precision resistance decades precision rtd simulators m ... · basic technical data resistance...

11
Image sheet Precision Resistance Decades Precision RTD Simulators M-622, M-612 M-602, M-602A

Upload: others

Post on 09-Apr-2020

8 views

Category:

Documents


0 download

TRANSCRIPT

Page 1: Precision Resistance Decades Precision RTD Simulators M ... · Basic technical data Resistance decade M-602 Precision RTD simulator M-612 Precision resistance decade M-622 Resistance

Image sheet

Precision Resistance DecadesPrecision RTD Simulators

M-622, M-612M-602, M-602A

Page 2: Precision Resistance Decades Precision RTD Simulators M ... · Basic technical data Resistance decade M-602 Precision RTD simulator M-612 Precision resistance decade M-622 Resistance

Basic technical data

Resistance decade M-602

Precision RTD simulator M-612

Precision resistance decade M-622

Resistance range from 0.1 Ohm to 10 MOhmAccuracy 0.02 %Temperature coefficient < 25ppm/ CMax. load 2 W (120 V or 0.6 A)Simulation of RTD (Pt, Ni) temperature sensorsAccuracy +/- 0.2 C2, 3, 4 - wire connectionPower line adapter (Internal accumulator optionally)Interface RS-232 (IEEE 488 optionaly)

o

o

Resistance range from 16.0000 Ohm to 10 kOhmAccuracy 0.005 %Temperature coefficient < 1ppm/ CSimulation of RTD (Pt, Ni) temperature sensorsSimulation of cable resistanceAccuracy +/- 0.02 C2, 3, 4 - wire connectionPower line adapter (Internal accumulator optionally)Interface RS-232 (IEEE 488 optionaly)

o

o

Resistance range from 1.000 00 Ohm to 1.200 000 MOhmAccuracy 0.005 %Temperature coefficient < 1ppm/ COperating voltage 120VSimulation of RTD (Pt, Ni) temperature sensorsAccuracy +/- 0.02 C2, 3, 4 - wire connectionInternal accumulator / power line adapterInterface RS-232 (IEEE 488 optionaly)

o

o

Page 3: Precision Resistance Decades Precision RTD Simulators M ... · Basic technical data Resistance decade M-602 Precision RTD simulator M-612 Precision resistance decade M-622 Resistance

Multifunction display shows:

Output terminals

Numerical keypadwith cursor keys

Housing holder

* status of the decade * preset values * SETUP and Calibration menu* calibration data * local or remote control

* 2W, 4W resistance* Simulation RTD sensors ( Pt, Ni ) * Open/Short functions

GND terminalprotection earth Power on/off button

Front panel ergonomy

Series of M6x2 Precision resistance decades/RTD simulators offers decades of electric resistance up 0,1 Ohm to 10 MOhm.

Basic application fields are calibration laboratories, manufacturers of electric resistance meters regulators and process meters.

Front panel is divided into logical areas :

* output terminals field * numerical keypad with cursor keys * display with contact and visibility from all directions

LED indicated4W function

LED indicated2W function

Page 4: Precision Resistance Decades Precision RTD Simulators M ... · Basic technical data Resistance decade M-602 Precision RTD simulator M-612 Precision resistance decade M-622 Resistance

Rear panel with interface

Power supply connector

Power line adapter

Series of M6x2 Precision resistance decades/RTD simulators offers decades of electric resistance up 0,1 Ohm to 10 MOhm.

Basic application fields are calibration laboratories, manufacturers of electric resistance meters, regulators and process meters.

RS-232 interface

GPIB interface

Page 5: Precision Resistance Decades Precision RTD Simulators M ... · Basic technical data Resistance decade M-602 Precision RTD simulator M-612 Precision resistance decade M-622 Resistance

2 W Resistance calibration

Simple connection and one touch buttons simplifytwo-terminal calibration of UUTs: * Connect UUT input to the precision decade 2W output terminals * Set R-function and requested value * Press ENTER

LED indicated2W function

1. Select function R2. Select 4W / 2W 3. Set value 0 Ohm

4. Set requested value using numerical keypad5. Press ENTER

Decade display R mode.

Page 6: Precision Resistance Decades Precision RTD Simulators M ... · Basic technical data Resistance decade M-602 Precision RTD simulator M-612 Precision resistance decade M-622 Resistance

4 W Resistance calibration

Simple connection and one touch buttons simplify four-terminal calibration of UUTs: * Connect UUT input to the precision decade 4W output terminals * Set R-function and requested value * Press ENTER

LED indicated4W function

1. Select function R2. Select 4W / 2W 3. Set value 0-10 kOhm

4. Set requested value using numerical keypad5. Press ENTER

Decade display R mode.

