Flow measurementSelection guide
Instruments International
Endress+HauserInstruments International AGKaegenstrasse 24153 ReinachSwitzerlandTel. +41 61 715 81 00Fax +41 61 715 25 00http://[email protected]
12.06/I.I.
Diameter Measuring range
Elec
trom
agne
tic (
EMF)
Prol
ine
Prom
ag
Promag WWater/Waste water
DN 25 to 2000 (1" to 78")
0 to 10 m/s 0 to 110,000 m³/h
Promag PChemical/Process applications
DN 15 to 600 (½" to 24")
0 to 10 m/s 0 to 9600 m³/h
Promag HHygiene
DN 2 to 100 (1/12" to 4")
0 to 10 m/s 0 to 282 m³/h
Promag S Inhomogeneous fluidsHigh solid content
DN 15 to 600 (½" to 24")
0 to 10 m/s 0 to 9600 m³/h
Ultr
ason
ic
Prol
ine
Pros
onic
Flo
w
Prosonic Flow UClamp-on sensor Pharma, semiconductor ind.
DN 15 to 100 (½" to 4")
0 to 10 m/s 0 to 424 m³/h
Prosonic Flow WClamp-on sensor Standard applications
DN 50 to 4000 (2" to 156")
0 to 15 m/s 0 to 678,550 m³/h
Prosonic Flow WInsertion sensor Standard applications
DN 200 to 4000 (8" to 156")
0 to 15 m/s 0 to 678,550 m³/h
Prosonic Flow PClamp-on sensor Process applications
DN 50 to 4000 (2" to 156")
0 to 15 m/s 0 to 678,550 m³/h
Prosonic Flow CInline sensor Water/Waste water
DN 300 to 2000 (12" to 78")
0 to 10 m/s 0 to 47,400 m³/h
Prosonic Flow FInline sensorTwo-wire technology
DN 25 to 150(1" to 6")
0 to 10 m/s 0 to 600 m³/h
Diff
. pre
ssur
e Deltatop/-set DPOOrifice plate
DN 10 to 1000 (½" to 40")
Liquids: 0 to 18,280 m³/h Gases: 0 to 169,650 m³/h Steam: 0 to 169,650 m³/h
Deltatop/-set DPPPitot tube
DN 40...12’000 (1½" to 480")
Liquids: 0 to 327,500 m³/h Gases: 0 to 9,767,000 m³/h Steam: 0 to 4,400,000 m³/h
Requirements Fluids (examples) Measuring principle
Elec
tro-
m
agne
tic
Cor
iolis
Ultr
ason
ic
Vort
ex
Ther
mal
Diff
eren
tial
pres
sure
Liqu
ids
Liquids in general Water • • • • •Flow rates < 2 l/h Additives, flavors • •Non-conductive liquids Oil, hydrocarbons • • • •Viscous liquids > 500 mPa�s Molasse, shower gel • •Cryogenic fluids Liquid O2 , Ar, N2 • • •Volume flow Water, milk, solvents • • • • •Mass flow Chemicals • • •Accuracy ≤ 0.2% • •Nominal pipe diameter DN > 500 • • •Operable flow range > 1:30 • • •Pressure loss negligible • • •Density, concentration, viscosity metering Foods, shower gel •Process temperature > 200 °C • • •Process pressure > 100 bar Crude oil • • • •
An example of using this selection guide
Selecting the measuring principle
Define all the primary requirements for your measuring point: fluid type, nominal diameter, accuracy, pressure, temperature, etc.
In the table on Page 4, select the category for the type of fluid in question: Liquid – Gas – Steam – Special applications.
In the table, specify all the requirements (see Point ) you defined for your measuring point.
Now select the measuring principles to which all the requirements apply (•).
Selecting the flowmeter
Now open out this brochure completely. The overview contains all the measuring principles you selected under Point and the corresponding Endress+Hauser flowmeters.
Now mark the sensors that can be installed in the pipe of your choice (DN = 1600 mm).
Mark those transmitters that can be combined with the sensors in question and that meet the required accuracy specifications (<0.3% o.r.)
You have found the two flowmeters that suit your measuring point.
