radiaation protection in diagnostic and interventional imaging p18_qc_for_ct_web - copy
TRANSCRIPT
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
1/63
IAEAInternational Atomic Energy Agency
RADIATION PROTECTION INDIAGNOSTIC ANDINTERVENTIONAL RADIOLOGY
Part 18: Optimization of protection in CT scannerPractical exercise - Quality Control
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
2/63
IAEA
18: Optimization of protection in CT scanner 2
Contents / Overview
Quality control test description on: CT accuracy, uniformity, linearity and noise, Low and high contrast resolution Z-axis sensitivity Alignment, Couch travel accuracy Gantry tilt measurement Dosimetry
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
3/63
IAEAInternational Atomic Energy Agency
Part 18: Optimization of protection in CTscanner
Topic 1: Quality Control
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
4/63
IAEA
18: Optimization of protection in CT scanner 4
Physical parameters of CT image
Image quality May be expressed in terms of physical parameters such
as uniformity, linearity, noise, spatial resolution, lowcontrast resolution
It depends on the technological characteristics of the CTscanner, the exposure factors used and image viewingconditions.
Quality may be assessed by quantitative measurementusing test phantoms, and by the appearance of artifacts.
Measurements should be conducted regularly
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
5/63
IAEA
18: Optimization of protection in CT scanner 5
Scanner performance: technicalparameters (I)
Test Phantoms: Test phantoms of a standardized human shape
or test objects of a particular shape, size and
structure, are used for the purposes ofcalibration and evaluation of the performancesof CT scanners
They should allow for the parameters to be
checked: CT number; uniformity; noise; spatialresolution; slice thickness; dose; positioning ofcouch
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
6/63
IAEA
18: Optimization of protection in CT scanner 6
Scanner performance: technicalparameters (II)CT Number Accuracy
CT number depends on tube voltage, filtration, object thickness CT number of water is by definition equal to 0 Measured CT number should be < 4 HU in the central ROI
CT Number Linearity
It concerns the linear relationship between the calculated CTnumber and the linear attenuation coefficient of each element of theobject
Deviations from linearity should be < 5 HUCT Number Uniformity
It relates to the fact that a CT number of each pixel in the image ofan homogeneous object should be the same over various regions The difference in the CT number between a peripheral and a central
region of an homogeneous test object should be < 8HU Differences are largely due to beam hardening phenomenon
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
7/63IAEA 18: Optimization of protection in CT scanner 7
The noise Is the local statistical fluctuation (standard deviation) ofCT numbers of an homogeneous Region Of Interest(ROI)
It strongly affects the low contrast resolution Noise is above all dependent on the radiation dose
Image noise should be measured over an area of about10% of the cross-sectional area of the test object.
The medical problem : to obtain an image with anacceptable level of noise while keeping the patient doseas low as reasonably achievable
Scanner performance: technicalparameters (III)
d o s e N o i s e 1 =
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
8/63
IAEA 18: Optimization of protection in CT scanner 8
Scanner performance: technicalparameters (IV)
Spatial Resolution The high contrast resolution determines the minimum
size of detail visualized in the plane of the slice with acontrast >10%.
It is affected by: the reconstruction algorithm the detector width the effective slice thickness
the object to detector distance the X-ray tube focal spot size the matrix size.
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
9/63
IAEA 18: Optimization of protection in CT scanner 9
Scanner performance: technicalparameters (V)
Spatial Resolution The low contrast resolution determines the size
of detail that can be visibly reproduced when
there is only a small difference in density relativeto the surrounding area
Low contrast resolution is considerably limited bynoise.
The perception threshold in relation to contrast anddetail size can be determined, for example, by meansof a contrast-detail curve.
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
10/63
IAEA 18: Optimization of protection in CT scanner 10
Scanner performance: technicalparameters (VI)
Slice Thickness The slice thickness is determined in
the center of the field of view as thedistance between the two points onthe sensitivity profile along the axisof rotation at which response has
fallen to 50%. The use of post-patient collimation
to reduce the width of the imageslice leads to very significantincreases in the patient dose
Positioning of couch The accuracy of positioning of the
patient couch is evaluated bymoving the loaded couch a defineddistance and subsequently moving itback to the start position
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
11/63
IAEA 18: Optimization of protection in CT scanner 11
Minimum requirements: CT scanner(I)
Image noiseThe Standard Deviation of CT numbers in the central 500mm2 ROI for a water or tissue equivalent phantom shouldnot deviate more than 20% from the baseline.
