SERVICE MANUAL

IES 21.3.x

10.1680116-443

Page 2: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference
Page 3: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

SERVICE MANUALIES 2

Page 4: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

Registered trademarks of Erbe Elektromedizin GmbH: AXUS®, BICISION®, BiClamp®, classicCOAG®, classicCUT®, CLEVERCAP®, DeCo®, dryCUT®, endoCUT®, endoCOAG®, ERBE®, ERBECRYO®, ERBEFLO®, ERBEFLO AeroRinse®, ERBEJET®, ERBELift®, ERBOKRYO®, FIAPC®, forcedCOAG®, Hybrid knife®, HybridKnife®, ICC 200®, ICC 80®, KYRON®, NESSY®, NESSY Ω®, PLURA®, preciseAPC®, Preflow®, pulsedAPC®, ReMode®, REMODE®, softCOAG®, sprayCOAG®, swiftCOAG®, The Color Blue®, thermoSEAL®, twinCOAG®, VIO®.

Service manual Art. No. 80116-443

All rights to this manual, in particular rights of duplication, dissemination and translation, are reserved. No part of this man-ual may be reproduced in any form (by photocopying, microfilming or other methods) or processed, duplicated or dissem-inated by the use of electronic systems without the written consent of Erbe Elektromedizin GmbH.

The information contained in this manual may be amended or supplemented without prior notice and represents no obli-gation on the part of Erbe Elektromedizin GmbH.

Printed by Erbe Elektromedizin

Printed in Germany

Copyright © Erbe Elektromedizin GmbH, Tübingen 2016

Page 5: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

Table of Contents80

116-

443

10.1

6

Table of Contents

1 Safety instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Classification of the safety information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Knowledge of the User Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Protection from the risk of electric shock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7Liability and warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

2 Unit description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Unit variants . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9Factory setting and Reset. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10Controls at the front . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11Controls at the rear . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13Description of function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14

3 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15

4 Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17Installation possibilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17

5 Circuit Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19Block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20Description of the various assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21

Processor board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Pressure sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Temperature sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Controller Supervisor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21Power switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Transformer (only for 100 and 120 V unit variants) . . . . . . . . . . . . . . . . . . . 22Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Potential equalization socket . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22Power connection with fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Line filter motor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23Motor control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23EEProm filter. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23ECB interface (ECB = ERBE Communication Bus). . . . . . . . . . . . . . . . . . . . . . 23Buttons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23LED display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

6 Settings and service functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25General information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Setting mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Service mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27

5 / 46

Page 6: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

Table of Contents

8011

6-44

3

10.1

6

7 Remedying malfunctions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29

8 Maintenance and servicing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31Who is allowed to perform servicing and maintenance work? . . . . . . . . . . . . . 31What is a technical safety check?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31How often does a technical safety check have to be performed? . . . . . . . . . . . 31Technical safety check – step by step . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

Safety information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32User manual and visual inspections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33Electrical safety checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33Performance tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33Activation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34Activation with VIO electrosurgical unit (via the ECB) . . . . . . . . . . . . . . . . . .35Operation suction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36Blocked suction detection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .36Basic suction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37Filter recognition . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37

9 Replacing components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39Replacing components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39

10 Spare parts. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41

6 / 46

Page 7: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

1 • Safety instructions80

116-

443

10.1

6

CHAPTER 1Safety instructions

Classification of the safety information

Knowledge of the User Manual

The User Manual for this unit constitutes an integral part of this Service Manual. For performing servicing activities it is assumed that the reader has knowledge of the User Manual, especially procedures for installation, putting into operation, and handling.

Protection from the risk of electric shock

WARNING! The WARNING! safety indication refers to a risk of personal injury.

CAUTION! The CAUTION! safety indication refers to a risk of damage to property.

ATTENTION! The ATTENTION! safety indication refers to a risk which can cause equipment to become unserviceable.

IMPORTANT! The IMPORTANT! designation indicates application information and other particularly important information.

WARNING! The supply voltage must match the voltage specified on the rating plate. Connect the unit / the equipment cart to a properly installed grounded outlet. Only use the Erbe power cord or an equivalent power cord for this purpose. The power cord must bear the national test sym-bol.

