pressure measuring system for wind-induced …

76
PRESSURE MEASURING SYSTEM FOR WIND-INDUCED PRESSURE ON BUILDING SURFACES by HOWARD HO-TAK NG, B.S. in C.E. A THESIS IN CIVIL ENGINEERING Submitted to the Graduate Faculty of Texas Tech University in Partial Fulfillment of the Requirements for the Degree of MASTER OF SCIENCE IN CIVIL ENGINEERING Approved Accepted August, 1988

Upload: others

Post on 08-Nov-2021

8 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

PRESSURE MEASURING SYSTEM FOR WIND-INDUCED

PRESSURE ON BUILDING SURFACES

by

HOWARD HO-TAK NG, B.S. in C.E.

A THESIS

IN

CIVIL ENGINEERING

Submitted to the Graduate Faculty of Texas Tech University in

Partial Fulfillment of the Requirements for

the Degree of

MASTER OF SCIENCE

IN

CIVIL ENGINEERING

Approved

Accepted

August, 1988

Page 2: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

/'r

ACKNOWLEDGEMENTS

I would like to express my sincere gratitude to Dr. Kishor C.

Mehta, my coimittee chairman, for his valuable guidance, advice, and

encouragement. I would also like to thank Dr. James R. McDonald and

Dr. W. Pennington Vann for their constructive suggestions. I wish to

acknowledge the National Science Foundation for providing financial

support under the Grant No. CES8611601 for this research.

Deepest appreciation confer to my teaitmates and colleagues in the

NSF Field Experiment project, C. V. Chok, B. D. Lakas, M. L. Levitan

and S. H. Liew, for their encouragement and help during the course of

my research. Special thanks go out to iny beloved one, Christina K.

M. Wan, for her patience and moral support. Mostly, I wish to thank

my parents and family, for their constant support and enduring love.

It is to them that I dedicate this thesis.

11

Page 3: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

CONTENTS

ACKNOWLEDGEMENTS 11

TABLES

FIGURES VI

CHAPTER

INTRODUCTION

Objectives

II. REVIEW OF PREVIOUS EXPERIMENTS

Aylesbury Experiment

Mobile Home Experiment

Single-story Houses Experiment

Solar Collector Experiment

Agricultural and Horticultural Buildings Experiment

III. PRESSURE MEASURING SYSTEM

Pressure Transducer

Pressure Tap

Pressure Transmitting System

Reference Pressure

Data Acquisition S>rstem

IV. ^/ERIFICATION AND SENSITIVrTY OF PRESSURE MEASURING SYSTEM

1

7

9

9

11

12

13

14

16

17

17

19

25

27

30

Static Calibration

D3mamic Calibration

30

32

111

Page 4: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

Examination of Reference Pressure Fluctuation 47

Temperature Effect on Transducer 58

V. CONCLUSIONS AND RECOMMENDATIONS 64

Conclusions 64

Recommendations 65

REFERENCES 68

IV

Page 5: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

TABLES

4.1 Information of Dynamic Calibration 39

4.2 Analysis of Reference Pressure Data 56

Page 6: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

FIGURES

1.1 30 X 45 X 13 ft Test Building and the 160 ft 3 High Meteorological Tower

1.2 Map of Area Surrounding Test Building 4

3.1 Pressure Measuring System 18

3.2 Pressure Tap Assembly 20

3.3 Pressure Transmitting Sj^tem 21

3.4 Valve Position under Normal Operating Condition 23

3.5 Valve Position under Zero Drift Test Condition 23

3.6 "Reservoir" for Collecting Water in Pressure Tubing 24

3.7 Reference Pressure Box 26

3.8 Data Acquisition System Arrangement 28

4.1 Static Calibration Kit 31

4.2 Dynamic Calibration Kit 34

4.3 Dynamic Calibration Setup 35

4.4 Dynamic Calibration of Wall Transducer 37

4.5 Dynamic Calibration of Roof Transducer 38

4.6 Frequency Response of Wall Transducer Setup With 41 Short Tube (Test 1)

4.7 Frequency Response of Roof Transducer Setup With 42 Short Tube (Test 2)

4.8 Frequency Response of Wall Transducer Set-up With 43 25 ft Tube (Test 3)

4.9 Frequency Response of Roof Transducer Setup With 44 25 ft Tube (Test 4)

4. 10 Frequency Response of Wall Transducer Setup With 45 50 ft Tube (Test 5)

vi

Page 7: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

4. 11 Frequency Response of Roof Transducer Setup With 46 50 ft Tube (Test 6)

4. 12 Setup for Assessment of Reference Pressure Fluctuation 49

4. 13 Time History of the Pressure Data taken at Zero 51 Differential Pressure Condition

4. 14 Time Histories of Reference Pressure and Wind Speed 52 (Data Set 1)

4. 15 Time Histories of Reference Pressure and Wind Speed 53 (Data Set 2)

4. 16 Time Histories of Reference Pressure and Wind Speed 54 (Data Set 3)

4.17 Relationship of Mean Wind Speed and Reference 57 Pressure Standard Deviation

4. 18 Deviation of Pressure due to Temperature Variance 60 on one Transducer

4. 19 Percentage Deviation of Pressure due to Temperature 61 Variance (-lOOC and 400C)

4.20 Percentage Deviation of Pressure due to Temperature 62 Variance (SO C and 600C)

Vll

Page 8: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

CHAPTER I

INTRODUCTION

Wind is the movement of air; it induces pressures on building

surfaces. As wind travels around and over the building, it ac­

celerates and creates outward-acting pressures across all surfaces

except the windward surface. Moreover, the air flow streamlines

separate at the sharp comers such as wall comers, eaves, roof

ridges, and roof comers. Relatively low pressures occur downstream

of these streamline separation points, causing locally severe,

outward-acting pressures.

Every year a significant amount of damage to buildings is caused

by wind-induced pressures. Estimated economic loss of wind-induced

building damage in this country may reach eight billion dollars (1978

dollars) by the year 2000 (Wiggins, 1978). It is also recognized

that most of the damage to the structures and most of the injuries to

people occur in low-rise ordinary buildings. Hence, understanding

the nature of wind-induced pressures on buildings will help civil

engineers to design safer as well as more economical buildings.

Assessment of wind-induced pressures on building surfaces is

generally obtained through wind tunnel tests. Although wind tunnel

results contribute significant information, they depend on simulation

of wind and use of small-scale models of buildings. It is difficult

to simulate the wind profile for a low-rise building in wind tunnel

because the building model is very close to the bottom boundairy

Page 9: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

layer. In view of the uncertainties of wind tunnel results, field

measurement of wind-induced pressures on building surfaces are

needed. Field measured results can provide base line data for wind

tunnel simulation, and may help to advance the wind tunnel

technology.

Recently a project is undeirtaken at Texas Tech University to

acquire wind data and building surface pressure data in the field.

The major objective of this project is to obtain a reliable data base

for external pressures on building surfaces. The anticipated wind-

induced pressures are in the range of ±.10 psf. However, pressures as

high as +20 psf may occur in a strong windstorm.

A 30 X 45 X 13 ft test building for pressure measurement and a

160 ft high meteorological tower for wind data are seen in Figure

1. 1. Figure 1.2 shows a map of the field site and the surrounding

area. The field site for the test building is located in a flat open

terrain. Low-rise residential areas are at least one-half mile from

the test building to the south and to the west.

The test building is constructed on a concrete pad. It is a

commercially available metal building sitting on a rigid fraras

undercarriage. Four wheels, one at each comer, are part of the

undercarriage. The wheels rest on an embedded circular track that is

flush with the surface of the concrete pad. Hydraulic jacks can

raise the building to support it on four wheels. The building

Page 10: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

3

Figure 1. 1 30 x 45 x 13 ft Test Building and tl>5 160 ft High Meteorological Tower

Page 11: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

Figure 1.2 Map of Area Surrounding Test Building

Page 12: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

can be rotated on track to obtain desired angle of attack of wind.

