ScanTel SCANTEL.EXE Installation and Operation V1.03 01/2018

Upload: others

Post on 20-Aug-2020

21 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

ScanTel SCANTEL.EXE

Installation and OperationV1.03

01/2018

Page 2: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

Table of Contents

Program Information....................................................................................................................................................1

Installation....................................................................................................................................................................1Download from Scanivalve website................................................................................................................1Installation from install disk.............................................................................................................................1Start the ScanTel Program..............................................................................................................................1

Operation......................................................................................................................................................................2

Configure ScanTel ......................................................................................................................................................3

Connect.......................................................................................................................................................................3

Configure the Device ...................................................................................................................................................4Binary UDP Data Transfers.............................................................................................................................4

DSA 3000 Series................................................................................................................................4DSA 3200 Series................................................................................................................................4DTS3250 and DTS4050 Series .........................................................................................................4DSM3000/3200 Series.......................................................................................................................4DSM3400 Series................................................................................................................................5RAD3200 Series................................................................................................................................5RAD4000 Series................................................................................................................................6

ASCII Data Transfers......................................................................................................................................7DSA 3000 Series................................................................................................................................7DSA 3200 Series................................................................................................................................7DTS3250 and DTS4050 Series .........................................................................................................7DSM3000/3200 Series.......................................................................................................................7DSM3400 Series................................................................................................................................7DSAENCL3200..................................................................................................................................8DSAENCL4000..................................................................................................................................8RAD3200 Series................................................................................................................................8RAD4000 Series................................................................................................................................8

File Capture..................................................................................................................................................................9Binary Data File...............................................................................................................................................9ASCII Data File...............................................................................................................................................11

Convert Binary File......................................................................................................................................................12

Multiple Unit Binary Control.........................................................................................................................................14 Configuring MPS Multiple Unit Binary Control................................................................................................14 MPS Multiple Unit Binary Control....................................................................................................................17

Upload ASCII File.........................................................................................................................................................20

Examples......................................................................................................................................................................22DSA3200 Series UDP Binary High Speed.......................................................................................................22DTS3250 and DTS4050 UDP Binary...............................................................................................................23RAD4000 UDP Binary High Speed..................................................................................................................24DSA3200 Series ASCII TCP/IP........................................................................................................................25

APPENDIX A - ENGINEERING UNIT CONVERSION CONSTANTS...........................................................................26

APPENDIX B - SOFTWARE CHANGE LOG................................................................................................................27

Page 3: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

1

Program InformationThis program, ScanTel, is designed to replace the Binary Telnet program, BTEL.EXE. ScanTel is a support program for all DSM DSA, RAD, MPS and DTS modules. It supports communication in ASCII or BINARY formats. SCANTEL is capable of running in Windows XP, Windows 7, and Windows 10 operating systems.

NOTE: A user should have a good working knowledge of the hardware and software of the device used with this software. Please refer to the applicable hardware and software manuals for information on the scan speeds and data types available.

Installation

ScanTel is available on a disk or by download from the Scanivalve Corp website: www.scanivalve.com.

Download from Scanivalve website

The installation s program is in a zip file named ScanTel.zip. The file contains the installation programs: ScanTelinstall.msiSetup.exe.

1. Unzip the files to a temporary folder2. Select: Start3. Select: Run4. Highlight the file: setup.exe in the folder where the ScanTel installation files are stored and click OK to

start the installation.

Installation from install disk

1. Insert the CD in a CD drive. If the installation does not autorun2. Select: Start3. Select: Run4. Highlight the file: setup.exe on the CD and click OK to start the installation.

Start the ScanTel Program

Click on the ScanTel icon on the desktop.

Page 4: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

2

Operation

To start the program, click on the desktop icon. The main window will open:

This window has six drop down menu choices:

File This menu contains the commands for the main program tasks. This menu will be used tp start and stop file captures, upload files, and convert

Connect Click this to connect to the Scanivalve Device. The device type and IP Address must be specified in the configuration window before Connect is selected

Configuration Click this to set the Device Type, IP Address, and UDP Receive Port

Screen Click this to set the font size and color in the main window

Display This menu option contains the commands to clear the screen or clear the byte counter.

