testo · web viewpos: 1 /td/kurzanleitungen/testo 480/behaglichkeitssonde 0628 0143 @...

8
Comfort probe Application information 0970 0472

Upload: others

Post on 03-Mar-2021

4 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Testo · Web viewPos: 1 /TD/Kurzanleitungen/testo 480/Behaglichkeitssonde 0628 0143 @ 9\mod_1309359060255_79.docx @ 81219 @ 555 @ 4 Application The comfort probe 0628 0143 is used

Comfort probe

Application information

0970 0472

Page 2: Testo · Web viewPos: 1 /TD/Kurzanleitungen/testo 480/Behaglichkeitssonde 0628 0143 @ 9\mod_1309359060255_79.docx @ 81219 @ 555 @ 4 Application The comfort probe 0628 0143 is used

2

Page 3: Testo · Web viewPos: 1 /TD/Kurzanleitungen/testo 480/Behaglichkeitssonde 0628 0143 @ 9\mod_1309359060255_79.docx @ 81219 @ 555 @ 4 Application The comfort probe 0628 0143 is used

Pos: 1 /TD/Kurzanleitungen/testo 480/Behaglichkeitssonde 0628 0143 @ 9\mod_1309359060255_79.docx @ 81219 @ 555 @ 4

ApplicationThe comfort probe 0628 0143 is used in conjunction with testo 480 to determine the air velocity and draught risk in accordance with DIN EN 13779.

Observe the information on the measuring process in the instrument instruction manual.

Overview

1 Sensors

3

Page 4: Testo · Web viewPos: 1 /TD/Kurzanleitungen/testo 480/Behaglichkeitssonde 0628 0143 @ 9\mod_1309359060255_79.docx @ 81219 @ 555 @ 4 Application The comfort probe 0628 0143 is used

CAUTION

Damage to sensor!> Do not touch the sensor.> Cover over with the protective sleeve after the measurement.

2 Handle 3 Flow direction (flow against testo logo)4 Connection for plug-in head cable (art. no. 0430 0100)

Technical data

Feature Values

Flow measuring range

0 to +5 m/s

Temperature measuring range

0 to 50 °C (probe head)

Operating range, handle

0 to 40 °C

Accuracy (at nominal temperature 22 °C) ± 1 digit

±0.5 °C

±(0.03 m/s + 4% of meas. val.)

At low air velocities, slightly increased temperature display!

Absolute pressure sensor

measuring range: +700...+1100 hPa

Accuracy: ±3 hPa

Adjustment conditions

Wind tunnel, adjusted in free jet Ø 350 mm, reference pressure 1013 hPa, flow against testo logo

The digital probe allows measurement values to be processed directly in the probe. This technology eliminates instrument measurement uncertainty.For calibration, the probe alone (without the hand instrument) can be sent away.

Entering the determined adjustment data via the

4

Page 5: Testo · Web viewPos: 1 /TD/Kurzanleitungen/testo 480/Behaglichkeitssonde 0628 0143 @ 9\mod_1309359060255_79.docx @ 81219 @ 555 @ 4 Application The comfort probe 0628 0143 is used

EasyClimate software generates a zero-error display.

=== Ende der Liste für Textmarke Inhalt ===

5

Page 6: Testo · Web viewPos: 1 /TD/Kurzanleitungen/testo 480/Behaglichkeitssonde 0628 0143 @ 9\mod_1309359060255_79.docx @ 81219 @ 555 @ 4 Application The comfort probe 0628 0143 is used

0970 0472 en 02