CT number valuesThe deviation in the CT number values for water or tissueequivalent material and materials of different densitiesshould
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
12/63
IAEA 18: Optimization of protection in CT scanner 12
Minimum requirements: CT scanner(II)
Computed tomography dose index (CTDI)The CTDI for a single slice for each available beam shaping filter andfor each available slice thickness should not deviate more than 20%from the baseline.
Irradiated slice thickness
The FWHM of the dose profile should not differ more than 20% frombaseline.
High contrast resolution (spatial resolution)The FWHM of the point spread function of a pin, or the edge responsefunction of an edge should not differ more than 20% from baseline.
Low contrast resolutionPolystyrene pins of 0.35 cm diameter inserted in a uniform body waterphantom should be visible in the image.
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
13/63
IAEAInternational Atomic Energy Agency
Part 18: Optimization of protection in CTscanner
Topic 2: Noise
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
14/63
IAEA 18: Optimization of protection in CT scanner 14
Imaging performance (Noise)
Noise is generally assessed using cylindricalphantoms, which are either filled with wateror made of a tissue equivalent material
Once an axial image of the phantom hasbeen acquired, noise is obtained from thestandard deviation in CT number in a regionof interest (ROI) placed centrally within theimage
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
15/63
IAEA 18: Optimization of protection in CT scanner 15
Region ofinterest(ROI)
Imaging performance (Noise)
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
16/63
IAEA 18: Optimization of protection in CT scanner 16
Imaging performance (Noise)
Noise figures given in manufacturersspecifications are quoted for a specific phantom(e.g. manufacturers QA phantom) and forspecified scan parameters
These conditions must be matched exactly for thepurposes of the acceptance test
Manufacturers often quote noise at a particularsurface dose
If this is the case, dose for axial scans can bemeasured by taping an ion chamber to the surfaceof the phantom
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
17/63
IAEA 18: Optimization of protection in CT scanner 17
Imaging performance (Noise)
Baseline noise values should be obtained forseveral scan protocols that will be usedclinically , using the routine QC noise
phantom
To ensure that noise figures are bothaccurate and representative, it is essential tofind the mean value from several scans (10scans.)
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
18/63
IAEAInternational Atomic Energy Agency
Part 18: Optimization of protection in CTscanner
Topic 3: CT uniformity and linearity
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
19/63
IAEA 18: Optimization of protection in CT scanner 19
CT number uniformity
CT number uniformity can be assessed at thesame time as measuring noise, by placing fouradditional ROI (N, E, S and W) at positions nearthe edge of the image of a uniform phantom
Mean CT number is then noted for these fourregions, as well as the central one
The deviation from the central value should becalculated
It can be valuable to check CT number uniformityfor large fields of view
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
20/63
IAEA 18: Optimization of protection in CT scanner 20
Axial imageof anhomogenousphantom
CT number uniformity
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
21/63
IAEA 18: Optimization of protection in CT scanner 21
CT number uniformity
CT number linearity is assessed using aphantom containing inserts of a number ofdifferent materials (materials should cover a
wide range of CT numbers) One example of a suitable phantom to useat acceptance is the Catphan (The PhantomLaboratory, Salem, NY), which contains fourinserts with CT numbers ranging fromapproximately -1000HU to +1000HU
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
22/63
IAEA 18: Optimization of protection in CT scanner 22
CT number uniformity
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
23/63
IAEAInternational Atomic Energy Agency
Part 18: Optimization of protection in CTscanner
Topic 4: Low and high contrast resolution
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
24/63
IAEA 18: Optimization of protection in CT scanner 24
Low contrast resolution
Low contrast resolution (or low contrastdetectability) is often quoted in specificationdocumentation, as the smallest visible object at agiven contrast for a given dose
Since this measurement relates directly to imagingperformance, it is an important parameter to verifyat acceptance
At least 20 images of the low contrast insert (LCR)should be acquired and then viewed by at least 3observers under optimal viewing conditions, so asto obtain an average
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
25/63
IAEA 18: Optimization of protection in CT scanner 25
Typical imageof theCatphan LCRinsert
Low contrast resolution
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
26/63
IAEA 18: Optimization of protection in CT scanner 26
Spatial resolution (high contrast)
There exist two broad categories of measuringtechniques : those involving analysis of the point spread function,
usually by calculation of the modulation transfer function(MTF)
those involving either objective analysis or visualassessment of images of a resolution bar phantom.
