Page 1: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

Technical Manual

EM0281-02

Q YU TA L I

ENTM SE YG SA TN E

A M

M

R

Temperature AdvancedCalibrators

TA-25NL and TA-45NL

presys

Page 2: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

EC Declaration of Conformity

We declare under our sole responsability that the CE marked products, are in conformity with the essential requirements of the following EC Directives when installed in accordance with the installation instructions contained in the product documentation:

Series TA-25NL, TA-45NL Description Dry-Block Temperature Calibrator

LVD Low Voltage Directive

2006/95/EC of the European Parliament and of the Council of 12 December 2006 on the harmonization of the laws of Member States relating to Electrical Equipment designed for use within certain voltage limits.

EN 61010-1:2011 Safety requirements for electrical equipment for measurement, control and laboratory use

EN 61010-2:010 Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-010: Particular requirements for laboratory equipment for the heating of Materials.

EMC directive 2004/108/EC of the European Parliament and of the Council of 15 December 2004 on the approximation of the laws of the Member States relating to electromagnetic compatibility and repealing Directive 89/336/EEC

EN 61326-1:2003 Electrical equipment for measurement, control and laboratory use - EMC requirements

São Paulo, 8 September 2015

Vinicius José Gomes Nunes Antonio Rafael Sito Antunes CEO Engineering Manager

presys

Page 3: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

The warranty conditions are available in our sites: www.presys.com.br/warranty

presys

Page 4: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

EM0281-02

Table of Contents

1.0 - Introduction ............................................................................................................ 1 1.1 - Technical Specifications ........................................................................................ 2

1.1.1 - Input Technical Specifications ........................................................................ 3 1.1.2 - Special Software Features.............................................................................. 4

1.2 - Order Code ........................................................................................................... 4 1.3 - Accessories .......................................................................................................... 5 1.4 - Parts Identification ................................................................................................ 2 1.5 - Instructions for Optional ........................................................................................ 3

2.0 - Calibrator Operation ............................................................................................... 6 2.1 - Calibrator Menu .................................................................................................... 7

2.1.1 - Probe Settings ................................................................................................ 9 2.1.2 - Input Settings ............................................................................................... 11 2.1.3 - Special Function ........................................................................................... 14 2.1.4 - Saving Current Configuration (Memory Manager) ........................................ 15

2.2 - HART® Configuration .......................................................................................... 16 2.2.1 - HART® Connections ..................................................................................... 16 2.2.2 - Starting Communication ............................................................................... 17 2.2.3 - Adjusting the Measurement Range of a HART® Transmitter ........................ 17 2.2.4 - Adjusting the Measurement Range of a HART® Transmitter with Reference 18 2.2.5 - Checking / Adjusting HART® Transmitter mA Output .................................... 19

2.3 - Automatic Tasks ................................................................................................. 20 2.3.1 - Creating Tasks ............................................................................................. 20 2.3.2 - Performing Tasks ......................................................................................... 22 2.3.3 - Viewing Results ............................................................................................ 23

2.4 - Data-Logger ........................................................................................................ 24 2.5 - Videos ................................................................................................................. 26 2.6 - Settings ............................................................................................................... 26

2.6.1 - System ......................................................................................................... 26 2.6.2 - Network ........................................................................................................ 27 2.6.3 - Built-in Web Server ...................................................................................... 27

3 - Safety Instructions .................................................................................................. 29

4 - Recommendations as regards Accuracy of Measurements ................................ 29

5 - Calibration (Adjustment) ......................................................................................... 30 5.1 - Input Calibration .................................................................................................. 30 5.2 - Probe Calibration ................................................................................................ 32

6.0 - Maintenance .......................................................................................................... 33 6.1 - Instructions for Hardware Maintenance ............................................................... 33 6.2 - Instructions for Insert Jamming ........................................................................... 33

presys

Page 5: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 1

1.0 - Introduction

TA-25NL / TA-45NL

Temperature Advanced Calibrators TA-25NL and TA-45NL generate temperature in the insert in order to calibrate thermocouples, RTDs, glass thermometers, thermo-switches. Besides providing high accuracy temperature values, it also allows the measurement of signals generated by the thermo-element which is being calibrated. This is possible due to an embedded calibrator specific for these types of signal, including 4-20 mA. Thus, they incorporate the functions of dry block, standard thermometer and calibrator for RTD, TC and mA, stirred liquid and blackbody source.

TA-45NL calibrator model generates temperatures from -45 ºC to 155 ºC

TA-25NL calibrator model generates temperatures from -25 ºC to 155 ºC.

The calibrator also provides an input for an external probe to perform the temperature control from a standard sensor (optional) inserted in the same measuring zone of the sensor to be calibrated, increasing the accuracy and decreasing setpoint errors and loading effects. The standard sensor calibration curve follows the parameterization of Callendar-Van Dusen.

They present a wide variety of programming resources, allowing the performance of automatic calibrations. In this case, the sensor is placed in the insert and its electrical terminals are connected to the embedded calibrator. The operator defines the calibration points and the number of repetitions (task), then the process is started and all the sequence is automatically accomplished. After completing the task, a Calibration Report is issued and it can be printed directly on a USB connected printer or a PDF document can be generated. It has HART® communication (optional) for reading and setting parameters of field devices that have this protocol.

presys

Page 6: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 2

Another way of performing automatic documented calibrations is by means of Calibration Software for PC/WindowsTM, which uses USB communication or Ethernet to connect the computer.

TA Calibrators have also many other features, such as:

RTD input for 2, 3 and 4 wires. Table IEC 60751, JIS or Callendar-Van Dusen user-configurable. Engineering units configurable to °C, °F and K.

