Electrical Safety Tester
GPT-9000 Series
USER MANUALGW INSTEK PART NO. 82PT-99040EA1
ISO-9001 CERTIFIED MANUFACTURER
-
7/25/2019 Safety Tester Manual
2/149
This manual contains proprietary information, which is protected bycopyright. All rights are reserved. No part of this manual may bephotocopied, reproduced or translated to another language withoutprior written consent of Good Will company.
The information in this manual was correct at the time of printing.However, Good Will continues to improve products and reserves therights to change specification, equipment, and maintenanceprocedures at any time without notice.
Good Will Instrument Co., Ltd.No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan
-
7/25/2019 Safety Tester Manual
3/149
SAFETY INSTRUCTIONS
3
Table of ContentsSAFETY INSTRUCTIONS .................................................. 5
GETTING STARTED .......................................................... 9
GPT-9000 Series Overview ................... 10
Appearance .......................................... 14
Set Up .................................................. 20
OPERATION .................................................................. 27
Menu Tree ............................................ 29
Test Lead Connection ......... ........ ......... . 33
Manual Testing .................................... 35
Common Utility Settings ...................... 71
Automatic Tests ................................... 78
EXTERNAL CONTROL .................................................... 95
External Control Overview .................... 96
REMOTE CONTROL ...................................................... 101
Interface Configuration ...................... 102
Command Syntax ............................... 106
Command List ................................... 109
Error Messages .................................. 136
FAQ .............................................................................. 137
APPENDIX .................................................................... 139
Fuse Replacement .............................. 139
Error Messages .................................. 140
GPT-9000 Specifications .................... 142
GPT-9000 Dimensions ....................... 146
Declaration of Conformity .................. 147
-
7/25/2019 Safety Tester Manual
4/149
GPT-9000 Series User Manual
4
INDEX .......................................................................... 148
-
7/25/2019 Safety Tester Manual
5/149
SAFETY INSTRUCTIONS
5
SAFETY INSTRUCTIONSThis chapter contains important safetyinstructions that you must follow duringoperation and storage. Read the following beforeany operation to ensure your safety and to keepthe instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on theinstrument.
WARNINGWarning: Identifies conditions or practices thatcould result in injury or loss of life.
CAUTIONCaution: Identifies conditions or practices thatcould result in damage to the GPT-9000 or to otherproperties.
DANGER High Voltage
Attention Refer to the Manual
Protective Conductor Terminal
Frame or Chassis Terminal
Earth (ground) Terminal
-
7/25/2019 Safety Tester Manual
6/149
GPT-9000 Series User Manual
6
Do not dispose electronic equipment as unsortedmunicipal waste. Please use a separate collectionfacility or contact the supplier from which this
instrument was purchased.
Safety Guidelines
GeneralGuideline
CAUTION
Do not place any heavy object on the GPT-9000.
Avoid severe impact or rough handling thatleads to damaging the GPT-9000.
Do not discharge static electricity to the GPT-9000.
Use only mating connectors, not bare wires, forthe terminals.
Do not block the cooling fan opening.
Do not disassemble the GPT-9000 unless you arequalified.
(Measurement categories) EN 61010-1:2010 specifies themeasurement categories and their requirements as follows. the GPTfalls under category I.
Measurement category IV is for measurement performed at thesource of low-voltage installation.
Measurement category III is for measurement performed in thebuilding installation.
Measurement category II is for measurement performed on thecircuits directly connected to the low voltage installation.
Measurement category I is for measurements performed oncircuits not directly connected to Mains.
Power Supply
WARNING
AC Input voltage range:100/120/220/230VAC 10%
Frequency: 50Hz/60Hz
To avoid electrical shock connect the protectivegrounding conductor of the AC power cord toan earth ground.
-
7/25/2019 Safety Tester Manual
7/149
SAFETY INSTRUCTIONS
7
Cleaning theGPT-9000
Disconnect the power cord before cleaning.
Use a soft cloth dampened in a solution of milddetergent and water. Do not spray any liquid.
Do not use chemicals containing harsh materialsuch as benzene, toluene, xylene, and acetone.
OperationEnvironment
Location: Indoor, no direct sunlight, dust free,almost non-conductive pollution (Note below)
Relative Humidity: 70% (no condensation)
Altitude: < 2000m
Temperature: 0C~40C
(Pollution Degree) EN 61010-1:2010 specifies the pollution degreesand their requirements as follows. The GPT-9000 falls under degree2.
Pollution refers to addition of foreign matter, solid, liquid, orgaseous (ionized gases), that may produce a reduction of dielectricstrength or surface resistivity.
Pollution degree 1: No pollution or only dry, non-conductivepollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollutionoccurs. Occasionally, however, a temporary conductivity causedby condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-conductive pollution occurs which becomes conductive due tocondensation which is expected. In such conditions, equipmentis normally protected against exposure to direct sunlight,precipitation, and full wind pressure, but neither temperaturenor humidity is controlled.
Storageenvironment
Location: Indoor Temperature: -10C to 70C
Relative Humidity: 85% (no condensation)
Disposal Do not dispose this instrument as unsortedmunicipal waste. Please use a separate collectionfacility or contact the supplier from which thisinstrument was purchased. Please make surediscarded electrical waste is properly recycled toreduce environmental impact.
-
7/25/2019 Safety Tester Manual
8/149
GPT-9000 Series User Manual
8
Power cord for the United Kingdom
When using the safety tester in the United Kingdom, make sure the
power cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHEDIMPORTANT: The wires in this lead are coloured in accordance with thefollowing code:Green/ Yellow: EarthBlue: NeutralBrown: Live (Phase)
As the colours of the wires in main leads may not correspond withthe coloured marking identified in your plug/appliance, proceedas follows:
The wire which is coloured Green & Yellow must be connected tothe Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal
which is marked with the letter N or coloured Blue or Black.The wire which is coloured Brown must be connected to theterminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipmentor contact the supplier.
This cable/appliance should be protected by a suitably rated andapproved HBC mains fuse: refer to the rating information on the
equipment and/or user instructions for details. As a guide, a cableof 0.75mm2should be protected by a 3A or 5A fuse. Largerconductors would normally require 13A types, depending on theconnection method used.
Any exposed wiring from a cable, plug or connection that isengaged in a live socket is extremely hazardous. If a cable or plug isdeemed hazardous, turn off the mains power and remove the cable,any fuses and fuse assemblies. All hazardous wiring must beimmediately destroyed and replaced in accordance to the abovestandard.
-
7/25/2019 Safety Tester Manual
9/149
GETTING STARTED
9
GETTING STARTEDThis chapter describes the safety tester in anutshell, including its main features and front /rear panel introduction. After going through theoverview, please read the safety considerations inthe Set Up chapter.
POWER
START STOP
REMOTE
M A NU/ AUT O E DIT / SA V E UT IL I TY
ESC PAGE
GPT-9803 Tester
HIGH VOLTAGE
CAUTION 5.0kVAC MAX.6.0kVDC MAX.
AC/ DC Withstanding Voltage/
Insulation Resistance
RETURN
PASS FAIL READY TEST
GPT 9000 Series Overview .............................................. 10
Series lineup ..............................................................................................10Main Features ...........................................................................................11
Accessories ................................................................................................12Package Contents .....................................................................................13
Appearance .................................................................... 14
GPT-9801/9802/9803 Front Panel .........................................................14GPT-9000 Rear Panel ................................................................................18
Set Up ........................................................................... 20
Line Voltage Connection and Power Up ..............................................20Installing the Optional GPIB Card .........................................................22Workplace Precautions ............................................................................23Operating Precautions .............................................................................24Basic Safety Checks ..................................................................................26
-
7/25/2019 Safety Tester Manual
10/149
GPT-9000 Series User Manual
1
GPT-9000 Series Overview
Series lineupThe GPT-9000 Series Safety Testers are AC/DC withstandingvoltage, insulation resistance and ground bond safety testers. TheGPT-9801 is an AC withstanding voltage tester, the GPT-9802 is anAC/DC withstanding voltage tester and the GPT-9803 is anAC/DC withstanding voltage and insulation resistance tester. TheGPT-9804 includes all the functions of the other models as well asground bond testing. All models can operate at up to 5kVAC for
AC withstanding voltage testing. The GPT-9802, GPT-9803 andGPT-9804 can also operate at up to 6kVDC for DC withstandingvoltage testing.The GPT-9000 Series can store up to 100 manual tests, as well asrun up to 16 manual tests sequentially as an automatic test,allowing the safety testers to accommodate any number of safetystandards, including IEC, EN, UL, CSA, GB, JIS and others.
Model name ACW DCW IR GB
GPT-9801
GPT-9802
GPT-9803
GPT-9804
-
7/25/2019 Safety Tester Manual
11/149
GETTING STARTED
11
Main Features
Performance ACW: 5kVAC
DCW: 6kVDC
IR: 50V~1000V (50V steps)
GB: 3A~30A
Features Ramp up time control
Safety discharge
100 test conditions (MANU mode)
100 automatic tests (AUTO mode)
Over temperature, voltage and currentprotection
Pass, Fail, Test, High Voltage and Readyindicators
PWM output (90% efficiency, increasedreliability)
Interlock (configurable).
Interface Remote control start/stop interface terminal
RS232/USB interface for programming
Optional GPIB interface for programming
Signal I/O port for pass/fail/test monitoring
and start/stop control/interlock
-
7/25/2019 Safety Tester Manual
12/149
GPT-9000 Series User Manual
12
Accessories
StandardAccessories Part number Description
GHT-114 x1 Test lead
Region dependent Power cord
GTL-115 x1 GB Test leads (GPT-9804)
N/A Remote terminal male plug
N/A Interlock key
OptionalAccessories
Part number Description
GHT-205 High Voltage Test Probe
GHT-113 High Voltage Test Pistol
GTL-232 RS232C cable
GTL-248 GPIB cable
GTL-247 USB cableGRA-402 Rack Adapter Panel (19,
4U)
Options Part number Description
Opt.01 GPIB Interface GPIB module
-
7/25/2019 Safety Tester Manual
13/149
GETTING STARTED
13
Package Contents
Check the contents before using the GPT-9000.
