedonthevalueofthestandardusedandtheexistingslopeandintercept.
Attheendofthecalibrationcycletheuserispresentedwiththenewcalculateddilutionfactortoacceptorreject.Thedefaultdilutionfactoris9.0(range8to10).Therangedependsonthepumpflowrate(pump#1)usedtoaddthestandardusedfordilutiondosing.
TheusermayeditthedilutionfactorintheANALYSISPARAMETERSmenu:
MENU ➔ SETUP ➔ ANALYSIS PARAMETERS ➔ (+) (+) PAGE 3 OF 4 ➔ DILUTION PARAMETERS menu
Consultfactorybeforemodifyingthedefaultdilutionsteps.Press(+)toaccessthedilutionsteps.
TheautomaticdilutioncyclemaybeturnedoffusingeitherahigherthanexpectedAUlimit:
ABSHIGHLIMIT = 5AUorbyde-selectingthedilutionevent:DILUTIONEVENT = NOTASSIGNED
Duringtheanalysisofhighconcentrationsamples,thesystemchecksthefinalabsorbancevalueofthereactedsolutionandifitexceedsthesetlimit(ABSHIGHLIMIT)itwillexecutethedilutionstepsandusethecalibrateddilutionfactortocalculatethefinalconcentration.
6 .5 Hi Range Dilution-Factor Calibration
36 Thermo Scientific Orion 2230 Silica Analyzer User Guide
MeasurementintheOrion2230SilicaAnalyzerisdonebymeasuringtheabsorbanceofasinglewavelengthoflightthroughthesample,andcomparingtheresultwiththeabsorbanceofthesamewavelengthoflightthroughastandardsolutionofknownconcentration.Ifthetwoabsorbenciesmatch,thenthesampleconcentrationisequaltothestandardsolution.
Tofacilitatethecomparison,theAnalyzerisfirstcalibratedwithstandardstodeterminetheslopeandinterceptoftherangeofconcentrationitcananalyze,andthenthemeasurementoftheunknownsampleisinterpolatedalongthecalibrationcurvetocalculatetheunknownconcentration.
i.e., Concentration=(SlopexAbsorbance)+intercept.Where SlopeandIntercept-DeterminedfromCalibration. Absorbance-Calculatedthroughmeasurement.
Thetwovariables(SlopeandIntercept)areavailableforeditinginthecalibrationparametersmenuscreen
TheOrion2230SilicaAnalyzerisdesignedtoprovideonlineprocessmeasurementsithastwomodesofoperation.Themodeisalwaysdisplayedatthebottomright,i.e.,belowtheInfofieldandabovethestatusfieldofthedisplay.
ThetwomodesoftheinstrumentareAUTOrestartmode(theinstrumentresumesautoanalysismodeafter0.5Hrofinactivity)andMANUALrestartmode.TheinstrumentawaitstheusertoinitiatetheRUNorAUTOANALYSISmodebypressingtheRUNkey.
6 .6 Theory of
Measurement
6 .7 Mode of
Measurement
Absorbance
Conc
entra
tion
Conc
entra
tion
of S
ampl
e
Measured Absorbance of Sample
ÿ
37Thermo Scientific Orion 2230 Silica Analyzer User Guide
Tochangetheinstrumentmode,fromMAINMENUfollowthesequence:
MENU ➔ SETUP ➔ (+) ANALYSIS PARAMETERS ➔ MODE
Theinstrumenthastwospecialkeystostartandstopthemeasurement.PressingtheRUNkeystartstheautoanalysisimmediatelyandanalysisrepeatsattherateofthetimesetinmeasurementfrequency,whichisselectable.Duringthismode,theinstrumentdisplaysRUNMODEindicatingthattheinstrumentisrunningcontinuously.
Toeditthemeasurementfrequency,fromtheMAIN MENUfollowthesequence:
MENU ➔ SETUP ➔ (+) ANALYSIS PARAMETERS ➔ SAMPLE FREQUENCY
STOP -keyabortsthemeasurementatanystageirrespectiveofthemodeoftheinstrument.
NOTE: RUN is not the mode of the instrument. it indicates that the instrument is performing measurement continuously, at the rate of measurement frequency.
Thedefaultmodeoftheinstrumentisautorestart.Inthismode,iftheinstrumentkeypadisuntouched(i.e.staysidle)fortheperiodofautorestartidletime(defaultvalueof30minutes),thenitautomaticallystartsthemeasurement,themodeisdisplayedasRUNMODEandrepeatsthemeasurementaspertheMeasurementfrequencysetinAnalysisparameters.
IftheSTOPkeyispressedduringthismeasurement,itabortsthemeasurementandwillrestartaftertheidletimecountdownexpires.Tostoptheautorestartoftheinstrumentthemodeshouldbechangedtomanualrestart:
MENU ➔ SETUP ➔ (+) ANALYSIS PARAMETERS ➔ MODE
ThismodeallowsyoutoSTARTthemeasurementsmanually.ToperformmeasurementoncommandfromtheMAINMENUfollowthesequence:
MENU ➔ MEASUREMENT ➔ ANALYzE ONCE
PressingRUNkeyinthismodekeepstheinstrumentinRUNMODEandpressingSTOPkeyduringmeasurementgetsbacktotheMANUALRESTARTmode.Inthismodeiftheinstrumentisuntouchedforaperiodofthestandbymodeidletime(60minutesdefault),itswitchestostandbymode.
6 .8 Auto Restart Mode
6 .9 Manual Restart Mode
38 Thermo Scientific Orion 2230 Silica Analyzer User Guide
Duringstandbyoperation,allfluidicsdevicesareshutdownandtheAnalyzergoesintoastandbyhousekeepingmodeinwhichthepumpsandthevalvesareactivatedonanhourlybasistopreventtubepinching.
WhenmanualrestartmodeisselectedtheAnalyzerautomaticallygoesintostandbyoperationafteronehour(60minutes)ofinactivity.Consultthefactoryshouldyourequireadifferentstandbytimeout;fordetails,refertoChapter7.
NOTE:
Even though the instrument is running continuously and displays rUn MODE, to know the actual rESTArT mode of the instrument from MAin MEnU follow the sequence:
MENU ➔ SETUP ➔ (+) ANALYS�S PARAMETERS ➔ MODE
TheOrion2230SilicaAnalyzerfollowsasequenceofstepsoreventsforperformingtheanalysiscycle.Theyare:
1.CellWash2.SampleCapture3.OpticsTesting4.TestReactionTemperature(whenenabled)5.AddsReagent1and26.BlankMeasurement7.AddsReagent38.CollectDataandOutputResult9.PostWash
Incellwashthesampleispumpedintothereactioncell.TheuserisallowedtoeditthevolumeofthesampletobepumpedbytheparametercalledCELLWASHVOLUME.Toeditthecellwashvolume,fromtheMAINMENUfollowthesequence:
MENU ➔ SETUP ➔ METhOD ➔ CELL wASh VOLUME
Incellwashvolume,thepositivevolumeistopumpthesampleintothereactioncell,andthenegativevolumeistopullbackanamountbelowthedrainafterthecellwashiscompleted.
DuringthecellwashthestirrerisonandtheInfofield(ifEnabled)showsthesamplepumpingprogresswithaprogressbar,alongwiththevolumepumped.
Inthesamplecapturestepthestirrerisswitchedoffandfluidisallowedtosettlefor5secondstodrainanyexcessfromthereactioncelloverflowdrain.
DuringthesamplecapturetheInfofield(ifEnabled)showsthe5secondscountdown.
6 .10 Measurement Cycle
6 .10 .1 Cell Wash
6 .10 .2 Sample Capture
39Thermo Scientific Orion 2230 Silica Analyzer User Guide
Inopticstesting,thesystemcheckstheopticalpathandstatusoftheLEDsource.
DuringthisstageinthecycletheInfofield(ifEnabled)showsthetestopticsprogressbarwhichtimesoutatamaximumperiodof90seconds.