For safety reasons, multiple outlets and extension cords should not be used. If their use is unavoidable, they also must be provided with proper grounding.

WARNING! Unplug the power cord from the outlet before exchanging parts of the unit or cleaning it.

WARNING! Do not plug a wet power cord into the unit or into an outlet.

7 / 46

Page 8: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

1 • Safety instructions

8011

6-44

3

10.1

6

Liability and warranty

WARNING! Do not touch any unprotected wires or conductive surfaces while the unit is opened and under voltage.

WARNING! Blown line fuses may only be replaced by a competent technician. Only replacement fuses of the rating specified on the unit's name plate may be used. Before resuming operation the unit must be subjected to a performance test by a competent technician.

ATTENTION! Adjustments, tests, modifications, maintenance and repair work may only be performed by Erbe or persons trained by Erbe. If the work is not performed by trained persons, Erbe accepts no liability and warranty rights become void.

It is recommended that the technical safety check also be performed by Erbe or persons trained by Erbe.

ATTENTION! Only use original Erbe spare parts. The manufacturer accepts no liability and the warranty rights becomes void if original spare parts are not used.

8 / 46

Page 9: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

2 • Unit description80

116-

443

10.1

6

CHAPTER 2Unit description

Unit variants

The IES 2 is avaiable in the following unit variants:

• 230 V• 120 V• 100 V

9 / 46

Page 10: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

2 • Unit description

8011

6-44

3

10.1

6

Factory setting and Reset

Factory setting The IES 2 has the following settings when it is delivered by the manufac-turer:

• Operation suction: 80 %• Basic suction: 10 %• Run-on time: 2 s• Occlusion / system blocked (sensitivity): 5• Occlusion / system blocked (retrigger hold-time): 0 s• Display brightness: 10• Sound: on (=1)• Basic suction time: 30 s• Setting mode inhibit: off (=0)• Basic setting inhibit: off (=0)

Reset (reset unit to factory settings)

IMPORTANT! The unit can only be reset once the "Basic setting inhibit" service function is off (=0), see page 27.

Fig. 2-1

1. When switching on the unit, keep the power switch and Operation suction button pressed until <S. 0> appears in the display.

2. Press the START Suction button within 5 seconds. The unit is reset to the factory settings if:– <P. 1> appears in the display,

– the Basic suction button illuminates and

– the %-s-display flashes3. To exit the Setting mode, switch off the unit.

10 / 46

Page 11: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

2 • Unit description80

116-

443

10.1

6

Controls at the front

Fig. 2-2

IMPORTANT! This chapter contains an overview of the controls of the unit. The rele-vant User Manual for the unit, knowledge of which is assumed for ser-vicing work, provides detailed information about how to use the unit.

1 Power Switch

2 Main filter

3 Main filter lock

4 Main filter input (applied part)

11 / 46

Page 12: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

2 • Unit description

8011

6-44

3

10.1

6

Fig. 2-3

1 Operation suction

2 Basic suction

3 Run-on time

4 START Suction

5 Display

6 Display for ECB connection made to VIO electrosurgical unit

7 System blocked indicator/Occlusion

8 Filter exhausted indicator

9 % and/or second display

10 Higher/Lower-Buttons

12 / 46

Page 13: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

2 • Unit description80

116-

443

10.1

6

Controls at the rear

Fig. 2-4

1 Power supply module with fuses

2 Footswitch socket

3 ECB socket (Erbe Communication Bus)

4 Potential equalization terminal

13 / 46

Page 14: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

2 • Unit description

8011

6-44

3

10.1

6

Description of function

The IES 2 is an electrical medical device which purifies the air of smoke and gas particles which result from the vaporization of human and animal tis-sue. This smoke typically arises during the use of laser and electrosurgical devices and generally consists of water vapor, aerosols and organic gases.

The IES 2 improves the ambient conditions in the OR by:

• reducing the amount of dust by removing respirable particles,• improving vision,• removing unpleasant-smelling gases,• evacuating and filtering bioaerosols.

