proof of concept trial for “az modification” of

1
Proof of Concept Trial for “AZ Modification” of Supraglottic Airways Rafi Avitsian MD, Andrew Zura MD Department of General Anesthesiology Cleveland Clinic , Cleveland OH ABSTRACT The AZ modification provides an airway management tool that can be used in a supraglottic and endotracheal manner interchangeably. We present a clinical trial where the concept of an exchange between two ventilation methods using this modification is investigated. Modification of two supraglottic airway platforms confirmed ease of placement and exchange between the two ventilation modes, allowing true staged extubation and intubation.. BACKGROUND A novel AZ- airway concept introduced at the 2016 DAS annual meeting was well received encouraging a proof of concept study. The device allows interchangeable supraglottic (SG) and endotracheal (ET) ventilation, provides rescue supraglottic airway device (SAD), ability to convert to endotracheal tube (ETT), and safer smooth or true staged extubation from an existing ETT without use of an exchange catheter or interrupting ventilation. This allows “tracheal rest” and “staged extubation” in difficult airways or intensive care setting1-3. The unique design includes a removable (R-piece) tube replaced by an ETT, enabling ventilation in a SG or ET mode depending on ETT depth. The goals were to: 1) Implement AZ- airway concept on commercially available SAD platforms 2) Test claimed capabilities on manikin 3) Clinical trial on patients METHODS The study was conducted after Institutional Review Board ethics approval. Two SAD models were chosen, the Ambu AuraGain as an inflatable cuff, and the i-gel as a non-inflatable cuff platform and were altered according to a 3D printed prototype. These concepts were verified: a) SG ventilation with R-piece, b) R-piece to ETT change for SG ventilation, c) Alternating ET and SG ventilation with existing ETT without interruption, d) Removal of SAD while keeping ETT in place e) Conversion of existing ETT to SAD In the departmental simulation center structurally altered SADs were tested on manikins. Fiberoptic bronchoscopy and videolaryngoscopy confirmed optimal positioning. Next, the concept was tested on seven consented patients with sterile manner of SAD alteration in the operating room. An anesthesiology staff, resident or nurse anesthetist did the initial insertion to test ease of insertion. Positioning was confirmed by fiberoptic bronchoscopy. REFERENCES RESULTS 1) Biro P, Priebe H. Staged Extubation Strategy: Is an Airway Exchange Catheter the Answer? Anasthe Analg 2007; 105(5): 1182-5 2) Mort TC. Continuous airway access for the difficult extubation: the efficacy of the airway exchange catheter. 2007; 105(5):1357-62 3) Perello-Cerda L, Fabregas N, Lopez AM, et al. ProSeal Laryngeal Mask Airway Attenuates Systemic and Cerebral Hemodynamic Response During Awakening of Neurosurgical Patients: A Randomized Clinical Trial. J Neurosurg Anesthesiol. 2015; 27(3):194-202 R- Adapter Currently there is no single device that can be used as both a supraglottic and infraglottic airway, interchangeably, without compromising ventilation. The concept of staged extubation as currently mentioned in some literature using an exchange catheter is a misnomer since it involves a period of unprotected airway with only the exchange catheter in place. Other scenarios of use for a device that can provide both types of ventilation include: Failed intubation rescue with a need for intubation Unclear if intubation will be necessary Need for Smooth Intubation Need for Smooth Extubation (e.g. in neurosurgery or ENT) Intraoperative change from intubation to supraglottic ventilation or vice versa. Staged Extubation and tracheal rest in ICU patients Unstable cervical spines Intubation by low trained personnel – field We were interested to see if the concept could be implemented in currently commercially available SAD which included either a cuff based or gel based mask portion. The manikin results were promising encouraging a trial in live patients. Our results demonstrated that in both SAD models AZ alteration can be easily performed. The claims were tested in simulation center using both platforms of SADs successfully. The clinical trial showed same results as that of manikins for ease of insertion, functionality, possibility of interchanging SAD to ETT and staged extubation. Rafi Avitsian MD FASA, Andrew Zura MD Cleveland Clinic, Cleveland OH, USA DISCUSSION At the time of this report, 7 out of 10 IRB approved patients were included. The “AZ modification” was achieved by a longitudinal opening in the body of the SAD shaft transforming it to a non- cylindrical, incomplete tube. An R-piece was made using the proximal portion of a larger ETT. The table shows the patient population, primary diagnosis, the supraglottic and R-piece used, and the operator. The initial placement was mentioned as easy and no different in effort or time compared to a conventional SAD. In all cases an ETT was passed without difficulty. In one case (case 6) only supraglottic part was tested and ETT placement was not performed. No Pt. SAD Rpiece ETT Operator 1 60 y/o 110 kg F i-gel 4 ETT 9.0 7.5 CRNA 2 55 y/o 102 kg M i-gel 5 ETT 9.5 7.5 CRNA 3 49 y/o 97 kg M i-gel 4 ETT 9.5 7.5 Staff 4 69 y/o 77kg F AuraGain 4 ETT 9.5 7.5 Resident 5 64 y/o 70kg M AuraGain 4 ETT 9.5 8.5 Resident 6 70 y/o 70 kg M AuraGain 4 ETT 8.5 ---- Resident 7 52 y/o 81 kg M i-gel 4 Ett 9.5 7.5 Resident

Upload: others

Post on 24-Dec-2021

2 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Proof of Concept Trial for “AZ Modification” of

Printing:This poster is 48” wide by 36” high. It’s designed to be printed on a large

Customizing the Content:The placeholders in this formatted for you. placeholders to add text, or click an icon to add a table, chart, SmartArt graphic, picture or multimedia file.

Tfrom text, just click the Bullets button on the Home tab.