The resolution is quoted as the spatial frequency (inline pairs / cm) at which the modulation falls to 50%,
10% or 2% MTF. These figures are often given for more than onereconstruction algorithm, e.g. for standard and high-resolution scans.
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
27/63
IAEA 18: Optimization of protection in CT scanner 27
Spatial resolution (high contrast)
The number of linepairs per cm justvisible in the image isapproximatelyequivalent to the 2%value of the MTF
This result can then becompared with the 2%MTF, if this is quoted inthe manufacturersspecification
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
28/63
IAEAInternational Atomic Energy Agency
Part 18: Optimization of protection in CTscanner
Topic 5: Z-axis sensitivity
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
29/63
IAEA 18: Optimization of protection in CT scanner 29
Z-Sensitivity (Imaged slice width)
Axial mode measurements Phantoms used for axial measurements may
contain thin metal plates, wires or arrays of airholes, inclined at an angle to the image plane
Manufacturers should be able to supply anappropriate phantom or, alternatively, the Catphancontains an insert suitable for this test
Note: to obtain meaningful measurements, the
thickness of the plates, wires or holes cannot begreater than the nominal slice width concerned.There may be problems for the sub-millimetre slicewidths offered on multi-slice scanners.
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
30/63
IAEA 18: Optimization of protection in CT scanner 30
Z-Sensitivity (Imaged slice width)
Axial mode measurements Phantoms Manufacturers may quote the tolerance
for each nominal slice width setting in theirspecification documentation
Z-sensitivity measurements in axial mode can beused to check that imaged slice widths are withinthe tolerances given
They can also be used in conjunction withirradiated slice width measurements to assess theaccuracy of post patient collimation and tocalculate the geometric efficiency for the scanner
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
31/63
IAEA 18: Optimization of protection in CT scanner 31
Plan view of a test object used to measure imagedslice widths for axial scans
Z-Sensitivity (Imaged slice width)
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
32/63
IAEAInternational Atomic Energy Agency
Part 18: Optimization of protection in CTscanner
Topic 6: Alignment
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
Alignment of indicating lights with
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
33/63
IAEA 18: Optimization of protection in CT scanner 33
Alignment of indicating lights withscan, coronal and sagittal planes
Several methods canbe used to performthese tests
The techniquesdescribed here arestraightforward toimplement and require
little specialist testequipment.
Agreement between internal and
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
34/63
IAEA 18: Optimization of protection in CT scanner 34
Agreement between internal andexternal scan plane lights
Use the envelope-wrapped film recommended forthat measurement, however, a piece of paper orcard can also be used
The wrapped film is placed flat on the table andilluminated by the external scan plane light
The position of the light is marked on the filmenvelope and the table is moved automatically tothe scan plane
If the distance between the internal and externallights is correct, the internal light should nowcoincide with the mark on the film envelope.
Co incidence of internal scan plane
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
35/63
IAEA 18: Optimization of protection in CT scanner 35
Co-incidence of internal scan planelights and scan plane
Pin pricks are made in a piece of therapyverification film (or similar) along the line of theinternal scan plane light, and the film is exposed toa narrow axial scan and developed
Coincidence between the pin pricks and the x-raybeam exposure indicates alignment between theinternal lights and the scan plane
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
36/63
Co incidence of internal scan plane
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
37/63
IAEA 18: Optimization of protection in CT scanner 37
Z
X
Pin pricks made infilm at position ofscan plane light
X-ray beamexposure
Co-incidence of internal scan planelights and scan plane
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
38/63
IAEA 18: Optimization of protection in CT scanner 38
Coronal and Sagital plane lights
A long, thin object, with a high CT number relativeto air, such as the lead in a pencil or astraightened paper clip, can be used as a markerto perform this test
The marker is supported above the patient tableand aligned, using the indicating lights, so that it ispositioned at the isocentre, parallel to the z-axisand perpendicular to the scan plane
If indicating lights are accurately aligned to thecoronal and sagital planes, the marker shouldappear as a dot at exactly x = 0, y = 0 on the axialimage.
