SIMPLE LOGGER ® DC MODULE L320 L410 L430 ENGLISH User Manual

Upload: others

Post on 21-Oct-2020

7 views

Category:

Documents


0 download

TRANSCRIPT

  • SIMPLE LOGGER®

    DC MODULEL320L410L430

    E N G L I S H User Manual

  • Owner’s RecordThe serial number for the Models L320, L410 and L430 is located on the side of the case. Please record this number and purchase date for your records.

    SIMPLE LOGGER DC MODULE

    MODEL #: _______________________________________________

    SERIAL #: _______________________________________________

    PURCHASE DATE: ________________________________________

    DISTRIBUTOR: ___________________________________________

  • Table of Contents

    1. INTRODUCTION ............................................................................... 41.1 International Electrical Symbols ................................................41.2 DefinitionofMeasurementCategories .....................................51.3 ReceivingYourShipment ..........................................................51.4 OrderingInformation .................................................................5

    1.4.1 Accessories and Replacement Parts ............................5

    2. PRODUCT FEATURES ...................................................................... 62.1 Indicators and Buttons ..............................................................72.2 Inputs and Outputs ...................................................................72.3 Mounting ...................................................................................7

    3. SPECIFICATIONS............................................................................. 83.1 ElectricalSpecifications ............................................................83.2 MechanicalSpecifications ........................................................93.3 EnvironmentalSpecifications ....................................................93.4 SafetySpecifications ................................................................9

    4. OPERATION .................................................................................. 104.1 Software Installation ...............................................................104.2 RecordingData .......................................................................114.3 UsingtheSoftware .................................................................11

    4.3.1 ScaleandEngineeringUnitProgramming ..................124.3.2 CreatingaLibraryofScales .......................................134.3.3 CreatingScalesforConnectedLoggers .....................14

    5. MAINTENANCE ............................................................................. 165.1 Battery Installation ..................................................................165.2 Cleaning ..................................................................................16

    APPENDIX A ....................................................................................... 17Importing.TXTFilesintoaSpreadsheet.........................................17

    OpeningaSimpleLogger.TXTfileinExcel ...........................17FormattingtheDateandTime ................................................18

  • APPENDIX B ....................................................................................... 20Time-ExtensionRecording(TXR™) ................................................20

    Repair and Calibration ...........................................................................22Technical and Sales Assistance ............................................................22Limited Warranty ...................................................................................23Warranty Repairs ...................................................................................23

  • Simple Logger® Models L320 / L410 / L430 3

    ** IMPORTANT NOTICE BEFORE OPERATING **

    A scale must be set prior to savinga downloaded DC Logger file

    for the logger to function correctly

    To create scales with the Logger connected:ThefirsttimeaDCloggerisusedwithafreshinstalloftheSimpleLoggersoftwareascalemustbesetforeachdifferentDCloggermodel.Onceascale issetforaspecificmodel, thesoftwarewilldefault to this scale each time that model is connected. To set a scaleforthemodelinuse,connecttheloggerandaScale menu will appear. Click on Scaleandthescalingwindowfor themodelusedwillappear.Fromthescalingwindowyoumaysetacustomscaleoropenapredefinedscale from theFile-Open menu. The predefinedscaleisinthedirectoryfortheSimpleLoggersoftware.

    To create scales without the Logger connected:Go to Scaling from the Filemenuandselecttherangeforthemodelused.FromthescalingwindowyoumaycreateacustomscaleandsaveitusingtheFile-SaveCommand.ThepredefinedscalesaresavedinthedirectoryfortheSimpleLoggersoftware.

    A scale may be set prior to or after downloading,but before saving.

  • 4 Simple Logger® Models L320 / L410 / L430

    CHAPTER 1

    INTRODUCTION

    WARNING Thesesafetywarningsareprovidedtoensurethesafetyofper-sonnel and proper operation of the instrument.• Read the instructionmanualcompletelyand followallsafety

    informationbeforeoperatingthisinstrument.• Usecautiononanycircuit:Potentiallyhighvoltagesandcur-

    rents may be present and may pose a shock hazard.• Readthespecificationssectionpriortousingthedatalogger.