Opening the box
Contents(single unit)
GPT-9000 unit
Quick Start guide
User manual CD
CTC (CalibrationTraceable Certificate)
Power cord x1 (regiondependent)
GHT-114 test leads x1
GTL-115 test leads x1(GPT-9804)
Remote terminal male
plug
Interlock key
NoteKeep the packaging, including the box, polystyrenefoam and plastic envelopes should the need ariseto return the unit to GW Instek.
-
7/25/2019 Safety Tester Manual
14/149
GPT-9000 Series User Manual
14
Appearance
GPT-9801/9802/9803 Front Panel
POWER
START STOP
REMOTE
M AN U/ AU TO E DI T/ SA VE U TI LI TY
ESC PAGE
GPT-9803 Tester
HIGH VOLTAGE
CAUTION 5.0 kVAC MAX.6.0 kVDC MAX.
AC / DC Withstanding Voltage /
Insulation Resistance
RETURN
PASS FAIL READY TEST
PASS/FAIL indicatorsDirectional keysFunction keys
HIGH VOLTAGEindicator
RETURNterminal
REMOTE terminalSTOP button
START button
POWERbutton
Display
Scroll wheelConfiguration keys
READY indicator
TEST indicator
HIGH VOLTAGEoutput terminal
GPT-9804 Front Panel
l
r
SENSE H Rx SENSE L
SOURCE LSOURCE H
RETURN
T
I . .. .
T T
T T T T
SOURCE L
SENSE L &
RETURN TerminalSENSE H
SOURCE H
Display 240 X 64 dot matrix display (LCD)
Function keys The function keys correspond to the soft-keysdirectly above on the main display.
Pass/Failindicators
PASS FAIL
The PASS and FAIL indicatorslight up upon a PASS or FAIL testresult at the end of a manual testor automatic test.
ESC key ESC
The ESC key is used to exit out ofa menu or cancel a setting.
-
7/25/2019 Safety Tester Manual
15/149
GETTING STARTED
15
PAGE key PAGE
The PAGE key is used to viewautomatic test information andtest results.
Directional arrowkeys
The directional arrow keys areused to navigate menus andparameter settings.
READY indicator READY
The READY indicator is lit whenthe tester is ready to begin testing.The STOP button is used to put
the tester into READY status.
TEST indicator TEST
The TEST indicator is lit when atest is on. The START button isused to put the tester into TESTstatus.
HIGH VOLTAGEindicator
HIGH VOLTAGE
CAUTION 5.0 kVAC MAX.6.0 kVDC MAX.
The HIGH VOLTAGE indicatorwill light up when an output
terminal is active. Only after thetest has finished or stopped willthe indicator turn off.
HIGH VOLTAGEoutput terminal
HIGH VOLTAGE
. .
6.0 kVDC MAX.
The HIGH VOLTAGE terminaloutput is used for outputting thetesting voltage. The terminal isrecessed for safety. This terminal
is used in conjunction with theRETURN terminal.
WARNING
USE EXTREME CAUTION.Do not touch the HIGH VOLTAGE terminalduring testing.
-
7/25/2019 Safety Tester Manual
16/149
GPT-9000 Series User Manual
16
RETURN terminalGPT-9801/9802/9803
RETURN
The RETURN terminal is thereturn terminal for all tests.
For GB testing, the RETURNterminal is also the SENSE Lterminal.
RETURN/SENSE L,SENSE H,SOURCE L,SOURCE Hterminal(GPT-9804)
SENSE H Rx SENSE L
SOURCE LSOURCE H
RETURN
The RETURN terminal isused for IR, DCW and ACWtests. For GB tests, it is usedas the SENSE L terminal.
SENSE H terminal for GBtests. GPT-9804 only.
SOURCE L terminal for GBtests. GPT-9804 only.
SOURCE H terminal for GBtest. GPT-9804 only.
Scroll wheel The scroll wheel is used to editparameter values.
UTILITY key UTILITY
Used to enter the MANU Utility or
Common Utility menu.
EDIT/SAVE key EDIT/SAVE
Used to start editingMANU/AUTO tests as well assave settings and parameters.
MANU/AUTOkey
MANU/AUTO
The MANU/AUTO key is used toselect manual tests (MANU) orautomatic tests (AUTO).
-
7/25/2019 Safety Tester Manual
17/149
GETTING STARTED
17
REMOTEterminal
REMOTE
The REMOTE terminal is used toconnect to a remote controller.
STOP button STOP
The STOP button is used tostop/cancel tests. The STOPbutton will also put the safetytester in the READY status tobegin testing.
START button START
The START button is used to starttests.
The START button can be used tostart tests when the tester is in theREADY status. Pressing theSTART button will put the testerin the TEST status.
POWER switch POWER
Turns the power on. The safetytester will always start up with thelast test setting from when theinstrument was last powereddown.
-
7/25/2019 Safety Tester Manual
18/149
GPT-9000 Series User Manual
18
GPT-9000 Rear Panel
SIGNAL I/O USB A port RS232 port Fan
Fuse selectorLine voltageOptional GPIB port
GND
TOAVOIDELECTRICSHOCKTHEPOWERCORD
PROTECTIVEGROUNDINGCONDUCTORMUSTBE
ONLYWITHSPECIFIEDTYPEANDRATEDFUSE.NOOPERATORSERVICEABLE COMPONENTSINSIDE
.DONOTREMOVECOVERS REFERSERVICINGTO
FORCONTINUEDFIRE PROTECTION.
REPLACE
CONNECTEDTOGROUND .
QUALIFIEDPERSONNEL.
WARNING
AC
LINEVOLTAGE
100V
230V
220V
120V
SELECTION
207~250V
RANGE
90~110V
198~242V
108~132V
(50/60Hz)
T5A250V
FUSE
POWER MAX.500VA
GND
SIGNAL I /O RS 232
SER.NO.L B
ENSURETHEPOWERISREMOVEDFROMTHEINSTRUMENTBEFOREREPLACINGTHEFUSE
GPIB
T2.5A250V
SIGNAL I/O port SIGNAL I / O
The SIGNAL I/O port is used tomonitor the tester status (PASS,FAIL, TEST) and input (START/STOP signals). It is also used withthe Interlock key.
USB A port Used for remote control.
RS232 interfaceport
RS232
Used for remote control andfirmware updates.
FanExhaust fan.
GND GND
Connect the GND (ground)terminal to the earth ground.
Line voltage input 220230
100
120
Line voltage input:100/120/220/230VAC 10%
-
7/25/2019 Safety Tester Manual
19/149
GETTING STARTED
19
Line voltage fuse220
230
100
120
Line voltage selector and fuse:
100V/120V T5A 250V220V/230V T2.5A 250V
Optional GPIBport
GPIB
Optional GPIB interface for remotecontrol.
-
7/25/2019 Safety Tester Manual
20/149
GPT-9000 Series User Manual
2
Set Up
Line Voltage Connection and Power Up
Background Before powering up the GPT-9000 ensure thecorrect voltage has been selected on the rearpanel. The GPT-9000 supports line voltages of100V/120V/220V and 230V.
Steps 1. Check the line voltage and the fusein the fuse holder.
Page139
The desired line voltageshould line up with thearrow on the fuse holder.
220
100120
230
2. Connect the power cord tothe AC voltage input.
220
100120
230
3. If the power cord does nothave an earth ground,ensure the groundterminal is connected to an
earth ground.
GND
WarningEnsure the power cord is connected to an earthground. Failure could be harmful to the operatorand instrument.
4. Press the Power button. POWER
-
7/25/2019 Safety Tester Manual
21/149
GETTING STARTED
21
5. When the unit is powering up, all the LEDindicators will light. Check to make sure all 5LED indicators are working.
6. Check to make sure the System Self Test passeswithout errors.
S y s e mt
H a r w a r d e C
h
C h e c i g . .n .
F i r w a r m e C h
SYS T E SM E L F T E S Tk
e c i g . .n .k
e c i g . .n .k
After the System Self Test completes, the tester willgo into VIEW status and be ready to operate.
I R G BA CW D CW
E
F R E Q = 0 H z6
0 100k V RA C = F FO
EIV MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
W
VIEW status
MM A UN N A_2M A N U = * * * - 0 0
m A
WARNINGSee the Appendix on page140 for details if a self-test error is detected.
-
7/25/2019 Safety Tester Manual
22/149
GPT-9000 Series User Manual
22
Installing the Optional GPIB Card
Background The optional GPIB is a user-installable option.Follow the instructions below to install theGPIB card.
WARNINGBefore installing the optional GPIB card ensure theGPT-9000 turned is off and disconnected frompower.
Steps 1.
Remove the screws from the rear panel coverplate.
2. Insert the GPIB card into the two slots on eitherside of the opening. Push the card gently until itis fully inserted.
-
7/25/2019 Safety Tester Manual
23/149
GETTING STARTED
23
Workplace Precautions
Background The GPT-9000 is a high voltage instrument thatoutputs dangerous voltages. The followingsection describes precautions and proceduresthat must be followed to ensure a safe workenvironment.
WARNINGThe GPT-9000 generates voltages in excess of5kVAC or 6kVDC. Follow all safety precautions,
warnings and directions given in the followingsection when using the GPT-9000.
1. Only technically qualified personnel should beallowed to operate the safety tester.
2. The operating workplace must be fully isolated,especially when the instrument is in operation.The instrument should be clearly labeled withappropriate warning signage.
