Radiotherapy Treatment PlanningRadiotherapy Treatment Planning
Anwar 2
Treatment planning is the task to make Treatment planning is the task to make sure a prescription is put into practice sure a prescription is put into practice in an optimized wayin an optimized way
Prescription
Planning
Treatment
Anwar 3
ObjectivesObjectives Understand the general principles of Understand the general principles of
radiotherapy treatment planningradiotherapy treatment planning Appreciate different dose calculation Appreciate different dose calculation
algorithms algorithms Understand the need for testing the treatment Understand the need for testing the treatment
plan against a set of measurementsplan against a set of measurements Be able to apply the concepts of optimization Be able to apply the concepts of optimization
of medical exposure throughout the treatment of medical exposure throughout the treatment planning processplanning process
Anwar 4
Contents of the lectureContents of the lecture
A. Radiotherapy treatment planning A. Radiotherapy treatment planning conceptsconcepts
B. Computerized treatment planningB. Computerized treatment planning
Anwar 5
The need to understand The need to understand treatment planningtreatment planning IAEA Safety Report Series 17 “IAEA Safety Report Series 17 “Lessons Lessons
learned from accidental exposures in learned from accidental exposures in radiotherapyradiotherapy “ (Vienna 2000): “ (Vienna 2000):
About 1/3 of problems directly related to About 1/3 of problems directly related to treatment planning!treatment planning!
May affect individual patient or cohort of May affect individual patient or cohort of patientspatients
Anwar 6
A. Basic Radiotherapy Treatment A. Basic Radiotherapy Treatment Planning ConceptsPlanning Concepts
i. Planning process overviewi. Planning process overview
ii. Patient data required for planningii. Patient data required for planning
iii. Machine data required for planningiii. Machine data required for planning
iv. Basic dose calculationiv. Basic dose calculation
Anwar 7
i. Planning process overviewi. Planning process overview Combine machine parameters and individual patient Combine machine parameters and individual patient
data to customize and optimize treatmentdata to customize and optimize treatment Requires machine data, input of patient data, Requires machine data, input of patient data,
calculation algorithm calculation algorithm Produces output of data in a form which can be used Produces output of data in a form which can be used
for treatment (the ‘treatment plan’)for treatment (the ‘treatment plan’)
Patient information
Planning
Treatment unit data
Treatment plan
Anwar 9
ii. Patient information requiredii. Patient information required
Radiotherapy is a localized treatment of Radiotherapy is a localized treatment of cancer - one needs to know not only the dose cancer - one needs to know not only the dose but also the accurate volume where it has but also the accurate volume where it has been delivered to.been delivered to.
This applies to tumor as well as normal This applies to tumor as well as normal structures - the irradiation of the latter can structures - the irradiation of the latter can cause intolerable complications. Again, both cause intolerable complications. Again, both volume and dose are important.volume and dose are important.
Anwar 10
One needs to knowOne needs to know Target locationTarget location Target volume and shapeTarget volume and shape Secondary targets - potential tumor Secondary targets - potential tumor
spreadspread Location of critical structuresLocation of critical structures Volume and shape of critical structuresVolume and shape of critical structures Radiobiology of structuresRadiobiology of structures
Anwar 11
Target delineation Target delineation ICRU 50 & 62ICRU 50 & 62
Gross Tumor Volume (GTV) = clinically demonstrated tumor
Clinical Target Volume = GTV + area at risk (eg. potentially involved lymph nodes)
It all comes down to the correct It all comes down to the correct dose to the correct volumedose to the correct volume
Dose Volume Histograms are a way to Dose Volume Histograms are a way to summarize this information summarize this information
0
20
40
60
80
100
120
0 20 40 60 80
Dose (Gy)
Vo
lum
e (%
)
0
20
40
60
80
100
120
0 20 40 60 80
Dose (Gy)
Vo
lum
e (%
)Comparison of three differenttreatment techniques (red, blue and green)in terms of dose to the target and a critical structure
Target dose
Criticalorgan
Anwar 13
The ideal DVHThe ideal DVH
Tumor:Tumor: High dose to all High dose to all Homogenous doseHomogenous dose
Critical organCritical organ Low dose to most of Low dose to most of
the structurethe structure
100%
dose
100%
dose
Anwar 15
Need to keep in mindNeed to keep in mind
Always a 3D problemAlways a 3D problem Different organs may respond differently Different organs may respond differently
to different dose patterns. to different dose patterns. Question: Is a bit of dose to all the Question: Is a bit of dose to all the
organ better than a high dose to a small organ better than a high dose to a small part of the organ?part of the organ?