• Electromagnetic• Ultrasonic• Differential pressure
• Promag W• Prosonic Flow W (clamp-on)• Prosonic Flow W (insertion)• Prosonic Flow P• Prosonic Flow C• Deltatop/Deltaset DPP
• 50 (W)• 53 (W)
• Promag 50W (standard)• Promag 53W (high-end)
Example
• Fluid: drinking water (urban area)• Pipe diameter: DN 1600• Required accuracy: <0.3% o.r.• Measured variable: Volume flow • Liquids
• Liquids in general • Volume flow• Nominal diameter DN > 500
• Electromagnetic• Ultrasonic• Differential pressure
➊
➋
➌
➍
➎
➏
➐
➑
➊
➍
➋ ➌ ➍
➎ ➏ ➐ ➐➑➑
Promag WWater/Waste water
2 3
Combinable with
Accuracy (best possible accuracy)
10Basic
SensorW, P, H ±0.5%
50Standard
Sensor W, P, H
±0.5% Optional: ±0.2%
53High-end
Sensor W, P, H ±0.2%
51Custody transfer
Sensor W, P
±0.5% Optional: ±0.2%
23Two-wire
Sensor P, H ±0.5%
55Special applications
Sensor S ±0.2%
91 Basic
Sensor W
Onsite: <2.0% Under reference conditions: ±0.5%
90 Standard
Sensor U, W, P
Onsite: <2.0% (U, W, P) Under reference conditions: ±0.5%
±0.5% (C)
93 High-end
Sensor U, W, P, C
Onsite: <2.0% Under reference conditions: ±0.5%
92 F Inline
Sensor F
Liquids: For 2 or 3 beams ±0.5%
For 4 beams ±0.3%
92 T Portable
Sensor U, W, P
Onsite: <2.0% Under reference conditions: ±0.5%
Deltabar SPMD 70/75
Deltatop PMD 70/75
Deltaset PMD 70/75
Orifice plate DPO: ±1.0%
Pitot tube DPP: ±1.5%
50Standard
53High-end
Requirements Fluids (examples) Measuring principle
Elec
tro-
m
agne
tic
Cor
iolis
Ultr
ason
ic
Vort
ex
Ther
mal
Diff
eren
tial
pres
sure
Liqu
ids
Liquids in general Water • • • • •Flow rates < 2 l/h Additives, flavors • •Non-conductive liquids Oil, hydrocarbons • • • •Viscous liquids > 500 mPa�s Molasse, shower gel • •Cryogenic fluids Liquid O2 , Ar, N2 • • •Volume flow Water, milk, solvents • • • • •Mass flow Chemicals • • •Accuracy ≤ 0.2% • •Nominal pipe diameter DN > 500 • • •Operable flow range > 1:30 • • •Pressure loss negligible • • •Density, concentration, viscosity metering Foods, shower gel •Process temperature > 200 °C • • •Process pressure > 100 bar Crude oil • • • •
Gas
Gas flow in general Natural gas, air • • • •Contaminated gas Biogas, exhaust gas • •Volume flow Exhaust gas • •Mass flow CO2 • • • •Accuracy ≤ 0.5% Rare gas (e.g. argon) •Nominal pipe diameter DN > 500 • •Operable flow range > 1:30 • •Pressure loss negligible Air • •Process temperature > 200 °C Exhaust gas • •Process pressure > 100 bar CNG • • •
Stea
m
Volume flow • •Mass flow • •Nominal pipe diameter DN > 300 •Operable flow range > 1:10 •Pressure loss negligible •Process temperature > 300 °C Overheated steam • •Process pressure > 100 bar Overheated steam • •
Spec
ial
appl
icat
ions
Hygienic applications (SIP/CIP) Milk, beer • • •Slurries, suspended solids Thick sludge • •Liquid/liquid mixtures Oil/water, emulsions •Highly abrasive fluids (e.g. in mining) Ore slurry •Bidirectional metering (forward/reverse) • • • •Measurement from outside without process interruption/retrofitting/temporary metering •
Filling/bottling applications see Flow measurement OEM (inside)
Flow monitoring see Flow monitoring (inside)
Selection table “Measuring principles”
Please note:The selection table above provides a rough overview and is used for quick clarification of suitable measuring principles. The table is not exhaustive. If further information has to be clarified, feel free to contact Endress+Hauser any time. We’ll be glad to be of assistance.
Supplementary documentation
Flow measurement For liquids, gases and steam
Products and services at a glance
FA005D/06/en
Metering flow the cost-optimised way
Comprehensive metering point solutions for planning, commissioning and maintenance
CP006D/06/en
Flow measurement in utilities
Energy management in your hands –customized solutions for compressed air, gas, steam and water
CP007D/06/en
Flow handbook
A practical guide: Measurement technologies –applications – solutions
ISBN 3-9520220-4-7(Order number 50104766)
4 5
Project documentation – data collection, archival, exportThe use of additional expansion mod-ules permits cost-saving administration and documentation of data for entire plant projects:• Clear presentation in a tree structure
of any desired number of metering points (e.g. by building, floor, meter-ing point), according to ISA 88.
• Uncomplicated inclusion of measur-ing devices directly from the Selection and Sizing functions. Also possible for instruments from other manufactur-ers or projects.
• Detailed listing of project manage-ment data (responsible person, cost-ing and budget data, etc.)
• Multiple data exchange possibilities in Excel or XML format: data export to other programs, data import from other projects, etc.
