satelec ® range. treat pulp and periapical diseases endo treatment (root canal) should be performed...
TRANSCRIPT
SATELEC®
RANGE
Treat pulp and periapical diseases
Endo treatment (root canal) should be performed in case of:- Pulp necrosis or irreversible pulpitis- Vital pulp(but poor prognosis of pulp vitality, high risk of pulp exposure during the coronal restoration, root amputation or hemisection)
Contra-indications:• Patient with high risk of infective endocarditis• Tooth with no functional future (wisdom teeth) • Tooth with inadequate periodontal support
ENDO basisENDO basis
Endo BasisEndo BasisTechniques comparisonTechniques comparison
Manual technique+: Tactile sense +: less risk of perforation (NiTi)-: Numerous files are needed-: Time consuming-: Access difficulties (posterior areas)
Rotating technique+: Fast canal preparation+: Auto-reverse- : Fatigue of the files
Reciprocity technique+: Few files needed (1hand file + 1 mechanical file)+: Same movement as manual (F+R)+: Secure (less breakage, less apical pressure, etc)
Canal morphology variesCurved / straightLarge / SmallLong / short
The canals’ common feature is their taper
Prepared canals: Taper 10 % in coronal area, 2 to 4% in the apical zone
To fit all canals, files To fit all canals, files present present different shapes: different shapes:
Different lengthDifferent lengthMost popular length:
21mm, 25mm, 29mm Other: 15mm, 19mm,
23mm
Different diameters (ISO)Different diameters (ISO)
Different tapersDifferent tapers4%, 6%, 8%, 12%
Different files shape + Clockwise and counterclockwise cutting edges
Radial-landed (Smooth) ex. ProFile (Dentsply), HERO (Micro-Méga)
Straight edges (Cut) Ex. ProTaper (Dentsply), RaCe (FKG), Sequence (Brasseler)
Ex. EndoEZE AET (Ultradent)
Other shapes
Bibliography:A la recherche des limes endodontiques idéales, Dr. Mc Spadden, Dental Tribune, 2009
Ø15 Ø20 Ø25 Ø30 Ø35 Ø40
Endo basisEndo basis
ENDO treatment protocolENDO treatment protocol
SOPIX² Pierre RollandPrimacaineTM
XMind
PSP !X
Pierre RollandEDETATSalvizol EDTA Pierre Rolland
Sealite - Cortisomol
SO4 Lateral Gutta percha condensation
ENDO treatment protocolENDO treatment protocol
irrisafe
Obturation - Performed on dry canal.Lateral condensation (use of spreaders)Lateral warm condensation (SO4 ultrasonic tip with no irrigation)Vertical warm compaction (manually or mechanically), etc.
Coronal obturation + Prosthetics
1 2 3 4
5 6 7 8
9
ENDO retreatmentENDO retreatment
Any failure of canal treatment will lead to a root canal retreatment
SATELECSATELEC®® in ENDO in ENDO
- Excellent reputation in endodontics
- The convergence of years of technical
endo experience
- Wide range of specialized tips
SATELECSATELEC®® in ENDO in ENDO
irrisafe
ENDO MOTORENDO MOTOR
Endodontic motor with rotating and reciprocating motions. Replacing the actual I-ENDO DUAL
- Encoder technology included - E-type (ISO 3964) connection micromotor - 6 contra-angle ratios pre-set: (1:1 ; 6:1 ; 8:1 ; 16:1 ; 20:1 ; 32:1) - Speed without contra-angle: 1,000 to 10,000Rpm Speed with a 16:1 100 to 600Rpm - Torque with a 16:1 0.5 to 7.5Ncm + MAX (10Ncm) - Many new Rpm settings to respond to all file systems
The ENDODUAL is the first unit having an encoder coupled with an accurate micromotor.
This patented encoder calculates an algorithm ensuring:
- In rotating mode high speed accuracy.
- In reciprocating (F+R) mode an exact angle measurement angle measurement for:
- Evaluating the angular position,
- Getting the best performances for each file system.
SATELEC SATELEC ExclusivityExclusivity
Encoder
256 positions analyzed
Signal
Analysis of:128 positions (forward)128 positions (reverse)
SATELEC ExclusivitySATELEC Exclusivity
Precision and rapidityPrecision and rapidity
Versatile, complete, light and compact, the ENDO
micromotor offers
a new vision of endodontics.
EncoderThe encoder calculates an algorithm for greater precision,
Integrated calibration systemThe integrated calibration system checks the contra-angle’s mechanical inertia for a perfect micro-motor/contra-angle adjustment.
MicroprocessorAdvanced microprocessor for an extremely short response time.