    Neverexceedthemaximumvoltageratingsgiven.• Safetyistheresponsibilityoftheoperator.• Formaintenance,useonlyoriginalreplacementparts.• NEVERopenthebackoftheinstrumentwhileconnectedto

    any circuit or input.• ALWAYSconnecttheleadstotheloggerbeforeinsertingthe

    leadstothetestvoltage• ALWAYS inspect the instrument and leads prior to use.

    Replace any defective parts immediately.• NEVERusetheSimpleLogger® Models L320, L410, L430 on

    electricalconductorsratedabove600VCat.III.

    1.1 International Electrical SymbolsThissymbolsignifiesthattheinstrumentisprotectedbydoubleorreinforcedinsulation.Useonlyspecifiedreplacementpartswhenservicingtheinstrument.This symbol on the instrument indicates a WARNING and that the operator must refer to the user manual for instructions before operating the instrument. In this manual, the symbol precedinginstructions indicates that if the instructions are not followed, bodily injury,installation/sampleandproductdamagemayresult.Riskofelectricshock.Thevoltageat thepartsmarkedwith thissymbolmaybedangerous.

  • Simple Logger® Models L320 / L410 / L430 5

    1.2 Definition of Measurement CategoriesCat. IV: For measurements performed at the primary electrical supply

    (

  • 6 Simple Logger® Models L320 / L410 / L430

    CHAPTER 2

    PRODUCT FEATURES

    Models L410 and L430:

    Model L320:

    !INPUT

    (10

    V D

    C m

    ax)

    1

    3

    2

    (1) Start/Stop Button

    (2) Input Safety Plugs

    (3) Red LED Indicator

    (4) RS-232 Interface

    !RS-232

    PRESS

    LOGGER OPERATION:

    1. Press to Start Logging(Logging = Two Blinks)

    2. Press to Stop Logging(Stand-By = One Blink)

    3. Press & Hold 5 sec. to Shut Off(Off = No Blinks)

    Warning: All Data will be Deleted!

    4

    (1) Start/Stop Button

    (2) Input Terminal Strip

    (3) Red LED Indicator

    (4) RS-232 Interface

    !INPUT

    (25

    mA

    ma

    x, R

    in =

    10

    0Ω)

    !RS-232

    PRESS

    LOGGER OPERATION:

    1. Press to Start Logging(Logging = Two Blinks)

    2. Press to Stop Logging(Stand-By = One Blink)

    3. Press & Hold 5 sec. to Shut Off(Off = No Blinks) Warning: All Data will be Deleted!

    1

    3

    2 4

  • Simple Logger® Models L320 / L410 / L430 7

    2.1 Indicators and ButtonsTheSimpleLogger® has one button and one indicator. Both are located onthefrontpanel.ThePRESSbuttonisusedtostartandstoprecordingsandtoturntheloggeronandoff.

    TheRedLEDindicatesthestatusofthelogger:

    • Single Blink: STAND-BYmode• Double Blink: RECORD mode• Continuously On: OVERLOADcondition• No Blinks: OFFmode

    2.2 Inputs and OutputsThe left side of the Simple Logger® incorporates 4mm safety banana jacks for the Models L410 and L430 and a screw connector for the Model L320.

    Therightsideoftheloggerhasafemale9-pin“D”shellserialconnectorusedfordatatransmissionfromtheloggertoyourcomputer.

    2.3 MountingYourSimpleLogger® is equipped with clearance holes in the base plate tabsformounting.Forlesspermanentmounting,theVelcro®pads(sup-plied loose)canbeattachedto the loggerandthesurfacetowhichtheloggerwillbemounted.

