Topics
• How does EO sterilize• Basic factors affecting sterilization• Calculating D value• Biological Indicators• An EO Cycle• Cycle Validation• EO Residues• Product Release
![Page 3: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/3.jpg)
Governing Standard• ANSI/AAMI/ISO 11135: Medical Devices –
Validation and routine control of ethylene oxide sterilization.
• www.aami.org
![Page 4: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/4.jpg)
How Does EO Sterilize?
![Page 5: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/5.jpg)
EO as a Sterilant
EO is the simplest epoxy• Boiling point 10.4°C (50.7°F)• Extremely reactive• Kills by alkylation
Irreversibly binds to key molecules in cellDNAProteins (enzymes)
![Page 6: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/6.jpg)
EO as a Sterilant
Molecules no longer function Cell no longer reproduces Cell dies
![Page 7: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/7.jpg)
Basic Factors Affecting EO Sterilization
![Page 8: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/8.jpg)
Four Basic Factors to Microbial Kill
•EO concentration•Water Vapor (Humidity)•Temperature•Time
![Page 9: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/9.jpg)
EO Concentration
•Commonly 400 and 700 mg/l •As EO concentration increases at a
given temp and RH, ▫microbial inactivation (kill) rate
increases
![Page 10: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/10.jpg)
EO Concentration
•Killing concentration required at the site where the microbes (BI’s) are located
![Page 11: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/11.jpg)
Water Vapor (Humidity)
•Required for EO to react with the critical cell molecules
Generally measured as Relative Humidity
![Page 12: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/12.jpg)
Temperature
•Kill rate increases with temperature▫D-value decreases (time) with
temperature▫Exponential function
•For each 10°C (18°F) rise in temperature, the spore inactivation rate will generally double
![Page 13: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/13.jpg)
Time
•Amount of kill increases with exposure (EO gas dwell) time
•90 percent of surviving microbes are killed for each D-value time in gas dwell
![Page 14: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/14.jpg)
D-Value
•Determined to prove predictable logarithmic death kinetics of challenge microorganisms (BIs) or natural bioburden
![Page 15: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/15.jpg)
D-Value
•“Decimal” Reduction Time - Measure of the biological organism’s resistance to the sterilant
•Time in minutes necessary to reduce (kill) a microbial population by one logarithm or 90%
![Page 16: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/16.jpg)
A Simple D-Value Calculation
TimeLog of Starting Population – Log of Final
Population
![Page 17: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/17.jpg)
Example•1,000,000 or 106 spores. (the Log = 6)•20 minute gas dwell•Result is 10,000 or 104 spores
(the Log = 4)
![Page 18: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/18.jpg)
Time = 20 minutesLog of Starting Population (which equals 6) – Log of Final Population
(which equals 4)
![Page 19: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/19.jpg)
20 minutes6 – 4
Thus: D-Value is 20 divided by 2 = 10 minutes
![Page 20: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/20.jpg)
What does that mean?
•It takes 10 minutes to go from:▫1,000,000 to 100,000▫ or 10,000 to 1,000▫ or 10 to 1▫ Etc.
![Page 21: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/21.jpg)
Biological Indicators
![Page 22: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/22.jpg)
Biological Indicators• In EO sterilization,
we use a paper strip impregnated with about a million spores . The strip is in a gas-permeable, bacteria-impermeable glassine envelope
![Page 23: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/23.jpg)
Biological Indicators (BIs)
•More difficult to kill than natural occurring bioburden (microrganisms in or on product)
![Page 24: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/24.jpg)
Biological Indicators (BIs)
•Traditionally, >106 spores of an organism (i.e., Bacillus atrophaeus) highly resistant to EO process
Spores
![Page 25: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/25.jpg)
Bacillus???
•The name of the bacterium is commonly Bacillus subtilis or B. subtilis.
• It has been renamed and is officially B. atrophaeus.
![Page 26: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/26.jpg)
Biological Indicators (BIs)
•Provided in various forms for industrial applications:▫Placed in product to represent
most resistant location in product
![Page 27: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/27.jpg)
Examples of Biological Indicators
SCBI Disc Strip SCBI
![Page 28: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/28.jpg)
Placing the BI: The BI is positioned in the
Most Difficult Location for gas to reach
![Page 29: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/29.jpg)
Testing EO-exposed BIs
•The BI strip or disc is planted into a test tube containing Tryptic Soy Broth (TSB).
•The strips are incubated at 30-35oC for seven days (USP).
![Page 30: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/30.jpg)
A Positive
Orange pellicle
Cloudy
Sediment
TSA Agar(orange)
Tyrosine Agar (black)
![Page 31: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/31.jpg)
The EO Cycle
![Page 32: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/32.jpg)
Conventional Cycle•Three steps▫Preconditioning▫Sterilization ▫Aeration
![Page 33: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/33.jpg)
Preconditioning•First “processing” step of EO
sterilization.•Ensures the load is always
processed at roughly the same temperature and humidity upon entry to the chamber.