Page 7: Precision Resistance Decades Precision RTD Simulators M ... · Basic technical data Resistance decade M-602 Precision RTD simulator M-612 Precision resistance decade M-622 Resistance

2 W / 4W Temperature calibration

Simple connection and one touch buttons simplify two/four-terminal calibration of UUTs: * Connect UUT input to the precision decade output terminals * Set RTD simulation and requested value * Press ENTER

LED indicated2W function

1. Select function and temperature scale Pt(68),Pt(90),Pt(US), Ni2. Set value of Ro(Pt,Ni) 3. Set T.unit (°C), (°F)

4. Set requested value of temperature using numerical keypad5. Press ENTER

Decade display RTD mode.

Page 8: Precision Resistance Decades Precision RTD Simulators M ... · Basic technical data Resistance decade M-602 Precision RTD simulator M-612 Precision resistance decade M-622 Resistance

Heat meters calibration

LED indicated2W function

1. Select function and temperature scale Pt(68),Pt(90),Pt(US), Ni2. Set value of Ro(Pt,Ni) 3. Set T.unit (°C), (°F)

4. Set requested value of temperature using numerical keypad5. Press ENTER

Simple connection and one touch buttons simplify two/four-terminal calibration of heat meters: * Connect UUT input 1 to the RTD simulator output terminals * Connect UUT input 2 to the RTD simulator output terminals * Set RTD simulation and requested value T1, T2 * Press ENTER

1

2

1

2

Input 1Input 2

Decade display RTD mode.

Page 9: Precision Resistance Decades Precision RTD Simulators M ... · Basic technical data Resistance decade M-602 Precision RTD simulator M-612 Precision resistance decade M-622 Resistance

Requested accuracy

1 mOhm1 mOhm1 mOhm

.

.

.20 Ohm

Standard(terminals)

R00 (4W)R01 (4W)R02 (4W)

.

.

.R35 (R2W/4W)

Nominal value

2.0 Ohm3.9 Ohm7.8 Ohm

.

.

.602 kOhm

Decade recalibration

Simple procedure of recalibration is accesible using decade CALIBRATION mode: - correct password must be entered befor calibration

Decade display password mode.

Decade display calibration mode.

- recalibration procedure consist of measuring basic resistance values and entering their actually measured data

- in following table nominal values of calibration points and requested recalibration accuracy are described

Page 10: Precision Resistance Decades Precision RTD Simulators M ... · Basic technical data Resistance decade M-602 Precision RTD simulator M-612 Precision resistance decade M-622 Resistance

Remote control & Automatic calibration

Precision decades full automatic operatio using remote control mode. Following intrefaces are available for connection the calibrator to PC:

- GPIB ( IEEE-488 ) interface. Connector is located on the rear panel Nationa instruments GPIB card is requested to be installed in the computer.

- RS-232 serial line. connector is located on the rear panel.

- USB interface using RS-232-USB converter

- by user created application SW, using control commands SCPI

- using Meatest CALIBER application SW

Decades can be applied in systems for automatic and semiautomatic calibration ofUUTs:

CALIBER application software for automatic and semiautomatic calibration of meters of electric quantities.

CALIBER ProcedureCalibrationcertificate

Standard unit

Unit under test

CALIBER application software is based on instrument cards. Basic instrument card are delivered with decades,however new card be created by calibration laboratory.

- output from the CALIBER is list of calibration points with measured and evaluated results including uncertainty of calibration

- calibration procedures can be easily created by calibration laboratory

Page 11: Precision Resistance Decades Precision RTD Simulators M ... · Basic technical data Resistance decade M-602 Precision RTD simulator M-612 Precision resistance decade M-622 Resistance

Ordering information - options

Resistance decade M-602 M602-Vxxxx - basic version (10W - 300kW)

M602A-Vxxxx - extende d version (0.1W - 10MW) Bus M602-V1xxx - RS232 M602-V2xxx - IEEE488

Addit ional functions M602-Vx0xx - none M602-Vx1xx - Short/Open function

Housing M602-Vxx0x - table version M602-Vxx1x - module 19“, 3HE M602-Vxx2x - table version with holde r

Power supplying M602-Vxxx0 - power line adaptor M602-Vxxx1 - battery + power line adapt or Example of order:

M602A-V2110 - resistance decade 0.1W - 10MW, IEEE488, Short/Open, 19“ rack, without battery

Precision RTD simulator M-612

Bus M612-V1xxx - RS232 M612-V2xxx - GPIB

Addit ional functions M612-Vx0xx - none M612-Vx1xx - Short/Open function

Housing M612-Vxx0x - table version M612-Vxx1x - module 19“, 3HE

Bus M622-V1xxx - RS232 M622-V2xxx - IEEE488

Addit ional functions M622-Vx0xx - none M622-Vx1xx - Short/Open function

Housing M622-Vxx0x - table version M622-Vxx1x - module 19“, 3HE

Precision resistance decade M-622