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
39/63
IAEAInternational Atomic Energy Agency
Part 18: Optimization of protection in CTscanner
Topic 7: Couch travel accuracy
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
40/63
IAEA 18: Optimization of protection in CT scanner 40
Couch travel accuracy
A ruler or tape measure placed alongside thetable, can be used to check that the degree ofcouch movement indicated on the gantry agreeswith the actual distance moved.
A load of approximately 70- 80 kg should beplaced on the table in order to simulate the weightof a patient.
The test should be performed twice : by driving thetable top both away from and towards the gantry.
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
41/63
IAEA 18: Optimization of protection in CT scanner 41
Assessment of distance indicator accuracy
Couch travel accuracy
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
42/63
IAEA 18: Optimization of protection in CT scanner 42
Axial scan incrementation accuracy
Verification of incrementation accuracy betweensuccessive axial slices can be achieved by placingenvelope-wrapped film on the couch (in theisocentre plane) and exposing it to an axial scansequence
Narrow slices separated by a couch incrementgreater than 1 slice width can be used, and thedistance between the lines on the film measured
Couch travel accuracy for helical
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
43/63
IAEA 18: Optimization of protection in CT scanner 43
Couch travel accuracy for helicalscans
In helical scanning, it is not sufficient to use a simplemechanical test because the distance imaged depends oncouch speed and scanner software
One method of assessing imaged distance accuracy is touse a Perspex test object containing two small radio-opaque markers, separated by a fixed distance (ex:20 cm)
The test object is scanned in Scan Projection Radiography(SPR) mode and a helical run is planned to start at the firstmarker and to end at a distance x from the first marker
If couch travel is accurate during the helical scan, themarkers should be clearly seen on the first and final imagesof the series.
Couch travel accuracy for helical
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
44/63
IAEA 18: Optimization of protection in CT scanner 44
Couch travel accuracy for helicalscans
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
45/63
IAEAInternational Atomic Energy Agency
Part 18: Optimization of protection in CTscanner
Topic 8: Gantry tilt measurement
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
46/63
l
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
47/63
IAEA 18: Optimization of protection in CT scanner 47
Gantry Tilt
Three axial exposures are made using the samefilm: one for the maximum superior gantry tilt,
one for the maximum inferior gantry tilt
one at 0 gantry tilt The three scan planes should then be visible on
the developed film
The angles + and - between scan planes atmaximum tilt relative to that at 0 tilt should equaltilt angles displayed on the gantry.
f f il
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
48/63
IAEA 18: Optimization of protection in CT scanner 48
Assessment of accuracy of gantry tilt
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
49/63
IAEAInternational Atomic Energy Agency
Part 18: Optimization of protection in CTscanner
Topic 9: Dosimetry
IAEA Training Material on Radiation Protection in Diagnostic and Interventional Radiology
D i CTDI i i
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
50/63
IAEA 18: Optimization of protection in CT scanner 50
Dosimetry - CTDI in air
The Computed Tomography Dose Index(CTDI) in air can be measured using a 10cmpencil ionization chamber, bisected by the
scan plane at the isocentre, supported fromthe patient table
The ion chamber can be supported using aretort stand and clamp, if a dedicated holderis not available
D i CTDI i i
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
51/63
IAEA 18: Optimization of protection in CT scanner 51
Ionizationchamber
Table
Dosimetry - CTDI in air
D i CTDI i i
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
52/63
IAEA 18: Optimization of protection in CT scanner 52
Dosimetry - CTDI in air
When commissioning a CT scanner, it isvaluable to measure the CTDI in air underthe following conditions:
For all beam shaping filters For all nominal slice widths For all clinical kV settings
For a range of scan times For a range of mA settings
D i t CTDI i i
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
53/63
IAEA 18: Optimization of protection in CT scanner 53
Dosimetry - CTDI in air
An additional test : comparison of dose from ahelical scan with that from an axial scan
A helical CTDI value cannot be measured directly,as CTDI is a single slice measurement