NOTE:
Optics test may flag the following error messages:
“TESTinG OPTiCS, PLEASE CHECK FOr OPTiCS FAiLUrE” Optics test failed due to the non-functionality of the LED source or detector, measurement is aborted, and the info field will flash the error message.
“TESTinG OPTiCS PLEASE CHECK FOr STrAY LiGHT” Stray light reading outside the allowed limits: will flash with a warning message.
“ TESTinG OPTiCS LOw SiGnAL LEVEL . . . CELL MAY BE COATED AnD MAY rEQUirE CLEAninG“
if the test determines that the light energy reading is outside the allowed limits, the measurement is aborted. The cell may require cleaning, or you many need to re-calibrate the optics.
Whenreactioncelltemperaturecontrolisenabled,atthisstageinthecyclethereactioncelltemperatureistestedagainstthesetpoint.
ToenableordisabletheREACTIONTEMPERATURECONTROL,fromtheMAINMENUfollowthesequence:
MENU ➔ SETUP ➔ METhOD ➔ REACTION TEMPERATURE CONTROL
Ifreactiontemperaturecontrolisenabledthenthetemperaturesetpointiseditable.Toeditthetemperaturesetpoint,fromtheMAINMENUfollowthesequence:
MENU ➔ SETUP ➔ METhOD ➔ REACTION TEMPERATURE
ThetemperaturestabilizationisattainedthroughPIDcontrol.Onceitreachesthesetpointitstaysalmostatthesametemperature,untilandunlessthereactiontemperaturecontrolisdisabledorsetpointisaltered.
Thistestisallowedamaximumof240secondsforthereactioncelltoattainthesettemperature.Ifitdoesnotreachstabilityorsetpointintheallowedtimeitwillflagawarningmessageandproceedtothenextstep.TheInfofield(ifEnabled)showsthetemperatureprogressbar.
6 .10 .3 Optics Testing
6 .10 .4 Checking
Reaction Temperature
40 Thermo Scientific Orion 2230 Silica Analyzer User Guide
NOTES:
no variation in temperature for a period of 80 seconds aborts the measurement, and the info field will flash with the error message: “PLEASE CHECK FOr HEATEr FAiLUrE”.
if the temperature is not stabilized within 200 seconds and below the set point, then the info field will flash with the warning message: “CHECKinG THE rEACTiOn TEMPErATUrE. TEMPErATUrE BELOw SET POinT”.
if the temperature is above the set point, then the info field will flash with the warning message: “CHECKinG THE rEACTiOn TEMPErATUrE. TEMPErATUrE ABOVE SET POinT”.
InthiseventareadingisrecordedcalledBLANKABSORBANCE.
Thestabilityoftherecordedreadingdependsonthespecifiedstabilitycriteria.Youhavethechoiceofeditingthestabilitycriteria.TodothisfromtheMAINMENUfollowthesequence:
MENU ➔ SETUP ➔ ANALYSIS PARAMETERS ➔ STAbILITY CRITERIA
DuringthisstagetheInfofield(ifEnabled)showstheblankmeasurementprogressbarwiththeBLANKABSORBANCEreadinginAU.(ThisisarelativevalueandNOTanabsolutevalue.)
FortheSilicameasurement,theeventofblankmeasurementanditssequenceinthemeasurementcycleisdonebeforeaddingthethirdcolordevelopingreagent.
NOTE: when unstable and not meeting the stability criteria, the info field will flash with the warning message “TAKinG BLAnK MEASUrEMEnT, UnSTABLE SiGnAL”.
Inaddreagent,thereagent(s)ispumpedintothereactioncell.YouareallowedtoeditthevolumeofthereagenttobepumpedbytheparametercalledREAGENTVOLUME.
Toeditthereagentvolume,fromtheMAINMENUfollowthesequence:
MAIN MENU ➔ SETUP ➔ METhOD ➔ (+)togotopage2 ➔REAGENT VOLUME(1,2,3)
Inreagentvolume,thepositivevolumeistopumpthereagentintothereactioncell,andthenegativevolumeistopumpbackanyreagentdropsinthepumpdeliverytubeafterthereagentadditioniscompleted.Afterpumpingthereagent,thestirreristurnedonforaneditableperiodoftimecalledSTIRTIMEtomixthereagentthoroughlywithsample.
6 .10 .5 Blank Measurement
6 .10 .6 Add Reagents
41Thermo Scientific Orion 2230 Silica Analyzer User Guide
ToedittheReagentvolume,fromtheMAINMENUfollowthesequence:
MAIN MENU ➔ SETUP ➔ METhOD ➔ (+)togotopage2 ➔ STIR TIME(1,2,3)
DuringtheadditionofreagentthestirrerisonandtheInfofield(ifEnabled)showsthereagentpumpingprogress,withabarandthevolumepumped.Afteraddingthereagentthecountdownofstirtimeisalsoshownalongwiththeabsorbanceofthesamplemixture.
NOTE: if the pump is not assigned to a reagent, it will abort the measurement and the info field will flash with the error message: “PUMP iS nOT ASSiGnED”.
Incollectdata,thefinalreadingisrecordedandthefinalabsorbanceisdisplayedattheendofthecycle.
Thestabilityoftherecordedreadingdependsonthespecifiedstabilitycriteria.Youhavethechoiceofeditingthestabilitycriteria.TodothisfromtheMAINMENUfollowthesequence:
MENU ➔ SETUP ➔ ANALYSIS PARAMETERS ➔ STAbILITY CRITERIA
Nowtheconcentrationiscalculatedasfollowsandthecurrentconcentrationisupdated.
Concentration=(SlopexAbsorbance)+Intercept
WhereslopeandinterceptarefromtheCURRENTcalibratedvaluesoftheinstrumentformeasurement.
NOTE: if unstable and didn’t meet the stability criteria, the info field will flash with the warning message “COLLECTinG DATA, UnSTABLE SiGnAL”.
TheprogrammablePOSTWASHistowashthereactioncellafterthemeasurementandevacuateanyresidueofthereactionmixedsolution.Thiseventisoptionalandcanbedisabled,orthereactioncellcanbewashedwithselectablesourceswitheditablevolumesandstirtimes,withuptofiveprogrammablesteps.Toperformpostwashaftermeasurement,selectANALYSISPARAMETERSforthePOSTWASHEVENT.
6 .10 .7 Collect Data
6 .10 .8 Post Wash
42 Thermo Scientific Orion 2230 Silica Analyzer User Guide
TosetthepostwashparametersfromtheMAINMENUfollowthesequence:
MAIN MENU ➔ SETUP ➔ ANALYSIS PARAMETERS ➔ (+) POST wASh EVENT ➔ POST wASh CYCLES
TheOrion2230SilicaAnalyzerproductstoresmeasurementandcalibrationresultsforupto31days.ThedataforeachsampleanalysiscycleorcalibrationmayberetrievedunderREVIEWRESULTSmenuscreenwiththecorrespondingtimestampandotherrelatedoperatingparameters.
Sampleanalysisdatastoresthedate,time,concentration,units,temperature,wavelengthandstatus(GoodorFault)alongwiththefaultsforeachmeasurementcycle.
Calibrationdatastoresdate,time,standardconcentration,slope,intercept,temperature,wavelengthandstatus.
Thestoredresultsaredisplayedwithrespecttotheselectableperiodoftimesuchas:Last12Hours,Last1Day,Last2Days,Last3Days,Last1week,Last2weeks,Last3weeks,Last4weeks,All,whichisselectedasDATARANGE.Todisplaythemeasurementresultsfromthemainmenufollowthesequence:
MENU ➔ VIEw DATA AND ALARMS ➔ REVIEw ANALYSIS RESULTS ➔ DISPLAY
NOTE: Validation results are OnLY reported on the display and the rS-232 at the time of running the validation. Validation cycles are not stored and are not reported on the analog outputs.