Present inputs for mA, thermocouples and thermo-switches. Thermo-element reading scaled to ITS-90 or IPTS-68. Use of internal standard thermometer. Accuracy to 0.1 ºC, stability of 0.02 ºC and resolution of 0.01ºC. Carry out completely automatic calibrations without the use of a computer. Built in Web Server, Ethernet communication and USB serial Communication. USB port for software/firmware upgrade. HART® Communication Protocol (optional) with internal resistance configurable,

transmitter power supply and latest DD as option. The electric signal calibrator is independent from the dry block function. Display indication when the temperature reaches the desired value. 5.7 inches touchscreen display that eases the operation and configuration of the

calibrator. Internal regulated 24 Vdc power supply for 2-wire transmitters. Independent circuit for over-temperature protection and safety. Insert to choose, soft carrying case, handles and test leads included.

1.1 - Technical Specifications

TA-25NL TA-45NL Operating Range -25 ºC to 155 ºC (1) -45 ºC to 155 ºC (1) Power Supply 115 Vac or 230 Vac 50/60 Hz, according to order code. Well Diameter / Depth 35 mm x 160 mm Resolution 0.01 ºC (0.01 ºF) Accuracy

internal reference with external reference

with external standard

0.10 ºC 0.07 ºC 0.05 ºC

Stability (after 10 min) 0.02 ºC (0.04 ºF) Power Consumption 300 W (rated) 450 W (rated) Heating Rate 25 minutes (25 ºC to 140 ºC) 20 minutes (25 ºC to 140 ºC) Cooling Rate 20 minutes (25 ºC to -25 ºC) 40 minutes (25 ºC to -45 ºC) Dimensions (H,W,D) 315 x 200 x 305 mm Weight 12 kg Warranty 1 year, except for Peltier effect elements

(1) Ambient Temperature: 23 ºC.

presys

Page 7: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 3

1.1.1 - Input Technical Specifications

Input Ranges Resolution Accuracy Remarks millivolt -150 mV to 150 mV 0.001 mV 0.01 % FS* Rinput > 10 M

150 mV to 2450 mV 0.01 mV 0.02 % FS auto-ranging mA -1 mA to 24.5 mA 0.0001 mA 0.01 % FS Rinput < 120

resistance 0 to 400 400 to 2500

0.01 0.01

0.01 % FS 0.03 % FS

Excitation current 0.85 mA auto-ranging

Pt-100 -200 to 850 C / -328 to 1562 F 0.01 C / 0.01 F 0.1 C / 0.2 F IEC 60751 Pt-1000 -200 to 400 C / -328 to 752 F 0.1 C / 0.1 F 0.1 C / 0.2 F IEC 60751

Cu-10 -200 to 260 C / -328 to 500 F 0.1 C / 0.1 F 2.0 C / 4.0 F Minco 16-9 Ni-100 -60 to 250 C / -76 to 482 F 0.1 C / 0.1 F 0.2 C / 0.4 F DIN-43760

TC-J -210 to 1200 C / -346 to 2192 F 0.1 C / 0.1 F 0.2 C / 0.4 F IEC 60584 TC-K -270 to -150 C / -454 to -238 F 0.1 C / 0.1 F 0.5 C / 1.0 F IEC 60584 TC-K -150 to 1370 C / -238 to 2498 F 0.1 C / 0.1 F 0.2 C / 0.4 F IEC 60584 TC-T -260 to -200 C / -436 to -328 F 0.1 C / 0.1 F 0.6 C / 1.2 F IEC 60584 TC-T -200 to -75 C / -328 to -103 F 0.1 C / 0.1 F 0.4 C / 0.8 F IEC 60584 TC-T -75 to 400 C / -103 to 752 F 0.1 C / 0.1 F 0.2 C / 0.4 F IEC 60584 TC-B 50 to 250 C / 122 to 482 F 0.1 C / 0.1 F 2.5 C / 5.0 F IEC 60584 TC-B 250 to 500 C / 482 to 932 F 0.1 C / 0.1 F 1.5 C / 3.0 F IEC 60584 TC-B 500 to 1200 C / 932 to 2192 F 0.1 C / 0.1 F 1.0 C / 2.0 F IEC 60584 TC-B 1200 to 1820 C / 2192 to 3308 F 0.1 C / 0.1 F 0.7 C / 1.4 F IEC 60584 TC-R -50 to 300 C / -58 to 572 F 0.1 C / 0.1 F 1.0 C / 2.0 F IEC 60584 TC-R 300 to 1760 C / 572 to 3200 F 0.1 C / 0.1 F 0.7 C / 1.4 F IEC 60584 TC-S -50 to 300 C / -58 to 572 F 0.1 C / 0.1 F 1.0 C / 2.0 F IEC 60584 TC-S 300 to 1760 C / 572 to 3200 F 0.1 C / 0.1 F 0.7 C / 1.4 F IEC 60584 TC-E -270 to -150 C / -454 to -238 F 0.1 C / 0.1 F 0.3 C / 0.6 F IEC 60584 TC-E -150 to 1000 C / -238 to 1832 F 0.1 C / 0.1 F 0.1 C / 0.2 F IEC 60584 TC-N -260 to -200 C / -436 to -328 F 0.1 C / 0.1 F 1.0 C / 2.0 F IEC 60584 TC-N -200 to -20 C / -328 to -4 F 0.1 C / 0.1 F 0.4 C / 0.8 F IEC 60584 TC-N -20 to 1300 C / -4 to 2372 F 0.1 C / 0.1 F 0.2 C / 0.4 F IEC 60584 TC-L -200 to 900 C / -328 to 1652 F 0.1 C / 0.1 F 0.2 C / 0.4 F DIN-43710 TC-C 0 to 1500 C / 32 to 2732 F 0.1 C / 0.1 F 0.5 C / 1.0 F W5Re / W26Re TC-C 1500 to 2320 C / 2732 to 4208 F 0.1 C / 0.1 F 0.7 C / 1.4 F W5Re / W 26Re

Accuracy values are valid within one year and temperature range of 20 to 26 °C. Outside these limits add 0.001 % FS / C taking 23 °C as the reference temperature. For thermocouples, using the internal cold junction compensation add a cold junction compensation error of 0.2 C or 0.4 F max. Note: The times for cooling and heating rate presented refer exclusively to the use of the solid (metallic) insert. For use with stirred liquid insert the time may vary according to the fluid used and its viscosity. pr

esys

Page 8: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 4

1.1.2 - Special Software Features

- Special Functions:

SCALE: makes the scaling of mA input. STEP: steps or set points with configurable time.