If you need more placeholders for titles, make a copy of what you need and drag it into place. PowerPoint’s Smart Guides will help you align it with everything else.

Want to use your own pictures instead of ours? No problem! Just rightChange Picture. Maintain the proportion of pictures as you resize by dragging a corner.

Proof of Concept Trial for “AZ Modification” of Supraglottic Airways

Rafi Avitsian MD, Andrew Zura MDDepartment of General Anesthesiology

Cleveland Clinic , Cleveland OH

ABSTRACT

The AZ modification provides an airway management tool that can beused in a supraglottic and endotracheal manner interchangeably. Wepresent a clinical trial where the concept of an exchange between twoventilation methods using this modification is investigated.Modification of two supraglottic airway platforms confirmed ease ofplacement and exchange between the two ventilation modes,allowing true staged extubation and intubation..

BACKGROUNDA novel AZ- airway concept introduced at the 2016 DAS annualmeeting was well received encouraging a proof of concept study. Thedevice allows interchangeable supraglottic (SG) and endotracheal(ET) ventilation, provides rescue supraglottic airway device (SAD),ability to convert to endotracheal tube (ETT), and safer smooth ortrue staged extubation from an existing ETT without use of anexchange catheter or interrupting ventilation. This allows “trachealrest” and “staged extubation” in difficult airways or intensive caresetting1-3. The unique design includes a removable (R-piece) tubereplaced by an ETT, enabling ventilation in a SG or ET modedepending on ETT depth. The goals were to:

1) Implement AZ- airway concept on commercially available SADplatforms

2) Test claimed capabilities on manikin

3) Clinical trial on patients

METHODS

The study was conducted after Institutional Review Board ethics

approval. Two SAD models were chosen, the Ambu AuraGain as an

inflatable cuff, and the i-gel as a non-inflatable cuff platform and

were altered according to a 3D printed prototype. These concepts

were verified:

a) SG ventilation with R-piece,

b) R-piece to ETT change for SG ventilation,

c) Alternating ET and SG ventilation with existing ETT without

interruption,

d) Removal of SAD while keeping ETT in place

e) Conversion of existing ETT to SAD

In the departmental simulation center structurally altered SADs were

tested on manikins. Fiberoptic bronchoscopy and videolaryngoscopy

confirmed optimal positioning. Next, the concept was tested on

seven consented patients with sterile manner of SAD alteration in the

operating room. An anesthesiology staff, resident or nurse

anesthetist did the initial insertion to test ease of insertion.

Positioning was confirmed by fiberoptic bronchoscopy.

REFERENCES

RESULTS

1) Biro P, Priebe H. Staged Extubation Strategy: Is an Airway Exchange

Catheter the Answer? Anasthe Analg 2007; 105(5): 1182-5

2) Mort TC. Continuous airway access for the difficult extubation: the

efficacy of the airway exchange catheter. 2007; 105(5):1357-62

3) Perello-Cerda L, Fabregas N, Lopez AM, et al. ProSeal Laryngeal Mask

Airway Attenuates Systemic and Cerebral Hemodynamic Response

During Awakening of Neurosurgical Patients: A Randomized Clinical

Trial. J Neurosurg Anesthesiol. 2015; 27(3):194-202

R-

Adapter

Currently there is no single device that can be used as both asupraglottic and infraglottic airway, interchangeably, withoutcompromising ventilation. The concept of staged extubation ascurrently mentioned in some literature using an exchange catheter isa misnomer since it involves a period of unprotected airway with onlythe exchange catheter in place. Other scenarios of use for a devicethat can provide both types of ventilation include:• Failed intubation rescue with a need for intubation• Unclear if intubation will be necessary• Need for Smooth Intubation• Need for Smooth Extubation (e.g. in neurosurgery or ENT)• Intraoperative change from intubation to supraglottic ventilation

or vice versa.• Staged Extubation and tracheal rest in ICU patients• Unstable cervical spines• Intubation by low trained personnel – fieldWe were interested to see if the concept could be implemented incurrently commercially available SAD which included either a cuffbased or gel based mask portion. The manikin results were promisingencouraging a trial in live patients.Our results demonstrated that in both SAD models AZ alteration canbe easily performed. The claims were tested in simulation centerusing both platforms of SADs successfully. The clinical trial showedsame results as that of manikins for ease of insertion, functionality,possibility of interchanging SAD to ETT and staged extubation.

Rafi Avitsian MD FASA, Andrew Zura MDCleveland Clinic, Cleveland OH, USA

DISCUSSION

At the time of this report, 7 out of 10 IRB approved patients wereincluded. The “AZ modification” was achieved by a longitudinalopening in the body of the SAD shaft transforming it to a non-cylindrical, incomplete tube. An R-piece was made using the proximalportion of a larger ETT. The table shows the patient population,primary diagnosis, the supraglottic and R-piece used, and theoperator. The initial placement was mentioned as easy and nodifferent in effort or time compared to a conventional SAD. In allcases an ETT was passed without difficulty. In one case (case 6) onlysupraglottic part was tested and ETT placement was not performed.

No Pt. SAD Rpiece ETT Operator

1 60 y/o 110 kg F i-gel 4 ETT 9.0 7.5 CRNA

2 55 y/o 102 kg M i-gel 5 ETT 9.5 7.5 CRNA

3 49 y/o 97 kg M i-gel 4 ETT 9.5 7.5 Staff

4 69 y/o 77kg F AuraGain 4 ETT 9.5 7.5 Resident

5 64 y/o 70kg M AuraGain 4 ETT 9.5 8.5 Resident

6 70 y/o 70 kg M AuraGain 4 ETT 8.5 ---- Resident

7 52 y/o 81 kg M i-gel 4 Ett 9.5 7.5 Resident