TECHNICAL REPORT ITAK
RT-138-2020 PEP
PROFICIENCY TEST FOR NICKEL ORE ANALYSIS
32nd ROUND – AUGUST 2020
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 2 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
1) Introduction
This Technical Report presents the obtained results, its evaluation and interpretation for
the participant laboratory.
The description of the statistical criteria and techniques used in the evaluation as well as
general information about this proficiency test is in a separate module and available to
each participant in the ITAK PTP Management System.
ITAK does not issue preliminary report. This current report is characterized as Final
Technical Report.
2) Preliminary Information
Tab. 1: Round information and participant laboratory identification
Round Identification: 32nd Round – PT in Nickel Ore Analysis
Period of execution: June and July 2020
Laboratory Identification: PT Vale Indonesia Tbk
Unit/Site: Sorowako – Indonesia
Laboratory Code: 06
Test Item Codes for this round: Ni57 and Ni58
Issue Data of this Report: August 07th of 2020
3) Participants of the Proficiency Testing
The participating laboratories of this inter-comparison PT are listed below:
Participating Laboratory Name Unit/Site
Anglo American Brasil Ltda – Barro Alto Barro Alto/GO – Brazil
Anglo American Brasil Ltda – Niquelândia Niquelândia/GO – Brazil
ITAK – Instituto de Tecnologia August Kekulé Ltda João Monlevade/MG – Brazil
Nilab Pouembout – New Caledonia
PT Carsurin Kendari Sulawesi Tenggara – Indonesia
PT Geoservices Geoassay Laboratory Bekasi City – Indonesia
PT Vale Indonesia Tbk Sorowako – Indonesia
Vale – Unidade Operacional Onça Puma Ourilândia do Norte/PA – Brazil
4) Appeal
It is afforded to the participant of Proficiency Testing Programs - ITAK the right of appeal
as to the evaluation of its performance given in this Technical Report. For that, the
participating laboratory who have doubts or does not agree with the performance
evaluation should register an appeal on the ITAK PTP Management System. According to
a requirement of ISO/IEC 17043 – Conformity assessment - General requirements for
Proficiency Testing, ITAK maintains a documented procedure governing all the systematic
including deadlines for appeal, analysis and response to participating client.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 3 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
5) Performance Commentary
Tab. 2: Performance evaluation of the participating laboratory – consolidated information.
Test
Item
Analyzed
Parameter
Assigned
Value
obtained by
Standard
Deviation
obtained by
Reference
Tables
Z Score
Performance
Youden
Performance
Ni57 Al2O3 Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard Deviation
ISO 5725-2
5 Unsatisfactory Unsatisfactory
Ni58 Al2O3 6 Satisfactory
Ni57 CaO Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard Deviation
ISO 5725-2
8 Unsatisfactory Insufficient
results Ni58 CaO 9 Satisfactory
Ni57 Co Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard Deviation
ISO 5725-2
11 Satisfactory Satisfactory
Ni58 Co 12 Satisfactory
Ni57 Cr2O3 Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard Deviation
ISO 5725-2
14 Unsatisfactory Unsatisfactory
Ni58 Cr2O3 15 Unsatisfactory
Ni57 Cu Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard Deviation
ISO 5725-2
17 Lab has not
participated
Lab has not
participated Ni58 Cu 18
Ni57 Fe Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard Deviation
ISO 5725-2
20 Unsatisfactory Unsatisfactory
Ni58 Fe 21 Satisfactory
Ni57 LOI Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard Deviation
ISO 5725-2
23 Lab has not
participated
Lab has not
participated Ni58 LOI 24
Ni57 MgO Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard Deviation
ISO 5725-2
26 Unsatisfactory Unsatisfactory
Ni58 MgO 27 Unsatisfactory
Ni57 MnO Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard Deviation
ISO 5725-2
29 Unsatisfactory Unsatisfactory
Ni58 MnO 30 Satisfactory
Ni57 Ni Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard Deviation
ISO 5725-2
32 Unsatisfactory Unsatisfactory
Ni58 Ni 33 Satisfactory
Ni57 SiO2 Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard Deviation
ISO 5725-2
35 Unsatisfactory Unsatisfactory
Ni58 SiO2 36 Satisfactory
Ni57 TiO2 Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard Deviation
ISO 5725-2
38 Lab has not
participated
Insufficient
results Ni58 TiO2 39
Ni57 Zn Mean of data set
means; Outliers
according to
ISO 5725-2
Reproducibity
Standard Deviation
ISO 5725-2
41 Lab has not
participated
Insufficient
results Ni58 Zn 42
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 4 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
6) Summary of Assigned Values, EP Standard Deviations and Uncertainties
Tab. 3: Assigned Values, EP Standard Deviations and Standard Uncertainties – Summarized
Information.
Parameter Test Item Unit Assigned Value
(XD)
Standard
Deviation
(S)
Standard
Uncertainty
(U)
Al2O3 Ni57 % 5.218 0.037 0.016
Ni58 % 3.711 0.068 0.030
CaO Ni57 % 0.4825 0.010 0.0046
Ni58 % 0.0094 0.0077 0.0043
Co Ni57 % 0.0567 0.0055 0.0024
Ni58 % 0.0292 0.014 0.0060
Cr2O3 Ni57 % 1.672 0.028 0.012
Ni58 % 4.384 0.079 0.035
Cu Ni57 % 0.163 0.021 0.010
Ni58 % 0.0184 0.0095 0.0045
Fe Ni57 % 15.154 0.10 0.046
Ni58 % 39.57 0.84 0.37
LOI Ni57 % 9.44 0.90 0.43
Ni58 % 9.27 0.75 0.36
MgO Ni57 % 8.007 0.14 0.060
Ni58 % 0.979 0.024 0.010
MnO Ni57 % 0.4924 0.0085 0.0038
Ni58 % 0.4389 0.013 0.0058
Ni Ni57 % 1.943 0.045 0.020
Ni58 % 0.9458 0.016 0.0069
SiO2 Ni57 % 50.06 0.26 0.12
Ni58 % 22.29 0.35 0.16
TiO2 Ni57 % 0.1416 0.0023 0.0015
Ni58 % 0.1239 0.013 0.0083
Zn Ni57 % 0.0401 0.0083 0.0052
Ni58 % 0.0267 0.0048 0.0030
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 5 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
7) Mandel’s Graphs
Pic. 1: Graph with Mandel’s h statistics for sample Ni57 for all the involved parameters.
