setup for cosmic ray test of vpa panel, using minidaq d. levin, f eb 8, 2012

11
Setup for Cosmic Ray Test of VPA panel, using MiniDAQ D. Levin, Feb 8, 2012 VPA: 600 Torr Ar CO2 (1%), filled on Jan 17, 2012 1” DESY mini-scintillator hodoscope placed under panel at intersection of HV and readout lines. 24 RO lines, 14, HV lines. Hodoscope HV = 850 V., trigger rate = 3.6 muons/min, located 2.5 inches under panel, and offset Run: run72406 VPA HV = 830 V. After 5 hr, 15 min, 1019 triggers, only 7 have coincident TDC hits. This is very low rate- less than 1%. However, the hodoscope is offset from the panel, and removed too far away. At 5 hr, 38 minutes into run, I bring it closer and directly under active region. Place Ru-106 source in holder, about 2.0 inches over active region on top of panel. Trigger rate is now 153/4 ~ 38 CPM. But panel coincidence rate is still low- don’t see any additional hits beyond the 7 obtained with cosmics.

Upload: chessa

Post on 23-Feb-2016

25 views

Category:

Documents


0 download

DESCRIPTION

Setup for Cosmic Ray Test of VPA panel, using MiniDAQ D. Levin, F eb 8, 2012 VPA: 600 Torr Ar CO2 (1%), filled on Jan 17, 2012 1” DESY mini- s cintillator hodoscope placed under panel at intersection of HV and readout lines. 24 RO lines, 14, HV lines. - PowerPoint PPT Presentation

TRANSCRIPT

Page 1: Setup for Cosmic Ray Test of VPA panel, using  MiniDAQ D. Levin,  F eb 8, 2012

Setup for Cosmic Ray Test of VPA panel, using MiniDAQ

D. Levin, Feb 8, 2012

VPA: 600 Torr Ar CO2 (1%), filled on Jan 17, 2012

1” DESY mini-scintillator hodoscope placed under panel at intersection of HV and readout lines.

24 RO lines, 14, HV lines.

Hodoscope HV = 850 V., trigger rate = 3.6 muons/min, located 2.5 inches under panel, and offset

Run: run72406 VPA HV = 830 V. After 5 hr, 15 min, 1019 triggers, only 7 have coincident TDC hits.

This is very low rate- less than 1%. However, the hodoscope is offset from the panel, and removed too far away.At 5 hr, 38 minutes into run, I bring it closer and directly under active region.

Place Ru-106 source in holder, about 2.0 inches over active region on top of panel.Trigger rate is now 153/4 ~ 38 CPM. But panel coincidence rate is still low- don’t see any additional hits beyond the 7 obtained with cosmics.

Page 2: Setup for Cosmic Ray Test of VPA panel, using  MiniDAQ D. Levin,  F eb 8, 2012
Page 3: Setup for Cosmic Ray Test of VPA panel, using  MiniDAQ D. Levin,  F eb 8, 2012
Page 4: Setup for Cosmic Ray Test of VPA panel, using  MiniDAQ D. Levin,  F eb 8, 2012

PDP pick-off card:Configuration: x200 attenuation. 50 ohms termination

200 49

1

HV

Signal out to 50 ohm ribbon coax to Hedgehog Card

70

pixel

Heavy, 1” Ground braid

Page 5: Setup for Cosmic Ray Test of VPA panel, using  MiniDAQ D. Levin,  F eb 8, 2012

PDP VPA 600 Torr Ar CO2 (1%)

Row 24

Row 1

Col 1 Col 100-114

Pickoff card200x

attenuation

Mezzanine hedgehog50 X attenuation

CSM

Scintillator Hodoscope

To trigger logic and TTCvi

HV

Panel hookup to Minidaq

Page 6: Setup for Cosmic Ray Test of VPA panel, using  MiniDAQ D. Levin,  F eb 8, 2012

Setup for Cosmic Ray Test of VPA panel, using MiniDAQ

D. Levin, Feb 8, 201218:15

Make the following changes:

1) Place source directly on the panel- so tha tit radiates lines 4-7, and also the hodoscope.2) Increase HV to 8403) Lower minidaq threshold from 200 to 50 mV4) Take data run72410

After 2hr, 26 minutes, trigger rate = 4.5 Hz

Events from Ru-106 – Triggered by hodoscopeHit distribution- corresponds to source location:Made with ADC > 50 counts (see next page)

Time of hit relative to trigger.

