GESensing
Optica™
Optica is a General Eastern product. General Eastern has joinedother GE high-technology sensingbusinesses under a new name_GE Industrial, Sensing.
The Next Generation HumidityReference Standard
GE's Optica Series chilled mirror hygrometers offerNational Institute of Standards and Technologies (NIST)_traceable humidity, temperature and pressuremeasurements for the connected generation. Now datais accessible from anywhere, any time, from within abrowser over the Internet or from your intranet.
• Temperature
• Relative humidity (% RH)
• Dew/frost point (Td)
• Absolute humidity (m/v)
• Mass mixing ratio (m/m)
• Volume mixing ratio (v/v)
Optica Measures
• Wet bulb (Tw)
• Enthalpy (h)
• Water vapor pressure (e)
• Pressure
• Alarm relays
• Analog outputs
Fundamental dew point measurement is a primarymeasurement used as a transfer standard for calibratingother humidity instruments and sensors. Chilled mirrorsare also the sensors of choice when process andlaboratory measurements call for high precision withoutlong term drift. Optica may be used with five, fullyinterchangeable chilled mirror sensors to provide ameasurement range from -112°F to 185°F (-80°C to 85°C)dew point with 0.36°F (0.2°C) or better accuracy. Inputsfor a 100 S RTD and silicon-based piezoresistive pressuretransducer ensure precision measurements, which areused to convert the dew point to any metric, English oruser-defined unit of humidity measure.
g
Communications
• Ethernet port
• Java-based applet loads in web browsers
• Data logger 6 megabyte memory
• Recorded data uploads in ASCII format
• Exports to spreadsheets
• Real time "strip chart" graphing
• Color VGA or 4 x 40 matrix display
Software
The Optica software is easy to navigate. Users selectreadout display, strip chart, analog output scale, digitalcommunications, self-diagnostics, cleaning and datalogging with intuitive pull-down menus. Setup values canbe saved in memory and loaded anytime, either locally orfrom anywhere via the Ethernet port.
In the Lab or on the Plant Floor
• Calibration labs
• Process control
• Clean rooms
• Environmental test chambers
• Precision HVAC monitoring and control
• Fuel cells
• Heat exchanger and refrigerant coil calorimeters
• Thermal processing/heat treating
• Semiconductor manufacturing
• Storage areas
• Pharmaceutical validation chambers
• Engine test cells and emissions testing
• Aircraft engines and turbines
Functionality
The Optica simultaneously measures dew point,temperature and pressure. The analyzer is equipped withprogrammable math functions to produce custom units.In, exp +, -, x and / functions enable derived engineeringunits to be displayed, recorded and transmitted to dataacquisition systems. The analyzer input channel is astandard 4 to 20 mA/0 to 5 VDC input, so the Optica canbe connected to any type of process transmitter and beconfigured to display engineering units.
Chilled Mirror Sensors Theory ofOperation
GE’s chilled mirror hygrometers are used in standardsand metrology labs as well as in industrial applicationswhere precise and repeatable humidity measurementand control is required. The inherent accuracy and longterm stability provides many advantages over othertypes of humidity measurement technologies. Chilledmirrors fundamentally measure the dew or frost pointtemperature directly by controlling a reflective surface toan equilibrium temperature between dew/frostformation and evaporation, and precisely measuring thetemperature of the mirror at this point.
“Plug and play” setup is easily accomplished withstandard connectors for power, sensor and analog outputs.
GESensing
GE's chilled mirrors consist of a small polished hexagonalrhodium, or platinum mirror attached to thermoelectriccooling module (TEC). The Optica analyzer's servocontroller applies current to the TEC, which causes themirror to cool. The mirror is illuminated with a regulatedGas emitter, which transmits light in the infraredspectrum. The light reflected by the mirror is received bya photodetector. When water vapor condenses on themirror as water or frost (ice crystals), the light receivedby the photodetector is reduced due to scattering. Thisresults in the servo controller reducing the powercausing the mirror to slightly warm. The Optica's controlsystem will modulate the amount of current flowingthough the TEC to maintain a temperature where therate of condensation and evaporation of watermolecules and the mass of water on the mirror isconstant. The resulting temperature of the mirror is then,fundamentally by definition, equal to the dew or frostpoint temperature. A precision four-wire platinum RTDimbedded in the mirror measures the temperature. Theaccuracy of the dew point measurement has beenvalidated to an accuracy of ±0.36°F (±0.2°C) dew/frostpoint. The precision can also be enhanced to ±0.27°F(±0.15°C) dew/frost point.