Help Click this to verify the ScanTel version number

This window has a status bar with the following information

IP Address The IP address of the device

Connect Status Connected or Disconnected

File Capture The destination file name will be displayed when a file capture is started. This window will only show the file name. It will not show the path or the capture type.

Byte Count Displays the number of Bytes collected during the current file capture.

Page 5: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

3

Configure ScanTel Click Configuration, the configuration window will open

Device The device type. This is very important as there are differences in the packets returned by each device.

Expect Binary Header Check this box if a Binary header will be included in the data packet.

Debug Output During Conversion Check this box if a file conversion fails and retry the conversion.

IP Address The IP address of the Device Specified

UDP Receive Port Identifies the port where the device will transmit Binary UDP packets.

Upload Line Delay in MS This should be set to regulate the transmission speed of a file upload to the device. This can be set from 0 to 100 milliseconds.

1. Select the Device Type by left clicking one of the modules listed in the window.2. If a Binary header will be included in the data packet, click the Expect Binary Header box3. Enter the IP Address of the device4. Enter the UDP port number for the data transmission5. Click OK to save the settings. The settings will be shown in the Status bar.

Connect Click Connect to connect to the device. The Status Bar should show the connection.

Page 6: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

4

Configure the Device

Binary UDP Data TransfersIn order for the ScanTel program to acquire UDP Binary data from a DSA, RAD, Enclosure, DSM, or DTS module, the module must be configured correctly. Changes to the module configuration may be made in the ScanTel window after the connection is made. The recommended settings for each type module are listed below. The setting s for PERIOD and AVERAGE reflect the maximum possible scan speed. This speed may not be obtained in all systems.

DSA 3000 SeriesSET PERIOD 325SET AVG 1SET BIN 1SET FORMAT 0SET PAGE 0SET NETTYPE UDPNOTE: When the NETTYPE variable is modified, the module must be rebooted before the change takes

effect.

DSA 3200 Series SET PERIOD 125SET AVG 1SET BIN 1SET FORMAT 0SET PAGE 0SET TIME 0SET PORT 23SET HOST <IP Address> <port> U Where: IP Address is the IP address of the host computer.

Port is the UDP data port U is UDP data transmission

NOTE: When the HOST variable is modified, the module must be rebooted before the change takes effect.

DTS3250 and DTS4050 Series SET PERIOD 325SET AVG 1SET BIN 1SET FORMAT 0SET PAGE 0SET TIME 0SET HOST <IP Address> <port> U Where: IP Address is the IP address of the Host Computer.

Port is the UDP data port to be usedU is UDP data transmission

NOTE: When the HOST variable is modified, the module must be rebooted before the change takes effect.

DSM3000/3200 SeriesSET PERIOD 50SET AVGn 1SET BIN 1SET FORMAT 0SET PAGE 0SET BINADDR <port> <IP Address> Where: IP Address is the IP address of the Host Computer.

Port is the UDP data portNOTE: When the BINADDR variable is modified, the module must be rebooted before the change takes

effect.

Page 7: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

5

DSM3400 SeriesSET PERIOD 30SET AVGn 1SET BIN 1SET CONOUT 2SET FORMAT 0SET PAGE 0SET BINADDR <port> <IP Address> Where: IP Address is the IP address of the Host Computer.

Port is the UDP data portNOTE: When the BINADDR variable is modified, the module must be rebooted before the change takes

effect.

DSAENCL3200SET PERIOD 30SET AVGn 1SET BIN 1SET CONOUT 2SET FORMAT 0SET PAGE 0SET BINADDR <port> <IP Address> Where: IP Address is the IP address of the Host Computer.

Port is the UDP data portNOTE: When the BINADDR variable is modified, the module must be rebooted before the change takes

effect.

DSAENCL4000SET PERIOD 25SET AVG1 1SET BIN 1 Set BIN to 4 if a header is to be added to the fileSET FORMAT 0SET PAGE 0SET BINADDR <port> <IP Address> Where: IP Address is the IP address of the Host Computer.

Port is the UDP data portNOTE1: When the BINADDR variable is modified, the module must be rebooted before the change takes

effect.NOTE2: When BIN is set to 4, the Expect Binary Scan Header Box in the Configuration Window must be

checked before a binary file is converted to ASCII.