Applicator® • Optimal planning and
sizing of flowmeters• Administration and
documentation for entire plant projects
FieldCare® • Easy and dependable commissioning and
configuring of measuring devices via PC• Complete documentation of device settings• Storing device data (upload/download)• Simplified trouble-shooting
• Process-optimization through the visualization of measurement data in realtime (line plotting function)
• Analysis of diagnostic data for maintenance purposes• Administration and analysis of verification results
from the “Fieldcheck” test instrument
Fieldcheck® • Simulation and
testing of alarm/process conditions directly in the plant (without flow)
• Verification of flow-meters directly in the piping (without removal)
W@M W@M – Life cycle manage-ment from Endress+Hauser is an open information system which provides technical and opera-tive management with complete
data flow and archival – from commissioning on through to maintenance and service within a plant, at any time or location. This is attained within an open
system based on Internet/intranet technology, and incorporating all software programs, products and services from Endress+Hauser.
Applicator ® – Optimal planning and sizingThe first steps are always the most important when it comes to planning metering points. An optimally dimensioned flowmeter at the right location provides many advantages: • Consistent process quality• Diminished risk of instrument failure at a later time• Product savings due to error-free, precise measurement
Proven since more than 20 yearsApplicator is a software for selecting and specifing flowme-ters that has proven its abilities over many years. A wealth of industry experience and specialized know-how is embedded in Applicator. It enables people with limited technical knowledge to specify metering points quickly and easily, without time- consuming research.
The most recent version of Applicator has a raft of new func-tions, including selection functions for all the other areas cov-ered by Endress+Hauser – namely level, pressure, temperature, liquid analysis, etc. You’ll find that it opens up an enormous range of application possibilities in metrology and control.
Selection – straight to your objectiveWhen starting out to select the ideal flowmeter, Appli-cator guides through all the important parameters. Just enter the type of fluid, requisite process data or meter-ing point attributes and Applicator already displays the measuring principles and families of meters best suited to the case in hand.
Sizing – safe specificationBy entering further application-specific values, such as process pressure, process temperature or measuring range, Applicator immediately determines the ideal flowmeter type – and along with the suitable nomi-nal diameter, working range, pressure loss, accuracy, material loads, conformity with the Pressure Equipment Directives (PED) or corrosion resistance. Messages and warnings make it easy to select the right meter, help avoid common mistakes, and enable optimization of the measuring point to suit specific requirements.
6 7
Diameter Measuring range Pressure PN Processtemperature
Materials(wetted parts)
Process connection Protection Hygienic certification
Elec
trom
agne
tic (
EMF)
Prol
ine
Prom
ag
Promag WWater/Waste water
DN 25 to 2000 (1" to 78")
0 to 10 m/s 0 to 110,000 m³/h
PN 6 to 40, Cl 150 to 300 JIS 10 to 20 K AWWA Class D
