r. l. sukovaty, s. k. roby, c. e. sheldon, a. cunningham celerion

1
The Development and Validation of Specific Ligand Binding Methods for the Determination of Parathyroid Hormone (1-34) and (1-84) in Human Plasma for Pharmacokinetic Sample Analysis R. L. Sukovaty, S. K. Roby, C. E. Sheldon, A. Cunningham Celerion, Lincoln, NE was less than 9.7% C.V. Inter-batch accuracy (% Bias) of the same quality controls samples was between -8.3 and +12.2% (Table 2). Sample collection and handling stability was established in whole blood for 2 hours at ambient temperature under UV-shielded light for PTH (1-84) and under white light for PTH (1-34) (Tables 3 and 4). Short-term stability in plasma was established for 24 hours for PTH (1-84) and 13 hours for PTH (1-34) at ambient temperature under white light. Cumulative short-term stability for PTH (1-84) was established for 39 hours at ambient temperature under white light. Freeze and Thaw stability was established for six freeze (-80°C for PTH (1-84) and -20°C for PTH (1-34)) and thaw (ambient tempera- ture) cycles for both PTH (1-84) and PTH (1-34). Long term stability of matrix samples was established for 98 days for PTH (1-84) at -80°C, 10 days for PTH (1-84) at -20°C, and for 259 days for PTH (1-34) at -20°C and -80°C. The quantitative integrity of PTH (1-84) and PTH (1-34) samples prepared in multiple lots of non-hemolyzed, non-turbid matrix was verified. The quantitative integrity of PTH (1-84) samples prepared in three lots of hemolyzed matrix and in three lots of turbid matrix was verified. Hemolyzed and turbid matrix did have an effect on the measurement of prepared PTH (1-34) samples. An evaluation of dilution integrity demonstrated that a dilution factor of 200 can be applied to PTH (1-84) samples and a dilution factor of 150 can be applied to PTH (1-34) samples to dilute them into the quantifiable range without impacting linearity. The absence of a high dose hook effect was demonstrated for both PTH (1-84) and PTH (1-34) by assaying a sample with a con- centration higher than the ULOQ at a minimum of 3 dilution levels above the ULOQ. All samples back-calculated with concentrations greater than the ULOQ for the respective assays. CONCLUSIONS: The validated methods allow for rapid, selective, accurate and reproducible quantitation of PTH (1-84) and PTH (1-34) in human plasma samples for pharmacokinetic evaluation. PURPOSE: Parathyroid hormone (PTH) is an 84 amino acid polypeptide produced in the parathyroid gland and plays a major role in regulating serum calcium concentrations. PTH (1-34) is a polypeptide consisting of the 34 N-terminal amino acids of the intact PTH (1-84) which shows similar effects on serum calcium regulation. Methods for determining PTH (1-84) in human plasma need to account for endogenous concentrations of the intact peptide when preparing controls and calibrators in plasma. Methods for determining PTH (1-34) in human plasma also need to account for antibody cross-reactivity with endogenous PTH (1-84). Commercially available PTH (1-84) kits and PTH (1-34) kits were adapted and validated to meet FDA guidelines for validation of bioanalytical methods. METHODS: PTH (1-84): Unknown samples and calibration samples prepared in human plasma (treated with an antibody against the C-terminal portion of PTH (1-84) attached to a solid substrate to remove endogenous PTH (1-84)) were added to test tubes containing a fixed antibody (C-terminal specific) and a radioactively labeled antibody (N-terminal specific). After washing, the tubes were counted in a gamma counter. The counts detected were direct- ly proportional to the concentration of PTH (1-84) present in the sample (Figure 1). PTH (1-34): Unknown samples and calibration samples in human plasma were pre-treated with an antibody against the C-terminal portion of PTH (1-84) attached to a solid substrate to remove PTH (1-84). The pre-treated samples and calibration samples were pipetted into microtiter plates coated with the appropriate capture antibodies. The wells were washed to remove unbound material and enzyme-labeled antibody was added. Unbound labeled anti- body was removed and a chromogenic substrate was added to the bound labeled