- Memory Manager: stores configuration types predefined by the user.

- Automated Tasks: creating of calibration work orders and automatic execution of calibration services, storage of data and reporting.

- Data Logger: monitoring of input or output signals, storage and visualization of data in chart or table.

- Videos: storage and viewing videos on the calibrator screen. 1.2 - Order Code

Notes: * HART® is a FieldComm Group trademark. pr

esys

Page 9: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 5

1.3 - Accessories Dry Block Insert:

Inserts Holes Order Code IN01 1 x 3/4" 06.04.0041-00 IN02 1 x 1/2" 06.04.0042-00 IN03 1 x 6.0mm and 3 x 1/4" 06.04.0043-00 IN04 3 x 6.0mm and 1 x 1/4" 06.04.0044-00 IN05 4 x 6.0mm 06.04.0045-00 IN06 2 x 6.0mm and 2 x 1/4" 06.04.0046-00 IN07 1 x 6.0mm, 1 x 8.0mm and 1 x 3/8" 06.04.0047-00 IN08 1 x 6.0mm, 1 x 3.0mm and 2 x 1/4" 06.04.0048-00 IN09 Without hole, to be drilled by the client 06.04.0049-00 IN10 Others, under ordering 06.04.0050-00 IN1P 1 x 3mm, 1 x 6mm, 1 x 8mm, 1 x 1/4" 06.04.0125-00 IN1A 1 x 1/8", 1 x 3/16", 2 x 1/4", 1 x 3/8" 06.04.0126-00 IN1E 1 x 4mm, 1 x 6mm, 2 x 8mm, 1 x 10mm, 1 x 1/4" 06.04.0127-00 BB Black Body Insert 06.04.0072-00

AG Stirred Liquid Kit: Stainless-steel cuplike insert, magnetic stirrer, sensor guide and support.

06.09.0029-00

presys

Page 10: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 6

Description Order Code Soft Carrying Case for TA-NL models 06.01.1032-00 Insert Extractor 02.06.0085-20 Lead Cable Kit 06.07.0018-00

Power Cable Type J – Brazil

01.14.0008-21

Power Cable Type B – US

01.14.0100-21

Power Cable Type F – Europe Universal

01.14.0089-21

Power Cable Type J – UK

01.14.0117-21

Accessories included:

- 01 x Soft Carrying Case; - 01 x Insert Extractor; - 01 x Insert; - 01 x Power Cable;

- 01 x Lead Cable Kit; - 01 x Technical Manual; - 01 x Factory Calibration

Certificate.

Note: When asked, the calibration certificate will be provided for the first insert ordered.

Dry Block Insert

Stirred Liquid Kit Insert (Optional)

Order Code: AG Kit Includes: magnetic stirrer, sensor guide,

support kit for sensors, specific kit for glass thermometers and carrying case. For use with two or more different fluids, it is recommended to purchase a kit for each fluid. This reduces

the time spent with fluid exchange and kit cleaning.

Black Body Insert (Optional) Order Code: BB

Includes: Insert type jug completely anodized - wall and special bottom in black, creating a cavity of high-emissivity blackbody for calibration of infrared thermometers.

presys

Page 11: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 2

1.4 - Parts Identification

Fig. 01 - Parts Identification

presys

Page 12: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 3

1.5 - Instructions for Optional

Stirred Liquid Kit

Fig. 02 - Uses of Stirred Liquid Kit

presys

Page 13: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 4

Thermal Fluids suitable for use with the insert (AG)

Low viscosity and safe operating temperature are the most important characteristics when choosing a heat transfer fluid to produce high homogeneity in the bath. How less viscous, how better the fluid circulation will be and homogeneous heat produced in the middle. We recommend using the following heat transfer fluids for models TA-25NL and TA-45NL: Viscosity Fluid Flash point* Usable range

Viscosity increases

ethyl alcohol 16 °C -50 °C to 15 °C Mixture of ethylene glycol and water (50% of each)

------------- -30 °C to 90 °C

Water ------------- 5 °C to 90 °C Silicone oil type 200.05 133 °C -40 °C to 130 °C

Silicone oil type 200.10 211 °C -30 °C to 209 °C

*Flash point is the temperature at which a vapor will ignite if exposed to a flame. When the flame is removed, the vapor stops burning.

At temperatures near the lower end of the range, the thermal fluid becomes very

viscous. It can disturb the circulation and increase the cooling time. Above the upper end, there is high evaporation of the fluid compromising the stability of the thermal medium.

If you use the insert with stirred liquid it is necessary to insert the magnetic stirrer (cross shape) at the bottom of the provided insert. The speed of the stirrer is regulated through the STIRRER option in Calibrator mode. Press STIRRER to turn on and off the stirrer, and select the speed between Low, Medium and High options in - and + buttons.

A good fit ensures good uniformity for various types of fluids: alcohol (low temperature), silicone oil (high temperature) etc. Be careful not to accelerate too much the motor so that the magnetic stirrer stops rotating at the bottom of the insert. This will worsen the good homogeneity of the calibrator.

To use the insert with the stirred liquid a good practice is to place first the sheath of the probes or thermometers to be calibrated and then fill up the fluid to a level around 2.5 cm below the top of the insert. This ensures that no overflow occurs either by the introduction of sensors, or by thermal expansion of the fluid. If the level starts to fall during use due to evaporation of the fluid, fill up again adding more fluid. Thus the minimum immersion will not be compromised.

