scintillation cameras for single-photon imaging (planar
TRANSCRIPT
1
Review of Last Week:���Counting Statistics & Error Propagation
• Types of Error – systematic error; eg bias in experiment, error with fixed tendency, cannot be treated with
statistics, can be corrected for if known (calibration) -- effects accuracy – random error; inherent to instruments and/or nature of measurement (eg biological
variability), cannot be ‘corrected’ but can be minimized with large samples/trials, treated with statistical approaches -- effects precision
• Statistical Descriptions of Random Error – probability distribution functions
• Binomial; probability of observing certain outcome in N attempts when probability of single outcome is p.
• Poisson (limiting case of binomial: p small, N large); mean = variance = (standard deviation)2, asymmetric, positive means only
• Gaussian (Normal); widely used, symmetric, approximates Poisson for large means, mean and SD can be independent
• Reporting Statistics/Error - consistency in significant figures • Propagation of Error - errors from independent sources add in ‘quadrature’;
error in f(x, y) due to errors in x (x) and y (y) is f2 = |∂f/∂x|2x
2 + |∂f/∂y|2y2
simplest case: f = ax ± by, then error in f: f2 = (ax)2 + (by)2 (a, b = const.)
This week :���
Scintillation Camerasfor single-photon imaging
(Planar and SPECT) William Hunter, PhD#
University of Washington#Department of Radiology#
Suggested reading: Chapters 13&14.Physics in Nuclear Medicine by Cherry et al.
Planar and SPECT Cameras - Summary • System components:
– CollimatorDetectorElectronics
• Collimator – Types: Parallel, Converging, Diverging, Pinhole, Multi-pinhole – Performance: Penetration, Resolution, Efficiency
• Detector: – Components: Scintillator crystal, Optical spacer, PMTs – Performance: Efficiency, Intrinsic (spatial) resolution, Energy resolution
• Acquisition modes: – Frame vs List mode – Static (single), Dynamic (TAC), Gated (cardiac / respiration)
• Camera QA corrections: – Uniformity, Linearity, Photo-peak window, Multi-energy registration
• SPECT QA/QC: – Center-of-rotation, Head tilt, uniformity
The Planar Gamma Camera
Siemens e.cam
2
acquisition &#processing#computer#
Gamma Camera Instrumentation#
Signal#processing#electronics#
Scintillator#Lightguide#
PMT#
Collimator#
The Scintillation Camera: Detector System#
Detection Efficiency#
0.64
Light response function versus position#(spatial resolution)#
x
E
x7 x6 x5 x4 x3 x2 x1
Crystal#LG#
PMT#
Intrinsic spatial Resolution: <4 mm FWHM
3
Scatter#
detector - NaI(Tl)
PMTs
Gamma Camera Energy Spectra#
0
1 104
2 104
3 104
4 104
5 104
6 104
0 50 100 150 200
Nai(Tl) Energy Spectra (140 keV)
10% ER20% ER
Count
s
energy (keV)
Gamma Camera Energy Spectra#
140 keV photons, 9.5 mm crystal#
List of Nuclear Medicine ʻSingle Photonʼ Radionuclides • Tc99m # #140.5 keV # # 6.03 hours#• I-131 # #364, 637 keV # 8.06 days#• I-123 # #159 keV # # 13.0 hours#• I-125 # #35 keV # # 60.2 days#• In-111 # #172, 247 keV # 2.81 days#• Th-201 # #~70, 167 keV # 3.044 days#• Ga-67 # #93, 185, 300 keV# 3.25 days#
From: Physics in Nuclear Medicine (Sorenson and Phelps)
4
The Scintillation Camera: Collimators#
Parallel Hole Collimator#
detector - NaI(Tl)#
PMTs#
le#
Collimators - Septal Penetration#
Minimum septa thickness, t, for <5% septal penetration:#
From: Physics in Nuclear Medicine (Cherry, Sorenson and Phelps)#
Detector
l
d t
Collimator septa
Collimator Efficiency#
Collimators typically absorb well over 99.95% of all photons emitted from the patient. Trade-off between spatial resolution and detection efficiency.#
5
From: The Essential Physics of Medical Imaging (Bushberg, et al)#
Collimator Resolution Gamma Camera - spatial resolution
Rs = Rint2 + Rcol
2( )
From: Physics in Nuclear Medicine (Cherry, Sorenson and Phelps)
Types of Collimators#
From: The Essential Physics of Medical Imaging (Bushberg, et al)#
magnification
From: Physics in Nuclear Medicine (Cherry, Sorenson and Phelps)
Collimator: Resolution and Sensitivity#
6
The Scintillation Camera: Corrections and QA
How was this image produced?