RAD3200 SeriesSET PERIOD 30SET AVGn 1SET BIN 1SET FILEOUT 0SET FORMAT 0SET PAGE 0SET BINADDR <port> <IP Address> Where: IP Address is the IP address of the Host Computer.

Port is the UDP data portNOTE: When the BINADDR variable is modified, the module must be rebooted before the change takes

effect.

Page 8: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

6

RAD4000 SeriesSET PERIOD 25SET AVG1 1SET BIN 1 Set BIN to 4 if a header is to be added to the file.SET FORMAT 0SET PAGE 0SET BINADDR <port> <IP Address> Where: IP Address is the IP address of the Host Computer.

Port is the UDP data portNOTE1: When the BINADDR variable is modified, the module must be rebooted before the change takes

effect.NOTE2: When BIN is set to 4, the Expect Binary Scan Header Box in the Configuration Window must be

checked before a binary file is converted to ASCII.

Page 9: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

7

ASCII Data TransfersIn order for the ScanTel program to acquire ASCII TCP/IP data from a DSA, RAD, Enclosure, DSM, or DTS module, the module must be configured correctly. Changes to the module configuration may be made in the ScanTel window after the connection is made. The recommended settings for each type module are listed below. The settings for PERIOD and AVERAGE reflect a scan speed that should work in most systems.

DSA 3000 SeriesSET PERIOD 325SET AVG 4SET BIN 0SET FORMAT 0SET PAGE 0SET NETTYPE TCPNOTE: When the NETTYPE variable is modified, the module must be rebooted before the change takes

effect.

DSA 3200 SeriesSET PERIOD 150SET AVG 4SET BIN 0SET FORMAT 0SET PAGE 0SET TIME 0SET PORT 23SET HOST 0.0.0.0 0 T T is TCP data transmission NOTE: When the HOST variable is modified, the module must be rebooted before the change takes effect.

DTS3250 Series SET PERIOD 7812SET AVG 4SET BIN 0SET FORMAT 0SET PAGE 0SET TIME 0SET HOST 0.0.0.0 0 T T is TCP data transmissionNOTE: When the HOST variable is modified, the module must be rebooted before the change takes effect.

DSM3000/3200 SeriesSET PERIOD 50SET AVGn 4SET BIN 0SET FORMAT 0SET PAGE 0SET BINADDR 0 0.0.0.0 NOTE: When the BINADDR variable is modified, the module must be rebooted before the change takes

effect.

DSM3400 SeriesSET PERIOD 50SET AVG 4SET BIN 0SET CONOUT 2SET FORMAT 0SET PAGE 0SET BINADDR 0 0.0.0.0NOTE: When the BINADDR variable is modified, the module must be rebooted before the change takes

effect.

Page 10: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

8

DSAENCL3200SET PERIOD 50SET AVGn 4SET BIN 0SET CONOUT 2SET FORMAT 0SET PAGE 0SET BINADDR 0 0.0.0.0NOTE: When the BINADDR variable is modified, the module must be rebooted before the change takes

effect.

DSAENCL4000SET PERIOD 50SET AVG1 41SET BIN 0 SET FORMAT 0SET PAGE 0SET BINADDR 0 0.0.0.0NOTE: When the BINADDR variable is modified, the module must be rebooted before the change takes

effect.

RAD3200 SeriesSET PERIOD 50SET AVGn 4SET BIN 0SET FILEOUT 0SET FORMAT 0SET PAGE 0SET BINADDR 0 0.0.0.0NOTE: When the BINADDR variable is modified, the module must be rebooted before the change takes

effect.

RAD4000 SeriesSET PERIOD 50SET AVG1 4SET BIN 0SET FORMAT 0SET PAGE 0SET BINADDR 0 0.0.0.0NOTE: When the BINADDR variable is modified, the module must be rebooted before the change takes

effect.

Page 11: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

9

File Capture

Binary Data File

1. Configure the device for a Binary UDP data transfer. Refer to the configuration settings and the device software manual for more information.

2. Select: File3. Select: Start Binary Capture File

4. The Save File window will open. Enter the file name for the test and click OK

Page 12: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

10

5. The status bar will show the file capture is on and the file name.

6. Type: Scan to start the data collection. If FPS is set to a finite number, the data collection will stop at the last frame. The Byte counter in the STATUS bar will count up as data are collected. When the scan stops, the total number of bytes will be displayed. If FPS is set to 0, the data collection must be stopped by the user by issuing a STOP command, or by pressing the escape key.