  • 8 Simple Logger® Models L320 / L410 / L430

    CHAPTER 3

    SPECIFICATIONS

    3.1 Electrical SpecificationsNumber of Channels: 1

    Measurement Range: L320:0to25mADCL410:0to100mVDCL430:0to10VDC

    Input Connection: L320:twopostscrewterminalstripL410andL430:recessedsafetybananajacks

    Input Impedance: L320:100ΩL410andL430:1MΩ

    L320: 8Bit(12.5µAminresolution)

    Scale Range Maximum Input Resolution100% 25.5mA 0.1mA50% 12.75mA 0.05mA25% 6.375mA 0.025mA

    12.5% 3.1875mA 0.0125mA

    L410: 8Bit(50µVminresolution)

    Scale Range Maximum Input Resolution100% 102mV 0.4mV50% 51mV 0.2mV25% 25.5mV 0.1mV

    12.5% 12.75mV 0.05mV

    L430: 8Bit(5mVminresolution)

    Scale Range Maximum Input Resolution100% 10.2V 40mV50% 5.1V 20mV25% 2.55V 10mV

    12.5% 1.275V 5mV

    Reference condition: 23°C ± 3K, 20 to 70% RH, Frequency 50/60Hz, No AC external mag-netic field, DC magnetic field ≤ 40A/m, battery voltage 9V ± 10%.

    Accuracy: 1% ± 2cts

  • Simple Logger® Models L320 / L410 / L430 9

    Sample Rate: 4096/hrmax;decreasesby50%eachtimememoryisfull

    Data Storage: 8192readings

    Data Storage Technique: TXR™TimeExtensionRecording™

    Power:9VAlkalineNEDA1604,6LF22,6LR61

    Battery Life Recording: Upto1yearrecording@77°F(25°C)

    Output: RS-232viaDB9connector(1200Baud)

    3.2 Mechanical SpecificationsSize: 2-7/8x2-5/16x1-5/8"(73x59x41mm)

    Weight (with battery):5oz(140g)

    Mounting: BaseplatemountingholesorVelcro® pads

    Case Material:PolystyreneULV0

    3.3 Environmental SpecificationsOperating Temperature: -4to158°F(-20to70°C)

    Storage Temperature:-4to176°F(-20to80°C)

    Relative Humidity: 5to95%non-condensing

    3.4 Safety Specifications

    Working Voltage: EN61010,30VCat.III

    *All specifications are subject to change without notice

  • 10 Simple Logger® Models L320 / L410 / L430

    CHAPTER 4

    OPERATION

    4.1 Software Installation

    Minimum Computer Requirements• Windows®7, 8, 8.1, 10• Processor-486orhigher• 8MBofRAM• 8MBofharddiskspaceforapplication,400Kforeachstoredfile• One9-pinserialport;oneparallelportforprintersupport• CD-ROMdrive

    1. InserttheSimpleLogger® CD into your CD-ROM drive.If auto-run isenabled, theSetupprogramwill start automatically. Ifauto-run is not enabled, select Run from the Start menu and type inD:\SETUP (ifyourCD-ROMdrive isdriveD. If this isnot thecase,substitutetheappropriatedriveletter).

    2. The Set-up window will appear.

    Figure 1

  • Simple Logger® Models L320 / L410 / L430 11

    There are a several options to choose from. Some options(*) require an internet connection.

    • Simple Logger, Version 6.xx-InstallstheSimpleLogger® soft-ware to the computer.

    • *Acrobat Reader - Links to the Adobe® web site to download themost recent version of Adobe® Acrobat Reader. Acrobat Reader isrequired for viewing PDF documents supplied on the CD-ROM.

    • *Check for Available Software Updates - Opens the AEMCSoftware update web site, where updated software versions areavailablefordownloading,ifnecessary.

    • View User Guide and Manuals - Opens Windows®Explorerforviewingofdocumentationfiles.

    3. To install the software, select Simple Logger Software Setup inthe top section of the Set-up window, then select Simple Logger,Version 6.xx in the Options section.

    4. Click the Install button and follow the on-screen prompts to install thesoftware.

    4.2 Recording Data• Connecttheloggertothecircuittobetested.

    NOTE: Be sure to observe the polarity oryoumaynotgetareading.• PressthePRESSbuttononthetopoftheloggertobegintherecording

    session. The LED indicator will double-blink to indicate that the record-ingsessionhasstarted.