ThestoredmeasurementresultscanbeprintedontotheHyperTerminal(minicominLinux)throughtheRS-232portwithintheselecteddatarange.Theselecteddatarangeisapplicabletodisplaying,printingandplottingofmeasurementresults.
6 .10 .10 Printing Results
6 .10 .9 Measurement Data
Thermo Scientific Orion 2230 Silica Analyzer User Guide 43
Chapter 7 Menu OverviewThe four major menu sections are described as follows:
Setup - Forconfiguringtheinstrument’sfunctionalityandoperationalparameterssuchas:measurementrange,measurementfrequency,displaycontrast,date,time,etc.DetailsofthefourmajorMenusectionsaredescribedinSection7.5.
Measurement - Forperformingthemanualanalysis,calibration,validationandcleaningcycles.
View Data and Alarms - Forreviewingthecollecteddataofuptothelast31daysintextandgraphicsmode,andforreviewingandresettingthevariousAlarmsandFaultstriggeredduringthelatestcycleperformed.
Services - Forsystemadministration,hardwaresetup,primefluidicsanddiagnostics;andcontrolofthevariousdevicesincludingLED,pumps,valves,inputlevelsensors,relaysandanalogoutputs.
NOTE: Press the STOP key to exit Auto RUN mode to access the Service Menu.
Alarms Menu
Allowsconfigurationofalarmparameters(type,valueandhysteresis).Alsoallowsassignmentofindividualalarmstofaults.
Relays Mapping MenuAllowsmappingofmultiplealarmstoindividualrelays.
Analog Outputs MenuAllowsconfigurationofindividualanalogoutputs.
Inputs Menu Allowsconfigurationoffiveinternalinputs(forfluidicssensors)andtwoexternalinputs(forinstrumentcontrol).
Display Parameters MenuAllowsconfigurationofdisplayrelatedparameters.
7 .1 Setup Menu
44 Thermo Scientific Orion 2230 Silica Analyzer User Guide
Calibration Parameters MenuAllowsconfigurationofparametersforcalibration.ThemenushowstwocolumnstaggedasREFERENCEandCURRENT.TheparameterslistedundertheREFERENCEcolumnaresetforvalidatingautocalibrationsrepeats.ThenewcalibrationresultsaretestedagainstthereferencecalibrationparameterslistedundertheREFERENCEcolumn.TheCURRENTcolumnshowsthecurrentoperatingparametersthatarebeingusedbytheinstrumentwhencalculatingtheanalysismeasurementresult.CalibrationsaresimilartoanalysiscyclesandareonthemethodsetinMETHODPARAMETERSMenu.
Validation Parameters Menu Allowsconfigurationofparametersforvalidation,choiceofvalidationsource,andrepetitionsrequired.
Analysis Parameters Menu Allowsconfigurationofparametersforinstrumentmode,measurementfrequency,measurementunits,analysislabel,cellwash&reagentflowrates,stabilitycriteriaandpostwashevents.Separatepostwasheventscanbeprogrammedforanalysis,calibration,andvalidationcycles.AllpostwashcyclescanbeturnedOFFwhensettothe“NOTASSIGNED”.
Cleaner Parameters Menu Allowsconfigurationoftheparametersformanualorperiodiccleaningofthereactioncell.
Temperature Units Menu Allowsselectionofunitsfordisplayingthetemperaturereadings.ChoiceofFahrenheitorCelsius.
Method Parameters Menu Allowsconfigurationoftheparametersformethodandorderoftheeventsforanalysisandcalibration.Consultfactorybeforemodifyinganyparameter.
Date and Time Menu Allowsconfigurationofthedateandtimeformatandeditingofthecurrentdate.
Daylight Saving AllowsconfigurationofthedaylightsavingforUSA,Europeandcustomsettings.Whendaylightsavingisenabledthesystemwillcorrectthetimebasedonthesettingsshownbelowtheselection.
Recovery After Power Failure Menu Allowsconfigurationoftheinstrumentafterpowerfailurerecovery.
45Thermo Scientific Orion 2230 Silica Analyzer User Guide
Ethernet Parameters Menu AllowsconfigurationofEthernetParameters.
RS232 Parameters Menu AllowsselectionoftheBAUDRATEfortheRS232port.Defaultsettingof115200Baud.
Language Menu Allowsconfigurationoftheinstrument’slanguage.CurrentreleaseisENGLISHlanguage.(ConsultFactoryforotherlanguages).
Analyze Once Allowsonemanualanalysistobeperformed.Numberofrepeats(default1)canbeeditedintheANALYSISMENU.Whenperformingamanualanalysis,theuserisrestrictedfromnavigatingthroughothermenus.
Calibrate Allowsmanualcalibrationusingstandardsolutionstobeperformed.TheslopeandinterceptvaluesobtainedheremaybeeditedintheCALIBRATIONPARAMETERSmenu.Whenperformingamanualcalibration,theuserisrestrictedfromnavigatingthroughothermenus.
Validate Allowsmanualvalidationoftheinstrumentwithaknownsampleorstandard.Whenperformingvalidation,theuserisrestrictedfromnavigatingthroughothermenus.
Clean Allowsmanualcleaningofthereactioncell.Theparameterssetforthemanualcleanarealsousedforautoclean.Whenperformingmanualcleaning,theuserisrestrictedfromnavigatingthroughothermenus.
Review / Reset Alarms AllowsusertoreviewandselectivelyresetthefaultsandalarmstriggeredduringthelatestAnalysis/Calibration/Validation/Cleaningcycleperformed.
Review Calibration Results Allowstheusertoreviewandprintthecalibrationresultsfromthestoredcalibrationdatabase,usingaselectabletimescale.Thedatabaseholdsdataoflast31days.
Review Analysis Results Allowstheusertoreviewandprinttheanalysisresultsfromthestoredanalysisdatabase,usingaselectabletimescale.Theselectedresultscanbedisplayedintextorgraphmodes.Thedatabaseholdsdataoflast31days.
7 .2 Measurement Menu
7 .3 View Data and
Alarms Menu
46 Thermo Scientific Orion 2230 Silica Analyzer User Guide
Theviewgraphhasthefollowingoperationalkeys:F1 - EscF2 - Un-ZoomF3 - ScrollLeftF4 - ScrollRightENTER - Zoom- / + -CursormovementMENU - noaction
Zoomworksas(Xhrsdisplayed)/4,aroundthecursorposition.Un-zoomistheexactoppositeofzoom.
Delete All Calibration Results Allowstheusertodeletetheentirecalibrationdatabaseafterconfirmation.
Delete All Analysis Results Allowstheusertodeletetheentireanalysisdatabaseafterconfirmation.
NOTE: To enter the Service menu, the instrument must be in manual mode. (Press STOP)
System Administration Menu Allowsaccesscontroltotheinstrumentforuptosixusersandonetechnician,allwithindividualpasswords.
Technicianalonehasthepermissiontoadd,deleteandsetaccesspermissionofanyuser.Eachusercanhaveaccesstoeditownusernameandpassword.
Menusalsoallowtechniciantochangefactorysettings(factorypasswordrequired),printingoptions,instrumentlabel,andtoviewinstrumentsetupdetails.
hardware Setup Menu AllowstechniciantocalibrateseveralhardwarecomponentsoftheAnalyzer.
Calibration results are editable by the technician.
Prime Fluidics Menu AllowsusertoprimethefluidicslinesoftheAnalyzer.
Diagnostics Menu Diagnosticsprovideadirectmechanismtoindividuallytestdevicessuchaspumps,valves,relays,opticsandstirrer.Turnvariouspumpsandvalvesonandofffortestingorduringadjustment.Maximumoftwodevicescanbeturnedonatthesametime.Thisfunctionalsoprovidesaquickcheckoftheopticssubsystem.
7 .4 Service Menu
47Thermo Scientific Orion 2230 Silica Analyzer User Guide
Edit User MenuAllowstechniciantoeditusernames,userpasswordsanduserpermissions,formeasurement,calibration&validation.Allowsusertoedithisownusernameandpassword.