Dry bulb temperature is measured with a precise four-wire 100 S platinum RTD and pressure is measuredwith a piezoresistive silicon pressure transducer. Thedew/frost point and dry bulb RTD resistance signals areconditioned and amplified by the Optica monitor todisplay and transmit dew/frost point and temperature.The pressure sensor transmits an amplified 4 to 20 mAsignal that is powered by the Optica. The cardinalmeasurements of dew/frost point, dry bulb temperatureand pressure are used to calculate other humidityparameters such as relative humidity, wet bulb, massmixing ratio, volumetric mixing ratio, absolute humidity,enthalpy and water vapor pressure values usingpsychometric equations.
The RTD sensor is imbedded in the chilled mirror andnever comes in contact with the process or testenvironment. The wetted parts consist of the platinum orrhodium mirror, a stainless steel or mylar vapor barrierand epoxy sealant. The net result is drift-free humiditymeasurement designed to maintain accuracyspecifications for many years.
Chilled mirrors require a nominal flow rate across themirror to achieve the optimal dew/frost formation andresponse time. The sensor should either be installed in aduct with moving air or be equipped with a samplingpump. The optimum flow rate is 0.5 to 5 SCFH (0.25 to 2.5 L/min). GE provides sampling systems to temperature
condition, regulate the pressure and filter process airprior to reaching the chilled mirror. Our applicationengineers will discuss your application in detail andprovide a recommendation for the system best suited toyour needs.
Self Cleaning and Digital Control
LED regulation Thermoelectricheat pump control
OPAMP
Light emitting diodesPhoto detectors
Thermoelectriccooling module
Rhodium orplatinum mirror
Precision four-wire PRTD
Power
Optical balance
Gas flow
For environments where physical contaminants such asdust, oil mist and pollen are present, the use of asampling system with a filter media is recommended.The filter media must be hydrophobic so as to notabsorb or release water vapor into the test stream. Overtime the mirror can be etched or pitted by particulate,altering the light scattering characteristics. GE mirrorsare field replaceable. The standard rhodium mirror maybe upgraded to platinum for industrial applications.
GE developed a patented contamination compensationscheme called PACER® (Program Automatic ErrorReduction). The PACER cycle may be initiated manually orby programming a timed cycle. The cycle starts bycapturing the data (during the PACER cycle a constantvalue is transmitted) and cooling the mirror well belowthe dew point such that a thick dew layer forms on themirror. The mirror is then rapidly heated. During theheating a significant amount of soluble and some non-soluble contamination is flash evaporated. Thecontamination left on the mirror tends to aggregate indry islands or spots (much like a glass that comes out ofa dishwasher). This process leaves approximately 85% ofthe mirror clean. The light signal received by thephotodetector is compared against a referenceLED/photodetector and the two signals are "balanced,"effectively negating the effect of the residualcontamination left on the mirror. The PACER cycle worksvery well, but eventually manual cleaning may berequired. All of GE's chilled mirrors are accessible for
GESensing
PACER cycle
Before PACER cycle After PACER cycle
Data is sampledand held (3 minutes)
Mirror heatsto dry state(90 seconds) Mirror returns
to dew point(60 seconds)
Mirror coolsand coalesces
(30 seconds)
Optical system automatically adjustsfor correct reflectance(5 seconds)
Prevailingdew point
Cooling
Heating
Endof cycle
Startof cycle
Typical PACER cycle
Slew rate
Dewpoint
Oscillation
Time
Tem
pera
ture
Analog control
Time
Tem
pera
ture
Digiloop control
Slew rate
Dewpoint
0.05°C temperature steps
manual cleaning. Cleaning is a simple process consistingof wiping the mirror with a cotton swab wetted withcleaning solution or distilled water. (Distilled water isrecommended as the final cleaning agent.)