The IES 2 extracts the smoke-laden air via a device on the surgical applied part or via a hand-held tube which is separate from the device and posi-tioned near the applied part. The evacuated air is then purified of smoke and gas particles by means of a high-performance filter, and recirculated to the ambient air.

Air purification refers to particles such as aerosols, which are captured by an ULPA high-performance particle filter, and also to organic gases, which are detected by an odor that humans find unpleasant, among other things, and which can be extracted by a special gas filter..

14 / 46

Page 15: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

3 • Technical data80

116-

443

10.1

6

CHAPTER 3Technical data

Power connection

Rated supply voltage 230 V ± 10 % 120 V ± 10 % 100 V ± 10 %

Rated supply frequency 50 / 60 Hz 50 / 60 Hz 50 / 60 Hz

Current consumption 3.0 A 4.2 A 5.0 A

Power consumption 410 watts 370 watts 375 watts

Terminal for potential equalization yes yes yes

Power fuses T 3.15 A H / 250 V T 8 A H / 250 V T 8 A H / 250 V

Device parameters 230 V variant 100 / 120 V variant

Suction capacity System: 1600 l/min ± 10 %System with new filter: >550 l/min

System: 1600 l/min ± 10 %System with new filter: >450 l/min

Filter parameters ULPA, retention = 99.9999 % @ 0.01 μm DIN EN 1822-1 Filter Class H16

Dimensions and weight

Width x height x depth (all versions) 205 x 260 x 404 mm (without main filter)

Weight (230 V version) 10.7 kg (with main filter)

Weight (100 V / 120 V version) 14.1 kg (with main filter)

Ambient conditions for transport and storage of unit

Temperature -40 °C to +70 °C

Relative humidity 10 % – 95 %

Air pressure 54 kPa - 106 kPa

15 / 46

Page 16: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

3 • Technical data

8011

6-44

3

10.1

6

Ambient conditions for operation of the unit

Temperature +10 °C to +40 °C

Relative humidity 15% - 80%, non-condensing

Air pressure 54 kPa - 106 kPa

Acclimatizing

If the unit has been stored or transported at temperatures below +10 °C or above +40 °C, the unit will require approx. 3 hours to acclimatize at room temperature.

Standards

Classification according to EC Directive 93/42/EEC I

Protection class as per EN 60 601-1 I

Type as per EN 60 601-1 CF

16 / 46

Page 17: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

4 • Installation80

116-

443

10.1

6

CHAPTER 4Installation

Installation possibilities

For IES 2 there are two installation possibilities:

• Installation on an Erbe VIO Cart.• IES 2 as a stand alone variant.

These installation possibilities are described in the User Manual for the IES 2 and the User Manual for the above equipment cart.

17 / 46

Page 18: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

4 • Installation

8011

6-44

3

10.1

6

18 / 46

Page 19: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

5 • Circuit Descriptions80

116-

443

10.1

6

CHAPTER 5Circuit Descriptions

19 / 46

Page 20: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

5 • Circuit Descriptions

8011

6-44

3

10.1

6

Block diagram

Fig. 5-1

Potentialequalizationsocket

Power connetionwith fuses

Line filtermotor

Transformer*100 / 230 V120 / 230 V

MotorMotor control

Power switch

Powersupply

Pressuresensor

Temperaturesensor

Controllersupervisor

Processor board

+24VDCController

+12VDCMotor

+5VDCECB

EEPromfilter

ECBinterface Buttons

LEDdisplay

*only for 100 and 120 V unit variants

20 / 46

Page 21: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

5 • Circuit Descriptions80

116-

443

10.1

6

Description of the various assemblies

Processor board

Pressure sensor

The pressure sensor records the back pressure and transmits this using the internal AD converter to the processor.

The pressure sensor is a 68.8 hPa (1 psi) sensor and is neither compensat-ed nor calibrated. It is fed via a constant-current source. The pressure sen-sor measuring bridge is then read in via a difference amplifier with a gain of 133 by the processor AD converter. The pressure sensor has a full-scale output signal of 20 mV. A gain of 133 gives a voltage of 2.66 V for 69 hPa. To adjust the zero output a potentiometer is provided. The amplitude is not adjusted, since the absolute value is not decisive for the evaluation of the occlusion (system blocked).

