radiaation protection in diagnostic and interventional imaging p18_qc_for_ct_web - copy

Upload: govardhanyu

Post on 04-Jun-2018

221 views

Category:

Documents


0 download

TRANSCRIPT

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    1/63

    IAEAInternational Atomic Energy Agency

    RADIATION PROTECTION INDIAGNOSTIC ANDINTERVENTIONAL RADIOLOGY

    Part 18: Optimization of protection in CT scannerPractical exercise - Quality Control

    IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    2/63

    IAEA

    18: Optimization of protection in CT scanner 2

    Contents / Overview

    Quality control test description on: CT accuracy, uniformity, linearity and noise, Low and high contrast resolution Z-axis sensitivity Alignment, Couch travel accuracy Gantry tilt measurement Dosimetry

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    3/63

    IAEAInternational Atomic Energy Agency

    Part 18: Optimization of protection in CTscanner

    Topic 1: Quality Control

    IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    4/63

    IAEA

    18: Optimization of protection in CT scanner 4

    Physical parameters of CT image

    Image quality May be expressed in terms of physical parameters such

    as uniformity, linearity, noise, spatial resolution, lowcontrast resolution

    It depends on the technological characteristics of the CTscanner, the exposure factors used and image viewingconditions.

    Quality may be assessed by quantitative measurementusing test phantoms, and by the appearance of artifacts.

    Measurements should be conducted regularly

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    5/63

    IAEA

    18: Optimization of protection in CT scanner 5

    Scanner performance: technicalparameters (I)

    Test Phantoms: Test phantoms of a standardized human shape

    or test objects of a particular shape, size and

    structure, are used for the purposes ofcalibration and evaluation of the performancesof CT scanners

    They should allow for the parameters to be

    checked: CT number; uniformity; noise; spatialresolution; slice thickness; dose; positioning ofcouch

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    6/63

    IAEA

    18: Optimization of protection in CT scanner 6

    Scanner performance: technicalparameters (II)CT Number Accuracy

    CT number depends on tube voltage, filtration, object thickness CT number of water is by definition equal to 0 Measured CT number should be < 4 HU in the central ROI

    CT Number Linearity

    It concerns the linear relationship between the calculated CTnumber and the linear attenuation coefficient of each element of theobject

    Deviations from linearity should be < 5 HUCT Number Uniformity

    It relates to the fact that a CT number of each pixel in the image ofan homogeneous object should be the same over various regions The difference in the CT number between a peripheral and a central

    region of an homogeneous test object should be < 8HU Differences are largely due to beam hardening phenomenon

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    7/63IAEA 18: Optimization of protection in CT scanner 7

    The noise Is the local statistical fluctuation (standard deviation) ofCT numbers of an homogeneous Region Of Interest(ROI)

    It strongly affects the low contrast resolution Noise is above all dependent on the radiation dose

    Image noise should be measured over an area of about10% of the cross-sectional area of the test object.

    The medical problem : to obtain an image with anacceptable level of noise while keeping the patient doseas low as reasonably achievable

    Scanner performance: technicalparameters (III)

    d o s e N o i s e 1 =

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    8/63

    IAEA 18: Optimization of protection in CT scanner 8

    Scanner performance: technicalparameters (IV)

    Spatial Resolution The high contrast resolution determines the minimum

    size of detail visualized in the plane of the slice with acontrast >10%.

    It is affected by: the reconstruction algorithm the detector width the effective slice thickness

    the object to detector distance the X-ray tube focal spot size the matrix size.

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    9/63

    IAEA 18: Optimization of protection in CT scanner 9

    Scanner performance: technicalparameters (V)

    Spatial Resolution The low contrast resolution determines the size

    of detail that can be visibly reproduced when

    there is only a small difference in density relativeto the surrounding area

    Low contrast resolution is considerably limited bynoise.

    The perception threshold in relation to contrast anddetail size can be determined, for example, by meansof a contrast-detail curve.

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    10/63

    IAEA 18: Optimization of protection in CT scanner 10

    Scanner performance: technicalparameters (VI)

    Slice Thickness The slice thickness is determined in

    the center of the field of view as thedistance between the two points onthe sensitivity profile along the axisof rotation at which response has

    fallen to 50%. The use of post-patient collimation

    to reduce the width of the imageslice leads to very significantincreases in the patient dose

    Positioning of couch The accuracy of positioning of the

    patient couch is evaluated bymoving the loaded couch a defineddistance and subsequently moving itback to the start position

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    11/63

    IAEA 18: Optimization of protection in CT scanner 11

    Minimum requirements: CT scanner(I)

    Image noiseThe Standard Deviation of CT numbers in the central 500mm2 ROI for a water or tissue equivalent phantom shouldnot deviate more than 20% from the baseline.