GE developed Digiloop™ control to overcome thelimitations of traditional analog PID (Proportional Integraland Derivative) control. Analog temperature controlproduces oscillations particularly at high or tracehumidity levels. It is difficult to apply self-tuning or PIDconstants to analog control loops, which are overdamped or under damped. Digiloop utilizes digitalsampling and feed-forward control by taking time-basedsamples when the dew point is within a predeterminedproportional band. By recording the oscillations, thedigital control effectively predicts the magnitude ofchange and adjusts the current to the thermoelectriccooling module, which steps the mirror temperature in0.09°F (0.05°C) increments. This results in a significantlyimproved control and dew point precision.
GESensing
Models• Color VGA with data logger and Ethernet
communications available in bench, rack or wall mount styles
• 4 x 40 dot matrix available in bench, rack or wall mount styles
Power95 to 265 VAC, 50/60 Hz, 200 watts
Electrical I/OIEC-style AC receptacle (bench mount), screw terminalblock (wall mount), multipin chilled mirror andtemperature sensor/cable connectors, screw terminalsfor analog outputs, DB-9 for RS232 and 10 base-T forEthernet*
Measured ParametersDew point, temperature and 4 to 20 mA/0 to 5 VDC
Calculated ParametersRelative humidity, wet bulb, mass mixing ratio, volumetricmixing ratio, absolute humidity, water vapor pressureand enthalpy in English and metric units, as well as user-configured units
Inputs1/3 Class A DIN 43760, 100 S RTD and dew/frost pointand dry bulb temperature. Loop powered 4 to 20 mA DCat 500 S maximum load
AccuracySystem accuracy ±0.36°F (±0.2°C) for dew/frost point,±0.27°F (±0.15°C) for temperature, 0.5% full scale (FS) forpressure
RangeGoverned by sensor
HystersisNegligible
Sensitivity0.1% FS
A/D16 bit
OpticaSpecifications
Data Logger Memory6 megabytes*
Display1/4 color VGA enables up to six parameters to bedisplayed or 4 x 40 dot matrix enables three parametersto be displayed
Operating Temperature32°F to 122°F (0°C to 50°C)
Cooling Rate0.27°F (1.5°C )/sec typical above 32°F (0°C)
Digital InterfaceRS232 port, Ethernet port*
Digital Output FormatData ASCII text, Ethernet java applet password protected*
TCP/IP AddressUser programmable*
Analog Outputs(2) 4 to 20 mA and 0 to 5 VDC, user configurable andscalable
Analog Outputs(2) 5 A at 250 V, form C, (SPDT) relays
EnclosureBenchtop: Type 1Wall Mount: Type 4
WeightBenchtop: 8 lb (3.6 kg)Wall Mount: 10 lb (4.5 kg)
*Feature available with VGA models only
16.5 in (419 mm
)
13.5 in (343 mm) 13.5 in (343 mm)
6.5 in(165 m
m)
Bench Top Wall Mount
GESensing
Sensing ElementFour-wire 1/3 Class A DIN 43760 RPT, 100 S @ 32°F (0°C)
Dew/Frost Point AccuracyStandard: ±0.36°F (±0.2°C)Optional: ±0.27°F (±0.15°C)
Sensitivity>0.05°F (>0.03°C)
Repeatability±0.09°F (±0.05°C)
HysteresisNegligible
Cooling StagesSingle stage thermoelectric cooling (TEC) module
Auxiliary CoolingNot applicable
Depression81°F (45°C) at 77°F (25°C) dry bulb and atmospheric pressure
Typical Measurement Range5°F to 77°F (-15°C to 25°C) dew/frost point (Td) in air @77°F (25°C) and atmospheric pressure. Equivalent to 5%to 100% RH.
Other humidity parameters based on calculations.