3. The operator should not wear any conductivematerials, jewelry, badges, or other items, suchwrist watches.
4. The operator should wear insulation gloves forhigh voltage protection.
5. Ensure the earth ground of the line voltage isproperly grounded.
6. Ensure any devices that are adversely affectedby magnetic fields are not placed near thetester.
-
7/25/2019 Safety Tester Manual
24/149
GPT-9000 Series User Manual
24
Operating Precautions
Background The GPT-9000 is a high voltage instrument thatoutputs dangerous voltages. The followingsection describes precautions and proceduresthat must be followed to ensure that the testeris operated in a safe manner.
WARNINGThe GPT-9000 generates voltages of up to 5kVACor 6kVDC. Follow all safety precautions, warningsand directions given in the following section when
using the GPT-9000.
1. Never touch the safety tester, lead wires,terminals, probes and other connectedequipment when the tester is testing.
2. Do not turn the safety tester on and off quicklyor repeatedly. When turning the power off,
please allow a few moments before turning thepower back on. This will allow the protectioncircuits to properly initialize.
Do not turn the power off when a test isrunning, unless in an emergency.
3. Only use those test leads supplied with theinstrument. Leads with inappropriate gaugescan be dangerous to both the operator and theinstrument.For GB testing, never use the Sense leads on theSOURCE terminals.
4. Do not short the HIGH VOLTAGE terminalwith ground. Doing so could charge the chassisto dangerously high voltages.
-
7/25/2019 Safety Tester Manual
25/149
GETTING STARTED
25
5. Ensure the earth ground of the line voltage isproperly grounded.
6. Only connect the test leads to the HIGHVOLTAGE/SOURCE H/SENSE H terminalsbefore the start of a test. Keep the test leadsdisconnected at all other times.
7. Always press the STOP button when pausingtesting.
8.
Do not leave the safety tester unattended.Always turn the power off when leaving thetesting area.
9. When remotely controlling the safety tester,ensure adequate safety measures are in place toprevent:
Inadvertent output of the test voltage.
Accidental contact with the instrument duringtesting. Ensure that the instrument and DUT arefully isolated when the instrument is remotelycontrolled.
10.Ensure an adequate discharge time for theDUT.
When DCW or IR tests are performed, the DUT,test leads and probes become highly charged.The GPT-9000 has discharge circuitry todischarge the DUT after each test. The timerequired for a DUT to discharge is typically200ms.
Never disconnect the safety tester before a
discharge is completed.
-
7/25/2019 Safety Tester Manual
26/149
GPT-9000 Series User Manual
26
Basic Safety Checks
Background The GPT-9000 is a high voltage device and assuch, daily safety checks should be made toensure safe operation.
1. Ensure all test leads are not broken and are freefrom defects such as cracks or splitting.
2. Ensure the safety tester is always connected to
an earth ground.
3. Test the safety tester operation with a lowvoltage/current output:
Ensure the safety tester generates a FAILjudgment when the HIGH VOLTAGE andRETURN terminals are shorted (using thelowest voltage/current as the testingparameters).
WARNINGDo not use high voltages/currents when the HIGHVOLTAGE and RETURN terminals are shorted. Itmay result in damage to the instrument.
-
7/25/2019 Safety Tester Manual
27/149
OPERATION
27
OPERATIONMenu Tree...................................................................... 29
Menu Tree Overview ...............................................................................30
Test Lead Connection ..................................................... 33
ACW, DCW, IR Connection ....................................................................33
GB Connection ..........................................................................................34
Manual Testing .............................................................. 35
Choose/Recall a Manual Test Number .................................................36Edit Manual Test Settings .......................................................................37Setting the Test Function .........................................................................38Setting the Test Voltage or Test Current ...............................................38
Setting the Test Frequency ......................................................................39Setting the Upper and Lower Limits .....................................................40
Setting a Reference Value ........................................................................42Setting the Test Time (Timer) .................................................................43Setting the Ramp Up Time ......................................................................45Creating a MANU Test File Name .........................................................46Setting the ARC Mode .............................................................................47Setting PASS HOLD .................................................................................48
Setting FAIL HOLD .................................................................................50
Setting MAX HOLD .................................................................................51
Setting the Grounding Mode ..................................................................52Saving and Exiting EDIT Status .............................................................56Running a MANU Test ............................................................................57PASS / FAIL MANU Test .......................................................................61Zeroing of the Test Leads (GB only) ......................................................66Special MANU Test Mode (000) .............................................................69
Common Utility Settings ................................................ 71
LCD Settings .............................................................................................71
-
7/25/2019 Safety Tester Manual
28/149
GPT-9000 Series User Manual
28
Buzzer Settings ......................................................................................... 72Interface Settings ...................................................................................... 73Control Settings ........................................................................................ 75
Automatic Tests .............................................................. 78
Choose/Recall an Automatic Test ......................................................... 78Edit Automatic Test Settings .................................................................. 80Adding a Step to the Automatic Test .................................................... 81Creating an AUTO Test File Name ........................................................ 82Saving and Exiting EDIT Status ............................................................. 83Automatic Test Page View ...................................................................... 84Running an Automatic Test .................................................................... 87Automatic Test Results ............................................................................ 91
-
7/25/2019 Safety Tester Manual
29/149
OPERATION
29
Menu TreeThis section describes the overall structure of the operation statuses
and modes for the GPT-9000 safety testers. The testers have twomain testing modes (MANU, AUTO) and 5 main operation statuses(VIEW, EDIT, READY, TEST and STOP).
statusSET T
PASS/FAIL result
statusOTS P
status
YAER D
status
EIV W
status
IDE T
Press
START
Press
STOP
PressMANU/
AUTO
Save the MANU
test or AUTO test
Press
STOP
Press
STOP
PressEDIT/
SAVE
Press
STOP
Hold
MANU/AUTO
Switch betweenAUTO andMANU mode
PAGE View
(AUTO mode
only)*
Press
PAGE
Press
ESC
PressEDIT/
SAVE
Common Utility
Settings*
Press
UTILITY
MANU UTILITY
menu (MANU
mode only)*
Press
UTILITY
* Press EDIT/SAVE to save settings, or ESC to cancel and return to the previous screen.
Press the STOP key twice for a FAIL result.
-
7/25/2019 Safety Tester Manual
30/149
GPT-9000 Series User Manual
3
Menu Tree Overview
VIEW status VIEW status is used to view the parameters ofthe selected manual test/automatic test. TheVIEW status is also used to put the tester intoMANU or AUTO mode.
I R G BA CW D CW
E
F RE Q = 0 H z6
0
100k V
RA C = F FO
EIV
MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
A U T O = 0 0 1 - 0 0 2
W
VIEW status
MM A UN N A_2M A N U = * * * - 0 0
MA U OT N A_
m A
E
EDIT status EDIT status is used to edit the manual test orautomatic test parameters. Pressing theEDIT/SAVE key will save any changes.Pressing the ESC key will cancel any changes.
I R G BA CW D CW I / LH O I M ET R
E
F R E Q = 0 H z6
0 100k V RA C = F FO
IDE MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
T
2M A N U = * * * - 0 0
EDIT status
m A
READY status When the tester is in READY status, it is readyto begin testing. Pressing the START buttonwill begin testing and put the tester into TESTstatus. Pressing the MANU/AUTO key willreturn the tester to VIEW status.
I R G BA CW D CW
E
F RE Q = 0 H z6
0 100k V RA C = F FO
AER MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
D
MM A UN N A_2M A N U = * * * - 0 0
READY status
Ym A
-
7/25/2019 Safety Tester Manual
31/149
OPERATION
31
TEST status TEST status is active when a MANU test orAUTO test is running. Pressing STOP willcancel the MANU test or the remaining steps in
an AUTO test.
I R G BA CW D CW
E
F R E Q = 0 H z6
0 100k V RA C = F FO
SET MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
MM A UN N A_2M A N U = * * * - 0 0
TEST status
Tm A00 33
STOP status STOP status is shown when a manual test or
automatic test did not finish running and hasbeen stopped by the operator. Pressing STOPwill return the tester to READY status.
I R G BA CW D CW
E
F R E Q = 0 H z6
0 100k V RA C = F FO
OTS MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
MM A UN N A_2M A N U = * * * - 0 0
STOP status
P
m A00 33
Page View Up to 16 tests can be used to create anautomatic test. Page View is used to see whichmanual tests (steps) an automatic test iscomposed of. The steps can be re-arranged anddeleted in Page View.
O V EM WA PS K I PS D E L
EMA U OT N A_0A U T O = 0 0 1 - 0 1M A N U _ N M EA VA CW = 0 . 0 0 k1
# 0 1 : 0 1 0 # 0 2 : 0 0 1 # 0 3 : 0 0 3 # 0 4 : 0 0 4# 0 5 : 0 0 7 # 0 6 : 0 0 3 # 0 7 : 0 3 8 # 0 8 : 0 0 5# 0 9 : # 1 0 : # 1 1 : # 1 2 :# 1 3 : # 1 4 : # 1 5 : # 1 6 :
*
1 . 0 0 m AH I ES = 0T_
-
7/25/2019 Safety Tester Manual
32/149
GPT-9000 Series User Manual
32
AUTO mode AUTO indicates that the tester is in AUTOmode. AUTO mode is for creating/running asequence of up to 16 MANU tests.
I R G BA CW D CW A D D
E
F RE Q = 0 H z6
0 100k V RA C = F FO
IDE MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MA U OT N A_
1 . 0 0 mAIH ES = 0T
0 0 . 1 S0A M P =R
T
2A U T O = 0 0 3 - 0 0
AUTO mode
m A
MANU mode MANU mode is used to create and/or execute asingle test. MANU indicates that the manual
test mode is active.
I R G BA CW D CW I / LH O I M ET R
E
F RE Q = 0 H z6
0 100k V RA C = F FO
IDE MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
T
2M A N U = * * * - 0 0
MANU mode
m A
Common Utility
Settings
This utility controls the LCD, buzzer, interface
and control settings. These settings are systemwide.
