Detection of Mechanical Detection of Mechanical Failure in BSCC Artificial Failure in BSCC Artificial
Heart Valve using Heart Valve using EMAT TechniqueEMAT Technique
Materials Assesment Research Group, Department of Electrical and Computer Engineering, Michigan State University, East Lansing, MI 48824.
![Page 2: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/2.jpg)
TeamTeam
Naveen V Nair
Sneha Teresa Thomas
Sridhar Ramakrishnan
Robert Ahn
Ihuma Ugenyi
Pradeep Ramuhalli
Dr Satish S UdpaDr Lalita Udpa
![Page 3: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/3.jpg)
WHAT IS A BSCC HEART WHAT IS A BSCC HEART VALVE?VALVE?
A mechanical prosthetic heart valve that is famous not for an excellent design but for a history of failure .
Of the 86,000 patients who received these valves, four hundred died from a strut fracture, in the first year.
Two hundred additional patients survived similar strut fractures through open-heart surgery
![Page 4: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/4.jpg)
BSCC HEART VALVE
![Page 5: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/5.jpg)
The EMAT TechniqueThe EMAT Technique
The EMAT method is one of the possible ways of detecting SLS in BSCC heart-valves. The method involves use of an electro magnetically generated force to induce acoustic vibrations in the strut.
![Page 6: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/6.jpg)
ProcedureProcedure
A large magnetic field is produced with the help of a permanent magnet. In this field a electromagnet is placed and made to carry an alternating current.
The patient placed in the field of the electromagnet such that the plane of the ring is perpendicular to the alternating magnetic field.
This induces a current in the strut. The permanent magnetic field and the induced current in the strut produce a force on the strut, which causes it to vibrate
![Page 7: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/7.jpg)
AnalysisAnalysis
As for any resonating structure the heart valves also have their own resonant frequencies and the resonance frequencies of the BSCC valve with an SLS is found to be different from that of an IOS (Intact Outlet Strut).
![Page 8: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/8.jpg)
Challenges involvedChallenges involved
The proper excitation is possible only when the ring is properly aligned to the magnetic fields of both the permanent magnetic field and also the electromagnet
It is difficult to produce a permanent magnetic field of high intensity to produce detectable vibrations in the strut
The need of sophisticated instruments to pick up signals generated from the strut without invasion
![Page 9: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/9.jpg)
Contd..Contd..
The signals obtained do not represent signals from the strut alone. The signals from the strut are mixed with signals from the lungs, Gastro intestinal tracts and other parts of the body
![Page 10: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/10.jpg)
EMATEMAT
Signal Processing Algorithms
![Page 11: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/11.jpg)
ObjectivesObjectives
To detect signals from the heart valve at either of the two resonance frequencies (~7.5 kHz or ~2.3 kHz)
Separate these from the other acoustic interference signals. (Potential sources include lungs and GI tract).
Remove ambient noise in measurements
![Page 12: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/12.jpg)
Why a sophisticated Why a sophisticated algorithm? algorithm?
Low Signal to Noise Ratio (SNR)– The acoustic signals picked up by the sensors
are corrupted by noise and other interfering signals.
Frequency overlap– Traditional frequency domain filtering
techniques would not work as the desired acoustic signal and noise have overlapping frequency spectra.
![Page 13: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/13.jpg)
Algorithm - conceptsAlgorithm - concepts
Beam-forming: An algorithm that will enable us to selectively listen to only one source at a particular location and frequency.
![Page 14: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/14.jpg)
Dimensions in cm
Algorithm - conceptsAlgorithm - concepts
![Page 15: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/15.jpg)
Algorithm - conceptsAlgorithm - conceptsSensor Signals
Estimate Number of Sources
(Method: Minimum Description Length)
Estimate Locations and Frequencies of Sources
(Method: Wax Sub-optimal algorithm)
Estimate Desired Source Signal
(Method: Linearly Constrained Minimum Variance Beamforming)
Approximate Locations and Frequencies of Sources
Desired Source Signal
![Page 16: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/16.jpg)
Sub-Optimal Technique - Sub-Optimal Technique - problem modelproblem model
The sources are assumed to be characterized by the following parameters– X coordinate– Y coordinate– Z coordinate– Center frequency
![Page 17: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/17.jpg)
Beamforming conceptBeamforming concept
ARRAY OF SENSORS
Combine sensor signals
optimally
BEAMFORMER ESTIMATE OF THE DESIRED
SOURCE
NOISE / INTERFERING SOURCES
DESIRED SOURCE
![Page 18: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/18.jpg)
Experiments (known location & Experiments (known location & frequency)frequency)
1. Simulated Data
2. Acoustic experiment – in air
3. Acoustic experiment - with human body as the transfer medium
![Page 19: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/19.jpg)
Simulation - BeamformerSimulation - Beamformer
Dimensions in cm
![Page 20: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/20.jpg)
Original Source Signals - timeOriginal Source Signals - time
![Page 21: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/21.jpg)
Original Source Signals - freqOriginal Source Signals - freq
![Page 22: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/22.jpg)
Sensor Signals - timeSensor Signals - time
![Page 23: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/23.jpg)
Sensor Signals - freqSensor Signals - freq
![Page 24: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/24.jpg)
Simulation Results - timeSimulation Results - time
![Page 25: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/25.jpg)
Simulation Results -freqSimulation Results -freq
Using exact source positions
![Page 26: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/26.jpg)
Simulation Results -freqSimulation Results -freq
Using source locations with error
![Page 27: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/27.jpg)
Experiment 2. Acoustic Data –in airExperiment 2. Acoustic Data –in airTest set-up
Sensors
Sources
Source Freq:1. 50002. 60003. 3000
![Page 28: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/28.jpg)
Experiment 2. ResultsExperiment 2. Results
![Page 29: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/29.jpg)
Experiment – 3Experiment – 3
0.5 kHz
7.5 kHz
![Page 30: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/30.jpg)
Experiment 3 - Sensor Signals - freq
![Page 31: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/31.jpg)
Experiment 3- Estimation of location and Experiment 3- Estimation of location and frequencyfrequency
Parameters Source 1 Source 2
Error in X
(in cm)0.05 0.12
Error in Y
(in cm)0.1 0.1
Error in Z
(in cm)0.1 0.12
Error in
Frequency
(in Hz)40 40
![Page 32: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/32.jpg)
ExperimentExperiment 3 Results - freq 3 Results - freq
![Page 33: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/33.jpg)
•Initial results show promise
Future Plans
•More extensive studies on practical data•Optimize algorithm parameters
Conclusions
![Page 34: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/34.jpg)
Blind Source Separation of Acoustic Blind Source Separation of Acoustic Signals from BSCC Heart valveSignals from BSCC Heart valve
Techniques and Issues
![Page 35: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/35.jpg)
Blind Source WHAT ????Blind Source WHAT ????
