Page 1: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Pre-stent plaque modification in complex, calcified coronary

artery lesions

Pre-stent plaque modification in complex, calcified coronary

artery lesions

AngioSculpt®A New Scoring Balloon

Ehud Grenadier M.DEhud Grenadier M.D

Cardiac catheterization LabsCardiac catheterization LabsHaifa;Herzlia Haifa;Herzlia

ISRAEL ISRAEL

Page 2: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Why is Lesion Why is Lesion

Preparation Preparation

EssentialEssential??

Page 3: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

MSD as percent of RLD

MSD as a percentage of RLD following stent deployment. Optimum stent deployment : (MSD 90% RLD).

Patients not achieving criteria with the stent delivery system (71%)

Brodie B. et al POSTIT Trial, Cath. Card. Int. 2003: 59,184-192.

Per

cen

t o

f P

op

ula

tio

n

40 50 60 70 80 90 100 120 130

Patients who did not achieve optimal stent deployment with SDS

POSTIT TrialPOSTIT Trial : when using only the stent delivery balloon : when using only the stent delivery balloon over 70% of the pts. did not achieve optimal stent deploymentover 70% of the pts. did not achieve optimal stent deployment..

Page 4: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

++++++

++ ++++

++++

++++

+

++++ ++++ +

++xx x

xxxxxxxx

xxx

x xx

xxx

xxx

xxxx x

x

++

xxxx

2.5 3.0 3.5 4.0 4.5

1.5

2.0

2.5

3.0

3.5

4.0

4.5

Manufacturer’s Predicted MSD (mm)

IVU

S M

eas

ure

d M

SD

(m

m) Nominal S

tent Size

Manufacturer AManufacturer BManufacturer CManufacturer Dx

+

Costa et al, Am J Cardiol 2005:96,74-78

Manufacturer’s Predicted Minimal Stent Diameter Manufacturer’s Predicted Minimal Stent Diameter (MSD) is Usually Not Achieved With the Original (MSD) is Usually Not Achieved With the Original

Stent BalloonStent Balloon

Page 5: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

IVUS guidance may result in more effective

stent expansion and results in 44% reduction

in TVR

Only 22% of patients that experience SAT have an optimum PCI result as assessed by

IVUS

CRUISE study: Peter J. Fitzgerald et.al. “Final Results of the Can Routine Ultrasound Influence Stent Expansion (CRUISE) Study”

Circulation 2000 102:523-530

.Cheneau Study: Edouard Cheneau,et al. “Predictors of Subacute Thrombosis. Results

of a Systematic Intravascular Ultrasound Study.” Circulation 2003; 108:43-47

T

V

R

S

A

T

Non-uniform stent strut distribution and

malapposition may result in more intimal

hyperplasia after stenting

MSA and stent under expansion

are major predictors of

stent thrombosis

RESTENOSIS

Takebayastri Study: Hideo Takebayastri et al. “Non-uniform Strent Strut Distribution

Correlates with More Neointimal Hyperplasia after Sirolimus-Elluting Stent Implantation”

Circulation 2004;110:3430-3434

B.Brodie. Adjunctive balloon post dilatation after stent deployment. J Interven Cardiol

2006;19:43-50

THROBOSIS

Page 6: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Benefit of the AngioSculpt Benefit of the AngioSculpt for Pre-dilatation of DESfor Pre-dilatation of DES

Benefit of the AngioSculpt Benefit of the AngioSculpt for Pre-dilatation of DESfor Pre-dilatation of DES

Group I (n=145)Group II (n=117)

Group III (n=37)

p value

MSD/PSD76% ± 10%76% ± 13%88% ± 18%<0.001

MSA/PSA67% ± 16%70% ± 23%88% ± 22%<0.001

Final stent CSA < 5.0mm226.2%25.6%10.8%0.03

Costa et al, AHA 2006

Group I = direct stenting

Group II = pre-dilatation with POBA

Group III = pre-dilatation with AngioSculpt

CSA = cross sectional area, MSA = minimum stent area, MSD = minimum stent diameter,

PSA = predicted stent area, PSD = predicted stent diameter

Page 7: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

AngioSculpt Scoring Balloon AngioSculpt Scoring Balloon CatheterCatheter

Semi-compliant balloon with an Semi-compliant balloon with an external external

nitinol shape memory helical scoring nitinol shape memory helical scoring edgeedge

Semi-compliant balloon with an Semi-compliant balloon with an external external

nitinol shape memory helical scoring nitinol shape memory helical scoring edgeedge

