Choosing the right inlet liner and injection parameter can increase peak areas and reduce detection limits by up to 300%1. Customer research conducted by SGE found that a significant number of GC users don’t understand the importance of inlet liner selection, or how it contributes to their analysis. The SGE inlet liner range aims to make it simple for all gas chromatographers to select the right liner.

TRANSCRIPT

Spot the difference in the SGE lineup!

SGE GC inlet liner range:

•Easytochoose.•Easytouse.•Confidenceinyouranalysis.

1.Kende,Aetal.Chromatographia,2006;63(3/4):181-7

Easy to chooseForeaseofselectionsoyoucanoptimizeyourresults,SGEhascolorcodedthelinersbygeometry.Colorcodingandsampletypesarelistedinthetablebelow.

Color Injection Technique

Sample Types Liner Geometry

Dar

k G

reen

Splitless•Tracelevelanalyses.

•Activecompounds.

Taper / Gooseneck

Blu

e

Split

•Generalpurpose.

•Concentratedsamples.

•Dirtysamples.

FocusLiner™

Aq

ua

Splitless

•Tracelevelanalyses.

•Dirtysamples.

•Wideboilingpointrange.

Taper Focus

Ora

ng

e

Direct•Tracelevelanalyses.

•Activecompounds.

ConnecTite™

Purp

le

SplitSplitless

•Generalpurpose.

•Concentratedsamples.

•Dirtysamples(onlyifquartzwool is present)

•Gaseous samples (also Purge andTrap,Headspace).

Straight

Yel

low

SplitlessLVI

•Tracelevelanalyses.

•Lowboilingpointcompounds.

•Activecompounds.

Double Taper

Gre

y

PTVLVI

•Tracelevelanalyses.

•Largevolumeinjections.

PTV/LVI

Easy to choose•Color coded by geometry to

simplifyyourselection.

Easy to use•Containo-ringssoyou’reready

togo.

Confidence in your analysis•Certified deactivation gives you

confidenceinyouranalysis.

Spot the difference in the SGE lineup!

Choosing the right inlet liner and injection parameter can increase peakareasandreducedetectionlimitsbyupto300%1.

Customer research conducted by SGE found that a significant number ofGCusersdon’tunderstandtheimportanceofinletlinerselection,orhowitcontributestotheiranalysis.TheSGEinletlinerrangeaimstomakeitsimpleforallgaschromatographerstoselecttherightliner.

SGE has also created a liner selection tool that can be found on our website: www.sge.com/linertool

Whetherforroutineanalysis,difficultprobesortraceanalysishaveconfidenceinyouranalysiswithSGE’sinletliners.Whilebottomtaperwithwoolisconsideredthebestgeometryfortraceanalysis,theSGEFocusLiner™ delivers optimal performanceinallapplications.

Confidence in your analysis

SGE standard inlet liner deactivation

SGEtestseverybatchofinletlinersforactivityusingtheEPA8081Bmethod.Thisstandardmethodensuresthateachbatchofinletlinershaslessthana3%Endrinbreakdown(5ppmstandard).SGEvalidatesthisqualityassurancebyincludingabatchcertificatewitheverypack.

5 7 9 11 13 15 17 19 21 23

InjectionNoinjection–gasflowfor5min

InjectorTemp 350 °C Purgeflow 20 ml/min

Ovenprogram50 °C to 350 °C (20/min)hold2.2mins

Carrier gas He(1.2ml/min)Detector(MS) 50-650mzInstrument Agilent 6890

MS scan for Siloxane bleed - thick vs thin film deactivation

Confidence in your trace analysis

14000000

12000000

10000000

8000000

6000000

4000000

2000000

0

SGE

Brand X

Brand Y

• Low siloxane bleed due toSGE’sthinfilmdeactivation.