antibody. The development of the colored reaction product was directly proportional to the concentration of PTH (1-34) present in the sample and was detected using a spectrophotometric plate reader (Figure 2). RESULTS: The PTH (1-84) method uses a 4-parameter logistic regression weighted 1/y over the range of 10.0 – 1500 pg/mL. The PTH (1-34) method uses a 4-parameter logistic regression weighted 1/y 2 over the range of 6.00 – 150 pg/mL. The concentrations of PTH (1-84) and PTH (1-34) calibration standards were back-calculated from the regression equation of the experimental data. The coefficients of variation (C.V.) were less than or equal to 4.4% and 8.5%, respectively. Inter-batch precision of PTH (1-84) quality control samples between 10.0 and 2000 pg/mL was less than 11.6% C.V. Inter- batch accuracy (% Bias) of the same quality control samples was between -1.3 and +8.8% (Table 1). Inter-batch precision (%CV) of PTH (1-34) quality control samples between 6.00 and 150 pg/mL Figure 1. PTH (1-84) Example Standard Curve Instrument Response (CPM) Nominal Conc. Endogenous Endogenous + Low Spike Endogenous + High Spike PTH (1-84) 0 minutes 120 minutes 0 minutes 120 minutes 0 minutes 120 minutes 28.0 30.2 51.7 49.9 938 949 26.6 29.3 50.6 53.5 983 955 27.3 29.3 49.7 52.0 909 947 26.5 29.4 49.9 49.7 921 887 28.0 29.3 46.5 51.2 923 943 27.2 26.5 48.9 50.5 943 885 Mean 27.3 29.0 49.6 51.1 936 928 % CV 2.4 4.4 3.6 2.8 2.8 3.5 % of Control 106.2 103.0 99.1 n 6 6 6 6 6 6 Table 3: Stability of PTH (1-84) During Sample Collection and Handling from Human Whole Blood (EDTA) at Ambient Temperature Under UV Shielded Light Conditions Low Spike High Spike PTH (1-34) 0 minutes 120 minutes 0 minutes 120 minutes 33.3 46.9 219 212 33.7 31.8 219 211 34.7 25.9 222 215 34.5 27.5 218 216 34.8 25.1 223 214 34.5 24.2 221 214 Mean 34.3 30.2 220 214 % CV 1.8 28.5 0.9 0.9 % of Control 88.0 97.3 n 6 6 6 6 Table 4: Stability of PTH (1-34) During Sample Collection and Handling from Human Whole Blood (EDTA) at Ambient Temperature Under White Light Conditions Table 1. PTH (1-84) Inter-Batch Precision and Accuracy LLOQ ENDO QC A QC B QC C ULOQ PTH (1-84) 10.0 pg/mL 11.7 pg/mL 29.7 pg/mL 160 pg/mL 1510 pg/mL 2000 pg/mL Inter-Batch Mean 10.6 11.7 32.3 163 1490 1990 Inter-Batch SD 0.886 1.36 2.02 4.41 71.5 84.9 Inter-Batch % CV 8.4 11.6 6.3 2.7 4.8 4.3 Inter-Batch % Bias 6.0 0.0 8.8 1.9 -1.3 -0.5 n 18 18 18 18 18 18 Table 2: PTH (1-34) Inter-Batch Precision and Accuracy LLOQ QC QC A QC B QC C ULOQ QC PTH (1-34) 6.00 pg/mL 18.0 pg/mL 75.0 pg/mL 120 pg/mL 150 pg/mL Inter-Batch Mean 6.73 16.5 77.6 125 162 Inter-Batch SD 0.651 0.774 2.26 4.17 8.04 Inter-Batch % CV 9.7 4.7 2.9 3.3 5.0 Inter-Batch % Bias 12.2 -8.3 3.5 4.2 8.0 n 17 18 18 21 18 Table 5. PTH (1-84) Validation Summary Information Requested Data Analyte PTH (1-84) Method Description Direct analysis using immunoradiometric assay (IRMA) Limit of Quantitation (pg/mL) 10.0 pg/mL Standard Curve Concentrations (pg/mL) 4.00 (anchor point), 10.0, 25.0, 50.0, 100, 200, 400, 800, 1600, and 2000 pg/mL QC Concentrations (pg/mL) LLOQ QC, endogenous (11.7 pg/mL), 29.7, 160, 1510, and ULOQ QC pg/mL QC Intra-Batch Precision Range (% CV) 1.2 to 18.0% QC Intra-Batch Accuracy Range (% Bias) -12.0 to 16.0% QC Inter-Batch Precision Range (% CV) 2.7 to 11.6% QC Inter-Batch Accuracy Range (% Bias) -1.3 to 8.8% Bench-Top Stability (Hrs) Short-Term Stability: 24 hours in polypropylene tubes at ambient temperature under white light Cumulative Short-Term Stability: 39 hours in polypropylene tubes at ambient temperature under white light (total of all thaw cycles) Stock Stability (Days) Long-Term Stability for Stock Solutions (Stock): 132 days at approximately 100 µg/mL in phosphate-buffered saline/0.1% bovine serum albumin in polypropylene tubes at -20°C Freeze-Thaw Stability (Cycles) 6 freeze (-80°C)-thaw (ambient temperature) cycles in polypropylene tubes under white light Long-Term Storage Stability (Days) Long-Term Stability: 98 days in polypropylene tubes at -80°C, 10 days in polypropylene tubes at -20°C Dilution Integrity Samples diluted up to 200-fold can be quantified Assay Volume Required 0.200 mL Regression Type 4PL, 1/Y Matrix Effect No significant