For more accurate measurements use a calibrated probe placed together with the sensors to be calibrated and connected to the external probe input. The calibrator will now control the temperature of the block using this temperature as reference (see item 2.1.2 - External Probe Connection). It is also possible to read the indication of this probe in the RTD Pt-100 auxiliary input.

presys

Page 14: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 5

Black Body Insert (BB) Placing the insert BB inside the well of the block of the calibrator provides it with a

good black-body cavity with emissivity above 0.95, suitable for calibration of infrared thermometer.

For more accurate measurements use a calibrated probe placed in the side hole of the bath and connected to the external probe input. The calibrator will then control the block temperature using this temperature as a reference. It is also possible to read the indication of this probe in the Pt-100 RTD auxiliary input.

Align the infrared thermometer to be calibrated with a black body cavity of the calibrator in a vertical position.

The distance of the thermometer to be calibrated in relation to the bottom of the blackbody cavity should comply with specification of its technical manual.

Fig. 03 - Schematic View of the Black Body Insert Assembly

presys

Page 15: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 6

2.0 - Calibrator Operation

When powered on, the calibrator goes through a self-test routine and shows the

last adjustment date and it is ready to operate.

After the self-test is completed, the display shows the main menu:

Fig. 04 - Main Menu

The main menu is divided in 06 functions: CALIBRATOR - selects the probe and input functions, see section 2.1 HART® - optional module that allows communication with devices that have HART® protocol, see section 2.2. TASKS - performs calibrations automatically, see section 2.3. DATA LOGGER - record measurements, enabling visualization in chart or table, see section 2.4. VIDEOS - help videos to assist the user, see section 2.5. SETTINGS - general instrument settings, see section 2.6. pr

esys

Page 16: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 7

2.1 - Calibrator Menu

To select the probe set point or electrical input functions, from the main menu, press the CALIBRATOR button. The following screen is displayed.

Fig. 05 - Calibrator Function

At the top is shown the probe settings and values. The centered value shows the block temperature. The GREEN color indicates that

the temperature is stable, otherwise it is RED. The set point value appears on the top. Touch in the SET bar to change it. Pressing on the temperature unit it can be changed between °C (Celsius), °F

(Fahrenheit) or K (Kelvin).

Fig. 06 - Calibrator Mode

presys

Page 17: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 8

In the STEP function, a step value can be configured, and the steps can be changed through the up and down arrows.

In REFERENCE menu, you can configure the type of probe reference (see section 2.1.1 – Probe Reference). The chosen reference appears just below the REFERENCE button.

At the bottom, an input can be configured. When the input is selected, the screen will split automatically. To select an input, just touch the INPUT bar (see section 2.1.2 - MENU INPUT).

The icon shows a Quick Navigator, with the options for Main Menu (HOME), Data-Logger and Tasks. Pressing MENU, there is an option for the Memory Manager (see section 2.1.4). Furthermore, it brings information about the probe/auxiliary input configuration and IP address. Press BACK to return to Calibrator Mode or HOME to go to Main Menu.

Fig. 07 - Quick Navigator and Secondary Menu

presys

Page 18: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 9

2.1.1 - Probe Settings

There are two different references to control the thermal block: Internal

Reference and External Reference. The Internal Reference is a sensor built into the block. The External Reference is an option for more accurate measurements. The

control reference comes from a Standard Sensor placed inside the insert, among the DUT (devices under test). This Standard Sensor, with Callendar-Van Dusen parameters, eliminates adjustment errors and block loading effects.

It can be used both in control and measurement mode. When used in measurement mode, the probe indication is displayed on the screen and the control is made by the internal probe.

Fig. 08 - Choosing the Type of Temperature Reference

To select the Reference between Internal and External, touch the REFERENCE

bar. When selecting External Reference, the Callendar-Van Dusen parameters must be set. It is also necessary to choose if the External Probe is going to control the block or not.

ID: Set an identification for the sensor R0 (): The last resistance measurement in 0 °C of the sensor A, B, C: Callendar-Van Dusen coefficients Low (°C): Lower value of the reference sensor calibrated range High (°C): Higher value of the reference sensor calibrated range The coefficient values can be found in the Reference Sensor Certificate. pr

esys

Page 19: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 10

Fig. 09 - Adding a new Reference Sensor

After filling the blanks, click on SAVE button and confirm. The new sensor is now available to be chosen in the list. To edit data from a sensor, select it and press MANAGER button. To remove a sensor, select it and press REMOVE

Fig. 10 - Connecting the Standard Sensor for the External Reference

Note: the values corresponding to controlled temperatures appear in GREEN / RED. Values that show only the sensor indication appear in BLACK. pr

esys

Page 20: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 11

2.1.2 - Input Settings

The INPUT menu has the following options:

Fig. 11 - Input Menu Options

For OHM measurement, you should also select between 2, 3 or 4 wires options. For RTD input, it should be chosen the type between Pt-100, Pt-1000, Cu-10 or

Ni-100 (standard table used), the number of lead wires (2, 3 or 4 wires) and the temperature scale (ITS-90 or IPTS-68). There is also the option to configure the Callendar-Van Dusen coefficients of the sensor, selecting the option CVD and the desired curve in the list.

Fig. 12 - Options for RTD Input

To create new CVD coefficients, press (edit), and ADD button. The curves appear in the created list as identified in ID.

For TC (thermocouple), you must select the thermocouple type and the type of cold junction compensation (CJC): Internal or Manual. In Internal option, the

presys

Page 21: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 12

compensation is done internally; In Manual you must provide the value of the temperature of the cold junction to the calibrator.

The option SWITCH has two ways to be used. When entering the SWITCH function, the input works as a continuity measurement between RTD2 and RTD4 terminals. When there is continuity, the input shows CLOSED, if not, indicates OPEN. The input also records the temperature value of the block at the time of contact opening / closure.

Using the option THERMOSWITCH TEST, the calibrator performs cycles registering the thermoswitch opening and closure interactively, in order to find the setpoint temperature of the thermoswitch and its respective hysteresis. In Setpoint Hi set a temperature above the opening of the thermoswitch contact. In Setpoint Lo, use a value below the setpoint discounted hysteresis. E.g.: To test a thermostat of 50 °C setpoint and 5 °C hysteresis, Setpoint HI can be set to 55 °C and Setpoint LO to 45 °C.