AS Table E, PN 16
–20 to +80 °C Hard rubber liner Polyurethane liner Flange: DIN, ANSI, JIS, AWWA, AS IP67/IP68
(NEMA 4X/6P) –
Promag PChemical/Process applications
DN 15 to 600 (½" to 24")
0 to 10 m/s 0 to 9600 m³/h
PN 10 to 40 Cl 150 to 300 JIS 10 to 20 K
AS Table E, PN 16
–40 to +180 °C PFA liner PTFE liner Flange: DIN, ANSI, JIS, AS IP67/IP68
(NEMA 4X/6P) –
Promag HHygiene
DN 2 to 100 (1/12" to 4")
0 to 10 m/s 0 to 282 m³/h
PN 16 to 40 Cl 150
JIS 10 to 20 K–20 to +150 °C PFA liner
Flange: DIN, ANSI, JIS, external thread, internal thread,
weld connection, clamp connection, couplings
IP67 (NEMA 4X) 3-A, EHEDG
Promag S Inhomogeneous fluidsHigh solid content
DN 15 to 600 (½" to 24")
0 to 10 m/s 0 to 9600 m³/h
PN 10 to 40 Cl 150 to 300 JIS 10 to 20 K
AS Table E, PN 16
–40 to +180 °C PTFE, PFA, polyurethane, natural rubber liner Flange: DIN, ANSI, JIS, AS IP67/IP68
(NEMA 4X/6P) –
Cor
iolis
Prol
ine
Prom
ass
Promass FUniversal flowmeter
DN 8 to 250 (5/16" to 10")
DN 25, 50, 80 (HT)0 to 2,200,000 kg/h
PN 16 to 100 Cl 150 to 600 JIS 10 to 63 K
–50 to +200 °C –50 to +350 °C (HT)
904L (1.4539)/ 316L (1.4404)
Alloy C-22 (2.4602)
Flange: DIN, ANSI, JIS, Tri-Clamp, couplings, VCO
IP67 (NEMA 4X) 3-A, EHEDG
Promass MHigh pressure
DN 8 to 80 (5/16" to 3")
DN 8 to 25 (HP)0 to 180,000 kg/h
PN 40 to 100, PN 350 (HP) Cl 150 to 600 JIS 10 to 63 K
–50 to +150 °C DN 8...50: Titanium grade 9DN 80: Titanium grade 2
Flange: DIN, ANSI, JIS, Tri-Clamp, couplings
IP67 (NEMA 4X) 3-A
Promass ALow flow rates
DN 1 to 4(1/24" to 5/32") 0 to 450 kg/h
PN 16 to 400 Cl 150 to 600 JIS 10 to 63 K
–50 to +200 °C 904L (1.4539) Alloy C-22 (2.4602) Tri-Clamp, VCO IP67
(NEMA 4X) 3-A, EHEDG
Promass IHygiene, viscosity
DN 8 to 80 (5/16" to 3") 0 to 180,000 kg/h
PN 40 to 100 Cl 150 to 600 JIS 10 to 63 K
–50 to +150 °C Measuring tube: Titanium grade 9Flange: Titanium grade 2
Flange: DIN, ANSI, JIS, Tri-Clamp, couplings
IP67 (NEMA 4X) 3-A, EHEDG
Promass HChemically aggressive fluids
DN 8 to 50 (5/16" to 2") 0 to 70,000 kg/h
PN 40 Cl 150 to 300 JIS 10 to 20 K
–50 to +200 °C Zirconium 702/R 60702 Flange: DIN, ANSI, JIS IP67
(NEMA 4X) –
Promass EBasic applications
DN 8 to 50 (5/16" to 2") 0 to 70,000 kg/h
PN 40 to 100 Cl 150 to 600 JIS 10 to 63 K
–40 to +125 °C 904L (1.4539)316L (1.4404)
Flange: DIN, ANSI, JIS, Tri-Clamp, couplings, VCO
IP67 (NEMA 4X) –
Promass S Food & Beverages
DN 8 to 50 (5/16" to 2") 0 to 70,000 kg/h
PN 40 to 63 Cl 150 to 300 JIS 20 to 40 K
–50 to +150 °C 904L (1.4539)316L (1.4435)
Flange: DIN, ANSI, JIS, Tri-Clamp, couplings
IP67 (NEMA 4X) 3-A, EHEDG
Promass PLife Sciences Industries(Pharma, biotechnology)
DN 8 to 50 (5/16" to 2") 0 to 70,000 kg/h
PN 40 to 63 Cl 150 to 300 JIS 20 to 40 K
–50 to +200 °C 316L (1.4435) Flange: DIN, ANSI, JIS, Tri-Clamp, couplings
IP67 (NEMA 4X)
3-A, EHEDG,ASME BPE
Ultr
ason
ic
Prol
ine
Pros
onic
Flo
w
Prosonic Flow UClamp-on sensor Pharma, semiconductor ind.
DN 15 to 100 (½" to 4")
0 to 10 m/s 0 to 424 m³/h
No limit Non-invasive –20 to +80 °C not wetted
(Clamp-on sensor) Clamp-on sensor IP52 –
Prosonic Flow WClamp-on sensor Standard applications
DN 50 to 4000 (2" to 156")
0 to 15 m/s 0 to 678,550 m³/h
No limit Non-invasive –20 to +80 °C not wetted
(Clamp-on sensor) Clamp-on sensor IP67 (NEMA 4X) –
Prosonic Flow WInsertion sensor Standard applications
DN 200 to 4000 (8" to 156")
0 to 15 m/s 0 to 678,550 m³/h PN 16 –20 to +80 °C 316L (1.4404) Welding connector IP67
