qcinstruments gregori
TRANSCRIPT
-
8/11/2019 QCInstruments Gregori
1/22
Grald Grgori, Ph.DPurdue University Cytometry Laboratories
Hansen Life Sciences Research Building, B050
201 S. University Street,
West Lafayette, IN 47907-2064,USA
Ph: +1(765) 494-0757
Fax: +1(765) 494-0517
e-mail: [email protected]
Quality Controls:
Get your instruments
under control!
-
8/11/2019 QCInstruments Gregori
2/22
In flow cytometry scatter and fluorescence scales are in arbitrary units
Scatter and Fluorescence values = function of set up(Voltage and gain of PMT)
- +
FS-PMT Voltage or gain
Example : beads 10 m
GreenFluorescence(au)
Forward Scatter (au)
Forward Scatter (au)
Arbitrary units?
-
8/11/2019 QCInstruments Gregori
3/22
Confidence in the instrument performance
&
Confidence in the results
Use of a Standard
Quality Control Tests
Daily basis
-
8/11/2019 QCInstruments Gregori
4/22
In theory :A standard = a reference(defined by a user, a laboratory, or anyaknowledged authority)
Properties accurately known(i.e., provided by the manufacturer)
In practice :A manufactured particle(fluorescent beads: several sizes
and excitation or emission wavelengths)
A biological particle(i.e., chicken and trout erythrocytes
DNA measurements)
Used as an absolute reference for qualitative and
quantitative comparisons
Whats a Standard?
-
8/11/2019 QCInstruments Gregori
5/22
Check the alignment of the optical pathway
Check the stability of the machine (Quality Control)
Test the capacities of the cytometer (flow rate)
Set up the Sorting (define the delay)
Quality Control of Instruments
-
8/11/2019 QCInstruments Gregori
6/22
0 200 400 600 800 1000
FL1 Lin
0
200
400
600
800
100
Number
1.93
1.93
0 200 400 600 800 1000
FL2 Lin
0
200
400
600
800
1000
1200
140
Num
ber
1.99
1.99
0 200 400 600 800 1000
FS Lin
0
200
400
600
800
10
00
1200
Number
1.37
1.37CVs < 2% expected for
beads (1-10 m)
If CVs > 2%check the alignment
(flow cell or laser position,
dichroic mirrors)
Very useful on sorters
Beads
Parameter intensity (au)
Numberofeven
ts
To check the alignment
of the cytometer
Intensity is
constant with
time
-
8/11/2019 QCInstruments Gregori
7/22
Example of bad alignment
CV=4.5
-
8/11/2019 QCInstruments Gregori
8/22
0 200 400 600 800 1000
FL3 Lin
0
200
400
600
800
1000
120
Number 1.62
R4
0 200 400 600 800 1000
FL1 Lin
0
200
400
600
800
100
Number
1.93R2
0 200 400 600 800 1000
FL2 Lin
0
200
400
600
800
1000
1200
140
Number
1.99
R3
0 200 400 600 800 1000
FS Lin
0
200
4
00
600
800
1000
1200
Number
1.37R1
Once a protocol is defined for a
particular standard bead solution
Beads must always be plotted
in the same regions
Daily quality control
Parameter intensity (au)
Numberofevents
Internal quality control
Check the stability of the
cytometer over time
-
8/11/2019 QCInstruments Gregori
9/22
Many possible problems :
Leak in the fluidics modified flow rates; instability
Check for bubbles
Dirty flow cell
Laser dying beam intensity decreases
Bead solution too old or damaged photobleaching
Problem of stability of the
cytometer?
-
8/11/2019 QCInstruments Gregori
10/22
-
8/11/2019 QCInstruments Gregori
11/22
Interrogation
point
Flow
cell
Last attached
undulation
Delay(s)
Piezoelectric TransducerInput
Break-off pointFirst droplet
Electric pulse
+
+++
+
+
+++
+
Delay determined
using
fluorescent beads
Droplet formation and timing
-
8/11/2019 QCInstruments Gregori
12/22
Sort 50 beads on a slide with
different delays (ex : from 29 to 35)
If you achieve 50 out of 50 beads, set the delay to that setting (ex: 31).