matrix effect was observed in any of the 10 human plasma (EDTA) lots that were fortified at low and high concentrations Hemolyzed Sample Integrity No significant interference for PTH (1-84) was observed in any of the 3 hemolyzed human plasma (EDTA) lots that were fortified near the concentration of the LLOQ (10.0 pg/mL) and near the concentration of the high QC (1500 pg/mL) sample Turbid Sample Integrity No significant interference for PTH (1-84) was observed in any of the 3 turbid human plasma (EDTA) lots that were fortified near the concentra tion of the LLOQ (10.0 pg/mL) and near the concentration of the high QC (1500 pg/mL) sample Short-Term Stability for Stock Solutions 6 hours at approximately 100 µg/mL in phosphate- (Stock) buffered saline/0.1% bovine serum albumin in polypropylene tubes at ambient temperature under white light Short-Term Stability for Stock Solutions 6 hours at 1.00 µg/mL and 100 ng/mL in phosphate- (Substock) buffered saline/2% bovine serum albumin in polypropylene tubes at ambient temperature under white light Stability of Analyte During Sample up to 120 minutes in human whole blood (EDTA) Collections and Handling in polypropylene tubes at ambient temperature under UV shielded light light Batch Size 132 runs Table 6. PTH (1-34) Validation Summary Information Requested Data Analyte PTH (1-34) Method Description Sample pre-treatment and direct analysis using enzyme linked immunosorbent assay (ELISA) Limit of Quantitation (pg/mL) 6.00 pg/mL Standard Curve Concentrations (pg/mL) 3.00 (anchor point), 6.00, 10.0, 25.0, 40.0, 60.0, 75.0, 100, 130, and 150 pg/mL QC Concentrations (pg/mL) LLOQ QC, 18.0, 75.0, 120, and ULOQ QC pg/mL QC Intra-Batch Precision Range (% CV) 0.3 to 7.4% QC Intra-Batch Accuracy Range (% Bias) -16.1 to 21.7% QC Inter-Batch Precision Range (% CV) 2.9 to 9.7% QC Inter-Batch Accuracy Range (% Bias) -8.3 to 12.2% Bench-Top Stability (Hrs) Short-Term Stability: 13 hours in polypropylene tubes at ambient temperature under white light Stock Stability (Days) Long-Term Stability for Stock Solutions (Stock): 131 days at 1.00 mg/mL in water in a polypropylene container at -20°C Freeze-Thaw Stability (Cycles) 6 freeze (-20°C)-thaw (ambient temperature) cycles in polypropylene tubes under white light (LLOQ and ULOQ QCs), 5 freeze (-20°C)-thaw (ambient temperature) cycles in polypropylene tubes under white light (QC D) Long-Term Storage Stability (Days) Long-Term Stability: 284 days (QC A and ULOQ QC) and 259 days (QC D) in polypropylene tubes at -20°C and -80°C Dilution Integrity Samples diluted up to 150-fold can be quantified Selectivity No quantitation greater than the LLOQ of PTH (1-34) was observed from endogenous components in any of the 10 human plasma (EDTA) lots screened Assay Volume Required 0.600 mL Regression Type 4PL, 1/Y2 Known Metabolite PTH 1-84 100, 150, and 500 pg/mL Matrix Effect No significant matrix effect was observed in 8 of the 9 human plasma (EDTA) lots that were fortified near the concentration of the LLOQ (6.00 pg/mL) and in 8 of the 10 human plasma (EDTA) lots that were fortified near the concentration of the high QC (120 pg/mL) sample Hemolyzed Sample Integrity Hemolysis has an effect on the measurement of the analyte Turbid Sample Integrity Turbidity has an effect on the measurement of the analyte Long-Term Stability for Stock Solutions 285 days at 10,000 ng/mL in assay dilution buffer (Substock) polypropylene container at -20°C 99 days at 5000 ng/mL in human plasma (EDTA) in a polypropylene container at -20°C 285 days at 5.00 ng/mL in assay dilution buffer in a polypropylene container at -20°C Short-Term Stability for Stock Solutions 8 hours at 1.00 mg/mL in water in a polypropylene (Stock) container at ambient temperature under white light Short-Term Stability for Stock Solutions 8 hours at 5.00 ng/mL in assay dilution buffer in a (Substock) polypropylene container at ambient temperature under white light Stability of Analyte During Sample up to 120 minutes in human whole blood (EDTA) Collections and Handling polypropylene tubes at ambient temperature under white light Batch Size 1, 96 well plate Figure 2. PTH (1-34) Example Standard Curve Instrument Response (O.D.) Nominal Conc.

