side by side evaluation of high performance fume hoods · side by side evaluation of high...
TRANSCRIPT
![Page 1: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/1.jpg)
SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS
2008 Labs21 Annual ConferenceSan Jose, CA
Presented by:Bernard Bhatti, University of Texas at Austin
Kevin Fox, Jacobs Engineering Group
![Page 2: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/2.jpg)
Understand Project PurposeDefinition of High Performance Fume HoodFume Hood Energy PerformanceUnderstand Fume Hood Testing MethodologiesCompare Hood Containment Performance under different operating conditionsIdentify additional Fume Hood Performance ChallengesLife Cycle Cost Comparisons
Learning Objectives
![Page 3: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/3.jpg)
Campus BackgroundUT Campus Info− 16 Colleges− 50,000 students (39,000 undergrad)− 350-acre campus, 16 Million SF− Research: More than 3,500 projects, annual
funding > $400 million. − Fume Hood Users
College of Natural Sciences College of EngineeringSchool of GeosciencesCollege of Pharmacy
![Page 4: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/4.jpg)
Experimental Sciences BuildingProgram conducted to determine optimal fume hood design standard for use in the new Experimental Sciences BuildingESB designed to Labs21 performance guidelines for energy efficiency and sustainability. Over 100 new fume hoods plannedExpand program to campus retrofit analysis
![Page 5: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/5.jpg)
UT Campus Energy Master PlanDSEMC Annual Savings Targets− Lighting – 75M kWh− Steam – 26M Lbs− Water – 60M gal
Campus Utilities− 100% Self-generating− 120 MW electrical capacity− 40,800 Tons chilled water capacity− 1,150,000 PPH steam capacity− Largest university CHP plant in the nation
![Page 6: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/6.jpg)
DefinitionHigh Performance Fume Hood− A fume hood designed for superior fume capture
performance with face velocities of 60 FPM or less with a fully open sash
Why High Performance?− Labs are often the most demanding energy users
of any facility on a University campus− Size and quantity of hoods can dictate total lab air
volume requirements − Energy savings (w/o sacrificing safety!)
One of Labs21 “Five Big Hits”− Face Velocity ≠ Safety (ANSI Z9.5-2003)
![Page 7: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/7.jpg)
Energy PerformanceOver 1,000 fume hoods on campusCampus standard - 100 FPM face velocityMajority of UT hoods are constant volumeHood Energy Savings (Standard vs. HP)− Exhaust & Supply Fan Energy− Makeup Air− Reheat
UT Austin energy cost = $4.60/Hood CFM5 foot hood, 625 CFM (std) vs. 500 CFM (HP)$575 Annual Savings/Hood, >$500K Annually
![Page 8: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/8.jpg)
Hood Evaluation ProcessMatrix – 8 manufacturers offering “HP” hoodsSelection of Hoods (UT criteria)− Face Velocity− Overall Depth− Static Pressure− Available Sizes− Maintenance
Three HP Hoods Tested− Hoods A, B, C (60 FPM)− Standard Hood (100 FPM)
Evaluate Safety, then EnergyTesting Facility – J.J. Pickle Research Campus− Center for Energy & Environmental Research
Testing Firm - ENV Services, San Antonio
![Page 9: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/9.jpg)
Evaluation MethodologySafety Testing− Test static and dynamic
containment performance using Owner-developed testing methods adapted from similar established testing procedures used previously at University of Wisconsin at Madison and Lawrence Berkeley National Laboratory.