Freedom of settingsFreedom of settingsReduction rate
RPM
1:1 6:1 8:1 16:1 20:1 32:11000 150 125 100 100 1001500 200 150 150 120 1502000 250 200 200 150 1752500 300 250 250 200 1803000 350 300 300 250 2003500 400 350 350 280 2504000 450 400 400 300 2804500 500 450 450 350 3005000 550 500 500 4005500 600 550 550 4506000 650 600 600 5006500 700 650
7000 750 7007500 800 7508000 850 8008500 900 8509000 950 9009500 1000 950
10000 1100 1000
1200 11001300 12001400 125015001600
Ncm with a 16:10.5 1 1.5 2 2.
53 3.5 4 4.5 5 5.5 6 6.
57 7.5 MA
X
Torque is not needed in Reciprocating mode
(Max)
Open System
TAKE CONTROL OF THE WAY YOU PRACTICETAKE CONTROL OF THE WAY YOU PRACTICE
An open system compatible with all rotating and reciprocating files providing access to future endodontic instrument innovations via its:
- 2 modes available (Rotating and Reciprocating), - Wide choice of contra-angle ratios ISO 3964 connections,- Independent torque, speed and angle settings.
• ISO contra-angle chuck
• Rotating files:– All rotating files are compatible with the ENDODUAL– Follow the manufacturer settings
• Reciprocating files:– Wave one/Reciproc (30°FWD – 150°REV) Grip sensation and cutting differences
Wave One Ø25 Taper 8%, Ø21 Taper 6%, Ø40 Taper 8% - Reciproc Ø25 Taper 8%, Ø40 Taper 6%, Ø50 Taper 5%,
– In the near future, other file manufacturers will most probably launch their own reciprocating files
• Rotating files tested in reciprocating motion:– S2 and F2 from Denstply Maillefer (150° FWD – 30° REV)
Files compatible with the Files compatible with the ENDODUALENDODUAL
Reciproc® file WaveOne® fileProtaper ® file
360° endodontics360° endodonticsFreedom of motionFreedom of motion
Color-coded buttons:
• GREEN: ROTATING FUNCTIONS
• GREY: RECIPROCATING FUNCTIONS
• WHITE: WHITE: COMMON FUNCTIONS
RotatingRotating
CalibrationCalibration
Controls the contra-angle’smechanical inertia for a perfectmicromotor/contra-angle
match.
Repetitive sterilizations and contra-angle wear can lead to a lack of precision.
This function accurately measures the properties of the micromotor and contra-angle handpiece prior to use for exact values.
RotatingRotating10 customizable automatically saved programs
available in each mode.
Switching from Rotating to reciprocating
For micromotor/ contra-angle match
Select the contra-angle ratio (1:1 ; 6:1 ; 8:1 ;16:1 ; 20:1 ; 32:1)
Select the speed rotation 100 to 600Rpm (16:1)Select the torque 0.5 to 7.5Ncm + MAX (10Ncm) (16:1)
Select the AutoreverseSTOP, AR, SAR
Clockwise/counterclockwise rotation
Parameters selection
RotatingRotating
When file reaches the set torque value, the motor runs reverse until the motor has been stopped. Next start will be in Forward.
Stops without reverse rotation.
When file reaches the set torque value, the motor runs reverse until the file is freed andAutomatically returns to Forward.
STP - “STOP” Autoreverse
AR – “Standard” Autoreverse
Beeping sound notifies user when 75% of the preset torque has been reached and when preset torque limit has been exceeded.
SAR – “Special” Autoreverse
The time from « x » to « y » must be as short as possible
(less than 1/1000 sec. )
Microprocessor analysis
Between X and Y, the torque is increasing progressively
until MAX value
Rotating movementRotating movement
Manufacturer
Files Protocol Ncm Speed Rpm
DentsplyVDW
ProTaper®
MtwoClick to enlarge 3 - 4
1 - 1.51.5 - 22- 33 - 42 - 31.5 - 2
250
Manufacturer
Files Protocol Ncm Speed Rpm
DentsplyVDW
ProTaper®
MtwoClick to enlarge 3 - 4
1 - 1.51.5 - 22- 33 - 42 - 31.5 - 2
250
FKG Pre-RaCe®
RaCe®
Click to enlarge 0.5 - 1.5 500-600
Manufacturer
Files Protocol Ncm Speed Rpm
DentsplyVDW
ProTaper®
MtwoClick to enlarge 3 - 4
1 - 1.51.5 - 22- 33 - 42 - 31.5 - 2
250
FKG Pre-RaCe®
RaCe®
Click to enlarge 0.5 - 1.5 500-600
Micro-Mega Endoflare®
Hero®
Hero Shaper®
Hero Apical®
R-Endo®
Revo-S®
Click to enlarge 0.6 - 3 400
Manufacturer
Files Protocol Ncm Speed Rpm
DentsplyVDW
ProTaper®
MtwoClick to enlarge 3 - 4
1 - 1.51.5 - 22- 33 - 42 - 31.5 - 2
250
FKG Pre-RaCe®
RaCe®
Click to enlarge 0.5 - 1.5 500-600
Micro-Mega Endoflare®
Hero®
Hero Shaper®
Hero Apical®
R-Endo®
Revo-S®
Click to enlarge 0.6 - 3 400
Sybron endo Twisted file™ TF 4-5 500-600
VDW Mtwo® 1.77 250-350
Mani NRT™ 1.47-1,772.45-2.94
300
Komet AlphaKite™e-Kite™
3 250-500
ReciprocatingReciprocating
ReciprocatingReciprocating10 customizable automatically saved programs
available in each mode.