    CT number valuesThe deviation in the CT number values for water or tissueequivalent material and materials of different densitiesshould

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    12/63

    IAEA 18: Optimization of protection in CT scanner 12

    Minimum requirements: CT scanner(II)

    Computed tomography dose index (CTDI)The CTDI for a single slice for each available beam shaping filter andfor each available slice thickness should not deviate more than 20%from the baseline.

    Irradiated slice thickness

    The FWHM of the dose profile should not differ more than 20% frombaseline.

    High contrast resolution (spatial resolution)The FWHM of the point spread function of a pin, or the edge responsefunction of an edge should not differ more than 20% from baseline.

    Low contrast resolutionPolystyrene pins of 0.35 cm diameter inserted in a uniform body waterphantom should be visible in the image.

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    13/63

    IAEAInternational Atomic Energy Agency

    Part 18: Optimization of protection in CTscanner

    Topic 2: Noise

    IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    14/63

    IAEA 18: Optimization of protection in CT scanner 14

    Imaging performance (Noise)

    Noise is generally assessed using cylindricalphantoms, which are either filled with wateror made of a tissue equivalent material

    Once an axial image of the phantom hasbeen acquired, noise is obtained from thestandard deviation in CT number in a regionof interest (ROI) placed centrally within theimage

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    15/63

    IAEA 18: Optimization of protection in CT scanner 15

    Region ofinterest(ROI)

    Imaging performance (Noise)

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    16/63

    IAEA 18: Optimization of protection in CT scanner 16

    Imaging performance (Noise)

    Noise figures given in manufacturersspecifications are quoted for a specific phantom(e.g. manufacturers QA phantom) and forspecified scan parameters

    These conditions must be matched exactly for thepurposes of the acceptance test

    Manufacturers often quote noise at a particularsurface dose

    If this is the case, dose for axial scans can bemeasured by taping an ion chamber to the surfaceof the phantom

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    17/63

    IAEA 18: Optimization of protection in CT scanner 17

    Imaging performance (Noise)

    Baseline noise values should be obtained forseveral scan protocols that will be usedclinically , using the routine QC noise

    phantom

    To ensure that noise figures are bothaccurate and representative, it is essential tofind the mean value from several scans (10scans.)

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    18/63

    IAEAInternational Atomic Energy Agency

    Part 18: Optimization of protection in CTscanner

    Topic 3: CT uniformity and linearity

    IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    19/63

    IAEA 18: Optimization of protection in CT scanner 19

    CT number uniformity

    CT number uniformity can be assessed at thesame time as measuring noise, by placing fouradditional ROI (N, E, S and W) at positions nearthe edge of the image of a uniform phantom

    Mean CT number is then noted for these fourregions, as well as the central one

    The deviation from the central value should becalculated

    It can be valuable to check CT number uniformityfor large fields of view

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    20/63

    IAEA 18: Optimization of protection in CT scanner 20

    Axial imageof anhomogenousphantom

    CT number uniformity

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    21/63

    IAEA 18: Optimization of protection in CT scanner 21

    CT number uniformity

    CT number linearity is assessed using aphantom containing inserts of a number ofdifferent materials (materials should cover a

    wide range of CT numbers) One example of a suitable phantom to useat acceptance is the Catphan (The PhantomLaboratory, Salem, NY), which contains fourinserts with CT numbers ranging fromapproximately -1000HU to +1000HU

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    22/63

    IAEA 18: Optimization of protection in CT scanner 22

    CT number uniformity

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    23/63

    IAEAInternational Atomic Energy Agency

    Part 18: Optimization of protection in CTscanner

    Topic 4: Low and high contrast resolution

    IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    24/63

    IAEA 18: Optimization of protection in CT scanner 24

    Low contrast resolution

    Low contrast resolution (or low contrastdetectability) is often quoted in specificationdocumentation, as the smallest visible object at agiven contrast for a given dose

    Since this measurement relates directly to imagingperformance, it is an important parameter to verifyat acceptance

    At least 20 images of the low contrast insert (LCR)should be acquired and then viewed by at least 3observers under optimal viewing conditions, so asto obtain an average

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    25/63

    IAEA 18: Optimization of protection in CT scanner 25

    Typical imageof theCatphan LCRinsert

    Low contrast resolution

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    26/63

    IAEA 18: Optimization of protection in CT scanner 26

    Spatial resolution (high contrast)

    There exist two broad categories of measuringtechniques : those involving analysis of the point spread function,

    usually by calculation of the modulation transfer function(MTF)

    those involving either objective analysis or visualassessment of images of a resolution bar phantom.

