table 2.1-1: electrical receptacles for patient care areas

7
Chapter 2.1 tables DRAFT 2022 FGI Guidelines for Design and Construction of Outpatient Facilities 1 Table 2.1-1: Electrical Receptacles for Patient Care Areas in Outpatient Facilities 1 Section Room Type Number of Single Receptacles 12 Receptacle Locations 23 PATIENT BED LOCATIONS 2.15-2.2 Extended stay patient room 12 2 at each side of the head of the bed 2 on all other walls 1 for a television, where provided 1 for each motorized bed, where provided 2.15- 2.2.4.1 Extended stay airborne infection isolation (AII) room 12 2 at each side of the head of the bed 2 on all other walls 1 for a television, where provided 1 for each motorized bed, where provided PATIENT CARE AND DIAGNOSTIC AREAS 2.1-3.2.1 Examination room/observation room 8 4 convenient to head of exam table or gurney or on each lateral side of the imaging gantry Table 2.1- 4 Class 1 imaging room 8 4 convenient to head of exam table or gurney or on each lateral side of the imaging gantry 2.1-3.2.2 Procedure room (including endoscopy) 12 34 8 convenient to table placement At least 1 on each wall Table 2.1- 4 Class 2 imaging room 12 34 8 convenient to table placement At least 1 on each wall 2.1-3.2.3 Operating room 36 34 12 convenient to table placement 2 on each wall Table 2.1- 4 Class 3 imaging room 36 34 12 convenient to table placement 2 on each wall 2.1-3.7.3 Pre-procedure patient care station 4 Convenient to gurney, lounge chair, or bed 2.1-3.7.5 Phase II recovery patient care station 4 Convenient to gurney, lounge chair, or bed 2.1-3.7.4 Phase I post-anesthesia recovery (PACU) patient care station 8 Convenient to head of gurney or bed 2.4-2.2 Birthing room 8 4 convenient to head of the mother’s bed 2.8-3.4.2 Emergency facility treatment room 12 4 convenient to head of exam table or gurney 2.8-3.4.4 Trauma/resuscitation room 16 Convenient to head of gurney or bed 2.8- 3.4.8.2 Human decontamination room 4

Upload: others

Post on 12-Jan-2022

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Table 2.1-1: Electrical Receptacles for Patient Care Areas

Chapter 2.1 tables

DRAFT 2022 FGI Guidelines for Design and Construction of Outpatient Facilities 1

Table 2.1-1: Electrical Receptacles for Patient Care Areas in Outpatient Facilities1

Section Room Type Number of Single Receptacles12

Receptacle Locations23

PATIENT BED LOCATIONS

2.15-2.2 Extended stay patient room 12 2 at each side of the head of the bed 2 on all other walls 1 for a television, where provided 1 for each motorized bed, where provided

2.15-2.2.4.1

Extended stay airborne infection isolation (AII) room

12 2 at each side of the head of the bed 2 on all other walls 1 for a television, where provided 1 for each motorized bed, where provided

PATIENT CARE AND DIAGNOSTIC AREAS

2.1-3.2.1

Examination room/observation room

8 4 convenient to head of exam table or gurney or on each lateral side of the imaging gantry

Table 2.1-4

Class 1 imaging room 8 4 convenient to head of exam table or gurney or on each lateral side of the imaging gantry

2.1-3.2.2

Procedure room (including endoscopy)

1234 8 convenient to table placement At least 1 on each wall

Table 2.1-4

Class 2 imaging room 1234 8 convenient to table placement At least 1 on each wall