This unique feature of the building permits data to be collected from

any wind-approach angle.

The test building has smooth exterior skin. Self tapping

screws, at 1 ft on center, hold the exterior flat stock to the

corrugated steel panels. The heads of the screws protrude about 3/8

in. from the surface. Except for these protrusions, the exterior

surface of the building is smooth. A rubber gasket skirt is attached

to the bottom of the building to provide a seal between the building

and the concrete pad. The roof of the building is essentially flat;

it has a slope of 2 in. in 10 ft for drainage.

The meteorological tower is located approximately 150 ft west of

the building. The tower is guyed at heights of 70 and 130 ft. Wind

speed measuring instruments (anemometers) are placed at four levels:

160, 70, 33, and 13 ft. Wind direction, temperature, barometric

pressure and relative humidity measuring instruments are also placed

on the tower.

A fixed 12 X 12 X 8 ft data acquisition room is located on

concrete pad inside the test building. This room is constructed of

reinforced concrete masonry walls and a cast-in-place concrete roof.

The roof of the room has a 12 in. diameter hole. Data signal cables

and tubing of the pressure instruments can go through the hole into

the room.

Page 13: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

Wind-induced pressures on the test building surfaces are

measured through pressure tapping holes using differential pressure

transducers. The differential pressure transducer essentially

measures pressure difference between two pressure sources. Hence,

each pressure transducer needs a reference pressure.

The components of pressure measuring system include pressure tap

assembly, tubes, valves, differential pressure transducer, and

electronic data acquisition system. These components are readily

available, however the pressure measuring system is not available

conmercially. The system has to be custom assembled and its perfor­

mance has to be verified. In addition, the reference pressure system

has to be designed specially for this project.

In view of the custom assembly of the pressure measuring system

and low magnitude of pressures to be measured (in the range of 10

psf), it is necessary to conduct experiments to provide confidence in

measured pressures. Specific experiments need to be conducted for

assessing fluctuation in reference pressure, checking dynamic

response of pressure transmitting system and evaluating sensitivity

of pressure transducers to the environmental temperature encountered

in the field. The thrust of this research is to custom assemble the

pressure measuring system and to verify its perfomnnce.

Qb.iectiygs

The general objective of this study is to investigate the

accuracy and sensitivity of the building surface pressure measuring

Page 14: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

7

system. Specific objectives of this study include the following:

(1) To assemble the pressure measuring system including the

pressure tap, pressure transmitting lines, differential

pressure transducer, reference pressure system, and the data

acquisition system.

(2) To perform static calibration of the pressure measuring

system.

(3) To perform dynamic calibration for frequency'response of the

pressure transmitting system.

(4) To assess the fluctuations in the reference pressure.

(5) To evaluate the sensitivity of the pressure transducers to

the environmental temperatures encountered in the field.

These objectives are achieved through assembly of the pressure

measuring system on the test building and through specific experi­

ments in the laboratory and in the field as appropriate. The

procedures developed and the results obtained in this study are valid

for any building surface pressure measuring system in the field.

Previous full scale measurements of wind pressure are studied in

order to gain information for the present project. Experiments that

measured building or structure surface pressures are briefly reviewed

in chapter II. The project facility and special feature of the

pressure measuring system for the test building are presented in

chapter III. CJalibration and sensitivity assessment are accomplished

through a series of experiments. Details of the experiments setup

Page 15: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

8

and results are provided in chapter IV. CJhapter V contains the con­

clusion of this study and reconroendations for future development of

the field work.

Page 16: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

CHAPTER II

REVIEW OF PREVIOUS EXPERIMENTS

A number of full scale experiments in the field have been

conducted to measure wind effects on a variety of structures. The

structures in the experiments include solar collectors, agricultural

buildings, chimneys, tall buildings, bridges, and transmission line

systems. Among these experiments, some used differential pressure

transducejTs to measure wind-induced pressures on structure surfaces.

A review of major aspects of the instrumentation used in these

experiments is presented below. These aspects include the pressure

measuring instrument, pressure transmitting setup, reference pressure

provision, and calibration of the transducer.

Avlesburv Experiment

Among different projects on full scale measurement of wind-

induced pressures, perhaps the best known is the project of an

experimental building at Aylesbury in the U.K. (Eaton and Mayne,

1974). The project involved measurement of wind and wind-induced

pressures on surfaces of a two-stoiy residential building in the

field.

The pressure transducers used to measure surface pressures were

developed at the Building Research Station in England. The sensitive

part of these transducers was a 4 in. diameter pressure plate mounted

flush with the external surface of the building. Underneath the

Page 17: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

10

pressure plate were three cantilever bars on which resistance strain

gages were mounted. The interior of the transducer was vented to a

reference pressure via a nozzle on the back of the transducer body.

Wind-induced pressures were detennined from strain gage measurements

(Eaton and Mayne, 1971). Since it is an external pressure plate,

neither a pressure tapping hole nor tubing is required to transmit

the wind-induced pressures to the transducer. The project did not

have any transducer to measure internal pressure of the building.

Reference pressure for the transducer was provided by running

venting tubes of 0.2 in. bore from the transducers to a reference

pressure manhole. The distance between the manhole and the building

is not given in the reference. The manhole was sealed by a lid

except for a 1 in. diameter hole in the center. The manhole lid was

flush with the ground; grass nearby was kept short so that the

pressure inside the manhole represented static atmospheric pressure.

Fluctuations in the reference pressure are not mentioned in the

reference. It is not known whether there were any experiments

conducted to assess the fluctuations in the reference pressure.

Static calibration of each transducer was obtained by applying a

known dead weight to each transducer and adjusting the gain of the

signal conditioning equipment to maintain a standard voltage level at

the input to the recorders. The frequency response of the transducer

was checked up to about 200 Hz. The lowest resonant frequency was

found to be 70 Hz (Eaton and Mayne, 1971). At the beginning of each

record, a solenoid valve connected a zeroing port located besides

Page 18: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

11

each transducer to the back of the transducer. The concept was to

connect the active side and passive side of the transducer together

so that the differential pressure would be zerx3. Then the recorded

voltage would be the zero offset which could be subtracted from the

recorded voltages during analysis of the test data.

In the master's thesis by Vickery (1984) on analysis of the

Aylesbury data, it was concluded that there were significant problems

associated with apparent shifts in the reference static pressure.

Tbs author inferred that on some of the wall locations the common

backing tubes which supplied the reference pressure to each pressure

transducer were themselves leaking to the inside of the building.

Mobile Home ExTPeriment

Full scale wind pressure measurements were collected on a mobile

home located on the campus of National Bureau of Standards (NES) in

Gaithesburg, Maryland (Marshall, 1977). Pressures on the exterior

surfaces of the mobile home and the internal pressure were measured

by Validyne variable reluctance differential pressure transducers.

The pressure tap hole on the surface of the building was 3/8 in.

diameter. The flexible tubing that connected the pressure tap with

the pressure transducer was 1/8 in. inside diameter and 4 in. long.

Although the pressure taps were designed to operate under wind-driven

rain conditions, on occasion, a meniscus formed in the pressure lir.e

causing a pronounced shift in the pressure transducer recordings.

Page 19: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

12

Reference pressure was provided through a manifold system of

3/16 in. inside diameter flexible tubing connecting the pressure

transducer to an ambient pressure probe which was mounted on a mast

at the height of 33 ft. This prcbe was developed at the NES. The

probe registered true ambient pressure for all wind directions in the

horizontal and accoranodated vertical angles of attack up to ±.10

degrees. Estimated from a site plan, the distance between the probe

and the mobile home was 80 ft.