Delete User MenuAllowstechniciantodeleteauserfromtheuserslist.
Add User MenuAllowstechniciantoaddanewusertotheuser’slist.Uptosixdifferentusersmaybeadded.
Factory Settings MenuAllowstechniciantorestorefactorydefaults,updatefactorydefaultsandeditfactorysettings.
Instrument Details MenuAllowsviewingofInstrumentdetails:
•Softwareversion •Serialnumber •Modelnumber
Analysis Label Change MenuAllowstechniciantoedittheanalysislabel.
Printing Options MenuAllowstechniciantoprintthesystemconfiguration,analysisparametersandmethodparameters.Alsoallowstechniciantoenable/disableonlineprinting.SeeAppendixAforthedefaultanalyzerparameterslistingprintouts.
Calibrate ThermistorAllowstechniciantocalibratethethermistor.Onlyrequiredifthefactory-installedthermistorhasbeenservicedorreplaced.Tocalibratethethermistor,userisrequiredtoenterthereactionblocktemperatureusinganexternalreferencethermometer.
Calibrate PumpsAllowstechniciantocalibratetheflowrateofthepumps.
Calibrate OpticsAllowstechniciantocalibratetheoptics.
Calibrate Analog OutputsAllowstechniciantocalibratethetwoanalogoutputs.
Edit Calibrated ValuesAllowstechniciantoeditthecalibratedresultsforfine-tuning.
7 .5 System
Administration Menu
7 .6 Hardware
Setup Menu
48 Thermo Scientific Orion 2230 Silica Analyzer User Guide
Auto Prime AllThismenurunsanautomaticprimingsequenceforallreagents,standardsandsamplebasedontheprimingvolumespecifiedbytheuser.
Prime Input Source (Sample/Reagent/...)Allowstechniciantoselectivelyprimefromtheavailablelistofdevices.
Priming VolumeAllowstechniciantochangethevolumeusedforpriming.
Press + to navigate to NExT pageThenewreagentsblankistheZEROinterceptoroffset.Auserentryherewillaffectthezerocalibrationvalueofthecalibrationparameters.
Set NEw reagents bLANKAllowstheoperatortosettheBLANKvaluefortheNEWreagents.Thisvalueisusuallysuppliedwiththenewbatchofreagentsanditsvaluemayvarybetween0to-10.Defaultvalueis-1.0.PLEASENOTETHE(-)SIGN.
Pumps, Valves and StirrerAllowstechniciantoselectivelycheckthefunctionalityofthepumps,valvesandstirrer.
Inputs StatusAllowstechniciantocheckthestatusoftheexternalInputs.Onceafunctionhasbeenassignedtoadigitalinputitdisappearsfromthenextmenuofavailablefunctionsforthenextinput.
External ControlTheAnalyzercanbedirectedtoinitiateananalysisorcalibrationcycleorchangeoperatingmodeunderthecommandofanexternalcontrollerusingthedigitalinputlines.Acommandonthedigitalinputisrepresentedbyacontactclosure(astheprogrammedEVENT)thatoccursformorethan1second.Anewcommandreceivedsupersedesthepreviouscommand;inessenceitwillinterruptthepreviouscommandtoexecutethenewcommand.
•AnalyzeOnce•RUN/STOP•Calibrate•Validate(thesourceisdefinedinthevalidationmenu)•Notassigned
7 .7 Prime Fluidics Menu
and New Reagents Blank
7 .8 Diagnostics Menu
49Thermo Scientific Orion 2230 Silica Analyzer User Guide
Validate Grab SampleThisfunctionisthesameasanalyzeonceexceptthattheresultdatadoesnotgetsenttotheanalogoutput(4-20mA),itisreportedonthescreenandtheserialoutputs.
RelaysAllowstechniciantoselectivelycheckthefunctionalityofthefourrelays.Therelaystatusisshownnexttotheselectedrelay.PressingtheENTERkeytogglesthestateoftherelay.Therelayinitialstatuswillbeupdateduponleavingthisdiagnosticsmenu.
Analog OutputAllowstechniciantoselectivelycheckthetwoanalogoutputs.Previousanalogoutputsettingwillbeupdateduponleavingthisdiagnosticsmenu.
LED and Detector ControlAllowsthetechniciantocheckthefunctionalityoftheopticssystemanditsindividualcomponents.AlsoallowsadjustingthedetectorpreampgainfortheselectedLED.AnynewsettingofthegainoroftheintensityoftheLEDwilloverrideanypreviousopticscalibrationsetting.
CONSULTFACTORYbeforeoverridinganysetting.
ThisdiagnosticsmenuallowsdirectreadingofthemVleveldetectedatthepreampanditshowsthecorrespondingabsorbancecalculated.Theinternalelectronicsenclosuretemperatureandthereactioncelltemperaturearealsodisplayedforreference.
Thedefaultpasswordsareasfollows:Technician - 123***User 1 - 111111(6ones)User 6 - 666666(6sixes)
Ifauserpasswordisforgotten,thetechniciancandeletetheuserandthenaddtheuserback.
Ifthetechnicianpasswordisforgotten,callcustomersupport:1.800.225.1480.
7 .9 Default and Lost or
Forgotten Passwords
50 Thermo Scientific Orion 2230 Silica Analyzer User Guide
Figure 14 Shows the UnLOCK Keypad
51Thermo Scientific Orion 2230 Silica Analyzer User Guide
7 .10 User �nterface Tree
MENU NAVIGATION
Setup Measurement View Data & Alarms Services
(+) Page 1 Analyze Once Review / Reset Alarms System Administration
Alarms Calibrate Review Calibration Results Hardware Setup
Relays-Mapping Validate Review Analysis Results Prime Fluidics
Analog Outputs CleanDelete All Calibration Results
Diagnostics
Inputs Delete All Analysis Results
Display Parameters
Temperature Units
(+) Page 2
Analysis Parameters
Validation Parameters
Cleaner Parameters
Calibration Parameters
Method
(+) Page 3
Date & Time
Daylight Saving
Recovery after Power Failure
Ethernet Parameters
RS232 Parameters
Language
(+) Press + sign to go to next page
52 Thermo Scientific Orion 2230 Silica Analyzer User Guide
MENU
Measurement View Data & Alarms Services
Analyze Once Review / Reset Alarms System Administration
Press Enter Reset All Alarms Reset All Faults
Edit User User name and Password
Delete User User Name X
Calibrate Add User Name, Password, Permission
Calibrate Standards
Review Calibration Results Factory Settings Restore, Update, Edit **
Display Instrument Details Software VersionModel and Serial NumberValidate Print
Press Enter Data Range Selection
Clean Review Analysis Results Unit Label Change
Press Enter Display Printing Options Print System Configuration
View Graph Print Analysis Parameters
Print Print Method Parameters
Data Range Selection Print Cleaner Parameters
Online Print : Enable/Disable
Delete All Calibration Results
Delete All Analysis Results Hardware Setup
Calibrate Thermistor **
Calibrate Pumps Pump 1, 2, 3
Calibrate Optics
Calibrate Analog Outputs **
Edit Calibrated Values
Prime Fluidics
Auto Prime All
Prime Standard
Prime Reagent (1, 2, 3)
Prime Sample
Priming Volume (+/-)
Press (+) to go to page 2 New Reagents Blank
Diagnostics
Pumps, Valves and Stirrer
Inputs Status
Relays
Analog Outputs
LED & Detector Control ** Consult Factory first
53Thermo Scientific Orion 2230 Silica Analyzer User Guide
MENU
Setup
Page 1 (+) Page 2 (+) Page 3
Alarms Analysis Parameters Date & Time
Display All + Page 1 (+) Page 2
Alarm 1 thru 7Type, value, Hysterisis
Restart Mode Post Wash Event Daylight Saving
Measuring Frequency Post Wash Cycles
Relays-Mapping Measuring UnitsSource, Volume, Stir Time
Recovery after Power Failure
Ion Label Power Fail duration
Analog Outputs No of Repeats(+) Page 3 and (+) Page 4
Power Fail Clean Cycle
Analog Output 1, 2 Type, Low, high Stability Criteria Dilution Parameters
Ethernet Parameters
Inputs Validation Parameters IP Address
Level Sensors Sensor 1 – 5 Validation Source Server IP
External Inputs Input 1, 2 No of Repeats
RS232 Parameters
Display Parameters Cleaner Parameters Baud Rate
Contrast Cleaner Source
Back Light On time Volume to Add Language
Display Help Clean Flow Rate
Display Info Stir Time
Result Decimal Point Frequency
Large Font Screen
Calibration Parameters
Standard
Temperature Units Mode (+) Press + sign to go to next pageFahrenheit, Celsius Frequency
Slope
Intercept
Calibration Repeat Tolerance
Method
Cell Wash Volume
Reaction Cell Control
Reaction Cell Temperature
Wavelength
Blank Measurement (+) Page 2
Reagent 1-3 and Stir Time
(+) Page 3
Cell Wash Flow rate
Reagent 1-3 flow rate
54 Thermo Scientific Orion 2230 Silica Analyzer User Guide
55Thermo Scientific Orion 2230 Silica Analyzer User Guide
Terms Description Possible Values, Options
ALARM
Type Assigning a fault type to the Alarm. Low Limit, High Limit, Fluidics Fault, General Fault, not Assigned
ValueBoundary value for triggering an Alarm, Only for Low Limit and High limit type.