(NEMA 4X) –
Prosonic Flow PClamp-on sensor Process applications
DN 50 to 4000 (2" to 156")
0 to 15 m/s 0 to 678,550 m³/h
No limit Non-invasive –40 to +170 °C not wetted
(Clamp-on sensor) Clamp-on sensor IP68 (NEMA 6P) –
Prosonic Flow CInline sensor Water/Waste water
DN 300 to 2000 (12" to 78")
0 to 10 m/s 0 to 47,400 m³/h
PN 16 CI 150 –20 to +60 °C Carbon steel, epoxy coated Flange: DIN, ANSI, AWWA IP68
(NEMA 6P) –
Prosonic Flow FInline sensorTwo-wire technology
DN 25 to 150(1" to 6")
0 to 10 m/s 0 to 600 m³/h
PN 40CI 150 –40 to +150 °C
Meter body: CF3M (1.4404)Sensor: 316/3166 (1.4404)
Flange: F316/F316L (1.4404)Flange: DIN, ANSI, JIS IP68
(NEMA 6P) –
Vort
ex
Prol
ine
Prow
irl
Prowirl FFlanged version
DN 15 to 300 (½" to 12")
Liquids: 0.16 to 2360 m³/h Gases: 3 to 19,700 m³/h
PN 10 to 250 Cl 150 to 1500 JIS 10 to 40 K
–200 to +400 °COptional:
–200 to +450 °C
Meter body: CF3M (1.4404),Alloy C-22 (2.4602)
DSC sensor: 316L (1.4435), Inconel 718, Alloy C-22 (2.4602)
Flange: DIN, ANSI, JISOptional: with integrated reduction
of line size
IP67 (NEMA 4X) –
Prowirl WWafer version
DN 15 to 150 (½" to 6")
Liquids: 0.19 to 625 m³/h Gases: 4 to 5210 m³/h
PN 10 to 40 Cl 150 to 300 JIS 10 to 20 K
–200 to +400 °COptional:
–200 to +450 °C
Meter body: CF3M (1.4404)DSC sensor: 316L (1.4435) Wafer: DIN, ANSI, JIS IP67
(NEMA 4X) –
Ther
mal
Prol
ine
t-m
ass t-mass I
Insertion sensor DN 80 to 1500
(3" to 60") Gases: 0 to 70 m/s PN 20 –40 to +130 °C 316L (1.4404) Alloy C-22 Thread: NPT, BSP IP67 –
t-mass FFlanged version
DN 15 to 100 (½" to 4") Gases: 0 to 70 m/s PN 40
Cl 300 –40 to +100 °C 316L (1.4404) Alloy C-22 Flange: DIN, ANSI, JIS IP67 –
Diff
. pre
ssur
e Deltatop/-set DPOOrifice plate
DN 10 to 1000 (½" to 40")
Liquids: 0 to 18,280 m³/h Gases: 0 to 169,650 m³/h Steam: 0 to 169,650 m³/h
PN 160/PN 420 –200 to +1000 °CC22.8
316Ti (1.4571) 15Mo3 (1.5415)
Wafer (corner tapping) Flange (flange tapping) Pipe unit (meter run)
IP65 –
Deltatop/-set DPPPitot tube
DN 40...12’000 (1½" to 480")
Liquids: 0 to 327,500 m³/h Gases: 0 to 9,767,000 m³/h Steam: 0 to 4,400,000 m³/h
PN 160/PN 420 –200 to +1000 °C C22.8 316Ti (1.4571)
Cutting ring, flange, flow tap IP65 –
EMF Dosimag
Filling/bottlingConductive liquids
DN 4, DN 8, DN 15K, DN 15(5/32", 5/16", ½")
0 to 10 m/s0 to 1.66 l/s PN 16 –40 to +125 °C PFA liner Tri-Clamp, weld connection IP67
(NEMA 4X) 3-A, EHEDG
Cor
iolis
DosimassFilling/bottlingLiquids
DN 8 to 25 (5/16" to 1")
0 to 5 kg/s PN 16 –40 to +125 °C 904L (1.4539)316L (1.4404) Tri-Clamp, couplings IP67
(NEMA 4X) 3-A
CNGmassFuelingCompressed Natural Gas
DN 15, DN 25 (½", 1") 0 to 150 kg/min G¾" and G1" for
max. 350 bar –50 to +125 °C 316L (1.4404)316L (1.4435)
Internal thread: G¾" (DN 15)G1" (DN 25)
IP67(NEMA 4X) –
LPGmass FuelingLiquified Petroleum Gas
DN 8 to 50 (5/16" to 2") 0 to 70,000 kg/h
PN 40 Cl 150 to 300 JIS 10 to 63 K
–40 to +125 °C 316L (1.4404)904L (1.4539)
Flange: DIN, ANSI, JIS, VCO
IP67(NEMA 4X) –
EMF Magphant
Limit switch/Trend displayConductive liquids
DN 1 to 2000(½" to 78") 0 to 5 m/s PN 16 –20 to +120 °C PVDF Weld nipple IP66
(NEMA 4X) –
Ther
mal
Flowphant DTT31Standard applicationsLiquids
DN 8 to 2000(5/16" to 78")
0.03 to 3 m/s as a relative value between
0 and 100%max. PN 100 –20 to +85 °C AISI 316L Weld nipple, compression fitting,
threaded connection: ANSI, ISO IP65/ IP66 –
Flowphant DTT35Hygienic applications Liquids