Details of static calibration are not provided in the reference.

E)ynamic calibration of the tap-tube-transducer arrangement indicated

a good response up to 40 Hz. A solenoid valve was placed with each

transducer. The valve normally transmitted the reference pressure tc.

the passive side of the pressure transducers. When actuated, the

solenoid valve connected the passive side of the pressure transducers

to the active pressure lines. This arrangement provided the pressure

transducer electrical offset voltage for a zero pressure differen­

tial.

Single-storv Houses Experiment

Full scale measurements of building surface pressures on single

story houses were collected in Philippines (Marshall, 1977).

Pressures were measured with flush-diaphragm differential pressure

transducers developed at the Building Research Station in England.

These transducers were similar to the ones used in the Aylesbury

project (Eaton and Mayne, 1974). Hence, the setup and the calibra-

Page 20: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

13

tion procedures were the same as in the Aylesbury project. The

transducers were referenced to an ambient pressure probe which was

similar to the one used in the mobile home project by Marshall

(1977). The probe was mounted next to a propeller-vane anemometer at

the 33 ft (10 m) level on a tower. Literature did not provide

details on calibration and zero offset control. The problem of bats

biting the pressure transmitting tubing has been mentioned during

personal discussions with Dr. Majrshall. The collection of data was

affected because of bat damage to the tubing.

Solar Collector Experiment

An investigation of wind loads on solar collector was conducted

in the field in Virginia (Tieleman, 1980). In this study Validyne

reluctance differential pressure transducers were used to measure

pressures on the solar collector surfaces. The literature did not

provide the dimensions of the pressure taps. The tubing used in the

test was 0.2 in. diameter.

Reference pressure for the differential pressure transducers was

obtained from an ambient pressure probe mounted on a mast at the

height of 37 ft. The probe was similar to the one used in the mobile

home project (Marshall, 1977).

Each pressure transducer was subjected to both a static and a

dynamic calibration. For static calibration, a water manometer was

used to apply a static pressure to the system and adjustments were

made to the amplifier-demodulator associated with the transducer to

Page 21: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

14

provide an output of +10 volts for ±0. 1 pei. The dynamic calibration

was carried out using a calibration rig provided by the NES. No

resonance was found below 50 Hz in the tubing with 0.2 in. diameter.

Each pressure transducer was fitted with a solenoid valve in the

reference pressure line similar to the arrangement used in the mobile

home project (Marshall, 1977).

Agricultural and Horticultural Buildings

Experiment

Full scale study of wind pressures on surfaces of agricultural

buildings is conducted by Hoxey (1983) in England. Pressure trans­

ducers used in this project were differential pressure transducer of

capacitance bridge micromanometer type. Surface pressures were

transmitted from a 0.38 in. diameter tapping hole via 0.27 in. inside

diameter tube to the transducer at a central measurement and record­

ing station. Tube lengths ranged up to 98 ft.

The free stream atmospheric static pressure was used as a

reference pressure for transducers. In the initial stages of the

project, reference pressui^ was derived from the pressure within a

container with a perforated top, sunk flush into level ground ap­

proximately 50 ft from the building. For later measurements an

ambient pressure probe based on the NBS design was used. The ambient

pressure probe was mounted on a mast upwind at building height. The

reference pressure was transmitted to the transducers via 0. 27 in.

inside diameter plastic tubing that was 132 ft long.

Page 22: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

15

Static calibration of the transducers was performed regularly in

the laboratory by applying a constant pressure to the transducer

using a Betz manometer. The dynamic response of the tube system was

limited by the use of long tubing. The natural frequency of the tube

system induced an amplitude gain above unity due to pneumatic

resonance. To reduce this gain to unity a pneumatic filter, consist­

ing of a flow restrictor (small bore stainless steel tube) and a

volume of air, was installed between the pressure tap and the

transducer. The dimensions of the restrictor tube and the volume of

air used in the system are not mentioned in the literature.

A zero pressure signal was generated for a few seconds at the

beginning and end of each four minute recording duration to obtain a

zero drift record of the transducers, amplifier or data acquisition

tape recorder. The zero drift signal was obtained by connecting the

two sides of the transducer diaphragm chamber by means of solenoid

valves that could be energized automatically by the control unit.

The review of these field experiments published in the litera­

ture gives insight to the installation and the use of pressure

measuring instruments. This insight also assisted in assembling the

building surface pressure measuring system. In addition, it helped

in planning the experiments to verify and check sensitivity of the

pressure measuring system used in the current project.

Page 23: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

CHAPTER III

PRESSURE MEASURING SYSTEM

The present project requires measurement of building surface

pressures at as many as 20 points without unduly altering the smooth

surface of the building. This pressure measurement is accomplished

by using 3/8 in. diameter holes in the metal skin and measuring

pressures with differential pressure transducers. The magnitude of

the pressure to be measured is relatively small, in the range of ±10

psf for wind speeds of 50 raph. According to the manufacturer's

specifications, the transducers have an accuracy of +0.25% of the

full scale. For the transducers with +0.2 psi capacity, the accuracy

is +0. 14 psf; for the transducers with +0.32 psi capacity, the

accuracy is +0.23 psf. Based on the accuracy of the transducers and

the anticipated wind pressure range of ±10 psf, the desirable

accuracy of the system is ±.0.5 psf.

In addition, it is desired to measure instantaneous fluctuations

in pressures. To assure measurement of instantaneous fluctuations,

the measurements must be taken at a sampling rate of 10 Hz or higher.

This low magnitude of pressures and high sampling rate govern the

assembly of pressure measuring system. Several components make up

the pressure measuring system; these components include pressure

transducers, pressure tap, flexible tubing, reference pressure and

16

Page 24: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

17

data acquisition system. This chapter describes the assembly and

specifications of these components. A typical pressure measuring

system assembly is shown in Figure 3. 1.

Pressure Tr;^n5^Hnnf^r

Two types of differential pressure transducers are used in the

project, Validyne variable reluctance transducers and Omega transduc­

ers. In initial stages, thirteen Validyne transducers are used; this

report relates only to these Validjme transducers. Nine transducers

have the capacity of +0.2 psi (model DP103-22N-7S4H) and the other

four transducers have the capacity of +0.32 psi (model DP103-24N-

7S4H). The heart of the pressure transducer is a diaphragm clamped by

two inductance coils. This diaphragm can be replaced with different

capacity diaphragms, thus changing the transducer capacity. Under

wind pressure, the diaphragm deflects and changes the inductance of

the coils. A corresponding AC voltage signal is sent to a demodula­

tor to be rectified and filtered into ±IC voltage which represents

the pressure polarity and magnitude. The transducer output is ±.5

volts for ±0.2 psi and +8 volts for +0.32 psi.

Pressure Tap

Pressure on the wall or the roof surfaces are measured through a

3/8 in. diameter hole. The hole and the copper tube assembly is

termed pressure tap. This dimension of 3/8 in. diameter prevents

rain water from forming a meniscus across the hole and the copper

Page 25: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

1 c

Cable to Data Acquisition System

Figure 3.1 Pressure Measuring System

Page 26: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

19

tube. The copper tube is connected to a flexible tubing which runs

to the pressure transducer (see Figure 3. 1). In designing the

pressure tap, several different ideas were considered. In the early

stages, soldering was used to connect the copper tubing to the

building surface to act as a pressure tap. However, due to the

significant temperature change in the field, the solder failed after

few days. Attempts to attach copper pressure tap assembly to the

building skin with epoxy were unsuccessful.

The final design of the pressure tap consists of a brass

compression connector (3/8 x 1/4 Pipe Thread) connected to a 3/8 in.

diameter copper tubing. The pressure tap assembly is shown in Figure

3.2. Installation of the pressure tap is accomplished by creating a

recessed indentation in the exterior cladding skin where the pressure

tap is to be located. The brass fitting is fastened to the building

surface with two 1/8 in. thick nuts, one on each side of the metal

building skin. The outside nut seats in the recessed indentation

which make it essentially flush with the outside building surface.

