QSTAR Operation

May 3, 2005

Bob Seward

Page 2: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Presentation CD Contents

Page 3: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

QSTAR Hardware

Page 4: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

10-7

torr

QSTAR Schematic

Page 5: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

QSTAR Schematic

Chernushevich et al, JMS 2001, 36: 849-865

Page 6: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

ESI Source

Curtain Plate

Page 7: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

ESI Source

Curtain Gas

Curtain Plate

Page 8: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

QSTAR Mass Ranges

Upper m/z Ranges

QSTAR 1 QSTAR 2TOFMS 12,000 12,000

MSMS 6,000 3,000

Lower m/z Ranges = 5

Page 9: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

With Turbo IonSpray Source

Page 10: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

TOF Analyzer

Page 11: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Turbo Pumps

Page 12: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

ESI Source With Curtain Plate

Page 13: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Curtain Plate Removed

Page 14: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Source Removed Skimmer Facing Up

Page 15: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Skimmer Removed

Ring

Page 16: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Skimmer

Page 17: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Q0

Source Removed: Q0

Page 18: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Q0 Q1 Q2 - EnclosedTo Maintain CAD

Gas Pressure

Quadrupole Rail

Page 19: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Stubbies and Q1

Decreasing Pressure

Page 20: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Q2 and CAD Gas Supply

Page 21: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Q2 Exit Lens

Page 22: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

DC Quad Lens: Directs Ion Beam Into TOF

Page 23: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Nanospray

Page 24: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Nanospray Source Cameras

CenterSide

XYZ Stage

Page 25: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Move TipClose ToOrifice

Nanospray Tip Positioning 1: Be Sure That QSTAR Is In Standby Mode

Page 26: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Center Side

Off Axis

Nanospray Tip Positioning 2Move Nanospray Tip Off Axis

Page 27: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Nanospray Tip Positioning 3Move Tip Back

Center Side

Good Spraying Distance

Page 28: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Sample Preparation and Acquisiton

Page 29: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Orifice Plate Contamination

ContaminantBuildup Around

The Orifice

Page 30: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Nanospray Guidelines To Reduce Contamination

• Pre-clean capillaries before pulling tips

• Inspect every tip under the microscope (200X)

• Always spray off axis from the orifice

• Cleanup the sample (dirty samples introduce contamination and require longer acquisition time)

• Acquire only long enough to get a good spectrum

• Put QSTAR in standby when not acquiring data (IS voltage is still applied when acquisition is stopped)

Page 31: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Monitoring Ion Counts With The Display Meter

Keep Stop Rate Below

50,000: Dilute Or Cleanup Sample If Too High

Signal On Each Of The 4 Anodes Of

The MCP

Page 32: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Wash The Capillary With Solvent (50%ACN)

Page 33: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Dry The Capillaries Overnight Before Pulling The Nanospray Tips

Page 34: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Performance Standards• Run before and after sample set acquisition

• Positive Mode– [Glu1]-Fibrinopeptide B, Human, Sigma F-3261– EGVNDNEEGFFSAR– 1pmol/µl, 50%ACN / 0.5%Formic Acid

• Negative Mode– Sulfated Disaccharide, V-LABS, Inc.– Di-6S (C 3203) or Di-4S (C 3202), m/z 458.0604 (1-)– 1pmol/µl, 30%MeOH / 0.1% NH4OH (v/v, using 30% NH4OH

stock)

• Prepare concentrated aliquots, store frozen– Dilute a fresh aliquot for each day’s use

Page 35: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Sample Log Book• Name and project folder (if different from name)

• Sample names, descriptions, and concentrations (if known)

• Solvent composition

• Acquisition mode (+ or -)

• File names of all samples (including performance standards)

• Cleanup procedure (ex, ZipTip, HILIC, SEC)

• Any unusual occurrences (ex. software crash)

Page 36: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Analyst QS Software

Page 37: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Software Components• Analyst QS with oMALDI Server

– Instrument operation and data analysis• Service Pack 8

– Update to Analyst QS• Biotools

– Extension of Analyst QS for Protein/Peptide analysis • PepSea Server

– Database search using Biotools data• Scripts

– Relatively simple software additions for data analysis• Pro ID

– Protein/Peptide database search script• Install software in the order listed

Page 38: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Analyst QS Overview

Manual Acquisition

Information Dependent Acquisition (IDA)

Data Analysis

Hardware Setup, Printing Configuration, Security

Page 39: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Analyst QS Overview

Navigation Bar:Most functions accessible from toolbars and menus.Can be closed to save screen space.

Page 40: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Analyst QS Overview

Configure Mode: Security Configuration

Page 41: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Analyst QS Overview

Help Topics

Page 42: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Creating a Project

“New Project” Icon

Page 43: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Creating a Project

Type in Project Name – Press OK:D:\PE Sciex Data\Projects\My Project

Data FolderAcquisition Methods Folder

Page 44: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

(1) Open Project(2) Select Tune Mode

Startup: Open Project and Tune Mode

Instrument Status Indicator:Green -ReadyYellow -StandbyRed -Fault

Page 45: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Make Sure That The Instrument Is In Standby Mode Before

Touching the Source!