ASHRAE 110 (standard)ASHRAE 110 (modified)Human as Mannequin (LBL-inspired)
Provide data to allow objective evaluation by EHS staff
![Page 10: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/10.jpg)
Evaluation MethodologyASHRAE 110-Standard (As Installed- 4.0AI0.10)− Face Velocity Measurement (12 readings)− Local smoke test− Large volume smoke test− Tracer gas containment
4 LPM SF6Empty hoodFull open sashSash Movement Effect (SME) 67” mannequin height (breathing zone 26” above work surface)
![Page 11: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/11.jpg)
Evaluation MethodologyASHRAE 110-Modified (As Installed- 8.0AI0.10)− 8 LPM SF6 (2X ASHRAE 110)− 18” and 26” mannequin breathing zone heights
(above work surface)− 30 & 75 FPM Cross Drafts− Sash Movement Effect
18” & 26” heights30 & 75 FPM cross drafts
− Space Pressure Effect 18” & 26” heights30 & 75 FPM cross drafts
![Page 12: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/12.jpg)
Evaluation MethodologyHuman as Mannequin (HAM) (4.0AI0.10)− 4 LPM SF6− Wide open sash− 3 ejector positions (right, center, left)− Breathing zone tracer gas detector on tester− Apparatus in hood− Choreographed movements
Hand insertion and removalMove objects front to backMove objects right to leftRemove object from hood
![Page 13: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/13.jpg)
Containment PerformanceStandard ASHRAE test (3 ejector positions)Modified ASHRAE test (3 ejector positions)− 30 FPM Cross Draft – 18” & 26” heights− 75 FPM Cross Draft – 18” & 26” heights− SME: 30 FPM Cross Draft – 18” height− SPE: 30 FPM Cross Draft – 18” height
HAM (3 ejector positions)− Hand Insertion− Front to Back− Right to Left− Water Pour− Remove Object
* Focus on containment excursions > 0.10 PPM, in keeping with ANSI Z9.5 for As Installed (AI) condition, for comparison purposes
** All tests conducted with fully open sash, to set standard for “sash abuse”
![Page 14: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/14.jpg)
Standard ASHRAE Test10 50 90
130
170
210
250
290
30 70
110
150
190
230
270
10 50 90
130
170
210
250
290
HOO
0.00
0.10
0.20
0.30
0.40
0.50
PPM
t (sec)
HOOD AHOOD BHOOD CSTD HOOD
LEFT CENTER
RIGHT
0.10 PPM
• ASHRAE test simulates idealized conditions• All smoke visualization tests passed
![Page 15: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/15.jpg)
Modified – 30 FPM CD, 18” MH10 50 90
130
170
210
250
290
30 70
110
150
190
230
270
10 50 90
130
170
210
250
290
HOO
0.00
0.10
0.20
0.30
0.40
0.50
PPM
t (sec)
HOOD AHOOD BHOOD CSTD HOOD
0.10 PPM
LEFT EJECTOR CENTER EJECTORRIGHT EJECTOR
![Page 16: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/16.jpg)
Modified - 30 FPM CD, 26” MH10 50 90
130
170
210
250
290
30 70
110
150
190
230
270
10 50 90
130
170
210
250
290
HOOD
0.00
0.10
0.20
0.30
0.40
0.50
PPM
t (sec)
HOOD AHOOD BHOOD CSTD HOOD
0.10 PPM
LEFT EJECTORCENTER EJECTOR
RIGHT EJECTOR
![Page 17: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/17.jpg)
Modified – 75 FPM CD, 18” MH10 50 90
130
170
210
250
290
30 70
110
150
190
230
270
10 50 90
130
170
210
250
290
HOO
0.00
0.10
0.20
0.30
0.40
0.50
PPM
t (sec)
HOOD AHOOD BHOOD CSTD HOOD
0.10 PPM
LEFT EJECTORCENTER EJECTOR RIGHT EJECTOR
![Page 18: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/18.jpg)
Modified – 75 FPM CD, 26” MH10 50 90
130
170
210
250
290
30 70