The dent is filled with automobile body-repair putty and sanded

smooth. The two nuts provide a positive connection to the building

skin and the putty prevents leakage around the hole as well as

maintains a smooth exterior surface.

Pressure Transmitting System

A sketch of the pressure transmitting system is shown in Figui e

3.3. Approximately 4 in. of 3/8 in. inside diameter flexible plastic

Page 27: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

Sleeve

3/8 In. diameter copper tube

Fitting

Exterior Surface

corrugated Steel Panel

3/8 in. 1.0. flexible tube

/8 in. thickness

Figure 3.2 Pressure Tap Assembly

Page 28: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

21

Wind Induced Pressure

4 In .

y

/^

/

Reference

Pressure (-35 f t )

Pressure Tap

3/8 jn. I.D.

3/16 In. 1.0.

Active Line

10 In.

3:

3/16 In. I.O.

Reference Line

Pressure Transducer

Tee connector

Figure 3.3 Pressure Transmitting System

Page 29: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

tubing extends from the pressure tap to the three-way valve. About

10 in. of 3/16 in. inside diameter flexible plastic tubing is

connected from the three-way valve to the transducer. Choice of the

dimensions of 4 in. and 10 in. is arbitrary. These lengths of tubes

provide sufficient room to assemble the system. Approximately 35 ft

of 3/16 in. inside diameter flexible plastic pressure tubing is used

to connect the transducer to the reference pressure. Under normal

operating condition, the three-way valve separates the active line

and reference line for the transducer as shown in Figure 3.4. During

the zero drift test, the function of the three-way valve is to

connect reference pressure to both sides of the transducer as shown

in Figure 3.5. Three-way valve is operated manually. When the

three-way valve connects reference pressure to both sides of the

pressure transducer, it will have a zero pressure difference, thus

yield a zero drift. The drift of the transducer voltage may be due

to the temperature variance or electrical power drift in data

acquisition system.

Owing to the fact that the ends of the pressure taps are exposed

to rain, water may collect in the tubing. To eliminate this pos­

sibility, the roof pressure transmitting system has a "reservoir" as

shown in Figure 3.6. The water collected in the reservoir can be

removed by opening the drain plug. The flexible plastic tubing

connected to taps on the wall surfaces is tilted about 20 degree

upwai'd to prevent water from collecting in the active pressLire line.

Page 30: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

Exterior Wall

Pressure Transducer

Active

Line

< >

Reference

Line

Three way valve

Figure 3.4 Valve Position under Normal Operating (Condition

Exterior Wall

Pressure Transducer

Active

Line

€> Reference

Line

Three way valve

Figure 3.5 Valve Position under Zero Drift Test (Condition

Page 31: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

:4

cer

Figure 3.6 "Reservoir" for Collecting Water in Pressure Tubing

Page 32: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

The pressure transmitting tubing and connections are checked for

leakage. The system was pressurized with a hand pump. Soapy water

was applied to the connections to indicate leakage. No visible

leakage was observed.

Reference Pressure

Reference pressure is obtained from an underground wooden box,

located 50 ft west of the test building. Figure 3.7 shows the

picture of the box and a schematic drawing of the reference pressure

system. The box is 4 x 5 x 3 ft deep and is covered by an 8 x 8 ft

plywood lid with a 1/2 in. diameter hole at the center. The lid is

flush with the ground level. Soil around the box has been graded to

provide a relatively smooth surrounding so that the turbulence which

would affect the pressure inside the reference pressure box is

minimized.

The reference pressure is transmitted through an 8 in. diameter

PVC pipe. The pipe is buried 2 ft below ground in a trench. The end

of the pipe inside the box is covered by a perforated cap. The pipe

runs underneath the concrete pad and enters the data acquisition room

through the floor. The end of the PVC pipe in the data acquisition

room is sealed with a cap. Three sixteenth inch inside dianeter

flexible plastic tubes run from the cap of the pipe to the reference

side of each transducer. The flexible plastic tubes vary in length

up to 35 ft.

Page 33: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

26

Data Acqusttton Room

Test Building

Reference Pressure Box

-TT^^r^r-

Reference Pressure ' P^Pe ( 8 in. diameter)

Figure 3.7 Reference Pressure Box

Page 34: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

27

Data Acquisition System

Components of the data acquisition system include carrier

demodulators, expansion multiplexers (EXP-16) and a microcomputer

with an analog to digital (A/D) converter (DAS-8). The carrier

demodulator is unique to pressure data acquisition; other components

are also part of wind data acquisition system. Figure 3.8 shows the

a schematic of the instrument connection.

The microcomputer used is an IBM PC XT with an internal 20

megabyte hard drive and full color graphics capabilities, enhanced

with a 12 MHz 80286 main processor and a 80287 math coprocessor

supplied by Personal Computer Support Group. Analog voltages from

the instruments are converted to digital form by a MetraByte DAS-8

12-bit A/D converter. Although the DAS-8 has only 8 input channels,

the unit is expanded to handle 64 input channels through MetraBjrte

EXP-16 expansion mltiplexers. The DAS-8 has an input range of ±5

volts.

Wires from the transducer pass through an opening in the roof of

the data acquisition room and are connected to the module case. The

module case houses 13 carrier demodulators, one for each pressure

transducer. The module case provides a power supply for the trans­

ducer while the demodulator acts as a signal conditioning unit t^ the

transducer. A switch-selectable active low pass filter on each

carrier demodulator can be set at 1, 5, 10, 50, 200 and 1000 Hz. A

potentiometer on each carrier demodulator is used during the static

calibration of each transducer to relate voltages to pressures. Ail

Page 35: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

28

I

4-> . ^ .O

e\j ^ w

^ 4-> • ^

1 . 0) 4J L. 01 > c o u

p _

« 4-> •r" cn

*^" S O

1— o x + * o Q .

^* CO ^^

O l o ^ a c <C

• — e x

L. t/t O ^ « C-r- I > lO 4 .»0 .

•.- Q . » — X C X 3 U J

ft)'

o

o «/) s

u 0> 01

^* "^ T3 1 -O lO

L. 3 l/l (/I 0» i_

Q . U 01

0) U C 3 >>T3

"O «/» •»- c r— <0 IQ i -> >—

+3

.<D

s

S CO c o • H +3 •H •r-t 3

.a s a 00

crj

b

Page 36: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

2y

the instruments and the computer are connected to a common ground

source in order to eliminate potential differences which might induce

erroneous output voltages.

Page 37: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

CHAPTER IV

VERIFICATION AND SENSITIVTTY OF

PRESSURE MEASURING SYSTEM

Four experiments have been conducted to accomplish the verifica­

tion and the sensitivity assessment of the pressure measuring system.

The first three experiments included static calibration, dynamic

calibration, and examination of reference pressure fluctuation. The

last experiment focused on the sensitivity assessment of the tempera­

ture effect on transducers. This chapter presents experimental

procedures and the results of each of these experiments.

Static Calibration

The purpose of the static calibration is to relate pressure

experienced by the transducer to the voltage output. The data

acquisition system records the voltage output of the transducer. The

calibration is accomplished by using an Omega portable calibration

kit (Model PCL200C) which consists of a hand pump and a built-in

digital meter. Pressure applied through the hand pump is registered

by the digital meter. Figure 4. 1 shows the calibration kit.

(Calibration is accomplished at the field site on a calm day

(wind speed of less than 15 mph) to minimize pressure fluctuations

inside the test building. After installing the transducer at ap­

propriate location in the test building, the calibration kit is

connected to the active side of the transducer. The reference

30

Page 38: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

' < * ^ I : ) t \

MH I-

Figure 4.1 Static Calibration Kit

Page 39: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

pressure for the transducer is the pressure inside the building.