    • When thedesired recordingsessionhasfinished,press thePRESSbuttontoendtherecording.TheLEDindicatorwillsingle-blinktoindi-catetherecordingsessionhasendedandtheloggerisinStand-by.

    • Removethe logger fromthecircuitunder testandtransport it to thecomputerfordatadownloading.SeetheUserGuideontheCD-ROMfordownloadinginstructions.

    4.3 Using the SoftwareLaunch the software and connect the RS-232 cable from your computer tothelogger.

    NOTE:Alanguagewillneedtobeselecteduponthefirstlaunch.

  • 12 Simple Logger® Models L320 / L410 / L430

    Select PortfromthemenubarandselecttheComport(COM1,23or4)youwillbeusing(seeyourcomputermanual).Oncethesoftwareautomati-callydetectsthebaudrate,theloggerwillcommunicatewiththecomputer.(IDnumberoftheloggerandnumberofpointsrecordeddisplayed).Select Downloadtodisplaythegraph.(Takesabout90seconds.)Select Filefromthemenubar,thenScalingandtheRangeofyourlogger.

    4.3.1 Scale and Engineering Unit Programming

    TheSimpleLogger® Models L320, L410, and L430 allow the operator to pro-gramthevaluesforscaleandengineeringunitsfromwithinthesoftware.Thispermitstheusertodisplaytherecordeddataonthegraphorinthetabularlistingdirectly,intheunitsappropriatetothemeasurement,ratherthanmathematicallyconvertingavoltageoracurrenttotheproperscaleandvalueafterthegraphisdisplayed.

    Scales can be programmed from two locations in the software:• File menu option: Use this option to create a library of scales to use

    withDCvoltageandDCcurrent loggers.Thiswill allow theuser toselectanumberofpredefinedscales.

    • Scale menu option: Usethisoptiontocreatescalesforloggerscon-nectedtotheserialportfordownloading.

    TheSimpleLogger®softwareallowstheoperatortodefineupto17pointsalongthescaleforDCcurrentmeasurementandupto11pointsforDCvoltagemeasurementtypeloggers.Any combination of points may be used to create the scale, which allows theusertoplotbothlinearandnon-lineardata.(SeeFigures2and3).

    Figure 2 Figure 3

  • Simple Logger® Models L320 / L410 / L430 13

    4.3.2 Creating a Library of Scales• SelectFile and then Scaling from the main menu.• Selectthetypeofloggertobescaledfromthechoicespresented.• Awindowsimilar toFigure4willappearassoonasyoumakeyour

    selection. This window shows the programmable scale points andtheprogrammableunitsfield.Theleftscreenprovidesthescaleandunitprogramming,whiletherightsidedisplaystheprofileofthepro-grammedscaleinrelationshiptotheactualinputtothelogger.

    Figure 4

    Scalevaluesenteredherewillnoteffectthecurrentgraph,shouldonebeonthescreen.Thiswindowisstrictlyforcreatingtemplatestobeusedatalaterdatewithnewlydownloadedloggers.Creatingandstoringthescalesandunitsherewillsaveyoutimelateron,especiallyforfrequentlyusedscalesettings.

    Two buttons are available from within this window:• Clear All button: This will clear all scale numbers entered and any

    unitsenteredallowingyoutheopportunitytostartoveragain.

    • Close button: Returnstothemainmenuwithoutsavingthedata.

    Perform the following steps to create a Template:• Clickinanyoftheblankslotsandtypeinanumber(upto5characters)

    toenterascalevalue.Theminussignanddecimalpointcanbeusedasvalidcharacters(e.g.-10.0wouldbeavalid5characternumber).As you enter numeric data in the scale slots, the scale profile willappearonthesmallgraphintherightsideofthewindow.Bothlinearandnon-linearprofilesareacceptable.

