report no.n950/07/14005 - find the...

25

Upload: others

Post on 14-Apr-2021

3 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report
Page 2: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 2 of 22

CONTENTS

1. INTRODUCTION ............................................................................................................3

2. DESCRIPTION OF TEST SAMPLE................................................................................4

3. TEST RIG GENERAL ARRANGEMENT.........................................................................6

4. TEST SEQUENCE .........................................................................................................7

5. SUMMARY AND CLASSIFICATION OF TEST RESULTS..............................................8

6. WATERTIGHTNESS TESTING......................................................................................9

7. WIND RESISTANCE TESTING....................................................................................11

8. IMPACT TESTING........................................................................................................17

9. APPENDIX – DRAWINGS............................................................................................22

Page 3: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 3 of 22

1. INTRODUCTION

This report describes tests carried out at the Taylor Woodrow Technology Centre at therequest of Telling Architectural Limited acting on the instructions from Aquarian CladdingSystems Ltd, Lower Ground Floor, 14 Marine Parade, Clevedon, North Somerset, BS217QS.

The test sample consisted of a Gebrik Insulated Brick Cladding System. The window wasnot included in the test.

Taylor Woodrow is accredited by the United Kingdom Accreditation Service as UKAS TestingLaboratory No.0057 and is also approved with Lloyds Register of Quality Assurance for ad-hoc in-service inspections and tests to ISO 9001 2000.

All measuring instruments and relevant test equipment were calibrated and traceable tonational standards.

The tests were carried out on the 15th January 2008 and were to determine theweathertightness of the test sample. The test methods were in accordance with the CWCTStandard Test Methods for building envelopes, 2005, for:

Watertightness – static pressure.

Watertightness – dynamic pressure.

Wind resistance – serviceability & safety.

The sample was also subjected to the following non UKAS accredited tests in accordancewith the Taylor Woodrow Quality System:

Impact resistance (BS 8200).

This test report relates only to the actual sample as tested and described herein.

The results are valid only for the conditions under which the tests were conducted.

Page 4: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 4 of 22

2. DESCRIPTION OF TEST SAMPLE

2.1 GENERAL ARRANGEMENT

The sample was as shown in the photo below and the drawings included as an appendix tothis report.

The sample comprised a brick slip system with insulation fixed onto a plywood backing wall.

The system was mounted onto a light gauge steel frame substrate.

On both sides of the window at one point the perps created a stack bond rather thanmaintaining the stretcher bond as with the rest of the sample.

PHOTO 1150013

TEST SAMPLE ELEVATION

Page 5: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 5 of 22

PHOTO 14025

SAMPLE DURING CONSTRUCTION

2.2 CONTROLLED DISMANTLING

During the dismantling of the sample no discrepancies from the drawings were found.

Page 6: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 6 of 22

3. TEST RIG GENERAL ARRANGEMENT

The test sample was mounted on a rigid test rig with support steelwork designed to simulatethe on-site/project conditions. The test rig comprised a well sealed chamber, fabricated fromsteel and plywood. A door was provided to allow access to the chamber. Representatives ofTelling installed the sample on the test rig. See Figure 1.

FIGURE 1

TYPICAL TEST RIG GENERAL ARRANGEMENT

FAN

PRESSURE TRANSDUCER

DEFLECTION TRANSDUCER

COMPUTER CONTROLLED

TEST SAMPLE

WATER SPRAY GANTRY

TEST RIG SUPPORT STEELWORK,

SEALED TEST CHAMBER

TO SIMULATE ON-SITE CONDITIONS

DATA LOGGER

CONTROLLED AND METERED AIRSUPPLY GENERATING POSITIVEAND NEGATIVE PRESSURES

SECTION THROUGH TEST RIG

Page 7: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 7 of 22

4. TEST SEQUENCE

The test sequence was as follows:

(1) Watertightness – static

(2) Wind resistance – serviceability

(3) Watertightness – static

(4) Watertightness – dynamic

(5) Wind resistance – safety

(6) Impact resistance

Page 8: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 8 of 22

5. SUMMARY AND CLASSIFICATION OF TEST RESULTS

The following summarises the results of the tests carried out. For full details refer toSections 6, 7 and 8.