Fig. 13 - Switch Auto Test Parameters

It is important that the number of cycles be at least 3. By selecting this amount you can check the repeatability of the thermoswitch. For the accuracy, when choosing higher accuracy levels, the temperature ramp times will be higher too. If a Report is needed for this test, use the TASK function.

The NONE option turns off the input function. When the input sensor breaking occur (RTD, resistance or probe) the display will

show the burn-out warning identified by question marks illustrated below:

Fig. 14 - Burn-out Warning

If an out of range signal is injected, a message of UNDER or OVER range appears.

presys

Page 22: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 13

2.1.2.1 - Input Connections Diagrams

Fig. 15 - Input Connections presys

Page 23: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 14

2.1.3 - Special Function

SCALE: For the current input, it is possible to use the scale function:

Fig. 16 - Option for mA Input: SCALE

It establishes a linear relationship between the mA input signal and what is shown

at the display, according to the graphic below:

Fig. 17 - SCALE Function (LINEAR)

The scaled indication at the display (#) may represent any engineering unit, such as: °C, % etc.

The number of decimals, up to 4, shown at the display may be configured. The value for Input High must be necessarily higher than Input Low. On the

other hand, Scale High and Scale Low may have any relationship between themselves: higher than, lower than or equal to, and they may have a signal before them. Thus direct or reverse relationships may be established. pr

esys

Page 24: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 15

Fig. 18 - Scale Function Configuration

Note: To enable the Scale Function, turn the function ON before touching

OK button. To disable, turn the function OFF.

2.1.4 - Saving Current Configuration (Memory Manager)

The TA Series calibrators admit several special functions that may become of frequent use. In these situations, it is useful to store such settings in the instrument in order to save time. After setting the desired calibration mode (input type, probe reference or special

function), press the icon > MENU, and the button MEMORY MANAGER. On the option CREATE NEW can be given a name for this configuration and a description. Press the SAVE button. The operation that was being performed by the TA Calibrator shall be stored in memory identified by the name given to it. To use it again, even after the calibrator is turned off and on, select the name of the desired setting and press the LOAD button. The SAVE AS DEFAULT button sets the current configuration as the default configuration of the calibrator. Thus, every time the calibrator is turned on, this will be the initial configuration of the calibrator.

pr

esys

Page 25: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 16

2.2 - HART® Configuration The TA Series Calibrators can be used to read and set parameters in devices that

have HART® Communication Protocol. The HART® Protocol allows digital communication between master (in this case, the TA Calibrator) and the slave (field instrument) superimposed on the 4-20 mA analog signal. To access this function from the main menu, select the HART® option.

The HART® Communication of TA Calibrators is an optional module. The calibrator has three versions: NH (without HART® Communication), CH (HART® Calibrator) and FH (Full-HART® configurator with DD library).

The CH option has basic and universal commands for HART® communication (zero, span, trim mA etc.) that allow you to adjust the range of the instrument, monitoring the primary variable, current adjustment etc. The FH option, in addition to basic and universal commands, is provided with the DD library (Device Description) from FieldComm Group and allows the setting of specific parameters of each instrument.

The following description is valid for CH and FH options.

2.2.1 - HART® Connections

To the connections shown in Figures 19 and 20, use the mA INPUT + HART®

option and INTERNAL RESISTOR enabled. In this mode, the 250 resistor is activated internally in series with the calibrator mA input. The calibrator can measure current from the transmitter and also read and set parameters via HART®. If the internal resistor is not enabled, an external resistor of at least 150 Ω must be inserted in series with the mA input. To power the transmitter, can be used TPS source (Fig. 19) or an external source (Fig. 20).

Fig. 19 - Transmitter Powered by the Calibrator Itself (TPS)

mA INPUT + HART® (Internal Resistor Enabled)

presys

Page 26: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 17

Fig. 20 - Transmitter Powered by an External Power Supply

mA INPUT + HART® (Internal Resistor Enabled)

2.2.2 - Starting Communication

After defining HART® connection type, inform the ADDRESS of the HART® device

and press the CONNECT button. If the instrument address is unknown, you can use SEARCH button to search the device address in the range from 0 to 15.

Up to 15 devices are allowed on HART® network (address 1-15). In connection with a single field instrument with poll address 0 and mA INPUT + HART® connection, the primary variable can be read either in analog (4-20 mA) and digital form (HART®).

When connected, data appear in the DEVICE INFO tab, identifying the instrument such as TAG, manufacturer, description, message, date, measuring range and input filter (damping). Some of these parameters can be changed in the DEFAULT SETTINGS. 2.2.3 - Adjusting the Measurement Range of a HART® Transmitter

In DEVICE INFO tab, the MIN and MAX fields indicate the measuring range of the

HART® transmitter. For PV (primary variable) equal to the MIN value, the transmitter should generate 4 mA. For PV (primary variable) equal to MAX value, the transmitter should generate 20 mA. The maximum allowable range of the transmitter is shown just above (RANGE ...). To edit the range of the transmitter, just change the MIN and MAX values and press the SAVE RANGE button.

On this screen you can also edit the unit of the primary variable and the input filter (damping). pr

esys

Page 27: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 18

Fig. 21 - Adjusting the measuring range of the HART® transmitter

2.2.4 - Adjusting the Measurement Range of a HART® Transmitter with Reference

Another way to adjust the range of the transmitter is generating the minimum and maximum values of the desired range in the transmitter input and adjusting these values as minimum and maximum (set by reference). To adjust the range of a temperature transmitter, insert the transmitter in the thermal block, choose the Reference, and set the temperature. Select Input mA and press the HART® button. The temperature generation will work as the standard value for the adjustment range of the instrument.