NOTE: At very fast scan speeds, it is recommended that FPS be set to a number other than 0. At speeds greater than 100 samples/channel/second, the module may not be able to acknowledge the STOP command.

The byte counter will show the total number of bytes captured.

Page 13: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

11

7. Select: File8. Select: Close Binary Capture File

ASCII Data File

1. Configure the device for ASCII TCP/IP data transfer. Refer to the configuration settings and the device software manual for more information.

2. Select: File3. Select: Start ASCII Capture File4. The Save File window will open. Enter the file name for the test and click OK5. The status bar will show the file capture is on and the file name.6. Type: Scan to start the data collection. If FPS is set to a finite number, the data collection will stop at the

last frame. If FPS is set to 0, the data collection must be stopped by the user by issuing a STOP command, or by pressing the escape key.

7. Select: File8. Select: Close ASCII Capture File.

Page 14: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

12

Convert Binary FileWhen the Binary File Capture is complete, the file can be converted to a .csv file that can be imported into any spreadsheet program. If the Binary File was captured from a RAD4000 or DSAENCL4000, the file may have a header. If the file has a header, the Expect Binary Scan Header Box in the Configuration Window must be checked before the conversion process is started. Headers may only be added to RAD4000 or DSAENCL4000 devices when using this software.

1. To convert a binary file to a .csv file2. Select: File3. Select: Convert Binary File

Page 15: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

13

4. The File Open Window will open

5. Highlight the file to be converted and click Open.6. The file will be converted. A window will open to show the progress

7. The converted file will be written to the same folder where the binary file was located. The file will have the same name as the binary file with the .csv extension. If a header was included in the binary file, the header data will be written to a file with the same name as the binary file, but with a .txt extension.

NOTE: Headers may only be added to RAD4000 and DSAENCL4000 devices in UDP data transmissions.

Page 16: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

14

Multiple Unit Binary ControlWhen certain device types are selected on the Configuration screen, an option for Multiple Unit Binary Control may appear on the main ScanTel window. The Multiple Unit Binary Control allows a user to connect to multiple devices, initiate a scan, collect scan data, and merge the scan data to a single file. This feature requires the user to set one device as a PTP master, and the others as slaves. The following section will explain the setup procedure for Multiple Unit Binary Control of specific devices.

Configuring MPS Multiple Unit Binary ControlThis procedure requires MPS software version 2.05 or newer. Contact the factory at [email protected] for

the latest version of MPS software. ScanTel version 1.03 or newer is also required.1. Connect power and Ethernet cables to each MPS to be used. Power on each MPS and allow approximately 2

minutes for the MPS to complete their boot cycle. 2. Open a ScanTel instance for each MPS to be used3. In ScanTel, click “Help”-”About” to verify that ScanTel is version 1.03 or newer. If it is not, please download the

latest version of ScanTel from the Scanivalve website www.scanivalve.com4. Connect to each MPS in their own ScanTel instance. Click Configuration, input the IP Address of the scanner,

and select MPS in the conversion settings. See image below.

Page 17: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

15

5. Adjust the LIST S settings of each scanner to be used. It is recommended that the LIST S settings be the same for all scanners being used. In each ScanTel instance, type:

LIST S<ENTER>SET RATE <desired scan rate in Hz><ENTER>SET FPS <desired frames per scan><ENTER>SET UNITS <desired units><ENTER>SET FORMAT B B<ENTER>SET TRIG 0<ENTER>SET ENFTP 0<ENTER> SET OPTIONS <desired options, default is 0 0 16><ENTER>

6. Adjust LIST UDP settings. The LIST UDP settings of each MPS to be used must have ENUDP disabled. In each ScanTel instance, type:

LIST UDP<ENTER>SET ENUDP 0<ENTER>SET IPUDP 0.0.0.0 0<ENTER>

Page 18: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

16

7. Adjust LIST PTP settings of a single MPS. Select a single MPS to adjust the PTP settings as specified. Only one MPS may have PTPEN set to 2 while the others have PTPEN set to 1. The MPS with PTPEN set to 2 will act as the PTP master while the others with PTPEN set to 1 act as slaves. In only one ScanTel instance, type:

LIST PTP<ENTER> SET PTPEN 2<ENTER>SET SST 0:0:0.0<ENTER>SET SSD 1971/1/1<ENTER>SET UTCOFFSET 0:0:0.0<ENTER>

NOTE: It is important that SST, SSD, and UTCOFFSET variables of all scanners being used are the same.

