the effect of stirring rate on the detection of hydrogen peroxide for carbon paste electrode...

1
The Effect of Stirring Rate on the Detection of Hydrogen Peroxide for Carbon Paste Electrode Modi fied with Meldola’s Blue Chi-Wen Lo ( 羅羅羅 ) , Chih-Ying Wu ( 羅羅羅 ) , Hau Lin ( 羅羅 ) Department of Chemical and Materials Engineering, Southern Taiwan University 南南南南南南南南南南南南南南南南 ABSTRACT . A study was conducted to use the Meldola’s Blue to modify the carbon paste electrode. Because the Meldola’s Blue possesses the excellent catalytic characteristic, it can be used with the graphite carbon powders which possesses the excellent conductivity to make the carbon paste electrode and to elevate the responding current of hydrogen peroxide. The responding current of hydrogen peroxide is detected in the phosphate buffer solution(PBS) and then the concentration of the hydrogen peroxide can be obtained and consequently, the concentration of the glucose can be determined. At 30℃, -300 mV operating potential, and in 0.05 M phosphate buffer solution (PBS), th e TB (Time Base) graphs for the carbon paste electrode at different stirring rates were plotted to evaluate the effect of stirring rate s on the responding current of detection of hydrogen peroxide. At the optimum operating conditions -300 mV operating potential, 500 rpm stirring rate and in 0.05 M PBS buffer solution ( pH = 7.4 ) , when the carbon paste electrode was modified with Meldola’s Blue [Meldola’s Blue : carbon powders : carbon paste = 2 15 15 ( weight ratio )] , the detection limit was 0.02 mM H 2 O 2 , the linear range was 0.02 1.4 mM H 2 O 2 , R 2 = 0.9994 , the sensitivity was 159.39 µA/cm 2 mM H 2 O 2 . Because hydrogen peroxide is used in the food industry for the purpose of preservation occasionally, a rapid and convenient hydrogen peroxide sensor for detection of the hydrogen peroxide is an important research subject. In recent years, diabetes has become one of the top ten causes of death for the people in Taiwan and therefore developing a rapid and convenient glucose biosensor also has become an important research subject. The glucose and oxygen can be catalyzed by the glucose oxidase to produce the gluconic acid and hydrogen peroxide, and therefore, as the concentration of hydrogen peroxide can be determined, the concentration of glucose can also be determined. Because the Meldola’s Blue possesses the excellent catalytic characteristic it can be used with the carbon paste and carbon powders which possess the high conductivity to make the carbon paste electrode and elevate the responding current of hydrogen peroxide. The responding current of hydrogen peroxide is detected in the phosphate buffer solution (PBS), and then the sensitivity of detec tion of hydrogen peroxide can be determined. A study of the TB (Ti me Base) graphs for the carbon paste electrode at different stirring rates was conducted to evaluate the effect of stirring ra tes on the responding current of detection of hydrogen peroxide. INTRODUCTION EXPERIMENTAL 1. Equipment Electrochemical Analyzer (CHI 401A, CH Instruments, Inc) was used to measure the activity of electrode by Cyclic Voltammetry ( CV ) and Time Base ( TB ) mode ; Oven (DENG YNG) ; Electric Stirrer(Fa rgo) ; pH meter (Metrohm 731); Constant Temperature Thermal Bath (Wi sdom BC-2DT 10L); Carbon Paste Electrode was used as the working ele ctrodes, Coiled Platinum Wire was used as the counter electrode and Ag / AgCl was used as the reference electrode. 