Anwar 16
In practice not In practice not always that clear cutalways that clear cut
ICRU report 62ICRU report 62 Need to understand Need to understand
anatomy and anatomy and physiologyphysiology
A clinical decisionA clinical decision
Anwar 17
*Int. J. Radiat. Oncol. Biol. Phys., 1998; 41:84-92.
In many organs, dose and volume In many organs, dose and volume effects are linked - eg.effects are linked - eg.
Dose(Gy)
Rectalvolume(%)
>65 40
>70 30
>75 5
Boersma* et al., classified the following (Dose,Volume) regions to be regions of high risk for developing rectal bleeding:
Anwar 18
In EBT practiceIn EBT practice Need to know Need to know
where to direct beam to, andwhere to direct beam to, and how large the beam must be and how it how large the beam must be and how it
should be shapedshould be shaped
Anwar 19
Target design and reference Target design and reference imagesimages In radiotherapy practice the target is In radiotherapy practice the target is
localized using diagnostic tools:localized using diagnostic tools: Diagnostic procedures - palpation, X-ray, Diagnostic procedures - palpation, X-ray,
ultrasoundultrasound Diagnostic procedures - MRI, PET, SPECTDiagnostic procedures - MRI, PET, SPECT Diagnostic procedures - CT scan, simulator Diagnostic procedures - CT scan, simulator
radiographradiograph
Anwar 20
Selection of treatment approachSelection of treatment approach
Requires training and experienceRequires training and experience May differ from patient to patientMay differ from patient to patient Requires good diagnostic toolsRequires good diagnostic tools Requires accurate spatial informationRequires accurate spatial information May require information obtained from May require information obtained from
different modalitiesdifferent modalities
Anwar 21
Minimum patient data required for Minimum patient data required for external beam planningexternal beam planning
Target locationTarget location Patient outlinePatient outline
Anwar 22
Diagnostic tools which could be Diagnostic tools which could be used for patient data acquisitionused for patient data acquisition CT scanner, MRI, PET scanner, US,…CT scanner, MRI, PET scanner, US,… Simulator including laser system, optical Simulator including laser system, optical
distance indicator (ODI)distance indicator (ODI) Many functions of the simulator are also Many functions of the simulator are also
available on treatment units as an alternative available on treatment units as an alternative - simulator needs the same QA! - simulator needs the same QA!
Anwar 23
Note on the role of simulationNote on the role of simulation Simulator is often used twice in the Simulator is often used twice in the
radiotherapy processradiotherapy process Patient data acquisition - target localization, Patient data acquisition - target localization,
contours, outlinescontours, outlines Verification - can the plan be put into Verification - can the plan be put into
practice? Acquisition of reference images practice? Acquisition of reference images for verification.for verification.
Simulator may be replaced by other Simulator may be replaced by other diagnostic equipment or virtual diagnostic equipment or virtual simulationsimulation
Anwar 24
Virtual simulationVirtual simulation
All aspects of simulator work are All aspects of simulator work are performed on a 3D data set of the performed on a 3D data set of the patientpatient
This requires high quality 3D CT data of This requires high quality 3D CT data of the patient in treatment positionthe patient in treatment position
Verification can be performed using Verification can be performed using digitally reconstructed radiographs digitally reconstructed radiographs (DRRs)(DRRs)
Anwar 25
Virtual SimulationVirtual Simulation
3D Model of the patient and the Treatment Devices
Anwar 26
SimulatorSimulatorDiagnosticX-ray tube
Simulator couch
Rotatinggantry
Image intensifierand X-ray film
holder
Radiation beamdefining system
Nucletron/Oldelft
Anwar 27
Radiotherapy simulatorRadiotherapy simulator
Obtain images and Obtain images and mark beam entry mark beam entry points on the patientpoints on the patient
Anwar 28
Moveable Lasers
Isocenter Projection
Isocen
ter P
ositio
n
CT images
CT Simulation (Thanks to ADAC)Marking the Patient already during CT
Anwar 29
Patient markingPatient marking
Create relation Create relation between patient between patient coordinates and coordinates and beam coordinatesbeam coordinates