Temperature sensor

For the temperature sensor a temperature-dependent resistor is used. This is read into the processor via the AD converter.

Controller SupervisorThe assembly has three functions:

Function 1 The controller supervisor keeps the controller in Reset while the voltages are ramping up.

Function 2 The controller supervisor detects a power-fail and sends a message to the controller on the interrupt input. It is now possible for the controller to store the current settings and to operate in a safe status.

Processor: MC68HC912BC32Quartz: 16 MHzADC: The AD converter inputs are filtered by an RC element

Manufacturer: SensorTechnicsType: SX01DD4

Type: B57620-C103Resistor value: at 25 °C = 10 kOhmSeries resistor: 10 kOhm

21 / 46

Page 22: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

5 • Circuit Descriptions

8011

6-44

3

10.1

6

Function 3 The controller supervisor functions as a low-active reset output. This out-put leads to a defined input in the controller.

The power-fail signal is emitted when a voltage of less than 1.25 V is de-tected at the PFI input. This voltage is tapped by the unregulated and only weakly filtered 18 V via a voltage divider.

R1 = 100 KOhm; R2 = 7.5 KOhm, U PFI = 1.25 V, U = 17.9 V

If the voltage after the rectifier is lower than 17.9 V, the processor detects a power-fail.

Power switchThe power on/off switch disconnects the unit from the mains. The switch is soldered directly to the printed-circuit board. It is a momentary-contact switch.

Transformer (only for 100 and 120 V unit variants)The ring core transformer sets the input mains voltage up to 230 V and supplies the fan motor with the required energy via the mains filter.

Power supplyThe power supply is subdivided into three main areas: The three power supplies are effected via a transformer. The three voltag-es are electrically isolated from each other.

Low voltage supply of the electronics, display and

controller

This voltage is rectified via a bridge-connected rectifier and stabilized to 5 V using a switching controller. The switching controller has an input volt-age range of 7 – 40 V and a clock frequency of 52 kHz.

Power supply for controlling the motor

This voltage is rectified via a bridge-connected rectifier and stabilized to 12 V using a linear voltage regulator.

Power supply for the ECB interface

This voltage is rectified via a bridge-connected rectifier and stabilized to 5 V using a linear voltage regulator.

Potential equalization socketThe potential equilization socket is for connecting up an external ground conductor. This socket is grounded.

22 / 46

Page 23: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

5 • Circuit Descriptions80

116-

443

10.1

6

Power connection with fusesThe power connection is via a non-heating appliance socket. The unit fuses are also here.

Line filter motorThe line filter compensates for the EMC interference of the motor. The line filter is connected to the motor by a copper band.

Motor controlThe motor is controlled by a PWM signal. Control is via an optical coupler as well as another electrically isolated supply voltage.

EEProm filterIn order to query the filter status, an EEProm is integrated in the filter. In this EEProm, the operating time of the filter is monitored, among other things. The EEProm is addressed via an SPI bus.

ECB interface (ECB = ERBE Communication Bus)The IES 2 can be controlled entirely via the electrically isolated ECB inter-face. All IES 2 functions can be availed of. The signal is generated by the ECB interface contained in the processor. Electrical isolation occurs using two optical couplers and another transformer winding. The levels are gen-erated via a corresponding interface module.

ButtonsThe system is operated using six buttons. Some of the buttons additionally have an LED, which is controlled by the processor. The buttons are inter-rogated via the processor, while the software carries out debouncing.

Optical coupler: CNY174 KV separation

Electrical supply: Electrically isolated transformer windingLinear controller 78L12

Control: Basic frequency: 330 HzIncrements: 15 μm

23 / 46

Page 24: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

5 • Circuit Descriptions

8011

6-44

3

10.1

6

LED displayFor the display, a 7-segment display for 3 characters was chosen. In addi-tion, there are four other LEDs for lighting up the buttons and four for dis-playing the various statuses.

Control of the LED cathodes is via the data bus of Port B and a driver, ULN2803. The LED anodes are controlled by Port T via the transistors of type BC817. Control is sequential with a basic frequency of 62 Hz.

The LED "ON" time can be changed via the brightness setting in the service mode. The LED "ON" time can be adjusted to between 0.1 ms and 1 ms.

