Page 2: STAND ALONE STABILITY/HUMIDITY CHAMBER

EQUIPMENT TECHNICAL SPECIFICATIONS

Internal MOC Stainless Steel 304 grade with Ma� finish

External MOC Stainless Steel 304 grade with Ma� finish

Insula�on 60 mm thick Rigid PU Foam with density of 40kg / m3 ± 2 kgs

Tray Heavy Duty SS Perforated tray with addi�onal support and unique Highly efficient perfora�ons for be�er uniformity.

Door External - Heavy duty construc�on, insulated, imported hinges with Gasket Internal - Glass with Heavy gauge aluminium frame with SS Latches, Door access Magnet and Silicon Gasket.

U�lity Internal u�lity installed at rear of the chamber.

Refrigera�on System: Set of Energy Efficient Copeland compressor, fin and tube type evaporator & condenser. Auto Mode Instant Start (Applicable only with Stand by Refrigera�on system) of compressors to assure opera�onal con�nuity during power outage and minimise Alarm dura�ons.

Humidifica�on system: Set of reservoir tank, Steam generator tank, ke�le heater, safety thermostat, float valve.

Air hea�ng system: Built-in set of stainless-steel tubular heaters provided for efficient air hea�ng.

Control system:

Controller Advanced PLC based control system installed with HMI

PLC Make Delta with built-in Ethernet connec�vity

HMI Make Delta 4.3” colour TFT LCD touch screen

Controlling Sensor- 1 no. Imported Transmi�er type, 4 to 20 MA with Accuracy of +/- 0.3°C & +/- 3% for RH

Control Switchgear L&T Make Contactor & MCB, etc.

Auto Changeover Refrigera�on and Humidifica�on system in case of primary failure

PLC Controlled Auto & manual Changeover to standby

· Refrigera�on: Dura�on & failure inclusive based for refrigera�on · Humidifica�on: Failure based Auto Changeover for humidifiers

Data Storage 1500 Lines for Controlling and Monitoring Sensor data

Illumina�on LEDs for longer life and energy efficiency

Valida�on Access Port Hole will not be provided since Wireless Data loggers are used for calibra�on & Valida�on.

Temperature

Range → 20°C TO 60°C

Opera�ng Accuracy → ≤ +/- 0.2°C

Uniformity→ ≤ +/- 1.0°C

Range → 40%RH to 90%RH

Opera�ng Accuracy → ≤ +/- 1%RH

Uniformity → ≤ +/- 3.0%

Humidity Opera�ng Condi�ons As Per ICH

Special Condi�ons

25°C / 60%RH

30°C / 65%RH

40°C / 75%RH

30°C / 75%RH

25°C / 40%RH

35°C / 35%RH

40°C / 25%RH

Page 3: STAND ALONE STABILITY/HUMIDITY CHAMBER

Safety features PLC based 10 levels of Safety cut offs to protect undershoot and overshoot of Temperature and Humidity.

Temperature high cut 1st Level Temperature Posi�ve Excursion- Sum of Set Value and Process Value. off (°C.THC1) Ac�on- Steam generator and Air heater Cut-Off.

Temperature higher cut 2nd Level Temperature Posi�ve Excursion- Sum of Set Value and Process Value.off (°C.THC2) Ac�on- Mains Supply Cut-Off

Temperature low cut off 1st Level Temperature Nega�ve Excursion- Difference between Set Value and Process (°C.TLC1) Value. Ac�on- Compressor Cut-Off

Temperature lower cut off 2nd Level Temperature Nega�ve Excursion- Difference between Set Value and (°C.TLC2) Process Value. Ac�on- Mains Supply Cut-Off

Humidity high cut off 1st Level Rela�ve Humidity Posi�ve Excursion- Sum of Set Value and Process Value.(%RHC1) Ac�on- Alarm Buzzer ac�va�on

Humidity higher cut off 2nd Level Rela�ve Humidity Posi�ve Excursion- Sum of Set Value and Process Value.(%RHC2) Ac�on- Steam Generator (Primary and Stand By) Cut-Off

Humidity low cut off 1st Level Rela�ve Humidity Nega�ve Excursion- Difference between Set Value and (%RHLC1) Process Value, Ac�on- Mains Supply Cut-Off

Low water steam Decreased Water Level in Primary SG Reservoir below required level.generator cut off Ac�on- Primary SG Shut Off and Switch Over to Stand By.

Chamber High Chamber Temperature exceeding CHT Thermostat range.Temperature cut off Ac�on- Mains Supply Cut off, Alarm Buzzer Ac�va�on

U�lity high temperature Chamber Temperature exceeding UHT Thermostat range.cut off Ac�on- Mains Supply Cut off, Alarm Buzzer Ac�va�on

SAFETY FEATURES

CAPACITY

Complete so�ware module designed to provide seamless and comprehensive control in real �me, with g ra n u l a r l eve l d ata a n a l ys i s & t ra n s p a re nt representa�on with conforming to 21 CFR part 11 compliance requirements pertaining to Electronic Records and Electronic Signatures.

Auto Backup

Built-inLDAP

3 Tier Dura�on based equipment wise

alarm ageing A1, A2, A3

Automated Genera�on Of

Scheduled Reports.

Real Time Ac�vity

Log

Trend Analysis Reports

Models Cap. in Internal size in mm External size in mm Tray Tray size in mm Connected litres Qty. Load

AO/SC-15 150 500Wx400Dx750H 625Wx1050Dx1350H 2 405Wx400Dx30H 2.25 KW

AO/SC-25 250 700Wx400Dx900H 825Wx1050Dx1500H 2 605Wx400Dx30H 2.25 KW

AO/SC-40 400 700Wx500Dx1150H 825Wx1150Dx1750H 3 605Wx500Dx30H 2.25 KW

AO/SC-50 500 700Wx600Dx1200H 825Wx1250Dx1800H 3 605Wx600Dx30H 2.25 KW

AO/SC-60 600 700Wx600Dx1450H 825Wx1250Dx2050H 3 605Wx600Dx30H 2.25 KW

AO/SC-80 800 700Wx800Dx1450H 825Wx1450Dx2050H 4 605Wx800Dx30H 2.5 KW

AO/SC-100 1000 830Wx830Dx1450H 955Wx1480Dx2050H 4 735Wx830Dx30H 2.5 KW

Page 4: STAND ALONE STABILITY/HUMIDITY CHAMBER

B-239, MIDC Malegaon, Sinnar, Dist. Nasik, Maharashtra - 422113 (INDIA)

CORPORATE OFFICE

ALLYONE ENVIRONMENTAL TECHNOLOGIES INDIA PVT LTD

ELEM TECHNOLOGIES+91 [email protected]

Andhra Pradesh and Telangana

JABS PHARMATECH+880 01715572988 / [email protected]@jabspharmatech.com

Bangladesh

TRADE TIME INTERNATIONAL+977 [email protected]

Nepal

CHANNEL PARTNERSSERVICE SUPPORT

[email protected]+91 8767804736

[email protected]+91 8767804734

[email protected]+91 8767804733

SAMA SUPPLIES+962 [email protected]

Jordan

®