Gamma Camera Processing Electronics(energy correction)#
2
3
45
6
7
8
10
11
12
131415
9
16
17
18
19
1
0
1 104
2 104
3 104
4 104
5 104
6 104
0 50 100 150 200
Energy channel vs. event location
10% ER (between)10% ER (over)
Cou
nts
energy (keV)
Gamma Camera Processing Electronics(linearity correction)
2
3
45
6
7
8
10
11
12
131415
9
16
17
18
19
1
From: Physics in Nuclear Medicine (Cherry, Sorenson and Phelps)
Gamma Camera Processing Electronics(linearity correction)
7
Additional Gamma Camera Corrections (sensitivity / uniformity)
Acquired from long uniform flood after energy and linearity corrections have been applied##Multiplicative correction##Adjusts for slight variation in the detection efficiency of the crystal##Compensates for small defects or damage to the collimator##Should not be used to correct for large irregularities#
Daily Gamma Camera QA Tests#
Photopeak window#
Flood uniformity#
From: The Essential Physics of Medical Imaging (Bushberg, et al)#
Multienergy spatial registration (e.g., Ga-67 (93-, 185-, and 300 keV) gamma rays)#
From: The Essential Physics of Medical Imaging (Bushberg, et al)#
properly adjusted# improperly adjusted#
Spatial Resolution Test#
From: The Essential Physics of Medical Imaging (Bushberg, et al)#
FWHM of LSF = 1.7 x (size of smallest bar resolved) #
8
Pulse Pile-up#
From: Physics in Nuclear Medicine (Sorenson and Phelps) and (Cherry, Sorenson and Phelps)
Energy spectra
Pile-up in image
The Scintillation Camera: Image Acquisition#
Image Acquisition • Frame mode (data stored as an image)#
- static#- single image acquisition#- can have multiple energy windows#
- dynamic#- series of images acquired sequentially#
- gated#- repetitive, dynamic imaging#- used for cardiac imaging#
• List-mode (data stored event by event)#- time stamps are included within data stream#- allows for flexible post-acquisition binning#- can result in very large data files#
Gated Acquisition
From: The Essential Physics of Medical Imaging (Bushberg, et al)#
9
Region of Interest (ROI) and Time-Activity Curves (TAC)
From: The Essential Physics of Medical Imaging (Bushberg, et al)
The Scintillation Camera: SPECT Imaging
Gamma Camera Imaging (SPECT)
Siemens e.cam (http://www.medical.siemens.com)
GE MG (http://www.gehealthcare.com)
www.medical.philips.com
Gamma Camera Imaging (SPECT)
www.medical.philips.com
10
Gamma Camera Imaging (gated cardiac SPECT)
www.cardiology.utmb.edu/ slides/Gated%20Cardiac%20SPECT.htm
www.medical.siemens.com/siemens/en_INT/ gg_nm_FBAs/files/brochures/Cardiology_Insert_4DMSPECT.pdf
• Images obtained during individual phases of cardiac cycle
• Gating based on EKG • 8-16 frames per cycle • Allows assessment of global
and regional LV function
Additional QA/QC Tests for SPECT Center-of-Rotation (COR) Alignment Artifact
From: The Essential Physics of Medical Imaging (Bushberg, et al)
Small COR misalignement will blur image, while larger misalignement will cause donut shape artifacts.
Additional QA/QC Tests for SPECT Camera Head Tilt
From: The Essential Physics of Medical Imaging (Bushberg, et al)
Additional QA/QC Tests for SPECT
Detector non-uniformity artifact
From: The Essential Physics of Medical Imaging (Bushberg, et al)
Rings around center of rotation (transverse view)
This artifact differs from a COR-misalignment artifact in that it only appears around the AOR.
11
Planar and SPECT Cameras - Summary • System components:
– CollimatorDetectorElectronics
• Collimator – Types: Parallel, Converging, Diverging, Pinhole, Multi-pinhole – Performance: Penetration, Resolution, Efficiency
• Detector: – Components: Scintillator crystal, Optical spacer, PMTs – Performance: Efficiency, Intrinsic (spatial) resolution, Energy resolution
• Acquisition modes: – Frame vs List mode – Static (single), Dynamic (TAC), Gated (cardiac / respiration)
• Camera QA corrections: – Uniformity, Linearity, Photo-peak window, Multi-energy registration
• SPECT QA/QC: – Center-of-rotation, Head tilt, uniformity