24 / 46

Page 25: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

6 • Settings and service functions80

116-

443

10.1

6

CHAPTER 6Settings and service functions

General information

This device has two operating modes for changing settings or to call up in-formation for the service technician:

• Setting mode• Service mode

Setting mode

OverviewDisplay Assembly Setting Range

P. 1 General Display brightness 1 – 10

P. 2 General Sound 0 (=off)1 (=on)

P. 3 Occlusion Sensitivity of blocked suction detection. If there is ever blocked suction, the system switches off.

0 – 100 (=off)

P. 4 Occlusion Retrigger blocking time = time period after recognized blocked suction during which the system remains switched off.

0 – 100 (=off)

P. 5 Basic suction Time period of basic suction. The basic suction is active during the work breaks. The system works with reduced power.

1– 99 s continous (=on)

25 / 46

Page 26: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

6 • Settings and service functions

8011

6-44

3

10.1

6

Calling up Setting mode IMPORTANT! The Setting mode can only be called up once the "Setting mode inhibit" service function is off (=0), see page 27.

Fig. 6-1

1. When switching on the unit, keep the power switch and Basic suction button pressed until <S. 0> appears in the display.

2. Press the Activation on/off within 5 seconds. The Setting mode is called up if:– <P. 1> appears in the display,

– the Basic suction button illuminates and

– the %-s-display flashes

Change settings

1. Select the required setting (P. 1 to P. 6) with the Higher / Lower but-tons.

2. Press the START Suction button. The selected setting can now be changed with the Higher / Lower buttons.

3. To save the changed setting press the START Suction button again.

Quitting the setting mode 1. To exit the Setting mode, switch off the unit.

P. 6 Filter Filter status display 0 – 99 % 99 % (=new)3 % (=almost exhausted)0 % (=exhausted)

Display Assembly Setting Range

26 / 46

Page 27: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

6 • Settings and service functions80

116-

443

10.1

6

Service mode

Overview

Calling up Service mode

Fig. 6-2

1. When switching on the unit, keep the Higher / Lower buttons button pressed until < . 0> appears in the display.

2. Press the START Suction button within 5 seconds. The Service mode is called up if:– < . 1> appears in the display,

– the buttons illuminate and

– the %-s-display flashes

Display Assembly Setting Range

.1 Error Error list display 0.x last error

1.x last error but one

2.x to 9.x

Error log

.2 Error Delete error list 0 (=no)1 (=yes)

.3 Parameter Basic setting inhibit

0 (=off)1 (=on)

.4 Parameter Setting mode inhibit

0 (=off)1 (=on)

.5 Software Display version xx (00 – 99)

.

.

.

.

.

.

.

27 / 46

Page 28: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

6 • Settings and service functions

8011

6-44

3

10.1

6

Change settings

1. With the Higher / Lower buttons select the required setting ( . 1 to . 5).

2. Press the START Suction button. The selected setting can now be changed with the Higher / Lower buttons.

3. To save the changed setting press the START Suction button again.

Quitting the service mode 1. To exit the Setting mode, switch off the unit.

..

28 / 46

Page 29: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

7 • Remedying malfunctions80

116-

443

10.1

6

CHAPTER 7Remedying malfunctions

WARNING! Unplug the power cord from the outlet before exchanging parts of the unit or cleaning it.

ATTENTION! Adjustments, technical tests, modifications, maintenance and repair work may only be performed by Erbe or persons trained by Erbe. If the work is not performed by trained persons, Erbe accepts no liability and warranty rights become void.

ATTENTION! Only use original Erbe spare parts. The manufacturer accepts no liability and the warranty rights becomes void if original spare parts are not used.

Error Possible cause Action

Unit will not start up. Power plug not correctly plugged in.

Check connection at socket.

No line voltage. Check main fuse.

Fuse faulty. Change fuse. WARNING! Only spare fuses with the ratings indicated on the unit rating plate may be used.

The unit is not producing enough power.

Leakages in the hose lines. Check that the hose lines are securely attached. Replace hoses.

The main filter is blocked. Replace the main filter.

The in-line filter is blocked. Replace in-line filter.

