Failure of dual radius hydroxyapatite-coated acetabular
cups
Many kinds of hydroxyapatite-coated cups were used, with favorable results in short term studies; it was supposed that its use could improve osteointegration of the cup, enhancing the stability and survivorship. The purpose of this study is to analyze the long term behavior of the hemispheric HA coated, Dual Radius Osteonics cup and to discuss the way of failure through the exam of the revised components and of both periacetabular and osteolysis tissue.
Materials and Methods
Between 1994 and 1997, at the Department of Orthopedic Sciences of the Insubria University, using the posterolateral approach, were implanted 276 Dual Radius Osteonics in 256 patients, with mean age of 63 years.
Results
At a mean follow-up of 10 years (range 812 years), 183 cups in 165 patients, were available for clinical and radiographical evaluation. 22 Cups among the 183 were revised (11%). The cause of revision was aseptic loosening in 17 cases, septic loosening in one case, periprosthetic fracture in another case, osteolysis and polyethylene wear in two cases and, finally, recurrent dislocations in the last one. In the remaining patients, mean HHS increased from a preoperative value of 50,15 to a postoperative value of 92,69. The mean polyethylene wear was 1,25 mm (min. 0,08, max. 3,9 mm), with a mean annual wear of 0,17 mm. The mean acetabular migration on the two axis was 1,6 mm and 1,8 mm. Peri-acetabular osteolysis were recorded in 89% of the implants (163 cases). The cumulative survivorship (revision as endpoint) at the time was 88,9%.
Conclusion
Our study confirms the bad behavior of this type of cup probably related to the design, to the method of HA fixation. The observations carried out on the revised cup confirm these hypotheses but did not clarify if the third body wear could be a further problem. Another interesting aspect is the high incidence of osteolysis, which are often asymptomatic becoming a problem for the surgeon as the patient refuses the possibility of a revision.
Primary human osteoblast culture on 3D porous
collagen-hydroxyapatite scaffolds.
Guy Hilton Research Centre, Institute for Science and Technology in
Medicine, Keele University, Thornburrow Drive, Hartshill ST4 7QB,
Stoke-on-Trent, United Kingdom.
There is a need in tissue-engineering for 3D scaffolds that mimic
the natural extracellular matrix of bone to enhance cell adhesion,
proliferation, and differentiation. The scaffold is also required
to be degradable. A highly porous scaffold has been developed to
incorporate two of the extracellular components found in
bone-collagen and hydroxyapatite (HA). The scaffold's collagen
component is an afibrillar monomeric type I atelocollagen extracted
from foetal calf's skin. This provided a novel environment for the
inclusion of HA powder. Five hundred thousand primary human
osteoblasts were seeded onto 4 mm cubed scaffolds that varied in
ratio of HA to collagen. Weight ratios of 1:99, 25:75, 50:50, and
75:25 hydroxyapatite:collagen (HA:Collagen) were analysed.
The scaffolds plus cells were cultured for 21 days. DNA assays and
live/dead viability staining demonstrated that all of the scaffolds
supported cell proliferation and viability. An alkaline phosphatase
assay showed similar osteoblast phenotype maintenance on all of the
3D scaffolds analysed at 21 days. MicroCT analysis demonstrated an
increase in total sample volume (correlating to increase in
unmineralised matrix production). An even distribution of HA
throughout the collagen matrix was observed using this technique.
Also at 3 weeks, reductions in the percentage of the mineralised
phase of the constructs were seen. These results indicate that each
of the ratios of HA/collagen scaffolds have great potential for
bone tissue engineering. (c) 2010 Wiley Periodicals, Inc. J Biomed
Mater Res Part A, 2010.
CELL-DERIVED MATRIX ENHANCES OSTEOGENIC PROPERTIES OF
HYDROXYAPATITE.
National O.Bohomolets Medical University, Department of Oral and
Maxillofacial Surgery, Kyiv , Ukraine; [email protected].
The study aimed to evaluate osteogenic properties of hydroxyapatite
(HA) scaffold combined with extracellular matrix (ECM) derived in
vitro from rat primary calvarial osteoblasts or dermal fibroblasts.
The cellular viability, and the ECM deposited onto synthetic HA
microparticles were assessed by MTT, Glycosaminoglycan, and
Hydroxyproline assays as well as immunohistochemistry and scanning
electron microscopy after 21 days of culture. The decellularized
HA-ECM constructs were implanted in critical-sized calvarial
defects of Sprague-Dowley rats, followed by bone repair and local
inflammatory response assessments by histomorphometry and
immunohistochemistry at 12 weeks postoperatively.
We demonstrated that HA supported cellular adhesion, growth and ECM
production in vitro, and the HA-ECM constructs significantly
enhanced calvarial bone repair (p