0 to XXXXX (XXXXX Value depends on precision selected in Setup ➔ Display parameters ➔ result Decimal Point)
Hysteresis Value gap or difference before an event change.
0 to XXXXX (XXXXX Value depends on precision selected in Setup ➔ Display parameters ➔ result Decimal Point)
RELAY Up to FOUr spare relays available.
MODE Allows Fault On or OFF operation. normal or Fail safe
ANALOG OUTPUT
Type normal or reverse acting. 0 to 20, 4–20, 20-0 or 20–4
Low Low concentration value.
High Hi concentration value.
ExTERNAL �NPUT
Label User programmable label.
Event Dry Contact event. Close/Open or Open/Close
Action Selected function. Analyze Once, run/Stop, Calibrate, Validate, not assigned
D�SPLAY PARAMETERS
Contrast To adjust the display contrast. 120 to 200
Back Light On-Time Screen saver time duration for backlight of display.
0 to 255 minutes, 0 for Continuously On
Display Help Help message for the user to navigate between the menus. Enabled, Disabled
Display info Message showing the status of the current cycle performing. Enabled, Disabled
results Decimal Point Selectable precision to show the analysis result. XXXX.X, XXX.XX, XX.XXX, X.XXX, XXXXX, Auto
Large Fonts Screen To display the result with large fonts. Enabled, Disabled
CAL�BRAT�ON PARAMETERS
Standard The concentration of the standard used for calibration.
0 to XXXXX (XXXXX Value depends on precision selected in Setup ➔ Display parameters ➔ result Decimal Point)
Mode Mode represents the calibration type. Single-Point, Base-Line
Frequency The time duration between two calibration cycles.
Every 6Hr, 12Hr, 24Hr, Sun, Mon, Tue, wed, Thu, Fri, Sat, Manual and Editable 0Hr to 99.99Hr
Chapter 8 Terms Definition
56 Thermo Scientific Orion 2230 Silica Analyzer User Guide
Terms Description Possible Values, Options
Start Time Calibrations start time, except for manual and editable frequency. Depends on Time format in Setup ➔ Time Menu
Slope
The CUrrEnT Slope value represents the coefficient that is multiplied by the absorbance to calculate the concentration of the sample +/- intercept value.
intercept reference and current numbers based on the base line calibration cycle.
Calibration Tolerancerepeat tolerance tested during auto calibration cycles to accept or reject automatic calibrations.
VAL�DAT�ON PARAMETERS
Validation Source Fluid source for validation. Sample, Standard, Grab Sample, not Assigned
number of repeats number of times that the validation is to be performed. 0 to 8
ANALYS�S PARAMETERS
Mode instrument rESTArT mode. Auto restart, Manual restart
Measuring Frequencyin the rUn mode, the frequency is the time duration between two analysis cycles.
15, 30, 45 Minutes, 1Hour, 0.20 Hr are Selectable and editable Minutes range 00.00 to 99.99Hr
Measuring Units Units for the analyzed sample. Six characters long. Alphabets, numerals and special symbols are allowed.
Analysis Label Six characters long. Alphabets, numerals and +, - are allowed.
Cell wash Flow rate The rate at which the sample is pumped to wash the cell.
Depends on the inner diameter of the pump tubing. it is derived from the calibration of pumps.
reagent 1–3 Flow rate The rate at which the reagent is pumped to wash the cell.
Depends on the inner diameter of the pump tubing. it is derived from the calibration of pumps.
Stability Criteria
Defines the stability level of the data before accepting it. The stability criteria is a measure of stability of the absorbance level rate of 0.001 AU per sec as specified by the number 1 thru 10.
1 thru 10
Post wash Eventwashing the cell after completion of the Event selected with the following Source, Volume and Stir time.
Analysis, Calibration, Validation, not Assigned
Post wash Cycle Source Fluid for post washing the reaction cell. Sample, reagent, Cleaner, Standard
57Thermo Scientific Orion 2230 Silica Analyzer User Guide
Terms Description Possible Values, Options
Post wash Cycle Volume
The amount of fluid pumped for post washing. -99.9 to +99.9 ml
Post wash Cycle Stir Time
The amount of time the fluid is stirred for post washing the cell. 0 to 999 Sec
CLEANER PARAMETERS
Cleaner Source Fluid for cleaning the cell. Sample, reagent, Cleaner, Standard, not Assigned
Volume to add The amount pumped for cleaning. 0 to 255 ml
Clean Flow rate The rate at which the cleaner source is pumped to clean the cell.
Depends on the inner diameter of the pump tubing. it is derived from the calibration of pumps.
Stir Time The amount of time the fluid is stirred for cleaning the cell. 0 to 999 Sec
Frequency The time duration between two cleaning cycles. Manual, Daily, weekly, Monthly
Start Day/DateFor weekly cleaning the day of cleaning. For Monthly Cleaning the date of cleaning.
Day - week days Date - Current month dates (1 to 31)
Start Time The time to start the cleaning on the day or date selected. 00:00 Hr to 23:59 Hr
TEMPERATURE UN�TS
Units Units of temperature. Celsius, Fahrenheit
METHOD PARAMETERS
Cell wash Volume The amount of sample pumped to wash the cell. 1 to 100 ml
reaction Temperature Control PiD temperature control. Enable, Disable
reaction Temperature The temperature set point. 0 to 60
wavelength The LED wavelength. nir3 810nm
Blank Measurement Allows the blank measurement in the Analysis cycle.
Disable, Before reagent/1, reagent2, Before reagent3
reagent Volume The amount of reagent pumped in to the cell. 0 to 6.00 ml
Stir Time The amount of time the fluid is stirred after pumping the reagent. 0 to 999 Sec
DATE
Format Selectable date format. MM/DD/YY, YY/MM/DD, DD/MM/YY
Current Date Present Date, validated depending on the format. Legible date
T�ME
Current Time Present time. Legible time depending on time format
Format Selectable time format. 12 Hour, 24 Hour
58 Thermo Scientific Orion 2230 Silica Analyzer User Guide
Terms Description Possible Values, Options
RECOVERY AFTER POWER FA�LURE
Power Fail Duration recovery duration after power fail. 0 to 99.99 Hr
Power Fail Clean Cycle Performing Clean cycle after recovering from power fail. Enable, Disable
ETHERNET PARAMETERS
iP Address instrument iP address. Editable
Server iP
network iD
RS232 PARAMETERS
Baud rate The rate at which the data is transferred through serial port. 115200 (default)
LANGUAGE
Language Language in the instrument. English, other languages, consult factory
REV�EW / RESET ALARMS
reset All Alarms? resets all the raised alarms, which are shown below this option. none
reset All Faults? resets all the faults, which are shown below this option. none
REV�EW CAL�BRAT�ON RESULTS
Display Displays the Calibration results for the selected Data range. none
Print Prints the Calibration results on to the PC for the selected Data range. none
Data range The time interval in which the Calibration results are to be retrieved.
