the evaluation of pre-cut corneas prepared in the international eye bank of prague
DESCRIPTION
The evaluation of Pre-Cut corneas prepared in The International Eye Bank of Prague. Magdaléna Netuková, Yveta Urbanová, Josef Šach, Pavel Kuchynka Ocular Tissue Bank OTB01/The International Eye Bank of Prague, Czech Republic - PowerPoint PPT PresentationTRANSCRIPT
The evaluation of Pre-Cut corneas prepared in The International Eye
Bank of Prague
Magdaléna Netuková, Yveta Urbanová, Josef Šach, Pavel Kuchynka
Ocular Tissue Bank OTB01/The International Eye Bank of Prague, Czech RepublicDpt. of Ophthalmology Vinohrady Teaching Hospital Prague, Czech Republic
All authors have No finacial interest
Aim of the study
• To evaluate quality of Pre-Cut corneas for DSAEK prepared in the Interantional Eye Bank of Prague
• To analyse possible complications during preparation of the Pre-Cut corneas.
• To analyse possible complications after preparation of Pre-Cut corneas
Methods - preparation
• Pre-Cut corneas were prepared from 30 corneas – not suitable for surgery
• All Pre-Cut corneas were preapred in „Clean rooms“ by certified staff members
• All Pre-Cut corneas were prepared using standardized method for DSAEK – with Moria microkeratome
Ljfpqejqoj tpojJkpavjpjp ojp
Moria microkeratome
No major technical problems during preparation
Mounting
Epithelial abrasion
Pachymetry
Free flap reposition
Free flap evaluatin
Cut
Dismounting
Methods - Evaluation
• All Pre-Cut corneas were evaluated using– Slit lamp– Specular microscopy
• Selected Pre-Cut corneas were screend for – Microbial comtamination
Ljfpqejqoj tpojJkpavjpjp ojp
Results - Slit lamp evaluation
• All 30 Pre-Cut corneas were evaluated using slit lamp• In 28 Pre-Cut corneas we found good position of the corneal
free flap• In 2 cases we foud dislocation of the free flap into the
medium• No obvious stromal or endotehelial defects or opacity were
detected
Ljfpqejqoj tpojJkpavjpjp ojp
Incorrect position of
the free flap
Correct position of
the free flap
Results Specular microscopy evaluation
• The average endothelial cell censity prior PreCut was 2709±483(SD) cells/mm2,
• The average endothelial cell censity after PreCut was 2623±407(SD) cells/mm2
• We found no significant endothelial cell loss after preparation.
Corneal endothelium after Pre-Cut
Endothelial Cell Density
Ljfpqejqoj tpojJkpavjpjp ojp
Resuts - Microbiology
• 5 randomly selected samples of medium and corneal swaps were microbiologically tested for microbial and fungal contamination – all results were negative
• no medium showed signs of bacterial contamination 1 week after preparation of the Pre-Cuts.
Summary
• Preparation – In our serie we experienced no major technical
problems– The procedure is straight forward with relatively
short learning curve
• Evaluation– In our serie evaluation of prepared tissue showed
high quality and high safety
Conclusion
• We believe that our eye bank prepares Pre-Cut corneas for lamelar keratoplasty of very high quality and safety
• We also believe that in the near future Pre-Cut corneas will represent majority of tissues provided by our eye bank
Thank you Thank you