2. Chemicals and Reagents Meldola’s Blue ; Hydrochloric Acid (HCl); Sodium Hydroxide (N aOH) ; Hydrogen Peroxide (H 2 O 2 ); Graphite Carbon Powder( C ); Carbon Paste; Cyclohexanone(C 6 H 10 O); Potassium Dihydrogenphosphate (KH 2 PO 4 ); Potassium Chloride (KCl). 3. Preparation of the Carbon Paste Electrode (1) Take one section of 7 cm electric wire with 0.05 cm inside diameter. After depriving the coating 0.5 cm length from both ends, the nake-end ed wire was washed, dried and ready for use. 7 cm 0.05 cm 0.5 cm (2) Then the Meldola’s Blue powders, graphite carbon powders and c arbon paste were mixed with the appropriate ratio [Meldola’s Blue : carbon powders : carbon paste = 2 15 15 ( weight ratio )]. Aft er the mixing was complete, the mixture was evenly coated on the na ke-ended electric wire and dried in the oven and then we obtained the carbon paste electrode. Meldola’s Blue Carbon Powders Mixing with Carbon Paste DEPARTMENT OF CHEMICAL AND MATERIALS ENGINEERING, SOUTHERN TAIWAN UNIVERSITY RESULTS AND DISCUSSION Table 1 The sensitivities, responding currents and R 2 values of different stirri rates for the carbon paste electrode modified with Meldola’s Blue Fig 1. CV graphs for (A) unmodi fied carbon paste electrode( the range of scanning potential: -0. 8 +0.8 V) (B) carbon paste ele ctrode modified with Meldola’s Blue ( the range of scanning pot ential: -0.8 +0.8 V) Fig. 2The TB graphs of carbon paste electrodes for detection of H 2 O 2 at different stirring rates[ (A) 300 rpm (B) 400 rpm (C) 500 rpm (D) 600 rpm ] Fig. 3 The calibration curves of different stirring rates for the carbon paste electrode modified with Meldola’s Blue. Fig. 5 The TB graphs of carb on paste electrode for determ ining the linear range of H 2 O 2 [Meldola’s Blue : carbon pow ders : carbon paste = 2 15 15 ( weight ratio )] CONCLUSIONS The results showed that the responding current for the car bon paste electrode modified with the Meldola’s Blue was elevated significantly. The TB (Time Base ) graphs at different stirring ra tes were plotted to evaluate the effect of stirring rate on the r esponding current of detection of hydrogen peroxide and determin e the optimum operating conditions. The results showed that at t he optimum operating conditions –300mV operating potential, 500rp m stirring rate and in 0.05M phosphate buffer solution(pH=7.4), t he detection limit was 0.02 mM, the linear range was 0.02~1.4 mM H 2 O 2 , R 2 =0.9994 and the sensitivity was 159.39 μA/cm 2 羅mM H 2 O 2 . REFERENCES 1. L. Mao and K. Yamamoto,“Glucose and Choline On-Line Biosensors Based on Electropolymerized Meldola’s Blue, ”Talanta, 51, 187 (2000). 2. Y.-M. Uang and T.-C. Chou , “Fabrication of Glucose Oxidase/Po lypyrrole Biosensor by Galvanostatic Method in Various pH Aqueous Solutions,” Biosens ors and Bioelectronics, 19, 141 (2003). 3. F. Qu, M. Yang, J. Jiang, G. Shen and R.Yu, “Amperometric Bios ensor for Choline Based on Layer-by-Layer Assembled Functionalized Carbon Nanotube an d Polyaniline Multilayer Film,” Analytical Biochemistry, 344, 108 (2005). 4. S. Timur, U. Anik, D. Odaci and L. Gorton, “Development of a Microbial Biosensor Based on Carbon Nanotube (CNT) Modified Electrodes,” Electrochemistry C ommunications, 9, 1810 (2007). 5. C. Guo, F. Hu, C. M. Li and P. K. Shen, “Direct Electrochemist Fig. 4 The TB graphs of carbon paste electrode for determining the detection limit of H 2 O 2 [Meldola’s Blue : carbon powders : carbon paste = 2 15 15 ( weight ratio )]