Upload: truongtuyen

Post on 14-Feb-2017

214 views

Category:

Documents


2 download

TRANSCRIPT

The Development and Validation of Specific Ligand Binding Methods for the Determination of Parathyroid Hormone (1-34) and (1-84) in Human Plasma for Pharmacokinetic Sample Analysis R. L. Sukovaty, S. K. Roby, C. E. Sheldon, A. Cunningham Celerion, Lincoln, NE

was less than 9.7% C.V. Inter-batch accuracy (% Bias) of the same quality controls samples was between -8.3 and +12.2% (Table 2).

• Sample collection and handling stability was established in whole blood for 2 hours at ambient temperature under UV-shielded light for PTH (1-84) and under white light for PTH (1-34) (Tables 3 and 4).

• Short-term stability in plasma was established for 24 hours for PTH (1-84) and 13 hours for PTH (1-34) at ambient temperature under white light. Cumulative short-term stability for PTH (1-84) was established for 39 hours at ambient temperature under white light.

• Freeze and Thaw stability was established for six freeze (-80°C for PTH (1-84) and -20°C for PTH (1-34)) and thaw (ambient tempera-ture) cycles for both PTH (1-84) and PTH (1-34).

• Long term stability of matrix samples was established for 98 days for PTH (1-84) at -80°C, 10 days for PTH (1-84) at -20°C, and for 259 days for PTH (1-34) at -20°C and -80°C.

• The quantitative integrity of PTH (1-84) and PTH (1-34) samples prepared in multiple lots of non-hemolyzed, non-turbid matrix was verified. The quantitative integrity of PTH (1-84) samples prepared in three lots of hemolyzed matrix and in three lots of turbid matrix was verified. Hemolyzed and turbid matrix did have an effect on the measurement of prepared PTH (1-34) samples.

• An evaluation of dilution integrity demonstrated that a dilution factor of 200 can be applied to PTH (1-84) samples and a dilution factor of 150 can be applied to PTH (1-34) samples to dilute them into the quantifiable range without impacting linearity.

• The absence of a high dose hook effect was demonstrated for both PTH (1-84) and PTH (1-34) by assaying a sample with a con-centration higher than the ULOQ at a minimum of 3 dilution levels above the ULOQ. All samples back-calculated with concentrations greater than the ULOQ for the respective assays.