110
150
190
230
270
10 50 90
130
170
210
250
290
HO
0.00
0.10
0.20
0.30
0.40
0.50
PPM
t (sec)
HOOD AHOOD BHOOD CSTD HOOD
0.10 PPM
LEFT EJECTORCENTER EJECTOR
RIGHT EJECTOR
![Page 19: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/19.jpg)
Modified – SME, 30 FPM CD, 18” MH10 30 50 70 90
110
130
150
170
190
210
230
250
270
290
HOOST
0.00
0.10
0.20
0.30
0.40
0.50
PPM
t (sec)
HOOD AHOOD BHOOD CSTD HOOD
0.10 PPM
SASH
OPE
N
SASH
CLO
SED
SASH
OPE
N
SASH
OPE
N
SASH
OPE
N
SASH
OPE
N
SASH
CLO
SED
SASH
CLO
SED
SASH
CLO
SED
SASH
CLO
SED
![Page 20: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/20.jpg)
Modified – SPE, 30 FPM CD, 18” MH10 30 50 70 90
110
130
150
170
190
210
230
250
270
290
HOODSTD
0.00
0.10
0.20
0.30
0.40
0.50
PPM
t (sec)
HOOD AHOOD BHOOD CSTD HOOD
0.10 PPM
DO
OR
OPE
N
DO
OR
CLO
SED
DO
OR
OPE
N
DO
OR
OPE
N
DO
OR
OPE
N
DO
OR
OPE
N
DO
OR
CLO
SED
DO
OR
CLO
SED
DO
OR
CLO
SED
![Page 21: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/21.jpg)
HAM Test – Hand Insertion0:
00:0
00:
00:0
90:
00:1
80:
00:2
70:
00:3
60:
00:4
5
0:00
:54
0:01
:03
0:01
:12
0:00
:05
0:00
:14
0:00
:23
0:00
:32
0:00
:41
0:00
:50
0:00
:59
0:01
:08
0:00
:01
0:00
:10
0:00
:19
0:00
:28
0:00
:37
0:00
:46
0:00
:55
0:01
:04
0:01
:13
HOO
0.00
0.10
0.20
0.30
0.40
0.50
PPM
t (sec)
HOOD AHOOD BHOOD CSTD HOOD
LEFT EJECTOR CENTER EJECTORRIGHT EJECTOR
0.10 PPM
![Page 22: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/22.jpg)
HAM Test – Move Front to Back0:
00:0
00:
00:0
90:
00:1
80:
00:2
70:
00:3
60:
00:4
50:
00:5
4
0:01
:03
0:01
:12
0:00
:05
0:00
:14
0:00
:23
0:00
:32
0:00
:41
0:00
:50
0:00
:59
0:01
:08
0:00
:01
0:00
:10
0:00
:19
0:00
:28
0:00
:37
0:00
:46
0:00
:55
0:01
:04
0:01
:13
HO
0.00
0.10
0.20
0.30
0.40
0.50
PPM
t (sec)
HOOD AHOOD BHOOD CSTD HOOD
0.10 PPM
LEFT EJECTORCENTER EJECTOR
RIGHT EJECTOR
![Page 23: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/23.jpg)
HAM Test – Move Side to Side0:
00:0
00:
00:0
90:
00:1
80:
00:2
70:
00:3
6
0:00
:45
0:00
:54
0:01
:03
0:01
:12
0:00
:05
0:00
:14
0:00
:23
0:00
:32
0:00
:41
0:00
:50
0:00
:59
0:01
:08
0:00
:01
0:00
:10
0:00
:19
0:00
:28
0:00
:37
0:00
:46
0:00
:55
0:01
:04
0:01
:13
HOO
0.00
0.10
0.20
0.30
0.40
0.50
PPM
t (sec)
HOOD AHOOD BHOOD CSTD HOOD
0.10 PPM
LEFT EJECTORCENTER EJECTOR
RIGHT EJECTOR
![Page 24: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/24.jpg)
HAM Test – Remove Object0:
00:0
00:
00:0
90:
00:1
80:
00:2
70:
00:3
60:
00:4
5
0:00
:54
0:01
:03
0:01
:12
0:00
:05
0:00
:14
0:00
:23
0:00
:32
0:00
:41
0:00
:50
0:00
:59
0:01
:08
0:00
:01
0:00
:10
0:00
:19
0:00
:28
0:00
:37
0:00
:46
0:00
:55
0:01
:04
0:01
:13
HOO
0.00
0.10
0.20
0.30
0.40
0.50
PPM
t (sec)
HOOD AHOOD BHOOD CSTD HOOD
0.10 PPM
LEFT EJECTOR CENTER EJECTORRIGHT EJECTOR
![Page 25: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/25.jpg)
Safety Performance Summary
AI test containment, red values exceeded AI 0.10
TEST CONDITIONCROSS
DRAFT, FPMMANNEQUIN HEIGHT, IN HOOD A HOOD B HOOD C HOOD D
STANDARD ASHRAE 110 STANDARD 0 26 0.00 0.00 0.00 0.00STANDARD ASHRAE 110 SME 0 26 0.04 0.03 0.00 1.25