2.1-3.2.3

Operating room

3634

12 convenient to table placement 2 on each wall

Table 2.1-4

Class 3 imaging room 3634

12 convenient to table placement 2 on each wall

2.1-3.7.3 Pre-procedure patient care station

4 Convenient to gurney, lounge chair, or bed

2.1-3.7.5 Phase II recovery patient care station

4 Convenient to gurney, lounge chair, or bed

2.1-3.7.4 Phase I post-anesthesia recovery (PACU) patient care station

8 Convenient to head of gurney or bed

2.4-2.2 Birthing room 8 4 convenient to head of the mother’s bed 2.8-3.4.2 Emergency facility

treatment room 12 4 convenient to head of exam table or

gurney

2.8-3.4.4 Trauma/resuscitation room 16 Convenient to head of gurney or bed

2.8-3.4.8.2

Human decontamination room 4 —

Page 2: Table 2.1-1: Electrical Receptacles for Patient Care Areas

Chapter 2.1 tables

DRAFT 2022 FGI Guidelines for Design and Construction of Outpatient Facilities 2

2.8-3.4.10.2

Low-acuity patient treatment station

4 Convenient to patient chair

2.8-6.2.2 Emergency facility triage area

6 Convenient to head of gurney or bed (at least 3 outlets connected to emergency system power and so labeled)

2.10-3.2.2 Hemodialysis patient care station

8 4 on each side of a patient bed or lounge chair (2 on each side of the bed connected to emergency power)

1Receptable numbers reflect the total number of receptacles from normal power, emergency power, or any combination thereof. See Section 2.1-8.3.3.1 (Essential electrical system or emergency electrical power) to determine if an essential electrical system or emergency electrical power is required.

21Permanently installed single, duplex, or fourplex receptacles or a combination of these shall be permitted. Receptacles in relocatable power taps or mounted on portable equipment shall not be counted as part of the total minimum requirement.

32In this table, “convenient” means the cords from the equipment to be used in the room can reach the receptacles without causing a trip hazard.

43The number of receptacles for these spaces is intended to agree with the number required in the governing edition of NFPA 99: Health Care Facilities Code.

Notes 1. Consideration shall be given to providing some outlets on essential electrical power and some on normal

power in operating rooms and post-anesthesia recovery areas in case of transfer switch failure. 2. Each room in the table shall meet the requirements for connection to the essential electrical system in the

governing edition of NFPA 99: Health Care Facilities Code. 3. Branch circuits serving only special purpose receptacles shall be permitted to be served by other

panelboards.

Page 3: Table 2.1-1: Electrical Receptacles for Patient Care Areas

Chapter 2.1 tables

DRAFT 2022 FGI Guidelines for Design and Construction of Outpatient Facilities 3

Table 2.1-2: Station Outlets for Oxygen, Vacuum, Medical Air, and Instrument Air Systems in Outpatient Facilities

Section Location Oxygen Vacuum Nitrous Oxide Medical Air Instrument Air PATIENT CARE AND DIAGNOSTIC AREAS 2.1-3.2.2 Procedure room 11 11 — — —

Table 2.1-4 Class 2 imaging room 2 2 — 11 —

2.1-3.2.3.2 (1)(a)

Operating room (255-square-foot OR)

11 11 01, 2 — —

2.1-3.2.3.2 (1)(b)-(c)

Operating rooms2 22 32 01, 2 11,2 —

Table 2.1-4 Class 3 imaging room2 22 32 01, 2 11,2 —

2.1-3.2.4.4 Hyperbaric pre-procedure patient care area

21 21 — — —3

2.1-3.3.2 Airborne infection isolation room 01 01 — — —

2.1-3.5.5.2 Magnetic resonance imaging scanner room

01, 2 01, 2 01, 2 — —

2.1- 3.7.4 Phase I post-anesthesia recovery (PACU) patient care station

01 01 — — —

2.1-3.7.5 Phase II recovery patient care station

01 01 — — —

— Cast room 01 01 01, 2 — —

2.4-2.2 Birthing room 11 11 01, 2 — —

2.8-3.4.2 Emergency facility treatment room

1 1 — — —

2.8-3.4.4 Trauma/resuscitation room—per gurney

2 2 — 1 —

2.8-3.4.8.2 Human decontamination room 11 01 — — —

2.8-3.4.10.2 Low-acuity patient treatment station

—1 —1 — — —

2.8-6.2.2 Emergency facility triage area—per station

1 1 — — —

2.9-3.2.2 Endoscopy procedure room 1 3 — — —

2.11-3.2.9.2 (2)