Pic. 2: Graph with Mandel’s h statistics for sample Ni58 for all involved parameters.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 6 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 3: Graph with Mandel’s k statistics for sample Ni57 for all involved parameters.
Pic. 4: Graph with Mandel’s k statistics for sample Ni58 for all involved parameters.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 7 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
8) Results, Methods, Graphs and Interpretation.
Tab. 4 – Assigned Value, Standard Deviations and Standard Uncertainties for Al2O3.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard
Uncertainty
𝒖 (%) Sample
Ni57 5.218 0.037 0.016
Ni58 3.711 0.068 0.030
Tab. 5 – Results reported, average and Z Scores of laboratories participating in the PT Program
for Al2O3 analysis in the sample Ni57.
Results for Al2O3 – Sample Ni57 Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 Al2O3 5.170 5.180 5.170 5.180 5.170 % 5.174 -1.178 FUS-XRF Lab 02 Al2O3 5.230 5.210 5.260 5.240 5.240 % 5.236 0.497 FUS-XRF Lab 03 Al2O3 5.020 5.090 5.140 5.250 5.140 % 5.128 -2.422 FUS-XRF Lab 04 Al2O3 5.265 5.259 5.267 5.238 5.274 % 5.260 1.157 FUS-XRF Lab 05 Al2O3 5.236 5.224 5.236 5.205 5.232 % 5.227 0.243 FUS-XRF Lab 06 Al2O3 4.415 4.432 4.421 4.390 4.430 % 4.418 -21.622 FUS-XRF Lab 07 Al2O3 5.180 5.185 5.205 5.207 5.177 % 5.191 -0.724 FUS-XRF Lab 08 Al2O3 5.570 5.557 5.580 5.541 5.550 % 5.560 9.243 FUS-XRF
Tab. 6 – Results reported, average and Z Scores of laboratories participating in the PT Program
for Al2O3 analysis in the sample Ni58.
Results for Al2O3 – Sample Ni58 Code Parameter 1 2 3 4 5 Unit Average Z Score Method
Lab 01 Al2O3 3.650 3.660 3.660 3.650 3.650 % 3.654 -0.846 FUS-XRF Lab 02 Al2O3 3.800 3.800 3.810 3.820 3.800 % 3.806 1.394 FUS-XRF Lab 03 Al2O3 3.700 3.690 3.670 3.710 3.640 % 3.682 -0.433 FUS-XRF Lab 04 Al2O3 3.729 3.763 3.740 3.757 3.721 % 3.742 0.448 FUS-XRF Lab 05 Al2O3 3.699 3.716 3.694 3.696 3.698 % 3.701 -0.159 FUS-XRF Lab 06 Al2O3 3.603 3.626 3.647 3.597 3.630 % 3.621 -1.338 FUS-XRF Lab 07 Al2O3 3.790 3.768 3.767 3.779 3.769 % 3.775 0.931 FUS-XRF Lab 08 Al2O3 3.993 3.988 3.993 3.994 3.995 % 3.993 4.144 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 8 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 5: Graph with Z Scores for samples Ni57 and Ni58 for Al2O3 Analysis.
Laboratory 06 - PT Vale Indonesia Tbk presented UNSATISFACTORY performance for Al2O3
analysis in the sample Ni57 and SATISFACTORY performance for Al2O3 analysis in the sample
Ni58, according to the criterion for Z Score evaluating.
Pic. 6: Youden Plot for Al2O3 Analysis.
Performance Commentary – Confidence Ellipse:
The Confidence Ellipse obtained for Al2O3 analysis showed normal dispersion among the
participating laboratories. The laboratories 03, 06, 07 and 08 were excluded from the Confidence
Ellipse.
Laboratory 06 - PT Vale Indonesia Tbk showed an UNSATISFACTORY performance for Al2O3
analysis according to the criterion for Youden evaluation, because indicated possibility of
systematic negative errors being excluded from the Ellipse. It is recommended to check the
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 9 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
standards and calibration curves, the Certified Reference Materials used, the number of digits
used in cases of conversion / calculation and / or balance / weighing, and others factors that may
affect systematically the Al2O3 analysis.
Pic. 7: Ranking Graph according to the Mandel’s k statistic to sample Ni57 for Al2O3 Analysis.
Pic. 8: Ranking Graph according to the Mandel’s k statistic to sample Ni58 for Al2O3 Analysis.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 10 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Tab. 7 – Assigned Value, Standard Deviations and Standard Uncertainties for CaO.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard
Uncertainty
𝒖 (%) Sample
Ni57 0.4825 0.010 0.0046
Ni58 0.0094 0.0077 0.0043
Tab. 8 – Results reported, average and Z Scores of laboratories participating in the PT Program
for CaO analysis in the sample Ni57.
Results for CaO – Sample Ni57
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 CaO 0.500 0.480 0.500 0.500 0.490 % 0.494 1.099 FUS-XRF Lab 02 CaO 0.480 0.480 0.490 0.490 0.490 % 0.486 0.333 FUS-XRF Lab 03 CaO 0.470 0.480 0.490 0.480 0.490 % 0.482 -0.051 FUS-XRF Lab 04 CaO 0.490 0.544 0.488 0.485 0.487 % 0.498 1.521 FUS-XRF Lab 05 CaO 0.479 0.478 0.475 0.473 0.480 % 0.477 -0.530 FUS-XRF Lab 06 CaO 0.405 0.413 0.408 0.406 0.406 % 0.408 -7.179 FUS-XRF Lab 07 CaO 0.482 0.489 0.491 0.486 0.483 % 0.486 0.352 FUS-XRF Lab 08 CaO 0.470 0.460 0.470 0.465 0.465 % 0.466 -1.584 FUS-XRF
Tab. 9 – Results reported, average and Z Scores of laboratories participating in the PT Program
for CaO analysis in the sample Ni58.
Results for CaO – Sample Ni58
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 CaO 0.020 0.020 0.020 0.020 0.020 % 0.020 1.376 FUS-XRF Lab 02 CaO --- --- --- --- --- % --- --- FUS-XRF Lab 03 CaO 0.010 0.010 0.010 0.010 0.010 % 0.010 0.084 FUS-XRF Lab 04 CaO 0.007 0.004 0.009 0.007 0.007 % 0.006 -0.381 FUS-XRF Lab 05 CaO 0.004 0.005 0.005 0.004 0.005 % 0.005 -0.614 FUS-XRF Lab 06 CaO 0.003 0.003 0.002 0.003 0.003 % 0.003 -0.846 FUS-XRF Lab 07 CaO <0.01 <0.01 <0.01 <0.01 <0.01 % <0.01 --- FUS-XRF Lab 08 CaO --- --- --- --- --- % --- --- FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 11 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 9: Graph with Z Scores for samples Ni57 and Ni58 for CaO Analysis.