Page 8: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Centrally Located Within the Arterial Lumen

Page 9: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

The AngioSculpt is The AngioSculpt is Comprised ofComprised of : :

A - Rapid exchange or OTW semi-compliant A - Rapid exchange or OTW semi-compliant balloonballoon

B - Laser cut nitinol scoring elementB - Laser cut nitinol scoring element

A B

Page 10: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Acute histopathology specimen of a patient post conventional

balloon angioplasty demonstrating extensive dissection and laceration

(yellow arrows)

Post-AngioSculpt scoring of porcine ISR (yellow arrows)

Histopathology of the AngioSculpt vs. Histopathology of the AngioSculpt vs. POBAPOBA

Page 11: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Use of the Angiosculpt for Pre-dilation of Use of the Angiosculpt for Pre-dilation of Challenging Coronary Lesions Prior to Challenging Coronary Lesions Prior to

Stenting Stenting

Personal ExperiencePersonal Experience

Use of the Angiosculpt for Pre-dilation of Use of the Angiosculpt for Pre-dilation of Challenging Coronary Lesions Prior to Challenging Coronary Lesions Prior to

Stenting Stenting

Personal ExperiencePersonal Experience Ehud Grenadier, Arthur Kerner, Nabeel

Makhoul, Luis Gruberg, Ariel Roguin

Herzliya & Rambam Med. Ctrs. Herzliya & Haifa, ISRAEL

Page 12: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Aims of the StudyAims of the Study

To evaluate the Deliverability, Efficacy, and Safety of the

AngioSculpt Scoring Balloon Catheter

in patients with challenging and calcific CAD

To evaluate the Deliverability, Efficacy, and Safety of the

AngioSculpt Scoring Balloon Catheter

in patients with challenging and calcific CAD

Page 13: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Study Design

Two-center, prospective, consecutive, non-randomized single operator registry

Inclusion criteria: pts with single or multi-vessel coronary artery disease scheduled to undergo PCI

Population: 221 consecutive pts enrolled at 2 sites

Follow-up: outpatient visits, telephone interviews and re-catheterization if clinically indicated

Two-center, prospective, consecutive, non-randomized single operator registry

Inclusion criteria: pts with single or multi-vessel coronary artery disease scheduled to undergo PCI

Population: 221 consecutive pts enrolled at 2 sites

Follow-up: outpatient visits, telephone interviews and re-catheterization if clinically indicated

Page 14: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

I.V.U.S - VirtualI.V.U.S - Virtual HistologyHistology

Evaluated in 60/221 [ 27.1 %] pts. 76/279 [ 27.2 %] lesions

“Volcano Therapeutics” IVG - 3 / S-5

Eagle Eye Gold IVUS catheter

Evaluated in 60/221 [ 27.1 %] pts. 76/279 [ 27.2 %] lesions

“Volcano Therapeutics” IVG - 3 / S-5

Eagle Eye Gold IVUS catheter

Performed in 111/221 [ 49.7 % ] pts. 154/279 [ 55.2 % ] lesionsPerformed in 111/221 [ 49.7 % ] pts. 154/279 [ 55.2 % ] lesions

Page 15: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

AngioSculpt AngioSculpt Relative ContraindicationsRelative Contraindications

Unprotected left main diseaseUnprotected left main disease

Non-calcified SVG lesionNon-calcified SVG lesion

Thrombotic lesionThrombotic lesion

Unprotected left main diseaseUnprotected left main disease

Non-calcified SVG lesionNon-calcified SVG lesion

Thrombotic lesionThrombotic lesion

Page 16: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Primary Performance Primary Performance EndpointEndpoint

Procedural Success

Final diameter stenosis ≤10% in all AngioSculpt treated lesions following completion of the interventional procedure (including adjunctive stenting when used) without in-hospital major adverse cardiac events (MACE)

Procedural Success

Final diameter stenosis ≤10% in all AngioSculpt treated lesions following completion of the interventional procedure (including adjunctive stenting when used) without in-hospital major adverse cardiac events (MACE)

Page 17: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Primary Safety Primary Safety EndpointEndpoint

Clinical Success

Successful PCI and freedom from MACE at 30 day follow-up

MACE is defined as death, Q wave or non-Q wave MI, CABG or TLR of the index lesion

Clinical Success

Successful PCI and freedom from MACE at 30 day follow-up

MACE is defined as death, Q wave or non-Q wave MI, CABG or TLR of the index lesion

