rapid determination of inorganic ions in wastewater by ion chromatography
DESCRIPTION
Wastewater is produced by multiple sources, including chemical manufacturing, power generation, petroleum product extraction, and private residences. Specific industries can use knowledge of around the analytes present in wastewater to make decisions on reuse, treatment, or whether disposal is the most cost effective option. Prior to any discharge into ground or surface waters, the level of specific analytes must be determined to ensure that they do not exceed regulated limits. If these limits are being exceeded, treatment will be required. Ion Chromatography (IC) is the primary technique used for measuring the concentration of ions in wastewater and numerous methods have been developed that meet regulatory requirements. Learn about IC methods that enable accurate, consistent, and rapid measurement of both anions, such as chloride, sulfate, and bromate, and cations, such as sodium and magnesium.TRANSCRIPT
1
The world leader in serving science
Peter Bodsky
Field Marketing Manager
March 26, 2014
Rapid Determination of Inorganic Ions in Wastewater by Ion Chromatography
2
Agenda
• Wastewater sources• Regulations and methods• Analytes of interest in wastewater• Challenge of wastewater analysis• Ion Chromatography system innovations
• Inline conductivity and automated dilution
• Reagent-Free™ Ion Chromatography (RFIC™)
• High-Pressure™ Capillary IC (HPIC™)
• Conclusions
3
Examples of Wastewater Sources
• Manufacturing• Oil and gas extraction• Petroleum refining• Mining• Power generation• Household sewage• Agriculture
4
Reasons to Perform Wastewater Analysis
• Monitoring discharge • Regulatory limits
• Nutrient Analysis• Excessive plant growth in aqueous environments
• Known samples • Historical analysis
• E.g., High chloride level from a treatment facility with a water inlet near the sea/estuary
• Unknown samples • Investigative, pollution incident, farm run-off, milk spill, or industrial
discharge plant failure
5
International Wastewater Regulations
• ISO• International Organization for Standardization
• ASTM International• “American Society for Testing and Materials”
• DIN • Deutsches Institut für Normung
• German Standard
• U.S. EPA• Environmental Protection Agency
6
Controlling Water Pollution in the U.S.
• National Pollutant Discharge Elimination System (NPDES) requirements• EPA
• Industrial or municipal facilities must obtain a permit to discharge to surface water• Discharge limits
• Monitoring and reporting requirements
• Ensure that surface waters stay safe for marine life, surrounding vegetation, recreational use, and as a drinking water source
7
Regulatory Method for Anions: EPA Method 300.0
• Revision 2.1 Parts A and B published in 1993• Outlines the method for determination of inorganic anions by ion
chromatography• Specifies use of suppressed conductivity for determination of:
• Bromide (Part A) • Ortho-Phosphate-P (Part A)
• Chloride (Part A) • Sulfate (Part A)
• Fluoride (Part A) • Bromate (Part B)
• Nitrate (Part A) • Chlorate (Part B)
• Nitrite (Part A) • Chlorite (Part B)
• Applies to:• Drinking water
• Ground and surface water
• Wastewater (domestic and industrial)
• Raw water (unfinished drinking water)
8
Column: Thermo Scientific™ Dionex™ IonPac™ AG4A-SC, AS4A-SC,
4 mm i.d.Eluent: 1.7 mM sodium bicarbonate/
1.8 mM sodium carbonateFlow Rate: 2.0 mL/minInjection: 50 µLDetection: Suppressed conductivity,
Thermo Scientific™ Dionex™ ASRS™ ULTRA Anion Self-Regenerating
Suppressor™, recycle mode
Peaks: 1. Fluoride 2 mg/L2. Chloride 33. Nitrite 54. Bromide 105. Nitrate 106. Phosphate 157. Sulfate 15
0 2 4 6 8Minutes
0
µS
10
10
1
23 4 5
6
7
EPA Method 300.0 (A)
9
EPA Method 300.1
• Published in 1997• Refinement of Method 300.0
• Greater sensitivity for ions
• Single, higher capacity Dionex IonPac AG9-HC column• 2 mm or 4 mm i.d.