Through the hand pump of the calibration kit, a selected pressure is

applied to the transducer. The digital meter of the kit registers

the applied pressure. The potentiometer on the carrier demodulator

corresponding to the transducer being calibrated is adjusted to a

voltage output related to the pressure. The procedure is repeated

several times to obtain a desired voltage. The voltage output is set

at +5 V for +0.2 psi and +8 V for +0.32 psi for the transducers with

respective capacities. This static calibration should be checked

periodically to guarantee proper operation of the transducers and

respective potentiometers should be adjusted as appropriate.

Dynamic Calibration

The purpose of the dynamic calibration is to assess distortions

in measured pressures due to fluid wave resonance or friction between

the air and the tube wall. The pressure at the pressure tap may be

amplified or attenuated after going through the pressure transmitting

system of tubes and the three-way valve. The length of the tubes can

distort the pressure measured by the transducers. If feasible, it is

desirable to place the transducer in the data acquisition room under

a temperature controlled condition. On the other hand, the distor­

tions in the measured pressure is the least with short length of

tubes. In order to verify the feasibility of putting transducer in

the data acquisition room, dynamic calibration with tube lengths :f

25 ft and 50 ft long tubing are conducted. The dynamic calibraticri

Page 40: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

33

requires application of pressure pulses at several different frequen­

cies. Pressures are measured with two transducers, one is a refer­

ence transducers with short tube length and the other is a calibrat­

ing transducer with desired tube length. ( Dmparison of the recorded

pressures by the two transducers provide information on the distor­

tions, if any, in the measured pressures.

The calibration kit (TriTec Model AT31A) shown in Figure 4.2 has

the ability to provide pressure pulses at rates of up to 200 Hz. The

kit is driven by an electrical signal generator. Pressure source for

the kit comes from an air tank connected to the calibration kit. By

adjusting the electrical signal generator, the frequency of the

pressure pulses is controlled. The pressure pulses are sent in a

sinusoidal form. Magnitude of the pressure pulses can be adjusted by

the pressure regulator on the air tank, though it's value is not

critical to the test.

Figure 4.3 shows the experimental setup. A pressure transducer

is connected to the calibration kit with a short tube. This trans­

ducer is termed reference transducer; it is assumed that the refer­

ence transducer records the pressure without distortion. The trans­

ducer at the other end of the pressure transmitting system is termed

calibrating transducer. The differences in the measured pressures

from the two transducers indicate the influence of the length of the

tubing and the three-way valve.

Six calibration tests are conducted. Tests 1 and 2 use the wall

and roof setups as they are built with short tubes (see Figures 3. 1

Page 41: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

To Air Tank

To Electrical Signal Generator

•-. r"'-'-

\ To Pressure Transducer

Figure 4.2 Dynamic Calibration Kit

Page 42: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

35

Air Tank

Tube length varied

Signal Generator

TriTec Dynamic Calibration Kit

Reference Transducer

Three way C valve

Calibrating Transducer

Closed

Figure 4.3 Dynamic CJalibration Setup

Page 43: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

36

and 3.6). Tests 3 and 4 are related to 25 ft long tubes, and Tests 5

and 6 are with 50 ft long tubes. Connections and dimensions of

tubing for these six tests are shown in Figure 4.4 for wall setup and

Figure 4.5 for roof setup. The reference side of the reference

transducer and the calibrating transducer are exposed to the interior

of the test building. Since the experiments are conducted on calm

days, the pressure inside the test building are expected to be

constant. At the beginning of each test, the air is allowed to flow

through the system for a 4-minute duration to attain a stable condi­

tion. The voltage outputs of the reference transducer and the

calibrating transducer are recorded for 15 second duration. Each

test includes collection of data at several pressure pulse frequen­

cies. Different transducers have been used to check the repeatabil­

ity of the experiment. They show similar results.

Table 4. 1 shows the pressure pulse frequencies, data collecting

frequencies and low pass filter frequencies for each test. The data

collecting frequency is at least four times the pressure pulse

frequency with a minimum of 120 Hz. In each test, the mean and

standard deviation values of records from both the reference trans­

ducer and the calibrating transducer are obtained. Ratios of these

values of calibrating transducer to reference transducer are deter­

mined. These ratios provide the information on distortion of the

measured values. If the ratio is unity, both transducers measure the

same pressure. If the ratio is less than unity, the signal is

attenuated through the pressure transmitting system. If the ratio is

Page 44: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

37

To Dynamic Calibration Kit

'i 2-

u Three way valve

Reference Transducer

01 stance "a" In

Experiment 1 Is 4 in.

Experiment 3 Is 25 ft

Experiment 5 Is 50 ft

10"

Calibrating Transducer

- \

U

Figure 4. 4 Dynamic Calibration of Wall Transducer

Page 45: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

38

To Dynamic Calibration Kit

7"

Reference Transducer

Drainage Plug

To Three way valve and Calibrating Transducer as in Figure 4.4

10"

Distance "a" in Experiment 2 is 4 in. Experiment 4 Is 25 f t Experiment 6 Is 50 ft

Figure 4.5 Dynamic Calibration of Roof Transducer

Page 46: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

39

TABLE 4.1

Information of Dynamic Calibration

Experiment Tubing Pressure Pulse Data CkDllecting Low Pass Filter Number Length Frequency Frequency Frequency

(Hz) (Hz) (Hz)

1 4 in. 1 120 50 10 120 50 30 120 50 40 160 50 50 200 200 60 240 200 100 400 200

2 4 in . 1 120 10 5 120 10

10 120 50 30 120 50 60 240 200 70 280 200

100 400 200

3 and 4 25 f t 1 120 10 3 120 10 5 120 10 6 120 10 7 120 10 8 120 10

10 120 50 30 120 50 60 240 200

100 400 200

5 and 6 50 f t 1 120 10 3 120 10 5 120 10

10 120 50 20 120 50 30 120 50 60 240 200

100 400 200

Page 47: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

40

greater than one, the pressure is amplified by resonance in the

pressure transmitting system. Plots of these ratios are presented in

Figures 4.6 to 4.11 for each of the test.

The ratios of mean values for all experiments (short tube, 25 ft

and 50 ft) remain one. This indicates that the mean values of

pressure are neither attenuated nor amplified because of the tube

length. However, fluctuations of pressures are affected as evidenced

by the ratios of standard deviation (see Figures 4.6 through 4.11).

When the tube lengths are short (Tests 1 and 2), the ratios of

standard deviation values for wall and roof setups are more than one

for frequencies higher than 20 Hz (see Figures 4.6 and 4.7). The

setup of a three-way valve with short tube length apparently genera­

tes secondary waves at frequency in the neighborhood of 30 Hz to

increase standard deviation (see Figures 4.6 and 4.7) values recorded

by the calibrating transducer. This suggests that reliable measure­

ments can be obtained with this setup for frequencies up to 20 Hz.

It gives confidence to the pressure data recorded at the sampling

rate of less than 20 PIz in the field.

For tube length of 25 ft (Tests 3 and 4), the ratios of standard

deviation are significantly amplified in frequency range of 3 to 10

Hz (see Figures 4.8 and 4.9). Beyond frequency of 10 Hz, the

measured signals are attenuated. The tube length of 25 ft can prov­

ide reliable results only if data is recorded at rate of 2 Hz or

less.

