quantification of carbon nano materials in complex matrices · 2015-08-18 · sw- & mw-cnt :...
TRANSCRIPT
![Page 1: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/1.jpg)
Quantification of carbon nano materials in complex matrices
Paul Westerhoff, PhD, PE, BCEE
Pierre Herckes & Takayuki Nosaka Arizona State University (Tempe, AZ)
Kyle Doudrick (University of Notre Dame)
RD833322
Nano-Go Funding: RES018801Z
![Page 2: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/2.jpg)
Outline • 1D to 3D Types of Carbon • Range of analytical
techniques & response ranges
• Standards & Analytics • Extraction / Separation
Strategies • Paths forward for Exposure
monitoring
![Page 3: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/3.jpg)
Types of Carbon Nanomaterials
720 Dalton 10 Å (1nm) x µm2 1-10 nm by µm-mm 3 - D 2 - D 1 - D
n C60
C 60, 70, etc Few Layer Graphene Single Wall CNT Multi- Wall CNT
Wide Ranges of surface Functionalities can occur
Graphene-Oxide Fullerol (Hydroxy-fullerene)
![Page 4: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/4.jpg)
Quantification Techniques Technique C60 & nC60 FLG & GO SW- & MW-CNT Light scattering ✔ ✔
Absorbance - UV - NIR fluorescence - Gel electrophesis
✔
✔ ✔
HPLC-UV (FFF-UV) ✔
LC-Mass Spec ✔
Thermal - Combustion / CO/CH4 - Microwave induced heat - TGA mass loss
✔
✔
✔
✔ ✔ ✔
14C Scintillation Thermal-Mass spec
✔ ✔
✔
✔ ✔
Raman spec. ✔ ✔
Photo-acoustic/thermal ✔
Single particle ICP-MS of catalyst
✔
![Page 5: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/5.jpg)
Examples – Fullerenes (C60)
• Light scattering (λ347 nm) • Liquid phase combustion
(TOC) • Thermal optical
transmittance • HPLC plus λ347 nm • LC/MS using 720 m/z
Improving specificity and lower detection limits
Wavelength (nm)
300 400 500 600 700 800
Abs
orba
nce
0.0
0.2
0.4
0.6
0.8
1.0
347 nm
Concentration of aqueous n-C60 in water (mg/L)
0 5 10 15 20 25 30 35
UV
Abso
rban
ce a
t 347
nm
0.0
0.2
0.4
0.6
0.8
1.0
Y = 0.0263 XR2 = 0.9999
Chen et al., 2008
![Page 6: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/6.jpg)
C60 analysis by HPLC-UV or –MS Requires Solvent Extraction
Benn et al., ABC, 2011
![Page 7: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/7.jpg)
Application of LC/MS C60 detection in a) a parking garage air sample of PM2.5, b)a filtered air sample spiked with C60, c) a parking garage air sample of size > PM2.5, d) C60 standard; Detection limit ~500 ng/L
![Page 8: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/8.jpg)
Fullerenes From Cosmetics
• A common cosmetic formulation disperses fullerenes using polyvinylpyrrolidone (C60-PVP)- see TEM
• LC/SM was used to separate and specifically detect fullerenes (C60 and C70) from interfering substances typically present in cosmetics (e.g., castor oil).
• C60 was detected in 4/5 commercial cosmetics ranging from 0.04 to 1.1 µg/g, and C70 was qualitatively detected in 2/5 samples.
• A single-use quantity of cosmetic (0.5 g) may contain up to 0.6 µg of C60 and demonstrates a pathway for human exposure to engineered fullerenes.
C60 Magic Stuff
Benn et al., Environ. Poll. (2011)
Dark spheres are C60
Background grey is PVP encapsulating C60
![Page 9: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/9.jpg)
Fullerols (C60(O), etc)
• Light scattering (λ347 nm) • Liquid phase combustion
(TOC) • Thermal optical
transmittance • LC/MS using 720 m/z
Chao et al., 2011
![Page 10: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/10.jpg)
Fullerol – Comparison of Detection Methods
R2 a MDL b [pg/mL]
RSD c SRFA d
UV/Vis 0.999 42 780 n.d. n.d.