Liner Comparison -0.5 ng 2, 4-dinitrophenol

Instrument: 5973inertMSDand6890GC

Oven:40 °C to 80 °C at 10 °C a minute 20 °C a minuteto190°Choldfor2minutes30°Caminuteto350°Choldforoneminute

Inlet: 4.9psi200°CHeColumn: 30mBPX50.25mmx0.25µm(PN:054101)GCMS: Scan 50 to 550InjectionVolume:

1µL

TestMix:CustomSVOCstandardmixfromAccuStandard S-23011 2000 ug/mL (diluted to0.5ppm)

Confidence with difficult probes

• Reproducibleperformancewiththerightchoiceinlinergeometry.

-10000

40000

90000

140000

190000

240000

10.12 10.14 10.16 10.18 10.2 10.22 10.24 10.26 10.28 10.3

Time (min)

SGE Tapered Focus 1

SGE Tapered Focus 2

SGE Tapered Focus 3

BrandXBottomTaperwithwool1

BrandXBottomTaperwithwool2

BrandXBottomTaperwithwool3

Confidence for routine analysis - injection after injection

50 ppb 100 Repeat injections comparison Endrin and DDT % breakdown

Testing Parameters

50 ppm Endrin and DDT test mix1µLSplitlessinjection@250°CHT grade Septa (PN: 041898)

• SGE’s Tapered FocusLiner™ shows improved performancecompared with competitor’spremiumdeactivatedliner.

• ChoosetheTaperedFocusLiner™ to ensure your analysis is uncompromised injection afterinjection.

0

5

10

15

20

25

30

35

0 20 40 60 80 100

%B

reak

dow

n

NumberofInjections

BrandXEndrinKetoneandEndrinAldehyde(Bottomtaperwithwool)

Brand X DDD and DDE (Bottomtaperwithwool)

SGEEndrinKetoneandEndrinAldehyde (Tapered FocusLiner)

SGE DDD and DDE (Tapered FocusLiner)

BR-0

359-

A_R

evC

© S

GE

Ana

lytic

al S

cien

ce P

ty L

td 0

4/20

12

Easy to useSGE’s inlet liners come as a complete, packaged solution:•1,5and25packs.•Complete with instrument

appropriateo-ringsorsealingrings.•Each pack supplied with quality

assurancetestresults.•SGE blister packs are perforated

enablingeasydivisionofthe5and25packswhilemaintainingblisterintegrity.

•25 packs come in a re-usablecontainer,witharangeofattractivedesigns, thatwill behandy aroundthelab.

Easy to chooseEasy to use Confidence in your analysis

SGE inlet liners, spot the difference.

To select the right inlet liner for your instrument, injection and sample – use our selection tool: www.sge.com/linertool

When deactivation REALLY matters

Singlepackdeactivationcertificationoptions:•CustomerswhoareISOaccreditedorfollowGLPwillbenefitfromSGE’scertifiedsinglepacks. Add CERT when ordering.•MSreadylinersinsinglepacksconditionedandMStestedpostdeactivationsotheyarereadyforusestraight

outofthepack.Add MS when ordering.

AUSTRALIA & PACIFIC REGIONSGE Analytical Science Pty LtdToll Free: 1800 800 167Tel: +61 (0) 3 9837 4200Fax: +61 (0) 3 9874 5672Email: [email protected]

JAPANSGEJapanIncTel: +81 45 222 2885Fax: +81 45 222 2887Email: [email protected]

EUROPESGE Europe LtdEuropeanHeadOfficeToll Free: 00800 2790 8999Toll Free Fax: 00800 2626 2609Tel: +44 1908 568 844Fax: +44 1908 566 790Tel France: +33 1 69 29 80 90Fax France: +33 1 69 29 09 25Tel Germany: +49 (0) 6155 / 60746 0Fax Germany: +49 (0) 6155 / 60746 50Email: [email protected]

UNITED STATES OF AMERICASGEIncorporatedToll Free: (800) 945 6154Tel: +1 512 837 7190Fax: +1 512 836 9159Email: [email protected]

www.sge.com