5.1 TEST RESULTS

TABLE 1

Date Testnumber

Test description Result

15 January 2008 1 Watertightness – static Pass

15 January 2008 2 Wind resistance – serviceability Pass

15 January 2008 3 Watertightness – static Pass

15 January 2008 4 Watertightness – dynamic Pass

15 January 2008 5 Wind resistance – safety Pass

15 January 2008 6 Impact resistance Pass

5.2 CLASSIFICATION

TABLE 2

Test Standard Classification / Declared value

Watertightness CWCT R7

Wind resistance CWCT 2400 pascals

Page 9: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 9 of 22

6. WATERTIGHTNESS TESTING

6.1 INSTRUMENTATION

6.1.1 Water Flow

An in-line water flow meter was used to measure water supplied to the spray gantry to within5%.

6.1.2 Temperature

Platinum resistance thermometers (PRT) were used to measure air and water temperaturesto within 1C.

6.2 FAN

6.2.1 Static Pressure Testing

The air supply system comprised a variable speed centrifugal fan and associated ductingand control valves to create positive and negative static pressure differentials. The fanprovided essentially constant air flow at the fixed pressure for the period required by the testsand was capable of pressurising at a rate of approximately 600 pascals in one second.

6.2.2 Dynamic Pressure Testing

A wind generator was mounted adjacent to the external face of the sample and used tocreate positive pressure differentials during dynamic testing. The wind generator compriseda piston type aero-engine fitted with 4 m diameter contra-rotating propellers.

6.3 WATER SPRAY

The water spray system comprised nozzles spaced on a uniform grid not more than 700 mmapart and mounted approximately 400 mm from the face of the sample. The nozzlesprovided a full-cone pattern with a spray angle between 90° and 120°. The spray systemdelivered water uniformly against the exterior surface of the sample.

6.4 PROCEDURE

6.4.1 Watertightness – static

Three positive pressure pulses of 1200 pascals were applied to prepare the test sample.

Water was sprayed onto the sample using the method described above at a rate of at least3.4 litres/m2/minute for 15 minutes at zero pressure differential. With the water spraycontinuing the pressure differential across the sample was then increased in increments of:50, 100, 150, 200, 300, 450 and 600 pascals, each held for 5 minutes.

Throughout the test the interior face of the sample was examined for water penetration.

Page 10: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 10 of 22

6.4.2 Watertightness – dynamic

Water was sprayed onto the sample using the method described above at a flow rate of atleast 3.4 litres/m2/minute.

The aero-engine was used to subject the sample to wind of sufficient velocity to produce apressure differential of 600 pascals. These conditions were maintained for 15 minutes.Throughout the test the inside of the sample was examined for water penetration.

6.5 PASS/FAIL CRITERIA

There shall be no water penetration to the internal face of the sample throughout testing.

6.6 RESULTS

Test 1 Date: 15 January 2008

No water penetration was observed throughout the test.

Chamber temperature= 10°CAmbient temperature = 9°CWater temperature = 8°C

Test 3 Date: 15 January 2008

No water penetration was observed throughout the test.

Chamber temperature= 10°CAmbient temperature = 9°CWater temperature = 8°C

Test 4 Date: 15 January 2008

No water penetration was observed throughout the test.

Chamber temperature= 11°CAmbient temperature = 10°CWater temperature = 8°C

Page 11: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 11 of 22

7. WIND RESISTANCE TESTING

7.1 INSTRUMENTATION

7.1.1 Pressure

One static pressure tapping was provided to measure the chamber pressure and was locatedso that the readings were unaffected by the velocity of the air supply into or out of thechamber.

A pressure transducer, capable of measuring rapid changes in pressure to within 2% wasused to measure the differential pressure across the sample.