Fig 22 - Quick HART® Adjustment with Reference

Generate the temperature to the transmitter input corresponding to the lower range value wait the stabilization and press the button. Transmitter will generate 4 mA to this value. Generate the temperature to the transmitter input corresponding to the upper range value wait the stabilization and press button. Transmitter will generate 20 mA for this value.

presys

Page 28: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 19

Another way to do this adjust is entering in the HART® option through the MAIN MENU, set the connection type, address and then press CONNECT. Select the MONITOR tab. In this screen are shown the value of the primary variable (PV) read by HART® (digital), the current that the transmitter wants to generate (AO - DIGITAL OUTPUT), and the current measured by the TA Calibrator (ANALOGIC READ). Set the block temperature pressing OUTPUT and adjust the transmitter range pressing the ↓ Inf Range and ↑ Sup. Range buttons.

Fig. 23 - Adjusting the Measuring Range of the HART® Transmitter with Reference

2.2.5 - Checking / Adjusting HART® Transmitter mA Output

In DEFAULT SETTINGS, the output current of the HART® transmitter (output

trim) can be adjusted according to current measured by the TA Calibrator. You can make this adjustment only when the TA Calibrator is connected to a single transmitter with address 0, in the mA INPUT + HART® option of connection, since the calibrator needs to measure the current to make the adjustment.

Before performing the adjustment, a transmitter output current check can be performed by pressing the CHECK button. The transmitter will generate fixed current (4, 8, 12, 16, 20 mA) and the calibrator will show the measured values for each point.

To adjust automatically, simply press the AUTO button. The calibrator will send the command to the transmitter to generate 4 and 20 mA (fix), makes the measurement of these points, and adjusts the output (trim). The adjustment is completed when D/A Adjustment Completed message appears.

The LEAD TIME sets each points stabilization time (in seconds). pr

esys

Page 29: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 20

Fig. 24 - Checking / Adjusting the mA HART® Transmitter Output

2.3 - Automatic Tasks In TA Calibrators, automatic calibration tasks can be created and performed. This

option can be used to generate calibration work orders for sensors, transmitters and indicators.

2.3.1 - Creating Tasks

To create tasks from the main menu, select CALIBRATOR. Select the desired type of input and probe configuration. For example, to calibrate a temperature transmitter, select the probe configuration (Internal or External Reference) and mA input (which will be connected to the current output of the transmitter). For a temperature indicator, e.g., selected NONE for the input.

Press the icon, and select TASKS and CREATE NEW TASK. Fill at least the serial number of the instrument to calibrate, instrument TAG, lead

time (time, in seconds, for the complete stabilization of the system), maximum error allowed for the instrument to be calibrated (in % of the span, reading or full scale), calibration range (highlighted fields). For the SWITCH tasks, the error fields are filled in the next step.

Fig. 25 - Task Information

presys

Page 30: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 21

Go to the As Found/ As Left tab. Add each point to be generated by the TA Calibrator and the expected value for the UUT (unit under test) both As found (calibration done before adjustment) and As left (calibration done after adjustment) values. Points can also be generated with the help of AUTO button. By pressing this button, simply enter the maximum and minimum values of the calibration and the number of points that need to be generated considering the same steps and linear scale. Also fill the number of repetitions (REP) of the readings, and the calibration strategy (initial to the final point ↑, final to the initial point ↓, etc.). If 0 value is set in as found repetition, the task will execute only As-Left calibration.

Fig 26 - Task Points and Strategy

For SWITCH tasks, the screen is different, as shown in the figure below. It must be filled the TRIP POINT of the thermoswitch and its DEADBAND, as well as their respective error. The TIME RAMP is the time in seconds that the calibrator will take to tour the range and find the value of opening and closing the thermoswitch. The minimum value for this field is 300 s. Tip: If the Trip point and deadband are not known, try the THERMOSWITCH TEST to find an approximate value before creating the task.

Fig 27 – SWITCH Task Parameters

presys

Page 31: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 22

Go to the Review and Save tab and choose an identification name/number for your task. If you want to save the model of this task for later use to create other tasks, press SAVE TEMPLATE and give a name for it. When you want to open this model again, open the task creation screen and press OPEN TEMPLATE in Task info tab.

Click on CREATE button to create it. The task is now saved in the calibrator.

Fig 28 - Creating a Task

2.3.2 - Performing Tasks

To perform a task created from the main menu select TASKS >

EXPLORE TASKS. A list of the created W.O. that have not been performed yet (● WAITING) is shown. Select the desired task and press OK. Make the necessary connections between the calibrator and the UUT and press START.

Fig 29 - Exploring Tasks

presys

Page 32: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 23

The TA Calibrator automatically starts the execution of the calibration, generating the registered setpoints, waiting the stabilization of the system and reading the associated input values. If you selected the option NONE as input, for each generated point, the calibrator will require to digit the value of the temperature indicated by the instrument (thermometer, indicator…). The result will be displayed on the screen, and a progress bar indicates the calibration remaining time. At the end of the calibration, a report is showed with the generated values, the obtained values, the expected values, and the error. If the error is higher than the registered value for the task, the line appears in red. The first time that a task is performed, it will be saved as As-found (before adjustment). If it runs again, it will be saved as As-left (after adjustment). The results are saved in the calibrator and can be viewed at any time.

2.3.3 - Viewing Results

Once a task has been performed, it remains saved in the calibrator.

To view the results of a calibration by the calibrator, in the main menu select TASKS.

Enable the option ● PERFORMED. The list will show only the tasks that have been performed. Select the desired work order and press OK. On screen, the report with the calibration points, the obtained values, expected values and the errors will be shown. If the error is higher than the value registered for the task, the line appears in red.

Fig. 30 - Task Results

The task data is saved in a PDF file in the internal memory card of the Calibrator and can be accessed connecting the calibrator to a computer. To save the data in a

PenDrive or External HD on the host USB, press the PenDrive icon .

presys

Page 33: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 24

To print the Calibration Report, press the printer icon . The printer must have been previously configured in SETTINGS > SYSTEM > PRINTER CONFIG.