K e y L o k : Oc F F
S t a r t t r lC
D o u b l e A c t i no
L C D B U Z Z N T EI R C T R L
:
: F RO N T P AN
MOC MO UN T I L I T Y E L
FO F
I N T E R L C K :O O FF
MANU Utility
Settings
The Manu Utility settings are configured for
each MANU test separately. The settingsinclude: ARC MODE, PASS HOLD, FAILHOLD, MAX HOLD and GROUND MODE.
L I T YIM A UN U T2M A N U = * * * - 0 0
M A X H O L D : O F FF A I L H L D :O O FFP A S S H L D :O O FF
A R C M D E : OO FF
G RO U ND M OD E : NO
-
7/25/2019 Safety Tester Manual
33/149
OPERATION
33
Test Lead ConnectionThis section describes how to connect the GPT-9000 to a DUT for
withstanding, insulation resistance or ground bond testing.
ACW, DCW, IR Connection
Background ACW, DCW and IR tests use the HIGHVOLTAGE terminal and RETURN terminalwith the GHT-114 test leads.
ACW, DCW, IRConnection
High Voltage terminal
DUT
GPT-9000
Return terminal
Steps 1. Turn the power off on the safety tester.
2. Connect the high voltage test lead(red) to theHIGH VOLTAGE terminal and screw firmlyinto place.
3. Connect the return test lead(white) into theRETURN terminal and screw the protector barinto place, as shown below.
HIGH VOLTAGETerminal
RETURN
Terminal
-
7/25/2019 Safety Tester Manual
34/149
GPT-9000 Series User Manual
34
GB Connection
Background GB tests use the SENSE H/L and SOURCE H/Lterminals with the GTL-115 test leads.
GB Connection
Source H
DUT
GPT-9000
Sense L
Sense H
Source L
Steps 1. Turn the power off on the safety tester.
2. Connect the Sense H lead to the SENSE Hterminal.
3. Connect the Sense L lead to the SENSE L
terminal.
4. Connect the Source H lead to the SOURCE Hterminal.
5. Connect the Source L lead to the SOURCE Lterminal.
SENSE H Rx SENSE L
SOURCE LSOURCE H
RETURNSense H lead Sense L lead
Source L leadSource H lead
-
7/25/2019 Safety Tester Manual
35/149
OPERATION
35
Manual TestingThis section describes how to create, edit and run a single ACW,
DCW, IR or GB safety test. Each manual test can be stored/recalledto/from one of 100 memory locations. Each stored manual test canbe used as a test step when creating an AUTO test (page78).
Choose/Recall a Manual Test number from page36.
Edit Manual Test Settings from page37.
Setting the Test Function from page38.
Setting the Test Voltage or Test Current from page38. Setting the Test Frequency from page39.
Setting the Upper and Lower Limits from page40.
Setting a Reference Value from page42.
Setting the Test Time (Timer) from page43.
Setting the Ramp Up Time from page45.
Creating a MANU Test File Name from page46.
Setting the ARC Mode from page47.
Setting PASS HOLD from page49.
Setting FAIL HOLD from page50.
Setting MAX HOLD from page51.
Setting the Grounding Mode from page52.
Saving and Exiting EDIT Status from page56.
Running a MANU Test from page57.
PASS / FAIL MANU Test from page61.
Zeroing of the Test Leads (GB only) from page66
Special MANU Test Mode (000) from page69
Before operating the GPT-9000 please read the safety precautions asoutlined in the Set Up chapter on page20.
-
7/25/2019 Safety Tester Manual
36/149
GPT-9000 Series User Manual
36
Choose/Recall a Manual Test Number
Background ACW, DCW, IR and GB tests can only becreated in the MANU (manual) mode. MANUnumber 001 to 100 can be saved and thus beloaded when editing/creating a MANU test orAUTO test. MANU number 000 is a specialmode. See page66 for details on the specialmode.
Steps 1.
If the tester is in AUTO mode,press and hold the MANU/AUTOkey for three seconds to switch toMANU mode.
The tester can only switch betweenAUTO and MANU mode when inthe VIEW status.
MANU/AUTO
I R G BA CW D CW
E
F RE Q = 0 H z6
0 100k V RA C = F FO
EIV MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
A U T O = 0 0 1 - 0 0 2
W
VIEW status
MM A UN N A_2M A N U = * * * - 0 0
MA U OT N A_
m A
E
2. Use the scroll wheel to choose theMANU number.
MANU # 001~100
(MANU# 000 is a special mode)
I R G BA CW D CW
E
F R E Q = 0 H z6
0
100k V
RA C = F FO
EIV MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
W
2M A N U = * * * - 0 0
MANU number
m A
-
7/25/2019 Safety Tester Manual
37/149
OPERATION
37
NoteThe MANU number can only be chosen in VIEWstatus. If in the EDIT status, switch to the VIEWstatus by pressing the EDIT/SAVE or ESC key.
Edit Manual Test Settings
Background To edit any of the manual test settings, thetester must be in EDIT status.
Any settings or parameters that are edited only
apply to the currently selected MANU number.
Steps 1. Press the EDIT/SAVE key when inVIEW status to enter the EDITstatus. This will enter the EDITstatus for the chosen test number.
EDIT/SAVE
I R G BA CW D CW I / LH O I M ET R
E
F R E Q = 0 H z6
0 100k V RA C = F FO
IDE MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
T
2M A N U = * * * - 0 0
EIV W
m A
2. The Status changes from VIEW to EDIT.
NotePressing the EDIT/SAVE key again will save thesettings for the current test and return back to
VIEW status.
-
7/25/2019 Safety Tester Manual
38/149
GPT-9000 Series User Manual
38
Setting the Test Function
Background After a MANU number has been chosen andthe tester is in EDIT status, a test function canbe set.
There are four test functions, AC Withstand,DC Withstand, Insulation Resistance andGround Bond.
Steps 1.
To choose the test function, press the ACW,DCW, IR or GB soft-keys.
A CW D CW I R G B
2. The test function soft-key is highlighted.
I R G BA CW D CW I / LH O I M ET R
E
F R E Q = 0 H z6
0 100k V RA C = F FO
IDE MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
T
2M A N U = * * * - 0 0
test function
m A
NoteThe chosen test function only applies to thecurrent test.
Setting the Test Voltage or Test Current
Background The test voltage can be set from 0.100kV to 5kVfor ACW, 0.100kV to 6kV for DCW and 0.050 to1kV for IR (50V steps). For GB tests the testcurrent can be set from 3A to 30A.
-
7/25/2019 Safety Tester Manual
39/149
OPERATION
39
Steps 1. Press the UP / DOWN arrow keysto bring the cursor to the voltagesetting.
I R G BA CW D CW I / LH O I M ET R
E
F R E Q = 0 H z6
0 100k V RA C = F FO
IDE MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
T
2M A N U = * * * - 0 0
cursor
m A
2. Use the scroll wheel to set the
voltage level.
ACWDCWIRGB
0.100kV ~ 5kV0.100kV ~ 6kV0.05kV ~ 1kV (50V steps)3.00A ~ 30.00A
Note When setting the voltage, be aware that amaximum of 200VA can be set for ACW and 50Wfor DCW.
The ground bond voltage (GBV) is calculated asthe HI SET limit x Test Current.
Setting the Test Frequency
Background A test frequency of 60Hz or 50Hz can be set,regardless of the input line voltage. The testfrequency setting only applies to ACW and GBtests.
Steps 1. Press the UP / DOWN arrow keysto bring the cursor to the FREQ
setting.
-
7/25/2019 Safety Tester Manual
40/149
GPT-9000 Series User Manual
4
I R G B
m A
A CW D CW I / LH O I M ET R
E
F R E Q = 0 H z6
0 100k V RA C = F FO
IDE MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
T
2M A N U = * * * - 0 0
cursor
2. Use the scroll wheel to set the testfrequency.
ACW, GB 50Hz, 60Hz
NoteThe test frequency can only be set for ACW or GBtests.
Setting the Upper and Lower Limits
Background There is both a LO and HI judgment setting.When the measured value is below the LO SET
setting, the test will be judged as FAIL. Whenthe value exceeds the HI SET setting the testwill be judged as FAIL. Any measurementbetween the LO SET and HI SET setting is
judged as PASS. The LO SET limit cannot bemade greater than the HI SET limit.
Steps 1. Press the HI/LO soft-key or use
the UP / DOWN arrow keys tobring the cursor to the HI SET(ACW/DCW/GB) setting or theLO SET(IR) setting.
I / LH O
OR
-
7/25/2019 Safety Tester Manual
41/149
OPERATION
41
I R G BA CW D CW I / LH O I M ET R
E
F R E Q = 0 H z6
0 100k V RA C = F FO
IDE MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
T
2M A N U = * * * - 0 0
cursor
m A
2. Use the scroll wheel to set the HISET/LO SET limit.
ACW (HI)DCW (HI)
IR (LO)GB (HI)
0.001mA~040.0mA0.001mA~010.0mA
0001M~ 9500M000.1m ~ 650.0m
3. Press the HI/LO soft-key again orpress the DOWN arrow key toswitch between HI SET and LOSET.
I / LH O
OR
I R G BA CW D CW I / LH O I M ET R
E
F R E Q = 0 H z6
0 100k V RA C = F FO
IDE MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m A0L ES = 0T
0 0 . 1 S0A M P =R
T
2M A N U = * * * - 0 0
cursor
1 . 0 0 m AIH ES = 0T
m A
4. Use the scroll wheel to set the HI
SET/LO SET limit.
ACW (LO)DCW (LO)IR (HI)GB (LO)
0.000mA~039.9mA0.000mA~009.9mA0001M~9500M, 000.0m ~ 649.9m
Note
The LO SET setting is limited by the HI SET
setting. The LO SET limit cannot be greater thanthe HI SET limit.