• Different injection volumes allowed depending on analytes and columns
10
Regulatory Method for Cations: ASTM D6919-03
• Outlines the method for determination of cations by ion chromatography• Specifies use of suppressed and nonsuppressed conductivity for
determination of:• Lithium
• Sodium
• Ammonium
• Potassium
• Magnesium
• Calcium
• Applies to:• Drinking water
• Reagent water
• Wastewaters
11
0 5 10 15 20
Separation of Alkali and Alkaline Earth Metals and Ammonium
Column: Dionex IonPac CS12A, 4 mm i.d.Eluent: 18 mN Methanesulfonic acidFlow Rate: 1.0 mL/minInj. Volume: 25 µLDetection: Suppressed conductivity,
Thermo Scientific™ Dionex™ CSRS™ 300 Cation Self-Regenerating Suppressor, recycle mode
Peaks: 1. Lithium 1 mg/L 2. Sodium 4
3. Ammonium 5 4. Potassium 10 5. Rubidium 10 6. Cesium 10 7. Magnesium 5 8. Calcium 10 9. Strontium 1010. Barium 10
1
4
6
2
3
Minutes
25
10
µS
0
9
8
5
7
20
12
Common Ions in Wastewater Measured by IC
• Inorganic anions• Chloride
• Disrupts nitrification processes
• Sulfate• Disrupt anaerobic digestion processes
• Nitrate, nitrite, phosphate• Plant nutrients; algal blooms and deoxygenation
• Bromide• Ozonation, chlorination -> Disinfection By-Products: brominated
trihalomethanes, bromate- Carcinogenic
• Organic acids• Formic and acetic acids
• pH balance
13
Common Ions in Wastewater Measured by IC
• Cations• Potassium, sodium
• Disrupts osmotic balance of plants
• Lithium • Human toxicity
• Ammonium• Corrosive
• Magnesium, calcium, barium• Scale buildup
• Strontium• Radioactive
14
Challenge of Wastewater Analysis
High concentrations of dissolved salts:
• Exceed column capacity• Poor chromatography
• Peak suppression
• Inaccurate reporting
• Exceed linear calibration range• Analyte-specific
• Inaccurate results
• Decrease column lifetime
0 2 4 110
12,000
µS
Minutes6 8 10
0
50
µS
0 2 4 116 8 10
Minutes
Undiluted
Diluted
15
Obtaining Accurate Data From Wastewater
• Manual analysis• Post-run
• Determine concentration from chromatogram peak area- Exceed limit → dilute → re-run sample
• Pre-run
• Manual conductivity measurement - Exceed limit → dilute → run sample
• Tedious
• Dilutions prone to errors
16
Increasing Accuracy With Automation
• Automated analysis• “AutoDilution”
• Post-run analysis using ion chromatograph software to determine if too much sample was loaded
• In-line Conductivity• Conductivity measured prior to loading sample onto column
• Exceeding upper limit triggers loading of less sample
• Less sample injected• Smaller sample loop
• Partial loop
• Automated sample dilution• Loading of much lower sample amounts
17
Automated Analysis: AutoDilution
Wastewater
Centrifugation
Filtration
Automated Sample Dilution
Report
Chromatogram
Thermo Scientific Dionex AS-APAutosampler
No
Yes Does peak area or height exceed
cut-off?
AutoDilution
Thermo Scientific™ Dionex™
Chromeleon™ CDS Software
IC System
Thermo Scientific™ Dionex™ ICS-2100
RFIC™ System
18
Automated Analysis: In-line Conductivity and Automated Dilution
Wastewater
Centrifugation
Filtration
Automated Sample Dilution
Does conductivity exceed cutoff?