![Page 34: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/34.jpg)
Preconditioning
•Heat and Humidity Introduced to Load
•Air circulated to assist heat/humidity transfer
•Generally 18-24 hrs
![Page 35: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/35.jpg)
Chamber
![Page 36: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/36.jpg)
EO Sterilization Cycle
Generic 100 Percent Cycle
0.0
5.0
10.0
15.0
20.0
25.0
30.0
0 100 200 300 400 500 600 700
Time from Cycle Start (minutes)
Pres
sure
(inH
gA)
Initial Vacuum
Leak Test
Humidification
Humidity Dwell
Gas Inbleed
Gas Dwell
After Vacuum
Gas Wash
Final Release
![Page 37: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/37.jpg)
Cycle Validation
![Page 38: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/38.jpg)
•Rigorous microbiological and physical testing
•Beyond that used in routine monitoring for either:• Conventional• Parametric
Cycle Validation
![Page 39: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/39.jpg)
• Conventional Cycle Run 3 Half Cycles with parameters established during Cycle Development Culture BIs (2 to 7 days)
Run minimum 1 Full Cycle to evaluate functionality and residuals
Cycle Validation
![Page 40: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/40.jpg)
•Conventional Cycle (continued)▫Aeration Requirements
Develop dissipation curve to establish release time
Aerate long enough to assure safe residuals typically around 2 days on average
Cycle Validation
![Page 41: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/41.jpg)
Conventional Cycle (continued)• Residual allowable limits are based on
application of product ISO 10993-7 Qualify release time based on three (3) separate lots.• May be satisfied with first two routine runs with end-point residual testing (to supplement dissipation curve)
Cycle Validation
![Page 42: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/42.jpg)
Parametric ReleaseRelease on parameters, not BIAdded parameters; Load temp prior to Precon Load temp at end of Precon Chamber RH during humidity Chamber EO during dwell Load temp during dwell
Cycle Validation
![Page 43: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/43.jpg)
Parametric ReleaseBI Requirements NONE
Cycle Validation
![Page 44: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/44.jpg)
Critical step in overall validation • Periodic (typically annual)
requalification studies designed to detect inadvertant process changes
• Recommended upon introduction of new or significant changes in: Product Process Packaging Equipment
Revalidation
![Page 45: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/45.jpg)
•Options:
▫Perform one (1) micro PQ run in each qualified chamber
Revalidation
![Page 46: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/46.jpg)
•Options
▫Paper Revalidation: Documented evaluation of process to verify no process shifts throughout year. May be used as a tool, however not
recommended for continuous replacement of physical requalification.
Revalidation
![Page 47: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/47.jpg)
AERATION
![Page 48: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/48.jpg)
Aeration (Degassing)
•Last “processing” step.▫Aeration Room physically similar to
Preconditioning, but with no “Humidity”•Two methods of Aeration available▫Heated▫Ambient
![Page 49: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/49.jpg)
Aeration (Degassing)
Heated Aeration•Heat Introduced to Load▫assist in degassing process
•Typically 2 days on average for Heated Aeration
![Page 50: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/50.jpg)
Aeration (Degassing)
Ambient Aeration•generally has much longer minimum
time requirements
![Page 51: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/51.jpg)
EO Sterilant Residues
![Page 52: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/52.jpg)
EO Sterilant Residues
•What is a sterilant residue?•Why do we need to test?•Who tells us we have to test?•How do we test for these residues?
![Page 53: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/53.jpg)
What is a Sterilant Residue?
Sterilant residues are compounds that remain on product after ethylene oxide sterilization.
![Page 54: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/54.jpg)
Residues - What are They?
•Ethylene Oxide
•Ethylene Chlorohydrin - Formed when EO reacts with chlorine or chloride ions. Chlorine may come from salt or bleach.
•Ethylene Glycol - Formed when EO reacts with water.
![Page 55: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/55.jpg)
Why Test?
•Patient safety.
•Regulatory limits.
![Page 56: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/56.jpg)
Who Requires Testing?
•FR, Volume 43, No. 122- June 23, 1978 (FDA)
•ANSI/AAMI/ISO 10993-7
![Page 57: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/57.jpg)
Factors Affecting Levels of EO
Residual amount depends on:▫Product materials▫Chamber gas concentration▫Gas dwell time▫Aeration conditions and length of aeration time
![Page 58: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/58.jpg)
Product Release
![Page 59: "Terminal Sterilization Basics of EO"](https://reader034.vdocuments.site/reader034/viewer/2022051403/589ce0641a28abf86d8b5141/html5/thumbnails/59.jpg)
Product Release
•Before or after aeration BI’s are pulled from load and sent to lab for testing (7 day test)
• If Parametric Release has been validated, load can be released immediately after aeration. (No BI’s required)