"Filter exhausted" indicator con-stantly illuminated although the main filter is not exhausted.

Filter EEProm faulty. Replace the main filter.

Contact fault to the main filter. Clean contact surface (gold-plated) at the main filter. Clean contact springs in the filter channel. Take out the contact springs and check for ease of movement, and, if necessary, replace the contact board.

29 / 46

Page 30: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

7 • Remedying malfunctions

8011

6-44

3

10.1

6

Display y.1 ECB fault Check IES 2 and VIO electrosurgi-cal unit software versions for compatibility.

Display y.2 Excess temperature (> 60 °C) Switch off the unit and wait until the temperature in the interior of the unit has returned to normal.

Display y.3 Insufficient temperature (< 0 °C) Switch off the unit and wait until the temperature in the interior of the unit has returned to normal.

Display y.4 Software error (check total) Reinstall software. IMPORTANT! Software may only be re-installed and updated by Erbe Elektromedizin. Send unit for repair.

Display y.5 Setup parameter error Restart unit. Carry out re-initialisation.

Display y.6 Filter data erroneous Restart unit. Carry out re-initialisation. If necessary, replace filter.

Display y.7 Pressure sensor error Restart unit. Replace control board.

Error Possible cause Action

30 / 46

Page 31: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

8 • Maintenance and servicing80

116-

443

10.1

6

CHAPTER 8Maintenance and servicing

Who is allowed to perform servicing and maintenance work?

What is a technical safety check?

How often does a technical safety check have to be performed?

ATTENTION! Adjustments, tests, modifications, maintenance and repair work may only be performed by Erbe or persons trained by Erbe. If the work is not performed by trained persons, Erbe accepts no liability and warranty rights become void.

It is recommended that the technical safety check also be performed by Erbe or persons trained by Erbe.

IMPORTANT! The technical safety check is a preventive measure to examine whether the device is safe and ready for operation. In order to perform the vari-ous tests the current specifications and regulations of the particular country and the instructions in this service manual must be observed.

IMPORTANT! Erbe recommends performing a technical safety check after every repair, but at least once a year.

31 / 46

Page 32: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

8 • Maintenance and servicing

8011

6-44

3

10.1

6

Technical safety check – step by step

For simplification the device to be tested is referred to below as the "test specimen".

Safety informationWARNING! For safety reasons (personnel protection) the test specimen should

generally be operated by a suitable isolating transformer. An exceptional case is the tests for grounded conductor resistance, ground leakage current, and patient leakage current, in which the test specimen is supplied with current via the safety tester.

ATTENTION! In the event of a fault occurring in the test specimen or individual com-ponents during the technical safety check the test steps taken so far no longer apply. Remedy the defect and repeat the technical safety check from the beginning.

IMPORTANT! It is assumed that the user knows how to operate the test specimen, the test equipment, the measuring equipment, and auxiliary test equip-ment. The test instructions only apply in conjunction with the relevant test steps.

IMPORTANT! Test equipment, measuring equipment, and auxiliary test equipment (cables, test boxes, etc.) are listed separately at the beginning of each test unit. Where Erbe article numbers are specified, only original Erbe test equipment, measuring equipment, and auxiliary test equipment may be used.

IMPORTANT! The test report for the technical safety check can be requested from Erbe Technical Service Tübingen. For the address see address sheet on last page.

32 / 46

Page 33: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

8 • Maintenance and servicing80

116-

443

10.1

6

User manual and visual inspections• Test specimen and accessories (where enclosed) undamaged exter-

nally.• User manual present.• All labels on the test specimen (conformity declaration mark, rating

plate, and all wording) present and readily legible.

Electrical safety checks

List of checks The individual checks comprise:

• Ground terminal to chassis.• Ground terminal to potential equalization pin.• Measurement of the unit leakage current.

Performance tests

Test setup

• The test specimen is connected to the power supply via the power cord.

Test procedure

Power switch 1. Check power switch for smooth operation. The power switch must be easy to operate and must neither stick nor scrape.

2. Press power switch. The power switch must snap into the "ON" posi-tion and the test specimen must perform a system start.

Control buttons 1. Check the function of all control panel buttons. Press each button at least twice.