-
7/25/2019 Safety Tester Manual
42/149
GPT-9000 Series User Manual
42
Setting a Reference Value
Background The REF# acts as an offset. The REF# value issubtracted from the measured current (ACW,DCW) or measured resistance (IR, GB).
Steps 1. Press the UP / DOWN arrow keysto bring the cursor to theREF# setting.
I R G B
m A
A CW D CW I / LH O I M ET R
E
F R E Q = 0 H z6
0 100k V RA C = F FO
IDE MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m A0L ES = 0T
0 0 . 1 S0A M P =R
T
2M A N U = * * * - 0 0
cursor
2. Use the scroll wheel to set theREF# value.
ACWDCWIRGB
0.000mA~HI SET current-0.1mA0.000mA~HI SET current-0.1mA0000M~HI SET-1M000.0m~HI SET-0.1m
NoteFor GB tests, a reference offset can beautomatically created using the zeroing function.
See page66 for details.
-
7/25/2019 Safety Tester Manual
43/149
OPERATION
43
Setting the Test Time (Timer)
Background The TIMER setting is used to set the test timefor the current test. The test time determineshow long the test voltage or current is appliedto the DUT. This test time does not includeRamp , initial start time or discharge time(note: GB does not have Ramp or dischargetimes). The test time can be set from 0.5 secondsto 999.9 seconds for ACW, DCW and GB and
1.0 second to 999.9 seconds for IR, with aresolution of 0.1 seconds for all modes. Thetimer can be turned off when in the specialMANU test mode when using the ACW orDCW test functions.
Each test has an initial test time of 100ms and adischarge time (except GB) of 200ms.
Discharge time(200ms)
Test V
timeStart V
Initial time(100ms)
TEST TIMERAMP
Steps 1.
Press the TIMER soft-key or usethe UP/DOWN arrow keys tobring the cursor to the TIMERsetting.
I M ET R
OR
-
7/25/2019 Safety Tester Manual
44/149
GPT-9000 Series User Manual
44
I R G B
m A
A CW D CW I / LH O I M ET R
E
F R E Q = 0 H z6
0 100k V RA C = F FO
IDE MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m A0L ES = 0T
0 0 . 1 S0A M P =R
T
2M A N U = * * * - 0 0
cursor
2. Use the scroll wheel to set theTIMER value.
ACWDCW
IRGB
000.5s~999.9s000.5s~999.9s
001.0s~999.9s000.5s~999.9s
NoteWith the ACW test function, when the test currentis between 30mA and 40mA, the ramp time + testtime cannot exceed 240 seconds. At this currentlevel, the tester also needs to pause after a test fora time equal to or greater than the output time. Seethe specifications on page142 for details.
Special ManualMode
When in special MANU test mode (page66)theTimer can be turned off when using the DCWor ACW test function.
Hold the TIMER soft-key for 3seconds to turn the timer off.
I M ET R
NoteThe timer can only be turned off under specialMANU test mode, however there is a limitation:The timer cannot be turned off (limited to 240s) ifthe test current is between 30mA and 40mA inACW mode.
The discharge time and initial test time cannot beedited.
-
7/25/2019 Safety Tester Manual
45/149
OPERATION
45
Setting the Ramp Up Time
Background The Ramp Up time is the total time taken forthe tester to reach the test voltage level. TheRamp Up time starts after the initial time(100ms) with a start voltage of 50 volts. TheRamp Up time can be set from 000.1 to 999.9seconds. The Ramp Up time is only applicablefor ACW, DCW and IR tests.
Discharge time(200ms)
Test V
timeStart V
Initial time(100ms)
TEST TIMERAMP
Steps 1. Use the UP/DOWN arrow keys tobring the cursor to the RAMPsetting.
I R G B
m A
A CW D CW I / LH O I M ET R
E
F R E Q = 0 H z6
0 100k V RA C = F FO
IDE MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m A0L ES = 0T
0 0 . 1 S0A M P =R
T
2M A N U = * * * - 0 0
cursor
2. Use the scroll wheel to set theRAMP value.
ACWDCWIR
000.1s~999.9s000.1s~999.9s000.1s~999.9s
Note The discharge time and initial test time cannot beedited.
-
7/25/2019 Safety Tester Manual
46/149
GPT-9000 Series User Manual
46
Creating a MANU Test File Name
Background Each manual test can have a user-defined testfile name (default: MANU_NAME) up to 10characters long. See the character list below forthe allowed characters.
Character List
0 1 2 3 4 5 6 7 8 9
WX Y ZM VO P RQ T USNLA B C D E F G H I J K
w x y zm vo p rq t usnla b c d e f g h i j k
+ - * / _ = : ? ( ) < > [ ]
Steps 1. Use the UP/DOWN arrow keys tobring the cursor to the MANU testfile name at the top of the screen.The test file name is initially set asMANU_NAME.
I R G B
m A
A CW D CW I / LH O I M ET R
E
F R E Q = 0 H z6
0 100k V RA C = F FO
IDE MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m A0L ES = 0T
0 0 . 1 S0A M P =R
T
2M A N U = * * * - 0 0
cursor
2. Use the scroll wheel to scrollthrough the available characters.
3. Press the Left/Right arrow keys togo the next character.
4. The MANU test file name is set when thecurrent test setting is saved or when the cursoris moved to another setting.
-
7/25/2019 Safety Tester Manual
47/149
OPERATION
47
Setting the ARC Mode
Background ARC detection, otherwise known as flashoverdetection, detects fast voltage or currenttransients that are not normally detected. Arcingis usually an indicator of poor withstandinginsulation, electrode gaps or other insulatingproblems that cause temporary spikes in currentor voltage during ACW and DCW testing.
There are three ARC detection settings: OFF,ON AND CONTINUE, ON AND STOP.The ON AND CONTINUE setting will detectarcs over the ARC current level and continuethe test, the ON AND STOP setting will stop thetest when an arc is detected.
ARC mode settings only apply to ACW andDCW tests.
Steps 1. Press the UTILITY key on the frontpanel when the tester is in EDITstatus. The tester will go to theMANU Utility for the current test.
UTILITY
L I T YIM A UN U T2M A N U = * * * - 0 0
M A X H O L D : O F FF A I L H L D :O O FF
A R C M D E : OO FF
G R O U N D MO D E : NO
P A S S H L D :O O FF
2. Use the UP/DOWN arrow keys tomove to the ARC MODE setting.
3. Use the scroll wheel to set the ARCmode.
-
7/25/2019 Safety Tester Manual
48/149
GPT-9000 Series User Manual
48
ARC MODES: OFF, ON AND CONTINUE,ON AND STOP
4. Press the EDIT/SAVE key to save
and exit the MANU Utility and goback to EDIT status.
EDIT/SAVE
NoteThe ESC key can be pressed at any time in theUtility menu to cancel and exit.
5. If the ARC MODE was set to either ON ANDCONTINUE, or ON AND STOP, the ARC
current level can be edited.
6. Use the UP/DOWN arrow keys tomove the cursor to the ARC setting.
7. Use the scroll wheel to edit theARC level.
ACWDCW
0.000mA~080.0mA0.000mA~020.0mA
NoteThe ARC setting range is directly related to the HISET current limit.
ACW
HI SET Limit ARC range
0.001mA~0.999mA 0.000mA~2.000mA01.00mA~09.99mA 00.00mA~20.00mA010.0mA~042.0mA 000.0mA~080.0mA
DCW
HI SET Limit ARC range
0.001mA~0.999mA 0.000mA~2.000mA01.00mA~09.99mA 00.00mA~20.00mA010.0mA~011.0mA 000.0mA~020.0mA
-
7/25/2019 Safety Tester Manual
49/149
OPERATION
49
Setting PASS HOLD
Background The PASS HOLD settings only apply to theselected test in an AUTO test. When the PASSHOLD setting is set to ON, a PASS judgment isheld until the START button is pressed.
NoteThe PASS HOLD setting only applies to AUTOtests. This setting is ignored when runningMANU tests.
Steps 1. Press the UTILITY key on the frontpanel when the tester is in EDITstatus. The display will go from thenormal EDIT status to the MANUUtility menu for the current test.
UTILITY
L I T YIM A UN U T2M A N U = * * * - 0 0
M A X H O L D : O F F
A R C M D E : OO FF
G R O U N D MO D E : NO
P A S S H L D :O O FFF A I L H L D :O O F F
2. Use the UP/DOWN arrow keys tomove to the PASS HOLD setting.
3. Use the scroll wheel to set PASS
HOLD.
PASS HOLD OFF, ON
4. Press the EDIT/SAVE key to saveand exit the MANU Utility menu.
EDIT/SAVE
Note
The ESC key can be pressed at any time in the
MANU Utility menu to cancel and exit.
-
7/25/2019 Safety Tester Manual
50/149
GPT-9000 Series User Manual
5
Setting FAIL HOLD
Background The FAIL HOLD settings only apply to theselected test in AUTO tests. When the FAILHOLD setting is set to ON, a FAIL judgment isheld until the START button is pressed.
NoteThe FAIL HOLD setting only applies to AUTOtests. This setting is ignored when runningMANU tests.
Steps 1. Press the UTILITY key on the frontpanel when the tester is in EDITstatus. The display will go from thenormal EDIT status to the MANUUtility menu for the current test.
UTILITY
L I T YIM A UN U T2M A N U = * * * - 0 0
A R C M D E : OO FF
G R O U N D MO D E : NO
P A S S H L D :O O FFF A I L H L D :O O F F
M A X H O L D : O F F
2. Use the UP/DOWN arrow keys tomove to the FAIL HOLD setting.
3. Use the scroll wheel to set FAIL
HOLD.
FAIL HOLD OFF, ON
4. Press the EDIT/SAVE key to saveand exit the MANU Utility menu.
EDIT/SAVE
Note
The ESC key can be pressed at any time in the
MANU Utility menu to cancel and exit.
