thermetric_resistance_temperature_detector_rtd.pdf
TRANSCRIPT
-
7/27/2019 thermetric_resistance_temperature_detector_rtd.pdf
1/4
RESISTANCE TEMPERATURE DETECTORS (RTDs)
THERMETRICs Industrial RTDs are designed for outdoor heavy duty applications, with the exception of Typ
566, they are intended to be used without a thermowell. A cast iron screw cover head offers protection in mosrugged environments. Aluminum heads are also available (list part number 9313-2 after part number of basi
unit for this option. 1/4 diameter RTDs are standard). 100 OHM platinum sensors are used, and they followthe international Resistance vs Temperature Curve IEC 751 (Replaced DIN 43760), which has a coefficient oresistance (Alpha) of 0.00385 OHMS/OHM/C. We manufacture other types of RTDs but recommend thi
standard for most applications.
Type 561
STANDARD NIPPLE
Element is brazed to end of nipple
Nipples are 1/2 NPT Sch. 40 steel
Nipple can be supplied in extendedlengths for hard to get at location.
Note: To determine RTD length for Types 561 through 565 add
3 to length from top of NPT thread engagement to tip
of probe to allow for head assembly. See standard lengths
shown in How To Order.
Type 562
HEX NIPPLE Hex nipple for easy installation 1/2 NPT process connection
Type 563
STANDARD INDUSTRIAL RTD probe is silver brazed to brass plug
for support Optional mounting adaptors available.
-
7/27/2019 thermetric_resistance_temperature_detector_rtd.pdf
2/4
Temperature Sensing Devices & Assemblies For Industrial Applications
Type 564
SPRING LOADED ASSEMBLY
Recommended when vibration is present Ideal for measuring bearing temperature Can be shipped unassembled for simple
field setting of insertion length 1 spring travel
1/2" NPT process connection Carbon steel nipple standard. For
stainless steel add 9120 to part number.
Type 565SPRING LOADED WITH FLUID SEAL
Pressure rating 20 P.S.I. 1/2 NPT process connection
Type 566
EXTENDED HEAD ASSEMBLY Provides extended head assembly for
ease of installation 1/2" NPT brass compression fitting with
1/8 vent hole included
For stainless steel compression fittingadd 9121 to part number
36 of 22 gauge stranded silver platedcopper, Teflon insulated lead wire withPVC coated stainless steel flexible
conduit Can be used with thermowell Can be spring loaded, add 9211-S to part
number If assembly is required without head and
terminal block, specify type 567
-
7/27/2019 thermetric_resistance_temperature_detector_rtd.pdf
3/4
Industrial RTD AssembliesDAILY THERMETRICS Industrial RTD Assemblies consist of three basic components:
1. A connection head assembly
2. A thermowell or protection tube3. A RTD element
The lengths and other dimensions of all three components are standardized to provide complete interchangeability. Select bas
components from this page. Thermowells and protection tubes can be selected from the thermowell catalogue. Contact your local Da
Thermetrics representative to show you how to lower your total cost by using standard RTD assembly, element, and thermowell leng
and designs.
STANDARD:
Heavy duty cast iron head Nipples are Schedule 40galvanized steel
Aluminum head available Unions are 150# galvanized malleable
iron 3 cap diameter conduit connection
1/4" diameter probe Use with standard length and .260 bore
thermowells.
SPRING LOADED: (Not illustrated) Maintains positive spring contact to
thermowell
Recommended where vibration is
present 1/4 diameter probe Use with standard length and .260
bore thermowells
TYPE NUMBER
STANDARD SPRING LOADED
Cast Iron
Head
Aluminum
Head
Cast Iron
Head
Aluminum
Head
Component PartsLength from
Head toThermowell
Add to Well
Drill Depth toArrive at
Element Length
511 531 Head Only 0 3512 532 522 542 Head and Nipple 2 3
513 533 523 543 Head, Nipple, and Union 5 6
514 534 524 544 Head, Nipple, Union, Nipple 5 6NOTE: Dimensions are established to utilize standard RTD and thermowell lengths. RTD lengths are shown in How To Order. Standard thermowell drill depths are 4, 6, 9,12, 15, 18, and 24.
Deduct 1 if wells with 4 drill depth are used.
Contact Daily Thermetrics or your representatives for other models for special applications and designs.
-
7/27/2019 thermetric_resistance_temperature_detector_rtd.pdf
4/4
How to Order
Type No.
RESISTANCE AND ACCURACY
1 100 OHM .6 Standard Industrial
2 100 OHM .12 (DIN Class B) High Accuracy
3 100 OHM .06 (DIN Class A) Ultra high accuracy4 2 X 100 OHM .6 Dual element - Standard Accuracy5 2 X 100 OHM .12 Dual element - High Accuracy
6 2 X 100 OHM .06 Dual element - Ultra high accuracy0 Other resistance, accuracy or temperature coefficient
LEADWIRE CONFIGURATIONS
1 3 wire Standard
2 2 wire
3 4 wire High accuracy
4 6 wire Standard - Duplex
5 4 wire Duplex6 8 wire High accuracy
0
OUTSIDE DIAMETER
8 0.385
6 0.25 Standard5 0.1884 0.125 Fast Response
0 Other
STEM DESIGN
1 Standard
2 Tip Sensitive (3/16" & 1/4 O.D. only)3 Flexible Sheath
TEMPERATURE LIMIT
1 Standard temperature -58F (-50C) to +482F (250C)2 Mid temperature -58F (-50C) to +752F (400C)3 High temperature -58F (-50C) to +1112F (600C)
LENGTH CODE
1 6"
2 9"
3 12"4 15"
5 18"6 21"
7 24"8 27"
9 30"0 Other - Specify
Thermowell Part No. from Catalogue (if required)