    The resolution is quoted as the spatial frequency (inline pairs / cm) at which the modulation falls to 50%,

    10% or 2% MTF. These figures are often given for more than onereconstruction algorithm, e.g. for standard and high-resolution scans.

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    27/63

    IAEA 18: Optimization of protection in CT scanner 27

    Spatial resolution (high contrast)

    The number of linepairs per cm justvisible in the image isapproximatelyequivalent to the 2%value of the MTF

    This result can then becompared with the 2%MTF, if this is quoted inthe manufacturersspecification

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    28/63

    IAEAInternational Atomic Energy Agency

    Part 18: Optimization of protection in CTscanner

    Topic 5: Z-axis sensitivity

    IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    29/63

    IAEA 18: Optimization of protection in CT scanner 29

    Z-Sensitivity (Imaged slice width)

    Axial mode measurements Phantoms used for axial measurements may

    contain thin metal plates, wires or arrays of airholes, inclined at an angle to the image plane

    Manufacturers should be able to supply anappropriate phantom or, alternatively, the Catphancontains an insert suitable for this test

    Note: to obtain meaningful measurements, the

    thickness of the plates, wires or holes cannot begreater than the nominal slice width concerned.There may be problems for the sub-millimetre slicewidths offered on multi-slice scanners.

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    30/63

    IAEA 18: Optimization of protection in CT scanner 30

    Z-Sensitivity (Imaged slice width)

    Axial mode measurements Phantoms Manufacturers may quote the tolerance

    for each nominal slice width setting in theirspecification documentation

    Z-sensitivity measurements in axial mode can beused to check that imaged slice widths are withinthe tolerances given

    They can also be used in conjunction withirradiated slice width measurements to assess theaccuracy of post patient collimation and tocalculate the geometric efficiency for the scanner

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    31/63

    IAEA 18: Optimization of protection in CT scanner 31

    Plan view of a test object used to measure imagedslice widths for axial scans

    Z-Sensitivity (Imaged slice width)

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    32/63

    IAEAInternational Atomic Energy Agency

    Part 18: Optimization of protection in CTscanner

    Topic 6: Alignment

    IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

    Alignment of indicating lights with

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    33/63

    IAEA 18: Optimization of protection in CT scanner 33

    Alignment of indicating lights withscan, coronal and sagittal planes

    Several methods canbe used to performthese tests

    The techniquesdescribed here arestraightforward toimplement and require

    little specialist testequipment.

    Agreement between internal and

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    34/63

    IAEA 18: Optimization of protection in CT scanner 34

    Agreement between internal andexternal scan plane lights

    Use the envelope-wrapped film recommended forthat measurement, however, a piece of paper orcard can also be used

    The wrapped film is placed flat on the table andilluminated by the external scan plane light

    The position of the light is marked on the filmenvelope and the table is moved automatically tothe scan plane

    If the distance between the internal and externallights is correct, the internal light should nowcoincide with the mark on the film envelope.

    Co incidence of internal scan plane

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    35/63

    IAEA 18: Optimization of protection in CT scanner 35

    Co-incidence of internal scan planelights and scan plane

    Pin pricks are made in a piece of therapyverification film (or similar) along the line of theinternal scan plane light, and the film is exposed toa narrow axial scan and developed

    Coincidence between the pin pricks and the x-raybeam exposure indicates alignment between theinternal lights and the scan plane

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    36/63

    Co incidence of internal scan plane

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    37/63

    IAEA 18: Optimization of protection in CT scanner 37

    Z

    X

    Pin pricks made infilm at position ofscan plane light

    X-ray beamexposure

    Co-incidence of internal scan planelights and scan plane

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    38/63

    IAEA 18: Optimization of protection in CT scanner 38

    Coronal and Sagital plane lights

    A long, thin object, with a high CT number relativeto air, such as the lead in a pencil or astraightened paper clip, can be used as a markerto perform this test

    The marker is supported above the patient tableand aligned, using the indicating lights, so that it ispositioned at the isocentre, parallel to the z-axisand perpendicular to the scan plane

    If indicating lights are accurately aligned to thecoronal and sagital planes, the marker shouldappear as a dot at exactly x = 0, y = 0 on the axialimage.