Sample Flow0.5 to 5.0 SCFH (0.25 to 2.5 L/min)
Operating Temperature5°F to 176°F (-15°C to 80°C)
Pressure-3 to 200 psig (0.8 to 15 bar)
1111H Single-StageChilled MirrorSpecifications
PowerDerived from Optica
Sensor BodyEpoxy-coated aluminum
FilterPolytetrafluoroethylene (PTFE) filter (standard on 1111H-GE)
MirrorStandard: Rhodium-plated copperOptional: Solid platinum
Sensor Wetted MaterialAluminum, copper, mylar, PTFE, rhodium or platinum
Vapor BarrierMylar
Electrical ConnectorMS-style multipin connector
Weight1 lb (1.4 kg) net
Accessories
MB-11 Wall mounting bracket
PTFE-GE PTFE filter
P Platinum mirror
X Enhanced accuracy ±0.27°F (±0.15°C) Td
O111D Pressure boss (1111H only)
1.23 in(31 mm)
.62 in(16 mm)
2.57 in (65 mm)
3.5 in(89 mm)
Ø 1.2 in
(30 mm
)
1 in NPT
GESensing
Sensing ElementFour-wire 1/3 Class A DIN 43760 RPT, 100 S @ 32°F (0°C)
Dew/Frost Point AccuracyStandard: ±0.36°F (±0.2°C)Optional: ±0.27°F (±0.15°C)
Sensitivity>0.05°F (>0.03°C)
Repeatability±0.09°F (±0.05°C)
HysteresisNegligible
Cooling StagesTwo-stage TEC module
Auxiliary CoolingNot applicable
Depression117°F (65°C) at 77°F (25°C) and atmospheric pressure
Typical Measurement Range31°F to 77°F (-35°C to 25°C) dew/frost point (Td) in air @77°F (25°C) and atmospheric pressure. Equivalent to 0.7%to 100% RH.
Other parameters based on calculations.
Sample Flow0.5 to 5.0 SCFH (0.25 to 2.5 L/min)
Operating Temperature-13°F to 185°F (-25°C to 85°C)
Pressure150 psig (11 bar) maximum
PowerDerived from Optica monitor
D2 Two-StageChilled MirrorSpecifications
Sensor BodyCast aluminum with 314 stainless steel flow cell
MirrorStandard: Rhodium-plated copperOptional: Solid platinum
Sensor Wetted Material302, 316 stainless steel, silicone o-ring, BK-7 glass, rhodium or platinum mirror
Vapor BarrierStainless steel
Inlet/Outlet1/4 in OD tubing compression fittings
Electrical ConnectorSub-D 15-pin connector mates with 2130 cable
Weight4 lb (1.8 kg) net
Accessories
P Platinum mirror
X Enhanced accuracy ±0.27°F (±0.15°C) Td
4.5 in(114 mm)
square
Ø .107 in (4.75 mm) through 4 places
1/4 incompression
stainless steel
4 in(102 mm)
GESensing
Sensing ElementFour-wire 1/3 Class A DIN 43760 RPT, 100 S @ 32°F (0°C)
Dew/Frost Point AccuracyStandard: ±0.36°F (±0.2°C)Optional: ±0.27°F (±0.15°C)
Sensitivity>0.05°F (>0.03°C)
Repeatability±0.09°F (±0.05°C)
HysteresisNegligible
Cooling StagesTwo-stage TEC module
Depression117°F (65°C) at 77°F (25°C) and atmospheric pressure
Typical Measurement Range31°F to 77°F (-35°C to 25°C) dew/frost point (Td) in air @77°F (25°C) and atmospheric pressure. Equivalent to 0.7%to 100% RH.
Other parameters based on calculations.
