Therac-25
CS4001
Kristin Marsicano
Therac-25 Overview What was the Therac-25? How did it relate to previous models? In what
ways was it similar/different? Was the Therac-25 reliable?
Therac-25 Overview Linear accelerator used to create high-energy
electron beams to treat shallow tumors and x-ray beams to reach deeper tumors
Differed from Therac-6 and Therac-20: computer was coupled with the system such that the
hardware could not function without the computer (e.g. turntable set up)
relied on the computer for safety checks; did not include the hardware safety features of previous models (which allowed for cost savings)
Similar to Therac-6 and Therac-20: Shared a common code base Used a computer to augment user
Was Therac-25 reliable?
Was Therac-25 reliable? Worked tens of thousands of times before
overdosing anyone Over course of 20 months (June 1985-July
1987) it administered massive overdoses to 6 patients, resulting in 3 deaths
Was notorious for displaying non-descript errors that had no negative side-effects (e.g. up to 40 times a day)
Do not confuse reliability with safety!
Under what conditions did the lethal doses occur?
Under what conditions did the lethal doses occur? Fast-typing operators
Race condition between magnet positioning and screen edits
Software relies on positioning of cursor to determine if edits have been made
Change from X-Ray mode to Electron mode made before magnets finish moving; software doesn’t check cursor position until after magnets have stopped
Set button Race condition between “gun ready” variable, gun
positioning, and “Set” button 0 means gun is ready and will fire; 1-255 means not ready;
increments as gun is moving and rolls over as necessary (which means it might be 0 when the gun is not really ready!
What parties were involved? Patients and their families AECL (maker of the machine) Developers Hospital where machine was used (and the
technicians)
AECL Mistakes Assumed error was only in software Did not design system to be fail-safe (fail-safe means no
single point of failure will lead to catastrophe); instead the Therac-25 relied 100% on the software to ensure safety of the system
Lack of software and hardware devices to detect and communicate an overdoes
Presumed correctness of reused code; assumed there were no errors in the previous code base when indeed there were
Management allowed the software to be developed without adequate documentation (e.g. no user manual for error codes)
Did not communicate fully with its customers with regards to the accidents