Last 12 Hr, 1 Day, 2 Days, 3 Days, 1 week, 2 weeks, 3 weeks and All
REV�EW ANALYS�S RESULTS
Display Displays the Analysis results for the selected Data range. none
View Graphs Displays the Linear graph for the selected Data range of Analysis records. none
Print Prints the Analysis results on to the PC for the selected Data range. none
Data range The time interval in which the Analysis results are to be retrieved.
Last 12 Hr, 1 Day, 2 Days, 3 Days, 1 week, 2 weeks, 3 weeks and All
DELETE ALL CAL�BRAT�ON RESULTS
Deletes the entire calibration results database. none
DELETE ALL ANALYS�S RESULTS Deletes the entire analysis results database. none
59Thermo Scientific Orion 2230 Silica Analyzer User Guide
�tems Description Possible Values, Options
ED�T USER
User name Change the user. 6 users and one Technician are allowed
Password Change the password of current user. Six characters long A–Z, 0–9, * and space are allowed
Measure Permission Change the Measurement permission for the current user. Enabled, Disabled
Calibrate Permission Change the Calibration permission for the current user. Enabled, Disabled
Validate Permission Change the Validation permission for the current user. Enabled, Disabled
DELETE USER
User name Delete the current user displayed opposite this field. All users can be deleted except Technician
ADD USER
User name Add a new user to the users list. User name is 6 characters long and only 6 users are allowed.
Password Password of the new user entered. 6 characters long A–Z, 0–9 * and space are allowed
Measure Permission Measurement permission for the new user. Enabled, Disabled
Calibrate Permission Calibration permission for the new user. Enabled, Disabled
Validate Permission Validation permission for the new user. Enabled, Disabled
Factory Settings Select this option to enter into Factory setup. none
�NSTRUMENT DETA�LS
Software Version The Current running Software Version of the instrument. none
Model number Model number of the instrument. none
Serial number Serial number of the instrument. none
UN�T LABEL CHANGE
Label name Change the Title of the instrument. Maximum 9 characters A–Z, 0–9, / , - , and space are allowed
PR�NT�NG OPT�ONS
Print System Configuration Prints all the system parameters. none
Print Analysis Parameters Prints only the analysis parameters. none
Print Method Parameters Prints only the method parameters. none
Print Clean Parameters Prints only the clean parameters. none
Online Print Prints the Date, time and concentration after each analysis cycle. Enabled, Disabled
60 Thermo Scientific Orion 2230 Silica Analyzer User Guide
�tems Description Possible Values, Options
ED�T CAL�BRATED VALUES
Pump1, Pump2, Pump3 To change the flow rate of the pumps after their Calibration. 0.10 to 60.00 ml
Optics Dark The level detected as dark light level. 0 to 300 mv
Optics Xref The level used as reference potential for light energy level. 0 to 2400.00 mv
PR�ME FLU�D�CS
Priming Volume The volume of the fluid selected for priming. 0 to 99.9 ml
PUMPS, VALVES & ST�RRER
Pump1, Pump2, Pump3 To check the functionality of Pumps. On, OFF
Valve1, Valve2, Valve3 To check the functionality of Valves. On, OFF
Stirrer To check the functionality of Stirrer. On, OFF
RELAYS
relay1, relay2, relay3, relay4 To check the functionality of relays. On, OFF
ANALOG OUTPUTS
Channel1, Channel2 To test the functionality of Analog outputs for different selectable currents. 0mA, 4mA, 10mA, 12mA, 20mA
LED & ADC
Select LED To test the functionality of LED. nirx 810, none
intensity intensity of the LED. 0% to 100%
Optics Amplifier Gain Changing the gain of the amplifier. 0 to 255
61Thermo Scientific Orion 2230 Silica Analyzer User Guide
MAinTEnAnCE SCHEDULE FrEQUEnCY
COnDUCT rEGULAr MAinTEnAnCE TO EnSUrE THE AnALYZEr PErFOrMS PrOPErLY.
Every week inspect in / out tubes, reagent bottle, and all tubes for fouling.
Every month inspect tubing and ensure that it is not blocked. replace reagents if necessary.
Every 6 months replace Pump and valve tubing. Use factory supplied maintenance kit. inspect for any fouling of the reaction cell.
Every 12 months Check the tubing, replace as needed.
Chapter 9 Maintenance Schedule
9 .1 Changing Reagents
CAUTiOn
Wear Gloves or Protective Equipment your site may require when changing reagents
Press the STOP key.
Press The MENU key
remove the cap from each bottle and change them over one by one. Do nOT re-use or refill old or existing bottles. Dispose of the used bottles and contents safely and according to local regulations.
Tighten bottle caps and route the tubing properly making sure that it does not get pinched when placing the bottle in the enclosure.
Press MENU key and select:
SERV�CES ➔ PR�ME FLU�D�CS ➔ AUTO PR�ME
repeat AUTO PriME if you notice air bubbles in the tubing.
Each reagents kit comes with a rEAGEnT BLAnK value.
After changing the reagent you must enter the new value.
Press (+) to go to next page and ENTER the value of the reagent blank offset which is detailed on the form that comes with the new reagents kit.
NOTE: The reagent blank value has a (-) sign to it. Default value is -1 .0
62 Thermo Scientific Orion 2230 Silica Analyzer User Guide
9 .1 .1 �nstalling Tubing Maintenance kit
Changing the Instrument Tubing
Itisrecommendedthatthepumpheadtubingbereplacedataminimumtwiceayear,orasneeded.
Thesixmonthmaintenancekit(2230MK)containscompletelyassembledtubingsets.
Tochangethetubingproceedasfollows:
PresstheSTOPkey.
Tominimizecontactwiththereagents,itishighlyrecommendedthatyouemptyoutthecontentsofthetubingusingthePRIMEbackmethodasdescribedbelow:
PressMENUandscrolldowntoSERVICESandthenselectPRIMEFLUIDICSandthensetthePRIMINGVOLUMEto-5ml.Thisstepallowstheusertopumpbackthereagentscontentsoftheconnectingtubing,includingtheperistalticpumptubing.
ScrolluptoeachofthereagentsandpressENTERtopumpbacktheselectedreagentfluidintoitscontainer.Repeatforeachreagent(1,2&3).
Disconnectthetubing.Replaceoneatimetoreducetheriskoferrorsinconnection.
Removetubinginfrontofthevalvesbygentlypullingitoutofthegroove.
Pressinonthecenterbuttonandthenpullthetubeoutofthegroovetoremovethereartube.RefertoChapter3formoredetails.
ReplacethepumptubingasdescribedinSection3.2.
Replacewithidenticaltubingfromthemaintenancekitbyreversingtheprocedure.Applysiliconelubricant(suppliedwiththekit)tothepumprollersbeforere-installingthetubing.
NOTE: PR�ME FLU�D�CS menu the (-) sign for the priming volume will be reset to (+) upon next menu entry.
63Thermo Scientific Orion 2230 Silica Analyzer User Guide
Figure 15Maintenance Kit Assembly Detail
64 Thermo Scientific Orion 2230 Silica Analyzer User Guide
9 .2 Cleaning the Sample
Flow Chamber
Itisrecommendedthatyouinspectthesampleflowchamberatleastonceayeartocheckforanyflowblockage,andtoinspecttheoperationofthefloatlevelsensor.Thesamplechamberislocatedontherightlowercornerofthefluidicsenclosure.
Thechambercanberemovedforserviceandcleanup.