Sample Flow0.5 to 5.0 SCFH (0.25 to 2.5 L/min)
Operating Temperature5°F to 212°F (-15°C to 100°C)
Pressure300 psig (21 bar) maximum
PowerDerived from Optica
1211H Two-StageChilled MirrorSpecifications
GE InfrastructureSensing
1211H
Sensor BodyCast aluminum with 314 stainless steel flow cell
MirrorStandard: Rhodium-plated copperOptional: Solid platinum
Sensor Wetted Material302, 316 stainless steel, silicone o-ring, BK-7 glass,rhodium or platinum mirror
Vapor BarrierMylar (upgradeable to stainless steel)
Inlet/Outlet1/4 in OD tubing compression fittings
Electrical ConnectorMS-style multipin connector mates with 2120 cable
Weight4 lb (1.8 kg) net
Accessories
P Platinum mirror
X Enhanced accuracy ±0.27°F (±0.15°C) Td
S Stainless steel vapor barrier
5 in(127 mm)
3.6 in(91.44 m
m)5.7 in
(144.78 mm
)
3.1 in(78.74 mm)
1/4 in(6.35 mm)
compressionfitting
GESensing
Sensing ElementFour-wire 1/3 Class A DIN 43760 RPT, 100 S @ 32°F (0°C)
Dew/Frost Point AccuracyStandard: ±0.36°F (±0.2°C)Optional: ±0.27°F (±0.15°C)
Sensitivity>0.05°F (>0.03°C)
Repeatability±0.09°F (±0.05°C)
HysteresisNegligible
Cooling StagesTwo-stage TEC module
Auxiliary CoolingNot applicable
Depression153°F (85°C) at 167°F (75°C) body temperature andatmospheric pressure
Typical Measurement Range14°F to 167°F (-10°C to 75°C) dew/frost point (Td) in air @167°F (75°C) body temperature in 77°F (25°C) ambienttemperature and atmospheric pressure. Equivalent to 0.7% to 100% RH.
Other parameters based on calculations.
Sample Flow0.5 to 5.0 SCFH (0.25 to 2.5 L/min)
Operating Temperature5°F to 212°F (-15°C to 100°C)
SIM-12 HeatedTwo-StageChilled MirrorSpecifications
25
25 55 70 85
185160130105
77 25 25°C °F
Max DewpointOutlet
Ready ON
Inlet
°C
°F
Model SIM-12HHeated Sensor
Heater ControlThermostatically controlled. Set point of 77°F (25°C),104°F (40°C), 131°F (55°C), 158°F (70°C), 185°F (85°C) and212°F (100°C)
Pressure50 psig (4.5 bar) maximum100/115/230 VAC, 50/60 Hz, 75 watts
Sensor BodyCast aluminum with 314 stainless steel flow cell
Sensor Wetted Material302, 316 stainless steel, silicone o-ring, BK-7 glass, rhodium or platinum mirror
MirrorStandard: Rhodium plated copperOptional: Solid platinum
Vapor BarrierMylar (upgradeable to stainless steel)
Electrical ConnectorMS-style multipin connectorIEC receptacle for AC power
Weight7 lb (3.2 kg) net
Accessories
P Platinum mirror
X Enhanced accuracy ±0.27°F (±0.15°C) Td
S Stainless steel vapor barrier
HSS-12 Heated sampling system. SIM-12 heated chilled mirror, SIM-HFT heated filter module and SIM-HFM heated flowmeter mounted on a plate with SIM-HSL heated sample line.
4 in(102 m
m)
8.64 in (219 mm)3.5 in(89 mm)
.25 in compression fitting
GESensing
Sensing ElementFour-wire 1/3 Class A DIN 43760 RPT, 100 S @ 32°F (0°C)
Dew/Frost Point AccuracyStandard: ±0.36°F (±0.2°C)Optional: ±0.27°F (±0.15°C)
Sensitivity>0.05°F (>0.03°C)
Repeatability±0.09°F (±0.05°C)
HysteresisNegligible
Cooling StagesFour-stage TEC module
Auxiliary CoolingLiquid cooling jacket
Depression, Air Cooled171°F (95°C) at 77°F (25°C) and atmospheric pressure
Depression, Liquid Cooled189°F (105°C) with 59°F (15°C) cooling water
Typical Measurement Range• -85°F to 77°F (-65°C to 25°C) dew/frost point (Td) in air
@ 77°F (25°C) and atmospheric pressure.• -103°F to 59°F (-75°C to 15°C) in liquid-cooled mode
with 59°F (15°C) coolant.
Other parameters based on calculations.