Q1 scan 0.9996 125 2.9 % 29.4 %
MI scan 0.9999 1.5 0.7% 8.6 %
MRM scan 0.9999 0.19 0.5 % 2.5 %
single quad scan, Q1 multiple ion monitoring, MI MS with multiple reaction monitoring, MRM
No standards available
![Page 11: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/11.jpg)
Fullerene Summary
Pycke et al., 2011
![Page 12: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/12.jpg)
Quantification – CNTs
![Page 13: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/13.jpg)
CNT & Graphene UV/VIS Absorbance
0
0.1
0.2
0.3
0.4
0.5
0.6
200 300 400 500 600 700
Abso
rban
ce
Wavelength (nm)
0.3% O 3.5% O6.4% O 7.3% O8.3% O
SEM images of FLG (with gold sputtering)
0
0.5
1
1.5
2
2.5
200 300 400 500 600 700
Abso
rban
ce
Wavelength (nm)
0.5 mg/L 1 mg/L5 mg/L10 mg/L20 mg/L30 mg/L
GO Analysis
![Page 14: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/14.jpg)
Standard curve for GO at different wavelengths
300nm y = 0.030x
R² = 1
400 nm y = 0.010x R² = 0.999
570 nm y = 0.003x R² = 0.998
0
0.2
0.4
0.6
0.8
1
0 5 10 15 20 25 30 35
Abso
rban
ce
Concentration of GO (mg/L)
![Page 15: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/15.jpg)
Indirect Measurement • Single Particle ICP-MS (spICP-MS)
– Track residual catalyst rather than carbon
– Ranville et al. showed for SWCNT – Our group has evidence for MWCNT
Pulses ≈? ≈ CNTs
Reed et al., 2013
![Page 16: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/16.jpg)
Thermal Optical Transmittance (Temperature Programmed Oxidation)
0
500
1000
1500
2000
2500
0100200300400500600700800900
1000
0 100 200 300 400 500 600 700
FID
Sig
nal
Tem
pera
ture
(°C
)
Time (seconds)
OC PEC EC
Non-oxidizing Oxidizing
Sunset Laboratories Lab OC-EC Aerosol Analyzer
![Page 17: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/17.jpg)
Comparison of 15 CNTs
CNT ID CNT Type State Puritya Metal
Content
Outer Diameter
(nm)
Inner Diameter
(nm)
Length (µm)
MW-O MWCNT Raw >95% <5% 20-30 5-10 10-30 MW-P MWCNT Purified >98% <2% 20-30 5-10 10-30 MW-F MWCNT Functionalized >99.9% <0.01% 20-30 5-10 10-30 MW-100 MWCNT Raw >95% <5% 60-100 5-10 0.5-500 MW-30 MWCNT Raw >95% <5% 10-30 5-10 0.5-500 MW-20 MWCNT Raw >95% <5% 10-20 5-10 0.5-200 MW-15 MWCNT Raw >95% <5% 7-15 3-6 0.5-200 MW-Arc MWCNTc Raw <50% 0% 5-10b - - MW-15G MWCNTd Annealed >97% <1% 7-15 3-6 0.5-200 MW-Mitsui MWCNT - >98% <1% 20-70 - - MW-OH MWCNT Functionalized >95% <1.5% 8-15 3-5 10-50 MW-COOH MWCNT Functionalized >95% <1.5% 8-15 3-5 10-50 SW SWCNT Raw <50% <10% 1.1 - 0.5-100 SW-65 SWCNT Purified <75% <10% 0.8 - 0.45-2
![Page 18: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/18.jpg)
Thermal Properties of CNTs
0
100
200
300
400
500
600
700
800
900
0%
20%
40%
60%
80%
100%
0 500 1000
Tem
pera
ture
(°C
)
Perc
ent C
NT
Mas
s Rem
aini
ng
Time (seconds)
MW-ArcMW-15MW-FMW-OSW-65MW-P
0
200
400
600
800
1000
0 1000 20000%
20%
40%
60%
80%
100%
120%
Tem
pera
ture
(°C
)
Time (seconds)
Perc
ent C
NT
Mas
s Rem
aini
ng
SW-65SWMW-FMW-OMW-15MW-15G
Inert conditions (100% He)
Oxidizing conditions (90% He/ 10% O2)
Conclusion: Not all CNTs “burn” at the same
temperature
Conclusion: Surface oxygen groups allow some CNTs to burn in inert gas
environment
![Page 19: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/19.jpg)
Temperature Comparison for 50% CNT combustion
19
0
200
400
600
800
1000
0 500 1000 1500 20000%
20%
40%
60%
80%
100%
120%
Tem
pera
ture
(°C
)
Time (seconds)
Perc
ent C
NT
Mas
s Rem
aini
ng
SW-65SWMW-FMW-OMW-15MW-15GMW-MitsuiMW-ArcTemperature
Doudrick, K., P. Herckes, P. Westerhoff. Detection of carbon nanotubes in environmental matrices using programmed thermal analysis. Environ. Sci. Technol., 46(22), 12246–12253, 2012.