  • 14 Simple Logger® Models L320 / L410 / L430

    • Once the scale is defined, click in the Units box to program theengineeringunitstobedisplayedonthegraph.Upto5alphanumericcharactersmaybetypedinthisbox(e.g.PSIGorGPM,etc.).

    • Afterall thedata isenteredandyouaresatisfiedwith the template,click on Fileatthetopleftsideofthedialogboxwindow.

    • Chooseoneoftheavailableoptions:- Open: Retrieves a previously stored template.- Save: Saves the current template you just created for future use.- Print: Printsacopyofthescaleandunitprogrammingwindowas

    seen on the screen.

    4.3.3 Creating Scales for Connected Loggers

    • ConnecttheSimpleLogger® to the computer’s serial port for download-ing.Seemainmanualfordownloadinginstructions.

    • Once theproper port is selected, datawill appear in theupdate boxat the top righthandsideof thescreen.This isan indication that thesoftwarehasmadeconnectiontothelogger.TheScale command willalso appear on the task bar if the logger detected allows scale andengineeringunitprogramming.

    • AscreensimilartoFigure5willappear.Theleftsideofthescreenpro-videsthescaleandunitprogramming,whiletherightsidedisplaystheprofileoftheprogrammedscaleinrelationshiptotheactualinputtothelogger.

    Figure 5

  • Simple Logger® Models L320 / L410 / L430 15

    • Theoperatormaysetthescalebyprogrammingasfewastwopoints,thelowendandthehighend,orbyenteringasmanypointsasnecessarytodefinethescaleupto17pointsforthe4-20mAloggerandupto11pointsfortheDCvoltageloggers.ThepointsentereddonothavetobelinearbutshouldbeanaccuraterepresentationoftherelationshipoftheDCsignalto the scale points.

    • Toenterascalevalueinanyoftheslots,clickontheslotandtypeinanumberupto5characters.Theminussignanddecimalpointmaybeusedasvalidcharacters(e.g.-25.4wouldbeavalid5-characternumber).

    • Once the scale is defined, click in the Unit box to program theengineeringunitstobedisplayedonthegraph.Upto5alphanumericcharactersmaybetypedinthisbox.

    • Onceyouhaveenteredthecorrectscaleandunitdata,clickonOKtoproceed.ThescreeninFigure6willappeargivingyoutheopportunitytosavetheentereddataforfutureuse.ClickonYestosavethedataorNotobypasssavingthedataanduseonetimeonly.

    • IfyouclickYes,adialogboxwillopensimilartoFigure7whereyoucantypeinthename(upto8characters)youwishtouseforthefile.

    • ClickonOKtosavethefileandplotthegraphwiththenewscaleandunit data or click on Cancel to discard it and return to the scale and unitprogrammingscreen.

    Figure 6

    Figure 7

  • 16 Simple Logger® Models L320 / L410 / L430

    CHAPTER 5

    MAINTENANCE

    5.1 Battery InstallationUnder normal conditions, the battery will last up to a year of continuous recordingunlesstheloggerisrestartedveryfrequently.IntheOFFmode,theloggerputsalmostnoloadonthebattery.UsetheOFFmodewhentheloggerisnotinuse.Replacethebatteryonceayearin normal use.Iftheloggerwillbeusedattemperaturesbelow32°F(0°C)orisfrequentlyturnedonandoff,replacethebatteryeverysixtoninemonths.

    1. Makesureyourloggeristurnedoff(noblinkinglight)andallinputsare disconnected.

    2. Turntheloggerupsidedown.RemovethefourPhillipsheadscrewsfrom the base plate, then take off the base plate.

    3. Locatethetwo-wire(red/black)batteryconnectorandattachthe9Vbatterytoit.Makesurethatyouobservepolaritybyliningupthebattery posts to the proper terminals on the connector.

    4. Oncetheconnectorispluggedontothebattery,insertthebatteryintotheholdingcliponthecircuitboard.

    5. Iftheunitisnotinrecordmodeafterinstallingthenewbattery,dis-connect it and press the button twice then reinstall the battery.