Sample Flow0.5 to 5.0 SCFH (0.25 to 2.5 L/min)
Operating Temperature32°F to 95°F (0°C to 35°C)
1311-DRFour-StageChilled MirrorSpecifications Control
Air Cooling
Liquid Cooling
~ Input Power
1311DR
Pressure300 psig (22 bar) maximum
Power100/115/230 VAC, 50/60 Hz, 300 watts
Sensor BodyCast aluminum with 314 stainless steel flow cell
MirrorStandard: Rhodium-plated copperOptional: Solid platinum
Sensor Wetted Material302, 316 stainless steel, silicone o-ring, BK-7 glass, rhodium or platinum mirror
Vapor BarrierMylar (upgradeable to stainless steel)
Inlet/Outlet1/4 in (6 mm) OD tubing compression fittings
Electrical ConnectorMS-style multipin connector matesIEC receptacle for AC power
Weight34 lb (16 kg) net
Accessories
P Platinum mirror
X Enhanced accuracy ±0.27°F (±0.15°C) Td
S Stainless steel vapor barrier
8.77 in(223 mm)
1.8 in(46 mm)
7.17 in(182 mm)Ø 0.31 in
(7.9 mm)4 places
16.3 in(414 mm)
GESensing
Sensing ElementFour-wire 1/3 Class A DIN 43760 RPT, 100 S @ 32°F (0°C)
Dew/Frost Point AccuracyStandard: ±0.27°F (±0.15°C)
Sensitivity>0.05°F (>0.03°C)
Repeatability±0.09°F (±0.05°C)
HysteresisNegligible
Cooling StagesFive-stage TEC module
Auxiliary CoolingLiquid cooling jacket
Depression, Liquid Cooled202°F (112°C) with 59°F (15°C) cooling water
Typical Measurement Range-112°F to 59°F (-80°C to 15°C) dew/frost point (Td) in air@ 77°F (25°C) and atmospheric pressure with 59°F (15°C)cooling water.
Other parameters based on calculations.
Sample Flow0.5 to 5.0 SCFH (0.25 to 2.5 L/min)
Operating Temperature32°F to 95°F (0°C to 35°C)
Pressure0 to 100 psig (1 to 8 bar)
1311-XRFive-StageChilled MirrorSpecifications
g
g 1311XR
1311XR
Power115/230 VAC, 50/60 Hz, 700 watts
Sensor Body314 stainless steel
MirrorStandard: Rhodium-plated copperOptional: Solid platinum
Sensor Wetted Material302, 316 stainless steel, silicone o-ring, BK-7 glass, rhodium or platinum mirror
Vapor BarrierMylar (upgradeable to stainless steel)
Inlet/Outlet1/4 in OD tubing compression fittings
Electrical ConnectorMS-style multipin connector IEC receptacle for AC power
Weight50 lb (26 kg) net
Accessories
P Platinum mirror
S Stainless steel vapor barrier
6.95 in(176.5 m
m)
5.2 in(132 m
m)
19 in(483 mm)
5.3 in(135 mm)
9 in(129 mm)
GESensing
Sensing ElementFour-wire 1/3 Class A DIN 43760 RPT, 100 S @32°F ( 0°C)
AccuracySystem at 25°CStandard: ±0.27°F (±0.15°C)Optional: ±0.18°F (±0.1°C)
Measurement Range-148°F to 212°F (-100°C to 100°C)
Response Time7 seconds for 77°F to 158°F (25°C to 70°C) step change influid
Sensor BodyStainless steel sheath
T-100 Four-Wire PRTDTemperatureMirrorSpecifications
CablePTFE insulation rated to 302°F ( 150°C). 10 ft (3 m) standard length.
FittingsAdjustable 1/4 in NPT stainless steel compression fitting
Strain ReliefStainless steel spring
PowerLow voltage derived from Optica
Weight2 lb (0.9 kg) net
OptionsAdditional cable length
7.5 in (190.5 mm)
Ø .25 in(6.35 mm)1/4 in NPT
Connector
PTFE cable, 305°F (150°C), 10 ft (3 m) standard length
GESensing
Sensing ElementMicro-machined silicon strain gauge
AccuracySystem at 77°F (25°C), ±0.5% of full scale
Range PT-30A0 to 30 psia (0 to 2 bar)
Range PT-300A0 to 300 psia (0 to 21 bar)
Response Time1 second for 90% of steady state (10% to 90% change)
Operating Ranges• -4°F to 176°F (-20°C to 80°C) ambient temperature• -13°F to 248°F (-25°C to 120°C) process temperature
Temperature Effect<1% FS on accuracy from 14°F to 122°F (-10°C to 50°C)for pressure 6 psi (0.4 bar)
PT SeriesPressureTransducerSpecifications
Power9 to 30 VDC. Derived from Optica monitor.