Laboratory 06 - PT Vale Indonesia Tbk presented UNSATISFACTORY performance for CaO
analysis in the sample Ni57 and SATISFACTORY performance for CaO analysis in the sample
Ni58, according to the criterion for Z Score evaluating.
Performance Commentary – Confidence Ellipse:
The valid results are not sufficient to build the Confidence Ellipse for the parameter CaO.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 12 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 10: Ranking Graph according to the Mandel’s k statistic to sample Ni57 for CaO Analysis.
Pic. 11: Ranking Graph according to the Mandel’s k statistic to sample Ni58 for CaO Analysis.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 13 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Tab. 10 – Assigned Value, Standard Deviations and Standard Uncertainties for Co.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard
Uncertainty
𝒖 (%) Sample
Ni57 0.0567 0.0055 0.0024
Ni58 0.0292 0.014 0.0060
Tab. 11 – Results reported, average and Z Scores of laboratories participating in the PT Program
for Co analysis in the sample Ni57.
Results for Co – Sample Ni57
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 Co 0.059 0.058 0.059 0.059 0.058 % 0.059 0.351 FUS-XRF Lab 02 Co 0.055 0.055 0.055 0.056 0.055 % 0.055 -0.269 FUS-XRF Lab 03 Co 0.066 0.067 0.067 0.067 0.067 % 0.067 1.844 FUS-XRF Lab 04 Co 0.058 0.059 0.060 0.059 0.059 % 0.059 0.405 FUS-XRF Lab 05 Co 0.058 0.058 0.058 0.058 0.058 % 0.058 0.241 FUS-XRF Lab 06 Co 0.048 0.050 0.047 0.047 0.048 % 0.048 -1.580 FUS-XRF Lab 07 Co 0.055 0.056 0.055 0.055 0.054 % 0.055 -0.305 FUS-XRF Lab 08 Co 0.053 0.052 0.053 0.053 0.053 % 0.053 -0.706 FUS-XRF
Tab. 12 – Results reported, average and Z Scores of laboratories participating in the PT Program
for Co analysis in the sample Ni58.
Results for Co – Sample Ni58
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 Co 0.029 0.029 0.029 0.028 0.029 % 0.029 -0.030 FUS-XRF Lab 02 Co 0.030 0.030 0.029 0.030 0.029 % 0.030 0.030 FUS-XRF Lab 03 Co 0.030 0.028 0.029 0.028 0.029 % 0.029 -0.030 FUS-XRF Lab 04 Co 0.019 0.019 0.020 0.020 0.019 % 0.019 -0.748 FUS-XRF Lab 05 Co 0.014 0.015 0.014 0.014 0.014 % 0.014 -1.111 FUS-XRF Lab 06 Co 0.016 0.016 0.018 0.016 0.018 % 0.017 -0.918 FUS-XRF Lab 07 Co 0.041 0.041 0.041 0.041 0.042 % 0.041 0.888 FUS-XRF Lab 08 Co 0.054 0.055 0.055 0.055 0.055 % 0.055 1.895 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 14 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 12: Graph with Z Scores for samples Ni57 and Ni58 for Co Analysis.
Laboratory 06 - PT Vale Indonesia Tbk presented SATISFACTORY performance for Co analysis in
the samples Ni57 and Ni58, according to the criterion for Z Score evaluating.
Pic. 13: Youden Plot for Co Analysis.
Performance Commentary – Confidence Ellipse:
The Confidence Ellipse obtained for Co analysis showed high dispersion among the participating
laboratories. None laboratory was excluded from the Confidence Ellipse.
Laboratory 06 - PT Vale Indonesia Tbk showed a SATISFACTORY performance for Co analysis
according to the criterion for Youden evaluation.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 15 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 14: Ranking Graph according to the Mandel’s k statistic to sample Ni57 for Co Analysis.
Pic. 15: Ranking Graph according to the Mandel’s k statistic to sample Ni58 for Co Analysis.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 16 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Tab. 13 – Assigned Value, Standard Deviations and Standard Uncertainties for Cr2O3.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard
Uncertainty
𝒖 (%) Sample
Ni57 1.672 0.028 0.012
Ni58 4.384 0.079 0.035
Tab. 14 – Results reported, average and Z Scores of laboratories participating in the PT Program
for Cr2O3 analysis in the sample Ni57.
Results for Cr2O3 – Sample Ni57
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 Cr2O3 1.660 1.660 1.660 1.690 1.680 % 1.670 -0.055 FUS-XRF Lab 02 Cr2O3 1.660 1.660 1.660 1.650 1.660 % 1.658 -0.483 FUS-XRF Lab 03 Cr2O3 1.620 1.670 1.670 1.670 1.680 % 1.662 -0.340 FUS-XRF Lab 04 Cr2O3 1.676 1.687 1.685 1.677 1.689 % 1.682 0.384 FUS-XRF Lab 05 Cr2O3 1.678 1.681 1.678 1.670 1.683 % 1.678 0.230 FUS-XRF Lab 06 Cr2O3 1.529 1.574 1.517 1.517 1.543 % 1.536 -4.834 FUS-XRF Lab 07 Cr2O3 1.619 1.642 1.627 1.638 1.639 % 1.633 -1.375 FUS-XRF Lab 08 Cr2O3 1.710 1.705 1.725 1.715 1.730 % 1.717 1.621 FUS-XRF
Tab. 15 – Results reported, average and Z Scores of laboratories participating in the PT Program
for Cr2O3 analysis in the sample Ni58.
Results for Cr2O3 – Sample Ni58
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 Cr2O3 4.310 4.320 4.320 4.320 4.330 % 4.320 -0.809 FUS-XRF Lab 02 Cr2O3 4.520 4.460 4.480 4.480 4.480 % 4.484 1.257 FUS-XRF Lab 03 Cr2O3 4.380 4.390 4.370 4.380 4.360 % 4.376 -0.103 FUS-XRF Lab 04 Cr2O3 4.402 4.420 4.462 4.413 4.361 % 4.411 0.343 FUS-XRF Lab 05 Cr2O3 4.272 4.342 4.315 4.338 4.293 % 4.312 -0.910 FUS-XRF Lab 06 Cr2O3 3.942 3.943 3.965 3.931 3.955 % 3.947 -5.506 FUS-XRF Lab 07 Cr2O3 4.290 4.314 4.309 4.296 4.318 % 4.305 -0.993 FUS-XRF Lab 08 Cr2O3 4.469 4.495 4.497 4.479 4.462 % 4.480 1.212 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 17 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 16: Graph with Z Scores for samples Ni57 and Ni58 for Cr2O3 Analysis.