7.1.2 Deflection

Displacement transducers were used to measure the deflection of principle framing membersto an accuracy of 0.1 mm. The gauges were set normal to the sample framework at mid-span and as near to the supports of the members as possible and installed in such a waythat the measurements were not influenced by the application of pressure or other loading tothe sample. The gauges were located at the positions shown in Figure 2.

7.1.3 Temperature

Platinum resistance thermometers (PRT) were used to measure air temperatures to within1C.

7.1.4 General

Electronic instrument measurements were scanned by a computer controlled data logger,which also processed and stored the results.

7.2 FAN

The air supply system comprised a variable speed centrifugal fan and associated ductingand control valves to create positive and negative static pressure differentials. The fanprovided essentially constant air flow at the fixed pressure for the period required by the testsand was capable of pressurising at a rate of approximately 600 pascals in one second.

7.3 PROCEDURE

7.3.1 Wind Resistance – serviceability

Three positive pressure differential pulses of 1200 pascals were applied to prepare thesample. The displacement transducers were then zeroed.

The sample was subjected to one positive pressure differential pulse from 0 to 2400 pascalsto 0. The pressure was increased in four equal increments each maintained for 15 ±5seconds. Displacement readings were taken at each increment. Residual deformationswere measured on the pressure returning to zero.

Any damage or functional defects were recorded.

The test was then repeated using a negative pressure of -2400 pascals.

7.3.2 Wind Resistance – safety

Page 12: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 12 of 22

Three positive pressure differential pulses of 1200 pascals were applied to prepare thesample. The displacement transducers were then zeroed.

The sample was subjected to one positive pressure differential pulse from 0 to 3600 pascalsto 0. The pressure was increased as rapidly as possible but not in less than 1 second andmaintained for 15 ±5 seconds. Displacement readings were taken at peak pressure.Residual deformations were measured on the pressure returning to zero.

Any damage or functional defects were recorded.

The test was then repeated using a negative pressure of –3600 pascals.

Page 13: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 13 of 22

FIGURE 2

DEFLECTION GAUGE LOCATIONS

External View

1

2

3

5

6

4

Page 14: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 14 of 22

7.4 PASS/FAIL CRITERIA

7.4.1 Calculation of permissible deflection

Gaugenumber

Member Span(L)

(mm)

Permissibledeflection

(mm)

Permissibleresidual

deformation

2

5

Vertical wall

Vertical frame

2400

1370

L/200= 12.0

L/200= 6.8

1 mm

1 mm

7.5 RESULTS

Test 2 (serviceability) Date: 15 January 2008

The deflections measured during the wind resistance test, at the positions shown in Figure 2,are shown in Table 4.

Summary Table:

Gaugenumber

Member Pressuredifferential

(Pa)

Measureddeflection

(mm)

Residualdeformation

(mm)

2

5

Vertical wall

Vertical frame

2426-2363

2426-2363

2.6-2.1

1.1-0.9

0.00.0

0.00.0

No damage to the test sample was observed.

Ambient temperature = 9°CChamber temperature = 11°C

Test 5 (safety) Date: 15 January 2008

The deflections measured during the structural safety test, at the positions shown in Figure 2,are shown in Table 5.

No damage to the sample was observed.

Ambient temperature = 10°CChamber temperature = 12°C

Page 15: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 15 of 22

TABLE 4

WIND RESISTANCE – SERVICEABILITY TEST RESULTS

Position Pressure (pascals) / Deflection (mm)

573 1239 1817 2426 Residual

1 0.2 0.4 0.6 0.8 0.1

2 1.8 2.7 3.8 4.8 0.0

3 2.3 2.7 3.2 3.7 0.0

4 1.6 2.5 3.5 4.7 0.2

5 2.4 3.5 4.7 6.1 0.1

6 2.7 3.5 4.3 5.3 0.1

2 * 0.6 1.2 1.9 2.6 0.0

5 * 0.3 0.5 0.8 1.1 0.0

Position Pressure (pascals) / Deflection (mm)