Sensor Body Material316 stainless steel
Sensor Wetted Material316 stainless steel and Hastelloy diaphragm
CablePVC insulation. 10 ft (3 m) standard length process connection 1/4 in F-NPT.
Weight4 lb (1.8 kg) net
5.03 in(1.28 mm)
1.5 in(38 mm)
10 ft (3 m)standard
length
GESensing
EnclosurePolycarbonate
FittingsBrass 1/4 in OD tubing
Compression Internal TubingImpolene
Rotameter0.5 to 5 SCFH (0.2 to 2 L/min)
Maximum Vacuum-98 in (-250 mm) of Hg
SSM SamplingSystemSpecifications
Free Air Displacement14 SCFH (6.5 L/min)
Maximum Temperature100°F (38°C)
Power95 to 265 VAC, 50/60 Hz, 15 watts
3.30 in(84 mm)
6.70 in (170 mm
)
Powerinputmodule
SSM Sampling System
In Out
2.68 in(68 mm)
Output gasfittings
Input gas fittings
GESensing
Appendix1111H Single-Stage Chilled MirrorSensor Body Temperature °C
100
75
50
25
0
-25
-50
-75
-100-50 -25 0 25 50 75 100
Mirrordepression
Sensor Body Temperature °C
Dew
Fro
st P
oint
Tem
pera
ture
°F (°
C)
45°C depression
D2 Two-Stage Chilled MirrorSensor Body Temperature °C
100
75
50
25
0
-25
-50
-75
-100-50 -25 0 25 50 75 100
Mirrordepression
Sensor Body Temperature °C
Dew
Fro
st P
oint
Tem
pera
ture
°F (°
C)
65°C depression
1211H Two-Stage Chilled MirrorSensor Body Temperature °C
100
75
50
25
0
-25
-50
-75
-100-50 -25 0 25 50 75 100
Mirrordepression
Sensor Body Temperature °C
Dew
Fro
st P
oint
Tem
pera
ture
°F (°
C)
65°C depression
SIM-12 Heated Two-Stage Chilled MirrorSensor Body Temperature °C
100
75
50
25
0
-25
-50
-75
-100-50 -25 0 25 50 75 100
Mirrordepression
Sensor Body Temperature °C
Dew
Fro
st P
oint
Tem
pera
ture
°F (°
C)
65°C depression
GESensing
Appendix1311-DR Four-Stage Chilled MirrorSensor Body Temperature °C
50
25
0
-25
-50
-75
-100-50 -25 0 25 50 75 100
Mirrordepression
Ambient Air Temperature °C
Dew
Fro
st P
oint
Tem
pera
ture
°F (°
C)
95°C depression
1311-XR Five-Stage Chilled MirrorSensor Body Temperature °C
100
75
50
25
0
-25
-50
-75
-100-50 -25 0 25 50 75 100
Mirrordepression
Sensor Body Temperature °C
Dew
Fro
stPo
int T
empe
ratu
re °F
(°C)
112°C depression
GESensing
Recommended for dew points ≥32°F (≥0°C)
MaterialsAnodized aluminum head, nylon bowl, nylon internals,Buna-N Seals
Fittings1/4 in compression
Temperature-150°F to 220°F (-101°C to 104°C)
FilterBorosilicate glass
Porosity93% of particles over 0.1 µm
Pressure150 psig (10 bar) maximum
BF-12SSBox of 10 replacement filter elements
AccessoriesBF-10DX General Purpose Filter
Recommended for dew points ≤32°F (≤0°C)
Materials316 stainless steel head, bowl and internals. Viton seals
Fittings1/4 in (6 mm) compression
Temperature-150°F to 220°F (-101°C to 104°C)
FilterBorosilicate glass
Porosity93% of particles over 0.1 µm
Pressure340 bar (5000 psig) maximum
BF-12SSBox of ten replacement filter elements
BF-12SS Stainless Steel Filter
MaterialsClear acrylic body, Buna-N Seals, black glass float, brassvalves
Fittings1/4 in compression
Range0.5 to 5 SCFH (0.2 to 2 L/min)