8. Adjust LIST PTP settings of the other MPS to be used. In each ScanTel instance that isn’t the PTP master, type: LIST PTP<ENTER> SET PTPEN 1<ENTER>SET SST 0:0:0.0<ENTER>SET SSD 1971/1/1<ENTER>SET UTCOFFSET 0:0:0.0<ENTER>

NOTE: It is important that SST, SSD, and UTCOFFSET variables of all scanners being used are the same.

Page 19: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

17

9. Adjust LIST M settings. Each MPS must have SVRSEL set to 2 to enable the binary server. In each ScanTel instance, type:

LIST M<ENTER>SET SVRSEL 2<ENTER>

10. Save the changes to each MPS. A save is required if SVRSEL, ENUDP, or ENFTP are changed. After the intial save, subsequent Multiple Unit Binary scans will not require a SAVE or change of the variables listed in steps 1-9 in this procedure. In each ScanTel instance, type:

SAVE<ENTER> STATUS<ENTER>Allow ample time for the save to complete. The save can take over a minute. When each MPS returns “STATUS:

READY” you may proceed to the next step.

11. Disconnect from each ScanTel instance. Click “Disconnect” on each ScanTel instance once the save is complete.

12. Close each ScanTel instance.

13. Power cycle each MPS. Turn each MPS off for at least 10 seconds. Reapply power.

MPS Multiple Unit Binary Control1. Open a single ScanTel instance.

2. Select a single MPS. Click the “Configuration” tab. Input the IP Address of the MPS. Select MPS in conversion settings. Click “OK”. Click “Connect”

3. Select “Multiple Unit Control”. The Multiple Unit Binary Control window will open as shown below.

Page 20: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

18

4. Set the Base Directory for the Multiple Unit Binary Control. This directory will become populated with a timestamped folder with data from each MPS every time a scan occurs.

Click “File”Click “Set Base Directory”Select a suitable location and select OK.

5. Set the IP Addresses of the scanners you want to collect data from. Click the empty IP Address to select it. Click the IP Address again to get a cursor to change the IP Address. Input the IP Address of an MPS and hit ENTER. Check the box next to the IP Address to enable or disable the scanner. Do this for each MPS, inputting their discreet IP Address

Page 21: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

19

6. Click “Connect” on the Multiple Unit Binary Control window. The “Status” for each scanner enabled should change from “Disconnected” to “Connected”.

7. To start scanning:Cick “Scan Control”Click “Start Scan” Scanning will begin. Note the status change from “Connected” to “Scanning”. Also note the “Frames Transferred”

climbs until the scan stops.

8. To stop scanning:Click “Scan Control” Click “Stop Scan”Scanning will stop. Note the status change from “Scanning” to “Connected”. The binary .dat files can be located in

the timestamped folder in the Base Directory.

9. Merge data files. Click “File” Click “Merge Data Files”Slect the timestamped folder that you would like the .dat files converted and merged to a single .CSV. Click “OK”Allow time for the files to be converted and merged.

NOTE: This step can be taken at a time convenient to the user. Files can be merged after testing if desired.

10. Access data. Once the merge is complete, a data file titled Merged.csv will appear in the timestamped folder that was converted.

Page 22: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

20

Upload ASCII FileASCII files may be uploaded to any of the devices supported in this program. Generally, this feature would be used to upload configuration files to a DSM, RAD or Enclosure. It most likely would be used to upload calibration coefficients to a DSA3200 Series module.

1. Before uploading a coefficient file to a DSA module the existing coefficients must be deleted.2. After the connection is made, Type: DELETE 0 69. This will delete all existing master planes from RAM in

the module. 3. From the main screen, Select: Upload ASCII File

4. The file Open window will open. Browse to the folder where the coefficients are stored, highlight the correct file, and click Open.

Page 23: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

21

5. The upload will start immediately. The upload is complete when the last channel in the last temperature plane is displayed.