-606 -1216 -1867 -2363 Residual

1 -0.1 -0.3 -0.5 -0.6 -0.1

2 -1.0 -1.9 -2.8 -3.5 0.0

3 -0.6 -1.1 -1.7 -2.2 -0.1

4 -0.9 -1.8 -2.9 -3.9 -0.2

5 -1.1 -2.2 -3.4 -4.5 -0.2

6 -0.9 -1.6 -2.5 -3.2 -0.2

2 * -0.6 -1.2 -1.7 -2.1 0.0

5 * -0.3 -0.5 -0.7 -0.9 0.0

* Mid-span reading adjusted between end support readings

Page 16: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 16 of 22

TABLE 5

WIND RESISTANCE - SAFETY TEST RESULTS

Position Pressure (pascals) / Deflection (mm)

3623 Residual -3607 Residual

1 0.9 -0.1 -0.9 -0.1

2 6.6 -0.1 -5.3 0.0

3 4.6 0.0 -3.1 0.0

4 6.8 0.1 -6.0 -0.1

5 8.5 0.1 -6.7 -0.1

6 7.1 0.1 -4.6 -0.1

2 * 3.9 -0.1 -3.2 0.1

5 * 1.6 0.0 -1.4 0.0

* Mid-span reading adjusted between end support readings

Page 17: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 17 of 22

8. IMPACT TESTING

8.1 IMPACTOR

8.1.1 Soft body

The soft body impactor comprised a canvas spherical/conical bag 400 mm in diameter filledwith 3 mm diameter glass spheres with a total mass of approximately 50 kg suspended froma cord at least 3 m long.

8.1.2 Hard body

The hard body impactor was a solid steel ball of 70 mm diameter and approximate mass of2.2 kg. Note: This impactor was larger than the standard impactor (62 mm, 1.0 kg)

8.2 PROCEDURE (BS 8200)

8.2.1 Soft body

The impactor almost touched the face of the sample when at rest. It was swung in apendular movement to hit the sample normal to its face. The test was performed at locations1 and 2 shown in Figure 3. The impact energies were 120 and 500 Nm.

8.2.2 Hard body

The impactor almost touched the face of the sample when at rest. It was swung in apendular movement to hit the sample normal to its face. The test was performed at location3 shown in Figure 3. The impact energy was 22 Nm (standard requires only 10 Nm).

8.3 PASS/FAIL CRITERIA

8.3.1 At impact energies for retention of performance

There shall be no failure, significant damage to surface finish or significant indentation.

8.3.2 At impact energies for safety

The structural safety of the building shall not be put at risk, no parts shall be made liable tofall or to cause serious injury to people inside or outside the building. The soft body impactorshall not pass through the wall. Damage to the finish and permanent deformation on the farside of the wall may occur.

Page 18: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 18 of 22

8.4 RESULTS

Test 6 Date: 15 January 2008

Location 1.

120 Nm soft body. No damage to the sample was observed.

500 Nm soft body. No damage to the sample was observed.

Location 2.

120 Nm soft body. No damage to the sample was observed.

500 Nm soft body. No damage to the sample was observed.

Location 3.

22 Nm hard body. No damage to the sample was observed.

Ambient temperature = 11°C

Page 19: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 19 of 22

FIGURE 3

IMPACT TEST LOACTIONS

External View

1

2

3

Page 20: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 20 of 22

PHOTO IMPACT

SOFT BODY IMPACTOR

Page 21: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 21 of 22

PHOTO 115015

HARD BODY IMPACTOR

Page 22: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report

Report No.N950/07/14005Page 22 of 22

9. APPENDIX – DRAWINGS

The following 3 unnumbered pages are copies of drawings showing the test sample.

END OF REPORT

Page 23: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report
Page 24: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report
Page 25: Report No.N950/07/14005 - Find The Needlepdfs.findtheneedle.co.uk/6727-Gebrik-CWCT-Report-14005.pdf · 2014. 10. 1. · Report No.N950/07/14005 Page 3 of 22 1. INTRODUCTION This report