Pressure100 psig (6 bar) maximum
Temperature150°F (65°C) maximum
FM-1 Rotameter/Needle Valve
5.4 in(136 mm)
.6 in(14 mm)
2.7 in(70 mm)
1/4 incompression
outlet andinlet ports
1.75 in (44 mm)
3.53 in(90 mm)
1.19 in (30 mm)
GESensing
Optica Monitor
Color VGA Display With Logger/EthernetA BenchtopB Benchtop with rack mount bracketC Wall mount
Four-color, 40 Alphanumeric DisplayD BenchtopE Benchtop with rack mount bracketF Wall mount
Optica Channel One1 Unit to be sold with new chilled mirror
(see sensor section)2 Use the existing D2
(conversion cable 2160 required)3 Use with existing 1111H, 1211H, SIM-12H and
1311DR (conversion cable 2160 required)4 Use with existing 1311XR
(conversion cable 2140 required)
Default Code0 Default code
Ordering CodeGuide
Chiller Mirror Cable
Cable Type0 No cable requiredC Adaptor cable for existing sensor/cable
installation 2 ft (.61 m)A 10 ft (3 m) standard length X Specify length in ft (m)
O P A 1 O A C 1 A O
2120 For use with 1111H, 1211H, SIM12, HSS-12, 1311DR sensors
2130 For use with D22140 For use with 1311XR2150 For use with Optica CCD sensor2160 Adaptor cable for sensors except 1311XR
2 ft (0.61 mm)2170 Adaptor cable for 1311XR 2 ft (0.61 mm)
Chilled Mirror Sensor
Sensor0 No sensorA 1111HB 1111H-GEC D2D 1211H
SIM-12E 100 VACF 115 VACG 230 VAC
HSS-12H 100 VACJ 115 VACK 230 VAC
1311DRL 100 VACM 115 VACN 230 VAC
1311XRP 100 VACQ 115 VACR 230 VAC
GESensing
Accuracy And Mirror
1 S/R/M standard accuracy, rhodium mirror, mylar vapor barrier
2 S/P/M standard accuracy, platinum mirror, mylar vapor barrier
3 S/P/P standard accuracy, platinum mirror, stainless steel vapor barrier
4 X/R/M enhanced accuracy, rhodium mirror, mylar vapor barrier
5 X/P/M enhanced accuracy, platinum mirror, mylar vapor barrier
6 X/P/P enhanced accuracy, platinum mirror, stainless steel vapor barrier
7 S/R/P standard accuracy, rhodium mirror, stainless steel vapor barrier (D2)
8 X/R/P enhanced accuracy, rhodium mirror, stainless steel vapor barrier (D2)
9 S/P/P standard accuracy, platinum mirror,stainless steel vapor barrier (D2)
Ordering CodeGuide
Temperature Sensor
Sensor0 No temperature sensorA T-100 with 10 ft (3 m) cable (standard accuracy)B T-100 with 10 ft (3 m) cable (enhanced accuracy)X T-100 with build to fit cable (standard accuracy)Z T-100 with 10 ft (3 m) cable (enhanced accuracy)
Pressure Sensors0 No pressure sensorA PT-30A 10 ft (3 m) cable (standard)B PT-30A build-to-fit cableX PT-300A 10 ft (3 m) cable (standard)Z PT-300A build-to-fit cable
GESensing
©2005 GE. All rights reserved.920-076C
All specifications are subject to change for product improvement without notice.Optica™ is a trademark of GE. GE® is a registered trademark of General Electric Co.Other company or product names mentioned in this document may be trademarks orregistered trademarks of their respective companies, which are not affiliated with GE.g www.gesensing.com