MODIFIED ASHRAE 110 CROSS DRAFT 30 18 0.04 0.09 0.69 0.36MODIFIED ASHRAE 110 CROSS DRAFT 30 26 0.03 0.00 0.03 0.11MODIFIED ASHRAE 110 CROSS DRAFT 75 18 0.10 1.65 0.29 1.62MODIFIED ASHRAE 110 CROSS DRAFT 75 26 0.06 0.28 0.34 0.07MODIFIED ASHRAE 110 SME 30 18 0.12 0.10 3.03 0.00MODIFIED ASHRAE 110 SME 30 26 0.05 0.35 0.01 0.00MODIFIED ASHRAE 110 SME 75 18 0.05 3.93 0.29 0.99MODIFIED ASHRAE 110 SME 75 26 0.13 0.04 0.00 1.67MODIFIED ASHRAE 110 SPE 30 18 0.99 1.97 3.92 0.46MODIFIED ASHRAE 110 SPE 30 26 0.08 0.36 0.00 0.01MODIFIED ASHRAE 110 SPE 75 18 0.18 3.93 0.15 0.12MODIFIED ASHRAE 110 SPE 75 26 0.09 0.73 2.85 0.08
HAM HAND INSERTION 0 28 0.01 0.00 0.00 0.00HAM MOVE FRONT TO BACK 0 28 0.00 0.02 0.01 0.01HAM MOVE LEFT TO RIGHT 0 28 0.00 0.00 0.00 0.00HAM WATER POUR 0 28 0.01 0.01 0.00 0.01HAM OBJECT REMOVAL 0 28 0.03 0.00 0.06 0.07
OVERALL AVERAGE 0.11 0.71 0.61 0.36AS INSTALLED AVERAGE 0.03 0.21 0.14 0.23
MAXIMUM RECORDED PPM 0.99 3.93 3.92 1.67
AS INSTALLED READINGS (AI)†
![Page 26: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/26.jpg)
Safety Performance ConclusionsConclusions from safety testing:− Cross drafts challenge hood containment, must be
dealt with in design (architects and engineers must work together to mitigate!)
− 18” breathing zone height proved more challenging for containment performance
− Standard ASHRAE test may not predict performance in actual operating conditions
− Not all high performance hoods are equal!− Newer hoods performed better than “standard” 100
FPM hood
![Page 27: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/27.jpg)
Other ConsiderationsUnderstand advertised face velocityDifferences between manufacturers:− Dimensions− Static Pressure Requirements− Containment Strategies
Building System Infrastructure (fan capacity)Controls (VAV vs. CV)Trade-off between hood density and system capacityIs a fume hood required?Consider lab air change rates!
![Page 28: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/28.jpg)
Life Cycle Cost PerformanceLCCA for prototype lab between 3 options:− Standard CV, 100 FPM (18” sash)− High Performance CV, 80 FPM (18” sash)− VAV (50% sash open, 50% sash closed)
Inputs:− 15 year term− $0.077/kWh− $0.89/Therm - CHW− $0.91/Therm - Natural Gas− 6% discount rate, 3% escalation
BLCC5 program, NIST Office of Applied Economics (http://www1.eere.energy.gov/femp)
![Page 29: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/29.jpg)
Life Cycle Cost Performance
$-
$20,000
$40,000
$60,000
$80,000
$100,000
$120,000
$140,000
$160,000
$180,000
STD HOOD HP HOOD VAV HOOD
CAPITAL COST O&M ENERGY
![Page 30: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/30.jpg)
Conclusions
Safety− High performance fume hoods can operate
safely− Face velocity is not an indicator of safety− Cross drafts can compromise hood
performance− Not all hoods are equal
![Page 31: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/31.jpg)
Conclusions
Energy− High performance hoods offer energy savings
over standard hoods− Well suited for retrofit/replacement projects− High performance hoods yield best LCC
performance− Lab ACH drives energy cost more than hood
CFM
![Page 32: SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS · SIDE BY SIDE EVALUATION OF HIGH PERFORMANCE FUME HOODS 2008 Labs21 Annual Conference San Jose, CA ... yTesting Firm - ENV](https://reader033.vdocuments.site/reader033/viewer/2022053007/5f0b376b7e708231d42f6bd5/html5/thumbnails/32.jpg)
QUESTIONS?
Bernard Bhatti, P.E.The University of Texas at [email protected]
(512) 475-8476
Kevin Fox, P.E., CEM, LEEDJacobs Engineering Group
[email protected](817) 347-7600