6. Type: SAVE to save the coefficients to the device. 7. Click Disconnect8. Cycle power to complete the process.

Page 24: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

22

Examples

DSA3200 Series UDP Binary High Speed

A DSA 3200/16Tx at address 191.30.80.144 is to be configured to send binary data at 500 samples/channel/sec over the UDP port 23.The data are to be logged to a binary file named data.bin in a folder named: data. 1000 frames of data are to be sent. After the data are logged, they are to be converted to ASCII format..

1. Click on the ScanTel Icon2. Open the Configuration Window 3. Select the Device Type 4. Enter the IP Address of the device 5. Set the UDP Port - Port 23 is the default6. Click OK7. Click Connect to connect to the module.8. Configure the module for UDP Binary Data Transmission

SET PERIOD 125SET AVG 1SET BIN 1SET FORMAT 0SET FPS 1000SET PAGE 0SET TIME 0SET PORT 23SET HOST <IP Address> <port> U Where: IP Address is the IP address of the host computer.

Port is the UDP data port U is UDP data transmission

SAVE This will save the changes9. Click Disconnect10. Cycle the power to the module11. Click Connect to reconnect to the module12. Select: File13. Select: Start Binary Capture File14. Browse to the correct folder, Enter the file name and click OK15. Type: SCAN in the ScanTel window16. When the scan is complete, Select: File17. Select: Close Binary File18. Select: Convert Binary File19 Highlight the file to be converted and click Open20. Close the Convert Status Window.

Page 25: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

23

DTS3250 and DTS4050 UDP Binary

A DTS 3250/16Tx at address 191.30.85.146 is to be configured to send binary data over the UDP port 23.The data are to be logged to a binary file named data.bin in a folder named: data. 1000 frames of data are to be sent. After the data are logged, they are to be converted to ASCII format.

1. Click on the ScanTel Icon2. Open the Configuration Window 3. Select the Device Type 4. Enter the IP Address of the device 5. Set the UDP Port - Port 23 is the default6. Click OK7. Click Connect to connect to the module.8. Configure the module for UDP Binary Data Transmission

SET PERIOD 325SET AVG 1SET BIN 1SET FPS 1000SET FORMAT 0SET PAGE 0SET TIME 0SET HOST <IP Address> <port> U Where: IP Address is the IP address of the Host Computer.

Port is the UDP data port to be usedU is UDP data transmission

SAVE This will save the changes9. Click Disconnect10. Cycle the power to the module11. Click Connect to reconnect to the module12. Select: File13. Select: Start Binary Capture File14. Browse to the correct folder, Enter the file name and click OK15. Type: SCAN in the ScanTel window16. When the scan is complete, Select: File17. Select: Close Binary File18. Select: Convert Binary File19 Highlight the file to be converted and click Open20. Close the Convert Status Window.

Page 26: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

24

RAD4000 UDP Binary High Speed

A RAD4000 at address 191.30.30.20 is to be configured to send binary data over the UDP port 23.The data are to be logged to a binary file named data.bin. 6000 frames of data are to be sent. The data must include a header. After the data are logged, they are to be converted to ASCII format.

1. Click on the ScanTel Icon2. Open the Configuration Window 3. Select the Device Type 4. Enter the IP Address of the device 5. Set the UDP Port - Port 23 is the default6. Check the box: Expect Binary Scan Header7. Click OK8. Click Connect to connect to the module.9. Configure the module for UDP Binary Data Transmission

SET PERIOD 25SET AVG1 1SET BIN 4SET FPS1 6000 SET FORMAT 0SET PAGE 0SET BINADDR <port> <IP Address> Where: IP Address is the IP address of the Host Computer.

Port is the UDP data portSAVE This will save the changes

10. Click Disconnect11. Cycle the power to the module12. Click Connect to reconnect to the module13. Select: File14. Select: Start Binary Capture File15. Browse to the correct folder, Enter the file name and click OK16. Type: SCAN in the ScanTel window17. When the scan is complete, Select: File18. Select: Close Binary File19. Select: Convert Binary File20. Highlight the file to be converted and click Open21. Close the Convert Status Window.