-
7/25/2019 Safety Tester Manual
51/149
OPERATION
51
Setting MAX HOLD
Background The MAX HOLD setting will hold themaximum current measured in the ACW andDCW tests or the maximum resistancemeasured in IR and GB tests.
Steps 1. Press the UTILITY key on the frontpanel when the tester is in EDITstatus. The display will go from the
normal EDIT status to the MANUUtility menu for the current test.
UTILITY
L I T YIM A UN U T2M A N U = * * * - 0 0
A R C M D E : OO FFP A S S H L D :O O FFF A I L H L D :O O F FM A X H O L D : O F FG R O U N D MO D E : NO
2. Use the UP/DOWN arrow keys to
move to the MAX HOLD setting.
3. Use the scroll wheel to set MAXHOLD.
MAX HOLD OFF, ON
4. Press the EDIT/SAVE key to saveand exit the MANU Utility menu.
EDIT/SAVE
NoteThe ESC key can be pressed at any time in theMANU Utility menu to cancel and exit.
-
7/25/2019 Safety Tester Manual
52/149
GPT-9000 Series User Manual
52
Setting the Grounding Mode
Background When GROUND MODE is set to ON, the GPT-9000 grounds the return terminal to the ground.This mode is best for DUTs that are groundedto an earth ground by their chassis, fixtures oroperation environment. This mode measuresthe potential of the HIGH VOLTAGE terminalwith respect to earth ground. This means thatany stray capacitance/resistance that leaks to
earth ground will also be measured. This is thesafest testing mode, though potentially not asaccurate.
When GROUND MODE is set to OFF, thereturn terminal is floating with respect to theearth ground. This mode is for DUTs that arefloating and not directly connected to an earthground. This is more accurate than when
GROUND MODE is set to ON as any straycapacitance/resistance that leaks to the earthground from the DUT side of the testing circuitwill not be measured. For this reason, thistesting mode is able to measure to a higheraccuracy and resolution.
The GROUND MODE is always set to OFF for
IR and GB tests.
GROUND MODE = ON, DUT grounded
High Voltage terminal
Return terminalDUT
GPT-9000
strayresistance,capacitance
-
7/25/2019 Safety Tester Manual
53/149
OPERATION
53
GROUND MODE = ON, DUT floating
High Voltage terminal
DUT
GPT-9000
strayresistance,capacitanceReturn terminal
GROUND MODE = OFF, DUT floating
High Voltage terminal
DUT
GPT-9000
strayresistance,capacitanceReturn terminal
GPT-9804 (GB testing)
DUT
Source H
Source L
Sense H
Sense L
stray
resistance,capacitance
-
7/25/2019 Safety Tester Manual
54/149
GPT-9000 Series User Manual
54
GROUND MODE = OFF, DUT grounded
High Voltage terminal
DUT
GPT-9000
strayresistance,capacitanceReturn terminal
GPT-9804 (GB testing)
DUT
Source H
Source L
Sense H
Sense L
strayresistance,
capacitance
WarningWhen GROUND MODE is set to OFF, the DUT,fixtures or connected instrumentation cannot begrounded. This will short circuit the internalcircuitry during a test.
For ACW and DCW tests, if it is not known whetherthe DUT test setup is grounded or not, always setGROUND MODE to ON.
Only set GROUND MODE to OFF when the DUTis floating electrically.
Steps1.
Press the UTILITY key on the frontpanel when the tester is in EDITstatus. The display will go from thenormal EDIT status to the MANUUtility menu for the current test.
UTILITY
L I T YIM A UN U T2M A N U = * * * - 0 0
A R C M D E : OO FFP A S S H L D :O O FFF A I L H L D :O O F FM A X H O L D : O F FG RO U ND M OD E : NO
-
7/25/2019 Safety Tester Manual
55/149
OPERATION
55
2. Use the UP/DOWN arrow keys tomove the cursor to the GROUNDMODE setting.
3. Use the scroll wheel to set theGROUND MODE.
GROUND MODE OFF, ON
4. Press the EDIT/SAVE key to saveand exit the MANU Utility menu.
EDIT/SAVE
5. The GROUND MODE icon on the displaychanges accordingly.
G B I / LH
E
RA C =
DE MT I E R
ER F =#MN A_
1 . 0 0 m A= 0
0 0 . 1 S0=
GROUNDMODE = OFF
m A
G B I / LH
E
RA C =
DE MT I E R
ER F =#MN A_
1 . 0 0 m A= 0
0 0 . 1 S0=
GROUNDMODE = ON
m A
NoteThe ESC key can be pressed at any time in theMANU Utility menu to cancel and exit.
IR and GB tests can only have GROUND MODEset to OFF.
-
7/25/2019 Safety Tester Manual
56/149
GPT-9000 Series User Manual
56
Saving and Exiting EDIT Status
Background After all test parameters have been set, the testcan be saved. After a test is saved it can be usedwhen creating an AUTO test.
WarningThe special MANU number, 000, can be saved,however it cannot be used for AUTO tests. Seepage66 for details.
Steps 1.
When in EDIT status, press theEDIT/SAVE key to save thecurrent test. This will enter theVIEW status for the chosen testnumber.
EDIT/SAVE
I R G BA CW D CW
E
F R E Q = 0 H z6
0 100k V RA C = F FO
EIV MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
W
2M A N U = * * * - 0 0
IDE T
m A
2. The Status changes from EDIT to VIEW.
NotePressing the EDIT/SAVE key again will return thetester back to EDIT status for the current test.
-
7/25/2019 Safety Tester Manual
57/149
OPERATION
57
Running a MANU Test
Background A test can be run when the tester is in READYstatus.
NoteThe tester cannot start to run a test under thefollowing conditions:
A protection setting has been tripped; when aprotection setting has been tripped the
corresponding error message is displayed onthe screen. See page141 for a comprehensivelist of the all the setting errors.
The INTERLOCK function is ON and theInterlock key is not inserted in the signal I/Oport (page75).
The STOP signal has been received remotely.
If Double Action is ON, ensure the STARTbutton is pressed immediately after the STOPbutton (
-
7/25/2019 Safety Tester Manual
58/149
GPT-9000 Series User Manual
58
2. Press the STOP button to put thetester into the READY status.
STOP
I R G B
m A
A CW D CW
E
F RE Q = 0 H z6
0 100k V RA C = F FO
AER MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
D
MM A UN N A_2M A N U = * * * - 0 0
READY status
Y
3. The READY indicator will be litblue when in the READY status.
READY
4. Press the START button when thetester is in the READY status. Themanual test starts automaticallyand the tester goes into the TEST
status.
START
5. The TEST indicator will be litorange when in the TEST status.
TEST
I R G BA CW D CW
E
F RE Q = 0 H z6
0
100k V
RA C = F FO
SET MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
MM A UN N A_2M A N U = * * * - 0 0
TEST status
Tm A00 00
6. The test will start by showing the remainingramp up time, followed by the remaining testtime. The test will continue unit the test isfinished or the test is stopped.
-
7/25/2019 Safety Tester Manual
59/149
OPERATION
59
I R G BA CW D CW
E
F RE Q = 0 H z6
0 100k V RA C = F FO
SET MT I E = 0 0 3 . 2 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
MM A UN N A_2M A N U = * * * - 0 0
T
remaining TIMER time
00 37m A
remaining RAMP time
ACW Example
I R G BA CW D CW
E
F RE Q = 0 H z6
0
100k V
RA C = F FO
SET MT I E = 0 0 3 . 2 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
MM A UN N A_2M A N U = * * * - 0 0
T
Test Voltage
Measured Current
00 37m A
DCW Example
I R G BA CW D CW
E
0 100k V RA C = F FO
SET MT I E = 0 0 3 . 2 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
MM A UN N A_2M A N U = * * * - 0 0
T
Test Voltage
Measured Current
00 56m A
IR Example
I R G BA CW D CW
E
0 050k V
SET MT I E = 0 0 3 . 2 SR
0 0 MER F =# 0 0
1 1 0 0 MIH ES =T
0 0 . 1 S0A M P =R
MM A UN N A_2M A N U = * * * - 0 0
T
Test Voltage
Measured Resistance
1056 M
GB Example
I R G BA CW D CW
E
F RE Q = 0 H z6
0 3 01A
SET G B V = 0 . 6 2 2 V MT I E = 0 0 3 . 2 SR
. 0 mER F =# 0 00
0 0 . 0 m IH ES = 3T
MM A UN N A_2M A N U = * * * - 0 0
T
Test Current
Measured Resistance
200 7m
-
7/25/2019 Safety Tester Manual
60/149
GPT-9000 Series User Manual
6
Stop the Test 1. To stop the test at any time when itis running, press the STOP button.The test will stop immediately.
When the STOP button is pressed,a judgment is not made on the test.
All panel keys except the STOPbutton are locked when the testeris in STOP status.
STOP
I R G BA CW D CW
E
F RE Q = 0 H z6
0 100k V RA C = F FO
OTS MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
MM A UN N A_2M A N U = * * * - 0 0
STOP status
Pm A00 00
2. To put the tester back into READYstatus, press the STOP buttonagain.
STOP
Exit TEST Status To exit testing, press theMANU/AUTO key when thetester is in the READY status. Thetester will revert to the VIEWstatus for the current test.
MANU/AUTO
I R G B
m A
A CW D CW
EF RE Q = 0 H z6
0 100k V RA C = F FO
EIV MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .01 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
W
MM A UN N A_2M A N U = * * * - 0 0
NoteDo not touch any terminals, test leads or any otherconnections when the test is on.
-
7/25/2019 Safety Tester Manual
61/149
OPERATION
61
PASS / FAIL MANU Test
Background If the test is allowed to run to completion (thetest is not stopped or a protection setting is nottripped) then the tester will judge the test aseither PASS or FAIL.