Page 48: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

41

N

>v O

0)

3 +>

<D U 3

a 2

£ C CO

i l 0) E -

s -aii 8 ^

(X x-S Si | w

l i s [zi :z CO TT

Cb

sanLBA aouaaa^ay/SuL^eaqLLBO ^o sott ey

Page 49: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

4

N

o; :i-l.|.>^!/!.l. . {.'.I .;.:. J '.,4,^';., . i-^f-j^ |.Li..| .. pf- _ . . 1 . .^. .^j^^o. ...;,_

r r^ • / T -; ^ - f T - r t - •+-r I ' + ^ i O j •

i UJLJ ' _^ : U . J . __^.-it{-i. .: ; „ .

•"1 . 'S»i*\ - } : " " § * • " ! r U f - l \ - L "- •• • i!L • • • •

' 1 / • . -" .. - J . , • '.

J. K i M ' ' _ _ • - o < . '

1 4 1 - 7 . • • : . ^ •: : ; i I ..__ ......^ !\ . . . . . . .. 1 ! . . : :_;

- 'T \ 1 ' M i _ , . 1 «-y A . . . * _ _ O n 1.1 —< .

i •( • 1 . . ] • ; ,

m2, f^-l 1 : i : : 1 ! ; 1 1 ! : ; ;

1 X ' ' '

-• ^ ^ - • - - M ;

- : - • - • ' • - . . • M ;

— 1. . . 11 i —• ' • • 1 ! " 1 | _ ^ ' ' : i

r I ' " "f ' 1 1 ' I 1 f 1 . ; '

1 H-h4-i r -i - H - i H--^- 1- • • 1 1 ' ! M ' ' ' '• 1 1 1 ! 1 i l l J i • • :

I . - r f - • - _L. J4 1 ''

j i '"' 1 . l . i . l 1 .

1 ' ! 1 I - - : _ -•:" -

1 [ • - 4 - - - - — - - ^ - ^ H .

- r ++ -Hf ^ - •

, -• - - : . U • i 1

• - o

. . .iJ . .• . - ci • O . - <ai , , , $ ? - • - • , : = - ^ _ : . : : .

. . : . • - • - . ' i ! ' ; j ,

CZ4 2

m

santeA a^u^a^:^^y/6uL:^BJqLLe^ ^o soi^ey

Page 50: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

43

N

>> O c

CO

B

3 CO

O 10

ai 3 r=

CO

CJU 3 :

00

CD

lA

r- ^ o

sanLBA aouaaa^ay/6uL:iBaqLLB3 ^o soL^By

Page 51: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

44

N

U c: 3 O" Qi

I <D

U

E-

Is o cn

a " a.s en 3

OS •p

C IT) I CM

2.1 tzi S

03

Cb

. - •- o sanLBA aouaaa^^ay/SuLt^BaqLLBO : o soL^By

Page 52: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

45

N

>» U

0) 3 cr

I <D

U 0)

B

(0 -

o to

1) E-

a^ tn 3

So

l i s

9

ta

santBA aoua-iaj.ay/6iJL:;BaqLiB3 ^o soit^sy

Page 53: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

46

N

O c 3 o-<U

O . I - T T T - - - O r « - '

- ri-|T---tt N ; 1 ! 1. ' 1 1 l-i::.:-l

- - 1 — - H ^ ' * . . , 1 1 - ' : . . . . . 1 T

r

i . r . . .

'-It •~ ..^ .. . _p - l - - - o

1

- --T [ I - -

i , i

•• i • • /

m» 1

T - \ f. .• J - .

. . . .^| • -— 1 — | - i = < » - | —

1 • "

1

1

:TM^lt-.M4^ M.-,-,

\J"'IT''1"1"' 4 t - ^ • j V - ; - - - " : 1 i ; I ;

(1! \ - ^ 0 ' ^ ! '

» I- "^ I ' l l

\ - - • • • - • • - • , , • ' •

. . . - -|-|— '-.\_ - • ". : M

.. . ^ _ 0 I ' ; .

1 " M 1 ' i - 1 - 1 1 1 1 • 1

] ' hi ilM "eiJ / oi i M '

P / . . : _ " • ' : ' T i I !

. . . 1 ' '

- • • i 1

d -xiL"--''r -''"

W-t^-'m :.-.-i.::£i3:,--7 1 ; i • ; ( - [•• M M

: - - . 1 1 _ : - ^ - 1 - — l i . . , - . - .

^ i " . - " .M . ^ . . - 1 , 1

. . . . • - 1 : 1 : .

T-- - I© Mil • T- , . ; i

• • • • : ] _ MM'

" ' • • " " •" " - - - i ; 1

\ ! ! 1 1 1 , i • 1

c! a. I I I '

: : . 4 •: •= | ! M

1 1 1 1 1 ' ' 1 1 1 ' •—Ji—1—L1—

-p 0)

03

U <D

B 13

a 2

05 CO

«H +^

o tn 0) E -

a.s in 3

o I-

•H Czi Z

?:

9

Cb

t/> m

sanLBA aouaua^ay/BuitiBaqLLBO ^o souBy

Page 54: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

47

The tube length of 50 ft (Tests 5 and 6) attenuate signal for

frequencies above 5 Hz (see Figures 4. 10 and 4. 11). This tube length

can be used only if mean pressures are desired. Measured values of

fluctuating pressures are undependable with 50 ft tubing.

In conclusion, the regular wall and roof setups with short tube

show an undistorted dynamic response up to frequency of 20 Hz. Above

20 Hz, the pressure is either amplified or attenuated. In the case

of 25 ft and 50 ft long tubing, the results show a significant

amplification or attenuation at frequencies above 3 Hz. Hence, the

idea of putting transducer in the data acquisition room by using long

tubing to connect the pressure tap to the transducer is not feasible.

Examination of Reference Pressure Fluctuation

The purpose of the examination of reference pressure fluctuation

is to ascertain the extent of fluctuation in the reference pressure

during a period of time. As described in chapter III, the reference

pressure is obtained from an underground box located approximately 50

ft from the test building. Ideally, the reference pressure should be

steady and without fluctuations for the duration of recording period.

Fluctuation in reference pressure could be due to electrical noise in

the data acquisition system and because of turbulence in the box

caused by strong winds. Fluctuations in the reference pressure due

to these causes are ascertained through specific tests.

The test setup consists of a transducer, a constant pressure

source and the data acquisition system. A constant static pressure

Page 55: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

48

is applied to the transducer through a Dwyer inclined manometer. The

experimental setup is shown in Figure 4.12. The inclined manometer

has two openings A and B (see Figure 4. 12). At first, a hand pump is

connected to opening A and is used to apply pressure to the side A of

the manometer. Then the opening B is closed. The hand pump is

disconnected and the transducer is connected to the side A. Opening

B is released so that the higher level of liquid on side B is the

pressure exerted on the transducer. The transducer is connected to

the manometer on one side and the reference pressure PVC pipe located

in the data acquisition room on the other side. After all the

connections are made, the opening B of the manometer is closed to

prevent possible fluctuation in pressure inside the data acquisition

room from affecting the constant pressure. It was noted during

initial tests that the constant pressure source was very sensitive to

temperature change. Touching the manometer or flexible tubing by

hand caused a drift in the pressure recording. To prevent this

drift, a larger volume of air was provided by using 100 ft long

tubing between the manometer and the transducer. In addition, heat

sources in the data acquisition room are minimized by turning off the

lights and evacuating the room. Extra precautions were taken to keep

temiperature in the room as stable as possible during the tests.

The first test was conducted to ascertain electrical noise in

the data acquisition system. In this test both sides of the trans­

ducer were connected to two tubes from the PVC reference pressure

pipe; manometer was not used. Since the reference pressiure was

Page 56: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

Tube Length of About 35 ft

u Transducer

Tube Length of TOO f t

Inclined Manometer

Reference Pressure Pipe

Figure 4. 12 Setup for Assessn^nt cf Reference Pressure Fluctuati-.^n

Page 57: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

50

acting on both sides of the differential transducer, the pressure

recording should be zero. A time history of pressure data recorded

at the rate of 10 Hz for 10 minute duration on a calm day is shown in

Figure 4. 13. Even though the figure indicates a lot of fluctuation,

the peek-to-peak value is 0.2 psf. It is felt that fluctuations of

this magnitude can be attributed to noise in data acquisition system.