Page 27: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

25

DSA3200 Series ASCII TCP/IP

A DSA 3200/16Tx at address 191.30.80.144 is to be configured to send ASCII TCP/IP data at 50 samples/channel/sec over the TelNet port (Port 23). 23.The data are to be logged to a file named data.txt in a folder named: data. 10000 frames of data are to be sent.

1. Click on the ScanTel Icon2. Open the Configuration Window 3. Select the Device Type 4. Enter the IP Address of the device 5. Click OK6. Click Connect to connect to the module.7. Configure the module for ASCII Data Transmission

SET PERIOD 156SET AVG 8SET BIN 0SET FORMAT 0SET PAGE 0SET FPS 10000SET TIME 0SET PORT 23SET HOST 0.0.0.0 0 T T is TCP data transmission

SAVE This will save the changes8. If the module had been set up for UDP transmission, go to step 9

If not, go to step 129. Click Disconnect10. Cycle the power to the module11. Click Connect to reconnect to the module12. Select: File13. Select: Start ASCII Capture File14. Browse to the correct folder, Enter the file name and click OK15. Type: SCAN in the ScanTel window16. When the scan is complete, Select Close ASCII File

The data may be viewed in a test editor or spreadsheet program.

Page 28: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

26

APPENDIX A - ENGINEERING UNIT CONVERSION CONSTANTS

UNITSCAN Setting

Engineering Unit PSI to EU1 psi =

EU to PSI1 EU =

ATM Atmospheres 0.068046 A 14.6960 psiBAR Bars 0.068947 b 14.5039 psiCMHG Centimeter of Mercury 5.17149 cmHg 0.193368 psiCMH2O Centimeter of Water 70.308 cmH2O 0.014223 psiDECIBAR Decibar 0.68947 db 1.4504 psiFTH2O Foot of Water 2.3067 ftH2O 0.43352 psiGCM2 Gram per square Centimeter 70.306 g/cm2 0.014224 psiINHG Inch of Mercury @ 0°C 2.0360 inHg 0.491159 psiINH2O Inch of Water @ 4°C 27.680 inH2O 0.036127 psiKGCM2 Kilogram per square Centimeter 0.0703070 kg/cm2 14.2235 psiKGM2 Kilogram per square Meter 703.069 kg/m2 0.0014223 psiKIPIN2 kips per square inch(ksi) 0.001 kip/in2 1000.0 psiKNM2 Kilonewton per square Meter 6.89476 kN/m2 0.145038 psiKPA Kilopascal 6.89476 kPa 0.145038 psiMBAR Millibar 68.947 mb 0.014504 psiMH2O Meter of Water 0.70309 mH2O 1.42229 psiMMHG Millimeter of Mercury 51.7149 mmHg 0.0193368 psiMPA Megapascal 0.00689476 Mpa 145.038 psiNCM2 Newton per square Centimeter 0.689476 N/cm2 1.45038 psiNM2 Newton per square Meter 6894.76 N/m2 0.000145038 psiOZFT2 Ounce per square Foot 2304.00 oz/ft2 0.000434028 psiOZIN2 Ounce per square Inch 16.00 in/ft2 0.062500 psiPA Pascal 6894.76 Pa 0.000145038 psiPSF Pound per square Foot 144.00 lb/ft2 0.00694444 psiTORR Torr 51.7149 T 0.0193368 psi

Page 29: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

27

APPENDIX B - SOFTWARE CHANGE LOG

This section contains change information to assist a user in determining the differences between different versions of software.

Version 1.00 - Initial Release - April 2011

Version 1.01 - August 2011Added ability to send the <TAB> character to the device.Corrected a bug that did not insert the time stamp into RAD, Enclosure, or DSM CSV files except when the packet types were 3, 4, 8 or 9.Changed the binary scan header from expecting a unsigned long to expecting a float when reading Period.

Version 1.02 - March 2015Added functionality to convert binary files of raw (non engineering unit) data.Resolved a bug that sometimes prevented binary data from being captured.

Version 1.03 - January 2018Added Multiple Unit Binary ControlSupports multiple MPS data collection and synchronization

Page 30: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

28

Page 31: ScanTelscanivalve.com/wp-content/uploads/2018/01/ScanTel_V103.pdf · SET AVG1 1 SET BIN 1 Set BIN to 4 if a header is to be added to the file SET FORMAT 0 SET PAGE 0 SET BINADDR

29