7-segment displays and LEDs 1. When starting the system, all 7-segment displays and LEDs must briefly light up at the same intensity.

IMPORTANT! Electrical safety is verified in accordance with the national specifica-tions and regulations.

33 / 46

Page 34: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

8 • Maintenance and servicing

8011

6-44

3

10.1

6

Activation

Testing and measuring equipment

Test setup

• The test specimen is connected to the power supply via the power cord.

• The single pedal footswitch IES 2 is connected.

Test procedure

Activation with footswitch 1. Set IES 2 to:– Operation suction: 80%– Basic suction: 2%– Run-on time: ON (=continuous operation)

2. Press footswitch pedal. The suction must be activated.3. Press the footswitch pedal again. The suction must be deactivated.

Activation with the START Suction button

1. Press the START Suction button. The suction must be activated.2. Press the START Suction button again. The suction must be deacti-

vated.

Erbe Art. No. Description

20322-101 IES 2 one pedal footswitch

34 / 46

Page 35: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

8 • Maintenance and servicing80

116-

443

10.1

6

Activation with VIO electrosurgical unit (via the ECB)

Testing and measuring equipment

Test setup

• as with previous testing and also• The power cord connects a VIO D or VIO S to the mains power supply.• The IES 2 and the VIO are connected to one another via the ECB.• Both units are switched on.• The two-pedal footswitch with ReMode is connected to the VIO.

Test procedure

Activation via VIO 1. Perform the following settings on the VIO:– Assign two-pedal footswitch to the bipolar receptacle.– IES settings for bipolar receptacle: Cut suction: OFF, coag suction

80%, basic suction: OFF, run-on time 1 sec2. Keep the COAG pedal of the VIO two-pedal footswitch pressed. The

suction must be activated.3. Release the COAG pedal of the VIO two-pedal footswitch. The suction

must be deactivated.4. Switch off VIO.

Erbe Art. No. Description

20322-101 IES 2 one pedal footswitch

20189-303 Dual-pedal footswitch with ReMode

35 / 46

Page 36: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

8 • Maintenance and servicing

8011

6-44

3

10.1

6

Operation suction

Test setup

• As for the previous test.

Test procedure

Operation suction without run-on time

1. Set IES 2 to:– Operation suction: 80%– Basic suction: 2%– Run-on time: 0 sec

2. Keep footswitch pedal pressed. The suction must remain activated as long as the pedal is kept pressed.

3. Release the footswitch pedal. The suction must be deactivated imme-diately.

Operation suction with run-on time

1. Set IES 2 to:– Operation suction: 80%– Basic suction: 2%– Run-on time: 5 sec

2. Keep footswitch pedal pressed. The suction must remain activated as long as the pedal is kept pressed.

3. Release the footswitch pedal. The suction must be deactivated after approx. 5 sec.

Blocked suction detection

Test setup

• As for the previous test.

Test procedure

1. Ascertain and note the sensitivity of blocked suction detection in the Setting mode (P. 3), see page 26. Set the blocked suction detection to 7 for this test.

2. Exit the Setting mode. Switch unit off and back on for this purpose.3. Keep footswitch pedal pressed. The suction must be activated.4. Keep the main filter input covered up (to simulate blocked suction).

Blocked suction detection works if: – the System blocked / Occlusion display lights red– a warning signal sounds (1 sec on, 1 sec off)– the suction is switched off immediately (without run-over time).

5. The sensitivity of blocked suction detection is reset to the old value.

36 / 46

Page 37: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

8 • Maintenance and servicing80

116-

443

10.1

6

Basic suction

Test setup

• As for the previous test.

Test procedure

1. Ascertain and note the basic suction time in the Setting mode (P. 5), see page 26. Set the blocked suction time to 20 s for this test.

2. Exit the Setting mode. Switch unit off and back on for this purpose.3. Keep footswitch pedal pressed. The suction must be activated.4. Set IES 2 to:

– Operation suction: 80%– Basic suction: 30%– Run-on time: 0 sec

5. Release the footswitch pedal. The suction must change audibly from the operation suction to the lower suction power of basic suction. After approx. 30 s basic suction, the suction has to be switch off com-pletely.

