Gabriele Hessling Department of Electrophysiology German Heart Center, Technical University Munich, Germany Pedirhythm VII, Thessaloniki, 6th February 2017 The new leadless pacemakers- when will they be feasible in children ?

Upload: others

Post on 10-Feb-2021

4 views

Category:

Documents


0 download

TRANSCRIPT

  • GabrieleHesslingDepartmentofElectrophysiology

    GermanHeartCenter,TechnicalUniversityMunich,Germany

    PedirhythmVII,Thessaloniki,6thFebruary2017

    Thenewleadlesspacemakers-whenwilltheybefeasibleinchildren?

  • PacemakerTherapy

    1956 2013

    Eachyearnearly1millionpersonsworldwidereceivetransvenouscardiacpacemakers.

    SNllnopacemakersorleadsaredesignedspecificallyforchildren

  • WhyLeadlesspacing?

    Lead-associatedcomplica8ons-Pneumothorax-CardiacperforaNon-Dislodgement-Venousocclusion-Fracture,insulaNonfailure

    Pocket/Generator-relatedcomplica8ons

    -Hematoma-SkinErosion-InfecNon-CosmeNcconcerns

    Pacemakerrelatedadverseeventsin~1of10paNents

  • PotenNalbenefitsofleadlesspacing

    1

    2

    3

    Percutaneous,lesshardware,nocosmeNcissues

    ReducNonofacuteandchroniccomplicaNons

    ReducNonofcomplicaNonsShortin-hospitalstay

    Onlysinglechamberpacing(VVI/R)possible

  • MMicra™

    St.JudeMediaNanos8m™

    Size

    Longevity

    Accesssite

    FixaNon

    RetrievalopNon

    41,4mm,ø5.9mm 25.9mm,ø6.7mm

    >9,3years 10years

    V.femoralis(18Fr.) V.femoralis(23Fr.)

    Screw-inHelix(1,3mm) 4self-expand.Ni8nolTines

    Yes Yes

    CEMark/FDAApproval

    Oktober2013/No April2015/April2016

    2Systems

    MRIcompaNble 1.5and3Tesla1.5Tesla

    Source:MedtronicInc,St.JudeInc

    St.JudeMedical Medtronic

  • 2Systems

    Nanos8m™Micra™

    Source:,St.JudeMedical,Medtronic

  • St.JudeMediaNanos8m™

    41,4mm,ø5.9mm

    >9,3years

    V.femoralis(18Fr.)

    Screw-inHelix(1,3mm)

    Yes

    Oktober2013

    1.5Tesla

    St.JudeMedical

    28thOctober2016St.JudeMedicalhaspressedpauseonallimplantsofitsNanos8mleadlesscardiacpacemakersduetoaba_eryproblemthathasresultedinlossofpacingandtelemetryinafewdevices.Theissuehasbeenobservedinsevendevices(29-37monthsaaerimplant)outofapproximately1400implantsaroundtheworld—a0.5%rate.

  • Leadlesspacing-ImplantaNon

    Catheterbaseddeliveryinthelowerrightventricularseptum

  • Removalseemspossible

    RetrievaloftheLeadlessCardiacPacemakerAMul8centerExperienceCircArrhythmElectrophysiol2016,December;9(12)VivekY.Reddy,MarcA.Miller,ReinoudE.Knops,PetrNeuzil,PascalDefaye,WernerJung,RahulDoshi,MarkCastellani,AdamStrickberger,R.HardwinMead,HarishDoppalapudi,DhanunjayaLakkireddy,Ma_hewBenne_,JohannesSperzel• Overallleadlesspacemakerretrievalsuccessratewas94%:• Leadlesscardiacpacemakerimplanted6months• Noprocedure-relatedadverseeventsat30dayspostretrievalprocedure.

  • IndicaNons

    Pa8entswithIndica8onforVVI(R)Pacing

    • ChronicatrialfibrillaNonwith2or3°AVBlock

    • Sinusrhythmwith2or3°AVorBBBblock,lowlevelofphysicalacNvityorpaNentswithalifespan<10years

    • SinusBradycardiawithinfrequentpausesorunexplainedsyncope

    Poten8alindica8ons?PhysicallyveryacNvepaNents(avoidpocket)?Venousaccessproblems?S/ppocketinfecNon?PaNentsatincreasedriskforleadfailureorinfecNon?Neurocardiogenicsyncopecardioinhibitorytype

  • Data

    ReddyetalNEJM2015;373:1125

    Nanostim n= 526 Micra n= 725

    Reynoldsetal,NEJM2016;374:533

  • Efficacy

    NanosNm Micra

    Efficacyendpointreachedin>90%

  • Safety-NanosNm

    At6months:Devicerelatedseriouseventsin6.7%;-CardiacperforaNonin1.3%-  Devicedislodgementin1.7%-  ThresholdelevaNonin1.3%-  VascularcomplicaNons1.3%

  • Safety-Micra

    At6monthsAtotalof28eventsin25paNents(4%)-Cardiacinjuryin1.6%

  • Leadlesspacinginchildren

    •  SafetyandfeasibilityofusingthisleadlesspacemakerinpaNentsyoungerthan18yearsofageunknown

    •  Sizeoftheintroducersheath(18French/23French)maymakeitsuseinchildrenmoredifficult(complicaNonsrelatedtoeitherthefemoralaccesssiteorcathetermanipulaNonwithintherightventricle)

    •  DevicesplacedinthesmallerrightventriclesofchildrenðTV-problems,proarrhythmia?

    •  Furtherminiaturiza8onrequired;shorterba_erylife?•  Extractability(firstdatainhumansaboutremovalof

    chronicallyimplantedsystems;riskoffibrosishigherinchildren?)

  • LeadlesspacinginadultswithCHD

    •  SafetyandfeasibilityofusingthisleadlesspacemakerinadultCHDpaNentsunknown

    •  IndicaNonsbasedonadultpopulaNonindicaNons;singlechamberpacing

    UnsolvedIssues

    •  Morphologicallyleaventricles(S/patrialswitch)withouttrabeculaNon?

    •  Retrogradeplacementina(single)ventricle?

    •  RiskofThrombosis-anNcoagulaNonmanagement?

  • Summary

    •  LeadlesspacingseemsanexciNngnewdevelopment

    •  Electricalperformancecomparablewithtransvenouspacemakers

    •  AcutecomplicaNonssuchastamponadeorperforaNonneedtobeaddressedðsafe(r)implantaNontechniques

    •  Long-termissues(thrombogenity,proarrhyhtmia,extractability)needtobeaddressed/solved

    Systemsapplicableforchildrencurrentlylacking

  • Summary

    •  LeadlesspacingseemsanexciNngnewdevelopment

    •  Electricalperformancecomparablewithtransvenouspacemakers

    •  AcutecomplicaNonssuchastamponadeorperforaNonneedtobeaddressedðsafe(r)implantaNontechniques

    •  Long-termissues(thrombogenity,proarryhtmia,extractability)needtobeaddresed/solved

    Systemsapplicableforchildrencurrentlylacking

    What´snext?

    Smallerdevices?Leadlessatrialdevices?

    CommunicaNonwithventriculardevice?LeadlesspacerandsubcutaneousICD?

    IntravascularICDs?

  • Thankyouforyoura_enNonandthankyouChristofKolb!