pressure measurement 3

Upload: murali-dharan

Post on 06-Apr-2018

222 views

Category:

Documents


1 download

TRANSCRIPT

  • 8/3/2019 Pressure Measurement 3

    1/35

    DR.N.VISWANATHAN

  • 8/3/2019 Pressure Measurement 3

    2/35

    ` Recognizing the appearance of normal pressure

    waveforms is a prerequisite to identifying

    abnormalities that characterize certain

    cardiovascular disorders` Forward pressure and flow waves, as seen in the

    central aorta, are intrinsically identical in shape

    and timing.

  • 8/3/2019 Pressure Measurement 3

    3/35

  • 8/3/2019 Pressure Measurement 3

    4/35

    ` Forward pressure and flow waves, as seen in thecentral aorta, are intrinsically identical in shape andtiming.

    ` The pressure wave is modified by summation with a

    reflected pressure wave (Pbackward), and the resultantmeasured central aortic pressure wave shows asteady increase throughout ejection .

    ` The flow wave is also modified by summation with areflected flow wave (Fbackward), but because flow isdirectional, Fbackward reduces the magnitude of flow in

    late ejection, giving the characteristic Fmeasured as isseen with aortic flowmeters or Doppler signals

  • 8/3/2019 Pressure Measurement 3

    5/35

    ` The reflections for pressure occur from many sites

    within the arterial tree, but the major effective

    reflection site in humans appears to be the region

    of the terminal abdominal aorta.` Ascending aortic pressure is increased

    substantially within one beat after bilateral

    occlusion of the femoral arteries by external

    manual compression

  • 8/3/2019 Pressure Measurement 3

    6/35

  • 8/3/2019 Pressure Measurement 3

    7/35

    ` Ascending aortic (ASC Ao) pressure waveform in

    a patient before and after bilateral occlusion of the

    femoral arteries by external manual compression

    (left arrow).` On the right, high-speed recordings show that the

    major portion of the increase in pressure results

    from augmentation of the late (reflected) wave. .

  • 8/3/2019 Pressure Measurement 3

    8/35

    ` Factors that augment pressure wave reflections

    Vasoconstriction

    Heart failure

    HypertensionAortic or iliofemoral obstruction

    Valsalva maneuver after release

  • 8/3/2019 Pressure Measurement 3

    9/35

    ` Vasodilation

    Physiologic (e.g., fever)

    Pharmacologic (e.g., nitroglycerin,

    nitroprusside)Hypovolemia

    Hypotension

    Valsalva maneuvere strain phase

  • 8/3/2019 Pressure Measurement 3

    10/35

  • 8/3/2019 Pressure Measurement 3

    11/35

    ` Pressure reflections are diminished during the

    strain phase of the Valsalva maneuver with the

    result that pressure and flow waveforms become

    similar in appearance .

    ` After release of the Valsalva strain, reflected

    waves return and are exaggerated.

  • 8/3/2019 Pressure Measurement 3

    12/35

  • 8/3/2019 Pressure Measurement 3

    13/35

    ` The commonly noted late-peaking appearance ofcentral aortic and left ventricular pressure tracings inhumans referred to as the type A waveform pattern isa result of strong pressure reflections in late systole.

    ` In addition to the Valsalva maneuver, pressurereflections are diminished during hypovolemia,hypotension, and in response to a variety ofvasodilator agents .

    ` In these circumstances, the left ventricular and

    central aortic pressure waves exhibit a type C pattern

  • 8/3/2019 Pressure Measurement 3

    14/35

    ` A wedge pressure is obtained when an end-hole

    catheter is positioned in a designated blood vessel

    with its open end-hole facing a capillary bed, with

    no connecting vessels conducting flow into oraway from the designated blood vessel between

    the catheter tip and the capillary bed.

    ` A true wedge pressure can be measured only in

    the absence of flow

  • 8/3/2019 Pressure Measurement 3

    15/35

    ` In the absence of flow, pressure equilibrates

    across the capillary bed so that the catheter tip

    pressure is equal to that on the other side of the

    capillary bed.

  • 8/3/2019 Pressure Measurement 3

    16/35

    ` If minimal damping occurs between the cathetertip and the opposite side of the capillary bed thatis,

    ` 1.if there is a large, relatively dilated capillary bed,

    ` 2.if the precapillary arterioles and postcapillaryvenules are not constricted, and

    ` 3.if there is no other source of obstruction, such asthe presence of microthrombi

    phasic as well as mean pressure may betransmitted to the wedged catheter.

  • 8/3/2019 Pressure Measurement 3

    17/35

    ` Thus, an end-hole catheter wedged in a hepatic

    vein may be used to measure portal venous

    pressure.