NoteThe test will be judged PASS when:
The HI SET and LO SET limits have not beentripped during the test time.
The test will be judged FAIL when:
Either the HI SET or LO SET limit has beentripped during the test time.
A protection setting has been tripped duringthe test time. See page141 for a list of errormessages.
PASS Judgment 1.
When the test is judged as PASS,PASS will be displayed, the buzzerwill sound and the PASS indicatorwill be lit green.
PASS
I R G BA CW D CW
E
F RE Q = 0 H z6
0 100k V RA C = F FO
SAP MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
MM A UN N A_2M A N U = * * * - 0 0
Sm A00 37
2. The PASS judgment will be held on the displayuntil the STOP or START button is pressed.
Pressing the STOP button willreturn the tester to the READYstatus.
STOP
-
7/25/2019 Safety Tester Manual
62/149
GPT-9000 Series User Manual
62
Pressing the START button willrestart the test.
START
NoteThe buzzer will only sound if the Pass Sound is setto ON. See page72 for details.
The START button is disabled when the buzzer isbeeping.
PASS TimingDiagrams
The timing diagrams below show the ACW, DCW,IR and GB timing for the START status, TESTstatus and PASS judgment.
ACW PASSTiming
Discharge time(200ms)
Output V
time
Initial time(100ms)
START
TEST
PASS
RAMP TEST TIME
DCW PASSTiming
Discharge time(200ms)
Output V
time
Initial time(100ms)
START
TESTPASS
RAMP TEST TIME
-
7/25/2019 Safety Tester Manual
63/149
OPERATION
63
IR PASS Timing
Discharge time(200ms)
Output V
time
Initial time(100ms)
START
TEST
PASS
RAMP TEST TIME
GB PASS Timing
Output I
time
START
TEST
PASS
0.5s
Initial time(100ms)
Total Test Time
FAIL Judgment 1. When the test is judged as FAIL,FAIL will be displayed, the buzzerwill sound and the FAIL indicatorwill be lit red.
As soon as a test is judged FAIL,power is cut from the terminals.
FAIL
I R G BA CW D CW
F RE Q = 0 H z6
0 100k V RA C = F FO
IAF MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
2M A UN 0 0_2M A N U = * * * - 0 0
Lm A01 37
2. The FAIL judgment will be held onthe display until the STOP buttonis pressed. Pressing the STOPbutton will return the tester to theREADY status.
STOP
-
7/25/2019 Safety Tester Manual
64/149
GPT-9000 Series User Manual
64
3. The READY indicator will be litblue in the READY status.
READY
I R G B
m A
A CW D CW
E
F RE Q = 0 H z6
0 100k V RA C = F FO
AER MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
D
MM A UN N A_2M A N U = * * * - 0 0
READY status
Y
NoteThe buzzer will only sound if Fail Sound is set toON. See page72 for details.
FAIL TimingDiagrams
The timing diagrams below show the ACW, DCW,IR and GB timing for the START status, TESTstatus and FAIL judgment.
ACW FAIL Timing
START
TEST
FAIL
Discharge time(200ms)
Output V
time
Initial time(100ms)
RAMP TEST TIME
DCW FAIL Timing
FAIL
Discharge time(200ms)
Output V
time
Initial time(100ms)
RAMP
START
TEST
TEST TIME
-
7/25/2019 Safety Tester Manual
65/149
OPERATION
65
IR FAIL Timing
START
TEST
FAIL
Discharge time(200ms)
Output V
time
Initial time(100ms)
RAMP TEST TIME
GB FAIL Timing
Output I
time
START
TEST
FAIL
0.5s
Initial time(100ms)
Total Test Time
-
7/25/2019 Safety Tester Manual
66/149
GPT-9000 Series User Manual
66
Zeroing of the Test Leads (GB only)
Background The Zeroing function is used to determine theresistance of the test leads for GB tests. When azero check is performed, the reference isautomatically set to the measured resistance ofthe test leads.
This function is only available for GB testing.
Steps 1.
Ensure the tester is in VIEW statusfor the current GB test. Save thecurrent test if necessary.
Page56
I R G BA CW D CW
E
F RE Q = 0 H z6
EIV G B V = V MT I E = 0 0 3 . 2 SR
. 0 mER F =# 0 00
0 0 . 0 m IH ES = 3T
MM A UN N A_2M A N U = * * * - 0 0
W m0 3 01A
2.
Short the positive and negative alligator clips asshown below.
3. Press the STOP button to put thetester into the READY status.
STOP
-
7/25/2019 Safety Tester Manual
67/149
OPERATION
67
4. The ZERO function can beactivated by pressing thecorresponding soft-key in the
READY status. The ZERO soft-keywill be highlighted.
E ROZ
5. Press the START button to performthe zero check. The tester will gointo the ZERO status.
START
0 3 01AZ E R 0I R G BA CW D CW
E
F RE Q = 0 H z6
REZ G B V = 0 . 6 2 2 V MT I E = 0 0 3 . 2 SR
. 0 mER F =# 0 00
0 0 . 0 m IH ES = 3T
MM A UN N A_2M A N U = * * * - 0 0
O
Measured Resistance
200 7m
ZERO status
6. When the zero check has finished, the tester
will return back to the VIEW status. Theresistance of the test leads will be automaticallyset as the Reference value.
0
3 01A
I R G BA CW D CW
E
F RE Q = 0 H z6
EIV G B V = V MT I E = 0 0 3 . 2 SR
. 7 mER F =# 0 02
0 0 . 0 m IH ES = 3T
MM A UN N A_2M A N U = * * * - 0 0
Wm
Reference value
NoteRemember to replace the test leads to the properposition on the DUT before testing.
-
7/25/2019 Safety Tester Manual
68/149
GPT-9000 Series User Manual
68
I
-
7/25/2019 Safety Tester Manual
69/149
OPERATION
69
Special MANU Test Mode (000)
Background When MANU number 000 is selected, thespecial test mode is activated. Under the specialtest mode, the voltage can be changed during atest, in real time (ACW, DCW only). The testfunction can also be changed when in READYor VIEW status, unlike under normal operation.
Separate settings can be saved under the special
test mode for each of the testing functions:ACW, DCW, IR and GB. This means a differentACW, DCW, IR and GB test setup can be savedfor MANU number 000.
Steps 1. Choose MANU number 000 toenter the special test mode.
Page36
2. Set all the necessary parameters fora test and save. A different testsetup can be saved for each testfunction (ACW, DCW, IR and GB).
Pages37~56
I R G BA CW D CW
E
F R E Q = 0 H z6
0 100k V RA C = F FO
EIV MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
W
MM A UN N A_0M A N U = * * * - 0 0
m A
NoteThe TIMER settings can be set OFF when in thespecial test mode for ACW and DCW tests.
3. The settings of a different testfunction can be loaded by pressingthe corresponding soft-key in theVIEW or READY status.
A CW
Example:ACW
-
7/25/2019 Safety Tester Manual
70/149
GPT-9000 Series User Manual
7
4. Run a manual test for the specialtest mode (000). The test is startedand stopped in the same way as for
the normal manual test mode.
Page57
5. Use the scroll wheel to set thevoltage level in real-time as the testis running (this does not apply toIR or GB tests).
ACWDCW
0.100kV ~ 5kV0.100kV ~ 6kV
-
7/25/2019 Safety Tester Manual
71/149
OPERATION
71
Common Utility SettingsThe Common Utility settings are system-wide settings that apply to
both MANU tests and AUTO tests.
The Common Utility menu includes the following settings:
LCD settings from page71.
Buzzer Settings from page72.
Interface Settings from page73.
Control settings from page75.
LCD Settings
Description The LCD settings include contrast andbrightness controls.
Steps 1. Ensure the tester is in VIEW status.
Save the current test if necessary.
Page56
I R G B
E
A CW D CW
F RE Q = 0 H z6
0 100k V RA C = F FO
EIV MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
W
MM A UN N A_0M A N U = * * * - 0 0
m A
VIEW status
2. Press the UTILITY key. UTILITY
3. Press the LCD soft-key to bring upthe LCD Common Utility menu.
L C D
L C D C o t r an
L C D B r g h ti n se
L C D B U Z Z N T EI R C T R L
s : B R I G H T
s t : 5
MOC MO UN T I L I T Y
-
7/25/2019 Safety Tester Manual
72/149
GPT-9000 Series User Manual
72
4. Use the UP/DOWN arrow keys tochoose a menu item: LCDContrast, LCD Brightness.
5. Use the scroll wheel to select aparameter for the chosen menuitem.
LCD ContrastLCD Brightness
1(low) ~ 8(high)BRIGHT, DARK
6. Press EDIT/SAVE to save thesettings and exit to VIEW status.
EDIT/SAVE
NoteThe ESC key can be pressed at any time to canceland exit back to VIEW status.
Buzzer Settings
Description The Buzzer settings allow you to set whetherthe buzzer will sound for PASS/FAIL
judgments. The buzzer time can also be set forthe PASS/FAIL judgments. The buzzer settingsare system-wide.
Steps 1. Ensure the tester is in VIEW status.Save the current test if necessary.
Page56
I R G B
E
A CW D CW
F RE Q = 0 H z6
0 100k V RA C = F FO
EIV MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
W
MM A UN N A_0M A N U = * * * - 0 0
m A
VIEW status
2. Press the UTILITY key. UTILITY
-
7/25/2019 Safety Tester Manual
73/149
OPERATION
73
3. Press the BUZZ soft-key to bringup the Buzzer Common Utilitymenu.
B U Z Z
L C D B U Z Z N T EI R C T R L
MOC MO UN T I L I T Y P a s s S u n do : NO
F a i l S u n do : FO F
T I M E : 0 0 . 50 S
4. Use the UP/DOWN arrow keys tochoose a menu item: Pass Sound orFail Sound.
5. Use the scroll wheel to select aparameter for the chosen menuitem.
Pass SoundFail Sound
ON (000.2s~999.9s), OFFON (000.2s~999.9s), OFF
6. Press EDIT/SAVE to save thesettings and exit to the VIEWstatus.
EDIT/SAVE
NoteWhen in automatic tests, the Pass Sound and FailSound settings only apply to the overall PASS/FAILof the overallautomatic test, not each test stepthat make up the automatic tests.