Page 7: Setup for Cosmic Ray Test of VPA panel, using  MiniDAQ D. Levin,  F eb 8, 2012

These are the ADC distributions for the first 8 channels. The good hits appear to have levels over 50. The cuts on the previous page are ADC> 50 ADC Counts

Page 8: Setup for Cosmic Ray Test of VPA panel, using  MiniDAQ D. Levin,  F eb 8, 2012

Setup for Cosmic Ray Test of VPA panel, using MiniDAQ

D. Levin, Feb 8, 201218:47

Make the following changes:

1) Remove source2) Leave HV =840, threshold =-503) Take cosmic ray run: data run72411

…check back tomorrow

Page 9: Setup for Cosmic Ray Test of VPA panel, using  MiniDAQ D. Levin,  F eb 8, 2012

D. Levin, Feb 9-13 2012

The work for the past few days is summarized here: Most of the effort has been focused on getting the MiniDAQ to work – without cross talk, and to record accurately the source on the panel in the location where the source is actually positioned. This has proved to be very difficult. The problems:1) The PPS radiates and produces hits on Mezz card even when the mezz is not hooked up.

Remedies:a. ground everything, shield the mezz card, and especially ground the Pikov card.b. Tried shielding the panel by placing foli on all surfaces. This resulted in HV discharges thru the glass to the foil.c. Made foil covered box and placed over whole apparatus. No effect.d. In the end found that the Pikov was not properly grounded. Attach ground strap.

b. Use new Pikov card with proper impedance termination.2) The ribbon coax that convey the PPS signals for external trigger formation were universally bad. In every cable bundle there was at least

one cable with a poor ground.a. Remedy: found other higher quality 8 conductor ribbon cables- with lemo connectors.

3) Needed to properly attenuate at the pikov card and on the Hedeghog.1) For Ar CO2 signals (600 Torr), use 200X Pikov and 50X HH card.

4) The panel has wildly varying discharge voltages depending on the row and colum location. The HV is at column 100-114.5) For rows from 1-24 the HV needs to be varied by 10-20 V to achiev3 the same response. At other HV locations the panel easily enters

constant discharge mode.6) Tried using 200 Meg resistors- but the HV card I used seemed to radiate high noise levels. So still using 66 MegOhm.7) Spontaneous hit rate on panel VPA has become very high on some channels.- hundred or so Hz at 820 V.8) It is very hard to extract reliable results. 9) Enormous variation in rates across pixels. Cannot find an operating voltage that allows all channels to operate with even remotely the

same efficiency.10) This panel has become very difficult to use. I will re-evacuate, bake and refill with the same gas11) Re-pumping and baking panel, Will refill this panel tomorrow Feb 14, 29

Page 10: Setup for Cosmic Ray Test of VPA panel, using  MiniDAQ D. Levin,  F eb 8, 2012

Proper configuration for minidaq and reaodut:

1) 200 X pikov card2) 50 x attenuation on HH3) Set external discriminators to 200 mV.4) Form OR of 24 channels5) Send output to TTCvi trigger input6) With long blue ribbon cables (about 20 m) – set

1) Bunch Counter offset to 40002) Search window 753) Threshold -150 to -2504) Match window 50

Page 11: Setup for Cosmic Ray Test of VPA panel, using  MiniDAQ D. Levin,  F eb 8, 2012

Monda afternoon, 16:00 Feb 13:

Re-install Panel VP-C with the 600 torr CF4

Instrument with new Pikov4 card and Hv on channels 110-124

Ramp up HV until ~ 1700 where I hear crackling, frying sounds from panel.

This continues then stops. Increase voltage. Sounds return, then stop.

A few pulses observed- but mostly just hear this crackling sound.

Abandon further testing.