It may be possible to reduce this noise level by determining the

frequencies of the noise and by providing appropriate filters. No

attempt was made to determine specific frequencies of electrical

noise. This test, however, does indicate that electrical noise

fluctuations in pressure data are less than 0. 2 psf.

To ascertain fluctuations in reference pressure during strong

winds three sets of data, each 10-minute duration, were recorded over

a period of time. For each data set a constant pressure is applied

to the transducer through the manometer. Time histories of wind

speed measured at 13 ft and the reference pressure measured by the

transducer are shown in Figures 4. 14, 4. 15 and 4. 16. The data was

recorded at the rate of 10 Hz. As indicated in the figures, the gust

wind speed varied from 18 mph to 50 mph during the recording periods.

Two questions are raised with regard to reference pressure

fluctuations: (1) does mean value of reference pressure vary with

mean wind speed? and (2) are the fluctuations in reference pressure

related to wind speed? To respond to these two questions, several

segments of 50 second duration records are selected from the three

sets of data such that a range of mean wind speeds are obtained.

Page 58: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

51

o o so

o o

CM

O

^ Q

«n o o

(O o d

•* o d

<N

o (J

CV4 O

O 1

• ^

o o 1

\o o o 1

i n

o o 1

^ o 1

f N

"" 6 1

2

(0

o - o

CM

_,_ o

in

£ t—

a§ -H

(D +J

3 ^

S<3 £ <u d) 3

^ en ^ 8 0 fi

k (0

me

His

tc

ffere

nti

E-H Q

CO

1X4

( > $ d ) d j n s s d J d

Page 59: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

C O

a.4

Reference Pressure ril«-rW18

Wind speed at 13 ft

c «.

Figure 4.14 Time Histories of Reference Pressure and Wind Speed (Data Set 1)

Page 60: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

53

Reference Pressure ri«-raf19

Wind speed at 13 ft

Figure 4. 15 Time Histories of Reference Pressure and Wind Speed (Data Set 2)

Page 61: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

54

Reference Pressure fU-rml2a

Wind speed at 13 f l

r a.

V

c

ril«-ra>»

Figure 4.16 Tims Histories of Reference Pressure and Wind Speed (Data Set 3)

Page 62: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

55

Four segments are selected from each data set; the mean and the

standard deviation of wind speeds as well as accompanying pressures

are calculated. These values are shown in Table 4.2. The mean wind

speeds of these segments vary from 21 mph to 38 raph. The constant

pressure for each data set was fixed independently, hence the data of

three sets cannot be compared directly. Admittedly, there is a

paucity of data. However, from the data shown in Table 4.2, no

specific trend is evident between mean reference pressure and mean

wind speed.

Mean wind speed versus s1:andard deviation of reference pressure

for all segments are shown in Figure 4. 17. The data is largely

scattered. Figure 4. 17 shows that there may be a trend; as mean wind

speed increases, the standard deviation of pressure also increases.

This suggests that there may be larger fluctuations in reference

pressure as wind speed increases.

Dr. Richard Marshall (1988) of the National Bureau of Standards,

after reviewing Figures 4. 14 to 4. 16 and the reference pressure

measurement procedure, suggests that most of the fluctuations in the

reference pressure may be due to use of captive air for the constant

applied pressure. He feels that slightest variation in pressure and

temperature surrounding the captive air (100 ft of flexible tubin.g

and the manometer) would record fluctuations indicated in Figures

4. 14 to 4. 16. He reconinends additional tests using damping device

such as yam of 3 to 4 in. length inside the flexible tuoe. Actual

Page 63: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

56

Data Set

1

»

2

3

Analysis of

TABLE 4.2

Reference Pressure Data

Se lec ted Segment (sec)

50-100

110-160

250-300

400-450

50-100

300-350

320-370

500-550

50-100

150-200

300-350

500-550

Wind Speed(mph)

Mean Std.dev.

27.95

33.96

21.09

36.37

33.51

32.27

27.76

37.68

28.63

29.35

35.76

28.86

6.40

4.69

2.09

5.29

6 .31

4.42

4.99

5.03

4.57

5.32

5.46

4.75

Pressure (psf)

Mean Std. dev.

7 .81

7.71

7.93

7.84

8. 16

8.19

8.26

8. 16

8.53

8.63

8.37

8.51

0.10

0. 11

0.07

0.19

0. 13

0. 14

0.09

0. 13

0. 12

0. 11

0. 10

0. 10

Page 64: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

57

(D

Cx4

(>sd) -ABp'Pis » jnss»Jd »3UdJ«jey

Page 65: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

58

length and type of yam will have to be determined by trial and error

procedure.

In conclusion, time histories of reference pressure in Figures

4. 14, 4. 15 and 4. 16 show that peak-to-peak variation of reference

pressure in 10 minute duration is as high as 1. 0 psf. Magnitude of

the fluctuation is higher than desirable for wind induced pressure

measurement. A portion of the fluctuation (0.2 psf) can be at­

tributed to electrical noise in dai:a acquisition system. The

remaining variation could be due to turbulence in the reference

pressure box or due to use of captive air for the constant pressure.

Based on these observations, the reference pressure system needs

additional tests and a possible damping device.

Temperature Effect on Transducer

The purpose of examining the temperature effect on transducer is

to assess whether the temperature would affect the output of a trans­

ducer, and to what extent. The data on temperature effect provides a

measure of sensitivity of pressure measuring system. Five trans­

ducers are selected randomly among thirteen transducers for the

experiment, three transducers with the capacity of ±0.2 psi and two

transducers with the capacity of ±0.32 psi. These transducers are

calibrated for static pressure at 30 C and are tested at four other

temperatures: -lOOC, 400C, 500C, and 600C. To achieve 400C tc 600C,

the transducers are put in an oven and t stad. For -lOOC, the

transducers are placed in a freezer and t stad. By -jsing the t.>iT>g3

Page 66: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

59

calibration kit mentioned earlier, static pressure is applied to the

transducer and the output voltage is recorded. Data is obtained at

several pressures from +20 psf to -20 psf during each temperature

test. At the beginning and the end of each test, the voltage at zero

differential pressure is recorded. The average voltage of these two

zero differential pressure is subtracted from the voltage recorded to

yield the adjusted voltage. Analysis is based on the adjusted

voltage.

Linear regression analysis is performed on the data of each

temperature test. The results show a very good degree of linearity

with a correlation coefficient above 99%. Deviations occurring at

different pressures for each specific temperature are obtained by

calculating the differences between the linear regression prediction

and the expected output. The expected output is based on the linear

relationship of pressure to voltage in static calibration such as 5 V

to 0.2 psi.

Deviations recording in psf for the full range of pressures of

-20 psf to +20 psf at the four temperatures for one of the transduc­

ers are shown in Figure 4.18. The figure shows a consistent linear

relationship between deviation and pressure. The closer to the

maximum capacity of the transducer, the higher the deviation. The

same observation is found in other tests. Figures 4. 19 .and 4.20 show

the deviation as percentage of full scale occurring at different

pressures for the three transducers with ±0.2 psi capacity. In the

graph, the negative pressures have a higher percentage deviation than

Page 67: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

60

o CM

O

_ o

a 0 o in

at a.