6. Reset the basic suction time is reset to the old value.

Filter recognition

Test setup

• As for the previous test.

Test procedure

1. Keep footswitch pedal pressed. The suction must be activated.2. During activation, remove the main filter. The suction must switch off

immediately and the Filter expended display must light up red.3. End the activation.4. Wait at least 10 s. Then replace the main filter. The Filter expended

display must go out.

37 / 46

Page 38: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

8 • Maintenance and servicing

8011

6-44

3

10.1

6

38 / 46

Page 39: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

9 • Replacing components80

116-

443

10.1

6

CHAPTER 9Replacing components

Safety information

Replacing components

The "wearing parts", such as seals or similar components, are designed such that, given normal operating conditions, it should not be necessary to replace components in the first 10 years (except for filter and consum-ables).

The following parts may have to be replaced during servicing:

Contact pins on the contact circuit board

Remove main filter; do not open the housing! Remove the contact pins by pulling them out firmly by hand.

Boards, fan and other internal components

Having removed the housing cover, they can be removed with suitable tools. When opening and closing the housing cover, be sure to be careful of the grounded conductor fastening on the cover.

Software Software may only be updated by Erbe Elektromedizin using suitable pro-gramming tools.

WARNING! Unplug the power cord from the outlet before exchanging parts of the unit or cleaning it.

ATTENTION! Adjustments, technical tests, modifications, maintenance and repair work may only be performed by Erbe or persons trained by Erbe. If the work is not performed by trained persons, Erbe accepts no liability and warranty rights become void.

ATTENTION! Only use original Erbe spare parts. The manufacturer accepts no liability and the warranty rights becomes void if original spare parts are not used.

39 / 46

Page 40: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

9 • Replacing components

8011

6-44

3

10.1

6

40 / 46

Page 41: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

10 • Spare parts80

116-

443

10.1

6

CHAPTER 10Spare parts

Fig. 10-1

40322-049 (2)

40322-029 (1)

40322-044 (1)

40322-043 (2)40322-041 (3)

40322-042 (3)

40322-031 (1)

40322-002 (1)

40322-036 (6)40322-004 (4)40322-008 (4)

51611-100 100 / 120 V~ (2) 40322-048 230 V~ (2)

40322-030 (1)

40322-045 (1)

41 / 46

Page 42: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

10 • Spare parts

8011

6-44

3

10.1

6

Fig. 10-2

4032

2-01

4 (1

)

4032

2-02

6 (1

)

4032

2-02

3 (2

)

4032

2-02

0 (1

)

4032

2-04

7 (4

)40

322-

019

(1)

4032

2-02

2 (1

)

4032

2-05

8 10

0 V

~ (1

)

40

322-

021

120

V~

(1)

2

30 V

~ (1

)

4032

2-01

6 (1

)

4032

2-01

8 (1

)

42 / 46

Page 43: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

10 • Spare parts80

116-

443

10.1

6

Fig. 10-3

40322-036 (4)

40322-035 (4)

40322-010 (1) 40322-032 (1)

40322-027 (1)

40322-050 (1)

40322-011 (1)

40322-040 (1)

40322-005 (1)

40322-007 (1)

20321-000 (1)

40322-046 (4)

40322-034 (1)

40322-054 (1)only 100 / 120 V~

43 / 46

Page 44: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

10 • Spare parts

8011

6-44

3

10.1

6

Fig. 10-4

30322-030 100 V~ 30322-032 120 V~ 30322-001 230 V~

40000-100 (1)

40000-329 (1)

40000-285 (1)

40000-311 (1)

44 / 46

Page 45: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

10 • Spare parts80

116-

443

10.1

6

Fig. 10-5

40322-009 (1)

40322-012 (1)

40322-059 (1)only 100 / 120 V~

40322-038 (1)

45 / 46

Page 46: SERVICE MANUAL - Erbe Elektromedizin GmbH of the User Manual ... 3 Main filter lock ... sor measuring bridge is then read in via a difference

10 • Spare parts

8011

6-44

3

10.1

6

Fig. 10-6

40322-024 (1)only 230 V~

40322-025 (1)

40322-013 (1)

46 / 46