Page 1: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

The CHROMacademy Essential Guide Understanding GC-MS Analysis Part 1

Page 2: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

Speakers

John Hinshaw GC Dept. Dean CHROMacademy

Tony Taylor Technical Director Crawford Scientific

Moderator

Peter Houston Editorial Director LCGC Magazine

eMeM gg 2)()(

Page 3: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS
Page 4: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

• When to use GC-MS • Overview of a typical GC-MS

System • GC Considerations • Interfacing the GC and MS

Components • Ionisation • Electron Impact Ionisation

Aims & Objectives

• Chemical Ionisation • Introduction to Mass Analysers • Ion Detection Systems • Vacuum Systems • Introduction to GC-MS data

analysis

Page 5: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

• GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS)

• GC-MS is primarily used for: - Identification or characterisation of analytes within the sample - Increased analytical sensitivity in the absence of a compound or

element specific detector such as ECD, NPD, FPD or chemiluminesence

When to Use GC-MS

Typical Electron Impact Ionisation Spectrum of the Cocaine molecule

GC-MS Total Ion Chromatogram used to quantify an environmental pollutant

Page 6: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

1. Pneumatic controls 2. Injector 3. Oven 4. Column 5. Interface

A Typical GC-MS System

6. Ion Source 7. Mass Analyser 8. Detector 9. Vacuum System 10. Control Electronics

Page 7: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

The GC-MS Process

1. Sample Introduction 2. Sample components separation 3. Transfer from the GC Column into the high vacuum of the

mass analyser 4. Ionisation of sample components 5. Separation and detection of gas phase ions

Page 8: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

GC Carrier Gas Considerations • Gas Purity of 99.999% recommended • Oxygen, Moisture and Hydrocarbons should be scrubbed from the gas

supply • Carrier gas flow rates limited to <2ml/min. due to detector vacuum

system which limits column choice to some extent

Typical Column Flow Rates for a number of Carrier Gas / Column Geometry Combinations

Page 9: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

GC Ferrules & Septa • Ferrules form a seal between the column and the inlet and detector

connections • GC-MS has a special requirement for ferrules to be impermeable to air

/ oxygen (reduces air background signal in the mass spectrometer) • Preferred ferrule materials for GC-MS are graphite / vespel composites or special metal ferrules • Septum bleed products must also be reduced

Septa Material & GC-MS Compatibility

Page 10: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

Stationary Phase Selection for GC-MS • Due to the nature of the detector – low bleed phases are preferred for

GC-MS • This reduces the background signal and contributes spurious ions to

the spectra of compounds of interest • Column Bleed increases with – Temperature Column Length, Film

Thickness & Polarity • Special low bleed phases exist with Silphenylene chemistry and many

manufacturers offer ‘MS’ designated phases • Specially immobilised PLOT columns should be used to prevent

evolution of particulate material into the ion-source

Silphenylene low-bleed Phenyl phase chemistry Traditional Phenyl

phase chemistry

Page 11: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

Common Contaminant Ions in GC-MS

Page 12: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

Interfacing GC with MS Detection

• The column outlet needs to be connected to the ion source of the mass spectrometer and fulfill the following conditions:

- Analyte must not condense in the interface - Analyte must not decompose before entering the mass

spectrometer ion source - The gas load (dictated by the mobile phase gas flow rate) entering

the ion source must be within the pumping capacity of the mass spectrometer

• Most capillary GC-MS interfaces directly couple the column exit with MS ion source.

• The advent of capillary columns brought about a significant reduction in the volumetric gas flow exiting the column (typically 1ml/min or below for columns of 0.32mm id and less), and the need to split the analyte away from the carrier gas to reduce gas load into the ion source was eliminated.

Page 13: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

‘Direct’ GC-MS Interfaces

• Column is inserted directly into the mass spectrometer ionisation chamber

• This interface gives the highest sensitivity

• Changing the GC column may be a time consuming process unless curtain gas devices are fitted

• Usually heated >10oC above the final oven temperature program and at a lower temperature than the ion source

• Analyte degradation should be avoided

Temperature Zones: Column Oven > Interface > Ion Source > Quadrupole

Page 14: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

mov1

Page 15: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

• The connection must be gas tight

• Traditional nut and ferrule couplings into the transfer device are common

GC – Interface Connection

• Several devices are available to facilitate vent-free columnchange including:

- Curtain gas connectors - Couplings with restrictors - Couplings using deactivated

silica tubing ‘pig-tails’

Page 16: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

1. Sample introduced directly into the ion source

2. Electrons are emitted from a heated filament

3. Accelerated using an appropriate potential (5-100V) to achieve the required electron energy

4. The industry standard for the electron energy is 70ev (electron volts) which

5. Resulting ions are accelerated out of the source using an electrostatic potential

6. Ion beam is focused using electrostatic ‘lenses’

Analyte Ionisation – Electron Impact (EI)