Laboratory 06 - PT Vale Indonesia Tbk presented UNSATISFACTORY performance for Cr2O3
analysis in the samples Ni57 and Ni58, according to the criterion for Z Score evaluating.
Pic. 17: Youden Plot for Cr2O3 Analysis.
Performance Commentary – Confidence Ellipse:
The Confidence Ellipse obtained for Cr2O3 analysis showed normal dispersion among the
participating laboratories. The laboratories 02 and 06 were excluded from the Confidence
Ellipse.
Laboratory 06 - PT Vale Indonesia Tbk showed an UNSATISFACTORY performance for Cr2O3
analysis according to the criterion for Youden evaluation, because indicated possibility of
systematic negative errors being excluded from the Ellipse. It is recommended to check the
standards and calibration curves, the Certified Reference Materials used, the number of digits
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 18 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
used in cases of conversion / calculation and / or balance / weighing, and other factors that may
affect systematically the Cr2O3 analysis.
Pic. 18: Ranking Graph according to the Mandel’s k statistic to sample Ni57 for Cr2O3 Analysis.
Pic. 19: Ranking Graph according to the Mandel’s k statistic to sample Ni58 for Cr2O3 Analysis.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 19 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Tab. 16 – Assigned Value, Standard Deviations and Standard Uncertainties for Cu.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard
Uncertainty
𝒖 (%) Sample
Ni57 0.163 0.021 0.010
Ni58 0.0184 0.0095 0.0045
Tab. 17 – Results reported, average and Z Scores of laboratories participating in the PT Program
for Cu analysis in the sample Ni57.
Results for Cu – Sample Ni57
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 Cu 0.175 0.174 0.176 0.176 0.173 % 0.175 0.598 FUS-XRF Lab 02 Cu 0.177 0.176 0.177 0.176 0.177 % 0.177 0.683 FUS-XRF Lab 03 Cu 0.167 0.169 0.168 0.169 0.169 % 0.168 0.284 FUS-XRF Lab 04 Cu 0.163 0.163 0.180 0.159 0.166 % 0.166 0.167 FUS-XRF Lab 05 Cu 0.167 0.166 0.166 0.166 0.167 % 0.166 0.187 FUS-XRF Lab 06 Cu --- --- --- --- --- % --- --- --- Lab 07 Cu 0.125 0.128 0.127 0.128 0.125 % 0.127 -1.748 FUS-XRF Lab 08 Cu 0.075 0.074 0.075 0.075 0.075 % 0.075 -4.267 FUS-XRF
Tab. 18 – Results reported, average and Z Scores of laboratories participating in the PT Program
for Cu analysis in the sample Ni58.
Results for Cu – Sample Ni58
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 Cu 0.022 0.022 0.021 0.022 0.022 % 0.022 0.352 FUS-XRF Lab 02 Cu 0.032 0.032 0.032 0.032 0.032 % 0.032 1.429 FUS-XRF Lab 03 Cu 0.030 0.031 0.031 0.029 0.029 % 0.030 1.219 FUS-XRF Lab 04 Cu 0.034 0.043 0.040 0.039 0.033 % 0.038 2.017 FUS-XRF Lab 05 Cu 0.019 0.019 0.019 0.019 0.019 % 0.019 0.087 FUS-XRF Lab 06 Cu --- --- --- --- --- % --- --- --- Lab 07 Cu 0.007 0.007 0.007 0.007 0.008 % 0.007 -1.174 FUS-XRF Lab 08 Cu 0.012 0.012 0.012 0.012 0.012 % 0.012 -0.671 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 20 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 20: Graph with Z Scores for samples Ni57 and Ni58 for Cu Analysis.
Laboratory 06 - PT Vale Indonesia Tbk has not participated of the PT for the parameter Cu.
Pic. 21: Youden Plot for Cu Analysis.
Performance Commentary – Confidence Ellipse:
Laboratory 06 - PT Vale Indonesia Tbk has not participated of the PT for the parameter Cu.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 21 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 22: Ranking Graph according to the Mandel’s k statistic to sample Ni57 for Cu Analysis.
Pic. 23: Ranking Graph according to the Mandel’s k statistic to sample Ni58 for Cu Analysis.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 22 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Tab. 19 – Assigned Value, Standard Deviations and Standard Uncertainties for Fe.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard
Uncertainty
𝒖 (%) Sample
Ni57 15.154 0.10 0.046
Ni58 39.57 0.84 0.37
Tab. 20 – Results reported, average and Z Scores of laboratories participating in the PT Program
for Fe analysis in the sample Ni57.
Results for Fe – Sample Ni57
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 Fe 15.170 15.190 15.320 15.380 15.180 % 15.248 0.909 FUS-XRF Lab 02 Fe 15.140 15.140 15.070 15.160 15.160 % 15.134 -0.191 FUS-XRF Lab 03 Fe 14.820 15.220 15.320 15.290 15.380 % 15.206 0.504 FUS-XRF Lab 04 Fe 15.083 15.147 15.248 15.033 15.132 % 15.128 -0.246 FUS-XRF Lab 05 Fe 15.274 15.222 15.225 15.211 15.332 % 15.253 0.955 FUS-XRF Lab 06 Fe 13.110 13.430 13.040 13.030 13.170 % 13.156 -19.265 FUS-XRF Lab 07 Fe 15.138 15.150 15.154 15.146 15.118 % 15.141 -0.121 FUS-XRF Lab 08 Fe 15.056 15.021 15.045 14.916 15.052 % 15.018 -1.307 FUS-XRF
Tab. 21 – Results reported, average and Z Scores of laboratories participating in the PT Program
for Fe analysis in the sample Ni58.