    ` A catheter wedged in a distal pulmonary arterymeasures pulmonary venous pressure;

  • 8/3/2019 Pressure Measurement 3

    18/35

    In the absence of cor triatriatum or

    obstruction to pulmonary venous outflow, the

    pulmonary venous and left atrial pressures are

    equal, so that pulmonary artery wedgepressure may be used as a substitute for left

    atrial pressure

  • 8/3/2019 Pressure Measurement 3

    19/35

    ` Some common sources of error and artifact in

    clinical pressure measurement include

    1.Deterioration in frequency response,

    2.Catheter whip artifact,3. End pressure artifact,

    4.Catheter impact artifact,

    5.Systolic pressure amplification in the periphery,

    and6.Errors in zero level, balancing, and calibration.

  • 8/3/2019 Pressure Measurement 3

    20/35

    ` Although frequency response may be high and

    damping optimal during setup of the transducers,

    substantial deterioration in the characteristics can

    develop during the course of a catheterization study.

    ` Air bubbles may be introduced into the catheters,

    stopcocks, or tubing, or dissolved air may come out of

    the saline solution used to fill the transducer (just as

    dissolved air may come out of solution in a glass of

    water allowed to stand unperturbed for a few hours).

  • 8/3/2019 Pressure Measurement 3

    21/35

    ` Even the smallest air bubbles have a drastic effect

    on pressure measurement because they cause

    excessive damping and lower the natural

    frequency (by serving as an added compliance).

  • 8/3/2019 Pressure Measurement 3

    22/35

    Flushing out the catheter, manifold, and

    transducer dispels these small air bubbles and

    restores the frequency response of the pressuremeasurement system.

  • 8/3/2019 Pressure Measurement 3

    23/35

    Motion of the tip of the catheter within the heart

    and great vessels accelerates the fluid contained

    within the catheter.

    Such catheter whip artifacts may producesuperimposed waves of 10 mm Hg.

    Catheter whip artifacts are particularly common in

    tracings from the pulmonary arteries and are

    difficult to avoid.

  • 8/3/2019 Pressure Measurement 3

    24/35

  • 8/3/2019 Pressure Measurement 3

    25/35

    Catheter impact artifact is similar but not

    identical to catheter whip artifact. When a

    fluid-filled catheter is hit (e.g., by valves in

    the act of opening or closing or by the wallsof the ventricular chambers), a pressure

    transient is created.

  • 8/3/2019 Pressure Measurement 3

    26/35

    Catheter impact artifacts are common with pigtail

    catheters in the left ventricular chamber, where the

    terminal pigtail may be hit by the mitral valve

    leaflets as they open in early diastole.

  • 8/3/2019 Pressure Measurement 3

    27/35

    ` When radial, brachial, or femoral arterial

    pressures are measured and used to represent

    aortic pressure, it must be remembered that peak

    systolic pressure in these arteries may beconsiderably higher (e.g., by 20 to 50 mm Hg)

    than peak systolic pressure in the central aorta ,

    although mean arterial pressure will be the same

    or slightly lower.

  • 8/3/2019 Pressure Measurement 3

    28/35

  • 8/3/2019 Pressure Measurement 3

    29/35

    ` Pressure waveforms in a patient undergoing

    cardiac catheterization, Is a function of distance

    from the aortic valve (Ao V).

    `

    First vertical line marks onset of primary (forward)pressure wave, which occurs progressively later

    after the QRS complex with increasing distance

    from the aortic valve.

    ` Second vertical line marks onset of secondarypressure rise associated with the backward or

    reflected pressure wave

  • 8/3/2019 Pressure Measurement 3

    30/35

    ` The change in waveform of arterial pressure as it

    travels away from the heart is largely a

    consequence of reflected waves.

    `

    These waves, presumably reflected from the aorticbifurcation, arterial branch points, and small

    peripheral vessels, reinforce the peak and trough

    of the ante grade pressure waveform, causing

    amplification of the peak systolic and pulsepressures .

  • 8/3/2019 Pressure Measurement 3

    31/35

    ` The peripheral arterial systolic pressure may

    commonly appear to be 20 mm Hg higher than the

    left ventricular systolic pressure as a result of this

    phenomenon.

  • 8/3/2019 Pressure Measurement 3

    32/35

    ` Error in the quantitation of pressures because of

    improper zero reference is common.

    ` As mentioned earlier, in many laboratories thezero reference point is taken at the midchest with

    the patient supine, although some laboratories use

    a point 10 cm vertically up from the back or 5 cm

    vertically down from the sternal angle.

  • 8/3/2019 Pressure Measurement 3

    33/35

    ` All manometers must be zeroed at the same point

    , and the zero reference point should be changed

    if the patient's position is changed during the

    course of the study (e.g., if pillows are placed toprop up the patient).

  • 8/3/2019 Pressure Measurement 3

    34/35

    ` All manometers must be zeroed at the same

    point , and the zero reference point should

    Be changed if the patient's position is

    changed during the course of the study (e.g., ifpillows are placed to prop up the patient

  • 8/3/2019 Pressure Measurement 3

    35/35