9 .3 Stand-By Mode and Extended Shutdown
ToPuttheSysteminStand-bymodemakesuretopresstheSTOPkeyandchecktheDISPLAYfieldnexttoF4toshowMANUALRESTART.IfAUTORESTARTisselected,youneedtosettheModetoManual:
MENU ➔ SETUP ➔ + ANALYSIS PARAMETERS ➔ MODE
ThesystemwillswitchtoStand-Byafter60minutesofnokeypadactivity.InStand-BytheAnalyzerwillrotatethepumpsandexercisethepinchvalvesperiodicallytoavoidpinchedtubingfromsealing.
Extended Shutdown. Toavoidclogging,flushtheentiresystemwithdistilledwaterbeforeshutdown.DisconnectandemptyoutthereagentsbottlesandflushthelinesusingFLUIDICSPRIMEmenu.Releasethepressureontheperistalticpumptubingandremovethetubesfromthepinchvalves.
SAMPLE IN
SAMPLERETURN
SAMPLE OUTTO PUMP
LEVELSENSORFLOATSWITCH
SAMPLECHAMBERMOUNTINGCLIP
Figure 16Sample Chamber Detail
65Thermo Scientific Orion 2230 Silica Analyzer User Guide
UsethetablebelowtodiagnoseandcorrectAnalyzerproblems.Table 3 Error Messages
Messages Cause Solution
Login as Technician or Operator.
User has not logged in, and trying to access restricted operations.
User has to login either as a Technician or Operator to access the restricted operations.
User is logged in, but does not have adequate privileges to access the restricted operations.
User is not authorized to access the restricted operations.
Password mismatch, Enter to continue. incorrect password has been entered.
Press “Enter” to cancel the wrong entry and enable re-entry of correct password. if password is lost consult factory for backdoor access code.
Measurement is in progress. Trying to perform measurement when it is already under way. wait for Measurement to complete.
User programmable measuring frequency.
Editing option of Analysis frequency is selected. Edit the Analysis frequency.
User programmable calibration frequency.
Editing option of Calibration frequency is selected. Edit the Calibration frequency.
Two or more inputs are of same type. if same type of solution is assigned to two or more inputs. reassign solution to inputs.
Error in Date, Press F1 to Continue. Date entered is invalid. Press F1 and re-enter valid Date.
Error in Hours, Press F1 to Continue. Hours entered is invalid. Press F1 and re-enter valid Hours.
More than 6 users are not allowed, delete a user to add further. Technician trying to add more than 6 users.
More than 6 users are not permitted at any time. To add a new user after 6, delete an existing user.
no records to delete. User is trying to delete Calibration/Analysis records, when there are no records stored. none.
Are you sure you want to delete all analysis results?
User is trying to delete all Analysis records. (At least one Analysis record is available in the database.)
none. Deleted records are not recoverable.
Are you sure you want to delete all calibration results?
User is trying to delete all Calibration records. (At least one Calibration record is available in the database.)
none. Deleted records are not recoverable.
no records to display.User is trying to display records, and no records are available in the selected data range.
none.
no records to print.User is trying to print records, and no records are available in the selected data range.
none.
Only one record to display. User is trying to view graph when only one Analysis record is available in the database. none.
Problem with memory allocation insufficient memory in the system or system is busy.
Do not try to repeatedly view the graph. Try again after some duration or after performing one more measurement cycle.
66 Thermo Scientific Orion 2230 Silica Analyzer User Guide
Measurement/Calibration/Validation Messages
Messages Cause Solution
reagent1 empty.Performing Analysis/ Calibration/ Validation with two or three reagents and reagent1 is exhausted.
Fill the empty reagent1 bottle with reagent1.
reagent2 empty.Performing Analysis/ Calibration/ Validation with two or three reagents and reagent2 is exhausted.
Fill the empty reagent2 bottle with reagent2.
reagent3 empty.Performing Analysis/ Calibration/ Validation with two or three reagents and reagent3 is exhausted.
Fill the empty reagent3 bottle with reagent3.
reagent1, reagent2 are empty.Performing Analysis/ Calibration/ Validation with two or three reagents and reagent1 & 2 are exhausted.
Fill the empty reagent1 & 2 bottles with reagent1 & 2.
Standard empty. Performing single point or baseline calibration and the Standard is exhausted. Fill the empty Standard bottle with Standard.
Low standard empty. Performing two-point calibration and the Low Standard is exhausted.
Fill the empty Low-Standard bottle with Low-Standard.
Hi standard empty. Performing two-point calibration and the Hi Standard is exhausted.
Fill the empty Hi-Standard bottle with Hi-Standard.
Two standards are empty. Performing two-point calibration and both standards are exhausted.
Fill the empty Standard bottles with their respective Standards.
Sample Empty. Performing Analysis with single Sample and Sample Chamber is empty.
Check for Sample flow. Check for pipe blockages to Sample Chamber.
Testing optics, please check for optics failure. no light output from optics.
run a Clean Cycle and try again. inspect cell and clean, check for light detection under LED & DETECTOr COnTrOL in Diagnostics menu.
Testing optics, low signal level . . . Cell may be coated and may require cleaning.
Low light output from Optics. Optical window of cell may be coated.
run a Clean Cycle and try to clean the reaction cell with detergent.
Testing optics please check for stray light. External Light is entering the Cell. Check if overflow cap on top of the
reaction cell is loose.
Please check for heater failure.
if heater is enabled in Method, and the Set point temperature is higher than ambient and Cell temperature differs from set point by +/- 10 ºC for 90 secs.
Try a different set point temperature. Call a service agent.
Checking the reaction temperature. Temperature below set point.
if the temperature is not able to reach the set point and is below the set point.
re-calibrate the Thermistor. Try a different set point temperature. Call a service agent.
Checking the reaction temperature. Temperature above set point.
if the temperature is not able to reach the set point and is above the set point.
re-calibrate the Thermistor. Try a different set point temperature. Call a service agent.
Taking blank measurement, unstable signal.
Unstable temperature, Cell may be coated, or may be due to Optics failure.
Perform Clean Cycle and Calibrate Optics and thermistor.
Collecting data, unstable signal. Unstable temperature or Cell may be coated, or may be due to Optics failure.
Perform Clean Cycle and Calibrate Optics and thermistor.
67Thermo Scientific Orion 2230 Silica Analyzer User Guide
Calibrating Optics Messages
Messages Cause Solution
Unstable signal, stray light is high please check. Some external Light is entering in to the Cell. Check for possible external light
entering areas on the reaction cell.
Optimum light level. Calibration is aborted. Cell may be coated. Perform Clean Cycle
and then Calibrate Optics.
Calibrating Thermistor Messages
Messages Cause Solution
Temperature below set point please check for hardware malfunction.
May be due to the connectors or the thermistor or heater malfunctioning.
Check for thermistor and heater connections, and calibrate them.
Temperature above set point please check for hardware malfunction.
May be due to the connectors or the thermistor or heater malfunctioning.
Check for thermistor and heater connections, and calibrate them.
Measured temperature differs widely from set temperature. Please check for hardware malfunction.
May be due to the connectors or the thermistor or heater malfunctioning.
Check for thermistor and heater connections, and calibrate if necessary.