3 m m 4)

O e O

U o 7

o CM I

0 g

•H

I I (D

( ; s d ) UOI^DIAda

Page 68: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

Tennperature at —10^0

e e

I > 2

a. ta<anadue«r C

Temperature 40°C

3 tl

Figure 4. 19 Percentage Deviation of Pressure due to Temperature Variance (-lO' C and 40°C)

Page 69: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

62

a4

a c

Temperature 50 C

- 2 0

t i u i a d u t f A tiar«duc«r 8 (P^)

• r C

c o

I

7 C

• 0.

Temperature 60°C

- 2 0

bwsducar A

- 1 0

to«f«duc«r • (^)

tKai«duc«r C

Figure 4.20 Percentage Deviation of Pressure due to Temperature Variance (50°C and 60°C)

Page 70: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

63

the positive pressures. This higher deviation for negative pressures

may be due to the fact that the static calibration at 300C is done

for positive pressure only.

In conclusion, the deviation of positive pressure due to

temperature variance is less than 0.2% of full scale in the range of

0 to +10 psf. This deviation is within the manufacturer's specific­

ation of 0.25% of full scale. For negative pressure up to -10 psf,

the percentage deviation is as high as 0. 65% of full scale which is

equal to 0. 37 psf. The deviation of negative pressure of 0.37 psf is

within the desired accuracy of the pressure measurement which is 0. 5

psf. Figures 4. 19 and 4.20 do not shown any consistent trend of the

percentage deviation at a specific temperature. The two transducers

of 0.32 psi capacity show similar trend in temperature effect as the

ones with 0.2 psi capacity. These observations indicate that the

pressure measurements are not drastically affected in the temperature

at -lOOC to 600C when recorded pressure values are less than 10 psf.

If recorded pressures exceed 10 psf, correction may be desired

through calibration of individual transducer.

Page 71: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

CHAPTER V

CONCLUSIONS AND RECOMMENDATIONS

Three aspects about a building surface pressure measuring system

for wind-induced pressure measurements in the field are presented in

this thesis. These aspects are the assembly, the verification, and

the sensitivity assessment of the pressure measuring system. The

verification and the sensitivity assessment are accomplished through

a series of experiments. The assembly of the pressure measuring

system include installation of the pressure tap and construction of

the reference pressure system and assembly of the valve and flexible

tubing.

Conclusions

The conclusions of the experiments are as follow.

1. The leakage check using soapy water indicate that the

flexible tubing provides leak proof assembly of the pressure

measuring system.

2. Static calibration of each transducer with the accuracy of

+0.25% of full scale, as specified by the manufacturer, is

achieved using the Omega calibration kit and the poten­

tiometer of the data acquisition system.

3 The pressure transmitting system including pressure tap,

flexible plastic tubing of short length and three-way valve,

has an undistorted dynamic response of frequency up to 20 Hz.

This gives the confidence to the pressure data to be

64

Page 72: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

65

collected in the project at the rate of 10 Hz. However, the

use of 25 ft and 50 ft long tubing to connect the pressure

tap with the transducer shows a significant amplification and

attenuation in fluctuating pressures at frequencies above 3

Hz. This suggests that the idea of using long tubes so that

the transducer can be placed in the data acquisition room

should be abandoned, fean pressures do not show amplifica­

tion or attenuation for tube length up to 50 ft.

4. The reference pressure experiment show fluctuations of upto

1.0 psf (peak to peak) in 10-minute duration of strong winds.

Of this value, 0.2 psf is attributed to the electrical noise

in the data acquisition system. It is possible that captive

air used to apply constant pressure has aggravated fluctua­

tions higher than desired value of 0.5 psf; additional

experiments for reference pressure are necessary.

5. Environmental temperature range of -lOOC to 600C show devia­

tions in pressure measurement to be less than 0.2% of full

scale on postive pressure up to 10 psf. This is within the

manufacturer's specification of 0.25% of full scale. On the

other hand the deviation is 0.65% of full scale, which is

equal to 0.37 psf, for negtive pressure up to -10 psf. This

deviation occurring in the negative pressure is within the

desired accuracy of 0.5 psf of the pressure measurement.

Hence, pressure measurements are not drastically affect^

when recorded pressure values are less than 10 psf. For

Page 73: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

66

recorded pressures exceeding 10 psf, correcrtion may be

desired through calibration of individual transducer.

BgconiDendatdQDS.

The reconnmedations for future development are presented below.

1. It is desirable to further reduce the reference pressure

fluctuations as much as possible. This may be accomplished

by including different devic^es in the system such as restric­

tor, damper or long tubing. More experiments have to be

conducted to find a right combination of these devices.

2. Periodic static calibration of the transducer is recommended

to ensure the accuracy of the transducer. The temperature

inside the building should be noted every time the static

c^alibration is conducted.

3. Since the temperature has some effect on the transducer be­

havior, a temperature sensor is recornnended to be installed

to record the temperature inside the metal building at the

time the pressure data is collected.

4. To eliminate the amplification of the pressure data obtained

from a proposed system using 25 ft or 50 ft long plastic

tubing, pneumatic filters such as restrictor and volimie of

air may be installed in the pressure transmitting system.

Additional frequency response calibration experiments will be

necessary to assess accuracy of the recordings using long

tubes.

Page 74: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

b.'

5. The zero drift test of the transducer is accomplished by

manually switching a three way valve to connect the reference

pressure to both sides of the transducer. To eliminate this

inconvenience and inefficiency of the procedure, an automatic

electrically controlled solenoid valve should be used.

Page 75: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

REFERENCES

Eaton, K. J. , and Mayne, J. R. , 1971: "Strain measurements at the GPO Tower, London." BRE CP 29/71, Building Research Establish­ment, (3arston, Watford, England.

Eaton, K. J. , and Mayne, J. R. , 1974: "The Measurenent of Wind Pressures on Two-Storey Houses at Aylesbury. " BRE CP 70A4, Building Research Establishment, (5arston, Watford, England.

Hoxey, R., 1983: "A Rationalised Procedure for the Assessment of Wind Loads," thesis presented to the Cranfield Institute of Technology, National College of Agricultural Engineering, Silson, Bedford, England.

J. H. Wiggins Co. , 1978: "Building Losses from Natural Hazards: Yesterday, Today and Tomorrow. " J. H. Wiggins Cb. , Redondo Beach, CA.

Marshall, R. D. , 1975: "Wind Loads on Single-Story Houses." Proceed­ings, Second U.S. National Conference on Wind Engineering Research (June 1975, Fort Cbllins, Colorado), Colorado State University, Fort Ctollins, Colorado, pp. II1-4-1.

Marshall, R. D. , 1977: "The Measurement of Wind Loads on Full-Scale Mobile Homes." NBSIR 77-1289, National Bureau of Standards, Washington, DC.

Marshall, R. D. , 1988: "Oral Ck)mraunicationsM' National Bureau of Standards, Gaithesburg, MD.

Tieleman, H. W. , Akins, R. E. , and Sj^-ks, P. R. . 1980. ' An Inves­tigation of Wind Loads on Solar Collectors. /PI-E-80-1, Virginia Polytechnic Institute and State University, BlacksbLirg,

VA.

Virkerv P J 1984: "Wind Loads on the Aylesbury Experimental Hoi ise" A Comparison between Full-Scale and Two Different Scale Models," thesis presented to the Faculty of Graduate Studies, The University of Western Ontario, London, Canada.

68

Page 76: PRESSURE MEASURING SYSTEM FOR WIND-INDUCED …

PERMISSION TO COPY

In presenting this thesis in partial fulfillment of the

requirements for a master's degree at Texas Tech University, I agree

that the Library and my major department shall make it freely avail­

able for research purposes. Permission to copy this thesis for

scholarly purposes may be granted by the Director of the Library or

my major professor. It is understood that any copying or publication

of this thesis for financial ^ain shall not be allowed without my

further written permission and that any user may be liable for copy­

right infringement.

Disagree (Permission not granted) Agree (Permission granted)

Student's signature Student's signature

At..