Results for Fe – Sample Ni58
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 Fe 38.330 38.370 38.380 38.340 38.330 % 38.350 -1.449 FUS-XRF Lab 02 Fe 38.710 38.650 38.340 38.360 38.400 % 38.492 -1.280 FUS-XRF Lab 03 Fe 40.360 40.350 40.440 40.560 40.680 % 40.478 1.079 FUS-XRF Lab 04 Fe 40.080 40.251 40.192 40.154 40.102 % 40.155 0.695 FUS-XRF Lab 05 Fe 39.471 39.771 39.545 39.637 39.448 % 39.574 0.005 FUS-XRF Lab 06 Fe 38.660 38.800 39.530 39.700 39.810 % 39.300 -0.320 FUS-XRF Lab 07 Fe 40.118 40.074 40.121 40.011 40.167 % 40.098 0.628 FUS-XRF Lab 08 Fe 39.636 39.748 39.836 40.199 39.783 % 39.841 0.322 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 23 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 24: Graph with Z Scores for samples Ni57 and Ni58 for Fe Analysis.
Laboratory 06 - PT Vale Indonesia Tbk presented UNSATISFACTORY performance for Fe analysis
in the sample Ni57 and SATISFACTORY performance for Fe analysis in the sample Ni58,
according to the criterion for Z Score evaluating.
Pic. 25: Youden Plot for Fe Analysis.
Performance Commentary – Confidence Ellipse:
The Confidence Ellipse obtained for Fe analysis showed high dispersion among the participating
laboratories. The laboratory 06 was excluded from the Confidence Ellipse.
Laboratory 06 - PT Vale Indonesia Tbk showed an UNSATISFACTORY performance for Fe analysis
according to the criterion for Youden evaluation, because indicated possibility of systematic
negative errors being excluded from the Ellipse. It is recommended to check the standards and
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 24 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
calibration curves, the Certified Reference Materials used, the number of digits used in cases of
conversion / calculation and / or balance / weighing, and others factors that may affect
systematically the Fe analysis.
Pic. 26: Ranking Graph according to the Mandel’s k statistic to sample Ni57 for Fe Analysis.
Pic. 27: Ranking Graph according to the Mandel’s k statistic to sample Ni58 for Fe Analysis.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 25 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Tab. 22 – Assigned Value, Standard Deviations and Standard Uncertainties for LOI (Loss On
Ignition).
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard
Uncertainty
𝒖 (%) Sample
Ni57 9.44 0.90 0.43
Ni58 9.27 0.75 0.36
Tab. 23 – Results reported, average and Z Scores of laboratories participating in the PT Program
for LOI analysis in the sample Ni57.
Results for LOI – Sample Ni57
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 LOI 10.680 10.620 10.650 10.550 10.600 % 10.620 1.313 GRAV Lab 02 LOI 9.760 9.750 9.740 9.840 9.840 % 9.786 0.388 GRAV Lab 03 LOI 8.420 8.500 8.340 8.460 8.320 % 8.408 -1.140 TGA Lab 04 LOI 21.253 21.235 21.167 21.253 21.179 % 21.217 13.064 GRAV Lab 05 LOI 8.251 8.311 8.318 8.325 8.349 % 8.311 -1.248 TGA Lab 06 LOI --- --- --- --- --- % --- --- --- Lab 07 LOI 9.642 9.730 9.565 9.657 9.735 % 9.666 0.255 GRAV Lab 08 LOI 9.768 9.842 9.849 9.784 9.892 % 9.827 0.433 GRAV
Tab. 24 – Results reported, average and Z Scores of laboratories participating in the PT Program
for LOI analysis in the sample Ni58.
Results for LOI – Sample Ni58
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 LOI 9.000 9.000 9.110 9.060 9.080 % 9.050 -0.297 GRAV Lab 02 LOI 9.810 9.740 9.760 9.850 9.860 % 9.804 0.704 GRAV Lab 03 LOI 8.890 9.000 8.900 8.880 8.910 % 8.916 -0.475 TGA Lab 04 LOI 10.618 10.789 10.736 10.672 10.740 % 10.711 1.907 GRAV Lab 05 LOI 8.664 8.686 8.669 8.657 8.654 % 8.666 -0.806 TGA Lab 06 LOI --- --- --- --- --- % --- --- --- Lab 07 LOI 8.548 8.563 8.586 8.570 8.566 % 8.567 -0.938 GRAV Lab 08 LOI 9.211 9.193 9.192 9.216 9.200 % 9.202 -0.095 GRAV
Legend:
GRAV Gravimetric Method
TGA ThermoGravimetric Analyzer
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 26 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 28: Graph with Z Scores for samples Ni57 and Ni58 for LOI Analysis.
Laboratory 06 - PT Vale Indonesia Tbk has not participated of the PT for the parameter LOI.
Pic. 29: Youden Plot for LOI Analysis.
Performance Commentary – Confidence Ellipse:
Laboratory 06 - PT Vale Indonesia Tbk has not participated of the PT for the parameter LOI.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 27 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 30: Ranking Graph according to the Mandel’s k statistic to sample Ni57 for LOI Analysis.
Pic. 31: Ranking Graph according to the Mandel’s k statistic to sample Ni58 for LOI Analysis.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 28 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Tab. 25 – Assigned Value, Standard Deviations and Standard Uncertainties for MgO.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard
Uncertainty
𝒖 (%) Sample
Ni57 8.007 0.14 0.060
Ni58 0.979 0.024 0.010
Tab. 26 – Results reported, average and Z Scores of laboratories participating in the PT Program
for MgO analysis in the sample Ni57.
Results for MgO – Sample Ni57
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 MgO 7.890 7.850 7.880 7.870 7.880 % 7.874 -0.985 FUS-XRF Lab 02 MgO 8.030 8.040 8.050 8.030 8.060 % 8.042 0.255 FUS-XRF Lab 03 MgO 7.740 --- 7.990 7.970 8.010 % 7.928 -0.590 FUS-XRF Lab 04 MgO 8.016 8.052 8.049 8.050 8.026 % 8.039 0.229 FUS-XRF Lab 05 MgO 8.234 8.197 8.223 8.180 8.191 % 8.205 1.457 FUS-XRF Lab 06 MgO 6.590 6.730 6.530 6.540 6.610 % 6.600 -10.383 FUS-XRF Lab 07 MgO 8.010 8.030 8.053 8.077 8.060 % 8.046 0.284 FUS-XRF Lab 08 MgO 7.830 7.800 7.846 7.795 7.925 % 7.839 -1.241 FUS-XRF
Tab. 27 – Results reported, average and Z Scores of laboratories participating in the PT Program
for MgO analysis in the sample Ni58.