Note The ESC key can be pressed at any time to canceland exit back to VIEW status.
Interface Settings
Description The interface settings choose the remoteinterface configuration. USB, RS232 and GPIB
(optional) can be selected.
-
7/25/2019 Safety Tester Manual
74/149
GPT-9000 Series User Manual
74
Steps 1. Ensure the tester is in VIEW status.Save the current test if necessary.
Page56
I R G B
E
A CW D CW
F RE Q = 0 H z6
0 100k V RA C = F FO
EIV MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
W
MM A UN N A_0M A N U = * * * - 0 0
m A
VIEW status
2. Press the UTILITY key. UTILITY
3. Press the INTER soft-key to bringup the Interface Common Utilitymenu.
N T EI R
I n t
L C D B U Z Z N T EI R C T R L
MOC MO UN T I L I T Y
B u d :a 1 1 5 2 0 0
R S 2 3 2f a c :ee r
4. Use the scroll wheel to select USB,RS232 or GPIB.
5. For RS232 or GPIB, use theUP/DOWN arrow keys to chooseBaud or Address.
6. Use the scroll wheel to select thebaud rate or GPIB address.
Baud
GPIB address
9600, 19200, 38400, 57600,1152000~30
-
7/25/2019 Safety Tester Manual
75/149
OPERATION
75
7. Press EDIT/SAVE to save thesettings and exit to VIEW status.
EDIT/SAVE
Note Ensure the baud rate settings or GPIB addressmatches the host machine.
NoteThe ESC key can be pressed at any time to canceland exit back to VIEW status.
Control Settings
Description The Control settings are accessed in theCOMMON UTILITY menu. The Controlsettings include: Start Control, Double Action,Key Lock and Interlock.
Start Control is used to determine how a test isstarted. Tests can be started via the front panel(START/STOP buttons), from a remotecontroller or via the SIGNAL I/O port.
The Double Action function is a safety featureused to prevent accidentally starting a test.Normally to start a test, the START button ispressed when the tester is in the READY status.To start a test when Double Action is ON, theSTOP button must first be pressed, followed by
the START button within 500ms.
Key Lock disables the front panel keys fromchanging the test number, mode or testingparameters. Only the Utility menu and anykeys required for testing are not disabled.
-
7/25/2019 Safety Tester Manual
76/149
GPT-9000 Series User Manual
76
The Interlock function is a safety feature. Theinterlock function prevents a test from running,unless the interlock pins on the signal I/O port
connector are shorted. The included interlockkey can be used for this purpose. See page100for details.
Steps 1. Ensure the tester is in VIEW status.Save the current test if necessary.
Page56
I R G B
E
A CW D CW
F RE Q = 0 H z6
0 100k V RA C = F FO
EIV MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
W
MM A UN N A_0M A N U = * * * - 0 0
m A
VIEW status
2. Press the UTILITY key. UTILITY
3. Press the CTRL soft-key to bringup the Control Common Utilitymenu.
T R LC
K e y L o k : Oc F F
S t a r t t r lC
D o u b l e A c t i no
L C D B U Z Z N T EI R C T R L
:
: F RO N T P AN
MOC MO UN T I L I T Y E L
FO F
I N T E R L C K :O O FF
4. Use the UP/DOWN arrow keys tochoose a menu item: Start Ctrl,Double Action, Key Lock orINTERLOCK.
5. Use the scroll wheel to selectsetting for the chosen menu item.
-
7/25/2019 Safety Tester Manual
77/149
OPERATION
77
Start Ctrl
Double ActionKey Lock
INTERLOCK
FRONT PANEL, REMOTECONNECT, SIGNAL IOON, OFFON, OFF
ON, OFF
6. Press EDIT/SAVE to save thesettings.
EDIT/SAVE
NoteThe Double Action setting is ignored when theGPT-9000 is being controlled remotely using theUSB, RS232 or GPIB interfaces.
NoteIf a test is started with INTERLOCK ON, but theinterlock signal I/O pins are not shorted (eitherwith the included interlock key or manually), theINTERLOCK OPEN message will be displayed,preventing the test from starting.
I R G B
m A
A CW D CW
E
F RE Q = 0 H z6
0 100k V RA C = F FO
AER MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
D
MM A UN N A_2M A N U = * * * - 0 0
YI N T E R L OC K O P EN
Interlock open message
-
7/25/2019 Safety Tester Manual
78/149
GPT-9000 Series User Manual
78
Automatic TestsThis section describes how to create, edit and run automatic tests.
Automatic tests allow you to link together up to 16 differentMANU tests and run them sequentially. Each stored MANU test isused as a test step when creating an AUTO test.
Choose/Recall an Automatic Test from page78
Edit Automatic Test Settings from page80
Adding a Step to the Automatic Test from page81
Creating an AUTO Test File Name from page82
Saving and Exiting EDIT Status from page83 Automatic Test Page View from page84
Running an Automatic Test from page87
Automatic Test Results from page91
Before operating the GPT-9000 please read the safety precautions asoutlined in the Set Up chapter on page20.
Choose/Recall an Automatic Test
Background The tester must first be put into AUTO mode tocreate or run automatic tests.
Up to 100 automatic tests can saved/recalled.
Steps 1. If the tester is in MANU mode,press and hold the MANU/AUTOkey for three seconds. This will putthe tester into Auto mode.
The tester can only switch betweenAUTO and MANU mode when inthe VIEW status.
MANU/AUTO
-
7/25/2019 Safety Tester Manual
79/149
OPERATION
79
I R G BA CW D CW
E
F R E Q = 0 H z6
0 100k V RA C = F FO
EIV MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
M A N U = * * * - 0 0 1
W
MA U OT N A_0A U T O = 0 0 1 - 1 0
VIEW status
MM A UN N A_ E
m A
NoteIf the chosen automatic test has not yet beensetup, then the screen will be blank except for thestatus and mode.
E
EIV
M A N U = * * * - 0 0 1
W
MA U OT N A_0A U T O = 0 0 1 - 1 0
VIEW statusMM A UN N A_ E
2. Use the scroll wheel to choose theAUTO number.
AUTO #001~100
I R G BA CW D CW
E
F RE Q = 0 H z6
0 100k V RA C = F FO
EIV MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
W
MA U OT N A_0A U T O = 0 0 1 - 1 0
AUTO number
m A
Note
The AUTO number can only be chosen in VIEW
status. If in the EDIT status, switch to the VIEWstatus by pressing the EDIT/SAVE or ESC key.
-
7/25/2019 Safety Tester Manual
80/149
GPT-9000 Series User Manual
8
Edit Automatic Test Settings
Background To edit an automatic test, the tester must be inEDIT status.
Any settings or parameters that are edited onlyapply to the currently selected AUTO number.
Steps 1. Press the EDIT/SAVE key when inVIEW status to enter the EDIT
status. This will enter the EDITstatus for the chosen AUTOnumber.
EDIT/SAVE
I R G B
m A
A CW D CW A D D
E
F R E Q = 0 H z6
0 100k V RA C = F FO
IDE MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MA U OT N A_
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
T
1A U T O = 0 0 1 - 0 0
EIV W
2. The Status changes from VIEW to EDIT. Thetester is now ready to edit the current AUTOtest.
NotePressing the EDIT/SAVE key again will save thesettings or pressing the ESC will cancel thesettings for the current AUTO test and return back
to VIEW status.
-
7/25/2019 Safety Tester Manual
81/149
OPERATION
81
Adding a Step to the Automatic Test
Background Up to 16 MANU tests (steps) can be added toan automatic (AUTO) test. Each step is addedin a sequential order.
Steps 1. Press the DOWN arrow keys tobring the cursor to the MANUnumber.
I R G B
cursor MANU number
m A
A CW D CW A D D
E
F R E Q = 0 H z6
0 100k V RA C = F FO
IDE MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
T
1A U T O = 0 0 1 - 0 0
2. Use the scroll wheel to choose aMANU number to add to the
automatic test.
MANU number 001~100
3. Press the ADD soft-key to add theselected manual test to theautomatic test as another step.
A D D
4. Repeat steps 2 and 3 for any other tests that you
wish to add to the automatic test.
NoteAfter 16 steps have been added to an AUTO test,FULL will be shown on the display when youattempt to add another step to the AUTO test.
I R G B
m A
A CW D CW A D D
E
F RE Q = 0 H z6
0 100k V RA C = F FO
LUF MT I E = 0 0 1 . 0 SR
0 0 m AER F =# 0 .0MM A UN N A_
1 . 0 0 m AIH ES = 0T
0 0 . 1 S0A M P =R
L
1A U T O = 0 0 1 - 0 0
-
7/25/2019 Safety Tester Manual
82/149
GPT-9000 Series User Manual
82
NoteThe test order can be edited in the Page Viewmenu after the AUTO test is saved. See page84 for
details.
Creating an AUTO Test File Name
Background Each automatic test can have a user-defined testfile name (Default: AUTO_NAME) up to 10characters long. See the character list below for
the allowed characters.
Character List
0 1 2 3 4 5 6 7 8 9
W X Y ZM VO P RQ T USNLA B C D E F G H I J K
w x y zm vo p rq t usnla b c d e f g h i j k
+ - * / _ = : ? ( ) < > [ ]
Steps 1. Use the UP/DOWN arrow keys to
bring the cursor to the AUTOnumber. A small cursor will alsoappear under the first character ofthe AUTO test file name. This isinitially set as AUTO_NAME
I R G B
m A
A CW D CW A D D
E