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    39/63

    IAEAInternational Atomic Energy Agency

    Part 18: Optimization of protection in CTscanner

    Topic 7: Couch travel accuracy

    IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    40/63

    IAEA 18: Optimization of protection in CT scanner 40

    Couch travel accuracy

    A ruler or tape measure placed alongside thetable, can be used to check that the degree ofcouch movement indicated on the gantry agreeswith the actual distance moved.

    A load of approximately 70- 80 kg should beplaced on the table in order to simulate the weightof a patient.

    The test should be performed twice : by driving thetable top both away from and towards the gantry.

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    41/63

    IAEA 18: Optimization of protection in CT scanner 41

    Assessment of distance indicator accuracy

    Couch travel accuracy

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    42/63

    IAEA 18: Optimization of protection in CT scanner 42

    Axial scan incrementation accuracy

    Verification of incrementation accuracy betweensuccessive axial slices can be achieved by placingenvelope-wrapped film on the couch (in theisocentre plane) and exposing it to an axial scansequence

    Narrow slices separated by a couch incrementgreater than 1 slice width can be used, and thedistance between the lines on the film measured

    Couch travel accuracy for helical

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    43/63

    IAEA 18: Optimization of protection in CT scanner 43

    Couch travel accuracy for helicalscans

    In helical scanning, it is not sufficient to use a simplemechanical test because the distance imaged depends oncouch speed and scanner software

    One method of assessing imaged distance accuracy is touse a Perspex test object containing two small radio-opaque markers, separated by a fixed distance (ex:20 cm)

    The test object is scanned in Scan Projection Radiography(SPR) mode and a helical run is planned to start at the firstmarker and to end at a distance x from the first marker

    If couch travel is accurate during the helical scan, themarkers should be clearly seen on the first and final imagesof the series.

    Couch travel accuracy for helical

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    44/63

    IAEA 18: Optimization of protection in CT scanner 44

    Couch travel accuracy for helicalscans

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    45/63

    IAEAInternational Atomic Energy Agency

    Part 18: Optimization of protection in CTscanner

    Topic 8: Gantry tilt measurement

    IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    46/63

    l

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    47/63

    IAEA 18: Optimization of protection in CT scanner 47

    Gantry Tilt

    Three axial exposures are made using the samefilm: one for the maximum superior gantry tilt,

    one for the maximum inferior gantry tilt

    one at 0 gantry tilt The three scan planes should then be visible on

    the developed film

    The angles + and - between scan planes atmaximum tilt relative to that at 0 tilt should equaltilt angles displayed on the gantry.

    f f il

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    48/63

    IAEA 18: Optimization of protection in CT scanner 48

    Assessment of accuracy of gantry tilt

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    49/63

    IAEAInternational Atomic Energy Agency

    Part 18: Optimization of protection in CTscanner

    Topic 9: Dosimetry

    IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology

    D i CTDI i i

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    50/63

    IAEA 18: Optimization of protection in CT scanner 50

    Dosimetry - CTDI in air

    The Computed Tomography Dose Index(CTDI) in air can be measured using a 10cmpencil ionization chamber, bisected by the

    scan plane at the isocentre, supported fromthe patient table

    The ion chamber can be supported using aretort stand and clamp, if a dedicated holderis not available

    D i CTDI i i

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    51/63

    IAEA 18: Optimization of protection in CT scanner 51

    Ionizationchamber

    Table

    Dosimetry - CTDI in air

    D i CTDI i i

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    52/63

    IAEA 18: Optimization of protection in CT scanner 52

    Dosimetry - CTDI in air

    When commissioning a CT scanner, it isvaluable to measure the CTDI in air underthe following conditions:

    For all beam shaping filters For all nominal slice widths For all clinical kV settings

    For a range of scan times For a range of mA settings

    D i t CTDI i i

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    53/63

    IAEA 18: Optimization of protection in CT scanner 53

    Dosimetry - CTDI in air

    An additional test : comparison of dose from ahelical scan with that from an axial scan

    A helical CTDI value cannot be measured directly,as CTDI is a single slice measurement

    To obtain a relative dose for helical scanning, theentire length of the chamber can be scanned firstly in axial mode (contiguous slices) and secondly inhelical mode, at pitch 1

    The ratio of the 2 doses provides a correctionfactor, which can then be used to convert axialCTDIs into helical CTDIs

    D i t CTDI i i (h li l)

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    54/63

    IAEA 18: Optimization of protection in CT scanner 54

    Axial slice positions

    Helical scan (pitch 1)