Electroconvulsive therapy treatment room

11 11 — — —

2.14-3.3 Dental operatory room — — 01, 2 — — PATIENT SUPPORT FACILITIES 2.1-4.3.2.2 (2) Sterile processing

decontamination room — — — — 11, 4, 5

2.1-4.3.2.2 (3) Sterile processing clean workroom

— — — — —1, 4, 5

2.1-4.3.2.3 One-room sterile processing room — — — — —1, 4, 5

2.9-4.3.2 Endoscope processing room—decontamination area

— —3 — — —1, 3, 5

2.9-4.3.3 Endoscope processing room—clean work area

— —3 — — —1, 3, 5

Page 4: Table 2.1-1: Electrical Receptacles for Patient Care Areas

Chapter 2.1 tables

DRAFT 2022 FGI Guidelines for Design and Construction of Outpatient Facilities 4

1Use of portable equipment in lieu of a piped gas system shall be permitted. 2If Where inhalation anesthesia, including nitrous oxide, will be used, in an operating room or Class 3 imaging

room, installation of a waste anesthetic gas disposal (WAGD) system should shall be provided considered. Use of portable delivery and scavenging equipment shall be permitted in lieu of a permanently installed WAGD system.

3Vacuum and/or instrument air shall be provided if needed for cleaning methods used. 4In the one-room sterile processing facility and the clean workroom of the two-room sterile processing facility,

an instrument air outlet or portable compressed air shall be provided as required by the equipment used. In the decontamination room of the two-room sterile processing facility, an instrument air outlet or portable compressed air is required.

5NFPA 99 permits the use of portable medical compressed air for single applications. Where cylinders are used for non-respiratory purposes, such as air for blowing down scopes and/or running decontamination equipment, NFPA 99 should be consulted for cylinder air quality, placement, and handling.

Note The provision of medical gases shall also comply with NFPA 99: Health Care Facilities Code. The clinical risk assessment in NFPA 99 may result in more stringent requirements for locations where this table permits use of portable gases.

Page 5: Table 2.1-1: Electrical Receptacles for Patient Care Areas

Chapter 2.1 tables

DRAFT 2022 FGI Guidelines for Design and Construction of Outpatient Facilities 5

Table 2.1-3: Locations for Nurse Call Devices in Outpatient Facilities1, 2

Section Location Patient Station

Staff Assistance Station

Emergency Call Station

Toilet Room Call Station

Notes

PATIENT CARE AND DIAGNOSTIC AREAS

2.1-3.2.2 Table 2.1-4

Procedure room (including endoscopy)

Class 2 imaging room l l

2, 3

2.1-3.2.3 Operating room l l 2 Table 2.1-4 Class 3 imaging room l l 2 2.1-3.7.3 Pre-procedure patient care station l l l 1, 2 2.1-3.7.4 Phase I post-anesthesia recovery

(PACU) patient care station l l l 2, 3

2.1-3.7.5 Phase II recovery patient care station l l l 1, 2 2.8-3.4.2 Emergency facility treatment room l l 1, 2, 3 2.8-3.4.8.2 Human decontamination room l 2.8-3.4.10.2 Low-acuity treatment patient care

station l

2.8-6.2.2 Emergency facility triage area l l 1, 2, 3 2.10-3.2.2 Hemodialysis patient care station l l 1 2.10-3.3 Dialysis facility home training room l l 1, 4 2.10-3.10.2 Dialysis facility patient toilet room l l 1 2.11-3.2.9.2 (2)

Electroconvulsive therapy (ECT) room l l 2

2.11-3.2.9.3 (2)

ECT recovery patient care station l l 2

1It is recognized that staff other than nurses may respond to these devices, but the term “nurse call” is used here as it is an industry-accepted term.

2A Where patient toilet room call stations are optional, a safety risk assessment shall be performed. identify patient toilet rooms where a patient station is required.