DN 8 to 2000(5/16" to 78")
0.03 to 3 m/s as a relative value between
0 and 100%max. PN 100 –20 to +85 °C AISI 316L and FKM Viton,
EPDM (sealings)Conical metal-metal G½"
Adapter concept: Tri-Clamp IP65/ IP66 –
t-switch ATT11Limit switch Liquids, gases
DN 40 to 1000 (1½" to 40")
Liquids: 0 to 3 m/s Gases: 0 to 50 Nm/s PN 25 –10 to +80 °C 316L (1.4404/1.4435) Threaded connection: BSP, NPT,
couplings, Tri-ClampIP66
(NEMA 4X) –
t-trend ATT12Trend display Liquids, gases
DN 40 to 1000 (1½" to 40")
Liquids: 0 to 3 m/s Gases: 0 to 50 Nm/s PN 25 –10 to +80 °C 316L (1.4404/1.4435) Threaded connection: BSP, NPT,
couplings, Tri-ClampIP66
(NEMA 4X) –
Combinable with
Housings Display/Operation Ambienttemperature
Power supply Galvanic isolation Inputs Outputs Communication Accuracy (best possible accuracy)
Ex-approvals Protection
10Basic
SensorW, P, H
Compact AluRemote Alu
Two-line LCD/ push buttons
Operation via FieldCare–20 to +60 °C 85 to 250 V AC (45 to 65 Hz)
20 to 28 V AC (45 to 65 Hz) or 11 to 40 V DCElectronical connections are
galvanically isolateted from each other – 4 to 20 mA, pulse, status HART® ±0.5% FM, CSA Div. 2 IP67 (NEMA 4X)
50Standard
Sensor W, P, H
Compact SSCompact AluRemote Alu
Two-line, backlit LCD/ push buttons
Operation via FieldCare
–20 to +60 °COptional: –40 to +60 °C
85 to 260 V AC (45 to 65 Hz) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC
Electronical connections are galvanically isolateted from each other Status 4 to 20 mA, pulse/frequency
(1 kHz passive), statusHART®, PROFIBUS PA,
PROFIBUS DP±0.5%
Optional: ±0.2% ATEX, FM, CSA, TIIS, NEPSI
In preparation: IECIP67
(NEMA 4X)
53High-end
Sensor W, P, H
Compact SSCompact AluRemote Alu
Four-line, backlit LCD/ touch control
Operation via FieldCare
–20 to +60 °C Optional: –40 to +60 °C
85 to 260 V AC (45 to 65 Hz) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC
Electronical connections are galvanically isolateted from each other
Status, current input
4 to 20 mA, pulse/frequency (10 kHz, active/passive), relays, status
HART®, PROFIBUS PA, PROFIBUS DP, FOUNDATION
Fieldbus, MODBUS® RS485±0.2% ATEX, FM, CSA, TIIS, NEPSI
In preparation: IECIP67
(NEMA 4X)
51Custody transfer
Sensor W, P
Compact AluRemote Alu
Two-line, backlit LCD/ push buttons
Operation via FieldCare–20 to +60 °C 85 to 260 V AC (45 to 65 Hz)
20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DCElectronical connections are
galvanically isolateted from each other Status 4 to 20 mA, pulse/frequency (1 kHz passive), status HART® ±0.5%
Optional: ±0.2% ATEX
In preparation: IECIP67
(NEMA 4X)
23Two-wire
Sensor P, H Compact Alu
Four-line LCD/ touch control
Operation via FieldCare–20 to +60 °C
Two-wire12 to 30 V DC
13.9 to 30 V DC (Ex) (45 to 65 Hz)
Electronical connections are galvanically isolateted from each other – 4 to 20 mA, pulse/frequency, status HART® ±0.5% ATEX, FM, CSA IP67
(NEMA 4X)
55Special applications
Sensor S
Compact AluRemote Alu
Four-line, backlit LCD/ touch control
Operation via FieldCare
–20 to +50 °COptional: –40 to +50 °C 20 to 260 V AC/DC (45 to 65 Hz) Electronical connections are
galvanically isolateted from each otherStatus,
current input4 to 20 mA, pulse/frequency
(10 kHz, active/passive), relays, statusHART®, PROFIBUS PA, FOUNDATION Fieldbus ±0.2%
FM, CSA Div. 2 In preparation: ATEX II 3G,
IEC Zone 2
IP67 (NEMA 4X)
40 Basic
SensorE Compact Alu
Two-line, backlit LCD/ Operation via HART® resp.