    6. ReattachthebaseplateusingthefourscrewsremovedinStep2.

    Yourloggerisnowrecording(LEDblinking).PressthePRESS button for fivesecondstostoptheinstrument.

    NOTE:For long-termstorage, remove thebattery topreventdis-chargeeffects.

    5.2 CleaningThe body of the logger should be cleanedwith a clothmoistenedwithsoapy water. Rinse with a cloth moistened with clean water. Do not use solvent.

  • Simple Logger® Models L320 / L410 / L430 17

    APPENDIX A

    Importing .TXT Files into a Spreadsheet

    Opening a Simple Logger .TXT file in Excel

    ThefollowingexampleusedwithExcelVer.7.0orhigher.

    1. After opening the Excel program, select “File” from the mainmenu and then select “Open”.

    2. In the dialog box that appears, browse and open the folderwhere your logger .TXT files are stored. This will be locatedin C:\Program Files\Simple Logger 6.xx if you accepted thedefaultchoiceofferedbytheloggerinstallationprogram.

    3. Next,changethefiletypeto“Text Files”inthefieldlabeledFilesof Type.Allthe.TXTfilesintheloggerdirectoryshouldnowbevisible.

    4. Double-clickonthedesiredfiletoopentheTextImportWizard.

    5. Reviewtheselectionsinthefirstwizardscreenandmakesurethatthefollowingchoicesareselected:

    OriginalDataType: DelimitedStartImportatRow: 1FileOrigin: Windows(ANSI)

    6. Click the “NEXT”buttonatthebottomoftheWizarddialogbox.The second wizard screen will appear.

    7. Click on “Comma” intheDelimitersbox.Acheckmarkshouldappear.

    8. Click the “NEXT” buttonatthebottomoftheWizarddialogbox.The third wizard screen will appear.

    9. A view of the actual data to be imported should appear in thelowersectionofthewindow.Column1shouldbehighlighted.IntheColumnDataFormatwindow,select“Date”.

    10. Next,clickon“Finish” to complete the process and import thedata.

    11. Thedatawillnowappearinyourspreadsheetintwocolumns(AandB)andwilllooksimilartothatshowninTableA-1.

  • 18 Simple Logger® Models L320 / L410 / L430

    A B

    8 Arms 35401.49 3.5 35401.49 5 35401.49 9 35401.49 13.5 35401.49 17 35401.49 20 35401.49 23.5 35401.49 27.5 35401.49 31 35401.49 34.5 35401.49 38

    Table A-1 - Sample Data Imported into Excel.

    Formatting the Date and Time

    Column ‘A’ contains a decimal number that represents both date and time. Excelcanconvertthisnumberdirectlyasfollows:

    1. Click on column ‘B’ at the top of the column to select the data,then click on “Insert” from the main menu and select “Columns”from the drop-down menu.

    2. Next,clickoncolumn‘A’atthetopofthecolumntoselectthedata,thenclickon“Edit” from the main menu and select “Copy”to copy the entire column.

    3. Click on cell 1 of column ‘B’ and then click on “Edit” and select“Paste” to insert a duplicate of column ‘A’ into column ‘B’. This isnecessary if you want to show the date and time in two separatecolumns.

    4. Next,clickonthetopofcolumn‘A’,thenclickon“Format” andselect “Cells” from the drop-down menu.

    5. Inthedialogboxthatopens,selectthe“Date” option from thecategorylistontheleft.Selectthedateformatyoudesireandclick on “OK” to format the column.

  • Simple Logger® Models L320 / L410 / L430 19

    6. Click on the top of column ‘B’, then click on “Format” and select“Cells” from the drop-down menu.

    7. Inthedialogboxthatopens,selectthe“Time” option from thecategorylistontheleft.Selectthetimeformatyoudesireandclick on “OK” to format the column.

    Table A-2 shows a typical spreadsheet with date, time and value displayed. Itmaybenecessarytochangethecolumnwidthtoseeallthedata.