Results for MgO – Sample Ni58
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 MgO 0.980 0.980 0.960 0.960 0.960 % 0.968 -0.459 FUS-XRF Lab 02 MgO 0.990 0.990 0.970 0.970 0.990 % 0.982 0.132 FUS-XRF Lab 03 MgO 1.060 1.000 0.990 0.970 0.970 % 0.998 0.807 FUS-XRF Lab 04 MgO 0.997 0.991 1.010 1.016 1.005 % 1.004 1.039 FUS-XRF Lab 05 MgO 0.972 0.997 0.991 0.996 1.008 % 0.993 0.587 FUS-XRF Lab 06 MgO 0.840 0.870 0.860 0.870 0.880 % 0.864 -4.847 FUS-XRF Lab 07 MgO 0.953 0.950 0.943 0.946 0.947 % 0.948 -1.311 FUS-XRF Lab 08 MgO 1.226 1.229 1.215 1.210 1.222 % 1.220 10.191 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 29 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 32: Graph with Z Scores for samples Ni57 and Ni58 for MgO Analysis.
Laboratory 06 - PT Vale Indonesia Tbk presented UNSATISFACTORY performance for MgO
analysis in the samples Ni57 and Ni58, according to the criterion for Z Score evaluating.
Pic. 33: Youden Plot for MgO Analysis.
Performance Commentary – Confidence Ellipse:
The Confidence Ellipse obtained for MgO analysis showed normal dispersion among the
participating laboratories. The laboratories 06 and 08 were excluded from the Confidence
Ellipse.
Laboratory 06 - PT Vale Indonesia Tbk showed an UNSATISFACTORY performance for MgO
analysis according to the criterion for Youden evaluation, because indicated possibility of
systematic negative errors being excluded from the Ellipse. It is recommended to check the
standards and calibration curves, the Certified Reference Materials used, the number of digits
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 30 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
used in cases of conversion / calculation and / or balance / weighing, and others factors that may
affect systematically the MgO analysis.
Pic. 34: Ranking Graph according to the Mandel’s k statistic to sample Ni57 for MgO Analysis.
Pic. 35: Ranking Graph according to the Mandel’s k statistic to sample Ni58 for MgO Analysis.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 31 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Tab. 28 – Assigned Value, Standard Deviations and Standard Uncertainties for MnO.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard
Uncertainty
𝒖 (%) Sample
Ni57 0.4924 0.0085 0.0038
Ni58 0.4389 0.013 0.0058
Tab. 29 – – Results reported, average and Z Scores of laboratories participating in the PT Program
for MnO analysis in the sample Ni57.
Results for MnO – Sample Ni57
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 MnO 0.530 0.550 0.530 0.540 0.540 % 0.538 5.346 FUS-XRF Lab 02 MnO 0.500 0.510 0.510 0.500 0.510 % 0.506 1.592 FUS-XRF Lab 03 MnO 0.490 0.490 0.500 0.500 0.500 % 0.496 0.419 FUS-XRF Lab 04 MnO 0.490 0.490 0.494 0.488 0.489 % 0.490 -0.297 FUS-XRF Lab 05 MnO 0.488 0.486 0.486 0.485 0.488 % 0.487 -0.684 FUS-XRF Lab 06 MnO 0.412 0.420 0.408 0.410 0.412 % 0.412 -9.390 FUS-XRF Lab 07 MnO 0.492 0.490 0.495 0.491 0.493 % 0.492 -0.027 FUS-XRF Lab 08 MnO 0.483 0.481 0.484 0.484 0.486 % 0.484 -1.036 FUS-XRF
Tab. 30 – Results reported, average and Z Scores of laboratories participating in the PT Program
for MnO analysis in the sample Ni58.
Results for MnO – Sample Ni58
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 MnO 0.420 0.420 0.440 0.420 0.420 % 0.424 -1.132 FUS-XRF Lab 02 MnO 0.450 0.440 0.440 0.450 0.440 % 0.444 0.385 FUS-XRF Lab 03 MnO 0.460 0.460 0.460 0.450 0.460 % 0.458 1.447 FUS-XRF Lab 04 MnO 0.449 0.449 0.452 0.451 0.439 % 0.448 0.666 FUS-XRF Lab 05 MnO 0.428 0.432 0.428 0.430 0.428 % 0.429 -0.738 FUS-XRF Lab 06 MnO 0.418 0.420 0.429 0.421 0.425 % 0.423 -1.238 FUS-XRF Lab 07 MnO 0.442 0.439 0.443 0.441 0.443 % 0.442 0.203 FUS-XRF Lab 08 MnO 0.444 0.443 0.444 0.444 0.445 % 0.444 0.385 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 32 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 36: Graph with Z Scores for samples Ni57 and Ni58 for MnO Analysis.
Laboratory 06 - PT Vale Indonesia Tbk presented UNSATISFACTORY performance for MnO
analysis in the sample Ni57 and SATISFACTORY performance for MnO analysis in the sample
Ni58, according to the criterion for Z Score evaluating.
Pic. 37: Youden Plot for MnO Analysis.
Performance Commentary – Confidence Ellipse:
The Confidence Ellipse obtained for MnO analysis showed high dispersion among the
participating laboratories. The laboratories 01 and 06 were excluded from the Confidence
Ellipse.
Laboratory 06 - PT Vale Indonesia Tbk showed an UNSATISFACTORY performance for MnO
analysis according to the criterion for Youden evaluation, because indicated possibility of
systematic negative errors being excluded from the Ellipse. It is recommended to check the
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 33 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
standards and calibration curves, the Certified Reference Materials used, the number of digits
used in cases of conversion / calculation and / or balance / weighing, and others factors that may
affect systematically the MnO analysis.
Pic. 38: Ranking Graph according to the Mandel’s k statistic to sample Ni57 for MnO Analysis.
Pic. 39: Ranking Graph according to the Mandel’s k statistic to sample Ni58 for MnO Analysis.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 34 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Tab. 31 – Assigned Value, Standard Deviations and Standard Uncertainties for Ni.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard
Uncertainty
𝒖 (%) Sample
Ni57 1.943 0.045 0.020
Ni58 0.9458 0.016 0.0069
Tab. 32 – Results reported, average and Z Scores of laboratories participating in the PT Program
for Ni analysis in the sample Ni57.