FieldCare
–20 to +60 °C Optional: –40 to +60 °C
85 to 260 V AC (45 to 65 Hz) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC
Electronical connections are galvanically isolateted from each other Status 4 to 20 mA, frequency, status HART® Liquids: ±0.5%
Gases: ±1.0% ATEX, IEC, FM, CSA, TIIS, NEPSI IP67 (NEMA 4X)
80 Standard
Sensor F, M, A, I, H, E, S, P
Compact SSCompact AluRemote Alu
Two-line, backlit LCD/ push buttons
Operation via FieldCare
–20 to +60 °COptional: –40 to +60 °C
85 to 260 V AC (45 to 65 Hz) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC
Electronical connections are galvanically isolateted from each other Status 4 to 20 mA, pulse/frequency
(1 kHz passive), statusHART®, PROFIBUS PA,
PROFIBUS DPLiquids: ±0.15%Gases: ±0.35% ATEX, IEC, FM, CSA, TIIS, NEPSI IP67
(NEMA 4X)
83 High-end
Sensor F, M, A, I, H, E, S, P
Compact SSCompact AluRemote Alu
Four-line, backlit LCD/ touch control
Operation via FieldCare
–20 to +60 °C Optional: –40 to +60 °C
85 to 260 V AC (45 to 65 Hz) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC
Electronical connections are galvanically isolateted from each other
Status, current input
4 to 20 mA, pulse/frequency (10 kHz, active/passive), relays, status
HART®, PROFIBUS PA, PROFIBUS DP, FOUNDATION
Fieldbus, MODBUS® RS485
Liquids: ±0.1%Gases: ±0.35 % ATEX, IEC, FM, CSA, TIIS, NEPSI IP67
(NEMA 4X)
84 Custody transfer
Sensor F, M, A
Compact SS Compact AluRemote Alu
Four-line, backlit LCD/ touch control
Operation via FieldCare
–20 to +60 °COptional: –40 to +60 °C
85 to 260 V AC (45 to 65 Hz) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC
Electronical connections are galvanically isolateted from each other Status
4 to 20 mA, pulse/frequency (10 kHz, active/passive),
2 × pulse (90°/180°), relays, status
HART® MODBUS® RS485
Liquids: ±0.1%Gases: ±0.35 % ATEX, IEC, FM, CSA, TIIS, NEPSI IP67
(NEMA 4X)
91 Basic
Sensor W Remote Alu
Two-line LCD/ push buttons
Operation via FieldCare–20 to +60 °C 85 to 250 V AC (45 to 65 Hz)
20 to 28 V AC (45 to 65 Hz) or 11 to 40 V DCElectronical connections are
galvanically isolateted from each other – 4 to 20 mA, pulse, status HART® Onsite: <2.0% Under reference conditions: ±0.5% FM, CSA Div. 2 IP67
(NEMA 4X)
90 Standard
Sensor U, W, P Remote Alu
Two-line, backlit LCD/ push buttons
Operation via FieldCare
–20 to +60 °COptional: –40 to +60 °C
85 to 260 V AC (45 to 65 Hz) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC
Electronical connections are galvanically isolateted from each other Status 4 to 20 mA, pulse/frequency
(1 kHz passive), status HART®, PROFIBUS PAOnsite: <2.0% (U, W, P)
Under reference conditions: ±0.5%
±0.5% (C)FM, CSA Div. 2 IP67
(NEMA 4X)
93 High-end
Sensor U, W, P, C Remote Alu
Four-line, backlit LCD/ touch control
Operation via FieldCare
–20 to +60 °C Optional: –40 to +60 °C
85 to 260 V AC (45 to 65 Hz) 20 to 55 V AC (45 to 65 Hz) or 16 to 62 V DC
Electronical connections are galvanically isolateted from each other
Status, current input
4 to 20 mA, pulse/frequency (10 kHz, active/passive),
relays, status
HART®, PROFIBUS PA, FOUNDATION Fieldbus
Onsite: <2.0% Under reference conditions: ±0.5%
ATEX, FM, CSA, TIIS, NEPSIIn preparation: IEC
IP67 (NEMA 4X)
92 F Inline
Sensor F
Compact AluRemote Alu
Two-line LCD/ push buttons
Operation via FieldCare
Compact: –40 to +70 °CRemote: –40 to +80 °C
12 to 36 V DC, EEx i: 12 to 30 V DC, EEx d: 15 to 36 V DC
Electronical connections are galvanically isolateted from each other – 4 to 20 mA,
pulse HART®, PFM, PROFIBUS PALiquids:
For 2 or 3 beams ±0.5% For 4 beams ±0.3%
ATEX, FM, CSAIn preparation: IEC, TIIS, NEPSI
IP67 (NEMA 4X)
92 T Portable
Sensor U, W, P Remote Alu Graphic display/
push buttons –10 to +45 °C Rechargeable battery: 100 to 240 V AC (47 to 63 Hz)
Electronical connections are galvanically isolateted from each other 4 to 20 mA 4 to 20 mA Serial interface
(RS 232)Onsite: <2.0%
Under reference conditions: ±0.5% – IP50
72 Standard
Sensor F, W
Compact AluRemote Alu
Two-line LCD/ push buttons