To obtain a relative dose for helical scanning, theentire length of the chamber can be scanned firstly in axial mode (contiguous slices) and secondly inhelical mode, at pitch 1
The ratio of the 2 doses provides a correctionfactor, which can then be used to convert axialCTDIs into helical CTDIs
D i t CTDI i i (h li l)
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
54/63
IAEA 18: Optimization of protection in CT scanner 54
Axial slice positions
Helical scan (pitch 1)
Dosimetry - CTDI in air (helical)
Dosimetry - CTDI in Perspex
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
55/63
IAEA 18: Optimization of protection in CT scanner 55
Dosimetry CTDI in PerspexPhantoms
The body phantom placed on the patient table andthe head phantom is supported in the head rest
Phantoms are aligned centred at the scanisocentre
The ion chamber is inserted into either the centralor one of the peripheral cavities of the phantom (allother cavities being filled with Perspex rods)
Dosimetry - CTDI in Perspex
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
56/63
IAEA 18: Optimization of protection in CT scanner 56
Dosimetry CTDI in PerspexPhantoms
Dose measurements at the center are used tocalculate the central CTDI
Peripheral CTDI is measured in at least four
positions around the phantom, so as to achieve atrue averageNote: if gantry rotation is initiated from different
angular positions for successive scans it may be
necessary to take a number of measurements ateach position in order to get a representative meandose
Dosimetry - CTDI in Perspex
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
57/63
IAEA 18: Optimization of protection in CT scanner 57
Head phantom Body phantom(or annulus
to fit over head phantom)
Insert to plug holes
Dos et y C D e spePhantoms
Dosimetry - CTDI in Perspex
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
58/63
IAEA 18: Optimization of protection in CT scanner 58
Central and peripheral CTDIs are used tocalculate weighted CTDI, CTDIw:
CTDIws can be compared against diagnosticreference levels for standard scanexaminations
y pPhantoms
) ( CTDI 3 2 + CTDI 3 1 C 1 = CTDI p 100, c 100, w n
Example of CTDI Values for some CT
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
59/63
IAEA 18: Optimization of protection in CT scanner 59
punits
MODEL KV FAD S nCDTI (mGy/mAs)
Air weightedHead BodySiemensHIQ
133 700 10 0.195 0.161 (0.83) 0.093 (0.48)
Siemens
Plus S
120 700 10 0.128 0.110 (0.86) 0.062 (0.48)
SiemensPlus S
137 700 8 0.161 0.082 (0.51)
GE Pace 120 525 10 0.344 0.200 (0.58) 0.094 (0.27)
PhilipsLX
120 606 10 0.200 0.160 (0.80) 0.081 (0.41)
FAD(mm): Focus to Axis distanceS (mm): Slice thickness
Dose Profiles (Irradiated slice width)
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
60/63
IAEA 18: Optimization of protection in CT scanner 60
Dose Profiles (Irradiated slice width)
Measurement of irradiated slice widths, for allnominal slice width settings, provides a direct testof pre-patient beam collimation functionality andallows geometric efficiencies to be calculated for
the scanner Geometric efficiency (GE) is defined as:
GE = width slice irradiated x 100 %width slice imaged
It is recommended that GE value is displayed onthe console if it is less than 70%
Dose Profiles (Irradiated slice width)
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
61/63
IAEA 18: Optimization of protection in CT scanner 61
Dose Profiles (Irradiated slice width)
First method : expose envelope-wrapped x-ray film,supported in air at the isocentre, at each of the slice widthsettings
Once developed, optical density profiles may be plotted
using a scanning microdensitometer or the width measuredusing a ruler
To provide an accurate dose profile, a calibration curve canbe applied to convert optical density profiles into dose
profiles, from which irradiated slice widths (FWHM of doseprofiles) may be derived Alternative technique: measure dose profiles using TLDs.
Dose Profiles (Irradiated slice width)
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
62/63
IAEA 18: Optimization of protection in CT scanner 62
Measurement ofirradiated slicewidths for a range
of nominal slicewidth settings
Dose Profiles (Irradiated slice width)
Quality control program
-
8/13/2019 Radiaation Protection in Diagnostic and Interventional Imaging P18_QC_for_CT_WEB - Copy
63/63
Quality control program
Example of a quality control program for a CT unit:
Baseline 20%H1 yBHigh contrast resolution
> 20% or 1 mmM1 yBImage slice thickness
> 20% or 1 mmH1 yBIrradiated slice thickness
Baseline 20%H1 y ACTDI
> 1.5%H1 yBCT number uniformity
Water: 10 HUOther materials 20 HU*more scans/parameters
H
H
D/W
1 Y*
A
B
CT number values
Baseline 20%*more scans/parameters
HH
D/W1 y*
AB
Image noise
Remedial levelPriorityFrequencyLevel ofexpertise
Physical parameter