Page 46: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

(3) View Queue Standby / Ready Mode

Startup: Open the Queue Manager

Queue Manager Can Be Minimized

Page 47: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

“Manual Tune” Icon

Startup: Open Manual Tuning Window

Ready Mode

Page 48: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Startup

D:\PE Sciex Data\Projects\API Instrument\Tuning Cache

Temp Folder – Only 2 most recent files are saved – can copy and paste file to save

Wait a few seconds when starting/stopping or switching between standby/ready mode to prevent software freeze

Page 49: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

10-7

torr

Instrument Settings

***

**

*

*

* User Adjustable

Page 50: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

TOF MS Acquisition

Page 51: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Set Acquisition Parameters: Source/Gas & MS Tabs

Source/Gas Tab MS Tab

Set Polarity In The Software Before

Setting “TOF Offset”On The Instrument

Panel

Page 52: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Set to Nanospray and Adjust “TOF Offset” to Correct Value

There Are Separate “TOF Offset” Values

For Negative And Positive Modes

Page 53: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Set Acquisition Parameters: Compound & Advanced MS Tabs

Compound Tab Advanced MS Tab

Page 54: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Save The Acquisition Method

Page 55: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Save a Method for Each Frequently-Used Analysis Mode

Open a new method when switching between TOFMS and MSMS:-Faster-Less chance of making errors

Page 56: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Open Acquisition Method

(1) Ready Mode(2) Manual Tune Window Open

(3) Open File

Page 57: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Product Ion Mode

Page 58: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Set Acquisiton Parameters: Product Ion Mode

Compound Tab MS Tab

Page 59: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Set Acquisiton Parameters: Resolution & Advanced MS Tabs

Advanced MS TabResolution Tab

Low Q1 Resolution:- Increased sensitivity- Transmit isotope envelope- TOF resolution unaffected

Page 60: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Ion Transmission Efficiency

Page 61: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

400-1700

350-1700 (2 “Hops”)

80-1700 (3 “Hops”)

Max m/zMin m/z

> ~5

Quadrupole “Hops”To Cover m/z Range

Min – Max m/z

Quadrupole Transmission Windows, TOFMS (RF-Only)

Page 62: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

10-7

torr

QSTAR Schematic

Page 63: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Ion Beamfrom Quad

Detector

“slice” of ions injected

Orthogonal Injection TOF

Page 64: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

X =width ofejected beam slice

L = distance traveled by ions between pulses

Duty cycle = x/L= 4% (m/z 100)= 12% (m/z 1000)= 26% (m/z 5000)

10 eV

TOF Cycle

Pulse

Duty Cycle of Orthogonal Injection TOF

Page 65: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

400-2000

400-2100

TOF Pulser Frequency Changes With Maximum m/z

Slower Frequency = Lower Transmission Efficiency

TOF Analyzer “Waits” Longer BetweenCycles – Ion Beam From Quad is Lost

During This Time

Min – Max m/z

Pulser Frequency Range: 3-20 KHz

Page 66: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Q2 Pulsing (Enhancing)

Page 67: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Enhancing

Enhance a Single m/z

Enhance a m/z Range

Page 68: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Ions are Stored in Q2, then Pulsed into the TOF

Q2 collision cell

QSTAR Pulsar: Pulsing ON

Page 69: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Q2 Exit Lens

Page 70: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Enhancing a Single m/zIRD & IRW are Optimized for that m/z

No effect on the Quadrupole Transmission Windows

Page 71: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Enhance All

More Quadrupole Transmission Windows are Generated

IRD & IRW are Optimized for Each Window

No Enhancing (70-1800)

Enhance All (70-1800)

Page 72: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Enhance All

Page 73: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Enhancing: Glu-Fibrinopeptide B

No Enhance Enhance All Enhance 785.8

Page 74: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Increasing Quadrupole Residence Time at Higher Mass

Page 75: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Changing CE For Individual Quadrupole Transmission Windows

Right-ClickNext to %Column

Page 76: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

MALDI Analysis

Page 77: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

MALDI Target and AdapterUse DHB Matrix Only

Bruker AdapterQSTAR Target

Page 78: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

MALDI Source

Laser Guide

Page 79: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Installing the MALDI Target

Target

Page 80: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Installing the MALDI Target

Page 81: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

oMALDI Server Software

• Controls laser and sample target

• Sample acquisition controlled by Analyst QS

• Start acquisition first – then switch laser on

Page 82: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

oMALDI ServerStart From Desktop Icon

Page 83: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Pumping Down The MALDI Source

Turns Bright Green When Vacuum is

Ready

Page 84: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Removing The MALDI Plate

Page 85: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Laser Control

Pulse Rate: 20

Attenuator: 20,000

Laser On/Off

Page 86: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Positioning The Target and Sample