    Dosimetry - CTDI in air (helical)

    Dosimetry - CTDI in Perspex

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    55/63

    IAEA 18: Optimization of protection in CT scanner 55

    Dosimetry CTDI in PerspexPhantoms

    The body phantom placed on the patient table andthe head phantom is supported in the head rest

    Phantoms are aligned centred at the scanisocentre

    The ion chamber is inserted into either the centralor one of the peripheral cavities of the phantom (allother cavities being filled with Perspex rods)

    Dosimetry - CTDI in Perspex

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    56/63

    IAEA 18: Optimization of protection in CT scanner 56

    Dosimetry CTDI in PerspexPhantoms

    Dose measurements at the center are used tocalculate the central CTDI

    Peripheral CTDI is measured in at least four

    positions around the phantom, so as to achieve atrue averageNote: if gantry rotation is initiated from different

    angular positions for successive scans it may be

    necessary to take a number of measurements ateach position in order to get a representative meandose

    Dosimetry - CTDI in Perspex

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    57/63

    IAEA 18: Optimization of protection in CT scanner 57

    Head phantom Body phantom(or annulus

    to fit over head phantom)

    Insert to plug holes

    Dos et y C D e spePhantoms

    Dosimetry - CTDI in Perspex

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    58/63

    IAEA 18: Optimization of protection in CT scanner 58

    Central and peripheral CTDIs are used tocalculate weighted CTDI, CTDIw:

    CTDIws can be compared against diagnosticreference levels for standard scanexaminations

    y pPhantoms

    ) ( CTDI 3 2 + CTDI 3 1 C 1 = CTDI p 100, c 100, w n

    Example of CTDI Values for some CT

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    59/63

    IAEA 18: Optimization of protection in CT scanner 59

    punits

    MODEL KV FAD S nCDTI (mGy/mAs)

    Air weightedHead BodySiemensHIQ

    133 700 10 0.195 0.161 (0.83) 0.093 (0.48)

    Siemens

    Plus S

    120 700 10 0.128 0.110 (0.86) 0.062 (0.48)

    SiemensPlus S

    137 700 8 0.161 0.082 (0.51)

    GE Pace 120 525 10 0.344 0.200 (0.58) 0.094 (0.27)

    PhilipsLX

    120 606 10 0.200 0.160 (0.80) 0.081 (0.41)

    FAD(mm): Focus to Axis distanceS (mm): Slice thickness

    Dose Profiles (Irradiated slice width)

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    60/63

    IAEA 18: Optimization of protection in CT scanner 60

    Dose Profiles (Irradiated slice width)

    Measurement of irradiated slice widths, for allnominal slice width settings, provides a direct testof pre-patient beam collimation functionality andallows geometric efficiencies to be calculated for

    the scanner Geometric efficiency (GE) is defined as:

    GE = width slice irradiated x 100 %width slice imaged

    It is recommended that GE value is displayed onthe console if it is less than 70%

    Dose Profiles (Irradiated slice width)

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    61/63

    IAEA 18: Optimization of protection in CT scanner 61

    Dose Profiles (Irradiated slice width)

    First method : expose envelope-wrapped x-ray film,supported in air at the isocentre, at each of the slice widthsettings

    Once developed, optical density profiles may be plotted

    using a scanning microdensitometer or the width measuredusing a ruler

    To provide an accurate dose profile, a calibration curve canbe applied to convert optical density profiles into dose

    profiles, from which irradiated slice widths (FWHM of doseprofiles) may be derived Alternative technique: measure dose profiles using TLDs.

    Dose Profiles (Irradiated slice width)

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    62/63

    IAEA 18: Optimization of protection in CT scanner 62

    Measurement ofirradiated slicewidths for a range

    of nominal slicewidth settings

    Dose Profiles (Irradiated slice width)

    Quality control program

  • 8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy

    63/63

    Quality control program

    Example of a quality control program for a CT unit:

    Baseline 20%H1 yBHigh contrast resolution

    > 20% or 1 mmM1 yBImage slice thickness

    > 20% or 1 mmH1 yBIrradiated slice thickness

    Baseline 20%H1 y ACTDI

    > 1.5%H1 yBCT number uniformity

    Water: 10 HUOther materials 20 HU*more scans/parameters

    H

    H

    D/W

    1 Y*

    A

    B

    CT number values

    Baseline 20%*more scans/parameters

    HH

    D/W1 y*

    AB

    Image noise

    Remedial levelPriorityFrequencyLevel ofexpertise

    Physical parameter