Notes 1. One device shall be permitted to accommodate patient station, staff assistance station, and emergency call

station functionality. 2. A visible signal shall be activated in the corridor at the door to the room from which the signal was initiated

and at the nurse/control station where one is provided. 3. One device shall be permitted to accommodate both staff assistance and emergency call station

functionality. 4. Provision of toilet room call stations shall be predicated on the fall prevention assessment portion of the

safety risk assessment.

Page 6: Table 2.1-1: Electrical Receptacles for Patient Care Areas

Chapter 2.1 tables

DRAFT 2022 FGI Guidelines for Design and Construction of Outpatient Facilities 6

Table 2.1-4: Examination/Treatment, Procedure, and Operating Room Classification1

[no changes; not in draft Table 2.1-5: Classification of Room Types for Imaging Services1 [no changes; not in draft] Appendix Table A2.1-a: Waiting Area Seating Capacity1 Chapter Facility Type Seating Calculation2 Wheelchair

Waiting Spaces3 2.2 General and specialty medical

facilities Primary care center and small neighborhood clinic

1.5 per patient care room 1

2.3 Outpatient imaging facility 1.5 per patient care room 1 2.4 Birth center 2 per birthing room 1 2.5 Urgent care center 2 per patient care room 1 2.6 Infusion and cancer treatment

center 2 per bay, cubicle or bed 1

2.7 Outpatient surgery facility 1.5 per procedure/operating room 1 2.9 Endoscopy facility 1.5 per endoscopy room 1 2.10 Dialysis center 2 .75 per patient care station 2 2.11 Psychiatric center 1.5 per patient care room 1 2.12 Rehab therapy facility 1.5 per simultaneous patient visits 5% total seats 2.14 Dental facility 1.5 per patient care room/station 1

1The total seating capacity should include 10 percent seating to accommodate individuals of size (patients and those who accompany them). See Section 2.1-2 (Accommodations for Care of Individuals of Size) for more information.

2Seating capacity should be rounded up to the next whole number when calculating the total number of seats required.

3Open spaces should be included for patients in wheelchairs who are waiting for care; these spaces may be counted as part of the total number of seats provided.

Page 7: Table 2.1-1: Electrical Receptacles for Patient Care Areas

Chapter 2.1 tables

DRAFT 2022 FGI Guidelines for Design and Construction of Outpatient Facilities 7

Appendix Table A2.1-b: Maximum Length of Hot Water System Pipe or Tube [no changes; not in draft] Appendix Table A2.1-c: Resources for Grab Bar Configurations

Accessibility Standards

(ANSI/ADAAG) Georgia Tech Study*

Preferred alternative configuration for one-person, two-person, or

equipment-assisted transfer Wall Partition Location Behind and adjacent Behind and adjacent, where

provided Grab Bar Dimensions Centerline of toilet from side wall or permanent fixture

16 to 18 inches (40.64 to 45.72 centimeters)

24 inches (60.96 centimeters) for independent resident transfer

Side wall partition grab bar length

42 inches (1.07 meters) long 12 inches (30.48 centimeters)

maximum from rear wall 54 inches (1.37 meters) minimum

from rear wall

Not addressed

Rear wall partition grab bar length

36 inches (91.44 centimeters) long minimum

12 inches (30.48 centimeters) from centerline of toilet on one side and 24 inches (60.96 centimeters) on the other side

Eliminate in favor of installing swing-up grab bars

Fixed horizontal grab bar height

33 to 36 inches (83.82 to 91.44 centimeters) above finished floor to top of gripping surface

Not addressed

Swing-up grab bar height Not applicable 31 to 33 inches (78.74 to 83.82 centimeters) above finished floor to top of gripping service

Swing-up grab bar length Not applicable Extend 6 to 9 inches (15.24 to 22.86 centimeters) in front of toilet

Swing-up grab bar from centerline of toilet

Not applicable 13 to 15 inches (33.02 to 38.1 centimeters)

*This data is based on the Mayer-Rothschild Foundation report “Determination of Grab Bar Specifications for Independent and Assisted Transfers in Residential Care Settings.” A related article on the research appeared in the HERD Journal September 2017 under the title “Beyond ADA Accessibility Requirements: Meeting Seniors’ Needs for Toilet Transfers.”