Page 18: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Secondary Secondary EndpointsEndpoints

Flow limiting dissectionFlow limiting dissection

Vessel perforationVessel perforation

Page 19: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Patient Patient DemographicsDemographics

221 pts, 279 lesions221 pts, 279 lesions Age: 63.8±19.4 yrs (range 34 – 88)Age: 63.8±19.4 yrs (range 34 – 88) Males: 72%Males: 72% Severe Angina (CCS class 3-4): 56%Severe Angina (CCS class 3-4): 56% Prior PCI: 43%Prior PCI: 43% Prior MI: 34%Prior MI: 34% Prior CABG: 6%Prior CABG: 6%

Age: 63.8±19.4 yrs (range 34 – 88)Age: 63.8±19.4 yrs (range 34 – 88) Males: 72%Males: 72% Severe Angina (CCS class 3-4): 56%Severe Angina (CCS class 3-4): 56% Prior PCI: 43%Prior PCI: 43% Prior MI: 34%Prior MI: 34% Prior CABG: 6%Prior CABG: 6%

Page 20: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Risk Factors

0

10

20

30

40

50

60

70

Hypercholesterolemia Diabetes Family history of CAD

5.7%5.7%

NIDDMNIDDM

58.7%58.7%

35.8%35.8%

41.2%41.2%%%

IDDMIDDM

Page 21: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Vessel DistributionVessel Distribution

2

36

4

1715

30

0

5

10

15

20

25

30

35

40

LM protected LAD Diagonal LCx LCxM RCA

%

Page 22: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Lesion CharacteristicsLesion Characteristics

ISR: 51/279 [18%], De-Novo: 228/279 [82%] Bifurcation: 69/279 [25%] Eccentric: 117/279 [42%] Ulcerated: 31/279 [11%] >30° angulation: 154/279[55%]

Mean lesion angulation: 34.3±17˚

ISR: 51/279 [18%], De-Novo: 228/279 [82%] Bifurcation: 69/279 [25%] Eccentric: 117/279 [42%] Ulcerated: 31/279 [11%] >30° angulation: 154/279[55%]

Mean lesion angulation: 34.3±17˚

Page 23: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Lesions Classification (AHA/ACC)

Page 24: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Calcification Severity Calcification Severity

Page 25: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Pre-Procedure QCAPre-Procedure QCA

Page 26: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Pre-Procedure Lesion Pre-Procedure Lesion Measurements (QCA)Measurements (QCA)

Length(mm)

MLD(mm)

RVD [d](mm)

17.1+11.40.26±0.292.69±0.35

Page 27: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Primary Endpoint Primary Endpoint ResultsResults

Procedural Success: 96.0% [268/279]

Clinical Success: 94.6% [209/221]

Procedural Success: 96.0% [268/279]

Clinical Success: 94.6% [209/221]

Page 28: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

AngioSculpt “Only” Procedural Success

Diameter stenosis <50%: 224/279 [80%]

Diameter stenosis ≤20%: 158/279 [57%]

Final failure to cross with the AngioSculpt occurred in 8/279 lesions [2 .9%], all occurred in severe, calcified

lesions with pronounced vessel/lesion angulation

Diameter stenosis <50%: 224/279 [80%]

Diameter stenosis ≤20%: 158/279 [57%]

Final failure to cross with the AngioSculpt occurred in 8/279 lesions [2 .9%], all occurred in severe, calcified

lesions with pronounced vessel/lesion angulation

Page 29: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

QCA ResultsQCA ResultsDiameter StenosisDiameter Stenosis)%( )%(

Pre-ASCPre-ASCPost-ASCPost-ASCPost-StentPost-Stent

82.982.9++11.11.2 2 23.523.5++13.613.65.95.9++4.4 4.4

Page 30: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

QCA Results (cont’d)QCA Results (cont’d)Minimum Luminal Diameter Minimum Luminal Diameter

(mm)(mm)

Pre-ASCPre-ASCPost-ASCPost-ASCPost-Post-StentStent

0.260.26++0.20.299

2.072.07++0.680.682.432.43++0.70.722

Page 31: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

IVUS Results IVUS Results MLD (mm) and MLD (mm) and CSA (mmCSA (mm22))

MLD, CSA Ref. Vessel

(d)