Operation via FieldCare
Compact: –40 to +70 °CRemote: –40 to +80 °C
Two-wire12 to 36 V DC,
EEx i: 12 to 30 V DC, EEx d: 15 to 36 V DC
Electronical connections are galvanically isolateted from each other – 4 to 20 mA,
pulse/statusHART®, PFM, PROFIBUS PA,
FOUNDATION FieldbusLiquids: ±0.75% Volume
Gases/steam: ±1.0% VolumeATEX, FM, CSA, TIIS, NEPSI
In preparation: IECIP67
(NEMA 4X)
73 With flow computer
Sensor F, W
Compact AluRemote Alu
Two-line LCD/ push buttons
Operation via FieldCare
Compact: –40 to +70 °CRemote: –40 to +80 °C
Two-wire 12 to 36 V DC,
EEx i: 12 to 30 V DC, EEx d: 15 to 36 V DC
Electronical connections are galvanically isolateted from each other
Pressure using HART®, PROFIBUS PA, FF
Temperature, density using HART®
4 to 20 mA, pulse/status,
frequency
HART®, PFM, PROFIBUS PA,FOUNDATION Fieldbus
Liquids: ±0.75% Volume Gases/steam: ±1.0% Volume Saturated steam: ±1.7% Mass
ATEX, FM, CSA, TIIS, NEPSIIn preparation: IEC
IP67 (NEMA 4X)
65 Standard
Sensor I, F
Compact AluRemote Alu
Two-line, backlit LCD/ push buttons
Operation via FieldCare
–20 to +60 °COptional: –40 to +60 °C
85 to 260 V AC20 to 55 V AC or 16 to 62 V DC
Electronical connections are galvanically isolateted from each other –
4 to 20 mA, pulse/frequency,
relays, status
HART®, PROFIBUS DP, MODBUS® RS485 Gases: ±1.5% ATEX, FM, CSA, NEPSI
In preparation: IEC, TIIS IP67
Deltabar SPMD 70/75
Deltatop PMD 70/75
Deltaset PMD 70/75
Compact SSRemote SS
Compact AluRemote Alu
4-digits LCD/ push buttons
Operation via FieldCare–40 to +85 °C Two-wire
11.5 to 45 V DC No galvanic isolation – 4 to 20 mA HART®, PROFIBUS PA, FOUNDATION Fieldbus
Orifice plate DPO: ±1.0%
ATEX, IEC, FM, CSA, TIIS, NEPSI IP65
Pitot tube DPP: ±1.5%
Compact version Compact SS Operation via FieldCare –20 to +60 °C 20 to 30 V DC Electronical connections are galvanically isolateted from each other – Pulse, status – ±0.5%
Optional: ±0.25% – IP67(NEMA 4X)
Compact version Compact SS Operation via FieldCare –20 to +60 °C 20 to 30 V DC Electronical connections are galvanically isolateted from each other – Pulse, status – ±0.15% ATEX II 3G IP67
(NEMA 4X)
Compact version Compact Alu Operation via FieldCare –40 to +60 °C 20 to 28 V AC 20 to 30 V DC
Electronical connections are galvanically isolateted from each other – Pulse/frequency, status MODBUS® RS485 ±0.5% ATEX, FM, CSA
In preparation: NEPSI, TIIS, IECIP67
(NEMA 4X)
Compact version Compact Alu Operation via FieldCare –40 to +60 °C 20 to 28 V AC 20 to 30 V DC
Electronical connections are galvanically isolateted from each other – Pulse/frequency, status MODBUS® RS485 ±0.2% ATEX, FM, CSA
In preparation: NEPSI, TIIS, IECIP67
(NEMA 4X)
Compact version Compact Alu LED –20 to +60 °C 20 to 30 V DC No galvanic isolation – 4 to 20 mA, relays – ±2.0% FM, CSA Div. 2 IP66(NEMA 4X)
Compact version Compact SSOne-line, backlit LCD –
Color change on faults, LED’s for status indication, push buttons
–40 to +85 °C 10 to 30 V DC (reverse polarity protection) No galvanic isolation –
1× or 2 × PNP switching outputs,optional: 1× PNP switching output and 1× 4 to 20 mA output, active
Programmable by a PC(ReadWin® 2000) ±2.0% –
IP65/IP66, dependent
on electrical connection
Compact version Compact SSOne-line, backlit LCD –
Color change on faults, LED’s for status indication, push buttons
–40 to +85 °C 10 to 30 V DC (reverse polarity protection) No galvanic isolation –
1× or 2 × PNP switching outputs,optional: 1× PNP switching output and 1× 4 to 20 mA output, active
Programmable by a PC(ReadWin® 2000) ±2.0% –
IP65/IP66, dependent
on electrical connection
Compact version Compact SSCompact Alu
LED, 4-digits, LCD (option)DIL switches
–10 to +65 °C without LCD 18 to 30 V AC/DC No galvanic isolation – Relays – ±5.0% – IP66
Compact version Compact SSCompact Alu
LED, 4-digits, LCD (option) push buttons
–10 to +65 °C without LCD 18 to 30 V AC/DC No galvanic isolation – 4 to 20 mA, pulse,
open collector – ±5.0% – IP66
Flow measurement
Subj
ect t
o m
odifi
catio
n
Flow measurement OEM