If beads are split between 2 drops, adjust the flow cell vertically
3332313029
Slide
Drop
Sorting process on the Altr a (Coulter, USA)
Green Fluorescence (au)
ForwardScatte
r(au)
beads
Beads (10 m)
Delay value
Count beads
(epifluorescent
Microscope)
Set up the delay with fluorescent beads
-
8/11/2019 QCInstruments Gregori
13/22
The necessary step to convert arbitrary units
into absolute physical values
Calibration
Calibration =adjustment of an instrument in order to
express the results in some accurate
physical measure.
Calibrator = a material known to have accurate
measured valuesfor one or
several characteristics
-
8/11/2019 QCInstruments Gregori
14/22
Example:Quantitation of antibody binding
capacity of cell populationsby flow cytometry
Quantum Simply Cellular (QSC)
-
8/11/2019 QCInstruments Gregori
15/22
Identical microbeads with various calibrated binding capacitiesof goat-anti-mouse IgG on their surface :
Ev
ents
Mean fluorescence intensity (au)
1 2
3 4
Blank
QSC, Cat. No. 815
Bangs Laboratories, Inc.
www.bangslabs.com
Antibody binding capacity (ABC) provided
by the manufacturer :
Blank. 0 MESF
1. 6851 MESF
2. 23379 MESF
3. 58333 MESF
4. 213369 MESF
bead
Ab site
MESF=Molecules of equivalent soluble fluorochrome
QSC Beads (Quantum Simply Cellular)
-
8/11/2019 QCInstruments Gregori
16/22
y = 32790x - 3926.1
R2
= 0.9981
0
50000
100000
150000
200000
250000
0 1 2 3 4 5 6 7
mean fluorescence- HLA-DR-FITC (au)
Moleculesof
equivalentsoluble
fluoroch
rome(MESF)
Courtesy of K. Rhageb
measure by FCM
corresponding MESF
Binding capacity
of cells in your sample
Quantum Simply Cellular Beads
Calibration Curve
-
8/11/2019 QCInstruments Gregori
17/22
Absolute Counts
Determination of cell concentrationby flow cytometry
Get the control of the volume analyzed
Four different methods
-
8/11/2019 QCInstruments Gregori
18/22
Sample
Cell concentration(# of events / volume)
CYTORON ABSOLUTE
(Ortho Diagnostic Systems)
I. Direct Absolute Count
To analyze a bead solution of known concentration
Control of the fluidic
-
8/11/2019 QCInstruments Gregori
19/22
Sample
Weigh the sample
before analysis
(Weight W0)
Weigh the sample after
analysis (Weight W1)
Analysis
(Flow Cytometer)
Volume analyzed
V = (W0W1)/
Cell concentration
= N/V
cell number (N)
density
DisadvantagesTime consuming
Less accurate
(back flow of sheath fluid
in the sample)
Analysis in one run
II. Weigh a Sample
-
8/11/2019 QCInstruments Gregori
20/22
Bead solution (known concentration)
count by microscopyTruCountTMbeads (BD)
Sample
Add a very
accurate volume
calculate the bead concentrationin the sample (1-10% of the
expected density of the target cells)
Flow Cytometer
cell number (N)
bead numberVolume
analyzed (V)
Cell concentration
= N/VStewart & Steinkamp, 1982, Cytometry 2 : 238-243
III. Add Beads in the Sample
-
8/11/2019 QCInstruments Gregori
21/22
Hypothesis:flow rate (l/s) through a flow cytometer = constantBergeron et al, 2003, Cytometry 52B:37-39
Conclusion:volume (V) analyzed in a fixed time = constant
No need to add beads in the samples.
If N events are analyzed by the cytometer in the fixed time
then cell concentration = N/V (cells/l)
Hint!
Analyzes must be done with the same flow rate
Volume accurately determined (microscopy, TruCountTMbeads) and controlled
Beads not necessary in the sample, but can be used as internal standard
IV. Determination of the Flow Rate
-
8/11/2019 QCInstruments Gregori
22/22
Quality Control Tests are mandatory
To assess the alignmentTo assess and control instrument
performance
for quantitative and reproducible applications on anyflow cytometer.
ReferencesR.A. Hoffman, Current Protocols in Cytometry, 1997 : 1.3.1-1.3.19
J.C.S. Wood, Current Protocols in Cytometry, 1997 : 1.4.1-1.4.12
Cytometry, Volume 33, Number 2, 1998 :
Conclusion