The Report logo can be changed for one of your own company. For this, connect a USB cable between TA Calibrator USB Device port and a computer. Change the file LOGO.bmp for your logo file (must be .bmp extension). We recommend an image close to 200 x 200 pixels.

Fig. 31 - Example of a Printed Calibration Report

2.4 - Data-Logger The TA Series Calibrators allow you to record series of measurements over time to display data in chart or table format. Select CALIBRATOR from the main menu and select the desired configuration for Probe and Input.

Press the icon and select DATA LOGGER. The calibrator automatically starts the measurements and displays each

measured point on the chart. For measurements to be saved, you must press the REC button (see Figure 32).

With this option selected, all points (measurement and time) are saved in an internal file in TA Calibrator, which can be used to generate a table or chart.

presys

Page 34: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 25

Fig. 32 - Data Logger

In configuration menu (icon , you can edit the background color of the chart, color and line thickness, sampling rate (in seconds) and set the x (time) and y (measurements) axis of the chart.

Fig. 33 - Data-Logger Configuration Menu

Recording can also be programmed to start at a certain date and time in the LOGGER option. Just set the start time and end time of recording. During the defined range, the measured points are saved in an internal file.

To view a saved file press the OPEN button, select the desired file, and press LOAD. The file name contains the date and time of the measurements.

The SHEET button allows the visualization of data in table format, with the date and time of the measurement and the measured values.

If the user wants to export the current data to a .csv file that can be opened in spreadsheet softwares, press the SAVE button and indicate the name and location where it will be saved. The button saves the current screen image as a .png file. All saved screens can be viewed in the IMAGE menu. These files are saved in the internal SD card of the calibrator. To access the files saved on the TA Calibrator, connect the USB cable to the computer (type A USB) and to the TA Calibrator (Micro-B USB, see Figure 1).

presys

Page 35: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 26

2.5 - Videos TA Series calibrators have a video player. These videos can be viewed while a

calibration is performed and are used to help operator to learn about some specific internal procedures.

From the main menu, selecting VIDEOS a list of video categories appears. Select the category and the desired video. Press the FULL SCREEN button to view the video in full screen and the WINDOW button to reduced screen.

To add new videos on the calibrator, connect the USB cable to the computer (type A USB) and to the TA Calibrator (Micro-B USB, see Figure 1). Open VIDEOS folder. Copy the new video to any sub-folder (category) of the VIDEOS folder. If you prefer to create a new category, simply create a new folder inside VIDEOS with the name of the desired category and copy the video to this folder.

2.6 - Settings

2.6.1 - System

In the SYSTEM tab, you can set the volume of the calibrator, the touchscreen calibration, identification of the calibrator, language, printer and security options.

Touchscreen Options

To adjust the touchscreen, press the TOUCHSCREEN OPTIONS button. Press on the screen the places indicated by + (it is recommended to use the stylus for touchscreen). After the calibration, press again on the screen at any point. Confirm the calibration to return to SYSTEM Menu.

Brightness

Select the display brightness. The options are 25%, 50%, 75% and 100%.

Language Setting Press the desired language for the system and confirm in OK button. The system

must be restarted to apply the configuration.

Calibrator Identification In this option is possible to identify your calibrator, choosing a TAG name, Owner

name and Location.

Sound Options Press + or - to configure a value for the system audio volume.

Printer Config. Select the Configuration for the Printer and connect it to the USB port.

presys

Page 36: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 27

Security Options The instrument initially has no access password. This setting can be changed in SECURITY OPTIONS.

To create a new user, press the key icon and then users icon . Fill the blanks and press CREATE. It is possible to add a signature that will be used on the calibration report. Pay attention to the functions that each user level has access in the table below:

User Level Function Calibrator Tasks HART® Data-Logger Settings

Operator Tec

Admin

To lock the system, press the padlock icon on system menu. The next time the TA Calibrator is turned on, it will request login and password. To unlock the system, login as an Admin Level user and press the padlock icon on system menu again.

Adjust Cal. Adjustment level protected by password. See section 5.0 - Calibration (Adjustment) for more information.

2.6.2 - Network

In NETWORK tab you can configure the IP address for the Ethernet communication with the computer. The IP address can be dynamically configured (DHCP) or may have a fixed address (disable DHCP option and edit the desired address). By connecting the calibrator to the network you can view and print reports of the tasks on the computer. 2.6.3 - Built-in Web Server

Connect the network cable into the Ethernet port of TA Calibrator on the side (see Fig. 1).

To access the built-in webserver open the web browser on your computer and enter the following address.

<calibrator_IP_address>:5000/taserver/pages/main.cgi User: admin Password: xvmaster

presys

Page 37: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 28

To verify the IP address press the button indicated below.

Fig. 34 - IP address

Fig. 35 - TA Calibrator Web Server In the Web Server, you can monitor the calibrator screen, change the setpoint and see the auxiliary input readings.

presys

Page 38: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 29

3 - Safety Instructions

If the calibrator is turned on, do not leave the room without an identification or warning

about the high temperature hazard. Before turning the calibrator off, return the block temperature to values close to the

ambient temperature. Never remove the insert from the dry block or the thermo-elements from the insert,

while they are in temperatures far from the ambient. Wait until they reach the ambient temperature so that the heterogeneous cooling of the parts do not jam each other.

4 - Recommendations as regards Accuracy of Measurements

PRESYS Temperature Advanced Calibrators are instruments of high accuracy

level, requiring the observation of all the procedures described in this section, in order to achieve the necessary conditions to get the accuracy levels during the calibrations. Special attention should be paid in relation to the insert cleanliness. When necessary,

it should always be washed with water and soap, well rinsed and dried. Oil, grease, solid particles can hinder the heat transference to the insert and even jam the insert inside the block.