Page 17: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

mov2

Page 18: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

Electron Impact Ionisation Process

eMeM gg 2)()(

1. The bombarding electron at 70eV abstracts an electron from the analyte molecule (M)

2. All 70eV is transferred to the analyte molecule which becomes a Radical Cation

3. Depending upon the magnitude of the analyte 1st Ionisation Potential (energy required to remove 1 electron from an atom within the molecule) there may be excess energy remaining

4. This remaining energy may cause the analyte to fragment through further bond cleavage

5. The molecule may also undergo simple intermolecular re-arrangement reactions

6. Any charged fragments and re-arrangement products will appear as signals within the mass spectrum

Page 19: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

Typical EI Spectra

1. EI is a relatively ‘harsh technique’ – inducing fragmentation in many cases

2. The intensity of the molecular ion varies depending on analyte radical cation stability

3. The fragmentation pattern can be used to identify the analyte molecule and acts as a ‘fingerprint’ of the molecule

4. Commercial (70eV) or in-house libraries may be searched by the instrument data system to provide a tentative ‘match’ for the unknown spectrum

5. Experienced users may undertake ab. initio. interpretation to elucidate the molecular structure from the mass spectrum

Page 20: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

Analyte Ionisation – Chemical Ionisation (CI) 1. Chemical ionisation involves the

ionisation of a reagent gas, such as methane or ammonia at relatively high pressure (~1 mbar) in an electron impact source

2. Once produced, the reagent gas ions collide with the analyte molecules producing ions through gas phase reaction processes such as proton transfer and adduction

3. Reagent gas, gas pressure and stability of gas pressure are critical

Page 21: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

mov3

Page 22: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

Chemical Ionisation Process

eNHeNH 244

][34 HMNHMNH

][ 44 NHMMNH

(Production of the reagent gas ion)

(Analyte ionised by Proton Transfer)

(Analyte ionised by Adduct Formation)

1. CI is a ‘soft’ ionisation technique compared to EI 2. All 70eV is transferred to the analyte molecule which becomes a

Radical Cation 3. Appearance of the spectra heavily influence buy the choice of

reagent gas – methane / isobutane / ammonia / combinations 4. Tuning the mass spectrometer to calibrate the mass axis / resolution

and relative response across the mass range is more difficult in CI

Page 23: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

Typical CI Spectra Afalatoxin B1

1. Energy of reagent gas rarely exceeds 5eV

2. Softer ionisation technique gives rise to more intense molecular ions and fewer fragments (library searching more difficult)

3. Useful for obtaining the molecular weight of the analyte 4. Operates in two modes – negative ion and positive mode 5. Negative ion mode is highly sensitive used for ultra-trace analysis of

halogenated compounds (capable of electron capture) 6. Positive ion mode less sensitive as reagent gas is also detected

Page 24: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

Mass Analysers – The Quadrupole

1. Ions are separated according to their mass-to-charge ratio (m/z) as they pass along the central axis of four parallel equidistant rods (or poles)

2. Ion separation is performed using controlled voltages applied to the mass analyser rods which impart an electrostatic field inside the analysing device.

3. As long as x and y, which determine the position of an ion from the centre of the rods, remains less than r0, the ion will be able to pass through the quadrupole without touching the rods. This is known as a non-collisional or stable trajectory.

Page 25: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

Mass Analysers – The Quadrupole (II)

4. Where the ion is caused to oscillate with a trajectory whose amplitude exceeds r0 it will collide with a rod, and become discharged and subsequently pumped to waste

5. This is known as an unstable or collisional trajectory.

Page 26: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

Quadrupole – Advantages & Limitations

Advantages Disadvantages

Reproducibility

Low cost

Low resolution

Mass discrimination. (Peak height

vs. mass response must be 'tuned‘)

Limited scanning speeds compared

to other analyser types

(Not suiable for very fast GC analysis

with very narrow peaks)

Page 27: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

Mass Analysers - Time of Flight (ToF)

1. Ions are extracted (or produced) in short bursts or packets within the ion source and subjected to an accelerating voltage.

2. The ions then ‘drift’ or ‘fly’ down an evacuated tube of a set length (‘d’). Once free from the region of accelerating voltage the speed at which the ions travel down the tube is dependant upon their mass (m) and charge (z).

3. All ions are detected (almost) simultaneously. 4. Scanning the mass range of all ions is very rapid and as such the

inherent sensitivity of the instrument is increased.