Yes
No
Report
Chromatogram
Thermo Scientific Dionex AS-APAutosampler
Thermo Scientific™ Dionex™
Chromeleon™ CDS Software
IC System
Dionex ICS-2100
RFIC System
Thermo Scientific Dionex
Sample Conductivity and pH Accessory
19
Analysis of Anions in Automatically Diluted Fracking Flowback Wastewater
Peaks: Measured Undiluted
1. Acetate < 0.05 mg/L < 52. Formate < 0.05 < 53. Chloride 940.0 94,000 4. Sulfate 0.12 12 5. Bromide 8.90 890
0.0
0.65
µS
Minutes
0 2 4 8
0
2,400
µS
Minutes
3
1 2
3
4
5
6
0 2 4 86
5
4
1 2
Column: Dionex IonPac AG18/AS18, 4 mm i.d.
Eluent Source: Thermo Scientific Dionex
EGC III KOH cartridgeEluent: 39 mM KOHFlow Rate: 1 mL/minInj. Volume: 25 µLCol. Temp.: 30 °CDetection: Suppressed conductivity,
Dionex ASRS 300 Anion Self-Regenerating Suppressor, recycle mode
Sample: 100-fold diluted fracking flowback, filtered, 0.2 µm
20
Analysis of Cations in Automatically Diluted Fracking Flowback Wastewater
Peaks: Measured Undiluted
1. Lithium < 0.05 mg/L < 52. Sodium 28.0 28,0003. Ammonium 0.35 350 4. Potassium 0.50 500 5. Magnesium 1.1 1,100 6. Calcium 10.0 10,000
0.0
8.2
µS
Minutes
0 5 10 23
0
80
µS
Minutes
3
Column: Dionex IonPac CG16/CS16,0.4 mm i.d.
Eluent Source: Dionex EGC-MSA (capillary) cartridgeEluent: 30 mM MSAFlow Rate: 0.01 mL/minInj. Volume: 0.4 µLCol. Temp.: 40 °CDetection: Suppressed conductivity,
Dionex CCES 300 Cation Self-Regenerating Suppressor,
recycle modeSample: 1000-fold diluted fracking flowback, filtered, 0.2 µm
1
2
3
4
5
15
5
4
1
2
6
20
6
0 5 10 2315 20
21
Innovations That Have Made IC Easier
• Packed Bed Suppression • Reduced background due to mobile phase for improved signal
• Electrolytic Suppression • Greater flexibility in mobile phase selection/strength; no chemical
regeneration
• Reagent-Free IC (RFIC) • Electrolytic eluent generation makes gradient separations as easy as
isocratic; just add water
• Capillary IC• 18 months continuous operation, infrequent calibration; IC on Demand
• High-pressure IC (HPIC)• Higher flow rate, faster runs; use of 4 µm particle columns for improved
efficiency, resolution, and chromatographic flexibility
22
• High purity eluents generated on line• Accurate, precise, reproducible• Just add water
Precise, Contamination Free Eluents Day after Day, Analyst to Analyst, Lab to Lab
Pump
H2O
K+
Purified KOH
[KOH] Current
Flow Rate
Pt Cathode(2H20 + 2e- 20H- + H2)-
Reagent-Free IC with Eluent Generation (RFIC-EG)
23
RFIC
Dionex ICS-900System
Dionex ICS-1100 System
Dionex ICS-4000 System
Dionex ICS-2100 System
Dionex ICS-1600 System
Dionex ICS-5000+
System
HPIC
The Dionex Ion Chromatography Product Line
24
Capillary HPIC Advantages
• “IC on Demand”• System is always ready for your samples
• Higher laboratory productivity
• Less labor needed for calibration
• High-pressure• Up to 5000 psi
• Faster runs
• Lower cost of ownership• Less eluent consumed and waste generated
25
Thermo ScientificDionex ICS-5000+
HPIC System
Thermo ScientificDionex ICS-4000
Capillary HPIC System
Capillary HPIC Systems
26
Capillary Technology – The Dionex IC Cube Module
Guard and Separation Column
Suppressor
CRD 200
Injection ValveEG Degas
Side View of CapSuppressor
8.9 cm
16.5 cm
27
5 25-1.0
60
µS
Minutes
Column: Dionex IonPac AG19/AS19,0.4 mm i.d.