SeparationConcept of Spatial LocalisationLack of frequency clarity -overlaps in freq
domainUtilise location and frequency information Conduct a sine before the actual test is
performed to determine actual position of the sources
![Page 36: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/36.jpg)
Location and FrequencyLocation and Frequency
Wax’s algorithm Cocktail party problem
– Selective listening
MDL principleOptimisation of the likelihood parameter
![Page 37: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/37.jpg)
Details of the problemDetails of the problem
Consider a set of sources emitting sounds (or any signals) which are band limited and have a single center frequency.
The signal can be written as
– Where the parameters a, w and are slow moving functions of time.
This signal is generated by a number of sources and picked up by a number of sources.
))(*)(2sin()()( tttwtats
![Page 38: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/38.jpg)
Details – cont’dDetails – cont’d
At the sensor what is obtained is a mixture of the signals. – The sensor signal looks like
– Where n is a noise parameter dependent on sensor characterestics only .
– is the time of flight– a is the transfer parameter relating the source signal to
the sensor. This is dependent only on the direction of arrival.
![Page 39: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/39.jpg)
Likelihood parameterLikelihood parameter
The likelihood parameter is found to be dependent only on the first term and is found to be
– In this the l the eigen values of which is the largest eigen value and u are the corresponding eigen vectors
Maximizing this yeilds the optimal set of b() which give the optimal , or direction of arrival
)(.)(
.)()()( 2
bb
ublL
H
iH
is
![Page 40: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/40.jpg)
Troubles and TribulationsTroubles and Tribulations
Heavily dependent on sensor locations How do we get this signal with almost –50dB
SNR out Lack of knowledge of source locations eg. the GI
tract is nowhere near a stationary source infact moves in an approximately gaussian path
Near zero information about the acoustic transfer parameters of the human torso…
![Page 41: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/41.jpg)
Ways around stumbling blocksWays around stumbling blocksSensor location optimisation…
– Maximize a cost function made up by summing energy in the 1.5-3.5 and 6-8 KHz freq range
Require transfer function of body– Make one ????
• Nearly 200 different entities with diff parameters !!!
– Do somekind of ray tracing to arrive at optimum location
Again requires some idea of geometry…
Round and round we go ’round the mulberry bush
![Page 42: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/42.jpg)
Current SolutionCurrent Solution
Use the same cost function but for data, use the Wax algo to determine the source location for a particular sensor location and then use the same algo inverted to find out the signal spectrum for various sensor locations…
![Page 43: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/43.jpg)
Other issues (non signal processing)Other issues (non signal processing) Sensors… ? If u want high sensitivity (like stethoscopes) then
freq range is very low… For these freq ranges no sensors with high enough
sensitivity… Have to get things custom made.. How do we know what
sensitivity we need ?
We rented hydrophones for around $1700 then found out one day before delivery that we couldn’t use them… got lucky this time.
People (sridhar) still working on modifying stethoscopes and so on…
- We don’t
![Page 44: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/44.jpg)
Synchronous data input Synchronous data input
We need to synchronise data received from signal for beamforming to work.– We used a software to split up the left and right channel
of the computer and used MATLAB to generate emulate two sources through each speaker.
– Used atomic clock to sync two machines to get 4 sources.
– Did some more vague stuff.. Split up a microphone using two wires and one way connectors… !!!!
![Page 45: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/45.jpg)
Positions of sensorsPositions of sensors
Measure accurately positions of the sensors– Used Flock of Birds.. – Have to do better
![Page 46: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/46.jpg)
Generating magnetic fieldGenerating magnetic field
Use a superconducting magnet pulled out of an MRI machine.
Incidentally it cupped all our computers
Magnetic noise…Flock of Birds refused to fly
![Page 47: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/47.jpg)
Other methodsOther methods
GradiometerInvasiveUse a magnet to generate current in strut.
Measure magnetic field peturbationUse a catheter
![Page 48: presentation](https://reader035.vdocuments.site/reader035/viewer/2022062706/55764065d8b42ac31b8b48b8/html5/thumbnails/48.jpg)