Results for Ni – Sample Ni57
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 Ni 2.000 2.000 1.970 1.990 1.990 % 1.990 1.034 FUS-XRF Lab 02 Ni 1.960 1.960 1.950 1.950 1.960 % 1.956 0.282 FUS-XRF Lab 03 Ni 1.890 1.930 1.940 1.940 1.970 % 1.934 -0.204 FUS-XRF Lab 04 Ni 1.985 1.984 2.016 1.965 1.984 % 1.987 0.959 FUS-XRF Lab 05 Ni 1.947 1.938 1.939 1.938 1.954 % 1.943 -0.001 FUS-XRF Lab 06 Ni 1.656 1.670 1.643 1.641 1.670 % 1.656 -6.353 FUS-XRF Lab 07 Ni 1.877 1.873 1.882 1.887 1.883 % 1.880 -1.390 FUS-XRF Lab 08 Ni 1.900 1.905 1.900 1.905 1.905 % 1.903 -0.890 FUS-XRF
Tab. 33 – Results reported, average and Z Scores of laboratories participating in the PT Program
for Ni analysis in the sample Ni58.
Results for Ni – Sample Ni58
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 Ni 1.010 1.020 1.000 1.010 1.000 % 1.008 4.004 FUS-XRF Lab 02 Ni --- 0.950 0.950 0.950 0.940 % 0.948 0.108 FUS-XRF Lab 03 Ni 0.960 0.960 0.960 0.970 0.960 % 0.962 1.041 FUS-XRF Lab 04 Ni 0.931 0.946 0.945 0.939 0.916 % 0.935 -0.684 FUS-XRF Lab 05 Ni 0.958 0.967 0.961 0.963 0.959 % 0.962 1.016 FUS-XRF Lab 06 Ni 0.940 0.948 0.972 0.944 0.978 % 0.956 0.681 FUS-XRF Lab 07 Ni 0.937 0.947 0.943 0.937 0.945 % 0.942 -0.259 FUS-XRF Lab 08 Ni 0.925 0.920 0.925 0.935 0.930 % 0.927 -1.213 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 35 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 40: Graph with Z Scores for samples Ni57 and Ni58 for Ni Analysis.
Laboratory 06 - PT Vale Indonesia Tbk presented UNSATISFACTORY performance for Ni analysis
in the sample Ni57 and SATISFACTORY performance for Ni analysis in the sample Ni58,
according to the criterion for Z Score evaluating.
Pic. 41: Youden Plot for Ni Analysis.
Performance Commentary – Confidence Ellipse:
The Confidence Ellipse obtained for Ni analysis showed normal dispersion among the
participating laboratories. The laboratories 01 and 06 were excluded from the Confidence
Ellipse.
Laboratory 06 - PT Vale Indonesia Tbk showed an UNSATISFACTORY performance for Ni analysis
according to the criterion for Youden evaluation, because indicated possibility of random errors
being excluded from the Ellipse. It is recommended to check the possible factors that may cause
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 36 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
random errors for Analysis of Ni, such as operator, instruments, calculation and results
transcription.
Pic. 42: Ranking Graph according to the Mandel’s k statistic to sample Ni57 for Ni Analysis.
Pic. 43: Ranking Graph according to the Mandel’s k statistic to sample Ni58 for Ni Analysis.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 37 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Tab. 34 – Assigned Value, Standard Deviations and Standard Uncertainties for SiO2.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard
Uncertainty
𝒖 (%) Sample
Ni57 50.06 0.26 0.12
Ni58 22.29 0.35 0.16
Tab. 35 – Results reported, average and Z Scores of laboratories participating in the PT Program
for SiO2 analysis in the sample Ni57.
Results for SiO2 – Sample Ni57
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 SiO2 50.210 50.220 50.250 50.210 50.190 % 50.216 0.598 FUS-XRF Lab 02 SiO2 50.010 49.890 49.890 49.910 49.970 % 49.934 -0.483 FUS-XRF Lab 03 SiO2 49.670 50.060 50.420 50.290 50.540 % 50.196 0.521 FUS-XRF Lab 04 SiO2 50.000 50.407 50.338 50.159 49.966 % 50.174 0.436 FUS-XRF Lab 05 SiO2 49.853 49.812 49.800 49.664 49.967 % 49.819 -0.923 FUS-XRF Lab 06 SiO2 42.200 42.540 42.040 42.050 42.350 % 42.236 -29.977 FUS-XRF Lab 07 SiO2 49.735 49.606 49.923 49.849 49.817 % 49.786 -1.050 FUS-XRF Lab 08 SiO2 50.380 50.530 50.209 50.285 50.070 % 50.295 0.900 FUS-XRF
Tab. 36 – Results reported, average and Z Scores of laboratories participating in the PT Program
for SiO2 analysis in the sample Ni58.
Results for SiO2 – Sample Ni58
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 SiO2 22.450 22.430 22.470 22.430 22.450 % 22.446 0.435 FUS-XRF Lab 02 SiO2 21.980 22.050 21.980 22.040 22.020 % 22.014 -0.795 FUS-XRF Lab 03 SiO2 23.020 22.800 22.860 22.870 22.980 % 22.906 1.745 FUS-XRF Lab 04 SiO2 22.418 22.487 22.566 22.499 22.166 % 22.427 0.381 FUS-XRF Lab 05 SiO2 22.386 22.573 22.366 22.407 22.410 % 22.428 0.385 FUS-XRF Lab 06 SiO2 21.610 21.710 21.970 21.710 21.850 % 21.770 -1.490 FUS-XRF Lab 07 SiO2 22.250 22.187 22.210 22.195 22.207 % 22.210 -0.238 FUS-XRF Lab 08 SiO2 22.258 22.140 22.055 22.100 22.169 % 22.144 -0.424 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 38 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 44: Graph with Z Scores for samples Ni57 and Ni58 for SiO2 Analysis.
Laboratory 06 - PT Vale Indonesia Tbk presented UNSATISFACTORY performance for SiO2
analysis in the sample Ni57 and SATISFACTORY performance for SiO2 analysis in the sample
Ni58, according to the criterion for Z Score evaluating.
Pic. 45: Youden Plot for SiO2 Analysis.
Performance Commentary – Confidence Ellipse:
The Confidence Ellipse obtained for SiO2 analysis showed normal dispersion among the
participating laboratories. The laboratory 06 was excluded from the Confidence Ellipse.
Laboratory 06 - PT Vale Indonesia Tbk showed an UNSATISFACTORY performance for SiO2
analysis according to the criterion for Youden evaluation, because indicated possibility of
systematic negative errors being excluded from the Ellipse. It is recommended to check the
standards and calibration curves, the Certified Reference Materials used, the number of digits
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 39 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
used in cases of conversion / calculation and / or balance / weighing, and other factors that may
affect systematically the SiO2 analysis.