Upload: pierce-powers

Post on 04-Jan-2016

218 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: The Effect of Stirring Rate on the Detection of Hydrogen Peroxide for Carbon Paste Electrode Modified with Meldola’s Blue Chi-Wen Lo ( 羅濟玟 ), Chih-Ying

The Effect of Stirring Rate on the Detection of Hydrogen Peroxide for Carbon Paste Electrode Modified with Meldola’s Blue

Chi-Wen Lo ( 羅濟玟 ) , Chih-Ying Wu ( 巫致穎 ) , Hau Lin ( 林浩 )

Department of Chemical and Materials Engineering, Southern Taiwan University 南台科技大學化學工程與材料工程系

ABSTRACT

. A study was conducted to use the Meldola’s Blue to modify the carbon paste electrode. Because the Meldola’s Blue possesses the excellent catalytic characteristic, it can be used with the graphite carbon powders which possesses the excellent conductivity to make the carbon paste electrode and to elevate the responding current of hydrogen peroxide. The responding current of hydrogen peroxide is detected in the phosphate buffer solution(PBS) and then the concentration of the hydrogen peroxide can be obtained and consequently, the concentration of the glucose can be determined. At 30 , -300 mV operating potential, and in 0.05 M ℃ phosphate buffer solution (PBS), the TB (Time Base) graphs for the carbon paste electrode at different stirring rates were plotted to evaluate the effect of stirring rates on the responding current of detection of hydrogen peroxide. At the optimum operating conditions -300 mV operating potential, 500 rpm stirring rate and in 0.05 M PBS buffer solution ( pH = 7.4 ) , when the carbon paste electrode was modified with Meldola’s Blue [Meldola’s Blue : carbon powders : carbon paste = 2 : 15 :15 ( weight ratio )] , the detection limit was 0.02 mM H2O2 , the linear range was 0.02 ~ 1.4 mM H2O2 , R2 = 0.9994 , the sensitivity was 159.39 µA/cm2 . mM H2O2.

Because hydrogen peroxide is used in the food industry for the purpose of preservation occasionally, a rapid and convenient hydrogen peroxide sensor for detection of the hydrogen peroxide is an important research subject. In recent years, diabetes has become one of the top ten causes of death for the people in Taiwan and therefore developing a rapid and convenient glucose biosensor also has become an important research subject. The glucose and oxygen can be catalyzed by the glucose oxidase to produce the gluconic acid and hydrogen peroxide, and therefore, as the concentration of hydrogen peroxide can be determined, the concentration of glucose can also be determined. Because the Meldola’s Blue possesses the excellent catalytic characteristic it can be used with the carbon paste and carbon powders which possess the high conductivity to make the carbon paste electrode and elevate the responding current of hydrogen peroxide. The responding current of hydrogen peroxide is detected in the phosphate buffer solution (PBS), and then the sensitivity of detection of hydrogen peroxide can be determined. A study of the TB (Time Base) graphs for the carbon paste electrode at different stirring rates was conducted to evaluate the effect of stirring rates on the responding current of detection of hydrogen peroxide.

INTRODUCTION

EXPERIMENTAL

1. Equipment Electrochemical Analyzer (CHI 401A, CH Instruments, Inc) was used to measure the activity of electrode by Cyclic Voltammetry ( CV ) and Time Base ( TB ) mode ; Oven (DENG YNG) ; Electric Stirrer(Fargo) ; pH meter (Metrohm 731); Constant Temperature Thermal Bath (Wisdom BC-2DT 10L); Carbon Paste Electrode was used as the working electrodes, Coiled Platinum Wire was used as the counter electrode and Ag / AgCl was used as the reference electrode.

2. Chemicals and Reagents

Meldola’s Blue ; Hydrochloric Acid (HCl); Sodium Hydroxide (NaOH) ; Hydrogen Peroxide (H2O2); Graphite Carbon Powder( C ); Carbon Paste; Cyclohexanone(C6H10O); Potassium Dihydrogenphosphate (KH2PO4); Potassium Chloride (KCl).