CONCLUSIONS:• The validated methods allow for rapid, selective, accurate and

reproducible quantitation of PTH (1-84) and PTH (1-34) in human plasma samples for pharmacokinetic evaluation.

PURPOSE:• Parathyroid hormone (PTH) is an 84 amino acid polypeptide

produced in the parathyroid gland and plays a major role in regulating serum calcium concentrations. PTH (1-34) is a polypeptide consisting of the 34 N-terminal amino acids of the intact PTH (1-84) which shows similar effects on serum calcium regulation.

• Methods for determining PTH (1-84) in human plasma need to account for endogenous concentrations of the intact peptide when preparing controls and calibrators in plasma. Methods for determining PTH (1-34) in human plasma also need to account for antibody cross-reactivity with endogenous PTH (1-84).

• Commercially available PTH (1-84) kits and PTH (1-34) kits were adapted and validated to meet FDA guidelines for validation of bioanalytical methods.

METHODS:• PTH (1-84): Unknown samples and calibration samples prepared

in human plasma (treated with an antibody against the C-terminal portion of PTH (1-84) attached to a solid substrate to remove endogenous PTH (1-84)) were added to test tubes containing a fixed antibody (C-terminal specific) and a radioactively labeled antibody (N-terminal specific). After washing, the tubes were counted in a gamma counter. The counts detected were direct-ly proportional to the concentration of PTH (1-84) present in the sample (Figure 1).

• PTH (1-34): Unknown samples and calibration samples in human plasma were pre-treated with an antibody against the C-terminal portion of PTH (1-84) attached to a solid substrate to remove PTH (1-84). The pre-treated samples and calibration samples were pipetted into microtiter plates coated with the appropriate capture antibodies. The wells were washed to remove unbound material and enzyme-labeled antibody was added. Unbound labeled anti-body was removed and a chromogenic substrate was added to the bound labeled antibody. The development of the colored reaction product was directly proportional to the concentration of PTH (1-34) present in the sample and was detected using a spectrophotometric plate reader (Figure 2).

RESULTS:• The PTH (1-84) method uses a 4-parameter logistic regression

weighted 1/y over the range of 10.0 – 1500 pg/mL. The PTH (1-34) method uses a 4-parameter logistic regression weighted 1/y2 over the range of 6.00 – 150 pg/mL.

• The concentrations of PTH (1-84) and PTH (1-34) calibration standards were back-calculated from the regression equation of the experimental data. The coefficients of variation (C.V.) were less than or equal to 4.4% and 8.5%, respectively.

• Inter-batch precision of PTH (1-84) quality control samples between 10.0 and 2000 pg/mL was less than 11.6% C.V. Inter-batch accuracy (% Bias) of the same quality control samples was between -1.3 and +8.8% (Table 1). Inter-batch precision (%CV) of PTH (1-34) quality control samples between 6.00 and 150 pg/mL

Figure 1. PTH (1-84) Example Standard Curve

Inst

rum

ent

Res

po

nse

(C

PM

)

Nominal Conc.

Endogenous Endogenous+LowSpike Endogenous+HighSpike

PTH(1-84) 0minutes 120minutes 0minutes 120minutes 0minutes 120minutes

28.0 30.2 51.7 49.9 938 949

26.6 29.3 50.6 53.5 983 955

27.3 29.3 49.7 52.0 909 947

26.5 29.4 49.9 49.7 921 887

28.0 29.3 46.5 51.2 923 943

27.2 26.5 48.9 50.5 943 885

Mean 27.3 29.0 49.6 51.1 936 928

%CV 2.4 4.4 3.6 2.8 2.8 3.5

%ofControl 106.2 103.0 99.1

n 6 6 6 6 6 6

Table 3: Stability of PTH (1-84) During Sample Collection and Handling from Human Whole Blood (EDTA) at Ambient Temperature Under UV Shielded Light Conditions

LowSpike HighSpike

PTH(1-34) 0minutes 120minutes 0minutes 120minutes

33.3 46.9 219 212

33.7 31.8 219 211

34.7 25.9 222 215

34.5 27.5 218 216

34.8 25.1 223 214

34.5 24.2 221 214

Mean 34.3 30.2 220 214

%CV 1.8 28.5 0.9 0.9

%ofControl 88.0 97.3

n 6 6 6 6

Table 4: Stability of PTH (1-34) During Sample Collection and Handling from Human Whole Blood (EDTA) at Ambient Temperature Under White Light Conditions