Pic. 46: Ranking Graph according to the Mandel’s k statistic to sample Ni57 for SiO2 Analysis.
Pic. 47: Ranking Graph according to the Mandel’s k statistic to sample Ni58 for SiO2 Analysis.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 40 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Tab. 37 – Assigned Value, Standard Deviations and Standard Uncertainties for TiO2.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard
Uncertainty
𝒖 (%) Sample
Ni57 0.1416 0.0023 0.0015
Ni58 0.1239 0.013 0.0083
Tab. 38 – Results reported, average and Z Scores of laboratories participating in the PT Program
for TiO2 analysis in the sample Ni57.
Results for TiO2 – Sample Ni57
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 TiO2 --- --- --- --- --- % --- --- --- Lab 02 TiO2 --- --- --- --- --- % --- --- --- Lab 03 TiO2 0.140 0.140 0.140 0.140 0.140 % 0.140 -0.688 FUS-XRF Lab 04 TiO2 0.233 0.233 0.220 0.239 0.230 % 0.231 38.339 FUS-XRF Lab 05 TiO2 --- --- --- --- --- % --- --- --- Lab 06 TiO2 --- --- --- --- --- % --- --- --- Lab 07 TiO2 0.150 0.142 0.146 0.148 0.143 % 0.146 1.807 FUS-XRF Lab 08 TiO2 0.143 0.142 0.144 0.143 0.144 % 0.143 0.688 FUS-XRF
Tab. 39 – Results reported, average and Z Scores of laboratories participating in the PT Program
for TiO2 analysis in the sample Ni58.
Results for TiO2 – Sample Ni58
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 TiO2 --- --- --- --- --- % --- --- --- Lab 02 TiO2 --- --- --- --- --- % --- --- --- Lab 03 TiO2 0.120 0.120 0.120 0.110 0.110 % 0.116 -0.597 FUS-XRF Lab 04 TiO2 0.195 0.193 0.192 0.188 0.225 % 0.198 5.591 FUS-XRF Lab 05 TiO2 --- --- --- --- --- % --- --- --- Lab 06 TiO2 --- --- --- --- --- % --- --- --- Lab 07 TiO2 0.141 0.137 0.135 0.142 0.139 % 0.139 1.119 FUS-XRF Lab 08 TiO2 0.116 0.117 0.118 0.118 0.116 % 0.117 -0.522 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 41 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 48: Graph with Z Scores for samples Ni57 and Ni58 for TiO2 Analysis.
Laboratory 06 - PT Vale Indonesia Tbk has not participated of the PT for the parameter TiO2.
Performance Commentary – Confidence Ellipse:
The valid results are not sufficient to build the Confidence Ellipse for the parameter TiO2.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 42 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 49: Ranking Graph according to the Mandel’s k statistic to the sample Ni57 for TiO2 Analysis.
Pic. 50: Ranking Graph according to the Mandel’s k statistic to the sample Ni58 for TiO2 Analysis.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 43 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Tab. 40 – Assigned Value, Standard Deviations and Standard Uncertainties for Zn.
Assigned Value
𝑿𝑫 (%)
Standard Deviation
𝒔 (%)
Standard
Uncertainty
𝒖 (%) Sample
Ni57 0.0401 0.0083 0.0052
Ni58 0.0267 0.0048 0.0030
Tab. 41 – Results reported, average and Z Scores of laboratories participating in the PT Program
for Zn analysis in the sample Ni57.
Results for Zn – Sample Ni57
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 Zn --- --- --- --- --- % --- --- --- Lab 02 Zn --- --- --- --- --- % --- --- --- Lab 03 Zn 0.030 0.030 0.030 0.030 0.030 % 0.030 -1.220 FUS-XRF Lab 04 Zn --- --- --- --- --- % --- --- --- Lab 05 Zn 0.042 0.042 0.042 0.041 0.042 % 0.042 0.205 FUS-XRF Lab 06 Zn --- --- --- --- --- % --- --- --- Lab 07 Zn 0.050 0.050 0.050 0.050 0.050 % 0.050 1.196 FUS-XRF Lab 08 Zn 0.038 0.038 0.039 0.039 0.039 % 0.039 -0.181 FUS-XRF
Tab. 42 – Results reported, average and Z Scores of laboratories participating in the PT Program
for Zn analysis in the sample Ni58.
Results for Zn – Sample Ni58
Code Parameter 1 2 3 4 5 Unit Average Z Score Method Lab 01 Zn --- --- --- --- --- % --- --- --- Lab 02 Zn --- --- --- --- --- % --- --- --- Lab 03 Zn 0.030 0.030 0.030 0.030 0.030 % 0.030 0.687 FUS-XRF Lab 04 Zn --- --- --- --- --- % --- --- --- Lab 05 Zn 0.023 0.022 0.022 0.022 0.022 % 0.022 -0.937 FUS-XRF Lab 06 Zn --- --- --- --- --- % --- --- --- Lab 07 Zn 0.032 0.032 0.031 0.032 0.031 % 0.032 1.021 FUS-XRF Lab 08 Zn 0.023 0.023 0.023 0.023 0.023 % 0.023 -0.771 FUS-XRF
Legend:
FUS-XRF Fusion / X-Ray Fluorescence
X,XX Satisfactory Result
X,XX Questionable Result
X,XX Unsatisfactory Result
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 44 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 51: Graph with Z Scores for samples Ni57 and Ni58 for Zn Analysis.
Laboratory 06 - PT Vale Indonesia Tbk has not participated of the PT for the parameter Zn.
Performance Commentary – Confidence Ellipse:
The valid results are not sufficient to build the Confidence Ellipse for the parameter Zn.
RT-138-2020 Cgcre’s accredited ABNT NBR ISO/IEC 17043:2011 Proficiency Testing Provider. Accreditation
nº PEP 0021. Contacts: email: [email protected]; Phone nº: +55 31 3851-3166 / +55 31 3851-6952 45 / 45
PROFICIENCY TEST PROGRAM IN NICKEL ORE ANALYZES – 32nd ROUND
Excl
usi
ve u
se i
n d
igit
al m
edia
. Rep
rodu
ctio
n p
roh
ibit
ed.
Pic. 52: Ranking Graph according to the Mandel’s k statistic to sample Ni57 for Zn Analysis.
Pic. 53: Ranking Graph according to the Mandel’s k statistic to sample Ni58 for Zn Analysis.