Switching from Rotating to reciprocating
For micromotor/ contra-angle match
Select the contra-angle ratio (1:1 ; 6:1 ; 8:1 ;16:1 ; 20:1 ; 32:1)
Select the speed rotation 100 to 600Rpm (16:1)
Parameters selection
Dedicated to reciprocating (F+R) mode. Adjustment from 10° to 360° in forward and in reverse motion by step of 5°
Whatever the file used, the angle should be higher in cutting direction
than in non cutting edges direction. Ex : Wave One (150° REV – 30° FWD),
Protaper (150°FWD – 30° REV)
Reciprocating Reciprocating movement movement
Major Advantages
• Cyclic fatigue File is deformed due to tension and compression motion performed in curved canals Can be reduced:
- Following the file manufacturer settings (speed)
• Torsion File’s tip is blocked in the canal but the contra-angle continues its rotation. Can be reduced:
- Using system for limiting the torque- Following the suggested settings of the
manufacturer
FracturesFractures
Whatever the file used, the angle should be higher in cutting direction
than in non cutting edges direction. Ex : Wave One (150° REV – 30° FWD),
Protaper (150°FWD – 30° REV)
Reciprocating Reciprocating movement movement
Major Advantages
Single use
2 files brands dedicated to
reciprocating motions are today
existing:
- Wave one® from Dentsply Maillefer
- Reciproc® from VDW:
more aggressive but also more fragile
Wave one /Reciproc®:
Reverse helix
Fixed angulations:
150° REV - 30° FWD 150° REV - 30° FWD
Files - ReciprocatingFiles - Reciprocating
Extract from the article: Mise en forme canalaire mino-instrumentale en mouvement réciproque WaveOne® et reciproc®, W Pertot, J. Webber, P. Machtou, S. Kuttler
Standard protocol:Standard protocol:1.K10 file: Canal exploration2.WaveOne until penetration length of the manual file (back-and-forth motion, debris removal on the file, irrigation)1.Manual file to pre-enlarge until the apical third2.WaveOne until the working length + irrigation
Files - ReciprocatingFiles - ReciprocatingMaillefer
VDW
Canal preparation using only one Ni-Ti rotary instrument: preliminary observations. G. YARED, International Endodontic journal, 2007.
Files - ReciprocatingFiles - Reciprocating
Files pricesFiles prices• Rotating files:
– Hero Shaper Micro-Mega 26,42€ without tax for 3 files (8,80€/file)
– ProTaper 39€ without tax for 6 files 6,5€/file
According to studies, a file life cycle in continuous rotating motion According to studies, a file life cycle in continuous rotating motion variesvaries
between 4 to 10 uses (depending on the shape of the canal treated).between 4 to 10 uses (depending on the shape of the canal treated).
• Reciprocating files:– WaveOne 38,98€ without tax for 3 single use files
WaveOne (Small, primary, large) Reciproc R25, 40, 50 (length 21, 25, 31)
Files dedicated to the reciprocating motion (reverse helix)
Fixed angulations: 150° - 30°Price analysis:Price analysis:
Max 10€ without tax/file – Uses: max x10 - min x4 = [1€-2,5€]Max 10€ without tax/file – Uses: max x10 - min x4 = [1€-2,5€]WaveOne 13€ without tax/fileWaveOne 13€ without tax/fileWaveOne/Reciproc files are 5 to 13 times more expensive than traditional rotating filesWaveOne/Reciproc files are 5 to 13 times more expensive than traditional rotating files
• New design,• Larger display,• New PCB,• 6 ratios available (8:1 added),• New Ncm and Rpm settings added,• Simplified interface,• Improvements in micromotor manufacturing process and cord
resistance,– Metallic micromotor connector– New micromotor assembling
• Controlling equipment: – Special investment in an angle and torque controlling machine.– Soft changes in reciprocating mode : 1st focusing on the angles then on the
speed. Giving the priority to the angle will enable to have a better angular accuracy with and without load.
SUM UPSUM UP
F28770 (mains cord equipped with 4 international plugs)
Equipped with:Endo motor and cordSingle footswitch1 micromotor adhesive support
Freedom in
endodontics
ENDODUAL benefitsENDODUAL benefits
Thank you