    A B C

    12/02/04 11:45AM 1712/02/04 11:45AM 2012/02/04 11:45AM 23.512/02/04 11:45AM 27.512/02/04 11:45AM 3112/02/04 11:45AM 34.512/02/04 11:45AM 3812/02/04 11:45AM 41.512/02/04 11:45AM 45.512/02/04 11:46AM 4912/02/04 11:46AM 52

    Table A-2 - Shows Date, Time and Value

  • 20 Simple Logger® Models L320 / L410 / L430

    APPENDIX B

    Time-Extension Recording (TXR™)

    Timeextensionrecordingisanautomaticprocessthatupdatesthesamplerate and the number of stored data points based on the length of therecording.Themaximumnumberofstoreddatapointsis8192.

    When thedata logger starts a new recording session, it does soat itsfastestsamplerateof4096pointsperhour(0.88secondsperpoint).TheSimpleLoggers®canrecordatthisratefortwohours.Iftherecordingses-sioncontinuesbeyondtwohours,thetimeextensionrecordingtechniquebecomes active.

    Beginningwiththesample,afterthecompletionoftwohoursofrecording,theloggercontinuesrecordingbyselectivelyoverwritingpreviouslystoreddata.TheSimple Logger® also halves its sample rate to 2048/hr (1.76secondsperpoint) for thenewstoredvalues tobecompatiblewith thepreviously recorded values.

    Recording continues for the next two hours at this new rate until theremaining4096storagepointsarefilled.

    2048

    11st 2 hrs.

    2nd 2 hrs.

    2 3 4 5 6 4096 4097 8191 8192

    1 2 3 4096 4097 4098 8192

    Table B-1 - Time Extension Memory Allocation

    Thetimeextensionrecordingprocessofselectivelyoverwritingpreviouslystoreddataandhalving thesample rate fornewstoreddatacontinueseverytimethememoryfillsup.TableB-2showstherelationshipbetweenrecordingtimeandsamplerateforthedataloggerusingthistechnique.

  • Simple Logger® Models L320 / L410 / L430 21

    Sample Rateper hr.

    SecondsPer Sample

    TotalRecordingTime (hrs)

    TotalRecordingTime (days)

    4096 0.88 2 0.0832048 1.76 4 0.1671024 3.52 8 0.333512 7.04 16 0.667256 14.08 32 1.333128 28.16 64 2.66764 56.32 128 5.33332 112.64 256 10.66716 225.28 512 21.3338 450.56 1024 42.6674 901.12 2048 85.3332 1802.24 4096 170.6671 3604.48 8192 341.333

    Table B-2 - Sample Rate vs Recording Time

    Recordingcontinuesinthismanneruntilthebatterywearsoutorrecord-ing isstopped.Forconvenience indataanalysis, the recording intervaltakesonvaluesoffifteenminutes,one-halfhour,onehourandsoon.

    Likeautomaticscaling,timeextensionrecordingispracticallyinvisibletotheuser.Forbestresults,settheloggertostandbymodewhenthemea-surement iscompleted,both toavoid includingunwanted information intheplot,andtoprovidemaximumresolutionfortheperiodofinterest.

  • 22 Simple Logger® Models L320 / L410 / L430

    Repair and Calibration

    Toensurethatyourinstrumentmeetsfactoryspecifications,werecommendthat it be scheduled back to our factory Service Center at one-year intervals for recalibration, or as required by other standards or internal procedures.

    For instrument repair and calibration:Youmust contactourServiceCenter foraCustomerServiceAuthorizationNumber(CSA#).Thiswillensurethatwhenyourinstrumentarrives,itwillbetrackedandprocessedpromptly.PleasewritetheCSA#ontheoutsideoftheshipping container. If the instrument is returned for calibration,weneed toknow if you want a standard calibration, or a calibration traceable to N.I.S.T. (Includescalibrationcertificateplusrecordedcalibrationdata).