Table 1. PTH (1-84) Inter-Batch Precision and Accuracy

LLOQ ENDO QCA QCB QCC ULOQPTH(1-84) 10.0pg/mL 11.7pg/mL 29.7pg/mL 160pg/mL 1510pg/mL 2000pg/mL

Inter-BatchMean 10.6 11.7 32.3 163 1490 1990

Inter-BatchSD 0.886 1.36 2.02 4.41 71.5 84.9

Inter-Batch%CV 8.4 11.6 6.3 2.7 4.8 4.3

Inter-Batch%Bias 6.0 0.0 8.8 1.9 -1.3 -0.5

n 18 18 18 18 18 18

Table 2: PTH (1-34) Inter-Batch Precision and Accuracy

LLOQQC QCA QCB QCC ULOQQCPTH(1-34) 6.00pg/mL 18.0pg/mL 75.0pg/mL 120pg/mL 150pg/mL

Inter-BatchMean 6.73 16.5 77.6 125 162

Inter-BatchSD 0.651 0.774 2.26 4.17 8.04

Inter-Batch%CV 9.7 4.7 2.9 3.3 5.0

Inter-Batch%Bias 12.2 -8.3 3.5 4.2 8.0

n 17 18 18 21 18

Table 5. PTH (1-84) Validation Summary

InformationRequested Data

Analyte PTH(1-84)

MethodDescription Directanalysisusingimmunoradiometric assay(IRMA)

LimitofQuantitation(pg/mL) 10.0pg/mL

StandardCurveConcentrations(pg/mL) 4.00(anchorpoint),10.0,25.0,50.0,100,200, 400,800,1600,and2000pg/mL

QCConcentrations(pg/mL) LLOQQC,endogenous(11.7pg/mL),29.7,160, 1510,andULOQQCpg/mL

QCIntra-BatchPrecisionRange(%CV) 1.2to18.0%

QCIntra-BatchAccuracyRange(%Bias) -12.0to16.0%

QCInter-BatchPrecisionRange(%CV) 2.7to11.6%

QCInter-BatchAccuracyRange(%Bias) -1.3to8.8%

Bench-TopStability(Hrs) Short-TermStability:24hoursinpolypropylene tubesatambienttemperatureunderwhitelight

CumulativeShort-TermStability:39hoursin polypropylenetubesatambienttemperatureunder whitelight(totalofallthawcycles)

StockStability(Days) Long-TermStabilityforStockSolutions(Stock): 132daysatapproximately100µg/mLin phosphate-bufferedsaline/0.1%bovineserum albumininpolypropylenetubesat-20°C

Freeze-ThawStability(Cycles) 6freeze(-80°C)-thaw(ambienttemperature)cycles inpolypropylenetubesunderwhitelight

Long-TermStorageStability(Days) Long-TermStability:98daysinpolypropylene tubesat-80°C,10daysinpolypropylenetubes at-20°C

DilutionIntegrity Samplesdilutedupto200-foldcanbequantified

AssayVolumeRequired 0.200mL

RegressionType 4PL,1/Y

MatrixEffect Nosignificantmatrixeffectwasobservedinany ofthe10humanplasma(EDTA)lotsthatwere fortifiedatlowandhighconcentrations

HemolyzedSampleIntegrity NosignificantinterferenceforPTH(1-84)was observedinanyofthe3hemolyzedhuman plasma(EDTA)lotsthatwerefortifiednearthe concentrationoftheLLOQ(10.0pg/mL)andnear theconcentrationofthehighQC(1500pg/mL) sample

TurbidSampleIntegrity NosignificantinterferenceforPTH(1-84)was observedinanyofthe3turbidhumanplasma (EDTA)lotsthatwerefortifiedneartheconcentra tionoftheLLOQ(10.0pg/mL)andnearthe concentrationofthehighQC(1500pg/mL)sample

