slide 1 module noise studies. slide 2 possible run conditions temperaturerelative humidityair...
TRANSCRIPT
![Page 1: Slide 1 Module Noise Studies. Slide 2 Possible Run Conditions TemperatureRelative HumidityAir Pressure Warm (20º C)High (40-45% RH)Room (1 atm) WarmLow](https://reader036.vdocuments.site/reader036/viewer/2022081907/5516020a550346cf6f8b5bcf/html5/thumbnails/1.jpg)
Slide 1
Module Noise Studies
![Page 2: Slide 1 Module Noise Studies. Slide 2 Possible Run Conditions TemperatureRelative HumidityAir Pressure Warm (20º C)High (40-45% RH)Room (1 atm) WarmLow](https://reader036.vdocuments.site/reader036/viewer/2022081907/5516020a550346cf6f8b5bcf/html5/thumbnails/2.jpg)
Slide 2
Possible Run Conditions
Temperature Relative Humidity Air Pressure
Warm (20º C) High (40-45% RH) Room (1 atm)
Warm Low (<1% RH) Room (1 atm)
Warm Low Vacuum
Cold (-20º C) Low Room
Cold Low Vacuum
Cold High Room
From what I can determine there are 5 “safe” and one possible “non-optimal” combinations of RH%, temperature, and air pressure that the modules are operated in at CERN and Liverpool. Stand conditions are not the same (grounding, LV, HV) at either site.
The non-optimal combination (Cold, high humidity, atmospheric pressure) was used study the effects of deposition of ice onto the module surface on the noise
![Page 3: Slide 1 Module Noise Studies. Slide 2 Possible Run Conditions TemperatureRelative HumidityAir Pressure Warm (20º C)High (40-45% RH)Room (1 atm) WarmLow](https://reader036.vdocuments.site/reader036/viewer/2022081907/5516020a550346cf6f8b5bcf/html5/thumbnails/3.jpg)
Slide 3
Typical Run ConditionsTemperature Relative Humidity Air Pressure
Warm (20º C) High (40-45% RH) Room (1 atm)
Warm Low (<1% RH) Room (1 atm)
Warm Low Vacuum
Cold (-20º C) Low Room
Cold Low Vacuum
Cold High Room
At CERN assembly lab, majority of tests are cold, low RH% (sub-5%), 1 atm. We can not reproduce this condition at Liverpool. This is the only state in which the noise has been seen (note: stand condition in respect to grounding/shielding have changed during these runs).
Only the state combination both sites have tested at is cold, low RH%, vacuum. This is state the vac-test is done at Liverpool on all modules and for the test-beam. Neither site has every seen noise in these conditions
All laser tests at Liverpool are performed warm, high RH%, room pressure. No noise is seen.
![Page 4: Slide 1 Module Noise Studies. Slide 2 Possible Run Conditions TemperatureRelative HumidityAir Pressure Warm (20º C)High (40-45% RH)Room (1 atm) WarmLow](https://reader036.vdocuments.site/reader036/viewer/2022081907/5516020a550346cf6f8b5bcf/html5/thumbnails/4.jpg)
Slide 4
Extreme tests at Liverpool
• We have undertaken two extreme sets of tests in order to try to reproduce the noise seen at CERN with our two test systems: With NA60, we reduced the RH% rapidly from
45% to <1% in about 1 minute. This will cause typical humidity effect to be the largest
– If effect present, expect increasing bias current and localized noise
With vac-tank, we cooled to -30º C in 45% RH air. This was to study the effects of deposited ice on the detector surfaces
– Not sure what symptoms should be expected
• Neither test should any unusual noise effects
![Page 5: Slide 1 Module Noise Studies. Slide 2 Possible Run Conditions TemperatureRelative HumidityAir Pressure Warm (20º C)High (40-45% RH)Room (1 atm) WarmLow](https://reader036.vdocuments.site/reader036/viewer/2022081907/5516020a550346cf6f8b5bcf/html5/thumbnails/5.jpg)
Slide 5
Extreme humidity tests
• For an typical humidity effect, dropping the RH% from room to low will cause the greatest noise effect It traps charge in a configuration on the surface of the sensor which causes high
fields at the corner of the implant surface• On the laser stand, we dropped the RH from ~45% to <1% in ~1 minute on
two modules No change in bias current, noise, or pedestal seen I have no idea of a more extreme test which we can do to try to reproduce the
effect
![Page 6: Slide 1 Module Noise Studies. Slide 2 Possible Run Conditions TemperatureRelative HumidityAir Pressure Warm (20º C)High (40-45% RH)Room (1 atm) WarmLow](https://reader036.vdocuments.site/reader036/viewer/2022081907/5516020a550346cf6f8b5bcf/html5/thumbnails/6.jpg)
Slide 6
Extreme ice tests I
Next, cooling cookies frostFinally, cooling outlet
frosts
• The noise seen may be due to an extremely thin layer of ice forming on the sensor/Beetle surface May act as conductive or capacitive connection between sensor strips
• We sealed vac-tank full of 45% RH air and then cooled to -30º C while taking thermal images Not clear if ice on detector surface
First, the cooling inlet frosts(lower reflectivity, lower uncalibrated temp)
![Page 7: Slide 1 Module Noise Studies. Slide 2 Possible Run Conditions TemperatureRelative HumidityAir Pressure Warm (20º C)High (40-45% RH)Room (1 atm) WarmLow](https://reader036.vdocuments.site/reader036/viewer/2022081907/5516020a550346cf6f8b5bcf/html5/thumbnails/7.jpg)
Slide 7
Extreme ice tests II
• Noise measured on 2 modules at 250 V bias Module was left cold over 20 minutes, then run for less than 1 minute
• No noise effects seen No localized high noise, no noise wings
![Page 8: Slide 1 Module Noise Studies. Slide 2 Possible Run Conditions TemperatureRelative HumidityAir Pressure Warm (20º C)High (40-45% RH)Room (1 atm) WarmLow](https://reader036.vdocuments.site/reader036/viewer/2022081907/5516020a550346cf6f8b5bcf/html5/thumbnails/8.jpg)
Slide 8
Conclusions• At Liverpool, a series of tests where taken in
order to try to reproduce noise effects seen at CERNTargeted humidity/ice effects
• We were unable to reproduce any of the effects at LiverpoolNo explanation of the cause of the noise symptoms
seen at CERN are completely consistent– Humidity, ice, pick-up, shielding, power, …
• As the cause of the effects are not understood, we feel it is extremely important to study themWe have no idea of how it will evolve, etc.We don’t know what will make the problem
better/worse