DESCR�PT�ON . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . CATALOG NUMBER
Maintenance and Service �temsMaintenance Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2230MK3ea Pump Head Tubing Covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2230PAPower Supply Assy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2230PSPower Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2230FSAir Purge Fittings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2230PIMain Tubing Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2230TKEmpty Auto Cal Btl (Validation Sample Bottle) . . . . . . . . . . . . . . . . . . . . . . .2230BCFittings, Sample In, Sample Out and Drain . . . . . . . . . . . . . . . . . . . . . . . . . 2230SFReagent Bottle Cap - reworked for luer fittings . . . . . . . . . . . . . . . . . . . . . .2230RCCal Bottle Cap - reworked for luer fittings . . . . . . . . . . . . . . . . . . . . . . . . . .2230SC Reagents and StandardsReagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2230RE 200 ppb Calibration Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22302020 ppb Calibration Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223002100 ppb Calibration Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2230101000 ppb Calibration Standard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223030Auto Standard Solution, 200 ppb, 0 .5 L . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223020Standard Solution, 20 ppb, 1L . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223002Standard Solution, 100 ppb, 1L . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223010Standard Solution, 1ppm ppb, 1L . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 223030
Parts and Accessories
68 Thermo Scientific Orion 2230 Silica Analyzer User Guide
Thisanalyzermayradiateradiofrequencyenergyandmaycauseinterferencetoradiocommunicationsifnotinstalledandusedproperlyinstrictaccordancewiththemanufacturer’sinstructions.ThisanalyzerhasbeentestedandfoundtocomplywiththelimitsforaClassAdigitaldevicepursuanttoSubpartBofPart15ofFCCRules,whicharedesignedtoprovidereasonableprotectionagainstsuchinterferenceinacommercialenvironment.Operationoftheanalyzerinaresidentialareamaycauseinterference,inwhichcasetheuser,athisownexpense,willberequiredtotakewhatevermeasuresmayberequiredtocorrecttheinterference.
ThisClassAdigitalapparatuscomplieswithCanadianICES-003.CetappareilnumériquedelaclasseAestconformeàlanormeNMB-003duCanada.
TheThermoScientificOrion2230SilicaanalyzerisCEandcTUVuscertified,whichincludestestingtoEMCandUSandCanadianstandards.
Notice of Compliance
69Thermo Scientific Orion 2230 Silica Analyzer User Guide
Manufacturer: ThermoFisherScientific,Inc.166CummingsCenterBeverly,MA01915USA
Themanufacturerherebydeclaresthatthe2230Silicaanalyzerconformswiththefollowingstandardsanddocuments:
Safety 2006/95/ECLowVoltageDirective EN/IEC61010-1:2001EMC 2004/108/ECElectromagneticCompatibility EmissionsandImmunity EN61326-1:2006
ThisThermoScientificOrionproducthasbeenmanufacturedincompliancewiththeprovisionsoftherelevantmanufacturingandtestdocumentsandprocesses.ThesedocumentsandprocessesarerecognizedascomplyingwithISO9001:2008byQMI-SAIGlobal,listedasFile#001911.
PatrickChiu December17,2010SeniorQualityEngineer, Beverly,MA RegulatoryCompliance
Declaration of Conformity
70 Thermo Scientific Orion 2230 Silica Analyzer User Guide
ThermoFisherScientificwarrantstotheoriginalpurchaserofanynewmerchandisethatallitemswillbefreeofdefectsinmaterialandworkmanshipfortheperiodssetforthbelow,whenusedunderspecifiedandnormaloperatingconditions,inaccordancewiththeoperatinglimitationandproceduresgivenintheinstructionmanuals,andwhennotsubjectedtoaccident,alteration,abuseormisuseofThermoFisherScientific’sproductsinunspecifiedapplications,forunauthorizedprocedures,orwiththird-partyproductsmayvoidthewarranty.ThermoScientific’sinstrumentarewarrantedasfollows:
Allpartsincontactwiththesampleforaperiodofninety(90)daysfromthedateofinstallation.Samplemustbechemicallycompatiblewithpartsorderedwiththeproduct.Ifpartsarenotcompatibleorifaquestionexistsregardingcompatibility,notifythefactorybeforeplacingtheproductinservicetomaintainthewarranty.
Partsnotincontactwithsampleforaperiodofone(1)yearfromthedateofshipment.
Peristalticpumptubingforaperiodofthirty(30)daysundernormaloperation.
Thefollowinguser-serviceablepartsarenotwarranted:Probes,O-rings.
THEWARRANTIESDESCRIBEDABOVEAREEXCLUSIVEANDINLIEUOFALLOTHERWARRANTIESWHETHERSTATUTORY,EXPRESSEDORIMPLIED.ALLOTHERWARRANTIESOFMERCHANTABILITYANDFITNESSFORAPARTICULARPURPOSE,ANDALLOTHERWARRANTIESARISINGFROMCOURSEOFDEALINGORUSAGEOFTRADE,EXCEPTTITLE,AREHEREBYOVERRIDDENANDEXCLUDED.NOLIABILITYSHALLATTACHTOTHERMOSCIENTIFICEITHERINCONTRACTORINTORT,FORANYPERSONALINJURY,DEATH,DAMAGETOPROPERTY,LOSSOFPROFITS,DAMAGES,COSTS,CHARGES,LIABILITIESOREXPENSES,WHETHERDIRECTORINDIRECT,CONSEQUENTIALOROTHERWISE,WHICHARISEOUTOFORINCONJUNCTIONWITHTHESALEORUSEOFTHISPRODUCT.
Thecustomer’ssoleandexclusiveremedyisthereturnofdefectivecomponentsorsubassembliestoThermoFisherScientificforrepairorreplacementor,atThermoFisherScientific’soption,refundofthepurchaseprice.
Forthemostcurrentwarrantyinformation,visitwww.thermoscientific.com/water.
Warranty �nformation
71Thermo Scientific Orion 2230 Silica Analyzer User Guide
Warranty Shipments/Returns/Adjustments
AwarrantyclaimmustbemadepromptlyandmustbereceivedduringtheapplicablewarrantyperiodbyThermoScientificoryourauthorizedThermoScientificdistributor.
Ifitbecomesnecessarytoreturnaproductforrepairand/oradjustment,priorauthorizationfromThermoScientificoryourThermoScientific-authorizeddistributormustbeobtained.InstructionsastohowandwheretheseproductsshouldbeshippedwillbeprovidedbyThermoScientificoryourThermoScientific-authorizeddistributor.
Anyproductorcomponentreturnedforexaminationand/orwarrantyrepairshallbesenttoThermoScientificinMA,oranyoftheirauthorizedrepresentatives.Allitemsmustbereturnedatthecustomer’scost(freightprepaid),quotingareturnauthorizationnumberthatisavailablefromtheServicedepartment.AllproductsorcomponentsrepairedorreplacedunderwarrantywillbereturnedtothecustomeratThermoScientific’scostusingUPS(UnitedParcelService)oranequivalentservice.
Inallcases,ThermoScientificoryourThermoScientific-authorizeddistributorhassoleresponsibilityfordeterminingthecauseandnatureoffailure,andThermoScientific’sorthedistributor’sdeterminationwithregardtheretoshallbefinal.
AllpartsthatarereplacedunderwarrantywillbecomethepropertyofThermoScientific.
ConsumablescanbeorderedfromThermoScientificoryourThermoScientific-authorizeddistributor.UseonlyThermoScientificproductsorThermoScientific-approvedproductsinyourThermoScientificAnalyzer.ThermoScientificshallnotbeliablefordamagetoormalfunctionofthesystem,whichitdeemswascausedbytheuseofunauthorizedmaterials.
Aftertroubleshootingallcomponentsofyourmeasurementsystem,contactTechnicalSupport.WithintheUnitedStatescall1.800.225.1480andoutsidetheUnitedStatescall978.232.6000orfax978.232.6015.InEurope,theMiddleEastandAfrica,contactyourlocalauthorizeddealer.Forthemostcurrentcontactinformation,visitwww.thermoscientific.com/water.
ForthelatestapplicationandtechnicalresourcesforThermoScientificOrionproducts,visitwww.thermoscientific.com/water.
Consumables
Assistance
266076-001 A 0810
Process Water Instruments
North America 166 Cummings Center Beverly, MA 01915 USA Toll Free: 1-800-225-1480 Tel: 1-978-232-6000 [email protected]
NetherlandsTel: (31) 033-2463887 [email protected]
India Tel: (91) [email protected]
Japan Tel: (81) [email protected]
ChinaTel: (86) 21-68654588 [email protected]
Singapore Tel: (65) [email protected]
www.thermoscientific.com/water
© 2010 Thermo Fisher Scientific Inc. All rights reserved.