    Ship To: ChauvinArnoux®, Inc. d.b.a. AEMC® Instruments15FaradayDriveDover, NH 03820 USAPhone:(800)945-2362(Ext.360)

    (603) 749-6434(Ext.360)Fax: (603)742-2346or(603)749-6309E-mail:[email protected]

    (Orcontactyourauthorizeddistributor)Costs for repair, standard calibration, and calibration traceable to N.I.S.T. are available.NOTE: You must obtain a CSA# before returning any instrument.

    Technical and Sales Assistance

    Ifyouareexperiencinganytechnicalproblems,orrequireanyassistancewiththeproperoperationorapplicationofyourinstrument,pleasecall,mail,faxore-mailourtechnicalsupportteam:

    ChauvinArnoux®, Inc. d.b.a. AEMC® Instruments200FoxboroughBoulevardFoxborough,MA02035USAPhone:(800)343-1391

    (508) 698-2115Fax: (508)698-2118E-mail: [email protected]

    NOTE: Do not ship Instruments to our Foxborough, MA address.

  • Simple Logger® Models L320 / L410 / L430 23

    Limited Warranty

    The Simple Logger® Model L320/L410/L430 is warranted to the owner for a period of two years from the date of original purchase againstdefects in manufacture. This limited warranty is given by AEMC® Instruments, not by the distributor from whom it was purchased. This warranty is void if the unit has been tampered with, abused or if the defect is related to service not performed by AEMC® Instruments.

    For full and detailed warranty coverage, please read the Warranty Coverage Information, which is attached to the Warranty Registration Card (if enclosed) or is available at www.aemc.com. Please keep the Warranty Coverage Information with your records.

    What AEMC® Instruments will do:If a malfunction occurs withing the warranty period, you may return the in-strument to us for repair, provided we have your warranty registrationinformation on file or a proof of purchase. AEMC® Instruments will, at its option, repair or replace the faulty material.

    REGISTER ONLINE AT:www.aemc.com

    Warranty Repairs

    What you must do to return an Instrument for Warranty Repair: First, request aCustomerServiceAuthorizationNumber (CSA#) by phoneorbyfaxfromourServiceDepartment(seeaddressbelow),thenreturntheinstrumentalongwith thesignedCSAForm.Pleasewrite theCSA#on theoutsideoftheshippingcontainer.Returntheinstrument,postageorshipmentpre-paidto:

    Ship To: ChauvinArnoux®, Inc. d.b.a. AEMC® Instruments15FaradayDrive•Dover,NH03820USAPhone:(800)945-2362(Ext.360)

    (603) 749-6434(Ext.360)Fax: (603)742-2346or(603)749-6309E-mail:[email protected]

    Caution:Toprotectyourselfagainstin-transitloss,werecommendyouinsureyour returned material.

    NOTE: You must obtain a CSA# before returning any instrument.

  • 24 Simple Logger® Models L320 / L410 / L430

    Notes:

  • 08/18

    99-MAN 100212 v14

    Chauvin Arnoux®, Inc. d.b.a. AEMC® Instruments15FaradayDrive•Dover,NH03820USA•Phone:(603)749-6434•Fax:(603)742-2346

    www.aemc.com

    CREATING SCALES IMPORTANT NOTEINTRODUCTIONPRODUCT FEATURES2.1Indicators and Buttons2.2Inputs and Outputs2.3Mounting

    SPECIFICATIONS3.1Electrical Specifications3.2Mechanical Specifications3.3Environmental Specifications3.4Safety Specifications

    OPERATION4.1Software Installation4.2Recording Data4.3Using the Software4.3.1Scale and Engineering Unit Programming4.3.2Creating a Library of Scales4.3.3Creating Scales for Connected Loggers

    MAINTENANCE5.1Battery Installation5.2Cleaning

    APPENDIX AImporting .TXT Files into a SpreadsheetOpening a Simple Logger .TXT file in ExcelFormatting the Date and Time

    APPENDIX BTime-Extension Recording (TXR™)

    Repair and CalibrationTechnical and Sales AssistanceLimited WarrantyWarranty Repairs