Tattoos Skin markers
Marks on shell
Anwar 42
Beam placement and shapingBeam placement and shaping
simulator filmwith block
DRR withconformal shielding
Anwar 43
Tools for optimization of the Tools for optimization of the radiotherapy approachradiotherapy approach
Choice of radiation Choice of radiation qualityquality
Entry pointEntry point Number of beamsNumber of beams Field sizeField size BlocksBlocks WedgesWedges CompensatorsCompensators
Anwar 44
Optimization approachesOptimization approaches
patient
target
beam
patienttarget
beam
patient
target
wedge
Choice of bestbeam angle
Use of a beammodifier
Anwar 45
Beam number and weightingBeam number and weighting
patient
target
beam100%
patient
Beam 150%
Beam 2
50%
30%
40%
10%
20%
Anwar 46
A note on weighting of beamsA note on weighting of beams
30%
40%
10%
20%
Different approaches arepossible: 1. Weighting of beams asto how much they contributeto the dose at the target2. Weighting of beams asto how much dose isincident on the patient
These are NOT the same
25%
25%
25%
25%
Anwar 47
Use of wedgesUse of wedges
Wedged pairWedged pair Three field Three field
techniquestechniques
patient
Isodose lines
patient Typical isodose lines
Anwar 48
Beam placement and shapingBeam placement and shaping
Entry pointEntry point Field sizeField size BlocksBlocks WedgesWedges CompensatorsCompensators
a two-dimensional a two-dimensional approach?approach?
Anwar 49
Beam placement and shapingBeam placement and shaping
Entry pointEntry point Field sizeField size BlocksBlocks WedgesWedges CompensatorsCompensators
Multiple beamsMultiple beams Dynamic deliveryDynamic delivery Non-coplanarNon-coplanar Dose compensation Dose compensation
(IMRT) not just (IMRT) not just missing tissuemissing tissue
Biological planningBiological planning
This is actually a 3D approach
Anwar 50
Target LocalizationTarget Localization Diagnostic procedures - palpation, X-ray, Diagnostic procedures - palpation, X-ray,
ultrasoundultrasound Diagnostic procedures - MRI, PET, SPECTDiagnostic procedures - MRI, PET, SPECT Diagnostic procedures - CT scan, Diagnostic procedures - CT scan,
simulator radiographsimulator radiograph
Allows the creation of Reference Images for Treatment Verification:
Simulator Film, Digitally Reconstructed Radiograph
Anwar 51
Simulator imageSimulator image
During ‘verification During ‘verification session’ the treatment session’ the treatment is set-up on the is set-up on the simulator exactly like it simulator exactly like it would be on the would be on the treatment unit.treatment unit.
A verification film is A verification film is taken in ‘treatment’ taken in ‘treatment’ geometrygeometry
Anwar 52
Simulator FilmSimulator Film Shows relevant Shows relevant
anatomyanatomy Indicates field Indicates field
placement and sizeplacement and size Indicates shieldingIndicates shielding Can be used as Can be used as
reference image for reference image for treatment treatment verificationverification
Field defining wires
Anwar 57
iii. Machine data requirements for iii. Machine data requirements for treatment planningtreatment planning
Beam description (quality, energy)Beam description (quality, energy) Beam geometry (isocentre, gantry, table)Beam geometry (isocentre, gantry, table) Field definition (source collimator distance, Field definition (source collimator distance,
applicators, collimators, blocks, MLC)applicators, collimators, blocks, MLC) Physical beam modifiers (wedges, Physical beam modifiers (wedges,
compensator)compensator) Dynamic beam modifiers (dynamic wedge, Dynamic beam modifiers (dynamic wedge,
arcs, MLC IMRT)arcs, MLC IMRT) Normalization of doseNormalization of dose
Anwar 58
Machine data required for Machine data required for planningplanning Depends on Depends on
complexity of treatment complexity of treatment approachesapproaches
resources available for resources available for data acquisitiondata acquisition
May be from published May be from published data or can be acquireddata or can be acquired
MUST be verified...MUST be verified...
Quick Question:Quick Question:
Who is responsible for the Who is responsible for the preparation of beam data for preparation of beam data for
the planning process?the planning process?
Anwar 60
Acquisition of machine dataAcquisition of machine data
……from vendor or from vendor or publications (eg BJR 17 and publications (eg BJR 17 and 25) - this requires 25) - this requires verification!!!verification!!!