Short-TermStabilityforStockSolutions 6hoursatapproximately100µg/mLinphosphate-(Stock) bufferedsaline/0.1%bovineserumalbuminin polypropylenetubesatambienttemperatureunder whitelight

Short-TermStabilityforStockSolutions 6hoursat1.00µg/mLand100ng/mLinphosphate-(Substock) bufferedsaline/2%bovineserumalbuminin polypropylenetubesatambienttemperatureunder whitelight

StabilityofAnalyteDuringSample upto120minutesinhumanwholeblood(EDTA)CollectionsandHandling inpolypropylenetubesatambienttemperature underUVshieldedlightlight

BatchSize 132runs

Table 6. PTH (1-34) Validation Summary

InformationRequested Data

Analyte PTH(1-34)

MethodDescription Samplepre-treatmentanddirectanalysisusing enzymelinkedimmunosorbentassay(ELISA)

LimitofQuantitation(pg/mL) 6.00pg/mL

StandardCurveConcentrations(pg/mL) 3.00(anchorpoint),6.00,10.0,25.0,40.0,60.0, 75.0,100,130,and150pg/mL

QCConcentrations(pg/mL) LLOQQC,18.0,75.0,120,andULOQQCpg/mL

QCIntra-BatchPrecisionRange(%CV) 0.3to7.4%

QCIntra-BatchAccuracyRange(%Bias) -16.1to21.7%

QCInter-BatchPrecisionRange(%CV) 2.9to9.7%

QCInter-BatchAccuracyRange(%Bias) -8.3to12.2%

Bench-TopStability(Hrs) Short-TermStability:13hoursinpolypropylene tubesatambienttemperatureunderwhitelight

StockStability(Days) Long-TermStabilityforStockSolutions(Stock): 131daysat1.00mg/mLinwaterina polypropylenecontainerat-20°C

Freeze-ThawStability(Cycles) 6freeze(-20°C)-thaw(ambienttemperature)cycles inpolypropylenetubesunderwhitelight(LLOQ andULOQQCs),5freeze(-20°C)-thaw(ambient temperature)cyclesinpolypropylenetubesunder whitelight(QCD)

Long-TermStorageStability(Days) Long-TermStability:284days(QCAandULOQ QC)and259days(QCD)inpolypropylenetubes at-20°Cand-80°C

DilutionIntegrity Samplesdilutedupto150-foldcanbequantified

Selectivity NoquantitationgreaterthantheLLOQof PTH(1-34)wasobservedfromendogenous componentsinanyofthe10humanplasma (EDTA)lotsscreened

AssayVolumeRequired 0.600mL

RegressionType 4PL,1/Y2

KnownMetabolite PTH1-84100,150,and500pg/mL

MatrixEffect Nosignificantmatrixeffectwasobservedin8of the9humanplasma(EDTA)lotsthatwerefortified neartheconcentrationoftheLLOQ(6.00pg/mL) andin8ofthe10humanplasma(EDTA)lotsthat werefortifiedneartheconcentrationofthehighQC (120pg/mL)sample

HemolyzedSampleIntegrity Hemolysishasaneffectonthemeasurementof theanalyte

TurbidSampleIntegrity Turbidityhasaneffectonthemeasurementof theanalyte

Long-TermStabilityforStockSolutions 285daysat10,000ng/mLinassaydilutionbuffer(Substock) polypropylenecontainerat-20°C

99daysat5000ng/mLinhumanplasma(EDTA) inapolypropylenecontainerat-20°C

285daysat5.00ng/mLinassaydilutionbuffer inapolypropylenecontainerat-20°C

Short-TermStabilityforStockSolutions 8hoursat1.00mg/mLinwaterinapolypropylene(Stock) containeratambienttemperatureunderwhitelight

Short-TermStabilityforStockSolutions 8hoursat5.00ng/mLinassaydilutionbufferina(Substock) polypropylenecontaineratambienttemperature underwhitelight

StabilityofAnalyteDuringSample upto120minutesinhumanwholeblood(EDTA)CollectionsandHandling polypropylenetubesatambienttemperatureunder whitelight

BatchSize 1,96wellplate

Figure 2. PTH (1-34) Example Standard Curve

Inst

rum

ent

Res

po

nse

(O

.D.)

Nominal Conc.