Eluent Source: Dionex EGC-KOH cartridge (Capillary)Gradient: 10 to 25 mM KOH (0–10 min)25 to 70 mM KOH (10–20 min)
10 mM KOH (20–25 min) Flow Rate: 0.010 mL/minInj. Volume: 0.4 µLColumn Temp.: 30 °CDetection: Suppressed conductivity,
Dionex ACES 300, recyclemodeSample: Treated wastewater, filtered, 0.2 µm
Peaks:0.4 mm i.d. 4 mm i.d.*
1. Fluoride 1.76 mg/L 1.69 2. Chloride 180.00 180.003. Bromide 0.42 0.514. Nitrate 11.80 11.905. Sulfate 96.90 96.86. Phosphate 0.94 1.25
*Data from 4 mm i.d. column using appropriate run conditions (Dionex ICS-1100 System)
1
2
3
4 5
6
10 15 200
Determination of Common Anions in Treated Wastewater
28
Fast Determination of Inorganic Ions Using the Dionex IonPac AS18-4µm Column
Column: Dionex IonPac AG18-4µm/AS18-4µm, 0.4 150 mm
Eluent Source: Dionex EGC-KOH Cartridge (Capillary) Eluent: 23 mM KOH Flow Rate: A: 0.010, B: 0.015, C: 0.025 mL/minInj. Volume: 0.4 µLColumn Temp.:30 °C Detection: Suppressed conductivity,
Dionex ACES 300, recycle mode
Peaks:1. Fluoride 0.1 mg/L2. Chlorite 1.03. Chloride 0.6
4. Nitrite 1.05. Carbonate --6. Bromide 2.07. Sulfate 2.08. Nitrate 2.09. Chlorate 2.0
3
64
750
Minutes
4321
0
13
µS
78
6
9
5
2
1 C: 0.025 mL/min3800 psi
B: 0.015 mL/min2400 psi
A: 0.010 mL/min1600 psi
29
Fast Determination of Inorganic Anions in Municipal Wastewater
Peaks (Total):A B C D
1. Chloride 76.5 146 154 130 mg/L
2. Nitrite 1.5 2.1 37.41.6
3. Carbonate -- -- -- --
4. Sulfate 41.6 88.9 84.891.8
5. Nitrate 28.8 7.2 31.7128
3
4
40
Minutes
3210
1.7
µS
B
A
5
2
1
C
D
Column: Dionex IonPac AG18-4µm/AS18-4µm, 0.4 mm i.d.
Eluent Source: Dionex EGC-KOH Cartridge (Capillary) Eluent: 23 mM KOH Flow Rate: 0.025 mL/minInj. Volume: 0.4 µLColumn Temp.: 30 °C Detection: Suppressed conductivity,
Dionex ACES 300, recyclemodeSample Prep: Diluted 1000-fold, filtered, 0.2 µmSamples: A: Influent
B: Primary effluentC: Trickling effluentD: Final effluent
30
Fast Determination of Cations in Municipal Wastewater
Column: Dionex IonPac CG16/CS16, 0.5 mm i.d.
Eluent Source: Dionex EGC-MSA Cartridge (Capillary)
Eluent: 30 mM MSAFlow Rate: A: 0.010 mL/min,
B: 0.030 mL/minInj. Volume: 0.4 µLColumn Temp.: 40 °CDetection: Suppressed Conductivity,
Dionex CCES 300, recycle mode
Sample: Wastewater diluted 50-fold, filtered, 0.2 µm
Peaks:1. Sodium 195.9 mg/L2. Ammonium ---3. Potassium 11.6
4. Magnesium 38.05. Calcium 52.9
Minutes
µS30 µL/min 3720 psi
A
B
0 10 30-2
14
10 µL/min 1250 psi
1
2
4
4
1
2
3
5
5
3
20
31
Conclusions
• The high ion concentrations typical of wastewater presents a challenge to analysis
• In-line conductivity measurement and automated sample dilution combine to ensure that what is loaded onto an IC column is within the calibration range
• Reagent-Free IC removes the inconvenience and variability of manual eluent preparation
• High-pressure capillary IC allows the use of faster flow rates for quick run times, while producing very little waste
32