Done by physicistDone by physicist Some dosimetric equipment Some dosimetric equipment
must be available (water must be available (water phantom, ion chambers, film, phantom, ion chambers, film, phantoms,…)phantoms,…)
Documentation essentialDocumentation essential
Anwar 61
Machine data availabilityMachine data availability
Hardcopy (isodose charts, output factor Hardcopy (isodose charts, output factor tables, wedge factors,…) - for emergencies tables, wedge factors,…) - for emergencies and computer break downsand computer break downs
Treatment planning computer (as above or Treatment planning computer (as above or beam model) - as standard planning databeam model) - as standard planning data
Independent checking device (eg. mu Independent checking device (eg. mu checks) - should be a completely checks) - should be a completely independent set of dataindependent set of data
Anwar 62
Machine data availabilityMachine data availability
Hardcopy (isodose charts, output factor Hardcopy (isodose charts, output factor tables, wedge factors,…)tables, wedge factors,…)
Treatment planning computer (as above Treatment planning computer (as above or beam model)or beam model)
Independent checking device (eg. mu Independent checking device (eg. mu checks)checks)
The data must be dated, verified in
regular intervals and the source
(including the person responsible
for it) must be documented
Anwar 63
Machine data summaryMachine data summary
Need to include all beams and options Need to include all beams and options (internal consistency(internal consistency اللزوجة اللزوجة القوام، , ,القوام،conventions, collisionconventions, collisionتصادمتصادم protection, protection, physical limitations) physical limitations)
Data can be made available for planning in Data can be made available for planning in installments as requiredinstallments as required
Some data may be required for individual Some data may be required for individual patients only (eg. special treatments)patients only (eg. special treatments)
Only make available data which is verifiedOnly make available data which is verified
Quick Question:Quick Question:
What data is available for physical What data is available for physical wedges in your center?wedges in your center?
This should include at least:Wedge angle - and how it is definedWedge output modification factor - and to which depth and field sizes it appliesThe field sizes for which the wedge can be usedBeam hardening? Maybe a new beam must be defined by TMRs or percentage depth doseProfiles in both directions (wedged and un-wedged - the latter is affected by divergence related profile changes)Weight (eg for OHS restrictions on lifting)
Anwar 65
From single to multiple beamsFrom single to multiple beams Mainly an issue Mainly an issue
for megavoltage for megavoltage photons where we photons where we have significant have significant contribution of contribution of dose to the target dose to the target from many beamsfrom many beams
11
44
33 2260 Gy60 Gy
Beam weighting must be factored in !!!
Anwar 66
CompensatorsCompensators Physical compensatorsPhysical compensators
lead sheetslead sheets brass blocksbrass blocks customized millingcustomized milling
Intensity modulationIntensity modulation multiple static fieldsmultiple static fields arcsarcs dynamic MLCdynamic MLC
Anwar 67
Intensity modulationIntensity modulation Can be shown to allow optimization of Can be shown to allow optimization of
the dose distributionthe dose distribution Make dose in the target homogenousMake dose in the target homogenous Minimize dose out of the targetMinimize dose out of the target Different techniquesDifferent techniques
physical compensatorsphysical compensators intensity modulation using multi leaf intensity modulation using multi leaf
collimatorscollimators
Anwar 68
Intensity Intensity ModulationModulation
Achieved using a Achieved using a Multi Leaf Collimator (MLC)Multi Leaf Collimator (MLC)
The field shape can be The field shape can be altered altered either step-by-step oreither step-by-step or dynamically while dose is dynamically while dose is
delivereddelivered
MLC pattern 1
MLC pattern 3
MLC pattern 2
Intensitymap
Anwar 69
iv. Basic dose calculationiv. Basic dose calculation
Once one has the target volume, the Once one has the target volume, the beam orientation and shape one has to beam orientation and shape one has to calculate how long a beam must be on calculate how long a beam must be on (60-Co or kV X-ray units) or how many (60-Co or kV X-ray units) or how many monitor units must be given (linear monitor units must be given (linear accelerator) to deliver the desired dose accelerator) to deliver the desired dose at the target.at the target.
Anwar 70
3D display of 3D display of beam beam placement placement may help to may help to identify the identify the structures in structures in the field.the field.
Anwar 71
Dose calculationDose calculation
Anwar 72
Dose display optionsDose display options
Isodose lines Color wash
Anwar 73
Isodose display - can be Isodose display - can be complex and 3Dcomplex and 3D
Anwar 74