3. Preparation of the Carbon Paste Electrode

(1) Take one section of 7 cm electric wire with 0.05 cm inside diameter. After depriving the coating 0.5 cm length from both ends, the nake-ended wire was washed, dried and ready for use.

7 cm

0.05 cm

0.5 cm

(2) Then the Meldola’s Blue powders, graphite carbon powders and carbon paste were mixed with the appropriate ratio [Meldola’s Blue : carbon powders : carbon paste = 2 : 15 : 15 ( weight ratio )]. After the mixing was complete, the mixture was evenly coated on the nake-ended electric wire and dried in the oven and then we obtained the carbon paste electrode.

Meldola’s Blue

Carbon Powders

Mixing with Carbon Paste

DEPARTMENT OF CHEMICAL AND MATERIALS ENGINEERING, SOUTHERN TAIWAN UNIVERSITY

RESULTS AND DISCUSSION

Table 1 The sensitivities, responding currents and R2 values of different stirring rates for the carbon paste electrode modified with Meldola’s Blue

Fig 1. CV graphs for (A) unmodified carbon paste electrode( the range of scanning potential: -0.8 ~ +0.8 V) (B) carbon paste electrode modified with Meldola’s Blue ( the range of scanning potential: -0.8 ~ +0.8 V)

Fig. 2The TB graphs of carbon paste electrodes for detection of H2O2 at different stirring rates[ (A) 300 rpm (B) 400 rpm (C) 500 rpm (D) 600 rpm ]

Fig. 3 The calibration curves of different stirring rates for the carbon paste electrode modified with Meldola’s Blue.

Fig. 5 The TB graphs of carbon paste electrode for determining the linear range of H2O2 [Meldola’s Blue : carbon powders : carbon paste = 2 : 15 : 15 ( weight ratio )]

CONCLUSIONS

The results showed that the responding current for the carbon paste electrode modified with the Meldola’s Blue was elevated significantly. The TB (Time Base ) graphs at different stirring rates were plotted to evaluate the effect of stirring rate on the responding current of detection of hydrogen peroxide and determine the optimum operating conditions. The results showed that at the optimum operating conditions –300mV operating potential, 500rpm stirring rate and in 0.05M phosphate buffer solution(pH=7.4), the detection limit was 0.02 mM, the linear range was 0.02~1.4 mM H2O2 , R2=0.9994 and the sensitivity was 159.39 μA/cm2 ּmM H2O2 .

REFERENCES

1. L. Mao and K. Yamamoto,“Glucose and Choline On-Line Biosensors Based on Electropolymerized Meldola’s Blue, ”Talanta, 51, 187 (2000). 2. Y.-M. Uang and T.-C. Chou , “Fabrication of Glucose Oxidase/Polypyrrole Biosensor by Galvanostatic Method in Various pH Aqueous Solutions,” Biosensors and Bioelectronics, 19, 141 (2003). 3. F. Qu, M. Yang, J. Jiang, G. Shen and R.Yu, “Amperometric Biosensor for Choline Based on Layer-by-Layer Assembled Functionalized Carbon Nanotube and Polyaniline Multilayer Film,” Analytical Biochemistry, 344, 108 (2005). 4. S. Timur, U. Anik, D. Odaci and L. Gorton, “Development of a Microbial Biosensor Based on Carbon Nanotube (CNT) Modified Electrodes,” Electrochemistry Communications, 9, 1810 (2007). 5. C. Guo, F. Hu, C. M. Li and P. K. Shen, “Direct Electrochemistry of Hemoglobin on Carbonized Titania Nanotubes and Its Application in a Sensitive Reagentless Hydrogen Peroxide Biosensor,” Biosensors and Bioelectronics, 24, 819 (2008).

Fig. 4 The TB graphs of carbon paste electrode for determining the detection limit of H2O2 [Meldola’s Blue : carbon powders : carbon paste = 2 : 15 : 15 ( weight ratio )]