The sensor to be calibrated must fit snugly into the appropriate well. In case the sensor is loose, the measurement accuracy meaning can be completely senseless. The meaning of clearance between the sensor and the respective well should be understood in a subjective way and the common sense is very important. Thus, the sensor should enter the insert well (both completely clean) in such a way to stay snugly enough so that it cannot move or swing inside but it should not enter by force to get jammed.

In the specific case of generating negative temperature, the calibrations should be performed in a descending temperature order. This procedure is useful because in negative temperatures ice is formed in the insert surface and between the insert and the thermo-element being calibrated. This humidity changes the thermal contact and cause error in the calibration. After finishing the use of the insert below 0 C, increase the set point to a positive temperature value, remove the insert from the block and the sensor and dry completely all these parts before continuing the calibration. This procedure guarantees accuracy next to 0.1 C. In case you can accept higher values of accuracy, like 0.2 C or more, these cares can be left apart.

presys

Page 39: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 30

5 - Calibration (Adjustment)

Warning: Enter the following options only after understanding them completely.

Otherwise, it may be necessary to return the instrument to the factory for recalibration!

Select ADJUST/CAL option from the SETTINGS > SYSTEM menu. You should then enter the password 9875 to access the calibration menu.

The password functions as a protection to calibration ranges. After the password is entered, the menu displays the options GENERAL, INPUTS and PROBE.

Options for INPUTS are mV, mA, ohm and thermocouple.

5.1 - Input Calibration

Select the corresponding mnemonic and apply the signals presented in the tables below.

Note that the applied signals just need to be close to the values shown in the table.

Once the signal has been applied, store the values of the calibration points 1 and 2. Press SAVE to save the typed values

mV Input Point 1 Point 2 G4 0.000 mV 70.000 mV G3 0.000 mV 120.000 mV G2 0.000 mV 600.000 mV G1 600.000 mV 2400.000 mV

mA Input Point 1 Point 2 Single range 0.0000 mA 20.0000 mA

Input calibration for is performed in two steps: a) Application of mV signal:

For the calibration below, leave terminals RTD3 (+) and RTD4 (+) short-circuited.

mV Signal Terminals Point 1 Point 2 V_OHM3 RTD3(+) and mV(-) 90.000 mV 120.000 mV V_OHM4 RTD4(+) and mV(-) 90.000 mV 120.000 mV

b) Application of standard resistors: Connect a decade box or standard resistors on terminals RTD1, RTD2, RTD3 and RTD4 (4-wire connection).

presys

Page 40: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 31

resistors Point 1 Point 2 OHM3 20.000 50.000 OHM2 100.000 500.000 OHM1 500.000 2200.000

The cold junction calibration (Thermocouple) is performed measuring the mV(-) terminal temperature. Store only the point 1.

Cold Junction Point 1 CJC 32.03 ºC

(measured value)

presys

Page 41: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 32

5.2 - Probe Calibration

To readjust the internal Probe it is necessary to compare the value indicated by the calibrator (Probe) and the temperature value from a standard probe placed in the dry block insert. The temperature of the standard probe should have high accuracy.

The option to adjust the internal sensor has seven points of adjustment. These points are recorded via points 1 to 7.

Before starting the calibration (adjustment), record in these points the respective initial storing values, according to the table below:

For the TA-25NL

Calibration Point

Initial value to record (°C)

Standard indication

New value to record

New indication

of the Standard

Point 1: -25 °C -25.00 -24.780 -24.78 -24.995 Point 2: -2 °C -2.00 -2.103 -2.10 -2.005 Point 3: 30 °C 30.00 29.910 29.91 29.990 Point 4: 60 °C 60.00 59.771 59.77 60.009 Point 5: 90 °C 80.00 89.770 89.77 90.000 Point 6: 120 °C 120.00 119.630 119.63 119.995 Point 7: 140 °C 140.00 139.539 139.54 140.005

For the TA-45NL

Calibration Point

Initial value to record (°C)

Standard indication

New value to record

New indication

of the Standard

Point 1: -45 °C -45.00 -44.780 -44.78 -44.995 Point 2: -2 °C -2.00 -2.103 -2.10 -2.005 Point 3: 30 °C 30.00 29.910 29.91 29.990 Point 4: 60 °C 60.00 59.771 59.77 60.009 Point 5: 90 °C 80.00 89.770 89.77 90.000 Point 6: 120 °C 120.00 119.630 119.63 119.995 Point 7: 140 °C 140.00 139.539 139.54 140.005

Select the calibration point and then press CHANGE TEMPERATURE. Wait for the complete stabilization of the point. On the field Adjusted Point, write the value presented in the standard thermometer and confirm in SAVE button. Go to the next point and continue the adjustment until the last point. Note: It is recommended to start the adjustment from the highest temperature

point (point 7: 140 ° C) so that the humidity of the air does not interfere in the

results.

presys

Page 42: Temperature Advanced Calibrators TA-25NL, TA-45NL · Temperature Advanced Calibrators . TA-25NL . and . TA-45NL. generate temperature in the insert in order to calibrate thermocouples,

PRESYS Instruments TA-25NL / TA-45NL

Page 33

6.0 - Maintenance

6.1 - Instructions for Hardware Maintenance There are no parts or components inside the temperature calibrator that can be repaired by the user. Only the 10 Amp fuse for 115V models or 6 Amp fuse for 230 V models, placed within the socket on the rear can be replaced in case of blow.

The fuse may blow due to a voltage spike in the mains or a calibrator component fault. Replace the fuse once. If a second fuse blows again, it is because the fault is not that simple. In this case, contact the Presys technical support. In case of malfunction of mA input, the input fuse (250 V/32 mA) can be exchanged.

6.2 - Instructions for Insert Jamming

If it happens that an insert jams inside the block proceed as follows. 1 - apply a lubricant oil between the parts; 2 - apply cooling liquid inside the insert wells in order to contract the insert; 3 - try again to withdraw the insert. After taking the insert out, sand both the surfaces with thin sandpaper, polish

them with an suitable paste and finally clean the parts using alcohol or solvent.