Page 87: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Positioning The Sample

Page 88: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

TOF Calibration

Page 89: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Select Two Peaks – Right Click – Re-Calibrate TOF

Page 90: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Select Calibration File

Page 91: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Select Calibration File

Page 92: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Calculate New Calibrations

Page 93: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Calculate New Calibrations

Page 94: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Calibrate Spectrum

Page 95: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Calibrate Spectrum

Page 96: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Apply to Entire File

For Highest Accuracy: Calibrate Instrument and Acquire Samples with the Same TOF Pulser Frequency

Page 97: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Recalibration of a Spectrum

Page 98: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Apply Calibrations to Sample

Uncheck Boxes

Page 99: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Save As Recalibrated File

Page 100: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Precursor Ion Scanning• Q1 scans a m/z range

• Collision in Q2

• Characteristic fragment ions detected in TOF

• Output spectrum indicates precursors from Q1 scan that produce the characteristic fragment

• Low resolution (Q1 scan)

Page 101: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Precursor Ion Scanning

Page 102: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Precursor Ion Scan

●●

Compound Tab MS Tab

Stores TOF m/z Range: Can Generate New

Precursor Scan From Other Fragment Ions

Page 103: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Precursor Ion Scan

Resolution Tab Advanced Tab

Peak Hopping = Integer m/z ValuesProfile = Fractional m/z Values

Page 104: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Precursor Ion Scan

500Da (1Da Step Size) x 0.02s Accumulation Time = 10 s Scan Time

●●

Page 105: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620 640 660 680 700m/z, amu

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

8588

411.351

342.625

342.791

407.750

305.641415.749 514.192314.644 345.356

335.121 356.890442.821 519.940 693.587395.154

325.174418.902362.896

372.173

Beta Casein Trypsin DigestNegative Mode TOFMS

Page 106: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Beta Casein Trypsin DigestNegative Mode Precursor of 79

300 320 340 360 380 400 420 440 460 480 500 520 540 560 580 600 620 640 660 680 700m/z, amu

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

7579

341.854

410.662

513.843

519.568423.310352.112

331.364

Page 107: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Information-Dependent Acquision (IDA)

Page 108: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

LCMS IDA Cycle1. TOFMS Survey Scan

2. Dependent Product Ion Scan-Peak Intensity-Charge State-m/z Range

3. New TOFMS Survey Scan-Dynamic Exclusion

4. Repeat Cycle Until End of LC Run

Page 109: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Display With IDA Explorer

Page 110: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Rolling Collision Energy3

Col

lisio

n En

ergi

esR

ollin

g C

E

IDA CE Parameters.dll(Scripts Menu)

Page 111: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Include and Exclude Lists

Can Import a Tab-Delimited List

Include List Ions Get First Priority in Selection for IDA

Page 112: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Include and Exclude Lists

Page 113: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Include and Exclude Lists

Exclude For Entire RunExclude For First 5 MinutesExclude From 14 to 16 Minutes

Page 114: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Mass Difference (and Neutral-Loss) IDA

Induce Some In-Source Fragmentation For Neutral-Loss IDA

Can be Used to Detect

Isotope Pairs (ex. ICAT)

Page 115: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Troubleshooting

Page 116: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Troubleshooting Steps

1) Check nitrogen gas supply

2) Exit and restart Analyst QS

3) Stop Analyst Services and restart

4) Close Analyst QS and restart Windows

5) Deactivate and activate Hardware Profile

Page 117: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Stop and Restart Analyst Services

Page 118: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Deactivate and Reactivate the Hardware Profile

Page 119: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Protein ID Software Tools

Page 120: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Protein and PTM Identifications

• Individual Spectra– BioTools– Mascot

• IDA Data– BioTools– Mascot– Pro ID

Page 121: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Bioanalyst 1.1 Tutorial

Page 122: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Pro ID Tutorial

Page 123: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Mascot Tutorials

Page 124: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Pro ID Database Searching

Page 125: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Pro ID Script

Page 126: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Data Dictionary

Page 127: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Pro ID Search Results

Page 128: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

View Sequence

Page 129: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Show Peptide ID Evidence

Page 130: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

View Selected Ions

Page 131: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Printing: Report Templates

Page 132: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Report Template Editor

Page 133: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker
Page 134: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker
Page 135: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker
Page 136: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Report Template Margins

Page 137: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker
Page 138: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Select Report Template For Printing

Page 139: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Other Software Features

Page 140: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Highlight Peak

Graph Information Window

Page 141: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

List Data

Page 142: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

List Data

Set Threshold For Peak Labeling

Page 143: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Show File Information

Page 144: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Show File Information

Page 145: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Changing Graph Label Precision

Page 146: QStar Presentation 050503 · 2009. 11. 29. · Sample Preparation and Acquisiton. Orifice Plate Contamination Contaminant Buildup Around ... Use DHB Matrix Only QSTAR Target Bruker

Trace Color

Main

Alternate(BPC)