50% remaining
PTA method is a refinement to NIOSH Method 5040
![Page 20: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/20.jpg)
Thermal Properties of CNTs related to Structural Defects
0
100
200
300
400
500
600
1000 1250 1500 1750 2000
Inte
snity
Raman Shift (cm-1)
MW-PMW-Arc
G-band D-band
![Page 21: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/21.jpg)
CNTs & Other NM Detection in biomass
• Raman and imaging can detect CNTs, but not quantify them well
• Extraction protocol must: – Minimize oxidation of CNT – Remove interfering
background organic carbon (from rat lung tissue)
– Separate solid-phase CNT from liquid-phase dissolved tissue
![Page 22: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/22.jpg)
Selective digestion can remove organic matter & facilitate CNT quantification
• Can solvents remove organic matter? – Yes – Oxidants (H2O2) – Acids (HNO3, H2SO4) – Alkali (NaOH, KOH, NH4OH,
Solvable) – Enzymes (TMAH, ProtK)
• But, do solvents affect CNT detection? – Mostly, yes – Surface oxygenation results in
combustion at lower temperatures – Enymes and customized alkali fair
best
0%
20%
40%
60%
80%
100%
CN
T R
ecov
ery S-CNT
W-CNT
Doudrick, K., Corson, N., Oberdörster, G., Eder, A., Herckes, P., Westerhoff, P. Extraction of carbon nanotubes from rat lung tissue. ACS Nano (2013)
![Page 23: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/23.jpg)
Microsoft clipart
Alkali Treatment (Solvable)
Enzymatic Treatment (Proteinase K)
CNT Transfer
1 2
Collect CNTs
Result: CNT Dose
Analyze CNTs
23
2.9 ±0.19 µg CNTs, whole lung – 93% recovery
Wash/Centrifuge
Application of Extraction Method to Rat Lungs
Doudrick, K., Corson, N., Oberdörster, G., Eder, A., Herckes, P., Westerhoff, P. Extraction of carbon nanotubes from rat lung tissue. ACS Nano (2013).
![Page 24: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/24.jpg)
Hazard & Exposure Analyses
Silva et al., ACS Nano, 2013
![Page 25: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/25.jpg)
Similar Approach for Graphene (GO & FLG)
Step 1- Programmed Thermal Analysis (PTA)
0
100
200
300
400
500
600
700
800
900
1000
0
200
400
600
800
1000
1200
1400
0 200 400 600 800 1000
Tem
pera
ture
(°C)
(D
otte
d Li
ne)
FID
Sign
al
Instrument Time (s)
Graphene OxideFew-layer GrapheneTemperature
Helium Helium/Oxygen (90/10)
Area where background carbon also evolves
GO
FLG
![Page 26: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/26.jpg)
Step 2: Improve separation of GO signal from background organics
• Add reductant (NaBH4)
• Reduced graphene oxide (RGO) analysis by XPS yields decreases number of C-O & C=O bonds by > 5 fold
• PTA thermogram improves
0
100
200
300
400
500
600
700
800
900
1000
0
200
400
600
800
1000
1200
1400
0 100 200 300 400
Tem
pera
ture
(°C)
FID
Sig
nal
Time (s)
0.04%
0.4%
2%
8%
![Page 27: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/27.jpg)
Step 3: Adding SolvableTM to degrade organics
• Solvable is an alkaline digestate that degrades organics; surfactant helps separation
• Solvable + NaBH4 produces good pellet for separation & analysis
0
150
300
450
600
750
900
0.0
0.2
0.4
0.6
0.8
1.0
0 100 200 300 400 500 600 700
Tem
pera
ture
(°C)
Frac
tion
Gra
phen
e Re
mai
ning
Instrument Time (s)
GO/SolGO/Sol - BH (0.04%)GO/Sol - BH (2%)GOGO/Water - BH (0.4%)
![Page 28: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/28.jpg)
Final Digestion Method to Handle Separation of FLG, GO (or CNT) from High
Biomass Concentrations (1 g/L)
GO Recovery with 1 g/L biomass: 110% ± 13% Detection limit: 2 µg/L
![Page 29: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/29.jpg)
Carbon NM Monitoring in Air Samples?
3
Recovery of CNTs on air filter samples
(Conclusion: Excellent CNT recoveries indicates viability to monitor CNTs in workplace air)
Spiked CNT / ug TOT data / ug 1 1.00±0.15 5 4.35±0.32 10 9.59±0.58
Indoor air (MWCNT spiked onto filter)
Outdoor air Spiked CNT / ug TOT data / ug 1 0.80±0.17 5 4.48±0.36 10 10.06±0.63
• Goal: To quantify the presence of CNTs in the presence of background air particulates
• Samples analyzed for organic carbon and CNT by PTA – which is a refinement to NIOSH Method 5040
![Page 30: Quantification of carbon nano materials in complex matrices · 2015-08-18 · SW- & MW-CNT : Light scattering Absorbance - UV - NIR fluorescence - Gel electrophesis HPLC-UV (FFF-UV)](https://reader033.vdocuments.site/reader033/viewer/2022050501/5f93b10db860a7638d1663f5/html5/thumbnails/30.jpg)
Conclusions • C60 derivatives
– Extraction in solvent (toluene) gives lowest detection limits using LC-MS
– Solvent extraction from tissue and commercial products is possible
– Extraction from urine and fluids can use solid phase extraction • Graphene (FLG/GO) & CNTs (SW / MW)
– Thermal methods can be non-selective unless related to known CNT or Raman analysis
– spICP-MS emerging as potential indirect approach to quantify metal catalyst rather than carbon itself
– Extraction from tissues aim to minimize oxygen incorporation => Alkaline conditions are better + Enzymatic digestion
• Personal monitoring devices can collect Carbon Materials and filters can readily be extracted for analysis