MLD, CSA Pre-

dilatation

MLD, CSAPost-ASC

P-Value*MLD, CSAPost-stent

P-Value*

2.99±0.587.1±2.9

1.7±0.142.57±0.45

2.1±0.353.86±1.0

8

p<0.0001

p<0.0001

2.82±1.15.5±2.03

p<0.002p<0.000

1

*Paired t-Test

Page 32: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

IVUS ResultsIVUS ResultsLesion CharacteristicsLesion Characteristics

Mod-Severe Ca++Eccentric

Post-ASC “Controlled” Dissections

72.4 %39.1%82.7 %

Page 33: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

AngioSculpt Dimensions & AngioSculpt Dimensions & Inflation PressuresInflation Pressures

ASC Diameter(mm)

ASC Length(mm)

Inflation Pressure(atm)

2.51+0.3416.4+3.315.1+2.4

range: 10-18

Page 34: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Reduction in Usage Reduction in Usage of NC Balloonsof NC Balloons

Usage of NC Balloons Pre-ASC

Usage of NC Balloons in ASC

Study

Reduction in Usage of NC

Balloons

89% 43% 46%

Page 35: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

AngioSculpt CrossabilityAngioSculpt Crossability

Planned POBA Pre-

ASC

Successful Primary

ASC

POBA Pre-dilatation

After Primary ASC Failure to

Cross

Failure to Cross with ASC

After POBA Pre-dilatation

Predicted Failure to

Cross (ASC not

attempted)

113/279[40.5%]

104/279[37.3%]

31/279[11.1%]

8/279[2.9%]

1/279 [0.4%]

Page 36: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Device SlippageDevice Slippage

Device slippage was observed in 4/271 [1.5 %] of ASC treated lesions

Device slippage was observed in 4/271 [1.5 %] of ASC treated lesions

Page 37: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Double Wire TechniquesDouble Wire Techniques

In 11/279 [3.9 %] lesions the additional use of a buddy wire resulted in successful crossing of 6/11 lesions

The anchor wire [LCA] was used in 5/279 [1.8%] lesions resulting in successful crossing in 3/5 lesions

In 11/279 [3.9 %] lesions the additional use of a buddy wire resulted in successful crossing of 6/11 lesions

The anchor wire [LCA] was used in 5/279 [1.8%] lesions resulting in successful crossing in 3/5 lesions

Page 38: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Stent Types Used in the StudyStent Types Used in the Study

21%

33%17%

15%14%

Cypher Taxus Endeavor Titan Other stents

Stents were implanted in 264/279 [94.6%] lesions Stents were implanted in 264/279 [94.6%] lesions

Page 39: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Distal Left Main and Proximal LCx Lesions

Page 40: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

ANGIOSCULPT 2.0 x 15 mm

Page 41: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

POST-DES INSERTION

Page 42: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

IVUS Demonstrates Optimal Stent Apposition

Page 43: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Angiogram and IVUS Pre-Intervention

MLD = 1.2mmMLA = 1.5mm²

Plaque area = 5.72mm²Plaque burden = 83%

Page 44: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Angiogram and IVUS Post-AngioSculpt

MLD = 1.4 mm

MLA = 3.1 mm²

Plaque area = 4.8 mm²

Plaque burden = 63%

Yellow arrows indicate “scoring marks”

Page 45: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Angiogram and IVUS Post-Stent

MLD = 2.6 mm

MSA = 5.2 mm²

Page 46: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Severe calcified Lcx LesionSevere calcified Lcx Lesion

Page 47: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Angiosculpt DilationAngiosculpt Dilation

Page 48: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Final Angiographic ResultFinal Angiographic Result [ Post Taxus Stent Insertion ] [ Post Taxus Stent Insertion ]

Page 49: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Distal long RCA lesion

Page 50: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Pre dilated RCA lesion - Q.C.A Analysis

Page 51: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

3.0/20 mm and 3.5/15 3.0/20 mm and 3.5/15 mm Angiosculpt mm Angiosculpt

PredilationPredilation

Page 52: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

3.53.5 mm Endeavor Stent Insertionmm Endeavor Stent Insertion

Page 53: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

FINAL Angiographic RESULTFINAL Angiographic RESULT

Page 54: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Cumulative MACE @ 30 day F/U Cumulative MACE @ 30 day F/U 215/221 [97.2%] pts215/221 [97.2%] pts

ComplicationRateTotal MACE6/221 [2.7 %]

Death at day 10 [ non cardiac ] Sepsis ? P.E ?