Page 28: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

Mass Analysers - Time of Flight (ToF) (II)

5. Most modern ToF instruments use orthogonal acceleration (‘oa’) pusher for improved instrument performance

6. Increasing the length of the flight tube using an ‘ion mirror’ or ‘reflectron’ leads to an increase in the resolution of the instrument

Page 29: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

mov4

Page 30: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

ToF – Advantages and Limitations Advantages Disadvantages

• High ion transmission

• Highest practical mass range of all MS analyzers

• Very Low Detection limits

• High mass accuracy and resolution

• (Will give elemental composition possibilities)

• Fast digitizers used in TOF can have limited dynamic range

• Cost (but this is becoming less of a barrier)

Page 31: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

Mass Analysers – Ion Traps

1. Ion trap mass spectrometers work on the basis of storing ions in a “trap”, and manipulating the ions by using applied DC and RF fields.

2. The amplitude of the applied voltages enables the analyser to trap ions of specified mass to charge ratio within the analysing device.

3. Non-selected ions are given a trajectory by the electrostatic field that causes them to exit the trap.

Page 32: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

Mass Analysers – Ion Traps (II)

4. By filling the trap with an inert gas fragmentation of selected ions is possible - useful when structural information is required.

5. Can perform multiple product ion scans with very good sensitivity (MSn)

6. Spectra acquired with an ion trap mass analyser may be significantly different to those acquired from a triple quadrupole system due to the different collision regimes within the systems (collision energy/gas).

Page 33: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

mov5

Page 34: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

Ion Traps – Advantages & Limitations

Advantages Disadvantages

• High sensitivity

• Multiple Product Ion

scan capability (MS)n

• High resolution

• Good specificity and capable of data dependant scanning (i.e. Automated MS/MS)

• Produces very unusual spectra if the

ions are stored in the trap too long.

• Easily saturated

• Poor for low mass work (below 100 Da)

• Poor dynamic range (except the most modern devices) and hence may have limited quantitative use

Page 35: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

GC-MS Detectors

• Once the ions have passed the mass analyser they have to be detected and transformed into a usable signal.

• The detector is an important element of the mass spectrometer that generates a signal from incident ions by either generating secondary electrons, which are further amplified, or by inducing a current (generated by moving charges)

Page 36: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

GC-MS Detectors

Point detectors: ions are not spatially resolved and sequentially impinge upon a detector situated at a single point within the spectrometer geometry. Array detectors: ions are spatially resolved and all ions arrive simultaneously (or near simultaneously) and are recorded along a plane using a bank of detectors.

Page 37: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

GC-MS Vacuum Systems

• The entire MS process must be carried out at very low pressures (~10-8 atm)

• A high level of vacuum within the instrument assists the processes of

ion movement and separation in the following ways: • By providing an adequate mean free path for the analyte ions • By providing collision free ion trajectories • By reducing ion-molecular reactions • By reducing background interference

• Vacuum systems consist of a differentially pumped arrangement: a foreline pump establishing a ‘rough’ vacuum and a high vacuum pump or pumps situated on the analyser body to establish the high levels of vacuum required for effective mass to charge ratio measurement.

Page 38: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

GC-MS Vacuum Systems

Oil Filled Rotary Foreline (Rough) Pump: 10-2 Torr

High Vacuum Turbomolecular Pump: 10-6 to 10-8 Torr

Page 39: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

GC-MS Data

• Quadrupole GC-MS systems operate in two distinct modes – ‘Scanning’ and ‘Selected Ion Recording’

• In SCAN mode, the quadrupole settings are ramped through a range of values which allow successively lower mass to charge ratio ions to pass through the analyser ‘scaning’ the mass range of ions emerging from the quadrupole device

• The scanning operation takes a finite time to complete (although scan rates of 5-20Hz are typical) and each individual m/z value is measured for only a fraction of the time that they elute into the mass analyser

• The Total Ion Chromtogram (Current) (TIC) is constructed by summing the abundance of all ions within a spectrum at a particular time point and plotting total abundance against the time at which the spectrum was acquired

Page 40: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

mov6

Page 41: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

GC-MS Data

• By choosing to set the quadrupole to certain voltage values, we are able to choose only certain masses for transmission through the mass analyser

• This type of spectral experiment, called selected or specific ion monitoring (SIM), or selected ion recording (SIR), has certain advantages over scanning wide mass ranges

• The advantage here is gained because the analyser concentrates on ‘useful’ ion signals from the analyte of interest and not on noise – gaining significantly on signal to noise ratio

• Because not all m/z values are recorded the mass analyser can carry out a SIM experiment very rapidly (102 - 104 increase over scanning experiment speeds) therefore acquiring more data points to accurately ‘model’ the peak shape

Page 42: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS

GC-MS Data

Page 43: The CHROMacademy Essential Guide Understanding GC-MS ... · •GC-MS combines the separating power of Gas Chromatography (GC), with the detection power of mass spectrometry (MS) •GC-MS