0.4% (1/221)

M.I (Q or Non-Q wave)0.9% (2/221)

Procedural complications (VF)0.4% (1/221)Chest pain, Re-cath, TIMI-2 , Patent0.4% (1/221)

Re-Hospitalization2.3 % (5/221)

Re-Catheterization2.3 % (5/221)

TLR

TLR- PCI0.9% (2/221)Vessel and Aortic Root - Dissection 0.4% (1/221)

Page 55: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Procedural ComplicationsProcedural Complications(per lesion)(per lesion)

Procedural ComplicationsProcedural Complications(per lesion)(per lesion)

Perforation0/279

Distal embolism 5/279 [1.8%]

Moderate side branch occlusion

10/279 [3.6%]

Thrombus formation1/279 [0.36%]

Dissections (> type C), no flow limiting dissection.

6/271 [2.2%]

Page 56: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

Cumulative MACE @ 19.2±11.2 mon. 187/221 [84.6%] pts.

ComplicationComplicationRateRate

Total MACETotal MACE4.9 % (11/221)4.9 % (11/221)

Death at day 10. Death at day 10.

non cardiac.non cardiac.0.4% (1/221)0.4% (1/221)

MI (Q or Non-Q wave)MI (Q or Non-Q wave)0.9% (2/221)0.9% (2/221)

Re-HospitalizationRe-Hospitalization18.6% (41/221)18.6% (41/221)

Re-CatheterizationRe-Catheterization15.4% (36/221)15.4% (36/221)

CABG [>6 months]CABG [>6 months]0.9% (2/221) [5.6%]0.9% (2/221) [5.6%]

Re-TLR/TVR - PCIRe-TLR/TVR - PCI2.7% (5/221) 2.7% (5/221) [13.8%][13.8%]

Vessel and aortic Vessel and aortic dissectiondissection

0.4% (1/221)0.4% (1/221)

Page 57: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

The The ‘AngioSculpt’‘AngioSculpt’ is highly is highly effectiveeffective for for the treatment of a the treatment of a broad spectrumbroad spectrum of of challenging challenging coronary artery lesionscoronary artery lesions

It is very useful to It is very useful to avoidavoid ““geographic geographic miss”miss” due to due to device slippagedevice slippage

A surprisingly low rate of TVR, A surprisingly low rate of TVR, no CARDIAC no CARDIAC relatedrelated DEATHDEATH and and no Late Stent no Late Stent THROMBOSISTHROMBOSIS was observed during long was observed during long term F/U in this study term F/U in this study

The The ‘AngioSculpt’‘AngioSculpt’ is highly is highly effectiveeffective for for the treatment of a the treatment of a broad spectrumbroad spectrum of of challenging challenging coronary artery lesionscoronary artery lesions

It is very useful to It is very useful to avoidavoid ““geographic geographic miss”miss” due to due to device slippagedevice slippage

A surprisingly low rate of TVR, A surprisingly low rate of TVR, no CARDIAC no CARDIAC relatedrelated DEATHDEATH and and no Late Stent no Late Stent THROMBOSISTHROMBOSIS was observed during long was observed during long term F/U in this study term F/U in this study

Conclusions IConclusions I

Page 58: Pre-stent plaque modification in complex, calcified coronary artery lesions AngioSculpt® A New Scoring Balloon Ehud Grenadier M.D Cardiac catheterization

NC balloonNC balloon usage was usage was reducedreduced by by 46%46% inin the lesions in whichthe lesions in which the the AngioSculptAngioSculpt was was successfully deployedsuccessfully deployed

Difficulty in device deliverability may be Difficulty in device deliverability may be anticipated in anticipated in severe, calcified lesions severe, calcified lesions associated with pronounced associated with pronounced vessel/lesion angulation vessel/lesion angulation (or when the (or when the IVUS catheter was unable to cross the IVUS catheter was unable to cross the lesion) lesion)

Deliverability enhancements including Deliverability enhancements including an improved crossing profile and an improved crossing profile and lubricious coating are in developmentlubricious coating are in development

NC balloonNC balloon usage was usage was reducedreduced by by 46%46% inin the lesions in whichthe lesions in which the the AngioSculptAngioSculpt was was successfully deployedsuccessfully deployed

Difficulty in device deliverability may be Difficulty in device deliverability may be anticipated in anticipated in severe, calcified lesions severe, calcified lesions associated with pronounced associated with pronounced vessel/lesion angulation vessel/lesion angulation (or when the (or when the IVUS catheter was unable to cross the IVUS catheter was unable to cross the lesion) lesion)

Deliverability enhancements including Deliverability enhancements including an improved crossing profile and an improved crossing profile and lubricious coating are in developmentlubricious coating are in development

Conclusions II